AND UA-651SL Instruction Manual

AND UA-651SL Instruction Manual

Digital Blood Pressure Monitor

Model UA-651SL

Instruction Manual Original

1WMPD4002763D

1. Dear Customers

The A&D blood pressure monitor is one of the most advanced monitors available and is designed for ease of use and accuracy. This device will facilitate your daily blood pressure regimen.

We recommend that you read through this manual carefully before using the device for the first time.

2. Preliminary Remarks

This device conforms to the European Directive 93/42 EEC for Medical Products. This is made evident by the mark of conformity. (0123: The reference number to the involved notified body)

The device is designed for use on adults, not newborns or infants. Environment for use. The device is for use to operate by yourself in the home healthcare environment.

This device is designed to measure blood pressure and pulse rate of people for diagnosis.

3. Precautions

Precision components are used in the construction of this device. Extremes in temperature, humidity, direct sunlight, shock or dust should be avoided.

Clean the device and cuff with a dry, soft cloth or a cloth dampened with water and a neutral detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean the device or cuff.

Avoid tightly folding the cuff or storing the hose tightly twisted for long periods, as such treatment may shorten the life of the components.

The device and cuff are not water resistant. Prevent rain, sweat and water from soiling the device and cuff.

Measurements may be distorted if the device is used close to televisions, microwave ovens, cellular telephones, X-ray or other devices with strong electrical fields.

When reusing the device, confirm that the device is clean.

Used equipment, parts and batteries are not treated as ordinary household waste, and must be disposed of according to the applicable local regulations.

When theAC adapter is used, make sure that theAC adapter can be readily removed from the electrical outlet when necessary.

Do not modify the device. It may cause accidents or damage to the device.

To measure blood pressure, the arm must be squeezed by the cuff hard enough to temporarily stop blood flow through the artery. This may cause pain, numbness or a temporary red mark to the arm. This condition will appearespecially whenmeasurement isrepeated successively. Anypain, numbness,orredmarkswilldisappearwithtime.

Take care to avoid accidental strangulation of babies or infants with the hose.

Do not twist the air hose during measurement. This may cause injury due to continuous cuff pressure.

Measuring blood pressure too frequently may cause harm due to blood flow interference. Check that the operation of the device does not result in prolonged impairment of blood circulation, when using the device repeatedly.

If you have had a mastectomy, please consult a doctor before using the device.

Donotletchildrenusethedevicebythemselvesanddonotusethedevice inaplacewithinthereachofinfants.Itmaycauseaccidentsordamage.

There are small parts that may cause a choking hazard if swallowed by mistake by infants.

Wireless communication devices, such as home networking devices, mobile phones, cordless phones and their base stations, walkie-talkies can affect this blood pressure monitor. Therefore, a minimum distance of 30 cm should be kept from such devices.

Unplug theAC adapter when not in use during the measurement.

Use of accessories not detailed inthis manualmay compromise safety.

Should the battery short-circuit, it may become hot and potentially cause burns.

Allow the device to adapt to the surrounding environment before use (about one hour).

Clinical testing has not been conducted on newborn infants and pregnant woman. Do not useonnewborn infantsorpregnant woman.

Do not touch the batteries, the DC jack, and the patient at the same time. That may result in electrical shock.

Do not inflate without wrapping the cuff around the upper arm.

Contraindications

The following are precautions for proper use of the device.

Do not apply the cuff on an arm with another medical electrical equipment attached. The equipment may not function properly.

People who have a severe circulatory deficit in the arm must consult a doctor before using the device, to avoid medical problems.

Donotself-diagnosethemeasurementresultsandstarttreatmentbyyourself. Alwaysconsultyourdoctorforevaluationoftheresultsandtreatment.

Do not apply the cuff on an arm with an unhealed wound.

Do not apply the cuff on an arm receiving an intravenous drip or blood transfusion. It may cause injury or accidents.

