Analytical Industries Inc (AII) AII-3000 AHC, AII-3000 MHC, AII-3000 M, AII-3000 A Operating Manual

Technical Specifications
Accuracy: < +2% of FS range under constant conditions Analysis: 0-100% oxygen Application: Personnel safety
Area monitoring O2 deficiency in confined spaces Checking breathing air tanks Checking tanks intended for scuba diving Confirming the O2 levels prior to welding
Approvals: ISO 9001:2000, MDD 93/42/Annex II, ISO 13485:2003 Area Classification: General purpose
Alarms: A models: None; M models: User adjustable HI 1-100%
and LO 0-99% alarms; 120 second alarm silence for cali­bration purposes; HI alarm defeat for 100% O2 analysis
Calibration: Air or certified 100% O2 every 8 hours Compensation: Temperature compensated Connections: A/M models: 1x16 mm thread; HC models: Tubing 1/4” Controls: Soft touch keypad for ON/OFF and menu function Dimensions: 3.6 x 5.9 x 1.6”; weight 10 oz. (280 grams)
Display: 3-1/2 digit backlit LCD 2.75 x 1.375; resolution 0.1% O2 Flow Sensitivity: None between 0.2 to 10 liters per minute Humidity: Non-condensing 0-95% RH LED Indicators: A models: None; M models: upon activation of alarms
Linearity: + 1% under constant conditions Pressure: Inlet - A/M models ambient, HC models - regulate; Vent
all models - atmospheric
Power: 2 AA Alkaline batteries; 1,200 hours continuous use Response Time: 90% of final FS reading in 9 seconds Sensitivity: < 0.5% of FS range Sensor: A/M models: AII-11-60; HC models: AII-11-60-HC Sensor Life: 60 months in air at 25ºC and 1 atmosphere Signal Output: None Storage Temp.: -20º to 60ºC (-4ºF to 140ºF) on intermittent basis Temp. Range: 5º to 45ºC (41ºF to 113ºF) Warm-up Time: None Warranty: 12 months analyzer; 12 months sensor
AII-3000 A
AII-3000 AHC
AII-3000 M
AII-3000 MHC
Accessories - see back page
ISO 9001:2008 Certified
INTERTEK Certificate No. 485
2855 Metropolitan Place, Pomona, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: info@aii1.com Rev 6/15
Spare Parts
AII-3000 A, AII-3000 M AII-3000 AHC, AII-3000 MHC
AII-11-60 Oxygen Sensor AII-11-60-HC Oxygen Sensor BATT-1008 Battery (2x) 1.5V AA BATT-1008 Battery (2x) 1.5V AA P-1087 Instructions for Use P-1087 Instructions for Use A--1162 PCB Assy Main A-1162 PCB Assembly Main CABL-1006 Coil Cable TUBE-1019 7/32” OD Tubing 3’
CC-1072 Carrying Case
O2
2855 Metropolitan Place, Pomona, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: info@aii1.com Rev 6/15
Copyright © 10/10 All Rights Reserved
Analytical Industries Inc., 2855 Metropolitan, Pomona, CA 91767 USA. Tel: 909-392-6900, Fax: 909-392-3665 e-mail: sales-medical@aii1.com, web: www.aii1.com
This manual may not be reproduced in whole or in part without the
prior written consent of Analytical Industries Inc.
ISO 9001:2008, Certificate #485A
AII-3000 AHC
AII-3000 MHC
AII-3000 M
AII-3000 A
Table of Contents
1 Introduction 1
1.1 Indications for Use
1.2 Intended Use
1.3 Device Description
1 2 2
2 Quality Control Certification 3
3 Safety Warnings 4
4 Start-up
4.1 Contents of Shipping Container
4.2 Install Batteries
4.3 Install Oxygen Sensor
4.4 Controls
4.5 Start-up Test
4.6 Alarms (AII-3000 M Oxygen Monitor)
4.7 Calibration
4.8 Mounting
5 Operation
5.1 Principle of Operation
5.2 Application Considerations
5.3 Calibration
5.4 Sampling
6 Maintenance
6.1 Serviceability
6.2 Battery Replacement
6.3 Oxygen Sensor Replacement
7 Troubleshooting
8 Specifications
8.1 Spare Parts & Accessories
9 Warranty
10 Material Safety Data Sheet (MSDS)
10.1 Disposal
i
1 Introduction
Congratulations on your purchase, these Instructions for Use describe the pre­cautions, set-up, operation, maintenance and specifications of the AII-3000 Series Oxygen Analyzers.
