Analytical Industries Palm O2 IW User Manual

Copyright © 6/11 All Rights Reserved
Analytical Industries Inc. dba Advanced Instruments Inc., 2855 Metropolitan, Pomona, CA 91767 USA. Tel: 909-392-6900, Fax: 909-392 -3665 Email: sales-industrial@aii1.com, Web: www.aii1.com
This manual may not be re produced in whole or in part wi thout the prior written consent of Analytical Industries Inc.
2
0
.9
Table of Contents
1 Introduction 1
1.1 Indications for Use
1.2 Intended Use
1.3 Device Description
2 Quality Control Certification 3 3 Safety Warnings 4 4 Start-up 6
4.1 Contents of Shipping Container
4.2 Controls
4.3 Start-Up Test
4.4 Calibration
4.5 Mounting
5 Operation 10
5.1 Principle of Operation
5.2 Application Considerations
5.3 Calibration
5.4 Sampling
6 Maintenance 14
6.1 Service ability
6.2 Battery Replacement
6.3 Oxygen Sensor Replacement
6.4 Reassembly
6.5 Remote Oxygen Sensor Replacement
7 Troubleshooting 18 8 Specifications 19
8.1 Spare Parts & Optional Accessories 20
9 Warranty 22
10 Material Safety Data Sheet (MSDS) 23
10.1 Disposal 23
1 2 2
6 6 6 7 9
10 11 12 12
14 14 15 16 17
1 Introduction
Congratul a tions o n your p urch ase, t hese Ins tr ucti ons fo r U se des cri be th e pr e­cautions, set-up, operation, maintenance and specifications of the Palm O2 Oxygen Analyzer.
This symbol means CAUTION – Failure to read and comply with the Instructions for Use could damage the device and possibly jeopardize the well being of the user.
Note: Adva nced Ins trume nts Inc. ca nnot warra nt any damag e res ulti ng from the misu se, unauthor ized repair or imp r o per maintenance of the device.
1.1 Indications for Use
The Pal m O2 O xyge n A nal yzer i s i ntend ed t o meas ur e an d di spl ay t he conc en­tration of oxyge n for pers onnel s afety, area mo nitoring, O2 defic iency in con­fined sp aces, ch ecking breathi ng air tanks a nd conf i r m i ng the O2 l ev el s pr i o r t o welding.
Users m ust read th e followi ng statements as they are es sential to re­ducing the r is k of us e erro r du e to ergo nomi c f eat ur es of th e devi ce o r the environment in which the device is intended to be used.
The device has be en desig ne d an d manufa ctured i n s uc h a way that when used under the conditions and for the purposes intended, they will not compromise the safety of the users or other persons.
Conform ity wi th ess enti al re quir eme nts has been demo nstr ate d by ver if ying th e perform ance of the device under normal co nditio ns, be nch testi ng and deter­mining that undesirable malfunctions constitute minimal risk to users.
Do not sterilize, autoclave, liquid sterilize, immerse in any liquid or expose the device or accessories to steam, ethylene ox ide or radiation s t er ilization.
The devic e is inten ded to be re-usabl e. Shoul d the de vice or access ories com e in conta c t w i t h pat ient bodily f l uids, either di s pos e of th e device or clean wi t h a soft cloth dampened with 70% isopropyl alcohol solution in water and allow the components to air-dry before re-use .
Do not o perat e t he an alyz er ne ar equi pme nt c ap able of em itti n g hi gh le vels of electromagnetic radiation as the reading may become unstable.
i
1
In order to obt ain op ti mum p erforman ce, t he o peration of th e d evice m ust b e performed in accordance with th ese Instructions for Us e. Maintenan ce should be performed only by trained personnel authorized by the manufacturer.
1.2 Intended Use
The Palm O2 Ox yge n Analyzer is intend ed to measur e and dis pl ay th e concen­tration of oxygen for personnel saf ety, are a monitoring, O 2 deficiency i n con­fined sp aces, chec king br eathi ng air tanks and confir ming the O2 le vels pri or to welding.