Do not use the device where flammable gases such as anesthetic gases are present. It may cause an explosion.

Do not use the device in highly concentrated oxygen environments, such as a high-pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion.

4. Parts Identification

Proper fit range

Index

Artery position mark

DC jack

Air Hose

 

AC adapter Plug

 

Display

 

START Button

Arm Cuff

Air Socket

 

Air Connector Plug

 

Battery Compartment

 

Battery Cover

Display

MEMORY

 

Systolic Pressure

 

Diastolic Pressure

 

Pulse Rate

 

I.H.B. symbol

 

(Irregular Heartbeat symbol)

 

Heart Mark

 

Battery Indicator

 

Pressure Bar Indicator and

 

WHO Classification Indicator

5. Symbols

Symbols that appear on the display

Symbols

Function / Meaning

RecommendedAction

 

 

 

Appears while measurement is

Measurement is in

 

 

 

in progress. It blinks when the

progress. Remain as still as

 

 

 

pulse is detected.

possible.

 

 

 

I.H.B. appears whenan irregular

 

 

 

 

 

 

heartbeat isdetected. Itmay light

 

 

 

 

 

 

whena very slight vibration like

 

 

 

 

 

 

shiveringor shaking isdetected.

 

 

 

 

 

 

Previous measurements stored

 

 

 

 

 

 

in memory.

 

 

 

 

 

 

FULLBATTERY

 

 

 

 

 

 

The battery power indicator

 

 

 

 

 

 

 

 

 

 

 

 

during measurement.

 

 

 

 

 

 

LOW BATTERY

Replace all batteries with

 

 

 

The battery power is low when it

new ones when the mark

 

 

 

blinks.

blinks.

 

 

 

Unstable blood pressure due to

Take another

 

 

 

movement during measurement.

measurement. Remain still

 

 

 

during measurement.

 

 

 

Thesystolicanddiastolicvalues

 

 

 

 

 

 

arewithin10mmHgofeachother.

 

 

 

 

 

 

The pressure value did not

 

 

 

 

 

 

increase during the inflation.

Apply the cuff correctly, and

 

 

 

The cuff is not applied correctly.

take another measurement.

 

 

 

 

 

 

 

 

 

 

PULDISPLAY ERROR

 

 

 

 

 

 

Thepulseisnotdetectedcorrectly.

 

 

 

Symbols

Function / Meaning

 

RecommendedAction

 

 

Remove the batteries and press

 

Device internal error

the

START

button, and then

 

install the batteries again. If the

 

 

 

error still appears, contact the

 

 

dealer.

Symbols printed on the device case.

Symbols

Function / Meaning

 

Standby and Turn the device on.

SYS

Systolic blood pressure in mmHg

DIA

Diastolic blood pressure in mmHg

PUL

Pulse per minute

 

Battery installation guide

 

Direct current

 

Type BF: Device, cuff and tubing are designed to provide

 

special protection against electrical shocks.

 

EC directive medical device label

 

EU-representative

 

Manufacturer

 

Date of manufacture

 

WEEE label

SN

Serial number

 

Refer to instruction manual/booklet

 

Polarity of DC jack

IP

International protection symbol

 

Keep dry

6.

Using the Monitor

 

 

6.1. Installing / Changing The Batteries

1.

Remove the battery cover.

 

Step 3

2.

Remove the used batteries from

 

 

 

the battery compartment in case of

 

 

 

changing them.

 

 

3.

Insert new batteries into the battery

 

Step 1

 

compartment as shown, taking

 

 

care that the polarities (+) and (-)

Step 2

Step 4

 

are correct.

 

 

4.

Replace the battery cover. Use only

 

 

 

R6P, LR6 orAAbatteries.

 

 

CAUTION

Insert the batteries as shown in the battery compartment. If installed incorrectly, the device will not work.

When blinks on the display and the device announces that the battery needs to be replaced, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. Replace the batteries two seconds or more after the device turns off.

does not appear when the batteries are drained.

