Amico Ceiling Column Motorized Retractable User Manual

Approval and Installation Booklet
Motorized Retractable Ceiling Column
www.amico.com
Table of Contents
Product Description 3
Cleaning 3
Inspection and Testing 3
Range of Column Retraction 4
Device Placement (Bottom View) 6
Electrical Diagram (Bottom View) 7
Riser Plate to Outlet Detail 8
Termination Box 9
Riser Pipe Assembly 10
Mounting of the Ceiling Column 11
Mounting Information 12
Typical Location View 13
Wiring Diagram 13
Remote (Optional) 14
Control Unit 15 - 16
Linear Drive 17 - 18
Servicing 19
*NOTE: Amico requires all pages to be signed and dated before fabrication will begin.
2 Amico Pipeline
Product Description
The motorized retractable ceiling column shall be an Amico Alert-1 series.
The motorized retractable ceiling column shall be of the motorized operated type. The column shall consist of an upper section for rigid mounting at the ceiling level and a telescopic lower section.
The lower section of the column is activated by a motorized actuator mounted into the internal shell of the column. With the aid of a wall switch or with a head mounted switch, the column can be raised or lowered to any desired position. With the aid of an internal tracking system, the movement of the telescoping portion will glide freely up or down.
The upper and lower section shrouds are made of 16 gauge stainless steel with a # 4 satin  nish and are complete with: Removable access panel, stainless steel ceiling collar and a heavy gauge steel mounting plate, equipped with hose assemblies (gas speci c) within the internal shell of the column and brazed type outlets for all medical gas connections above the ceiling line.
All electrical and medical gas services are provided to the face of the lower section by means of  exible hoses and conduits. All threaded connections comply with NFPA, CGA and DISS recommendations preventing interchanging of connections. All services are pre-assembled and factory tested. Standard dimensions are 13.56” x 13.56” [344.5mm x 344.5mm] for the upper section and 12” x 12” [304.8mm x 304.8mm] for the lower section.
NOTE: Electrical devices are pre-wired, unless otherwise stated.
Cleaning
The Amico Outlets are factory cleaned for oxygen service. Exposed surfaces of the outlet may be cleaned with a mild detergent solution or wiped with a disinfectant commonly used in patient rooms that is compatible with plastics, anodized aluminium and die cast zinc.
Inspection and Testing
Medical Gas Outlets should be inspected periodically or at least once a year. The test should be in accordance with NFPA 99 “Gas and Vacuum systems”, or CSA Z7396.1-06 “Non ammable Medical Gas Piping System”.
Test for leaks: Ensure that no leaks exist, with or without the adapter inserted. Test for Indexing: Only a mating gas speci c adapter should insert smoothly into the outlet, latch and be retained. Test for Flow:
- Gas Outlets: 120 l/min (4.2 scfm) @ 345 kPa (50 psi), maximum allowable pressure drop is 28 kPa (4 psi).
- Nitrogen Outlet: 400 l/min (14.1 scfm) @ 1,250 kPa (180 psi), maximum allowable pressure drop is 70 kPa (10 psi).
- Vacuum Outlet: 30 l/min (1.1 scfm) @ 54 kPa (16 inHg), maximum allowable pressure drop is 13 kPa (4 inHg).
Refer to the appropriate standards for the proper way of performing the  ow test.
Note: The Amico medical gas and vacuum outlets meet and exceed these requirements at the time of manufacture. However piping source capacity, sizing and restrictions may prevent outlets from attaining these values.
www.amico.com 3
Range of Column Retraction
To accommodate for di erent ceiling heights, Amico Corporation has standardized three di erent variations.
The LOW height is designed for ceilings below 9’ - 0” (2743mm) 10” (254mm) Retraction Only. The MEDIUM height is designed for ceilings between 9’ - 1” (2769mm) and 9’ - 11” (3023mm) 14 “ (355.6mm) Retraction Only. The HIGH height is designed for ceilings above 10’ - 0” (3048mm) 18” (457.2mm) Retraction Only.
Please select the nominal column height below that applies to you application.