1.3 Device Description
The Pal m O 2 Ox yg en A nal yzer is desi gne d to b e ha nd held but ca n be t em po ­rarily placed in a fixed posi tion on a pole or s helf w i th o ption al br ac ket a tta ch­ments as illustrated in Section 8.1. A remote sensor option is available which makes th e Palm O2 m ore flexi ble and eas ier to use. Either wa y, it provi des continuous, fast, reliable and accurate oxygen measurements.
The device utilizes an electrochemical galvanic fuel cell type oxygen sensor of the type t hat is extens ivel y used to meas ure o xygen co nce ntratio ns f rom 0% to 100% in gas streams. O xygen, the fuel f or this electroch emical transduc er, diffus ing into the s ensor thr ough a g as permeabl e membra ne reacts ch emi­cally at the sensi ng electro de to produ ce an elect rical curr ent o utput propo r­tional to the ox yg en con ce ntra tio n in the g as phas e. T he s enso r h as an abso­lute zero meaning that when no oxygen is present to be chemically reacted the LCD displays 00.0 oxygen.
The sensor’s signal output is linear over the entire range, remains virtually constant over t he specif ied us eful li fe and drop s off sharply at the e nd. The sensor itself requires no maintenance and is simply replaced at the end of its useful l ife like a ba ttery. In asmuch as t he sensor i s a transdu cer in its own right, its expected life is not affected by whether the analyzer is ON or OFF.
A batter y pow ered st ate-of -the-ar t mi cro-pr ocess or con verts t he s ensor’ s sig­nal output repr esentin g the parti al pressure of o x ygen in the gas stream being analyzed. The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display (LCD) that has a resolution of 0.1% oxygen. The microprocessor is controlled from a keypad and provides system diagnostics and warning indicators for continuous monitoring that enhance both safety and effectiveness.
Prior to shipment, every device is thoroughly tested at the factory and docu­mented in the form of a Quality Control Certification that is included in the Instructions for Use supplied with every device.
2 Quality Control Certification
Customer: ________________________ Order No. _____________ Date: _______
Model: Palm O2 Oxygen Analyzer ( ) Industrial
( ) Welding S/N _______________
Sensor: ( ) AII-11-75-PO2
Electronics: A-1190 PCB Assembly Ma in Software Version _______________ Accessories: BATT-1008 Battery, 1.5V AA Alkaline (Qty 2)
PASS QC Test: LCD display 3-1/2 digits ……………………………………………………. ______ Battery symbol displays when battery is low ……………………….. ______ Span adjustment + Following calibration with 99-100% oxygen and flushing with
Span adjustment + Following calibration with air (20.9% oxygen) and exposing
Options: Item No.
Delivery:
( ) AII-11-75-PO2R
P-0188 Manual, Ins tructions for Use ……………………. Included ________
ambient air, oxygen reading as displayed by LCD 20.9% +
to 99-100% oxygen, LCD displays 100% + Overall inspection for physical defects ………………………………... ______
Qty Item No. Qty
A-3692 Hand Pump Kit (A-3674, PUMP-1020, TUBE-1018) …………
FITN-1009 Tee Ad a pter …………. FITN-1112-1 Flow Diverter ……..
HRWR-1157 Screwdriver …………
HRWR- 1158 Lan yard ……………… Other:
(plus A-3654, CABL-1009) S/N _______________
10-30% FS with 100% oxygen calibration ______
2% ____
10-30% FS with air calibration ……………….. ______
2% ……………………. ______
A-3657-1 Dovetail Mounting Kit (A-3657, HRWR-1162) ..…………….
____
HRWR-1075 Dove ta il Cla mp ………
____
A-3388 Adapter, Dome to Sensor .
____
A-3676 Adapter, 1/8” to Sensor …
____
TUBE-1018 Tubing, 1/8” x 3’ ……..