The battery life varies with the ambient temperature and may be shorter at low temperatures. Generally, four new R6P batteries will last approximately for three months when used twiceformeasurement each day.

Use the specified batteries only. The batteries provided with the device are for testing the device performance and may have a limited life.

Remove the batteries if the device is not to be used for a long time. The batteries may leak and cause a malfunction.

6.2. Connecting The Air Hose

Insert the air connector plug

Air Hose

into the air socket firmly.

 

 

Air Connector Plug

6.3. Connecting The

Air Socket

AC Adapter

AC adapter Plug

Insert theAC adapter plug into the

DC jack. Then, insert theAC

DC Jack

adapter into an electrical outlet.

 

TheAC adapter, the model

 

TB-233, is sold separately.

 

When disconnecting theAC adapter from the

 

electrical outlet, grasp and pull theAC adapter body out of the outlet.

When disconnecting the AC adapter plug from the blood pressure monitor, grasp and pull theAC adapter plug out of the monitor.

6.4. Selecting The Correct Cuff

Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value.

The arm size is printed on each cuff.

The index and proper fit range, on the cuff, tell you if you are applying the correct cuff. Refer to "6.5 Applying The Arm Cuff".

If the index points outside of the range, contact your local dealer to purchase a replacement cuff.

The arm cuff is a consumable. If it becomes worn, purchase a new one.

Arm Size

Recommended Cuff Size

Catalog Number

31 cm to 45 cm

Large adult cuff

CUF-F-LA

23 cm to 37 cm

Semi large adult cuff

CUF-F-SL

Arm size:

The circumference at the biceps.

 

Note:

Model UA-651SL is not designed for using a small cuff.

Symbols printed on the cuff.

Symbols

Function / Meaning

RecommendedAction

 

 

 

 

 

 

 

Set the

mark on the

 

 

 

 

Artery Position Mark

arteryof the upper arm or

 

 

 

 

in line with the ringfinger

 

 

 

 

 

 

 

 

 

 

 

 

 

 

onthe insideof the arm.

 

 

 

 

Index

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REF

Catalog number

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Proper fit range for the semi large adult

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

cuff. It's printed on the adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Over range printed on the semi large

Use the large adult cuff

 

 

 

 

instead of the semi large

 

 

 

 

adult cuff.

adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Proper fit range for the large adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

It's printed on the large adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Under range printed on the semi large

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Under range printed on the large adult

Use the semi large adult

 

 

 

 

cuff instead of the large

 

 

 

 

cuff.

adult cuff.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lot number

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Semi large adult cuff

 

 

 

 

 

 

 

 

 

 

 

 

Proper fit range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6.5. Applying The Arm Cuff

1.Wrap the cuff around the upper arm, about 1 to 2 cm above the inside of the elbow, as shown.

Place the cuff directly against the skin, as clothing may cause a faint pulse and result in a measurement error.

2.Constriction of the upper arm, caused by rolling up a shirtsleeve, may prevent accurate readings.

3.Confirm that the index points within the proper fit range.

Note:During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed).

Do not roll up shirtsleeve tightly

 

Cuff

 

1 to 2 cm

 

Artery position mark

 

Secure the fabric fastener

 

Air hose

Index

and Proper fit range

 

 

 

 

 

 

 

 

 

 

7.

Measurements

 

 

 

 

 

 

7.1. Normal Measurement

At heart level

1. Place the cuff on the arm at heart

 

level (preferably the left arm). Sit

 

 

 

 

 

 

 

quietly during measurement.

Press the

 

button

 

 

 

 

START

 

 

 

2.

Press the

 

button.

