Amico Mira 45 User Manual

Page 1
Installation Instructions
www.amico.com
Mira 45
Page 2
Table of Contents
General Information 3
Safety instruction 3
Abbreviation and Symbols 4
Assembly and Installation MIRA 45 Single Ceiling 5
Assembly and Installation MIRA 45 Double Ceiling 6
Assembly and Installation MIRA 45 Wall Mount 7
Assembly and Installation MIRA 45 Floor Stand 8
Head Assembly 8
Putting into Operation 9
Care, Maintenance and Repair 9
Technical Data 11
Disposal 11
2 Amico Lighting Solutions
Page 3
General Information
UEnglish
1. General Information
UEnglish
1. General Information
1.1. Designated Use
1.1.1. Intended purpose
Minor surgical luminaire (Treatment light) – light for local illumination of the patient’s body during diagnosis or treatment. It is not intended for use in operating theatres.
1.1.2. Location
Only in dry rooms not subject to the risk of explosion. The minimum distance to inflammable parts has to be 0.5m.
1.1.3. Operation mode
The light is designed for continuous operation.
2. Safety Instruction
The light may only be operated in dry rooms, not subject to the risk of explosion.
Do not leave the light on without supervision.
The minimum distance to inflammable parts has to be 0.5 m.
Keep inflammable substances and objects away from the light and the light cone.
When using lights of protection class I make sure that the grounding conductor is connected with the lamp housing.
Do not touch the luminous medium while the light is switched on and for some time afterwards – danger of burns.
Defective apertures must be replacedRisk of injury.
Intended Purpose
Minor surgical luminaire (Treatment light) – light for local illumination of the patient’s body during diagnosis or treatment. It is not intended for use in operating theatres.
Location
Only in dry rooms not subject to the risk of explosion. The minimum distance to inammable parts has to be 0.5 m.
Operation Mode
The light is designed for continuous operation.
Safety Instructions
• The light may only be operated in dry rooms, not subject to the risk of explosion.
• Do not leave the light on without supervision.
• The minimum distance to inammable parts has to be 0.5 m.
• Keep inammable substances and objects away from the light and the light cone.
• When using lights of protection class I make sure that the grounding conductor is
connected with the lamp housing.
• Do not touch the luminous medium while the light is switched on and for some time afterwards – danger of burns.
• Defective apertures must be replaced – Risk of injury.
• Apart from changing the lamps, no further modications or work may be carried
out by the user.
• Defective lamps may only be replaced with lamps with the same power (W, Watt) and of the same type (specications see ’Technical Data’ at page 11 ).
• Do not kink the power supply cable to avoid damages.
• Do not touch the protective glass.
• The user has to comply with the requirements of the national institute for hygiene
and disinfection. Risk: Transmission of diseases.
www.amico.com 3
Page 4
Abbreviations and Symbols
3. Abbreviation and Symbols
Attention, observe accompanying documents
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
Defective shades must be replaced
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
Defective shades must be replaced
Symbol for separate disposal according the regulation 2002/96/EC (WEEE - Waste Electrical and Electronic Equipment)
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
Defective shades must be replaced
Symbol for separate disposal according the regulation 2002/96/EC (WEEE - Waste Electrical and Electronic Equipment)
UL-Conformity mark (USA + Canada)
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
Defective shades must be replaced
Symbol for separate disposal according the regulation 2002/96/EC (WEEE - Waste Electrical and Electronic Equipment)
UL-Conformity mark (USA + Canada)
VDE-Conformity mark
3. Abbreviation and Symbols
Attention, observe accompanying documents
Device of international protection class I (protective earth)
I
ON
0
OFF
I
ON (stage I = 60%)
II
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5m to inflammable parts
Defective shades must be replaced
Symbol for separate disposal according the regulation 2002/96/EC (WEEE - Waste Electrical and Electronic Equipment)
UL-Conformity mark (USA + Canada)
VDE-Conformity mark
CE-Conformity mark
Attention, observe accompanying documents
Device of international protection class I (protective earth)
ON
OFF
ON (stage I = 60%)
ON (stage II = 100%)
Attention! Hot surface
Keep the minimum distance of 0.5 m to inammable parts
Defective shades must be replaced
4 Amico Lighting Solutions
Separate disposal according the regulation 2002/96/EC (WEEE - Waste Electrical and Electronic Equipment)
UL-Conformity mark (USA + Canada)
VDE-Conformity mark
CE-Conformity mark
The manufacturer cannot be held liable for damages caused if the light is used for other purposes
!
than the intended purpose or if safety instruction and warnings are ignored.
