2855 Metropolitan Place, Pomona, California 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: info@aii1.com
Table of Contents
Introduction 1
Quality Control Certification 2
Safety & Installation 3
Features & Specifications 4
Operation 5
Maintenance 6
Spare Parts 7
Troubleshooting 8
Warranty 9
Material Safety Data Sheets 10
Advanced Instruments Inc.
1 Introduction
Your new oxygen analyzer is a precision piece of equipment designed to give you years of use in variety of industrial oxygen
applications.
This analyzer is designed to measure the oxygen concentration in inert gases, gaseous hydrocarbons, hydrogen, and a variety
of gas mixtures. In order to derive maximum performance from your new oxygen analyzer, please read and follow the
guidelines provided in this Owner’s Manual.
The serial number of this analyzer may be found on the inside the analyzer. You should note the serial number in the space
provided and retains this Owner’s Manual as a permanent record of your purchase, for future reference and for warranty
considerations.
Serial Number: _______________________
Every effort has been made to select the most reliable state of the art materials and components designed for superior
performance and minimal cost of ownership. This analyzer was tested thoroughly by the manufacturer for best performance.
However, modern electronic devices do require service from time to time. The warranty included herein plus a staff of trained
professional technicians to quickly service your analyzer is your assurance that we stand behind every analyzer sold.
Advanced Instruments Inc. appreciates your business and pledge to make effort to maintain the highest possible quality
standards with respect to product design, manufacturing and service.
( ) Stainless steel sensor housing, manual flow control and bypass valves, ¼” compression
( ) Temperature controlled heater system 85°F specify: ( ) 110VAC ( ) 220VAC Power: 100/120/220/250 VAC (universal without temperature controlled heater systems) Enclosure: ( ) Std. panel mount (“T”) 7.5x10.8x12”; ( ) “TO” option 7.75x 7.75x12”
Test System start-up diagnostics satisfactory Auto/manual range Alarm relays activate/deactivate with changes in O2 concentration Alarm bypass Analog outputs: Signal output 4-20mA Range ID: ( ) 4-20mA or ( ) 5x relay contacts plus 1x common Recovery from air to < 10 ppm in < 15 minutes Baseline drift on zero gas < ± 2% FS over 24 hour period Noise level < ± 1.0% FS Span calibration gas value Span adjustment within 10-50% FS Peak to peak over/under shoot < 0.5% FS Overall inspection for physical defects
Options Label analyzer “Oxygen Service” in accordance with P-1507 Rev-1; see certificate next page.
Notes
A-1146-10 PCB Assembly Main / Display Software V. ______
( ) A-1174-10 PCB Assy Power Supply / Interconnect, 4-20mA Range ID
( ) A-1174-10C PCB Assy Power Supply / Interconnect, 5x Relay Contacts Range ID
type fittings for sample inlet and vent
( ) Delete bypass valve from above (T and TO options)
( ) Sample, span, zero inlet solenoid valves
( ) Bezel for 19” rack mmount 19x12x12” option
( ) GPR-1600-W option general purpose wall mount 12x12x8”
( ) GPR-1600-W306 option general purpose panel mount 18.2x16x10”
Compressed Gas Association,
Publication: G-4.1 Edition 4,
Title: Cleaning Equipment for Oxygen Service,
Published 1/1/1996 and related publications
Mfg. Item No.: GPR-1600 Series
Description: ppm Oxygen Analyzer
Serial No.:
Customer:
Purchase Order:
Quantity: 1 of
Warranty Date: 12 months from ______________
The undersigned warrants on behalf of Manufacturer that the product identified above
conforms to the manufacturing, testing and packaging criteria set forth by the ‘Standard’
specified above.
______________
Place: Pomona, CA
By print name:
Signature:
Title:
4
Advanced Instruments Inc.
3 Safety Guidelines
Safety
This section summarizes the basic precautions applicable to all analyzers. Additional precautions specific to individual analyzer
are contained in the following sections of this manual. To operate the analyzer safely and obtain maximum performance follow
the basic guidelines outlined in this Owner’s Manual.
Caution: This symbol is used throughout the Owner’s Manual to Caution and alert the user to recommended safety and/or
operating guidelines.
Danger: This symbol is used throughout the Owner’s Manual to identify sources of immediate Danger such as the presence of
hazardous voltages.
Read Instructions: Before operating the analyzer read the instructions.
Retain Instructions: The safety precautions and operating instructions found in the Owner’s Manual should be retained for
future reference.
Heed Warnings Follow Instructions: Follow all warnings on the analyzer, accessories (if any) and in this Owner’s Manual.
Observe all precautions and operating instructions. Failure to do so may result in personal injury or damage to the analyzer.
Heat: Situate and store the analyzer away from sources of heat.
Liquid and Object Entry: The analyzer should not be immersed in any liquid. Care should be taken so that liquids are not
spilled into and objects do not fall into the inside of the analyzer.
Handling: Do not use force when using the switches and knobs. Before moving your analyzer be sure to disconnect the
wiring/power cord and any cables connected to the output terminals located on the analyzer.
Maintenance
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operator to service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in this Owner’s Manual. Avoid contact with any liquid or crystal
type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should be
disposed of in accordance with local regulations.
Troubleshooting: Consult the guidelines in section 8 for advice on the common operating errors before concluding that your
analyzer is faulty. Do not attempt to service the analyzer beyond those means described in this Owner’s Manual.
Do not attempt to make repairs by yourself as this will void the warranty, as detailed by section 9, and may result in electrical
shock, injury or damage. All other servicing should be referred to qualified service personnel.
Cleaning: The analyzer should be cleaned only as recommended by the manufacturer. Wipe off dust and dirt from the outside
of the unit with a soft damp cloth then dry immediately. Do not use solvents or chemicals.
Nonuse Periods: Disconnect the power when the analyzer is left unused for a long period of time.
5
Advanced Instruments Inc.
Installation
Gas Sample Stream: Ensure the gas stream composition of the application is consistent with the specifications and review the
application conditions before initiating the installation. Consult the factory to ensure the sample is suitable for analysis.
Contaminant Gases: A gas scrubber and flow indicator with integral metering valve are required upstream of the of the
analyzer to remove interfering gases such as oxides of sulfur and nitrogen or hydrogen sulfide that can produce false readings,
reduce the expected life of the sensor and void the sensor warranty if not identified at time of order placement. Installation of a
suitable scrubber is required to remove the contaminant from the sample gas to prevent erroneous analysis readings and
damage to the sensor or optional components. Consult the factory for recommendations concerning the proper selection and
installation of components.
Expected Sensor Life: With reference to the publish specification located as the last page of this manual, the expected life of
all oxygen sensors is predicated on oxygen concentration (< 1000 ppm or air), temperature (77°F/25°C) and pressure (1
atmosphere) in “normal” applications. As a rule of thumb sensor life is inversely proportional to changes in the parameters.
Deviations are outside the specifications and will affect the life of the sensor. Avoid exposure to oxygen levels above 1% (10000
ppm) for hours at a time.
Failure to do so will result in damage to the sensor(s) as follows:
¾ GPR Series ppm sensors – reduced sensor life and loss of low end sensitivity (GPR sensor exposed continuously to the
20.9% content of air will last approximately 3.5 months and will develop a low end offset > 10-20 ppm);
¾ XLT Series ppm sensors - reduced sensor life and loss of low end sensitivity (XLT sensor exposed continuously to the
20.9% O2 content of air will last approximately 7 days and will develop a low end offset > 10-20 ppm)
Accuracy & Calibration: Refer to section 5 Operation. The 0-25% Range is provided only for the purpose of air calibration
which is recommended only if span gas is not available. Bringing the analyzer back online after calibration with the 20.9% or
209,000 ppm oxygen content of air takes longer than calibrating the analyzer with a span gas containing 80 ppm or 800 ppm
oxygen.
Materials: Assemble the necessary zero, purge and span gases and optional components such as valves, coalescing or
particulate filters, and, pumps as dictated by the application; stainless steel tubing is essential for maintaining the integrity of
the gas stream for ppm and percentage range (above or below ambient air) analysis; hardware for mounting.
Operating Temperature: The sample must be sufficiently cooled before it enters the analyzer and any optional components.
A coiled 10 foot length of ¼” stainless steel tubing is sufficient for cooling sample gases as high as 1,800ºF to ambient. The
maximum operating temperature is 45º C on an intermittent basis unless the user is willing to accept a reduction in expected
sensor life – refer to analyzer specification - where expected sensor life is specified at an oxygen concentration less than 1000
ppm oxygen for ppm analyzers and air (20.9% oxygen) for percent analyzers, but in all instances at 25°C and 1 atmosphere of
pressure. Expected sensor varies inversely with changes in these parameters.
Pressure & Flow
All electrochemical oxygen sensors respond to partial pressure changes in oxygen. The sensors are equally capable of analyzing
the oxygen content of a flowing sample gas stream or monitoring the oxygen concentration in ambient air (such as a confined
space such in a control room or an open area such as a landfill or bio-pond).
Analyzers designed for in-situ ambient or area monitoring have no real inlet and vent pressure because the sensor is exposed
directly to the sample gas and intended to operate at atmospheric pressure, however, slightly positive pressure has minimal
effect on accuracy.
Sample systems and flowing gas samples are generally required for applications involving oxygen measurements below 1% and
at a pressure other than ambient air. In these situations, the use of stainless steel tubing and fittings is critical to mainta
the integrity of
which is normally at atmospheric pressure.
The sensor is exposed to sample gas that must flow or be drawn through metal tubing inside the analyzer. The internal sample
system includes 1/8” compression inlet and vent fittings, a stainless steel sensor housing with an o-ring seal to prevent the
leakage of air and stainless steel tubing.
the gas stream to be sampled and the inlet pressure must always be higher than the pressure at the outlet vent
ining
6
Advanced Instruments Inc.
Inlet Pressure: Analyzers designed for flowing samples under positive pressure or pump vacuum (for samples at atmospheric
or slightly negative atmospheres) that does not exceed 14” water column are equipped with bulkhead tube fitting connections
on the side of the unit (unless otherwise indicated, either fitting can serve as inlet or vent) and are intended to operate at
positive pressure regulated to between 5-30 psig although the rating of the fitting itself is considerably higher. Caution: If the
analyzer is equipped with an optional H2S scrubber, inlet pressure must not exceed 30 psig.
Outlet Pressure: In positive pressure applications the vent pressure must be less than the inlet, preferably atmospheric.
Flow Rate: Flow rates of1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH
generate backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample
gas at the higher flow rate. The direction the sample gas flows is not important, thus either tube fitting can serve as the inlet or
vent – just not simultaneously.
