Amico LDR Headwall User Manual

Installation Instructions
Reection Series
LDR (Labor/Delivery/Recovery) Headwall
Corporation
The installation instructions provided are for the Amico standard products. Slight modications to the installations may be required. The product shipped to the job site may not be exactly as shown on the installation drawings.
Before Installation
Note
Headwall installation should conform to local building codes
1.
Review a copy of the nal approved shop drawing(s) and/or submittal package. These documents will provide you with technical information specic to your installation, such as:
- equipment types and quantities
- room numbers and location
- shop drawings of each unit type
- wiring diagrams and rough-in locations
2.
Locate the headwall unit from the shipping boxes that you wish to install. The contents of each shipping box are listed on the shipping document(s).
1 Amico Architectural
Installation
Caution
Do not rest the cabinet on its movable bottom tray. Always lay cabinet on its back to avoid damage.
1.
Secure the LDR metal frame to the wall structure. Use hardware (by others) compliant to applicable building codes. Holes in mounting frame are to be drilled according to wall structure
(stud) location. Back plate is also recommended.
studs
drywall
electrical termination box (see detail B)
2.
Make all rough-in connections between wall and LDR frame for piping and electrical services. Electrical services are terminated at junction boxes, and gas terminations are located according to rough-in drawings (provided) to facilitate installation.
med gas termination (see detail A)
frame assembly
Please see details A and B for reference.
www.amico.com 2
Detail A
Figure 1
3 Amico Architectural
Detail B
www.amico.com 4
3.
Secure the LDR hanging bracket to the wall (see Detail below).
Hang the cabinet over the frame. Center the cabinet on the frame. Cabinet openings on the left
4. and rights sides should match with boxes on the frame.
Make all connections between light (if applicable) and bottom cabinet tray micro switches.
5. Wiring is labeled to facilitate electrical connections.
5 Amico Architectural
6.
Have all front access panels for your unit ready. Pull all electrical devices out of the boxes. Put one access panel on the cabinet and match the cut-outs with the device positions. Repeat for
the other panels. Adjust cabinet position, if needed, to ensure proper alignment.
Holding the front panel in the correct position, surface mount the devices in place and then
7. secure the cover to the cabinet. Drill pilot holes and attach with hardware provided.
8.
Install device cover plates.
9.
Insert trim caps where necessary.
10.
Install any accessories using hardware provided (if applicable).
www.amico.com 6
www.amico.com
Amico Architectural | www.amico.com
85 Fulton Way, Richmond Hill
Ontario, L4B 2N4, Canada
71 East Industry Court, Deer Park NY 11729, U.S.A
Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778
Fax: 905.763.8587 Email: info@amico.com
C US LISTED
INSTALL INSTR. REFLECTION SERIES LDR 2011
Loading...