LOW MEDIUM HIGH
CEILING HEIGHT-LOW
78
1981.2mm]
1625.6mm]
62
108"
[2743mm]
30
762mm]
1117.6mm]
44
114
[2895.6mm]
CEILING HEIGHT-MEDIUM
78
1981.2mm]
60
1524mm]
36
914.4mm]
1384mm]
54
120"
[3048mm]
CEILING HEIGHT-HIGH
42
1066.8mm]
60
1524mm]
78
1981.2mm]
60
1524mm]
4 Amico Pipeline
FINISHED FLOOR LINE
Approval Signature Date Phone No.
Device and Accessory Chart
*QD
XXX
Ohmeda / Medaes
Compatible
CH
XXX
Chemetron Compatible
PB
XXX
Puritan-Bennett
Compatible
*DI
XXX
DISS
PE
Ceiling Column
Passive Evacuation
(19mm/33mm)
IMP
Passive Evacuation
with Impedance
V al ve (19mm /2 2mm)
(Side Mount Only)
GJ
Ground Jack
Receptacle
Single Hampden
(SLR-3) GREEN
GJ2
Ground Jack
Receptacle
Dual Hampden
(SLR-3) GREEN
E9-RED
E10-IVO.
Duplex Receptacle
8300-HG
125V/20a
(US / International)
E11-RED E12-IVO.
ngle Receptacle
Si
8310-HG
125V/20a
(US / International)
E5-RED E6-IVO.
Duplex Receptacle
8200-HG
125V/15a
(CAN / International)
E7-RED E8-IVO.
Single Receptacle
8210-HG
125V/15a
(CAN / International)
E13-RED
E14-IVO.
TL
Oxequip
Compatible
SL
OX
XXX
Monitor Jack Duplex
(5-pin shown)
Amphenol Model
10-825806-05S
97-14S-5S
IV
MJ
E13C-RED
E14C-IVO.
Ground Fault Circuit
Interrupter
(C = Canadian)
DJ DP
Vacuum
Bottle Slide
ALL ELECTRICAL DEVICES ARE HUBBELL BRAND
NOTE: CARBON DIOXIDE OUTLETS ARE NOT AVAILABLE IN PURITAIN-BENNETT. NITROGEN OUTLETS ARE ONLY AVAILABLE IN DISS.
NOTE: PASSIVE EVACUATION FOR RETRACTABLE WILL PROVIDE 240” [6096mm] LENGTH OF 0.75” [19.05mm] PVC HOSE FOR CONNECTION IN CEIILNG.
Stai nless Steel
IV Hook
(Duplex only)
Data - Telephone
Communication Jack or
Duplex Receptacle
(Provision Only)
DP - Duplex Receptacle
Provision
Twist-Lock
Receptacle
Hubble 23000-HG
125V/20a
Approval Signature Date Phone No.
www.amico.com 5
Device Placement (Bottom View)
Please fill the devices that you require for your application. (Specify by using the symbols on Page 5 of this Approval Booklet.)
Is a Local Switch Required?
* if YES, please provide the location (ex. B1)
Yes o
No
Nitrogen Control Panel
Yes o
No
Location of Nitrogen Control Panel
Side A
Side B
Side C
Side D
Please Specify Outlet Language
US
CAN
FRENCH
SPANISH/US SPANISH/ISO
Gas Legend
Oxygen = O
Medical Air = A
Vacuum = V
Nitrous Oxide = 2
Nitrogen = N
Carbon Dioxide = C
WAGD (US) = W
AGSS (Canada) = E
NOTE:
1. Two electrical/communication/medical gas devices cannot be placed together at any corner.
2. Maximum of 12 gas outlets are allowed on the column. If 12 gas outlets are required, then side mount is optional.
3. All electrical devices must be located within section A, B, C, or D only.
4. The Nitrogen Control Panel will occupy three spaces of the bulkhead (E.G. A2 to A4 see Outline).
6 Amico Pipeline
Approval Signature Date Phone No.
