Installation Instructions
Noble Series
Neonatal Intensive Care Unit (NICU)
Corporation
Page 2
Before Installation
Note
The installation instructions provided are for the Amico standard products. Slight modications to the installations may be required. Please refer to project specic submittal package for
1 Amico Architectural
Page 3
Installation
1.
Review a copy of the nal approved shop drawing(s) and/or submittal package. These documents will provide you with technical information specic to your installation, such as:
- equipment types and quantities
- room numbers and location
- shop drawings of each unit type
- wiring diagrams and rough-in locations
2.
Locate the NICU unit from the shipping boxes that you wish to install. The contents of each shipping box are listed on the shipping document(s).
www.amico.com 2
Page 4
Other Installation Notes
Attach the top unit to the base using holes and hardware provided.
1.
Position the base in the desired location in the room and secure it to the wall. Access to the
2. inside of the base is via laminate panels, all of which are removable. To remove a laminate panel, use the suction cup tool provided to rst lift the panels vertically until resistance is felt, and then rotate the bottom of the panel out and pull down. Be careful not to drop panel.
tool provided:
suction cup
3 Amico Architectural
Page 5
Installation
Position the chase in place and secure it to the wall structure. To access to the inside of the
3. chase, remove any laminate panels necessary.
Make all gas and electrical connections between the base, the headwall sections, and the chase.
4.
www.amico.com 4
Page 6
Make all gas and electrical connections between the NICU unit and gas and electrical sources.
5. The electrical termination may be located in the base, in the chase, or could just be loose conduit extending outside the chase. See wiring diagrams provided with each project.
Other Installation Notes
Division 15 to make gas pipe connections from base to chase, and connect gas stubs from chase
1. to building medical gas supply lines previously roughed in.
2.
Division 16 to do all electrical connections required.
5 Amico Architectural
Page 7
www.amico.com
Amico Architectural | www.amico.com
85 Fulton Way, Richmond Hill
Ontario, L4B 2N4, Canada
71 East Industry Court, Deer Park NY 11729, U.S.A
Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778
Fax: 905.763.8587 Email: info@amico.com
C US LISTED
INSTALL INSTR. NOBLE NICU 2011
Loading...