Analytical Industries Inc (AII) PI2-UHP 50, PI2-UHP 100, PI2-MS 500, PI2-MS 1000 Specifications

Page 1
Oxygen Analyzers for Ultra High Purity Gases
PI2-UHP 50/100 and PI2-MS 500/1000
Based on the stable and long-life Pico-Ion oxygen sensors, the PI2 range of trace oxygen analyzers provides accurate and stable measurements of oxygen down to low parts per trillion. Low cost of ownership through minimal maintenance of this high-performance oxygen analyzer make measurement of oxygen down to ppt level easy and affordable for customers with limited budgets.
PPM
Highlights
Measuring trace O2 in ultra-pure Ar, H2, He and N
LDL to less than 100 ppt
Low cost of ownership
User friendly and easy to operate and maintain
Fast recovery from process upsets
Integrated bypass sample system as standard
Optional auto calibration system with true zero
feature via O2 scrubber
Applications
2
Quality checking of ultra-high purity gases
produced by cryogenic air separation
Validating high purity gases used in electronics
and semi-conductor manufacture
Commissioning UHP gas lines and processes
www.aii1.com
Page 2
30
4
56070
3030404505060
70
8
0
Second generation, no­maintenance Pico-Ion sensor technology
A high-performance, low ppb oxygen sensor with stable output requiring no maintenance or frequent electrolyte additions. The innovative design of the gas chamber maximizes the rate of oxygen reaction while the refined proprietary sensor design provides a high signal output per unit area to enhance sensitivity as well as a significantly higher output. Further advantages are:
Membrane
Electrolyte
0202020
0
0
2
2
-
0H
0H
i
-
0H
-
0H
-
0H
-
-
0H
-
0H
-
0H
-
0H
-
0H
Pb0 Pb0
-
0H
0H
0H 0H
0H 0H
02020
2
-
-
0H
-
0H
-
0H
-
-
0H
-
0H
-
0H
-
0H
0
2
0
2
2
-
0H
-
0H
-
0H
-
-
0H
-
0H
0H
-
0H
-
-
0H
-
-
0H
-
-
0H
-
0H
Layer
Cathode
Electrolyte
Anode
Lower detectable limit (LDL) of less than 100ppt
High signal to noise ratio
Excellent stability
Response time of less than 15 seconds
Minimizes the temperature dependence
Fast recovery from higher oxygen lev els during process
upset conditions
Approximately 12 months of continuous use
Sensor Construction
PICO ION Sensor
100
90
80
70
60
50
40
30
Oxygen Reading / Output
20.9%
20
15.2%
10
15.2%
0
0 10 20 30 40 50 60 70 80 90 100
60.6%
40.0%
40.3%
21.0%
% Oxygen Concentration
100.2%
60.3%
100.0%
Typical sensor output
www.aii1.com
Page 3
Recovery time
O
PI2-MS 1000 PI2-MS 500
PI2-UHP 100 PI2-UHP 50
Level Duration O2 Target Recovery on N2
2
9 ppm 2 minutes 10 ppb 10 minutes
Air 30 seconds 1 ppm 45 minutes 9 ppm 1 minute 10 ppb 15 minutes 9 ppm 1 minute 1 ppb 60 minutes
Simple, intuitive HMI
The analyzer is operated by proprietary easy-to-use menu-driven software and a large graphical LCD with four control keys. The analyzer can be operated remotely via USB or RS232 and allows the operator to obtain the data, change settings, calibrate and diagnose the instrument.
Standard Full Featured Automated Sample System
The PI2-UHP analyzers have a full featured ultra-clean sampling system designed for the measuring < 100 PPT oxygen:
• 316 stainless wetted parts including electro-polished tubing, Face Seal fittings or orbital welded connections
• Integral pressure regulator and flow control
• Pneumatic diaphragm valves that operate:
- Sample and span gas inlets
- Integral bypass system that isolates the sensor from high oxygen levels and
- An oxygen scrubber system for generating consistent zero gas
• Temperature control of the sample system limits drift during day-and-night cycles where ambient temperatures fluctuations were typically ±10F°.
www.aii1.com
www.aii1.com
Page 4
Technical Specifications
PI2-MS 1000 PI2-MS 500 PI2-UHP 100 PI2-UHP 50
Measurement range
Accuracy at constant conditions
Response time T90 <15 seconds Recovery time See separate table on inside pages Sensitivity (LDL) < 5 ppb < 2.5 ppb < 250 ppt < 100 ppt Linearity < 1% of scale Sensor GPR-12-2000 MS-2 GPR-12-2000 MS-2E GPR-13-2000 UHP-2 GPR-13-2000 UHP-2E Sensor life at 25ºC
(77°F) and 1 atm Calibration interval 30 days Inlet pressure 1.4-3.4 barg (20-50 psig) with atmospheric vent, max 10.3 barg (150 psig) Flow rate 0.5 - 1.0 Nl/m (1-2 SCFH)
Gas connections 1/4” compression tube fittings
Wetted parts Stainless Steel
0-1, 0-10, 0-100,
0-1000 ppm
< 3% of reading or ±5 ppb (whichever is greater) ±3% of reading or ±0.5 ppb (whichever is greater)
0-0.5, 0-1, 0-10, 0-100,
0-1000 ppm
12 months
(Up to 24 months for the MS versions)
0-100 ppb, 0-1, 0-10,
0-50, 0-100 ppb, 0-1,
0-100 ppm
Inlets: 1/4” face seal fittings;
Vent and pneumatic valves: 1/4”
compression tube fittings
0-10 ppm
Display Graphical LCD 12 x 7cm (5 x 2.75”); resolution 0.1
Bench top, painted sheet metal (35 x 25 x 34cm (13.9” x 9.9” x 13.4”))
Enclosure
Compensation Barometric pressure and temperature; heated sample system and sensor housing Signal output 4-20 mA isolated, 0-1V and 0-5V Range ID 1-5V DC or 4-20mA, optional relay contacts Communications Choice of USB, RS232 Alarms 2-off user-adjustable alarms form C relay contacts non-latching Operating
temperature Power 100-240 V AC
Options: panel or 19” rack mount, wall mount
(30.5 x 30.5 x 20.3cm (12” x 12” x 8”))
0° to 45°C (32°F to 113°F)
Bench top
Options: panel or 19” rack mounting
Analytical Industries Inc. 2855 Metropolitan Place, Pomona, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: info@aii1.com
Please note: Analytical Industries Inc. adopts a continuous development program which sometimes necessitates specification changes without notice.
www.aii1.com
Please contact us for the latest version. Issue No: Analyzers for Ultra High Purity Gases_partnumber_V2_UK_1018