This symbol means CAUTION – Failure to read and comply with the Instructions for Use could damage the device and possibly jeopardize the well being of the user.
Note: Advanced Instruments Inc. cannot warrant any damage resulting from the misuse, unauthorized repair or improper maintenance of the device.
1.1 Indications for Use
The AII-3000 Series Oxygen Analyzers are intended to measure and display the concentration of oxygen in compressed breathing air tanks intended for scuba diving, for personnel safety, area monitoring, O2 deficiency in confined spaces, checking breathing air tanks and confirming the O2 levels prior to welding.
Users must read the following statements as they are essential to re­ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used.
The devices have been designed and manufactured in such a way that when used under the conditions and for the purposes intended, they will not compro­mise the clinical condition or the safety of patients, or safety of the users or other persons.
Conformity with essential requirements has been demonstrated by verifying the performance of the device under normal conditions, bench testing and deter-
mining that undesirable malfunctions constitute minimal risk to users.
Do not sterilize, autoclave, liquid sterilize, immerse in any liquid or expose the device or accessories to steam, ethylene oxide or radiation sterilization.
The device is intended to be re-usable. Should the device or accessories come in contact with patient bodily fluids, either dispose of the device or clean with a soft cloth dampened with 70% isopropyl alcohol solution in water and allow the components to air-dry before re-use .
Do not operate the analyzer near equipment capable of emitting high levels of electromagnetic radiation as the reading may become unstable.
1
In order to obtain optimum performance, the operation of the device must be performed in accordance with these Instructions for Use. Maintenance should be performed only by trained personnel authorized by the manufacturer.
1.2 Intended Use
The AII-3000 Series Oxygen Analyzers are intended to measure and display the concentration of oxygen in compressed breathing air tanks intended for
scuba diving, for personnel safety, area monitoring, O2 deficiency in confined
spaces, checking breathing air tanks and confirming the O2 levels prior to welding.
1.3 Device Description
The AII-3000 Series Oxygen Analyzers can be positioned on a table top or pole (tripod wire stand and V-mount dovetail attachments are mounted on the back of the device) and are readily portable from one location to another. They provide continuous, fast, reliable and accurate oxygen measurements of up to respiratory care systems.
The devices utilize an electrochemical galvanic fuel cell type oxygen sensor of
the type that is extensively used to measure oxygen concentrations from 0%
to 100% in gas streams. Oxygen, the fuel for this electrochemical transducer, diffusing into the sensor through a gas permeable membrane reacts chemical­ly at the sensing electrode to produce an electrical current output proportional to the oxygen concentration in the gas phase. The sensor has an absolute zero meaning that when no oxygen is present to be chemically reacted the LCD displays 00.0 oxygen.
The sensor’s signal output is linear over the entire range, remains virtually
constant over the specified useful life and drops off sharply at the end. The sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery. Inasmuch as the sensor is a transducer in its own
right, its expected life is not affected by whether the analyzer is ON or OFF.
A battery powered state-of-the-art micro-processor converts the sensor’s sig- nal output representing the partial pressure of oxygen in the gas stream being analyzed. The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display (LCD) that has a resolution of 0.1% oxygen. The microprocessor is controlled from a keypad and provides features like system diagnostics, warning indicators, controls and an alarm capability for continu­ous monitoring that enhance both safety and effectiveness.
Prior to shipment, every device is thoroughly tested at the factory and docu­mented in the form of a Quality Control Certification that is included in the
Instructions for Use supplied with every device.