Page 5
Operating manual (GB)
4. Assembly and Installation MIRA 45 Single Ceiling
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Single Ceiling. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The ceiling fitting is to be used exclusively at ceilings that are either solid, or supplied with a suitable high-strength ceiling-mounting solution.
c Drill holes in accordance to the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque).
d Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screw has to be flush.
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and fix it temporarily, using supplied bolts (see detailed view 2). Align the support tube to the direction of the fixture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
4. Assembly and Installation MIRA 45 Single Ceiling
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Single Ceiling. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The ceiling fitting is to be used exclusively at ceilings that are either solid, or supplied with a suitable high-strength ceiling-mounting solution.
c Drill holes in accordance to the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque).
d Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screw has to be flush.
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and fix it temporarily, using supplied bolts (see detailed view 2). Align the support tube to the direction of the fixture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
f Head Assembly (see page 68)
Before mounting the MIRA 45 Single Ceiling, take heed of the operating range.
Assembly and Installation MIRA 45 Single Ceiling
Note: Only qualied technicians should carry out MIRA 45 installation.
Mounting steps
Detailed view 1
Detailed view 2
Before mounting the MIRA 45 Single Ceiling, take
Two people are required for assembling the MIRA 45 Single Ceiling. The supply voltage must correlate with the voltage noted on the xture’s rating label. The ceiling tting is to be used exclusively at ceilings that are either solid, or supplied with a suitable high­strength ceiling-mounting solution.
1. Drill holes in accordance to the enclosed drilling template (Ø=8 mm; depth=55 mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4 turns, 23Nm torque).
2. Perform this Step on a at and level surface. Pull the cable through the support tube. Insert the slider into the groove of the arm joint, noting the location of the xing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screw has to
be ush.
3. Place the O-ring and the ceiling cover over the support tube. Insert the support tube into the anchor plate and x it temporarily, using supplied bolts (see detailed view 2). Align the support tube to the direction of the xture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
4. Head Assembly (see page 68) Before mounting the MIRA 45 Single Ceiling, take heed of the operating range.
heed of the operating range.
www.amico.com 5
Page 6
Operating manual (GB)
5. Assembly and Installation MIRA 45 Double Ceiling
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Double Ceiling. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The ceiling fitting is to be used exclusively with ceilings that are either solid, or supplied with a suitable high-strength ceiling-mounting solution.
c Drill holes in accordance with the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque.
d Perform this Step on a flat and level surface. Pull the cable through the
bearing head. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (1a). Align the hole in the slider with the set- screw hole at the bottom of the bearing head. Insert the set-screw (18 mm) into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screws had to be flush. Pull the cable through the cover (1d) and the support tube (1e). Insert the slider into the groove of the arm joint, noting the location of the fixing hole (1f). Slide the support tube and the bearing head together (1e). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw (10 mm) into the hole to lock the slider in place (1g). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Fix the cover (1d) with the four screw (1h) Check: Set-screw has to be flush.
Detailed view 2
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and fix it temporarily, using supplied bolts (see detailed view 2). Align the support tube to the direction of the fixture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
5. Assembly and Installation MIRA 45 Double Ceiling
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Double Ceiling. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The ceiling fitting is to be used exclusively with ceilings that are either solid, or supplied with a suitable high-strength ceiling-mounting solution.
c Drill holes in accordance with the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque.
Detailed view 1
d Perform this Step on a flat and level surface. Pull the cable through the
bearing head. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (1a). Align the hole in the slider with the set- screw hole at the bottom of the bearing head. Insert the set-screw (18 mm) into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screws had to be flush. Pull the cable through the cover (1d) and the support tube (1e). Insert the slider into the groove of the arm joint, noting the location of the fixing hole (1f). Slide the support tube and the bearing head together (1e). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw (10 mm) into the hole to lock the slider in place (1g). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Fix the cover (1d) with the four screw (1h) Check: Set-screw has to be flush.
