ANDERSON-NEGELE MPP Installation And Startup Manual

Page 1
READ
Installation and Startup Guide
MPP Modular Pressure
THIS
Transmitter
FIRST
PRODUCT DESCRIPTION
The model MPP modular pressure transmitter has been specically designed for placement in sanitary process applications in the Life Sciences industry where
Anderson Instrument Co., Inc.
156 Auriesville Rd., Fultonville, NY 12072
Phone: 518-922-5315
Fax: 518-922-8997
www.anderson-negele.com
accurate and repeatable pressure measurements in difcult (vibration, wide temperature swings, humidity) environments are required. The modular design allows for eld conguration to best suit application needs and allows economical component replacement. State of the art performance yields class leading accuracy and stability while minimizing process and ambient temperature effects. A menu driven interface with an internal diagnostic display provides user adjustability of pressure range and a host of other control characteristics without tools or
pressure standards.
SENSOR WIRING
To facilitate electrical connections the MPP transmitter will be provided with either a 5 pin M12 quick disconnect receptacle, a M16 thread cable gland, or a ½” NPTF threaded adaptor. Shielded cable is recommended. See manual for
additional detail.
Field wireable connectors or molded cordsets are available as accessories from Anderson Instrument.
Document 1184
FIELD WIREABLE CONNECTOR ASSEMBLY - ORDERED AS ACCESSORY
1. Insert cable through Pressing Screw, Compression Ring, Seal Grommet, and Sleeve as shown below.
2. Strip back 1-1/4” of outer sheathing, cut off any excess wires, shield and ground. Strip off 1/4” insulation from remaining two wires. It is not necessary or recommended
Shown with
wire cover plate
removed
Swtiches and display only on
EX
MENU
_
UP
DOWN
HART version
LOOP
+
SENSOR
3. Orient Connector end so that center pin connecting screw is horizontal facing right (see detail).
Retaining Ring
Connector End
Pin 2 - Black
(Loop -)
DETAIL
Sleeve
Center screw pin
orientation reference
Pin 1 - Red
(Loop +)
MOLDED CORD SETS
Anderson Cord Set
Loop+(red) wire Pin 1
Loop- (black) wire Pin 2
Shield (clear or bare) wire
Note: Green and White not used
Typical Cord Set by others
4. Wire LOOP+ (red) wire to top-right terminal, and LOOP- (black) wire to top-left terminal. No connection is made to the center and bottom terminals.
5. Screw on the Sleeve. Hand-tighten only.
Compression
Ring
Seal
-2 included choose one to accommodate cable OD
Grommet
6. Press the Seal Grommet into the Sleeve and hand-tighten the Pressing Screw against the compression ring.
7. Use a wrench to tighten the Pressing Screw another 3/4 turn. Do not over-tighten!
Pressing Screw
Loop+(brown) wire Pin 1
Loop- (white) wire Pin 2
Shield (bare) wire
Page 2
USER INTERFACE GUIDE
The Anderson MPP Modular Pressure Transmitter is factory calibrated to the URL (upper range limit) and congured to the range and units specied by the order matrix number. Range, pressure units, output damping and analog scale (HART only) may be easily modied by the user. The Display Interface (DI) provides user feed back for menu functions, displays diagnostic error codes and nominally loop current or process variable.
The Modular Pressure Transmitter conguration parameters are sorted into three modes, and are accessible via the button along the bottom of the display interface. While the cover is removed, do not allow moisture to enter the housing.
RUN Mode
Zero
Adjust Display Decimal
Momentarily Display mA Output (HART only)
Descriptive Error Message
0%
Down
Execute
Display Interface
50%
PHARMACEUTICAL
SERIES
E
M
100%
Up
Menu
SENSOR CONFIG Mode
PSIG/BAR (native units)
4-20mA / 20-4mA (HART only)
LRV
URV
Damping
Alarm1 Alarm2
Display Units
Unit Description Scroll
Factory Restore
CALIBRATION Mode
2 Point Cal 4 Point Cal
SENSOR CONFIGURATION
Each instruction assumes starting from RUN mode which is default at power on and process value is displayed.
CALIBRATION – Zero
NOTE – For ABSOLUTE stems zero calibration is disabled as zeroing is not possible in atmospheric conditions.
Zeroing the sensor provides the best accuracy when clamped into the application therefore negates possible positioning and clamping errors. Be sure sensor is exposed to zero psig when performing this function.