A flow valve upstream (with flow indicator positioned downstream) of the sensor is recommended as a means of controlling the
flow rate of the sample gas, minimizing air leaks and produce optimum accuracy. A flow rate of 2 SCFH or 1 liter per minute is
recommended for optimum performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is flowing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty). To avoid generating a vacuum on the sensor (as described above) during operation, always select and install
the vent fitting first and remove the vent fitting last.
Application Pressure - Positive: A flow indicator with integral metering valve positioned upstream of the sensor is
recommended for controlling the sample flow rate between 1-5 SCFH. To reduce the possibility of leakage for low ppm
measurements, position a metering needle valve upstream of the sensor to control the flow rate and position a flow indicator
downstream of the sensor. If necessary, a pressure regulator (with a metallic diaphragm is recommended for optimum
accuracy, the use of diaphragms of more permeable materials may result in erroneous readings) upstream of the flow control
valve should be used to regulate the inlet pressure between 5-30 psig.
Caution: If the analyzer is equipped with a H2S scrubber as part of an optional sample conditioning system, inlet pressure
must not exceed 30 psig.
Application Pressure - Atmospheric or Slightly Negative: For accurate ppm range oxygen measurements, an optional
external sampling pump should be positioned downstream of the sensor to draw the sample from the process, by the sensor
and out to atmosphere. A flow meter is generally not necessary to obtain the recommended flow rate with most sampling
pumps.
Caution: If the analyzer is equipped with an optional flow indicator with integral metering valve or a metering flow control
valve upstream of the sensor - open the metering valve completely to avoid drawing a vacuum on the sensor and placing an
undue burden on the pump. If pump loading is a consideration, a second throttle valve on the pump’s inlet side may be
necessary to provide a bypass path so the sample flow rate is within the above parameters.
Recommendations to avoid erroneous oxygen readings and damaging the sensor:
¾ Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is flowing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (thus voiding
the sensor warranty).
¾ Assure there are no restrictions in the sample or vent lines
¾ Avoid drawing a vacuum that exceeds 14” of water column pressure – unless done gradually
¾ Avoid excessive flow rates above 5 SCFH which generate backpressure on the sensor.
¾ Avoid sudden releases of backpressure that can severely damage the sensor.
¾ Avoid the collection of liquids or particulates on the sensor, they block the diffusion of oxygen into the sensor
¾ If the analyzer is equipped with an optional integral sampling pump (positioned downstream of the sensor) and a flow
control metering valve (positioned upstream of the sensor), completely open the flow control metering valve to avoid
drawing a vacuum on the sensor and placing an undue burden on the pump.
- wipe away.
7
Advanced Instruments Inc.
Moisture & Particulates: Installation of a suitable coalescing or particulate filter is required to remove condensation, moisture
and/or particulates from the sample gas to prevent erroneous analysis readings and damage to the sensor or optional
components. Moisture and/or particulates do not necessarily damage the sensor, however, collection on the sensing surface can
block or inhibit the diffusion of sample gas into the sensor resulting in a reduction of sensor signal output – and the appearance
of a sensor failure when in fact the problem is easily remedied by blowing on the front of the sensor. Consult the factory for
recommendations concerning the proper selection and installation of components.
Moisture and/or particulates generally can be removed from the sensor by opening the sensor housing and either blowing on
the sensing surface or gently wiping or brushing the sensing surface with damp cloth. Caution: Minimize the exposure of ppm
sensors to air during this cleaning process. Air calibration followed by purging with zero or a gas with a low ppm oxygen
concentration is recommended following the cleaning process. Moisture and/or particulates generally can be removed from the
sample system by flowing the purge gas through the analyzer at a flow rate of 4.5-5 SCFH for an hour.
Mounting: The analyzer is approved for indoor use, outdoor use requires optional enclosures, consult factory. Mount as
recommended by the manufacturer.
Gas Connections: Inlet and outlet vent gas lines for ppm analysis require 1/8” or ¼” stainless steel compression fittings; hard
plastic tubing with a low permeability factor can be used percentage range measurements.
Power: Supply power to the analyzer only as rated by the specification or markings on the analyzer enclosure. The wiring that
connects the analyzer to the power source should be installed in accordance with recognized electrical standards. Ensure that is
properly grounded and meets the requirements for area classification. Never yank wiring to remove it from a terminal
connection. AC powered analog analyzers consume 5 watts, digital analyzers 50 watts without optional heaters. Optional 110V
and 220V heaters AC powered heaters consume an additional 100-150 watts; DC powered digital analyzers consume 30 watts,
40 watts with the optional DC powered heater.
8
Advanced Instruments Inc.
4 Features & Specifications
9
Advanced Instruments Inc.
10
Advanced Instruments Inc.
5 Operation
Principle of Operation
The GPR-1600 ppm Oxygen Analyzers incorporates a a variety of ppm range advanced galvanic fuel cell type sensors and is
configured for panel mounting and requires a 7.5x10.8” (T configuration) cutout with 4 holes for the analyzer’s front panel.
Optional configurations include a panel mount (TO configuration) 7.75x7.75” with cutout; 19” bezel for rack mounting either the
T or TO; 12x12x8” wall mount enclosure (GPR-1600W); 18.2x16x10” panel mount configuration (GPR-1600W-306) using the
wall mount enclosure. Contact the factory for additional information on options. All configurations are tested and calibrated by
the manufacturer prior to shipment. The GPR-1600 analyzers and sensors are CE certified and manufactured under a Quality
Assurance System certified by an independent agency to ISO 9001:2000 standards.
Advanced Galvanic Sensor Technology
The sensors function on the same principle and are specific for oxygen. They measure the partial pressure of oxygen from low
ppm to 100% levels in inert gases, gaseous hydrocarbons, helium, hydrogen, mixed gases, acid gas streams and ambient air.
Oxygen, the fuel for this electrochemical transducer, diffusing into the sensor reacts chemically at the sensing electrode to
produce an electrical current output proportional to the oxygen concentration in the gas phase. The sensor’s signal output is
linear over all ranges and remains virtually constant over its useful life. The sensor requires no maintenance and is easily and
safely replaced at the end of its useful life.
Proprietary advancements in design and chemistry add significant advantages to an extremely versatile oxygen sensing
technology. Sensors for low ppm analysis recover from air to ppm levels in minutes, exhibit longer life, extended operating
range of -20°C to 50°C, excellent compatibility with CO
significant advantage over the competition. Other advancements include extending the expected life of our new generation of
percentage range sensors now range to five and ten years with faster response times and greater stability. Another significant
development involves the first galvanic oxygen sensor capability of continuous oxygen purity measurements and expanding the
operating temperature range from -40°C to 50°C.
Oxygen, the fuel for this electrochemical transducer, reacts chemically at the sensing electrode to produce an electrical current
output proportional to the oxygen concentration in the gas phase. The sensor’s signal output is linear over all four ranges and
remains virtually constant over its useful life. The sensor requires no maintenance or electrolyte addition and is easily and safely
replaced at the end of its useful life.
and acid gases (XLT series) and reliable quality giving them a
2
11
Advanced Instruments Inc.
Electronics
The signal generated by the sensor is processed by state of the art low power micro-processor based digital circuitry. The first
stage amplifies the signal. The second stage eliminates the low frequency noise. The third stage employs a high frequency filter
and compensates for signal output variations caused by ambient temperature changes. The result is a very stable signal.
Sample oxygen is analyzed very accurately. Response time of 90% of full scale is less than 10 seconds (actual experience may
vary due to the integrity of sample line connections, dead volume and flow rate selected) on all ranges under ambient
monitoring conditions. Sensitivity is typically 0.5% of full scale low range.
Additional features of the micro-processor based electronics include manual or auto ranging, auto-zero and auto-cal, isolated 420mA signal for signal output and range ID, separate relay contacts rated 30VDC max @ 1A are provided for the alarm feature
and an optional range ID feature (auto-zero/auto-cal with relay contacts for Range ID is special order, so . Whenever the
analyzer is calibrated, a unique algorithm predicts and displays a message indicating a ‘weak sensor’ suggesting the sensor be
replaced in the near future.
Sample System
The sample must be properly presented to the sensor to ensure an accurate measurement. In standard form the GPR-1600 is
designed with a sample system that complements the performance capabilities of the advanced oxygen sensor and enables the
user to isolate the sensor from exposure to high oxygen concentration which results is a substantial increase is user
productivity. This bypass feature has two important features: one, the sensor can be isolated from exposure to h igh oxygen
levels when changing sample lines, during transport and during standby intervals making it ideal for mobile cart applications.
Two, it enables the user to purge newly connected gas lines of the oxygen trapped inside. The result is an analyzer that comes
on-line at ppb levels in a matter of minutes and provides users with a significant increase in productivity.
For ppb and ppm trace oxygen measurements, the sensor is exposed to sample gas that must flow or be drawn through the
analyzer’s internal sample system. This unique sample system, when operated accordingly to the instructions in this Owner’s
Manual, can significantly increase user productivity by minimizing the sensor’s exposure to ambient air or high oxygen
concentrations which contribute to the significant amount of downtime associated with competitive analyzers.
The advantages of the bypass sample system include:
¾ Supplying the analyzer with the sensor it was qualified with.
¾ Isolating the sensor during transport, calibration and maintenance intervals when changing gas line connections.
¾ Isolating the sensor from exposure to high oxygen levels during upset conditions which extends sensor life.
¾ Purging the air (or high oxygen levels above 1,000 ppm) trapped in the gas lines following a process upset.
Advanced Instruments Inc. offers a full line of sample handling, conditioning and expertise to meet your application
requirements. Contact us at 909-392-6900 or e-mail us at info@aii1.com
12
Advanced Instruments Inc.
Accuracy Overview
Single Point Calibration: As previously described
the galvanic oxygen sensor generates an electrical
current proportional to the oxygen concentration in
the sample gas. In the absence of oxygen the sensor
exhibits an absolute zero, e.g. the sensor does not
generate a current output in the absence of oxygen.
Given these linearity and absolute zero properties,
single point calibration is possible.
Pressure: Because sensors are sensitive to the
partial pressure of oxygen in the sample gas their
output is a function of the number of molecules of
oxygen 'per unit volume'. Readouts in percent are
permissible only when the total pressure of the
sample gas being analyzed remains constant. The
pressure of the sample gas and that of the calibration
gas(es) must be the same (reality < 1-2 psi).
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise known as an
'oxygen diffusion limiting barrier' and all diffusion processes are temperature sensitive, the fact the sensor's electrical output
will vary with temperature is normal. This variation is relatively constant 2.5% per ºC. A temperature compensation circuit
employing a thermistor offsets this effect with an accuracy of +
independent of temperature. There is no error if the calibration and sampling are performed at the same temperature or if the
measurement is made immediately after calibration. Lastly, small temperature variations of 10-15º produce < 1% error.
Accuracy:
1) those producing 'percent of reading errors', illustrated by Graph A below, such as +
tolerances of range resistors and the 'play' in the potentiometer used to make span adjustments and
2) those producing 'percent of full scale errors', illustrated by Graph B, such as +
are really minimal due to today's technology and the fact that other errors are 'spanned out' during calibration.