Detailed view 2
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and fix it temporarily, using supplied bolts (see detailed view 2). Align the support tube to the direction of the fixture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
f Head Assembly (see page 8)
Assembly and Installation MIRA 45 Double Ceiling
Note: Only qualied technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Double Ceiling. The supply voltage must correlate with the voltage noted on the xture’s rating label. The ceiling tting is to be used exclusively with ceilings that are either solid, or supplied with a suitable high­strength ceiling-mounting solution.
1. Drill holes in accordance with the enclosed drilling template (Ø=8 mm; depth=55
Detailed view 1
mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open­ended wrench (approx. 3-4 turns, 23Nm torque.
2. Perform this Step on a at and level surface. Pull the cable through the bearing head. Insert the slider into the groove of the arm joint, noting the location of the xing hole (1a). Align the hole in the slider with the set-screw hole at the bottom of the bearing head. Insert the set-screw (18 mm) into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Check: Set-screws had to be ush. Pull the cable through the cover (1d) and the support tube (1e). Insert the slider into the groove of the arm joint, noting the location of the xing hole (1f). Slide the support tube and the bearing head together (1e). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw (10 mm) into the hole to lock the slider in place (1g). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint. Fix the cover (1d) with the four screw (1h) Check: Set-screw has to be ush.
3. Place the O-ring and the ceiling cover over the support tube. Insert the support tube into the anchor plate and x it temporarily, using supplied bolts (see detailed view
2). Align the support tube to the direction of the xture’s general usage. Tighten the screws on the anchor plate. Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring upwards.
4. Head Assembly (see page 8)
6 Amico Lighting Solutions
Mounting steps
Page 7
6. Assembly and Installation MIRA 45 Wall Mount
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The wall fitting is to be used exclusively with walls that are either solid, or supplied with a suitable high-strength wall-mounting solution.
c Drill holes in accordance with the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts according to detailed view 2. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque.
d Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint.
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and install the locking bolt (Detailed view 2a) from top. Tighten the screws on the anchor plate. Observe the correct order of the washer (2b) and the serrated lock washer (2c). Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring to the anchor plate.
6. Assembly and Installation MIRA 45 Wall Mount
Note: Only qualified technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply voltage must correlate with the voltage noted on the fixture’s rating label. The wall fitting is to be used exclusively with walls that are either solid, or supplied with a suitable high-strength wall-mounting solution.
c Drill holes in accordance with the enclosed drilling template (=8mm;
depth=55mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts according to detailed view 2. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4turns, 23Nm torque.
d Perform this Step on a flat and level surface. Pull the cable through the
support tube. Insert the slider into the groove of the arm joint, noting the location of the fixing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint.
e Place the O-ring and the ceiling cover over the support tube. Insert the
support tube into the anchor plate and install the locking bolt (Detailed view 2a) from top. Tighten the screws on the anchor plate. Observe the correct order of the washer (2b) and the serrated lock washer (2c). Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring to the anchor plate.
f Head Assembly (see page 6 8)
Assembly and Installation MIRA 45 Wall Mount
Note: Only qualied technicians should carry out MIRA 45 installation.
Two people are required for assembling the MIRA 45 Wall Mount. The supply voltage must correlate with the voltage noted on the xture’s rating label. The wall tting is to be used exclusively with walls that are either solid, or supplied with a suitable high-strength wall­mounting solution.
1. Drill holes in accordance with the enclosed drilling template (Ø=8 mm; depth=55 mm). Clean out the drilled holes. Attach the anchor plate using the four anchor bolts according to detailed view 2. Tap the anchor bolts completely in, using a hammer. Tighten the nut with an open-ended wrench (approx. 3-4 turns, 23Nm torque.
Detailed view 1
2. Perform this Step on a at and level surface. Pull the cable through the support tube. Insert the slider into the groove of the arm joint, noting the location of the xing hole (see detailed view 1 part ‘a’). Slide the arm and support tube together (view 1, direction ‘b’). Align the hole in the slider with the set-screw hole at the bottom of the support tube. Insert the set-screw into the hole to lock the slider in place (1c). Tighten the set-screw completely, then loosen it by a maximum 1-quarter turn and check the operation of the arm joint.