 

 

 

 

 

 

 

 

 

 

START

 

 

 

 

 

 

 

All of the

display

segments are

All of the display segments

 

displayed. Zero is displayed

 

 

 

 

 

 

 

blinking briefly. Then the display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

changes, as indicated in the figure

 

 

 

 

 

 

 

at the right, as the measurement

 

 

 

 

 

 

 

begins. The cuff starts to inflate. It

 

Zero display

is normal for the cuff to feel very

 

tight.Apressure bar indicator is

 

displayed, as in the figure at the

 

right, during inflation.

Starts inflation

Note:If you wish to stop inflation at any time,

 

press the

START

button again.

 

11.About Blood Pressure

What is Blood Pressure?

Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).

12.WHO Classification Indicator

Each segment of the bar indicator corresponds to the WHO blood

pressure classification.

WHO Classification Indicator

: The indicator displays a

 

 

Severe hypertension

segment, based on the

 

 

Moderate hypertension

 

 

Mild hypertension

current data,

 

 

corresponding to the

 

 

High normal

WHO classification.

 

 

Normal

 

 

 

 

 

Optimal

 

 

 

13.Troubleshooting

3.When inflation is complete, deflation starts automatically and

the (heart mark) blinks, indicating that the measurement is in progress. Once the pulse is detected, the mark blinks with each pulse beat.

Note:If an appropriate pressure is not obtained, the device starts to inflate again automatically.

4.When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed. The cuff exhausts the remaining air and deflates completely.

5.Press the START button again to turn off the power.

Pressurizing

Measurement in progress

Systolic pressure Diastolic pressure WHO classification Pulse rate

Exhausts remaining air automatically

Turn off with the

START button

Note:Model UA-651SL is provided with an automatic power shut-off function. Allow at least 3 minutes between measurements on the same person.

7.2. Measurement with the Desired Systolic Pressure

If re-inflation occurs repeatedly, use the following methods. If your systolic pressure is expected to exceed 230 mmHg, use this procedure.

1.Place the cuff on the arm at heart level (preferably the left arm).

2. Press the START button.

3.During the zero blinks, press and hold the START button until a number about 30 to 40 mmHg higher than your expected systolic pressure appears.

4.When the desired number is reached, release the START button to start measurement. Then continue to

measure your blood pressure as described on the section "7.1 Normal Measurement".

At heart level

Press the START button

Press and hold the

START button

Inflation in progress

Release the START button to stop inflation

The desired pressure

Refer to the section "7.1 Normal Measurement"

7.3. Notes for Accurate Measurement

Sit comfortably on a chair. Rest your arm on the table. Do not cross your legs. Keep your feet on the floor and straighten your back.

Sit down in a comfortable position. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart.

Place the center of the cuff at the same level as your heart.

Relax for about five to ten minutes before measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.

Remain still and keep quiet during measurement.

Do not measure immediately after physical exercise or a bath. Rest for twenty or thirty minutes before taking the measurement.

An individual's blood pressure varies constantly, depending on what you are doing and what you have eaten. What you drink can have a very strong and rapid effect on your blood pressure.

This device bases its measurements on the heartbeat. If you have a very weak or irregularheartbeat,thedevicemayhavedifficultydeterminingyourbloodpressure.

Should the device detect a condition that is abnormal, it will stop the

measurement and display an error symbol. Refer to the section "5 Symbols" for the description of the symbols.

This device is intended for use by adults. Consult with your physician before usingthisdeviceonachild.Achildshouldnotusethisdeviceunattended.

Try to measure your blood pressure at the same time every day.

The automatic blood pressure monitor’s performance may be affected by excessive temperature or humidity, or altitude.

8. Recalling the Memory Data

The device automatically stores up to thirty blood pressure and pulse measurements in memory. Data stored in memory are assigned a data number in the order of the newest to the oldest. The oldest data displays as "". The symbol in the upper left corner of the display indicates that you are viewing previous data stored in memory.

8.1. Recalling Data

Press and hold the

No data

 

1.

When nothing is displayed, press

START button

 

 

and hold the

START

button to

Release the button

 

 

recall the stored data.