Graph C illustrates these 'worse case' specifications that are typically used to develop an analyzer's overall accuracy statement
of < 1% of full scale at constant temperature or < 5% over the operating temperature range. QC testing is typically < 0.5%
prior to shipment.
Example 1: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation circuit,
during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78 (100/20.9) if used for
measurements of 95-100% oxygen concentrations. Conversely, an error during a span adjustment at 100% of full scale range is
reduced proportionately for measurements of lower oxygen concentrations.
Refer to the Calibration section for additional details.
In light of the above parameters, the overall accuracy of an analyzer is affected by two types of errors:
5% or better and generates an output function that is
5% temperature compensation circuit,
1-2% linearity errors in readout devices, which
13
Advanced Instruments Inc.
Mounting the Analyzer
The standard GPR-1600 is designed to be panel mounted and requires a cutout that accommodates the enclosure and 4
mounting bolts. The design also lends itself to 19” rack mounting with an optional bezel or wall mount enclosures as illustrated
below.
Procedure:
1. The GPR-1600 is designed for panel mounting directly to any flat vertical surface, wall or bulkhead plate with the
appropriate cut out and four ¼” diameter holes for insertion of the mounting studs located on the back side of the front
panel.
2. When mounting the analyzer position it approximately 5 feet off the floor for viewing purposes and allow sufficient room for
access to the terminal connections at the rear of the enclosure.
3. Note: The proximity of the analyzer to the sample point and use of optional sample conditioning components have an
impact on sample lag time.
Mounting GPR-1600:
14
Mounting GPR-1600-W Option:
Advanced Instruments Inc.
15
Mounting GPR-1600-W-306 Option:
Advanced Instruments Inc.
16
Advanced Instruments Inc.
Gas Connections
The GPR-1600 with its standard flow through configuration is designed for positive pressure
samples and requires connections for incoming sample and outgoing vent lines, see
illustrations above.
The user is responsible for calibration gases and the components described below. Flow rates
of 1-5 SCFH cause no appreciable change in the oxygen reading. A flow valve upstream (flow
indicator downstream) of the sensor is recommended as a means of controlling the flow rate
of the sample gas. A flow rate of 2 SCFH is recommended for optimum performance.
Caution: Do not place your finger over the fitting designated as the vent (it pressurizes the
sensor) or to test the flow indicator when gas is flowing to the sensor. Removing your finger
(the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).
Procedure:
1. Caution: Do not change the factory setting until instructed.
2. Regulate the pressure and flow as described in Pressure & Flow above.
3. Install the sample out or vent line connection to the 1/8” dia. fitting labeled SAMPLE VENT.
4. Install the incoming sample or span gas line to the 1/8” dia. fitting labeled SAMPLE IN.
5. Set the flow rate to 1 SCFH (open the flow control valve completely if using an external sampling pump positioned
downstream of the sensor).
6. Allow gas to flow through the analyzer for 3-5 minutes and proceed to Calibration or Sampling.
17
Advanced Instruments Inc.
Electrical Connections:
Incoming power for the 100-250V AC powered analyzers is supplied through a universal power entry module. A standard
computer type power cord (P/N A-1008) is required for the universal power entry module. A well grounded insulated power
cable is recommended to avoid noise resulting from unwanted interference.
The appropriate AC power supply (110V or 220V) must specified be specified at order placement if the analyzer is to be
equipped with proper the temperature control heater system.
Power consumption is approximately 150-200 watts with the temperature control heater system and 30 watts without.
Caution: Integral 4-20mA converters are internally powered and do not require external power. DO NOT supply any voltage to
any of the terminals for 4-20mA signal output and range ID or the 4-20mA converters will be damaged.
Caution: To assure proper grounding, connect the 4-20mA signal output to the external device (PLC, DCS, etc.) before
attempting any zero or span adjustments.
Optional Range ID:
The standard 4-20mA output used for range identification, as described below, can be replaced by eliminating the alarms
feature and using the relay contacts associated with the alarms to provide a single common and four (4) normally open relay
contacts that close when the related range is active. The dry contacts are rated at 30VDC @ 1A and powering them is not
required if the PLC can distinguish contact closure via continuity check.
18
Advanced Instruments Inc.
Procedure:
1. As illustrated above the sensor, power and alarm relays and signal output connections are hard wired to screw type
terminal blocks located at the rear of the analyzer.
2. Use a small bladed screwdriver to loosen the appropriate terminal screws as illustrated above.
3. Strip the wires of the cable no more than 3/16 inch.
4. To connect to an active relay or “fail safe”, connect the live cable to the common terminal C and the secondary cable to the
normally open NO terminal.
5. To break the connection upon relay activation, connect the secondary cable to the normally closed NC terminal.
6. Insert the stripped end of the cables into the appropriate terminal slots assuring no bare wire remains exposed that could
come in contact with the back panel of the analyzer enclosure.
7. Tighten the terminal screws to secure the wires of the cable.
Danger: While connecting the cables to the relay terminals, ensure there is no voltage on the cables to prevent electric shock
and possible damage to the analyzer. Caution: Assure the stripped wire ends of the cable are fully inserted into the terminal
slots and do not touch each other or the back panel of the analyzer enclosure.
Interconnections for the optional wall mount enclosure pictured below.
19
Advanced Instruments Inc.
Alarms
The analyzer is configured with two user adjustable threshold type alarm relays that can be configured in the field from the
ALARM option on the MAIN MENU as follows:
¾ Establish independent set points
¾ Either Hi or Lo
¾ Either On or Off (enabled or disabled)
¾ Both temporarily defeated using a user entered ‘timeout’ period (normally minutes)
The alarm set point represents a value. When the oxygen reading exceeds (high alarm) or falls below (low alarm) the alarm set
point, the relay is activated and the LCD displays the alarm condition.
When activated the alarms trigger SPDT Form C non-latching relays @ 5A, 30VDC or 240VAC resistive. To prevent chattering of
the relays, a 2% hysteresis is added to the alarm set point. This means that the alarm will remain active until the oxygen
reading has fallen 2% below the alarm set point (high alarm) or risen 2% above the alarm set point (low alarm) after the alarm
was activated. The timeout feature is useful while replacing the oxygen sensor or during calibration when the oxygen reading
might well rise above the alarm set point and trigger a false alarm.
Note: When making connections the user must decide whether to configure/connect Alarm 1 and Alarm 2 in failsafe mode
(Normally Open – NO – where the alarm relay de-energizes and closes in an alarm condition) or non-failsafe mode (Normally
Closed – NC – where alarm relay energizes and opens in an alarm condition).
Power Failure Alarm
A dry contact rated at 30VDC @ 1A is provided as a power failure alarm that activates when power supplied to the analyzer’s
circuits is unacceptable. The contact is normally closed but opens when the power to the analyzer is switched off or interrupted
and cannot be disabled.
4-20mA Signal Output
The analyzer provides a 4-20mA full scale fully isolated ground signals for external recording devices. The integral IC on the
main PCB provides 4-20mA fully isolated signals for output and range ID. The 4-20mA current output is obtained by connecting
the current measuring device between the positive and negative terminals labeled OUTPUT 4-20mA. To check the signal output
of the 4-20mA E/I integrated circuit connect an ammeter as the measuring device and confirm the output is within +
4mA. A finer adjustment of the zero offset of the 4-20mA converter can be provided by a potentiometer mounted on the main
PCB Assembly. Consult factory for instructions
Range ID
For range ID the output of 4mA, 8mA, 12mA, 16mA, 20mA correspond to the most sensitive to least sensitive analysis range.
The standard 4-20mA output used for range indentification, as described below, can be replaced by eliminating the alarms
feature and using the relay contacts associated with the alarms to provide a single common and four (4) normally open relay
contacts that close when the related range is active. The dry contacts are rated at 30VDC @ 1A and powering them is optional
as some PLCs can distinguish contact closure via continuity check.
Caution:
voltage to any of the two terminals of the 4-20mA converter.
The integral 4-20mA converters are internally powered and do not require external power. DO NOT supply any
0.1mA of
20
Advanced Instruments Inc.
Temperature Controlled Heater System with Runaway Protection Circuit
The standard GPR-1600 Series analyzer is not equipped with the heater system. However, in anticipation of very low ppm (high
ppb) oxygen analysis, the user may elect to add the heater system. If the analyzer is equipped with an optional temperature
controlled heater system, open the front door of the analyzer to access it. This unit is a PID controller which operates between
0-99°F. The controller is programmed to maintain the temperature at 85°F.
Caution: Do not change this setting. A higher temperature
setting may drastically reduce sensor life and possibly cause
damage to the electronic circuitry of both the controller and the
analyzer.
Warning:
the temperature controller is ON.
When power is applied to the temperature controller, the controller
tunes itself to eliminate and/or minimize the over/under shoot of
temperature from the set point. It is recommended that at initial
start-up, when replacing the oxygen sensor or when trouble
shooting, turn off the power to the heater or set the temperature
set point at 60°F (to turn the heater off) to prevent overheating
the analyzer. When operating the analyzer under normal
conditions, set the temperature controller at 85⁰F.
Changing the display value from °F to °C:
1. Push the UP ARROW and ENTER buttons down for 5 seconds to access the SECURE MENU
2. Press INDEX to advance to the F-C MENU
3. Select °C or °F by pressing the UP ARROW key
4. Press the ENTER key when F-C starts flashing on the display
5. Press INDEX to exit the SECURE MENU
Heater Runaway Protection
Part of the optional temperature controlled heater system is a heater runaway protection circuit that protects the electronics in
the event the temperature controller should fail and thereby allowing the heater to runaway damaging the components inside
the analyzer.
The runaway protection is provided by a J2 type device positioned between the temperature controller and the heater. This
device cuts-off power to the heater if the temperature inside the analyzer exceeds 70°C. Should the J2 device cut power to the
heater, correct the problem and reset the runaway protector device by exposing it to 0°C for a few minutes (a refrigerator
freezer will do).
Keep the front door securely fastened and closed when
Installing the Oxygen Sensor
The analyzer is equipped with an internal oxygen sensor that has been
tested and calibrated by the manufacturer prior to shipment and is fully
operational from the shipping containers. The sensor has been installed at
the factory , however, it may be necessary to install the sensor in the field.
Caution: Review procedure before proceeding, mainly 2 and 9.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive
liquid electrolyte that could be harmful if touched or ingested, refer to the
Material Safety Data Sheet contained in the Owner’s Manual appendix.
Avoid contact with any liquid or crystal type powder in or around the sensor
or sensor housing, as either could be a form of electrolyte. Leaking sensors
should be disposed of in manner similar to that of a common battery in
accordance with local regulations.
Procedure:
21
Advanced Instruments Inc.
