3 3/4 DIGIT AUTO DIGITAL MULTIMETER
SAFETY RULES
λ This meter is designed and tested in accordance with IEC publication 1010, pollution
degree II and installation category (overvoltage category) III 600V.
λ This meter has been tested according to the following EC Directives
♦ 89/336/EEC (EMC of Nov., 1992, Electromagnetic Compatibility)
♦ 73/23/EEC (Product safety law of June 11, 1979, Low Voltage Directive of
February 19, 1973)
λ This meter is designed to be indoor use at temperature 41°F to 104°F (5°C to 40°C)
and altitude up to 2,000m.
λ To ensure that the meter is used safely, follow all safety and operating instructions in
this operation manual. If the meter is not used as described in this operation manual,
the safety features of this meter might be impaired.
λ Do not use the meter if the meter or test leads look damaged, or if you suspect that the
meter is not operating properly.
λ When using the probes, keep your fingers behind the finger guards on the probes.
λ Disconnect the live test lead before disconnecting the common test lead.
λ Make sure power is off before cutting, unsoldering, or breaking the circuit. Small
amount of current can be dangerous.
λ Do not apply more than 600Vdc or 600Vac rms between a terminal and earth ground.
λ To avoid electrical shock, use CAUTION when working above 60Vdc or 25Vac rms.
Such voltages pose a shock hazard.
λ Never make measurements with the battery cover or bottom case off.
λ To avoid electrical shock or damage to the meter, do not exceed the input limits.
INTERNATION SYMBOLS
Important information Diode
!
see manual Continuity
AC Ground
DC Double insulation
FEATURES
λ One rotary switch operation.
λ Auto polarity with engineering units annunciator
λ Dual polarity A-D converter.
λ Auto voltage ranges selection: DCV: 1mV - 600V, ACV: 1mV - 600V
λ Auto ampere ranges selection: DCA: 40mA - 10A, ACA: 40mA - 10A
λ Auto resistance ranges selection: 0.1Ω - 40MΩ
λ Diode & continuity test
3.1 General Specifications
Display : 3 3/4 digit LCD, 3999 counts with decimal point and
Polarity : Automatic, (-) negative polarity indication.
Zero adjustment : Automatic.
Measuring Method : Dual slop integration A to D converter system.
Over range indication : Only the “OL” is displayed.
Power : Two, standard 1.5 volt (AAA) battery, UM4 or equivalent.
Dimension : 70 (H) x 144 (W) x 40 (D) mm.
Weight : Approx. 280gm. (including batteries and packaging)
3.2 Electrical Specifications
Accuracies are ± (% of reading + number of least significant digits) at 23°C ± 5°C, <75% RH.
DC Voltage Range Resolution Accuracy Input
4V 1mV
40V 10mV ±(0.8%+1) 10MΩ 600V DC/AC rms
AC Voltage Range Resolution Accuracy Frequency
DC Current Range Resolution Accuracy Input resistor
AC Current Range Resolution Accuracy Input resistor
Resistance Range Resolution Accuracy Measurement
Diode Test Range Resolution
Continuity Range Resolution
Test 400Ω 100mΩ
PANEL DESCRIPTIONS
1) 3 3/4 digit LCD display
2) Continuity test
3) Resistance range
4) AC voltage range
5) DC voltage range
6) Power off
7) Rotary switch
8) Data hold / power resume button
9) Case
10) Diode range
11) DC current range
12) AC current range
600V 1V
4V 1mV Overload
40V 10mV
400V - 600V 1V ±(1.5%+5) 40-100 600V rms
10A 10mA
10A 10mA
400Ω 100mΩ VDDA 1.2mA or
4kΩ 1Ω
40kΩ 10Ω for 40MΩ from
400kΩ 100Ω
4MΩ 1kΩ
40MΩ 10kΩ±(1.5%+2)
OPERATION MANUAL
engineering units annunciator display
Impedance
(Hz)
±(1.2%+3)
±(1.8%+5) 0.01Ω
±(3.0%+5) 0.01Ω
±(1.0%+2)
0~3.999V 1mV
40-400
Voltage/Current
less
0.7V(TYP)
0.7mA or less
0.47V(TYP)
0.7mA or less
Open short or Normal Range
Buzzer sound generation resistance
Overload
Protection:
Input Impedance:
10MΩ
Protection:
Overload
Protection:
10A/600V fuse
Response time:
5 sec. max. on
40MΩ range
3 sec. on all other
range
Temperature range
18°C to 30°C
_50Ω
AM-34
OPERATION WARNING
1) When measuring voltage ensure that the instrument is not connected or switched to a
current or resistance or temperature or to the diode check/continuity range. Always ensure
that the correct terminals are used for the type of measurement to be made.
2) Use extreme care when measuring voltage above 50V, especially from sources where
high energy exists.
3) Avoid making connections to “live” circuits whenever possible.
4) When making current measurements ensure that the circuit is not “live” before opening it
in order to connect the test leads.
5) Before making resistance measurements or continuity / diode test, ensure that the circuit
under test is de-energised.
6) Always ensure that the correct function and range is selected. If in doubt about the correct
range, start with the highest and work downwards.
7) Extreme care should be taken when using the instrument to conjunction with a current
transformer connected to the terminals. High voltage may be produced at the terminals if
an open circuit occurs.
8) Ensure that the test leads and prods are in good condition with no damage to the
insulation.
9) Take care not to exceed the over-load limits as given in the specifications.
10) Fuse for replacement must be of the correct type and rating.
5.1 DC and AC Voltage measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to DC V or AC V position and connect the test prods across the
source or load under measurement.
5.2 DC and AC Current measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “10A” socket for
measurement up to 10A.
2) Set the selector switch to the DC current or AC current position.
3) Connect the prods in series with the current source to be measured.
5.3 Resistance measurement
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to resistance Ω position.
3) Connect the prods across the circuit to be tested.
!
CAUTION: Ensure that the circuit to be tested is “dead”.
Max. input over-load : 600V rms < 10sec.
5.4 Diode test
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to position.
3) Connect the black and red test prods to the cathode (-) and anode (+) ends of the diode to
be tested respectively.
4) Read the forward voltage drop (junction) value from the display. If reverse connected the
prods to diode, display shows over-load.
5.5 Continuity test
1) Connect the black test lead to the “COM” socket and red test lead to the “ ” socket.
2) Set the selector switch to position.
3) Connect the prods across the circuit to be tested, if the resistance less than approx. 50Ω,
buzzer will be activated.
5.6 Data hold
A push switch at the middle of rotary switch is used to hold data during measurement.
Pressing the key, reading is held and “H” appears on display. Resume active reading by
pressing switch gently once.
5.7 Auto power off
The meter will turn off automatically to extend the battery life after approx. 15 minutes without
operation. Resume power by pressing data hold push switch gently once.
6. MAINTENANCE
!
CAUTION
BEFORE ATTEMPTING BATTERY AND FUSE REMOVAL OR REPLACEMENT,
DISCONNECT TEST LEADS FROM ANY ENERGISED CIRCUITS TO AVOID SHOCK
HAZARD.
6.1 Fitting and replacing the battery and fuse
1) Ensure that the instrument is not connected to any external circuit, set the selector switch
to OFF position and remove the test leads from the terminals.
2) Remove the screw of the back case.
3) Replace the spent battery or fuse with the same type and rating.
4) Reinstate the back case, tighten and securing screw.
6.2 Cleaning
Periodically wipe the case with a soft damp cloth and mild household cleanser. Do not use
abrasives or solvents. Ensure that no water gets inside the equipment to prevent possible
shorts and damage.