1
Features:
Inlet and outlet pressure display gauges in psi/kPa
Manual shut-o valve
Outlet supply pipe for remote mounted outlets
3/8” [9.5 mm] internal tubing for high ow
High ow capacity regulator
Wall mount with adjustable brackets for di erent wall
thicknesses
Aluminum front simpli es cleaning
Maximum supply pressure 250 psi [1,724 kPa]
Maximum delivery pressure:
DISS = 200 psi
HP (Safety Swing Coupling) = 250 psi
Gas Control Panel
General Speci cations:
The Gas Control panel shall be an Amico Alert-1 series.
The control panel shall be supplied with a quarter turn
shut-o ball valve, rated at no less than 300 psi [2,069
kPa].
The unit shall have an inlet pressure indicator 0 - 400 psi
[0-2, 750kPa] valve to indicate the inlet supply pressure.
The pressure regulator shall be adjustable between 0 to
250 psi [0 to 1,724 kPa].
The outlet controlled pressure shall be monitored by a
pressure indicator with a range of 0 - 400 psi [0 - 2,750
kPa].
The DISS outlet shall be a Diameter Index Safety System (D.I.S.S.) For Air or Nitrogen Service Outlet or “Safety
Swing Coupling” for pressures above 200 psi. The outlets shall be used for connection to pneumatic surgical
tools.
The controls shall be mounted in an 1.3 mm [#18 gauge]
painted steel enclosure, with adjustable bracket for
different wall thicknesses.
The unit shall be factory tested for intended gas service.
The Gas Control Panel meets CSA and NFPA requirements.
Amico products comply with NFPA-99.