1. The sensor has not been installed at the factory (in standard configuration there are no valves to isolate the sensor) and it
will be necessary to install the sensor in the field.
2. As described above the following steps should already be completed:
a) Secure the sensor housing bracket with two 6/32 mounting screws, in the preferred position the upper section with the
interconnection cable should be facing the ceiling;
b) connect the gas lines;
c) electrical connections.
1. Caution: Do not change the factory settings until instructed to do in this manual.
2. Purge the oxygen trapped in the newly connected gas lines for 3-5 minutes.
3. Flow zero gas or sample gas with a low ppm oxygen concentration to the analyzer at the predetermined flow rate of 1
SCFH.
4. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located under the sensor housing, see photo.
5. Rotate the upper section of the sensor housing 90º to disengage from the clamp.
6. Remove the upper section by pulling it straight up and place it on a smooth
surface.
7. Select the AUTO RANGING option from the SAMPLE menu with gas flowing to the
analyzer.
8. Remove the oxygen sensor from the bag and remove the red shorting device
(including the gold ribbon) from the PCB located at the rear of the sensor.
9. Minimize the time the sensor is exposed to ambient air.
10. Immediately place the sensor in the bottom section of the sensor housing with the
PCB facing up.
11. Immediately place the upper section of the sensor housing over the sensor, gently
push the upper section downward and rotate 90º to engage the clamp.
12. Finger tighten the clamp bolt and then tighten it one full turn with the 5/16 wrench
to securely lock the two sections of the sensor housing.
13. The analyzer will OVER RANGE for a short period of time as indicated by the
graphical LCD display.
14. Wait until the display shows a meaningful oxygen reading and begins to approach
the expected oxygen content of the sample gas.
Span Gas Preparation
The analyzer must be calibrated periodically. See the Calibration section below for recommendations.
Caution: Do not contaminate the span gas cylinder when connecting the regulator. Bleed the air filled regulator (faster and
more reliable than simply flowing the span gas) before attempting the initial calibration of the instrument.
Required components:
¾ Certified span gas cylinder with an oxygen concentration, balance nitrogen, approximating 80% of the full scale range
above the intended measuring range.
¾ Regulator to reduce pressure to 30 psig.
¾ Flow meter to set the flow between 1 SCFH,
¾ Suitable fittings and 1/8” dia. 4-6 ft. in length of metal tubing to connect the regulator to the flow meter inlet
¾ Suitable fitting and 1/8” dia. 4-6 ft. in length of metal tubing to connect from the flow meter vent to tube fitting designated
SAMPLE IN on the GPR-1600.
Procedure:
1. With the span gas cylinder valve closed, install the regulator on the cylinder.
2. Open the regulator’s exit valve and partially open the pressure regulator’s control knob.
3. Open slightly the cylinder valve.
4. Loosen the nut connecting the regulator to the cylinder and bleed the pressure regulator.
5. Retighten the nut connecting the regulator to the cylinder
6. Adjust the regulator exit valve and slowly bleed the pressure regulator.
7. Open the cylinder valve completely.
8. Set the pressure between 5-30 psig using the pressure regulator’s control knob.
9. Caution: Do not exceed the recommended flow rate. Excessive flow rate could cause the backpressure on the sensor
and may result in erroneous readings and permanent damage to the sensor.
22
Advanced Instruments Inc.
Establishing Power to the Electronics:
Once the power to the electronics is established, the digital display responds instantaneously. When power is applied, the
analyzer performs several diagnostic system status checks termed “SYSTEM SELF TEST” as illustrated below:
System Self Test
CPU
Memory
RTC
Analog
GPR Series Oxygen Analyzer
Software Version X.XX
Advanced Instruments
2855 Metropolitan Place
Pomona, CA 91767
Tel: 909-392-6900
Fax: 909-392-3665
e-mail: info@aii1.com
After 3 seconds the system defaults to the STANDBY mode and the LCD displays the following:
OK
OK
OK
OK
* MAIN MENU
Sample
Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
Standby
12/31/07 12:00:00
Menu Format
Menu displayed – displayed on the top line in the upper left corner of the display.
Menu options available - displayed in the upper left corner of the display under the current menu on the top line.
Menu option selected - indicated by the cursor (*) positioned to the left of the menu option selected.
System mode - indicated at the top center of the display.
Range mode and current auto or fixed manual range - displayed on the first line at the bottom of the display.
Temperature inside the analyzer and ambient pressure - displayed on the second line at the bottom of the display.
23
Advanced Instruments Inc.
Note:
Menu Navigation
The four (4) pushbuttons located on the front of the analyzer control the system’s micro-processor:
1. Green - ENTER (select)
2. Yellow UP ARROW – advance cursor up
3. Yellow DOWN ARROW – advance cursor down
4. Red – ESC (menu)
Select menu option by advancing cursor (*) by repeatedly pressing the yellow UP/DOWN ARROW keys.
Accept the menu option selected with cursor (*) by pressing the green ENTER key.
Abort the menu option selected with cursor (*) and return to the previous menu by pressing the red ESC key.
Note: If a selection is not made within 30 seconds, the system returns to the MAIN MENU.
In the event power to the analyzer is interrupted, the system defaults to the “Standby” mode when power is
restored. To resume sampling, advance the cursor (*) to “Sample” mode, press ENTER to select and select the
range mode as described below.
24
Advanced Instruments Inc.
Range Mode Selection
Advance the cursor (*) to the “Sample” option as illustrated and press the green ENTER key to accept the selection.
MAIN MENU
* Sample
Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
* SAMPLE
Auto Range
Manual Range
Bypass
Standby
Auto Range
85⁰F 100Kpa
The analyzer is equipped with five (5) standard measuring ranges (see specification) and provides users with a choice of
sampling modes. By accessing the MAIN MENU, users may select either the Auto Range or a fixed Manual Range) mode.
Note: For calibration purposes, use of the Auto Range mode is recommended. However, the user can select a fixed Manual
Range (full scale) as dictated by the oxygen content of the calibration gas.
Standby
12/31/07 12:00:00
Standby
12/31/07 12:00:00
25
Advanced Instruments Inc.
Auto Range Sampling
The display will shift to the next higher range when the oxygen reading exceeds 99.9% of the current range. The display will
shift to the next lower range when the oxygen reading drops to 85% of the next lower range.
For example, if the analyzer is reading 1 ppm on the 0-10 ppm range and an upset occurs, the display will shift to the 0-100
ppm range when the oxygen reading exceeds 9.99 ppm. Conversely, once the upset condition is corrected, the display will shift
back to the 0-10 ppm range when the oxygen reading drops to 8.5 ppm.
Procedure: From the SAMPLE menu, advance the cursor (*) to the “Auto Range” option and press ENTER:
SAMPLE
* Auto Range
Manual Range
Bypass
Standby
Auto Range
85⁰F 100Kpa
Note: To provide for the possibility of an optional automated sample system, the system displays the message “Opening
Sample Valves” at this time. This message does not apply to analyzers equipped with the standard manually operated
sample system.
Similarly, the Bypass and Standby options do not apply to analyzers equipped with manual sample systems.
Within seconds the system assesses the oxygen concentration, selects the appropriate range (as described above) and returns
to the MAIN MENU in the “Sample” mode. On the top line at the bottom of the menu, the Auto Range mode is indicated as is
the current full scale range.
* MAIN MENU
Sample
Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
Standby
12/31/07 12:00:00
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
26
Advanced Instruments Inc.
Manual Range Sampling
The display will not shift automatically. Instead, when the oxygen reading exceeds 125% of the upper limit of the current range
an OVER RANGE warning will be displayed. Once the OVER RANGE warning appears the user must advance the analyzer to the
next higher range.
Procedure: From the SAMPLE menu, advance the cursor (*) to the “Manual Range” option and press ENTER:
SAMPLE
Auto Range
* Manual Range
Bypass
Standby
Auto Range
85⁰F 100Kpa
The following display appears:
MANUAL RANGE
0 to 25%
0 to 1%
0 to 1000 PPM
0 to 100 PPM
* 0 to 10 PPM
Auto Range
85⁰F 100Kpa
Advance the cursor (*) to the desired fixed manual range, e.g. 0 to 10 ppm and press ENTER.
Within seconds the system assesses the oxygen concentration and returns to the MAIN MENU in the “Sample” mode. On the top
line at the bottom of the menu, the Manual Range mode is indicated as is the fixed full scale range selected.
Standby
12/31/07 12:00:00
Standby
12/31/07 12:00:00
* MAIN MENU
Sample
Span
Zero
Alarm
System
Standby
Manual Range
85⁰F 100Kpa
Sample
5.00PPM
0 to 10 PPM
12/31/07 12:00:00
27
Advanced Instruments Inc.
If the oxygen reading exceeds 99.9% of the full scale fixed range manually selected, the system displays the following:
* MAIN MENU
Sample
Span
Zero
Alarm
System
Standby
Manual Range
85⁰F 100Kpa
Sample
5.00PPM
OVER RANGE
0 to 10 PPM
12/31/07 12:00:00
Alarms
The analyzer is configured with two user adjustable threshold type alarm relays that can be configured in the field from the
ALARM option on the MAIN MENU as follows:
¾ Establish independent set points
¾ Either Hi or Lo
¾ Either On or Off (enabled or disabled)
¾ Both temporarily defeated using a user entered ‘timeout’ period (normally minutes)
The alarm set point represents a value. When the oxygen reading exceeds (high alarm) or falls below (low alarm) the alarm set
point, the relay is activated and the LCD displays the alarm condition.
When activated the alarms trigger SPDT Form C non-latching relays @ 5A, 30VDC or 240VAC resistive. To prevent chattering of
the relays, a 2% hysteresis is added to the alarm set point. This means that the alarm will remain active until the oxygen
reading has fallen 2% below the alarm set point (high alarm) or risen 2% above the alarm set point (low alarm) after the alarm
was activated. The timeout feature is useful while replacing the oxygen sensor or during calibration when the oxygen reading
might well rise above the alarm set point and trigger a false alarm.
Note: When making connections the user must decide whether to configure/connect Alarm 1 and Alarm 2 in failsafe mode
(Normally Open – NO – where the alarm relay de-energizes and closes in an alarm condition) or non-failsafe mode (Normally
Closed – NC – where alarm relay energizes and opens in an alarm condition).
Procedure: Advance the cursor (*) to the “Alarm” option and press the green ENTER key to accept the selection.
MAIN MENU
Sample
Span
Zero
* Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
28
Advanced Instruments Inc.
ALARM Sample
* Set Alarm 1
Set Alarm 2
Alarm 1 HI
Alarm 2 HI
Alarm 1 ON
Alarm 2 ON
Alarm Timeout
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the “Set Alarm 1” option and press the green ENTER key to accept the selection.