____
A-3609 Adapter, 1/8” to Dome ……
____
____ ____
____ ____ ____
___
2
3
3 Safety Warnings
ALWAYS follo w the s tateme nts bel ow as they are ess enti al t o red uci ng the risk of use error due to ergonomic features of the device or the environment in which t he device is inte nded to be used.
Only trained personnel who have read, understand and agree to follow the
Inst ructions for Use shou l d operate the device .
Retain the Instructions for Use for future reference. Refer service needs to trained authorized personnel. Failure to do so may
cause the dev ice to fail and void the warranty.
Inspect the d e vice and acc es s or i es bef o re operatin g and ens ure: (a) there
is no evidence of physical damage; (b) the sensor (particularly the sensing surface) and electrical connections are dry; and, (c) the sensor is installed and is upstream from any h umidifying device for accura te calibration and oxygen readings.
Calibrate: (a) with a known source of dry air or 100% oxygen before using
each da y o r af t er 8 hours of co nt i n uo us us e; ( b) whe n t h e temp erature or pressure of the operating environment changes; (c) if the oxygen sensor has been disconn ected and r econne cted; (d) af ter the battery or oxyge n sensor has been replaced.
Sampling flowing gas: (a) install the optional accessories as shown in Sec-
tion 8.1 and (b) assure there is a tight fit between the components.
Sampling static, ambient or controlled atmospheres unscrew and remove
all components from the oxygen sensor.
 Clean the device and accessories in accordance with Section 6.1.2.  Batter y repla cement Section 6.2: (a) repl ace the batteri es whe n the ‘ LO’
message is displayed on the LCD and (b) calibrate the analyzer after re­placing the batteries.
Oxygen sensor installation or replacement Section 6.3 or 6.5: allow the
new sensor to stabilize for 15-20 minutes in ambient air before attempting to calibrate.
Store the device by turning the power OFF and removing the batteries if
the device will not be operated for over thir ty (3 0) days.
Attempt to repeat the procedure that caused a perceived malfunction and
refer to troubleshooting hints in Section 7 before concluding the device is faulty. If in doubt, contact the manufacturer for assistance.
NEVER o perate the d evice in any m anner des cribed below doing so may compromise the clinical condition or the safety of patients, users or other persons.
If the reading is unstable or a malfunction is suspected. After the ‘ERR’ or ‘LO’ messages are displayed on the LCD. Near equipme nt cap abl e o f emi tting hi g h level s of elect ro m a gnetic radia-
tion (EMI) or radio frequency interference (RFI).
Expose the device; particularly the LCD display or sensor to sources of
extreme heat, cold or excessive sunlight beyond the device’s storage temperature range, refer to Section 8 for extended periods of time.
In a gas stream with a vacuum greater than 14” water column. Immerse the device, oxygen sensor or optional coiled cab le in any liquid. Outside of the parameters specified in Section 8 particularly at flow rates
greater than 10 liters per min ute - the b ackpressure generated produces erroneously high oxygen readings.
Calibrate: (a) with 20.9% ox ygen or room air with the intent of taking
oxygen measurements at oxygen levels above 30% oxygen; (b) in a humidified gas stream or atmosphere; (c) without allowing a newly in­stalled sensor to stab ilize for 15-20 minutes in ambient air.
Attempt to sterilize, autoclave, liquid sterilize, immerse in any liquid or
expose the devic e or acces sories to steam, ethylene o xide or r adiation sterilization.
Open the main com par tm e nt of t he d evice, ex cept to ch a nge t he i nt egral
oxygen sensor.
Open the oxygen sensor or probe the sensing surface, refer to Section 10
in the event the sensor should leak and someone comes in contact with the electrolyte from inside the sensor.
Optional remote se nsor with a cable that a ppears wor n, torn or cr acked,
or, allow an excess length of cable near the patient’s head or neck; se­cure it to the bed rail or other suitable object to avoid the possibility of strangulation.
Allow the devi ce or o x ygen sens or to be s ervi ced, re pair ed or alter e d by
anyone except trained personnel – failure to do so may endanger th e patient or damage the device rendering the warranty null and void.