Data Sheet
TMD-10
Dual Temperature Meter
The Amprobe TMD-10 is a completely portable dual input
thermometer. With a digital dual display that displays T1, T2
and T1-T2. Choose between Celsius or Fahrenheit degrees
measurement results. The TMD-10 will store and later display
your minimum and maximum values. Will measure temperature with either Type K or Type J thermocouples. Ships as a
complete kit with carrying case, two Type K probes, battery
and manual.
Compatible with K and J type thermocouples ■
Dual temperature for easy Super-Heat and Sub-Cool ■
calculations
Selectable °F or °C ■
Min/Max recording ■
No hassle
warranty
No waiting.
No shipping
charges.
Our commitment to
high-quality products
and customer service
is demonstrated by
our industry exclusive
“No Hassle” warranty.
In the unlikely event
that an Amprobe Test
Tool requires warranty
service, any of our
local dealers are
authorized to replace
it, on the spot.
(note: $500 MSLP limit)
Delta reading (T1 – T2) ■
Data Hold function to retain reading on the screen for ■
later viewing
Low battery indication ■
Auto-Power off ■
Offset Mode to compensate for error in the thermocouple ■
Includes carrying case, battery, and two K-type probes, ■
Users Manual
Amprobe® Test Tools
TMD-10 Dual Temperature Meter
Data Sheet
Specifications
General Specifications
Display Dual 5-digit LCD displays
Overrange indication “OL” stands for positive temperature
Zero Automatic
Low battery indication The “”sign will appear while battery is low
Power Supply 9V NEDA 1604, IEC 6F22, JIS 006P battery.
Battery life approx. 200 hours
Auto power off approx. 15 minutes
Measurement rate 0.5 sec
Operating environment 0°C to 50°C (32°F to 122°F); < 80% RH
Storage temperature 0°C to 50°C (32°F to 122°F); < 80% RH
Sensor Type K and J thermocouples
Altitude 2000m, indoor operation
Dimensions 130 x 56 x 38 mm (5.1 x 2.2 x 1.5 in.)
Weight 170g (0.37 lb.) battery included
Electrical (23°C ± 5°C, < 80% RH)
Temperature scale °C or °F
Thermocouple type K-type: -200°C to 1370°C (-328°F to 2498°F)
Resolution 0.1 degree C or F
Accuracy ± (0.05% rdg + 0.7°C), ± (0.05% rdg + 1.4°F)
Input protection 24V DC or AC
Temperature Coefficient ± (0.01% rdg + 0.03°C per °C); ± (0.01% rdg + 0.06°F per °F) outside the specified + 18°C to + 28°C
Calibration Yearly
Agency Approvals & Certifications
“-OL” stands for negative temperature
J-type: -200°C to 1050°C (-328°F to 1922°F)
(+ 64°F to + 82°F) range
-EMC EN61326-1
©2008 Amprobe Test Tools. All rights re served.
10/2008 3360423 Rev A
Amprobe® Test Tools