3. Place the O-ring and the ceiling cover over the support tube. Insert the support tube
Detailed view 2
into the anchor plate and install the locking bolt (Detailed view 2a) from top. Tighten the screws on the anchor plate. Observe the correct order of the washer (2b) and the serrated lock washer (2c). Now connect to the electrical power supply, as per the wiring diagram (see sheet enclosed with the set of bolts). Slide the ceiling cover and the O-ring to the anchor plate.
4. Head Assembly (see page 8)
Mounting steps
www.amico.com 7
Page 8
7. Assembly and Installation MIRA 45 Floor Stand
c Insert the support tube into the castor base and tighten the set-screws
(support disc, screws - 1xM8, 1 x M5 – and serrated washer).
d Place the spring-loaded arm on the support tube and align the holes.
Attach the cable holder with the supplied screws and tighten.
e Head Assembly (see below)
f
7. Assembly and Installation MIRA 45 Floor Stand
Mounting steps
c Insert the support tube into the castor base and tighten the set-screws
(support disc, screws - 1xM8, 1 x M5 – and serrated washer).
d Place the spring-loaded arm on the support tube and align the holes.
Attach the cable holder with the supplied screws and tighten.
e Head Assembly (see below)
8. Head Assembly
Match the plug- and socket connectors as shown in Fig c completely. The active parts has to be total covered. Align the hole on the end of the top arm with the hole on the limiter and slide the limiter into the top arm (see Figs. d and e). Insert and tighten the two set-screws to hold the head in place (see Fig. f).
c
d
e
f
7. Assembly and Installation MIRA 45 Floor Stand
Mounting steps
c Insert the support tube into the castor base and tighten the set-screws
(support disc, screws - 1xM8, 1 x M5 – and serrated washer).
d Place the spring-loaded arm on the support tube and align the holes.
Attach the cable holder with the supplied screws and tighten.
e Head Assembly (see below)
8. Head Assembly
Match the plug- and socket connectors as shown in Fig c completely. The active parts has to be total covered. Align the hole on the end of the top arm with the hole on the limiter and slide the limiter into the top arm (see Figs. d and e). Insert and tighten the two set-screws to hold the head in place (see Fig. f).
Assembly and Installation MIRA 45 Floor Stand
1. Insert the support tube into the castor base and tighten the set-screws (support disc, screws - 1 x M8, 1 x M5 – and serrated washer).
2. Place the spring-loaded arm on the support tube and align the holes. Attach the cable holder with the supplied screws and tighten.
3. Head Assembly (see below)
Mounting steps
Head Assembly
Match the plug- and socket connectors as shown in Fig 1 completely. The active parts has to be total covered. Align the hole on the end of the top arm with the hole on the limiter and slide the limiter into the top arm (see Figs. 2 and 3). Insert and tighten the two set-screws to hold the head in place (see Fig. 4)
c
8 Amico Lighting Solutions
e
f
Page 9
Operating manual (GB)
9. Putting into Operation
Before placing the light into operation, the user has to check whether mains voltage and frequency specified on the rating plate are identical.
10. Care, Maintenance and Repair
Disconnect the light from the mains before carrying out any maintenance or repair work! This work must be carried out only by a qualified electrician!
As replacement parts only use part approved by the manufacturer.
10.1. Regular checks
9. Putting into Operation
Before placing the light into operation, the user has to check whether mains voltage and frequency specified on the rating plate are identical.
10. Care, Maintenance and Repair
Disconnect the light from the mains before carrying out any maintenance or repair work! This work must be carried out only by a qualified electrician!
As replacement parts only use part approved by the manufacturer.
10.1. Regular checks
The mains line must be checked minimum once a a year for damage (only MIRA 45 Floor Stand). In case of damage the manufacturer must replace the defective mains line.
10.2. Adjustment of the counter-balance (Fig. c)
Figure c
The counter-balance arm will already have been factory set for best possible operation. Should adjustment become necessary: Set the spring tension as illustrated, and using a clearance of 4 mm.
10.3. Changing the lamp
Note: Let the halogen lamp cool down before changing it!