 

 

 

mark

 

2.

Release the button when

 

 

Average data

 

 

displaying the average data.

 

 

 

 

<![if ! IE]>

<![endif]>To

3.

The data number and stored data

Data number

<![if ! IE]>

<![endif]>next

 

are automatically displayed in

 

 

 

 

order from the last measurement.

 

 

<![if ! IE]>

<![endif]>data

4.

The display will turn off

 

Stored Data

 

 

 

automatically after all data is

 

 

 

 

 

 

 

displayed.

 

 

 

 

 

Note:If you press the START button while

 

 

 

recalling data, the device turns off.

Turns off automatically

 

 

 

 

 

8.2. Clearing Data

Press and hold the

 

1.

When turning off the device,

START

button

 

 

press and hold the

START

Release the

 

 

button until the "

" is

 

 

displayed.

 

 

 

button

 

2.

Select "

" to clear data.

Select "

" to

5 seconds

 

 

 

 

3.

Data is cleared when the

 

clear data.

 

 

 

mark blinks.

 

 

The

mark blinks.

 

4.

The device turns off

 

 

 

Data is cleared.

 

 

automatically.

 

 

 

 

 

 

 

 

 

Turns off automatically

9. What is an Irregular Heartbeat

The UA-651SL provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.

Note: We recommend contacting your physician if you see this symbol frequently.

10.Pressure Bar Indicator

The indicator monitors the progress of pressure during measurement.

 

Pressurizing

 

 

 

Releasing air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement in progress

Inflation complete

Inflation in progress

Problem

Possible Reason

RecommendedAction

Nothing

Batteries are

Replace all batteries with new

appears on

drained.

ones.

the display,

Battery terminals

Reinstall the batteries with negative

even when

and positive terminals matching

the power is

are not in the

those indicated on the battery

turned on.

correct position.

compartment.

 

Battery voltage is

 

 

too low.

 

 

 

 

The cuff

blinks. If

the

 

 

Replace all batteries with new

does not

batteries are

ones.

inflate.

drained completely,

 

the mark does not

 

 

appear.

 

 

The cuff is not

Apply the cuff correctly.

 

applied properly.

 

 

 

You moved your

Make sure you remain still and

 

arm or body during

The device

measurement.

quiet during measurement.

does not

 

 

 

 

 

 

 

 

 

 

 

Sit comfortably and still. Place your

measure.

 

 

 

 

 

The cuff position is

arm on a table with your palm

Readings

not correct.

facing upward and the cuff at the

are too high

or too low.

 

 

 

 

 

same level as your heart.

 

 

 

 

 

If you have a very weak or irregular

 

 

 

 

 

 

 

 

 

 

 

 

heartbeat, the device may have

 

 

 

 

 

 

difficulty in determining your blood

 

 

 

 

 

 

pressure.

 

 

 

 

 

 

At a clinic or doctor's office, an

 

The value is

apprehension may cause an

 

elevated reading. Home

 

different from that

 

measurement reduces the effects

 

measured at a clinic

of outside influences on blood

Other

or doctor’s office.

pressure readings, supplements

 

 

 

 

 

 

the doctor's readings.

 

 

 

 

 

 

Remove the batteries. Place them

 

 

 

 

 

 

back properly and take another

 

 

 

 

 

 

measurement.

Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.

14.Maintenance

Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers.

The device was designed and manufactured for a long service life. However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized dealer in your area orA&D for maintenance.

15.Technical Data

Type

UA-651SL

 

 

Measurementmethod

Oscillometric measurement

Measurement range

Pressure:

 

0 - 299 mmHg

 

Systolic pressure:

60 - 279 mmHg

 

Diastolic pressure: 40 - 200 mmHg

 

Pulse:

 

40 - 180 beats / minute

Measurementaccuracy

Pressure:

 

±3 mmHg

 

Pulse:

 

±5 %

Power supply

4 x 1.5V batteries (R6P, LR6 orAA) or

 

AC adapter (TB-233C) (Not included)

Number of measurements

 

 

 

Approx. 700 times

LR6 (alkalinebatteries)

 

Approx. 200 times

R6P (manganesebatteries)

 

With pressure value 180 mmHg, room

 

temperature 23 °C.