The following menu appears:
Sample
20%
Press UP or DOWN
to change value
ENTER to Save
ESC to Return
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Follow the prompt above and press the ENTER key to save the alarm value (% full scale) or ESC to return to the MAIN MENU.
Within a few seconds after pressing the ENTER key, the system returns to the MAIN MENU.
Repeat the above steps for “Set Alarm 2”.
Configure Alarm 1 and Alarm 2 by advancing the cursor (*) to the desired feature as illustrated below.
ALARM Sample
Set Alarm 1
Set Alarm 2
* Alarm 1 HI / LO
* Alarm 2 HI / LO
* Alarm 1 ON / OFF
* Alarm 2 ON / OFF
Alarm Timeout
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Press the ENTER key to toggle between the settings: HI and LO and/or ON and OFF.
After toggling, the system returns to the MAIN MENU. To confirm selection, re-access the ALARM menu above.
Advance the cursor (*) to the “Alarm” option and press the green ENTER key to accept the selection.
29
Advanced Instruments Inc.
MAIN MENU
Sample
Span
Zero
* Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
ALARM Sample
Set Alarm 1
Set Alarm 2
Alarm 1 HI
Alarm 2 HI
Alarm 1 ON
Alarm 2 ON
* Alarm Timeout
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the “Alarm Timeout” option and press the green ENTER key to accept the selection.
The following menu appears:
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
Sample
20 Minutes
Press UP or DOWN
to change value
ENTER to Save
ESC to Return
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Follow the prompt above and press the ENTER key to save the alarm timeout value or ESC to return to the MAIN MENU.
Within a few seconds after pressing the ENTER key, the system returns to the MAIN MENU.
30
Advanced Instruments Inc.
System Menu
The analyzer is equipped with a wide range of features that enables users to enhance performance and tailor their interface
with the analyzer. The SYSTEM menu shown below lists the features available and is followed by a description of each function.
Most of the functions are initiated by toggling between options using the ENTER key as previously described.
Advance the cursor (*) to the “Alarm” option and press the green ENTER key to accept the selection.
MAIN MENU
Sample
Span
Zero
Alarm
* System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
* SYSTEM
Enable Low Flow Alarm
Disable Alarm During Cal
Signal Average
Range
Logging Interval
Temp Coefficient
View Data Graph
Set Clock (and Date)
Logging ON
Show Text
Display Negative (Reading) ON
Advance the cursor (*) to the desired option and press follow the instructions below.
Enable Low Flow Alarm If the analyzer is equipped with a low flow alarm, press ENTER key to toggle between
Disable Alarm During Cal Press ENTER key to toggle between ENABLE and DISABLE.
Signal Average Press ENTER key to select and choose Low, Medium (default) or High – functions allows
Range Same as Manual Range option found on SAMPLE menu.
Logging Interval Press ENTER key and a display appears similar to Alarm Timeout above for the user to
Temp Coefficient Not operable at this time.
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
ENABLE and DISABLE.
users to select their preference regarding the trade-off of response time vs. noise
filtering.
enter the interval in minutes for capturing data points for logging purposes.
31
Advanced Instruments Inc.
View Data Graph Provided “Logging” feature below is toggled ON, selecting this feature provides a
fullscreen display or graph of the data points.
Set Clock (and Date) Selecting this option generates a display for selecting Time or Date with each followed
by a detailed display for setting hour, minute, second or year, month, day.
Logging Press ENTER key to toggle between ON and OFF.
Show Text Press ENTER key to toggle between MAIN MENU display options:
1.) Menu text with large numbers (as illustrated herein)
2.) Menu text with small numbers and a small graph of O2 reading.
Display Negative (Reading) Press the ENTER key to toggle between ON and OFF.
Installation & Start-up is now complete . . .
32
Advanced Instruments Inc.
Calibration
The electrochemical oxygen sensors manufactured by Analytical Industries Inc. (dba Advanced Instruments) generate an
electrical current that is linear or proportional to the oxygen concentration in the sample gas. In the absence of oxygen the
sensor exhibits an absolute zero, e.g. the sensor does not generate a current output in the absence of oxygen. Given the
properties of linearity and an absolute zero, single point calibration is possible.
As described below, zero calibration is recommended only when the application (or user) demands optimum accuracy for
analysis below 5% of the most sensitive or lowest range available on the analyzer. Span calibration in one of the forms
described below is sufficient for all other measurements. When employed zero calibration should precede span calibration.
Zero Calibration
Despite the absolute zero inherent in electrochemical oxygen sensors, the reality is that analyzers display an oxygen reading
when sampling a zero gas due to:
¾ Contamination or quality of the zero gas
¾ Minor leakage in the sample line connections
¾ Residual oxygen dissolved in the sensor’s electrolyte
¾ Tolerances of the electronic components
The zero capability (low end sensitivity) of every analyzer is qualified prior to shipment. However, because the factory sample
system conditions differ from that of the user, no ZERO OFFSET adjustment is made to the analyzer by the factory
Span Calibration
Involves periodically, see Intervals section below, checking and/or adjusting the electronics to the sensor’s signal output at a
given oxygen standard or span. Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The
frequency of calibration varies with the application conditions (potential for contamination), the degree of accuracy required by
the application and the quality requirements of the user. However, the interval between span calibrations should not exceed
three (3) months.
Note: Regardless of the oxygen concentration of the standard used, the span calibration process takes approximately 10
minutes, however, the time required to bring a ppm analyzer back on-line can vary, see Online Recovery Time below.
Considerations
When it comes to the calibration of oxygen analyzers utilizing an electrochemical oxygen sensor circumstances vary widely from
the ideal conditions that exist at the factory to a variety of differing circumstances users encounter in the field. The following
describes the most common factors and reasons that they influence the calibration procedures.
Factor Reasons
Intervals:
Conditions:
Analysis Level Required:
All electrochemical sensor based analyzers require periodic, e.g. weekly intervals to a 3
month maximum, calibration to ensure accuracy and ascertain whether the sensor has been
contaminated or otherwise damaged while in service.
Calibrate at the temperature and pressure of the sample.
Continuous analysis below 5% of the most sensitive or lowest range available:
ZERO CALIBRATION followed by SPAN CALIBRATION with good quality gases is
recommended (for optimum accuracy) when:
- the analyzer and/or O2 sensor is initially installed,
- the sample system connections are modified,
- the O2 sensor is replaced.
Note: It is not necessary to repeat the ZERO CALIBRATION with subsequent periodic
SPAN CALIBRATION unless desired or one of the above events occurs.
All other analysis: SPAN CALIBRATION is sufficient. Procedure varies with factors.
33
Zero Calibration Offset
Adjustment Capability:
Type of Analyzer:
Online Recovery Time:
Advanced Instruments Inc.
Designed to facilitate precise analysis below 5% of the most sensitive or lowest range
available on the analyzer, this feature enables users’ to compensate for background offsets,
as defined above, of up to 50% of the most sensitive or lowest full scale range available on
the analyzer and bring analyzers online faster.
As described below, accomplishing either objective places a degree of responsibility on the
user. Determining the true offset requires the user to wait, see Online Recovery Time
section, until the analyzer reading is no longer trending downward (best evidenced by a
constant horizontal trend on an external recording device). Bringing the analyzer online
faster, basically the same as choosing not to wait for the stable horizontal trend reading,
requires the user to repeat the ZERO CALIBRATION function. The frequency of which is at
the user’s discretion, hourly is recommended but at least when the reading goes negative.
Advanced Instruments’ oxygen analyzers are capable of zero offset adjustments of 50% of
the most sensitive or lowest range available on the analyzer. Since every analyzer is QC
tested to 1% of the most sensitive or lowest range available, exercise CAUTION when
initiating the ZERO CALIBRATION function at 50% (prematurely) of the most sensitive or
lowest range available on the analyzer. If the anticipated O2 analysis level is less than the
offset value or if adequate time is not allowed for the analyzer to establish the true offset,
the analyzer will in all probability display a negative reading.
Note: From the SYSTEM menu option “Display Negative (Reading)” users can toggle
between ON and OFF choose by pressing the ENTER key and control whether
analyzer displays negative readings.
Online ppb or ppm process analyzers: Analysis below 5% of the most sensitive or
lowest range is normally limited to these analyzers. However, such analysis is possible with
portable analyzers from Advanced Instruments due to their 60 day duty cycle and/or ability
to operate during battery charging.
Portable analyzers: Typically used intermittently moving between different sample
points/systems for trend analysis above 5% of the most sensitive range and, therefore, they
fall into the “all other analysis” category requiring only span calibration.
Percentage analyzers: Generally used above 5% of the most sensitive range and,
therefore, fall into the “all other analysis” category requiring only span calibration.
The priority users place on getting or keeping an analyzer online is “the” most significant
factor involved in calibration and troubleshooting issues. The time it takes an analyzer to
come down to a specific level after exposure to high O2 concentrations or air is significantly
different if a sensor is being installed than if the sensor had been in-service at low ppm
levels for more than 1 week as illustrated below:
Sensor Calibration at Install In-service Calibration
ppm Fuel Cell Air to 100 ppm < 10 min
Air to 10 ppm < 1 hr
Air to 1 ppm < 6 hrs
Air to .1 ppm < 16 hrs
The above times assume the introduction of a purge gas, the lower of the available zero or
sample (if known and constant) gas, with a ppm level O2 concentration less than 0.5 ppm is
introduced to the analyzer following span calibration to purge (accelerate the reaction of)
the O2 that has dissolved into the electrolyte inside the sensor. If zero gas is not available,
substitute the sample gas and expect slightly longer recovery time.
34
Air to 1 ppm < 30 min
80 ppm to .1 ppm < 10 min
8 ppm to .05 ppm < 5 min
Span Gas Selection:
Type of Sensor:
Span Gas vs. Air
(Fuel Cell Sensors only):
Span Calibration
Adjustments:
Advanced Instruments Inc.
The O2 concentration of a span gas should approximate 70-90% of the full scale range
dictated by the span gas, e.g. 80 ppm O2 on the 0-100 ppm range. For optimum accuracy,
the full scale range dictated by the span gas should be at least one range higher than the
intended analysis range. Both of the aforementioned recommendations reduce the error
induced by the tolerance of the electronic components; also, span gases with higher O2
concentrations are more reliable and less expensive.
Conversely, if the recommended span gas is not available and air calibration is not an
option, a span gas of the same full scale range and near the anticipated analysis level
(approximately 10% of full scale) is acceptable with the understanding the accuracy will
suffer slightly.
Use of span gas near 10% of the same full scale range for measurements at the higher end
of the range has the effect of “expanding the error” by moving upscale as illustrated by
Graph A and Example 1 in the Accuracy section above and is not recommended. Of course
the user can always elect at his discretion to accept an accuracy error of +
range if no other span gas is available.
Galvanic Fuel Cell Sensors: Analyzers utilizing these sensors can be calibrated with
ambient air (20.9% or 209,000 ppm O2) and have been proven capable of precise analysis
below 0.5 ppm (500 ppb).
