Analytical Industries, Inc.
11
Other Sample Handling Considerations
The sample system may include optional components positioned upstream of the analyzer such as additional
pressure regulator, particulate filter, isolation/directional flow control valves etc. Check the QC sheet to confirm
the Sample System of your analyzer.
To avoid damaging the sensor or sample system, observe the following guidelines:
1. Do not draw a vacuum or over-pressure the sensor. When sample gas flows into the analyzer, any
restriction on the vent line will cause a pressure build up within the analyzer sample system, a sudden
release of the sample pressure will draw a vacuum on the sensor.
2. The vacuum generated in 1 above may cause the sensor to leak electrolyte (voiding the sensor
warranty). The electrolyte leakage may damage the electrical contacts of the upper section of the sensor
housing assembly, the electrolyte might enter in the sample system and if it does, it will require a
complete overhaul of the sample system.
3. Do not supply sample gas with pressure above the recommended limit. High positive pressure can
damage the integral pressure regulator, the sensor and or other sample system components.
4. Do not introduce calibration span gas from a pressurized cylinder without first regulating the pressure to
within the recommended limits (30-50 PSIG).
CAUTION: Do not vent the sample to a line with pressure above atmospheric pressure. If the vent line pressure
is above atmospheric (below 10-20 inches of water and is varying, use a back-pressure regulator set at 20 inches
of water to stabilize the pressure at the sensor.
Installation
The PI2-UHP 50/100 series Oxygen Analyzers consist of an electronic
module, sensor housing and sample system housed in a 13.9”W x
9.9”H x 13.4”D enclosure suitable for bench-top or cart applications.
The enclosure can be panel or rack mounted with an optional bezel.
The analyzer includes a temperature controlled heating system to
maintain the temperature of the sensor at a pre-set temperature of
85 ◦F. The analyzer has been tested and calibrated by the
manufacturer prior to shipment.
Note: Analyzer is shipped from the factory with sensor placed in a
separate bag sealed under nitrogen. Install sensor as instructed later
in this section.
Caution: Do not change the factory settings of the analyzer until instructed to do so in this manual.
Sample/Span Gas Connections Considerations
Assemble the necessary hardware for mounting the analyzer and optional components (if provided or required by
the user). Perform the following steps
1. Use ¼” stainless steel compression type fittings for air input and sample vent line connections. Use ¼”
Face Seal type fittings for sample and span gas inlet connections. Caution: use clean/fresh gaskets for
making Face Seal type connections.
2. Use cleaned, preferably electro polished SS tubing for all sample/span gas connections. Plastic tubing
rated at 150 psig may be used to provide instrument air.
3. Review the application/sample analysis conditions to ensure the sample gas is suitable for analysis.
Sample and Span gas pressure 30-50 psig
Recommended sample and span gas pressure 30-50 PSIG (maximum pressure of 150 psig is acceptable). The 3050 psig pressure is recommended for ease of controlling of the gas flow rate by using the integral pressure
regulator.