1. Press both “D” and “M” simultaneously for approximately 5 seconds – Sensor displays 4.00 milliamp (HART version), or
0.00 (FFB version) for gauge, for compound sensor displays appropriate output for set range
Native Units: PSI or Bar
1. Press “M” – Sensor displays “COnF”
2. Press “E” – Sensor displays “PSI” or “BAR”
3. Press “U” or “D” – Sensor toggles between “PSI” or “BAR”
4. Press “M” to proceed to the next operation
Note: When changing units conguring range is now required
Page 3
Output: 4Ma – 20Ma or 20mA – 4mA (HART only)
1. Press “M” – Sensor displays “COnF”
2. Press “E” – Sensor displays “PSI” or “BAR”
3. Press “M” – Sensor displays “4-20” or “20-4”
4. Press “U” or “D” – Sensor toggles between “4-20” or “20-4”
5. If no other operation is to be accessed exit to “RUN” mode via pressing “M” repeatedly until process value is displayed
PRESSURE RANGE
LRV: Lower range value
Sensors equipped with ”Compound” style measurement cells (stem) may congure the LRV. “Absolute” stems are predened as 0 and are not recongurable
1. Press “M” – Sensor displays “COnF”
2. Press “E” – Sensor displays “PSI” or “BAR”
3. Press “M” – Sensor displays “4-20” or “20-4” (HART only)
4. Press “M” – Sensor displays “LRV”
5. Press “E” – Sensor displays present LRV value – Example: “0”
6. Press “U” or “D” to set desired LRV value– Sensor display increases or decreases accordingly
Note: continue to URV
URV: Upper range value
Note: starting from LRV above
7. Press “M” – Sensor displays “URV”
8. Press “E” – Sensor displays present URV value – Example: “50”
9. Press ”U” or “D” to set desired URV value – Sensor display increases or decreases accordingly
10. Press “E” to store displayed value – Sensor displays “URV”
11. If no other operation is to be accessed exit to “RUN” mode via pressing “M” repeatedly until process value is displayed
Output Damping
1. Press “M” – Sensor displays “COnF”
2. Press “E” – Sensor displays “PSI” or “BAR”
3. Press “M” – Sensor displays “4-20” or “20-4” (HART only)
4. Press “M” – Sensor displays “LRV”
5. Press “M” – Sensor displays “URV”
6. Press “M” – Sensor displays “dMPg”
7. Press “E” – Sensor displays current damping value from 0 - 10
8. Press “U” or “D” to display desired damping value– Sensor display increases or decreases accordingly
9. Press “E” to store displayed value – Sensor displays “dMPg”
10. If no other operation is to be accessed exit to “RUN” mode via pressing “M” repeatedly until process value is displayed
Setting Display Process Variable - Functionality in display interface only.
For additional display interface functionalities see section 7.2 of the manual.
The following Engineering units may be selected: Millibar, Torr & Kg/cm2, PSI, BAR, kPA, In H2O, In Hg, mm H2O, mA
1. Press “M” – Sensor displays “CONF”
2. Press “E” – Sensor displays “PSIG” or “BARG”
3. Press “M” – Sensor displays “4-20” or “20-4”(HART only)
4. Press “M” – Sensor displays “LRV”
5. Press “M” – Sensor displays “URV”
6. Press “M” – Sensor displays “DAMP”
7. Press “M” – Sensor displays “ALRM1”
8. Press “M” – Sensor displays “ALRM2”
9. Press “M” – Sensor displays currently set engineering unit
10. Press “U” or “D” repeatedly to select the desired engineering unit
11. Press “M” to proceed to next operation or repeatedly to exit to “run” mode
Factory configuration reset
Perform if a return to the original factory conguration is desired
1. Press “M” – Sensor displays “COnF”
2. Press “E” – Sensor displays “PSI” or “BAR”
3. Press “M” – Sensor displays “4-20” or “20-4” (HART only)
4. Press “M” – Sensor displays “LRV”
5. Press “M” – Sensor displays “URV”
6. Press “M” – Sensor displays “dMPg”
7. Press “M” – Sensor display “FrES”
8. Press “E” – Sensor display “nO”
9. Press “U” or “D” to display “YES”
10. Press “E” – Sensor display “FrES”
11. exit to “RUN” mode via pressing “M” repeatedly until process value is displayed
Page 4
EXPLODED VIEW
3.55
6.59
3.55
6.59
Cap
Display Interface
Puck
DIMENSIONAL DRAWINGS
3.6"
(91mm)
Housing
M12 Quick
Disconnect
Stem
ACCESSORIES
Cord Sets Shielded Molded w/25’ cable 42117K0025 Shielded Molded w/50’ cable 42117K0050 Shielded Molded w/100’ cable 42117K0100
Clear Cap w/gaskets 56328P0001 Stainless Steel Cap w/gaskets 56329P0001 Enclosure w/Clear cap w/gaskets SP5632700066 Enclosure w/SS cap w/gaskets SP56327A0066 M12 Quick Disconnect Receptacle SP56726A0002 Cord Grip SP5633100000 1/2” NPTF adaptor SP5633200000 Seal Kit (6) gaskets 5633000001 Field Wireable Connector-Straight 42119B0000 Field Wireable Connector-90° 42119A0000 Display Kit - HART option SP56741P0066 Display Kit - FFB option SP56741F0066 5’ Remote Kit SP73228A0005 10’ Remote Kit SP73228A0010 15’ Remote Kit SP73228A0015 20’ Remote Kit SP73228A0020 25’ Remote Kit SP73228A0025
3.6"
(91mm)
6.6" (167mm)
Horizontal Orientation
Cable available
in 5’, 10’, 15’, 20’
and 25’ lengths
6.6" (167mm)
Vertical Orientation
7.87” (200mm)