1. Lift the hinged protective back cover of the head (see Fig. c)
2. The required Allen wrench is located inside the back cover (see Fig. d). Remove the lamp head retaining screw (see Fig. e)
3. Remove the cover ring and reflector
4. Change the halogen lamp (use a soft cloth and do not touch the glass surface of either the old or new lamp, see Fig. f)
5. Replace the reflector an cover ring. Replace an tighten the screws.
6. Replace the Allen wrench (see Fig, d)
Note: Only use lamp types approved of by the manufacturer (see ‚
Technical Data on page 11 ).
c
9. Putting into Operation
Before placing the light into operation, the user has to check whether mains voltage and frequency specified on the rating plate are identical.
10. Care, Maintenance and Repair
Disconnect the light from the mains before carrying out any maintenance or repair work! This work must be carried out only by a qualified electrician!
As replacement parts only use part approved by the manufacturer.
10.1. Regular checks
The mains line must be checked minimum once a a year for damage (only MIRA 45 Floor Stand). In case of damage the manufacturer must replace the defective mains line.
10.2. Adjustment of the counter-balance (Fig. c)
Figure c
The counter-balance arm will already have been factory set for best possible operation. Should adjustment become necessary: Set the spring tension as illustrated, and using a clearance of 4 mm.
10.3. Changing the lamp
Note: Let the halogen lamp cool down before changing it!
1. Lift the hinged protective back cover of the head (see Fig. c)
2. The required Allen wrench is located inside the back cover (see Fig. d).
Remove the lamp head retaining screw (see Fig. e)
3. Remove the cover ring and reflector
4. Change the halogen lamp (use a soft cloth and do not touch the glass
surface of either the old or new lamp, see Fig. f)
5. Replace the reflector an cover ring. Replace an tighten the screws.
6. Replace the Allen wrench (see Fig, d)
Note: Only use lamp types approved of by the manufacturer
Putting into Operation
Operating manual (GB)
9. Putting into Operation
Before placing the light into operation, the user has to check whether mains voltage and frequency specified on the rating plate are identical.
10. Care, Maintenance and Repair
Disconnect the light from the mains before carrying out any maintenance or repair work! This work must be carried out only by a qualified electrician!
As replacement parts only use part approved by the manufacturer.
10.1. Regular checks
The mains line must be checked minimum once a a year for damage (only MIRA 45 Floor Stand). In case of damage the manufacturer must replace the defective mains line.
10.2. Adjustment of the counter-balance (Fig. c)
The counter-balance arm will already have been factory set for best possible operation. Should adjustment become necessary: Set the spring tension as illustrated, and using a clearance of 4 mm.
10.3. Changing the lamp
Note: Let the halogen lamp cool down before changing it!
Before placing the light into operation, the user has to check whether mains voltage and frequency specied on the rating plate are identical.
Care, Maintenance and Repair
Disconnect the light from the mains before carrying out any maintenance or repair work! This work must be carried out only by a qualied electrician!
As replacement parts only use part approved by the manufacturer.
Regular Checks
The mains line must be checked minimum once a year for damage (only MIRA 45 Floor Stand). In case of damage the manufacturer must replace the defective mains line.
Figure c
c
Adjustment of the counter-balance (Fig. 1)
The counter-balance arm will already have been factory set for best possible operation. Should adjustment become necessary: Set the spring tension as illustrated, and using a clearance of 4 mm.
Changing the Lamp
Note: Let the halogen lamp cool down before changing it!
1. Lift the hinged protective back cover of the head (see Fig. 1)
2. The required Allen wrench is located inside the back cover (see Fig. 2). Remove the lamp head retaining screw (see Fig.3)
3. Remove the cover ring and reector
4. Change the halogen lamp (use a soft cloth and do not touch the glass surface of either the old or new lamp, see Fig. 4)
5. Replace the reector an cover ring. Replace an tighten the screws.
6. Replace the Allen wrench (see Fig. 2)
Note: Only use lamp types approved of by the manufacturer (see Technical Data on page
11)
d
e c
f
www.amico.com 9
Page 10
10.4. Troubleshooting
1. Check the on/off Switch
2. Check the electrical supply
3. Check the halogen lamp, change if necessary
4. Check all terminal connections in the fixture Also check the voltage on the connecting terminal.
10.5. Cleaning
Note: The periodical cleaning of the face-plate ensure a better light efficiency. Use a usual glass cleaner.