 

Classification

Internally powered ME equipment (Supplied by

 

batteries) / Class II (Supplied by adapter)

 

Continuous operation mode

Clinical test

According to ISO81060-2 2013

EMC

IEC 60601-1-2: 2014

 

Memory

Last 30 measurements

Operatingconditions

+10 to +40 °C / 15 to 85 %RH / 800 to 1060 hPa

Transport/Storageconditions -20 to +60 °C / 10 to 95 %RH/ 700to1060 hPa

Dimensions

Approx. 96 [W] x 68 [H] x 130 [D] mm

Weight

Approx. 240 g, excluding the batteries

Ingress protection

Device:

IP20

 

Applied part

Cuff

Type BF

 

Useful life

Device:

5 years (whenusedsixtimesaday)

 

Cuff:

2 years (whenusedsixtimesaday)

 

ACadapter:5years(whenusedsixtimesaday)

AccessoryAC adapter

The adapter is to connect the device to a power source at home. Please contact your localA&D dealer for purchasing. TheAC adapter is required to be inspected or replaced periodically. TB-233C Please contact your localA&D dealer for purchasing.

TheAC adapter is required to be inspected or replaced periodically.

Symbols that are printed on theAC adapter.

 

Symbols

Function / Meaning

 

 

 

 

 

 

 

For indoor use only

 

 

 

 

 

 

 

 

Class II device

 

 

 

 

 

 

 

 

Thermal fuse

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fuse

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EC directive device label

 

 

 

 

 

 

 

EAC certification device label

 

 

 

 

 

 

 

Polarity ofAC adapter plug

 

 

 

 

 

 

 

Accessories sold separately

 

 

Cuff

Catalog Number

Cuff Size

Arm Size

 

CUF-F-LA

Large adult cuff

31cmto45cm

 

CUF-F-SL

Semi large adult cuff

23cmto37cm

 

 

 

 

 

 

Arm

size: The circumference at the biceps.

AC adapter

Catalog Number

Plug (Outlet type)

 

 

 

 

 

 

 

TB-233C

Type C

Note: Specifications are subject to change without prior notice. IPclassification is the degrees of protection provided by

enclosures in accordance with IEC 60529.This device is protected against solidforeign objects of 12mm diameter and greater such as afinger.This device is not protectedagainst water.

A&D Company, limited

1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119

Emergo Europe B.V.

Prinsessegracht 20, 2514 AP The Hague, The Netherlands Telephone: [31] (70) 345-8570 Fax: [31] (70) 346-7299

A&D INSTRUMENTS LTD.

Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom

Telephone: [44] (1235) 550420

Fax: [44]

(1235) 550485

A&D ENGINEERING, INC.

 

 

1756 Automation Parkway, San Jose, California 95131 U.S.A.

Telephone: [1] (408) 263-5333

Fax: [1] (408)263-0119

A&D Australasia Pty Ltd.

 

 

32 Dew Street, Thebarton, South Australia

5031 AUSTRALIA

Telephone: [61] (8) 8301-8100

Fax: [61]

(8) 8352-7409

OOO A&D RUS OOO "ЭЙ энд ДИ

РУС"

121357,Российская Федерация, г.Москва, ул.Верейская, дом 17 ( Business-Center"Vereyskaya Plaza-2" 121357, Russian Federation, Moscow, Vereyskaya Street 17 )

тел.: [7] (495) 937-33-44 факс: [7] (495) 937-55-66

138 21 A 200120

: [86] (21) 3393-2340 : [86] (21) 3393-2347

: 91-124-4715555 :91-124-4715599