2
2 Quality Control Certification
3
Customer: ________________________ Order No. _____________ Date: _______ Model: ( ) AII-3000 A Oxygen Analyzer
( ) AII-3000 M Oxygen Analyzer ( ) AII-3000 AHC Oxygen Analyzer ( ) AII-3000 MHC Oxygen Analyzer S/N _______________
Sensor: ( ) AII-11-60 or ( ) AII-11-60-HC S/N _______________
Electronics: A-1152 PCB Assembly Main Software Version _______________ Accessories: AII-3000 A / M: CABL-1006 Cable, Coiled Phone Jack
AII-3000 AHC / MHC: TUBE-1019 7/32” OD Tubing 3 ft. All units: BATT-1008 Battery, 1.5V AA (Qty 2)
P-0187 Manual, Instructions for Use ……… Included ______
PASS QC Test: LCD display 3-1/2 digits ……………………………………………………. ______ Battery symbol displays when battery is low ……………………….. ______ Span adjustment +10-30% FS with 100% oxygen calibration ______ Following calibration with 99-100% oxygen and flushing with
ambient air, oxygen reading as displayed by LCD 20.9% +2% ____ Span adjustment +10-30% FS with air calibration ……………….. ______ Following calibration with air (20.9% oxygen) and exposing
to 99-100% oxygen, LCD displays 100% +2% ……………………. ______
Overall inspection for physical defects ………………………………... ______ Qty Options:
_____________________________________________ ____
See Sec 8.1 ____________________________________________________ _____ ____________________________________________________ _____
____________________________________________________
____
____________________________________________________ _____ Delivery: ____________________________________________________
3 Safety Warnings
ALWAYS follow the statements below as they are essential to reduc­ing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used.
Only trained personnel who have read, understand and agree to follow
the Instructions for Use should operate the device.
Retain the Instructions for Use for future reference. Refer service needs to trained authorized personnel. Failure to do so may
cause the device to fail and void the warranty.
Inspect the device and accessories before operating and ensure: (a)
there is no evidence of physical damage; (b) the sensor (particularly the sensing surface) and electrical connections are dry; and, (c) the sensor is installed and upstream from any humidifying device for accurate calibra­tion and oxygen readings.
Calibrate: (a) with a known source of air or dry 100% oxygen before
using each day or after 8 hours of continuous use; (b) when the temper-
ature or pressure of the operating environment changes; (c) if the oxy­gen sensor has been disconnected and reconnected; (d) after the battery or oxygen sensor has been replace.
Sampling flowing gas: (a) install the optional flow diverter and tee-
adapter in a vertical position as shown in Section 4.3 and (b) assure there is a tight fit between the flow diverter and tee adapter.
Sampling static, ambient or controlled atmospheres remove the flow
diverter.
Clean the device and accessories in accordance with Section 6.1.2.
Battery replacement Section 6.2: (a) replace the batteries within twenty-
four (24) hours of the battery symbol appearing on LCD display and (b) calibrate the analyzer after replacing the batteries.
Oxygen sensor installation or replacement Section 6.3: allow the new
sensor to stabilize for 15-20 minutes in ambient air before attempting to calibrate.
Store the device by turning the power OFF and removing the batteries if
the device will not be operated for over thirty (30) days.
Attempt to repeat the procedure that caused a perceived malfunction
and refer to troubleshooting hints in Section 7 before concluding the device is faulty. If in doubt, contact the manufacturer for assistance.
4
NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients, users or other persons.
If the reading is unstable or a malfunction is suspected. After the battery symbol appears in the LCD display. Near equipment capable of emitting high levels of electromagnetic radia-
tion (EMI) or radio frequency interference (RFI).
Expose the device; particularly the LCD display or sensor to sources of
extreme heat, cold or excessive sunlight beyond the device’s storage tem-
perature range, refer to Section 8 for extended periods of time.
In a gas stream with a vacuum greater than 14” water column. Immerse the device, oxygen sensor or coiled cable in any liquid. Outside of the parameters specified in Section 8 particularly at flow rates
greater than 10 liters per minute - the backpressure generated produces erroneously high oxygen readings.
Calibrate: (a) with 20.9% oxygen or room air with the intent of taking
oxygen measurements at oxygen levels above 30% oxygen; (b) in a hu­midified gas stream or atmosphere; (c) without allowing a newly installed sensor to stabilize for 15-20 minutes in ambient air.
Attempt to sterilize, autoclave, liquid sterilize, immerse in any liquid or
expose the device or accessories to steam, ethylene oxide or radiation sterilization.
In the presence of flammable gases. Open the main compartment of the device, except to change the integral
oxygen sensor of the AII-3000 AHC or AII-3000 MHC Oxygen Analyzers.
Open the oxygen sensor or probe the sensing surface, refer to Section 10
in the event the sensor should leak and someone comes in contact with the electrolyte from inside the sensor.
Operate with a cable that appears worn, torn or cracked, or, allow an
excess length of cable near the patient’s head or neck; secure it to the bed
rail or other suitable object to avoid the possibility of strangulation.
Allow the device or oxygen sensor to be serviced, repaired or altered by
anyone except trained personnel – failure to do so may endanger the patient or damage the device rendering the warranty null and void.
5
Loading...
+ 13 hidden pages