Troubleshooting, Cleaning and Disinfection
Troubleshooting
1. Check the on/o Switch
2. Check the electrical supply
3. Check the halogen lamp, change if necessary
4. Check all terminal connections in the xture Also check the voltage on the connecting terminal.
Cleaning
Note: The periodical cleaning of the face-plate ensure a better light eciency. Use a usual glass cleaner.
Do not use concentrated disinfectant or dissolver. For cleaning the outer surfaces, use a cloth moistened with usual cleaner. Observe the compatibility with varnishes an plastics of the used cleaners.
Disinfection
According to instruction diluted brands can used for cleaning the outer surfaces:
• Lysoformin®
• Dismozon®
• Hexaquart® plus
• Sagrotan® – Rapid disinfection cleaner
Observe all applicable local regulations.
The disinfection cleaner are applied only for wipe disinfection. Spraying disinfection is not allowed.
Sterilization
The grip can be sterilized for 30 minutes at 120° and 1.05 bar.
10 Amico Lighting Solutions
Page 11
11. Technical Data
Electrical values:
Nominal voltage:
120 V ~
Frequency range
50/60 Hz
11. Technical Data
Electrical values:
Nominal voltage:
120 V ~
Frequency range
50/60 Hz
Power consumption:
50 W
Integrated electronic transformer
120 V / 12 V
Luminous values:
Central illuminance (0.8 m)
45’000 lux
Central illuminance (1.0 m)
40'000 lux
Light field diameter 0.8 m D10 = 15cm Colour temperature
ca. 4000 K
Colour rendering index
Ra 90
Ambient conditions for transport, storage and operating:
Ambient temperature (transport and storage)
-20°C to +70°C
Ambient temperature (operating)
0°C to +35°C
relative humidity
max. 90%
Fitted with:
Halogen lamp
1 x 12 V / 50 W
Type
QT12 50 W
Classification:
Protection class (
MIRA 45 Floor Stan
II
Protection class (MIRA 45 Wall Mount, MIRA 45 Single Ceiling, MIRA 45 Double Ceiling)
I
Int. protection according to IEC 529
IP 20
Class I
Safety inspection according
EN 60601-1
IEC 601-1
UL 601-1
12. Disposal
At the end of its service life, the lamp and the halogen lamps must be disposed of in accordance with local regulations.
Technical Data and Disposal
Technical Data
Electrical values:
Nominal voltage: 120 V ~
Frequency range 50/60 Hz
Power consumption: 50 W
Integrated electronic transformer 120 V / 12 V
Luminous values:
Central illuminance (0.8 m) 45’000 lux
Central illuminance (1.0 m) ≥ 40’000 lux
Light eld diameter 0.8 m D10 Ø = 15 cm
Colour temperature ca. 4000 K
Colour rendering index Ra 90
Ambient conditions for transport, storage and operating:
Ambient temperature (transport and storage) -20°C to +70°C
Ambient temperature (operating) 0°C to +35°C
Relative humidity max. 90%
Fitted with:
Halogen lamp 1 x 12 V / 50 W
Type QT12 50 W
Classication:
Protection class (MIRA 45 Floor Stan II
Protection class (MIRA 45 Wall Mount, MIRA 45 Single Ceiling, MIRA 45 Double Ceiling) I
Int. protection according to IEC 529 IP 20
Rating according to 93/42 EWG - Appendix IX Class I
Safety inspection according EN 60601-1
IEC 601-1
UL 601-1
Disposal
At the end of its service life, the lamp and the halogen lamps must be disposed of in accordance with local regulations.
www.amico.com 11
Page 12
www.amico.com
Amico Lighting Solutions | www.amico.com
85 Fulton Way, Richmond Hill
Ontario, L4B 2N4, Canada
71 East Industry Court, Deer Park NY 11729, U.S.A
Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778
Fax: 905.763.8587 Email: info@amico.com
C US LISTED
INSTALL INSTR. MIRA 45 MAR 2012
Loading...