AD8526
The AD-8526 Ethernet Converter can connect the RS-232C interface of a weighing instrument to the Ethernet (LAN) port of a computer that is not equipped with an RS-232C interface. By using the AD-8526, weight data from a weighing instrument can be managed with a computer connected to the network. The weighing instrument can also be controlled by the connected computer.
Unpacking the AD-8526
MAC address
Ethernet converter
Note: Only the following cable is included if specified when ordering:
(Length: Approx. 2m. 2 pcs)
DIN 7-pin to D-Sub 9-pin cable (AX-KO1786-200)
If not specified, the following two communication cables are included: DIN 7-pin to D-Sub 9-pin cable (AX-KO1786-200), D-Sub 9-pin to D-Sub 9-pin cable (AX-KO2466-200)
Specification
1. Specification 3. RS-232C serial interface
Operating environment : -10°C to +40°C Connector: D-Sub 9-pin male
AC adapter:
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption: Approx. 11VA (supplied to the AC adapter Dimensions: 113(W) x 60(D) x 38(H) mm Parity: Even*, Odd, None Net weight: Approximately 250g Stop bits: 1 bit* or 2 bits
2. Ethernet interface
Connector: RJ45 * : Factory settings Protocol: TCP/IP 
Weighing Instrument Cables
Weighing instrument Interface opt ion GX, GF, GX-K, GF-K, GP, GR, HR, MC EK-i, EW- i, FC- i, FC-S i, GH, HR- i AD-4212C HV-G, HV-WP, HW-G, HW-WP FG FG-L, FG-M
1WMP
D4001294C
Positions to affix
AC adapter labels
(Example)
ffix them to the AC adapter and
the connector end of the cable.
Note Please confirm that the AC adapter type is correct for your
Write a preset IP address and affix it on the weighing instrument.
local voltage and receptacle type.
(5 labels)
ffix it on the IP address label.
Transmission form: Asynchronous, bi-directional
Baud rate:
)
Data format
Data bits: 7* or 8 bits
Required AD-8526 Cable by Weighing Instrument
Communications cable (
None (D-Sub 25-pin, standard accessory)
None (D-Sub 9-pin, standard accessory) None (D-Sub 9-pin, standard accessory)
AD4212C standard accessory
None (DIN 7-pin, standard accessory)
OP-03 (DIN 7-pin) OP-23 (DIN 8-pin)
Please refer to our website for the latest information.
(5 labels)
600, 1200, 2400*, 4800, 9600, 19200bps
Length 2 m
AX-KO1710-200
AX-KO2466-200
AX-KO1786-200 AX-KO1786-200 AX-KO1786-200
Connecting the Weighing Instrument to a Computer
Caution
Contact your network manager before connecting the AD-8526 to a network. The AD-8526 may generate a network error. A&D assumes no responsibility for any errors that may be caused. Affix the AC adapter labels to the AC adapter and the connector end of the cable. Set the IP address and subnet mask to the AD-8526 one at a time. The factory default IP address is 172.16.100.2. Do not duplicate an IP address. The IP address cannot be returned to the factory setting. We recommend that you write the IP address on the supplied label.
AD-8526
Preparation
Please download each software from A&D website
https://www.aandd.jp/products/software/software.htmlfor below.
(1) Setting procedure “WinCT-Plus” instruction manual (2) IP address setting software “Device Installer” (3) Data acquisition software
Step 1 Connect the AC adapter to the AD-8526.
Step 2 Connect the AD-8526 and the computer with a cross
“WinCT-Plus”
cable directly or connect them with a hub and straight cables.
Step 3 Input the IP addresses and subnet masks of the AD-8526
and computer. For the setting procedure, refer to the “WinCT-Plus” instruction manual.
Communication error may occur if set incorrectly.
Step 4 Write the IP address on the IP address label and affix it on
the AD-8526 (and the weighing instrument
if necessary)
where you can see the address easily.
Step 5 Install the RsMulti data acquisition software to download the
“WinCT-Plus” instruction manual.PDF form A&D website.
http
s://www.aandd.jp/products/software/software.html
Step 6 Connect the accessory RS-232C cable between the RS-232C
interface of the weighing instrument and the AD-8526’s port.
Step 2
Straight cable
AD-8526
Step 6
Data Communication Software
The RSMulti data acquisition software can be launched on a Windows PC. Menu: [Start] [Program] [A&D WinCT operation. This software can acquire data from multiple devices connected by LAN or RS-232C. The software can control these devices with commands. The software can acquire data transmitted from devices. balance, data is output and is acquired with the computer. Stored data can be used with Excel. (Install Microsoft Excel before use.)
Example A sample window of RsMulti
Two balances are connected.
)
Computer 172.16.100.1 GX-600 172.16.100.2 GP-30K 172.16.100.3
Hub
Computer
172.16.100.1 172.16.100.2 172.16.100.3
[RsMulti]. Refer to the Manual PDF file in the A&D WinCT-Plus for details on the
-Plus]
Example: When the
PRINT
IP address Subnet mask
GX-600
255.255.0.0
GX-600 GP-30K
GX-600 GP-30K
+0182.252 +0182.253 +0182.955 +0182.964 +0182.965 +0186.676 +0182.251 +0182.251 +0182.965 +0182.265 +0182.252 +0182.239
AC adapter
Step1
Cross cable
Step 2
Computer
Weighing
instrument
Step 6
RS-232C cable
key is pressed on the
GP-30K
+019218.1 +019218.2 +019218.2 +019289.8 +019218.2 +019218.4 +019218.0 +019218.0