Note: As described above, these oxygen sensors are capable of one point calibration.
Span gas calibration: Recommended for calibration of in-service analyzers, if the priority
is the fastest online recovery time.
Air calibration: Recommended for calibration an analyzer if a new oxygen sensor is being
installed or if the availability of span gas, the cost of span gas or the accuracy of a span gas
is an issue. An air calibrated analyzer can be used to reliably verify a “certified” span gas,
which has frequently been found to be inaccurate. For best results, select a recognized
name supplier.
Note: Galvanic Fuel Cell sensors can be calibrated (using air) and brought online (using
sample gas) without “purchased calibration gases” and without sacrificing accuracy provided the analyzer is given enough time to come down to the O2 concentration of the
sample gas.
Prematurely initiating the SPAN CALIBRATION function (when there is no intention of
performing a ZERO CALIBRATION) before the analyzer reading has stabilized can result in
erroneous readings as follows:
When purging an analyzer to lower ranges for span gas calibration: If the oxygen
reading reaches less than 2% of the intended calibration range, enter the value of the span
gas. If the oxygen reading is greater than 2% of the calibration range, add the O2 reading
to the value of the span gas (the impact of the offset on accuracy is minor but the addition
allows the oxygen sensor to continue to purge down and avoid negative readings).
Note: If ZERO CALIBRATION has been performed or the analyzer has been in service, the
analyzer reading should already be stable and below 2% of the calibration range.
When installing a new oxygen sensor and calibrating with air: Allow 5-10 minutes
for the sensor to equilibrate in ambient air from storage packaging. Failure to do so can
introduce a positive offset (electronic gain) that prevents the analyzer from displaying low
ppm O2 readings.
35
2-3% of full scale
Advanced Instruments Inc.
Menu Functions – Zero Calibration
Factory Default Zero: The system eliminates any previous zero calibration offset adjustment stored in
memory and displays the unadjusted oxygen reading of the gas currently flowing
through the analyzer.
This function is recommended before performing a manual ZERO CALIBRATION or
when troubleshooting an analyzer. This function is not recommended for subsequent
periodic SPAN CALIBRATION - see Analysis Level Required above.
Auto Zero: Applicable only to analyzers equipped with an optional automated sample system. The
system automatically closes the sample inlet valve and opens the zero inlet valve at a
predetermined interval in time, see System Menu section above. It also trends the
downward slope of the oxygen reading. When the slope of the oxygen reading trend
approaches the factory set value, the system initiates a ZERO CALIBRATION offset
adjustment and displays an oxygen reading of 0.000 or 00.00 ppm.
Manual Zero: Recommended for optimum accuracy. The user must ascertain that the oxygen reading
has reached a stable value, see Zero Calibration Offset Adjustment Capability above,
below 50% of the most sensitive or lowest range available on the analyzer before the
system will accept and make a ZERO CALIBRATION offset adjustment.
If the user attempts to initiate the ZERO CALIBRATION function while the oxygen
reading is above 50% of the most sensitive or lowest range, the system displays the
message “CALIBRATION FAILED” and returns to the “Sample” mode.
Timed Zero Cal: Subtitle only, this function represented by the following menu options are applicable
only to analyzers equipped with an optional automated sample system.
Timed Zero Cal in ___ days:
Calibration will occur at ____: Requires the user to enter a “military time in hours and minutes” representing the time
Menu Functions – Span Calibration
Factory Default Span: The system eliminates any previous span calibration adjustment stored in memory and
Span Gas Units/Value:
Requires the user to enter a “number” representing the interval in days before the next
ZERO CALIBRATION.
of day at which the next ZERO CALIBRATION will be initiated.
displays an oxygen reading +
analyzer.
If the oxygen reading is outside +
message “CALIBRATION FAILED” and returns to the “Sample” mode. This feature
allows the user to test the sensor’s signal output without removing it from the sensor
housing.
This function is recommended before performing a SPAN CALIBRATION or when
troubleshooting an analyzer.
After initiating either Auto or Manual Span from the SPAN CALIBRATION menu, the
system produces a display prompting the user to select span gas in ppb, ppm or %
units, which is followed by a second display prompting the user to enter a numerical
span gas value.
50% of the span gas value currently flowing through the
50% of the span gas value, the system displays the
36
Advanced Instruments Inc.
Auto Span: Applicable only to analyzers equipped with an optional automated sample system. The
system automatically closes the sample inlet valve and opens the span inlet valve at a
predetermined interval in time, see System Menu section above. It also trends the
downward slope of the oxygen reading vs. time curve. When the slope of the oxygen
reading reaches “0”, the system initiates a SPAN CALIBRATION adjustment and
displays an oxygen reading corresponding to the span gas value entered.
Manual Span: The user must ascertain that the oxygen reading has reached a stable value, see Span
Calibration Adjustment above, before the system will accept and make a SPAN
CALIBRATION adjustment.
Timed Span Cal: Subtitle only, this function represented by the following menu options are applicable
only to analyzers equipped with an optional automated sample system.
Timed Span Cal in ___ days: Requires the user to enter a “number” representing the interval in days before the next
SPAN CALIBRATION.
Calibration will occur at ____: Requires the user to enter a “military time in hours and minutes” representing the time
of day at which the next SPAN CALIBRATION will be initiated.
Timed Span Gas Value:
Procedure – Zero Calibration
The analyzer is online in the Auto Range mode as described above.
Review the Analysis Level Required, Online Recovery Time, Zero Calibration Offset Adjustment and Menu Functions – Zero
Calibration above before proceeding.
Assure there are no restrictions in vent line.
Regulate the pressure between 5-30 psig and set the flow rate to 2 SCFH for Fuel Cell sensors.
Allow the analyzer reading to stabilize below 50% of the most sensitive or lowest range available on the analyzer.
Advance the cursor (*) to the “Zero” option as illustrated and press the green ENTER key to accept the selection.
MAIN MENU
Sample
Span
* Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
Displays the numerical span gas value and units entered when the Auto Span option
was initiated.
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
37
Advanced Instruments Inc.
ZERO Sample
* Factory Default
Calibrate
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the Factory Default Zero option and press ENTER.
Within seconds the system returns to the MAIN MENU in the “Sample” mode.
MAIN MENU
Sample
Span
* Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
Advance the cursor (*) to the “Zero” option and press the green ENTER key to accept the selection.
The following menu appears:
ZERO Sample
Factory Default
* Calibrate
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the “Manual Zero” option and press the green ENTER key to accept the selection.
The following menu appears:
Sample
5.00 PPM
0 to 10 PPM
12/31/07 12:00:00
38
Advanced Instruments Inc.
ZERO CALIBRATION
IN PROGRESS . . .
5.00 PPM
After the oxygen reading has stabilized, press ENTER to complete the Zero Calibration. If the user attempts to initiate the ZERO
CALIBRATION function while the oxygen reading is above 50% of the most sensitive or lowest range, the system displays the
message “CALIBRATION FAILED” and returns to the “Sample” mode.
CALIBRATION
FAILED
CALIBRATION
COMPLETE
39
Advanced Instruments Inc.
Procedure – Span Calibration
The analyzer is online in the Auto Range mode as described above.
Review the Intervals, Online Recovery Time, Span Gas Selection, Type of Sensor, Span Calibration Adjustment and Menu
sections above, before proceeding.
Allow the analyzer reading to stabilize, Advance the cursor (*) to the “Span” option as illustrated and press the green ENTER
key to accept the selection.
MAIN MENU
Sample
* Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
SPAN Sample
* Factory Default
Calibrate
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the Factory Default Span option and press ENTER.
Within seconds the system returns to the MAIN MENU in the “Sample” mode.
Sample
1.00 PPM
0 to 10 PPM
12/31/07 12:00:00
MAIN MENU
Sample
* Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
At this point, the user must decide whether to perform a Span Gas or Air Calibration, see above.
Note: If the analyzer is to calibrated with instrument or compressed air follow the Span Gas Calibration procedure below.
Sample
1.00 PPM
0 to 10 PPM
12/31/07 12:00:00
40
Advanced Instruments Inc.
Span Gas Calibration: Assure there are no restrictions in vent line.
Regulate the pressure between 5-30 psig and set the flow rate to 2 SCFH for Fuel Cell sensors.
If the analyzer is equipped with a SAMPLE/BYPASS valve, place it in the BYPASS position.
Disconnect the sample gas line and install the span gas line.
Keep the SAMPLE/BYPASS valve in the BYPASS position.
Allow the span gas to flow for 1-2 minutes to purge the gas lines inside the analyzer.
Place the SAMPLE/BYPASS valve in the SAMPLE position.
If the analyzer is not equipped with a SAMPLE/BYPASS valve:
Allow the span gas to flow for 1-2 minutes to purge the air trapped in the span gas line.
Disconnect the sample gas line and install the purged span gas line.
Air Calibration: Access the interior of the analyzer by opening the
front door of the analyzer.
Using the 5/16 wrench supplied loosen but do not
remove the clamp bolt located in the center of
the bracket attached to bottom section with the
elbow fittings.
Rotate the upper section of the sensor housing
90º to disengage from the clamp.
Remove the upper section by pulling it straight up and let it rest on your 1
With your other hand remove the oxygen sensor, place it in the upper section of the sensor housing
ensuring the PCB contacts the two gold pins and use your thumb to hold the sensor and upper
section of the sensor housing together.
The sensor is now exposed to ambient air, connected to the analyzer electronics and ready for
calibration.
Allow the sample gas to continue to flow if possible, otherwise close the FLOW valve but do not
disconnect the sample gas line.
Caution: Allow the reading to stabilize before proceeding. A premature SPAN CALIBRATION adjustment can result in erroneous
readings, see the Span Calibration Adjustment section above.
It is highly recommended that the analyzer be connected to an external recording device, if practical, to ensure that the
analyzer reading reaches a stable value (10-20 minutes) before accepting the span calibration.
Advance the cursor (*) to the “Span” option as illustrated and press the green ENTER key to accept the selection.
st
and 2nd fingers.
MAIN MENU
Sample
* Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
Sample
80.0 PPM
0 to 10 PPM
12/31/07 12:00:00
41
Advanced Instruments Inc.
MAIN MENU
Sample
* Span
Zero
Alarm
System
Standby
Auto Range
85⁰F 100Kpa
The following menu appears:
SPAN Sample
Factory Default
* Calibrate
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the “Manual Span” option and press the green ENTER key to accept the selection.
The following menu appears:
Sample
20.9 %
0 to 10 PPM
12/31/07 12:00:00
SPAN GAS Standby
* Enter as PPM
Enter as PPB
Enter as %
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Enter span gas units appropriate to the numerical span gas value and press ENTER key.
Note: Select “Enter as %” for air calibration.
42
Advanced Instruments Inc.
