Product Speci cations
The Amico retro- t alarm will allow an existing Medæs integrated alarm to be replaced, without removing the existing
back box and pipe-stubs.
The ller frame enables the Amico Alert-2 Microprocessor alarm front assembly to sit on the wall of the existing alarm
panel box. The ller frame acts like a spacer between the integrated box and the front assembly. Please make sure that
the gas is turned o before the retro- t alarm is installed.
Installation Instructions
Remove all internal components of the integrated alarm (mother board,
power supply, sensors and all wiring). Remove the front frame and the
display modules from the back box.
Ensure that the back box is clear of ALL screws, studs, stand-o s and insert
tabs. Bend the lower front lip of the alarm box. Ensure that the lip is at.
Add the power supply to the far left corner of the back box assembly. Use
4 tapping screws to hold power supply into box. Use tapping screws for
grounding wires. Please note, tapping screws are not supplied by Amico.
Remove the DISS tting from the sensors that are mounted remotely.
Connect the Amico Alert-2 sensor, which contains a DISS tting (nut and
nipple).
Add the Amico ller frame to the wall surface. Ensure that the box on the
wall is in the center of the ller frame.
Install the front assembly and follow the “Installation & Operation Manual of
the Alert-2 Alarm”. Also supplied with the Retro-Fit alarm.
# 22 Gauge twisted
pair shielded cable
OXYGEN
6”-8” [0.1m-0.2m] supplied
2 Amico Pipeline

A complete alarm assembly conversion from Madaes to Amico.
Product Dimensions
Inch
[mm]
Front View
Filler Box
Filler Frame
Number of Gangs A
3 12 (305)
4 15 (381) 16 (406)
5 18 (465) 19.3 (490)
7 25 (635) 25.9 (658)
12
[304.8]
A
B
(Filler Box)B(Filler Frame)
10.25
[260.35]
11
[279.40]
Side View
LCD Display
Module
Filler Frame
Filler Box
Existing
Back Box
NOTE: No filler frame is required when converting 3 gang alarms.
Filler Frame Part Numbers
A3P-LCD-FILLER-4 Alert-3 LCD Conversion Alarm Filler Frame 4-Station
A3P-LCD-FILLER-5 Alert-3 LCD Conversion Alarm Filler Frame 5-Station
A3P-LCD-FILLER-7 Alert-3 LCD Conversion Alarm Filler Frame 7-Station
NOTE: Alert-2 Sensor Modules are not compatible with the Alert-3 LCD Alarm.
www.amico.com 3

www.amico.com
Amico Pipeline | www.amico.com
85 Fulton Way, Richmond Hill
Ontario, L
4B 2N4, Canada
71 East Industry Court, Deer Park
NY 11729, U.S.A
Toll Free Tel: 1.877.264.2697
Toll Free Fax: 1.866.440.4986
Tel: 905.763.7778
Fax: 905.763.8587
Email: info@amico.com