Electrical Diagram (Bottom View)
12 gauge
12 gauge
Black
White
10 gauge
Green
*Please Select One of the Following:
Amico recommended circuit schedule based on layout
Custom circuit schedule. Please specify on circuit chart
Receptacle Type:
US/CAN
*BRITISH (simplex only)
*GERMAN (simplex only)
*Please specify electrical details:
Type of wire:
NOTE:
Circuit Chart
THHN XHHW RW90
AWG Wire Gauge: Wire Color:
10 & 12 LIVE/HOT
10 & 12 NEUTRAL
10 GROUND
Isolated Power
Please select one of the following
All receptacles shall be mounted with ground pin “DOWN”
All receptacles shall be mounted with ground pin “UP” (Default)
1. Two electrical/communication/medical gas devices cannot be placed together at any corner.
2. If type of wire & gauge is not specified, Amico will use the following:
For US/International
LIVE/HOT - 12 Ga. THHN Wire Black NEUTRAL - 12 Ga. THHN Wire White GROUND - 10 Ga. THHN Wire Green
For Canada
LIVE/HOT - 12 Ga. RW90 Wire Black NEUTRAL - 12 Ga. RW90 Wire White GROUND - 10 Ga. RW90 Wire Green
3. Conduit: 9/16” [14.28mm]  exible metal conduit.
4. All conduits and wires will extend a minimum of12” [304.8mm] above  nish column for connection to riser plate.
5. A pull-string will be provided to be connected by others for all ground jacks and data provision.
6. British or German Receptacle cannot be positioned side by side ie: A1 and A2.
Approval Signature Date Phone No.
www.amico.com 7
Riser Plate to Outlet Detail C - Riser
VIEW COLUMN
Access Panel This Side
Riser Plate Bottom View Looking Up
17”
[431.8mm]
pipe stub
4.5” [114.3 mm] above
 nished ceiling
anti-sway support
(supplied by others)
17”
[431.8mm]
14.75”
[374.65mm]
For DISS riser pipe inlet (see page 9)
for column X4
14.75”
[374.65mm]
anti-sway support
(supplied by others)
7/8” [22.2mm] hole X8
1.00”
[25.4mm]
1.00”
[25.4mm]
Mounting holes for threaded rod X4
1A
see note #4
14.75”
[374.65mm]
Access panel This side
Access Panel This Side
Side A
Access Panel This Side
Side A
C
BOTTOM VIEW COLUMN
Looking Up
BOTTOM VIEW COLUMN
D
Looking Up
Termination X4 box use see page 10
DISS
for electrical conduit use
X4 See page 9
2A
BD
BC
NOTE:
1. Riser plate is completed with DISS pipe stubs, typical for each gas (For DISS riser pipe inlet see page 11).
2. Pipe stubs diameters are 3/8” [9.97mm] ID and 1/2” [12.7mm] OD for all gases. Refer to Riser Pipe cut sheet.
3. Exact locations of gas terminations on riser plate will be determined after receiving this booklet.
Approval Signature Date Phone No.
8 Amico Pipeline
Termination Box
Is a riser mounted termination box required? See diagram below
(refer to page 8 for locations).
Ye s (Amico provides termination box - default)
No (Termination box will be supplied by others)
4.773”
[121.23mm]
2 5/8”
[66.68mm]
5 x O .250”
10 - 32 TAPPED HOLE
1 1/2”
[38.1mm]
2 9/16”
[65.09mm]
4”
[101.6mm]
3/4”
[19.05mm]
3/4”
[19.05mm]
3/4”
[19.05mm]
DETAIL A
1 1/4”
[31.75mm]
5/8”
[15.88mm]
[6.35mm]
1/4”
4 - R .344” [R 8.73mm]
4.125” [104.78mm]
.409” [10.39mm]
.061” [1.55mm]
[19.05mm]
1 1/2”
[38.1mm]
0 0
.344” [ 8.74mm]
.376” [9.55mm]
R 1/16” [1.59mm]
1 3/16”
[30.16mm]
2 3/8”
[60.33mm]
4.125”
[104.78mm]
.203” [5.16mm]
4”
[101.6mm]
.629”
[15.97mm]
.394”
[10.02mm]
.188” [ 4.78mm]
0
0
.331” [8.41mm]
[17.09mm]
R 3/16” [17.09mm]
1.47”
[0.058mm]
.188”
[4.78mm]
.673”
.246”
[6.25mm]
8 - 32 UNC - 2B TAPPED HOLE
DETAIL A
3/4”
1 1/16”
11x 1/2” [12.7mm] KO, RECESSED 2 EACH SIDE, 3 IN BOTTOM
1/4”
[6.35mm]
1 1/8”
[28.575mm]
[26.99mm]
5/8”
[15.88mm]
7/64”
[2.769mm]
17/32”
[13.487mm]
8x 0.171” [4.34mm] 2 EACH SIDE
3/4”
[19.05mm]
6x 3/4” [19.05mm] KO, RECESSED 1 EACH SIDE, 2 IN BOTTOM
NOTE:
1. Maximum of four, Termination Boxes can be mounted on a riser.
Approval Signature Date Phone No.