The following display appears:
Sample
0.00 PPM
08
Press UP or DOWN keys to change values
Select ENTER to save, ESC to return
Auto Range 0 to 10 PPM
85⁰F 100Kpa 12/31/07 12:00:00
Press the ENTER key to move the cursor (underscore) to the right to the digit to be changed.
Press the UP or DOWN key until the desired number appears in digit field.
Repeat as necessary to enter the numerical span value.
Press the ENTER key to accept and save the span gas value.
The system then initiates the SPAN CALIBRATION function and the following menu appears:
CALIBRATION
IN PROGRESS . . .
SPAN VALUE ACTUAL O2 VALUE
After the oxygen reading has stabilized, press ENTER to complete the Span Calibration.
If the oxygen reading is outside +
returns to the “Sample” mode.
50% of the span gas value, the system displays the message “CALIBRATION FAILED” and
CALIBRATION
FAILED
43
Advanced Instruments Inc.
CALIBRATION
COMPLETE
Following a successful calibration, the system returns to the “Sample” mode and it will be necessary to reconnect the sample
gas line as follows:
Span Gas Calibration: Assure there are no restrictions in vent line.
Regulate the pressure between 5-30 psig and set the flow rate to 2 SCFH for Fuel Cell sensors.
If the analyzer is equipped with a SAMPLE/BYPASS valve, place it in the BYPASS position.
Disconnect the span gas line and install the sample gas line.
Keep the SAMPLE/BYPASS valve in the BYPASS position.
Allow the sample gas to flow for 1-2 minutes to purge the gas lines inside the analyzer.
Place the SAMPLE/BYPASS valve in the SAMPLE position.
If the analyzer is not equipped with a SAMPLE/BYPASS valve:
Allow the sample gas to flow for 1-2 minutes to purge the air trapped in the sample gas lines.
Disconnect the span gas line and install the purged sample gas line.
Air Calibration:
The sample gas should still be flowing, if necessary open the FLOW valve.
Reassemble the sensor housing.
Place the sensor face down in the bottom section of the sensor housing.
Ensure the o-ring is in place.
Install the upper section by gently pushing it
straight down.
Ensure the PCB contacts the two gold pins.
Rotate the upper section of the sensor housing
90º to engage the clamp.
Using the 5/16 wrench supplied tighten the clamp
bolt located in the center of the bracket attached
to bottom section with the elbow fittings to
secure the sensor housing.
Close the front door of the analyzer.
44
Advanced Instruments Inc.
Sampling
Process ppm oxygen analyzers require positive pressure to flow the sample gas by the sensor to measure the oxygen
concentration in a sample gas. See Pressure & Flow under Installation in section 3 Safety Guidelines. To assure optimal
performance: connect gas lines with metal tubing, quality compression type fittings to minimize leaks, follow pressure and flow
recommendations and avoid exposing the sensor to air and high oxygen concentrations for prolonged periods of time (this does
not include the 5 minutes it should take to air calibrate the analyzer once a week).
The sample must be properly presented to the sensor to ensure an accurate measurement. In standard form the GPR-1600 is
designed with a sample system that complements the performance capabilities of the advanced oxygen sensor and enables the
user to isolate the sensor from exposure to high oxygen concentration which results is a substantial increase is user
productivity.
For ppb and ppm trace oxygen measurements, the sensor is exposed to sample gas that must flow or be drawn through the
analyzer’s internal sample system. This unique sample system, when operated accordingly to the instructions in this Owner’s
Manual, can significantly increase user productivity by minimizing the sensor’s exposure to ambient air or high oxygen
concentrations which contribute to the significant amount of downtime associated with competitive analyzers.
The advantages of the bypass sample system include:
¾ Supplying the analyzer with the sensor it was qualified with.
¾ Isolating the sensor during transport, calibration and maintenance intervals when changing gas line connections.
¾ Isolating the sensor from exposure to high oxygen levels during upset conditions which extend sensor life.
¾ Purging the air (or high oxygen levels above 1,000 ppm) trapped in the gas lines following a process upset.
45
Advanced Instruments Inc.
Procedure:
Following calibration the analyzer returns to the SAMPLE mode after 30 seconds.
1. Select the desired sampling mode - auto or if manual, the range that provides maximum resolution – as described above.
2. Use metal tubing to transport the sample gas to the analyzer.
3. The main consideration is to eliminate air leaks which can affect oxygen measurements above or below the 20.9% oxygen
concentration in ambient air - ensure the sample gas tubing connections fit tightly into the 1/8” male NPT to tube adapter,
and, the NPT end is taped and securely tightened into the mating male quick disconnect fittings which mate with the
female fittings on the analyzer
4. Assure there are no restrictions in the sample line.
5. For sample gases under positive pressure the user must provide a means of regulating the inlet pressure between 5-30 psig
and setting the flow rate of the sample gas at 2 SCFH (flow rates between .5 and 5 SCFH are acceptable).
6. For sample gases under atmospheric or slightly negative pressure an optional sampling pump is recommended to draw the
sample into the analyzer. Generally, no pressure regulation or flow control device is involved.
7. Caution: If the analyzer is equipped with an optional sampling pump and is intended for use in both positive and
atmospheric/slightly negative pressure applications where a flow meter valve is involved – ensure the valve is completely
open when operating the sampling pump. Refer to the Pressure & Flow section above.
8. Assure the sample is adequately vented for optimum response and recovery – and safety.
9. Allow the oxygen reading to stabilize for approximately 10 minutes at each sample point.
To avoid erroneous oxygen readings and damaging the sensor:
¾ Do not place your finger over the vent to test the flow indicator when gas is flowing to the sensor. Removing your finger
generates a vacuum on the sensor and may damage the sensor (voiding the sensor warranty).
¾ Assure there are no restrictions in the sample or vent lines
¾ Avoid drawing a vacuum that exceeds 14” of water column pressure – unless done gradually
¾ Avoid excessive flow rates above 5 SCFH.
¾ Avoid sudden releases of backpressure that can severely damage the sensor.
¾ Avoid the collection of particulates, liquids or condensation collect on the sensor that cou ld block the diffusion of oxygen
into the sensor.
¾ If the analyzer is equipped with an optional integral sampling pump (positioned downstream of the sensor) and a flow
control metering valve (positioned upstream of the sensor), completely open the flow control metering valve to avoid
drawing a vacuum on the sensor and placing an undue burden on the pump.
Standby
The analyzer has no special storage requirements. The sensor should remain connected during storage periods. Store the
analyzer with the power OFF. If storing for an extended period of time protect the analyzer, cable and sensor from dust, heat
and moisture.
46
Advanced Instruments Inc.
6 Maintenance
There are no moving parts in the analyzer given the modular nature of the electronics and sensor. Cleaning the electrical
contacts when replacing the sensor is the extent of the maintenance requirements of this analyzer. Serviceability: Except for
replacing the oxygen sensor, there are no parts inside the analyzer for the operator to service. Only trained personnel with the
authorization of their supervisor should conduct maintenance.
Sensor Replacement:
Periodically, the oxygen sensor will require replacement. The operating life is determined by a
number of factors that are influenced by the user and therefore difficult to predict. The sections
dealing with Specification and Installation Considerations define the normal operating conditions
and expected life of the standard sensor utilized by the GPR-1600 analyzer. As a general
guideline, expected sensor life is inversely proportional to changes in oxygen concentration,
pressure and temperature.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte
that could be harmful if touched or ingested, refer to the Material Safety Data Sheet contained
in the Owner’s Manual appendix. Avoid contact with any liquid or crystal type powder in or
around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors
should be disposed of in accordance with local regulations.
Procedure:
1. Determine your calibration requirements by reviewing the ZERO CALIBRATION and SPAN
CALIBRATION discussions in section 5 Operation. Consult the analyzer specifications for
recovery times and span gas values.
2. Open the door of the analyzer to access the sensor housing.
3. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located in the
center of the bracket attached to bottom section with the elbow fittings.
4. Rotate the upper section of the sensor housing 90º to disengage from the clamp.
5. Remove the upper section by pulling it straight up and place it on a smooth surface.
6. Remove the old oxygen sensor and dispose of it as you would a battery.
7. Remove the o-ring from the bottom section of the sensor housing.
8. Wipe the o-ring with a damp lint free cloth.
9. Lightly lubricate the o-ring with vacuum grease for optimal seal.
10. Reinstall the o-ring into the bottom section of the sensor housing.
11. From the MAIN MENU select AUTO RANGING as described above.
12. If equipped with SAMPLE/BYPASS valve, place it in the SAMPLE
position.
13. Set the flow rate to 2 SCFH.
14. Connect zero gas or low oxygen content sample gas line to purge the lines and the sensor of oxygen (once reinstalled).
15. Caution: Minimize the time the new sensor is exposed to ambient air.
16. Remove the new oxygen sensor from the shipping bag.
17. Remove the red label and the gold ribbon (shorting device) from the PCB at the rear of the sensor.
18. Place the new sensor in the bottom section of the sensor housing with the PCB facing up.
19. Place the upper section of the sensor housing over the sensor.
Gently push the upper section downward and rotate 90º to engage the
20.
21. Finger tighten the clamp bolt and one full turn with the 5/16 wrench to compressed the o-ring seal.
22. Expect the analyzer reading to recover to ppb levels as described in the analyzer specification.
23. Perform the desired calibration(s).
24. Begin sampling once the analyzer has reached the value of the purge gas.
clamp.
47
Advanced Instruments Inc.
7 Spare Parts
Recommended spare parts for the GPR-1600 Oxygen Analyzer include:
A-1004-1-36 Housing Sensor Stainless Steel
A-1016-A Housing Sensor Bottom Assembly Stainless Steel
B-2762-A-1-36 Housing Sensor Upper Assembly Stainless Steel
MTR-1011 Meter Digital Panel LCD Backlight
ORNG-1007 O-ring 3/32 x 1-3/8 x 1-9/16 Viton
A-1146-10 PCB Assembly Main / Display
A-1174-10 PCB Assembly AC Power Supply / Interconnection Alarms, 4-20mA Range ID
A-1174-10C PCB Assembly AC Power Supply / Interconnection w/o Alarms, Relay Contacts Range ID
48
Advanced Instruments Inc.
