Amprobe VPC-10 User Manual

Instruction Manual
m Safety Information
• Never use the tester if it or its test leads appear damaged
• Never apply more than 600V DC or 600V AC RMS between any terminal and earth ground.
( If the LED turns on it indicates the presence of voltage at the probe tip.
• Disconnect the live test lead before disconnecting the common test lead.
• Before each use, verify the tester’s operation by measuring a known voltage.
Voltage and Continuity Tester
Model VPC-10
Testing for Continuity
Turn off circuit power before testing Beeper indicates shorts lasting 1ms or longer
OPEN SHORT or 0-85 KΩ
220V AC
Automatic Operation
The tester automatically turns on when you place the probes across a complete circuit. The tester selects continuity or dc or ac voltage mode based on the resistance or volt­age between the probes. The tester automatically turns off when you remove the probes from the complete circuit.
Measuring Voltage
Connect test leads across the source or load under measurement, the LEDs turn on at between 70% and 100% of their rated voltages.
DC Voltage
600V DC maximum
600
AC Voltage
600V RMS maximum
45Hz to 66Hz
Continuity
Continuity
r Beeping
Specifications
Function Range Accuracy
AC voltage(V) 24,120,208,240,277,480,600 -30% to 0% of reading DC voltage(V) 6,12,24,36,48,110,220 -30% to 0% of reading
Continuity test: Audible threshold 0-85kΩ, Continuity beeper 2KHz Maximum Voltage Between any Terminal and Earth Ground:
600 V Overvoltage Category III
Input Impedance: 1MΩ Temperature: Operating: 5˚C (41˚F) to 40˚C (104˚F) Storage : -30°C to +60°C
Altitude: Operating: 2000m; Storage:10,000m
Relative Humidity:
80% for Temp up to 31˚C decreasing linearly to 50% RH at 40˚C
Battery Type and Life: AA(2); 350 hours with NEDA 15˚F or IEC R6 Safety: Complies with EN61010-1:1993 for use in overvoltage CAT III environments Pollution Degree: 2
For indoor use only EMC Regulations: EN50081-1, EN50082-1
Battery Test and Replacement
Replace the batteries when touching the leads together no longer turns on the continuity LED.
2xAA
Disconnect probes from Voltag. Do not over-tighten screws
WARRANTY
Congratulations! Your new instrument has been quality crafted according to quality standards and contains quality components and workmanship. It has been inspected for proper operation of all of its functions and tested by qualified factory technicians according to the long-established standards of our company. Your instrument has a limited warranty against defective materials and/or workmanship for one-year from the date of purchase provided that, in the opinion of the factory, the instrument has not been tampered with or taken apart. Should your instrument fail due to defective materials, and/or workmanship during this one-year period, a no charge repair or replacement will be made to the original pur­chaser. Please have your dated bill of sale, which must identify the instrument model number and serial number and call the number listed below:
Repair Department ATP – Amprobe, TIF, Promax Miramar, FL Phone: 954-499-5400 800-327-5060 Fax : 954-499-5454 Website: www.amprobe.com
Please obtain an RMA number before returning product for repair. Outside the U.S.A. the local representative will assist you. Above limited warranty covers repair and replacement of instrument only and no other obligation is stated or implied.
Miramara, Florida 33025
P/N: ATP VPC-10Manual printed in China