www.amico.com 9
Riser Pipe Assembly
Ye s (Amico provides termination box - default)
No (Termination box will be supplied by others)
NOTE:
1. “X” represent US/ISO and GAS type i.e: C-PIPE-CHK-U-AIR
2. 3/8” [9.97mm] ID 1/2” [12.7mm] OD Type “K” Copper Pipe
C-PIPE-CHK-X-OXY
1/4”
[6.35mm]
(OXYGEN)
CAUTION
Before brazing the Riser an intermittent blow-out purge
*
is required to be done until all debris are cleared from the Pipeline System.
*
Installations of the Riser should be done in accordance with current NFPA 99 and CSA Z7396.1.
*
Risers with Demand checks should avoid excess or direct heat near its brass end.
C-PIPE-CHK-X-XXX (AIR, N2O, CO2, NIT)
C-PIPE-CHK-X-VAC
C-CHKVLV-OXY
3/16”
[4.76mm]
C-CHKVLV-GAS
3/16”
[4.76mm]
C-CHKVLV-VAC
7/32”
[5.56mm]
(AIR, N
0, CO2 & NIT)
2
C-CHKVLV-EVA
1/4”
[6.35mm]
C-CHKVLV-IAR
C-PIPE-CHK-X-XXX (WAGD, AGSS)
(VACUUM)
C-PIPE-CHK-X-IAR
(EVACUATION)
(INSTRUMENT AIR)
10 Amico Pipeline
Approval Signature Date Phone No.
Mounting of the Ceiling Column
There will be four (4) 1/2” - 13 x 8” threaded rods attached to the riser plate. Each rod contains four (4) 1/2” nuts, four (4) 1/2” washers and four (4) nylon retaining washer.
Remove the outer and inner access panels from the ceiling column.
NOTE: The gas hose(s) and electrical conduit(s) have been wrapped and attached to the top of the ceiling column to maximize the smooth movement of the lower telescopic portion of the ceiling column.
Remember the location of the attached electrical conduit(s) at the top of the ceiling column. Do not cut free the bundle(s) of gas hose(s) and/or electrical conduit(s) inside the ceiling column. Cut the cable tie(s) holding the gas hose(s) at the top of the ceiling column and pull the gas hose(s) into the ceiling column. Do not cut the cable tie(s) holding the electrical conduit(s) in place at the top of the ceiling column unless there is absolutely no room for the installer to attach the conduit(s) to the appropriate junction box(es) on the riser plate. Cut and remove the excess length of conduit(s) that is not required to attach to the appropriate junction box(es).
NOTE: The smooth movement of the retractable ceiling column will be a ected, if the above notes are not followed.
Remove the (4) nuts and  at washers from the bottom of the riser plate and hoist the ceiling column onto the riser plate. Ensure the open access panel of the ceiling column is aligned with the access panel label on the riser plate. Secure the ceiling column into place with the (4) washers and nuts. Ensure the ceiling column is level. Attach the conduit(s) to the appropriate junction box(es) with the ceiling column fully extended. Ensure the electrical conduit(s) are coiling around the handle shaft with very little slack as it heads up towards the top of the ceiling column. Attach the gas hoses to the appropriate outlets in the riser plate. Slowly retract the lower telescoping portion of the ceiling column and ensure that the electrical conduit(s) and gas hoses do not interfere with the smooth movement of the ceiling column as it retracts. Extend and retract the lower telescoping portion of the ceiling column a few times and ensure that the electrical conduit(s) and gas hoses do not get pinched by the springs, the lower telescoping portion itself or the riser plate mounting bolts.
Remove the plastic  lm from the access plates.
Attach the inner access plates onto the lower telescoping portion of the ceiling column and test the movement of the ceiling column to ensure that the movement is smooth. Attach the outer access plate onto the ceiling column.
Remove the plastic  lm from the mounted ceiling column.
Attach the (4) ceiling trim plates at the top of the ceiling column.
Approval Signature Date Phone No.
www.amico.com 11
Mounting Information
NOTE: Riser plate to be installed in ceiling with appropriate mounting structure and anti-sway sup­port. The methods of support shown are provided for information only.