8 Troubleshooting
Symptom Possible Cause Recommended Action
Slow recovery
High O
after installing
or replacing sensor
High O
Sampling
Response time slow Air leak, dead legs, distance of sample line,
O
agree to expected O
values
reading
2
reading
2
reading doesn’t
2
At installation, defective sensor
Air leak in sample system connection(s)
Abnormality in zero gas
Damaged in service - prolonged exposure to
air, electrolyte leak
Sensor nearing end of life
Analyzer calibrated before sensor stabilized
caused by:
1) Prolonged exposure to ambient air, worse
if sensor was unshorted
low flow rate, volume of optional filters and
scrubbers
Pressure and temperature of the sample is
different than span gas
2
Abnormality in gas
49
Replace sensor if recovery unacceptable or O
reading fails to reach 10% of lowest range
Leak test the entire sample system: Vary the flow
rate, if the O
change in flow rate indicates an air leak - correct
source of leak
Qualify zero gas (using portable analyzer)
Replace sensor
Replace sensor
Allow O
span/calibration adjustment
Continue purge with zero gas
Leak test the entire sample system (above)
Qualify zero gas (using portable analyzer)
Correct pressure and flow rate
Remove restriction on vent line
Replace GPR/PSR sensor with XLT sensor when
or acid gases are present
CO
2
Leak test (above), reduce dead volume or increase
flow rate
Calibrate the analyzer (calibrate at pressure and
temperature of sample)
Qualify the gas (use a portable analyzer)
reading changes inversely with the
2
reading to stabilize before making the
2
2
Advanced Instruments Inc.
Symptom Possible Cause Recommended Action
Erratic O
reading
2
or
reading
No O
2
Erratic O
reading
2
or
Negative O
reading
2
or
reading
No O
2
accompanied by
electrolyte leakage
Change in sample pressure
Dirty electrical contacts in upper section of
sensor housing
Corroded solder joints on sensor PCB from
corrosive sample or electrolyte leakage from
sensor
Corroded spring loaded contact in upper
section of sensor housing from liquid in
sample or electrolyte leakage from sensor
Liquid covering sensing area
Improper sensor selection
Presence of interference gases
Unauthorized maintenance
Sensor nearing end of life
Pressurizing the sensor by flowing gas to the
sensor with the vent restricted or SHUT OFF
valve closed and suddenly removing the
restriction draws a vacuum on the sensor
or
partially opening the valves upstream of the
analyzer when using a pump downstream of
the analyzer to draw sample from a process at
atmospheric pressure or a slight vacuum.
Placing a vacuum on the sensor in excess 4”
of water column is strongly discouraged.
A premature adjustment of the ZERO OFFSET
potentiometer is a common problem
Sensors without PCB use mV setting.
Calibrate the analyzer (calibrate at pressure and
temperature of sample)
Clean contacts with alcohol (minimize exposure time
of MS sensor to ambient air to extent possible)
Replace sensor and return sensor to the factory for
warranty determination
Upper section of sensor housing: Clean contacts
with alcohol, flow sample or zero gas for 2-3 hours
to flush sample system and sensor housing
Sensor: Replace if leaking and return it to the
factory for warranty determination
Wipe with alcohol and lint free towel or flow sample
or zero gas for 2-3 hours to flush
Replace GPR/PSR sensor with XLT sensor when CO
or acid gases are present. Consult factory.
Replace sensor and install scrubber
Consult factory.
Replace sensor
Zero the analyzer. If not successful replace the
sensor
Avoid drawing a vacuum on the sensor, a
pressurized sensor may not leak but still produce
negative readings.
From MAIN MENU select DEFAULT ZERO
2
50
Advanced Instruments Inc.
51
Advanced Instruments Inc.
52
Advanced Instruments Inc.
9 Warranty
The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are performed under a certified
Quality Assurance System that conforms to established standards and incorporates state of the art materials and components
for superior performance and minimal cost of ownership. Prior to shipment every analyzer is thoroughly tested by the
manufacturer and documented in the form of a Quality Control Certification that is included in the Owner’s Manual
accompanying every analyzer. When operated and maintained in accordance with the Owner’s Manual, the units will provide
many years of reliable service.
Coverage
Under normal operating conditions, the monitor, analyzers and sensor are warranted to be free of defects in materials and
workmanship for the period specified in accordance with the most recent published specifications, said period begins with the
date of shipment by the manufacturer. The manufacturer information and serial number of this analyzer are located on the rear
of the analyzer. Advanced Instruments Inc. reserves the right in its sole discretion to invalidate this warranty if the serial
number does not appear on the analyzer.
If your Advanced Instruments Inc. monitor, analyzer and/or oxygen sensor is determined to be defective with respect to
material and/or workmanship, we will repair it or, at our option, replace it at no charge to you. If we choose to repair your
purchase, we may use new or reconditioned replacement parts. If we choose to replace your Advanced Instruments Inc.
analyzer, we may replace it with a new or reconditioned one of the same or upgraded design. This warranty applies to all
monitors, analyzers and sensors purchased worldwide. It is the only one we will give and it sets forth all our responsibilities.
There are no other express warranties. This warranty is limited to the first customer who submits a claim for a given serial
number and/or the above warranty period. Under no circumstances will the warranty extend to more than one customer or
beyond the warranty period.
Limitations
Advanced Instruments Inc. will not pay for: loss of time; inconvenience; loss of use of your Advanced Instruments Inc. analyzer
or property damage caused by your Advanced Instruments Inc. analyzer or its failure to work; any special, incidental or
consequential damages; or any damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized
repair or modification of the analyzer; affixing of any attachment not provided with the analyzer or other failure to follow the
Owner’s Manual. Some states and provinces do not allow limitations on how an implied warranty lasts or the exclusion of
incidental or consequential damages, these exclusions may not apply.
Exclusions
This warranty does not cover installation; defects resulting from accidents; damage while in transit to our service location;
damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized repair or modification of the
analyzer; affixing of any label or attachment not provided with the analyzer; fire, flood, or acts of God; or other failure to follow
the Owner’s Manual.
Service
Call Advanced Instruments Inc. at 909-392-6900 (or e-mail info@aii1.com) between 7:30 AM and 5:00 PM Pacific Time Monday
thru Thursday or before 12:00 pm on Friday. Trained technicians will assist you in diagnosing the problem and arrange to
supply you with the required parts. You may obtain warranty service by returning you analyzer, postage prepaid to:
Advanced Instruments Inc.
2855 Metropolitan Place
Pomona, Ca 91767 USA
Be sure to pack the analyzer securely. Include your name, address, telephone number, and a description of the operating
problem. After repairing or, at our option, replacing your Advanced Instruments Inc. analyzer, we will ship it to you at no cost
for parts and labor.
53
Advanced Instruments Inc.
10 MSDS Material Safety Data Sheet
Product Identification
Product Name Oxygen Sensor Series - PSR, GPR, AII, XLT
Synonyms Electrochemical Sensor, Galvanic Fuel Cell
Manufacturer Analytical Industries Inc., 2855 Metropolitan Place, Pomona, CA 91767 USA
Emergency Phone Number 909-392-6900
Preparation / Revision Date January 1, 1995
Notes Oxyg en sensors are sea led, conta in protective c overings and in normal conditions do n ot present a
Specific Generic Ingredients
Carcinogens at levels > 0.1% None
Others at levels > 1.0% Potassium Hydroxide or Acetic Acid, Lead
CAS Number Potassium Hydroxide = KOH 1310-58-3 or Acetic Acid = 64-19-7, Lead = Pb 7439-92-1
Chemical (Synonym) and Family Potassium Hydroxide (KOH) – Base or Acetic Acid (CH
General Requirements
Use Potassium Hydroxide or Acetic Acid - electrolyte, Lead - anode
Handling Rubber or latex gloves, safety glasses
Storage Indefinitely
Physical Properties
Boiling Point Range
Melting Point Range
Freezing Point
Molecular Weight KOH = 56 or Acetic Acid – NA, Lead = 207
Specific Gravity
Vapor Pressure
Vapor Density KOH – NA or Acetic Acid = 2.07
pH KOH > 14 or Acetic Acid = 2-3
Solubility in H
% Volatiles by Volume None
Evaporation Rate Similar to water
Appearance and Odor Aqueous solutions: KOH = Colorless, odorless or Acetic Acid = Colorless, vinegar-like odor
O Complete
2
Fire and Explosion Data
Flash and Fire Points Not applicable
Flammable Limits Not flammable
Extinguishing Method Not applicable
Special Fire Fighting Procedures Not applicable
Unusual Fire and Explosion Hazards Not applicable
Reactivity Data
Stability Stable
Conditions Contributing to Instability None
Incompatibility KOH = Avoid contact with strong acids or Acetic Acid = Avoid contact with strong bases
Hazardous Decomposition Products KOH = None or Acetic Acid = Emits toxic fumes when heated
Conditions to Avoid KOH = None or Acetic Acid = Heat
health hazard. Information applies to electrolyte unless otherwise noted.
H) – Acid, Lead (Pb) – Metal
3CO2
KOH = 100 to 115° C or Acetic Acid = 100 to 117° C
KOH -10 to 0° C or Acetic Acid – NA, Lead 327° C
KOH = -40 to -10° C or Acetic Acid = -40 to -10° C
KOH = 1.09 @ 20° C, Acetic Acid = 1.05 @ 20° C
KOH = NA or Acetic Acid = 11.4 @ 20° C
54
Advanced Instruments Inc.
Spill or Leak
Steps if material is released Sensor is packaged in a sealed plastic bag, check the sensor inside for electrolyte leakage. If the
Disposal
Health Hazard Information
Primary Route(s) of Entry Ingestion, eye and skin contact
Exposure Limits Potass ium Hydroxide - ACGIH TLV 2 mg/cubic m eter or Acetic Acid - ACGIH TLV / OSHA PEL 10
Ingestion Electrolyte could be harmful or fatal if swallowed. KOH = Oral LD50 (RAT) = 2433 mg/kg or Acetic
Eye Electrolyte is corrosive and eye contact could re sult in permanent loss of vision.
Skin Electrolyte is corrosive and skin contact could result in a chemical burn.
Inhalation Liquid inhalation is unlikely.
Symptoms Eye contact - burning sensation. S kin contact - soapy slick feeling.
Medical Conditions Aggravated None
Carcinogenic Reference Data KOH and Ace tic Acid = NTP Annual Report on Carcinogens - no t listed; LARC Monographs - not
Other Lead is listed as a chemical known to the State of California to cause birth defects or other
Special Protection Information
Ventilation Requirements None
Eye Safety glasses
Hand Rubber or latex gloves
Respirator Type Not applicable
Other Special Protection None
Special Precautions
Precautions Do not remove the sensor’s protective Teflon and PCB coverings. Do not probe the sensor with
Transportation Not applicable
sensor leaks inside the plastic bag or inside an analyzer sensor housing do not remove it without
rubber or latex gloves and safety glasses and a source of water. Flush or wipe all surfaces
repeatedly with water or wet paper towel (fresh each time).
In accordance with federal, state and local regulations.
ppm (TWA), Lead - OSHA PEL .05 mg/cubic meter
Acid = Oral LD50 (RAT) = 6620 mg/kg
listed; OSHA - not listed
reproductive harm.
sharp objects. Wash hands thoroughly after ha nd ling. Avoid contact with eyes, skin and clothing.
Empty sensor body may contain hazardous residue.
55
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.