Important Note
Due to unknown factors of the existing equipment, projects requiring Retro-Fit Conversions must be fully investigated by
the installer.
It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are
required. Amico is not responsible for any additional components required.
Existing NCG
Retro-Fit Instructions
Step 1:
Remove all secondary check parts from the
existing outlets. Replace the NCG rubber seat with
Amico #15 O-Ring.
Rough-In
Step 2:
Thread the Amico Retro-Fit rough-in assembly to
the existing NCG rough-in.
O-VC-OH1-TOOL
Amico Tool
Hole for Screw Driver
(to loosen & tighten)
NCG Original Rubber
Seat (to be removed &
replaced by #15 O-Ring)
Existing NCG
Rough-In
O-Ring
(H-ORING-015VTN)
Back Rough-In
Assembly Kit
(O-VC-NCG-U-GAS)
2 Amico Corporation

Retro-Fit Instructions
Step 3:
Use the rivet provided to attach the upgrade
bracket to the existing console to ensure it is
gas-specic.
Note: Ensure the Amico rough-in and the NCG
rough-in are the same gas.
Step 4:
Place the pre-assembled ller frame and face
plate over the wall outlet and insert the latch
valve assembly through the face plate and
fasten with the latch valve screws.
Existing NCG
Rough-In
Aluminum Rivet
(H-RIVIT-H-W-END)
(H-RIVIT-H-W-END)
#6-32 Screw
(H-MPF-0612)
Latch Valve Screw
(H-1612-06-18A)
Existing NCG
Rough-In
Filler Frame
(O-CON-FIL-PB500)
Chromed Faceplate
(O-TRIM-M500C)
Front Latch Valve
(O-FASC-QD-U-GAS)
www.amico.com 3

www.amico.com
Amico Corporation | www.amico.com
85 Fulton Way, Richmond Hill
Ontario, L4B 2N4, Canada
Toll Free Tel: 1.877.462.6426
Toll Free Fax: 1.866.440.4986
Tel: 905.764.0800
Fax: 905.764.0862
Email: info@amico.com