Page 1
Retro-Fit Manual
www.amico.com
NCG Chemetron
Page 2
Important Note
Due to unknown factors of the existing equipment, projects requiring Retro-Fit Conversions must be fully investigated by the installer.
It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are required. Amico is not responsible for any additional components required.
Existing NCG
Retro-Fit Instructions
Step 1:
Remove all secondary check parts from the existing outlets. Replace the NCG rubber seat with Amico #15 O-Ring.
Rough-In
Step 2:
Thread the Amico Retro-Fit rough-in assembly to the existing NCG rough-in.
O-VC-OH1-TOOL
Amico Tool
Hole for Screw Driver (to loosen & tighten)
NCG Original Rubber Seat (to be removed & replaced by #15 O-Ring)
Existing NCG Rough-In
O-Ring (H-ORING-015VTN)
Back Rough-In Assembly Kit (O-VC-NCG-U-GAS)
2 Amico Corporation
Page 3
Retro-Fit Instructions
Step 3:
Use the rivet provided to attach the upgrade bracket to the existing console to ensure it is gas-specic.
Note: Ensure the Amico rough-in and the NCG rough-in are the same gas.
Step 4:
Place the pre-assembled ller frame and face plate over the wall outlet and insert the latch valve assembly through the face plate and fasten with the latch valve screws.
Existing NCG Rough-In
Aluminum Rivet
(H-RIVIT-H-W-END)
(H-RIVIT-H-W-END)
#6-32 Screw (H-MPF-0612)
Latch Valve Screw (H-1612-06-18A)
Existing NCG Rough-In
Filler Frame (O-CON-FIL-PB500)
Chromed Faceplate (O-TRIM-M500C)
Front Latch Valve (O-FASC-QD-U-GAS)