Amprobe ALC-110 Specifications

ALC-110
Leakage Current Clamp
Prevent unplanned downtime by detecting leakage currents with the ALC-110 Leakage Current Clamp. Rated CAT III 600 V to ensure electrical safety, the ALC-110 is designed to resist external magnetic fields for highly accurate readings.
Features
• Low influence of external low frequency
magnetic fields: @ I
≤ 30 A/m, operating class 2
• True-rms measurements for accuracy when
measuring complex, non-sinusoidal waveforms
• Mechanical jaw lock to assure accuracy of low
current measurements, avoiding disturbance from measured wires and magnetic fields; jaw opening 1.2 in (30 mm)
• Highest resolution of 0.001 mA,
measure up to 60 A
• Broad frequency range 15 Hz to 1 kHz for
a variety of different applications, including industrial and railway
• Selectable filters:
- Low Pass 50/60Hz filter to perform measurements on equipment driven by Variable Frequency Drives (VFDs) or affected by high
frequency noise
- Appliance Filter to conduct leakage current
measurements on appliances
- No filter
• Selectable limits:
0.25 mA, 0.5 mA, 3.5 mA, 10 mA, 12 mA
• Max readings and hold function
• High safety rating: CAT III 600 V
• Battery-saving design with an auto backlight off
and auto power off
N: 3.500 mA – 600.0 mA
amprobe.com
Mechanical jaw lock for high accuracy readings
Finger guard for safe usage
Robust, strong rubber housing for easy grip
Large digital display with bar graph and filter function
Applications
• Measure:
- Earth and differential leakage currents
- Leakage currents through the earth (PE) conductor
- Appliance current consumption without interruptions to the circuit
• Trace the source of earth leakage current
• Identify intermittent GFCI trips without
switching equipment off
• Check a system for an electrical break of
insulation where current is flowing to the ground
• Where industrial leakage current testing is required
Electrical
ALC-110
Safety
Leakage Current Clamp
6
Safety Certification
All Amprobe tools, including the Amprobe ALC-110, are rigorously tested for safety, accuracy, reliability, and ruggedness in our state-of-the-art test lab. In addition, Amprobe products that measure electricity are listed by a
3rd party safety lab, either UL or CSA. This system assures that Amprobe products meet or exceed safety regulations and will perform in a tough, professional environment for many years to come.
©2019 Amprobe® For detailed specications and ordering go to amprobe.com 6011715 AAmprobe® | info@amprobe.com | Fluke Corporation, Everett, WA 98203 | Tel: 877-AMPROBE (267-7623)
Specifications
Features ALC-110
Complies with IEC/EN 61557-13
Built according IEC/EN 61010-2-032
Low frequency magnetic fields ≤ 30 A/m, operating class 2 @ I
Safety, measurement category CAT III 600 V
True-rms
AC frequency 15 Hz to 1 kHz
Low pass filter 50 Hz to 60 Hz, for appliance measuring acc. EN61557-16 amendment A, no filter
Limit judgement 0.25 mA, 0.5 mA, 3.5 mA, 10 mA, 12 mA
Jaw opening 1.2 in (30 mm) max.
Jaw locking
Data hold
Backlight function
Overload protection 60 A
Auto power off Idle 20 minutes
Auto ranging
Max hold
AC current
[1]
Add 4.0% for C.F. 2.5 to 3.0. Position error of clamp: +/- 1% of LCD reading.
[2]
[3]
[1] [2] [3]
Range Resolution
6.000 mA 0.001 mA ±(5.0%+5D) ±(2.0%+5D) ±(1.0%+5D) ±(2.0%+5D)
60.00 mA 0.01 mA ±(5.0%+5D) ±(2.0%+5D) ±(1.0%+5D) ±(2.0%+5D)
600.0 mA 0.1 mA ±(5.0%+5D) ±(2.0%+5D) ±(1.0%+5D) ±(2.0%+5D)
6.000 A 0.001 A Unspecified ±(2.0%+5D) ±(1.0%+5D) ±(2.0%+5D)
60.00 A 0.01 A Unspecified ±(2.0%+5D) ±(1.0%+5D) ±(2.0%+5D)
Crest Factor (C.F.) may be up to 3.0 as 3000 counts. For non-sinusoidal waveforms: Add 1.0% for C.F. 1.0 to 2.0 / Add 2.5% for C.F. 2.0 to 2.5 /
Minimum reading is 0.010 mA. When operating in magnetic fields up to ≤ 30 A/m, add ±0.3 mA error to accuracy.
15 Hz to 40 Hz > 40 Hz to 50 Hz > 50 Hz to 60 Hz > 60 Hz to 1 kHz
N : 3.500 mA to 600.0 mA / 40 Hz to 1 kHz
Accuracy
amprobe.com
General Specifications
Display Digital: 6000 counts, updates 5/sec
Range Auto ranging
Over range indication OL or -OL
Operating temperature
Operating altitude ≤ 2000 m
Storage temperature -4 °F to 140 °F (-20 °C to 60 °C)
Pollution degree 2
Temperature coefficient 0.1 x (specified accuracy) / °C (<21 °C, >25 °C)
Vibration requirement acc. IEC 61557-1 clause 4.10
Drop proof 4 ft (1.2 m)
Battery 2 x batteries 1.5 V IEC LR03 AAA
Battery life 60 hours typical
Low battery indication
Safety compliance IEC/EN 61010-1, IEC/EN 61010-2-032, IEC/EN 61557-13
EMC compliance IEC/EN 61326-1, IEC/EN 61326-2-2
Agency approvals
Dimensions (H x W x L) 8.7 x 3.5 x 1.8 in (221 x 89 x 48 mm)
Weight 0.9 lb (410 g) with batteries installed
Included: 2 x AAA batteries, soft carrying case and user manual
32 °F to 86 °F (0 °C to 30 °C), ≤ 80% RH
86 °F to 104 °F (30 °C to 40 °C), ≤ 75% RH
104 °F to 122 °F (40 °C to 50 °C), ≤ 45% RH
©2019 Amprobe® For detailed specications and ordering go to amprobe.com 6011715 AAmprobe® | info@amprobe.com | Fluke Corporation, Everett, WA 98203 | Tel: 877-AMPROBE (267-7623)
Loading...