17” Sq.
[431.8mm Sq.]
14.75” Sq.
[374.65mm Sq.]
STRUCTURAL CEILING
ANTI-SWAY SUPPORT STRUCTURE (SUPPLIED BY OTHERS)
1.00” [25.4mm]
4.5”
[114.3mm]
FINISHED CEILING
COLUMN
MOUNTING
PLATE
9.5” Sq.
[241.3mm Sq.]
RISER PLATE
17” Sq.
[431.8mm Sq.]
CEILING COVER
(FIELD MOUNTED)
3/4” [19.5mm] or 7/8” [22.23mm] THREADED ROD (X4) (SUPPLIED BY OTHERS)
MIN. 13” [330.2mm] MAX. 16” SQ CEIILNG CUT-OUT FOR COLUMN
1/2” [12.7mm] DIA. MOUNTING BOLTS FUR­NISHED WITH RUOGH-IN PLATE. TYP. OF 4, See note 1.
NOTE:
1. All labour and material for installation of Riser to Mounting Structure supplied by others.
2. General Contractor has  nal responsibility for the strength and stability of the mounting structure that is
supplied by others.
3. Riser plate to be mounted 4.5” (114.3mm) above  nished ceiling.
4. Mounting of column unit to riser plate by (4) 1/2” threaded rods, provided by Amico.
5. Space between top of Riser Plate and structural ceiling must be minimum 10”(254mm).
6. Riser Pipe(s) extend 8” (203.2mm) above riser plate.
7. Pre-ship of the riser assembly is available upon customer’s request.
Approval Signature Date Phone No.
12 Amico Pipeline
Typical Location View
Wiring Diagram
Recommended location of columns in surgeries and special care areas.
3’ - 4’
[914.4 - 1219.2mm]
3' to 4'
[914.4 - 1219.2mm]
Access Panel
3' to 4'
Access Panel
3’ - 4’
[914.4 - 1219.2mm]
OPERATING
TABLE
3’ - 4’
3' to 4’
[914.4 - 1219.2mm]
OPERATING ROOM
3' to 4'
3’ - 4’
*Conduit and wiring to internal junction box by others. Flexible conduit must be used. Allow a minimum of 24” (609.6mm) of wiring to internal junction box for extension and retraction of column.
Approval Signature Date Phone No.
www.amico.com 13
Remote (Optional)
Terminal Connections
RJ45
Model Number
Remote = C-EL-SWITCH-W
Approval Signature Date Phone No.
14 Amico Pipeline
Control Unit Speci cations
The control unit shall be designed to operate a maximum of 4 linear drives or telescopic columns in 24 V direct cur­rent technology.
It shall have an integrated transformer and recti er with plug connection for the main cable and hand-held or foot switch. It will also have up to 3 motor connections or 2 motors and 2 limits with switch connections.
The switch shall be an optimal hand-held or foot switch or an infrared receiver that can be plugged in directly at the control element connection. All control, and motor supply cables are locked after correct connection and sealed by sealing rings.
When in no-load operation, the Amico control unit shall be the auxiliary transformer which is active and maintain only the control voltage for the operating element (standby circuit). The transformer shall have thermal protection to avoid overheating.
The control box shall be ultrasonically welded.
Amico products comply with NFPA-99 and CSA Z7396.1.
low voltage control unit
protected against overheating
• maintenance free
minimum electromagnetic interference  elds radiated.
total current cut-o prevents exceeding of current consumption and protects drives.
Motor Connection 3
Voltage/Frequency 120 / 60 V / HZ
Output Power 24 V DC / 6 A
Duty Cycle: Intermittent Mode 1 min. / 9 min. int.
Ambient Temperature 0 / +40C
Degree of Protection 66 IP
Weight 2.8 kg
Approval Signature Date Phone No.
www.amico.com 15
Control Unit Dimensions
7.05”
[179mm]
KOM 3/4 CONTROL UNIT KOM 3/4 CONNECTIONS
6.50”
[165mm]
6.57”
[167mm]
[228.6mm]
[153mm]
[139mm]
9”
6.02”
5.5”
3.15”
[80mm]
4.21”
[107mm]
5.5”
0
[139mm]
13.3”
9”
[337.82mm]
2.68”
[228.6mm]
[68mm]
9”
[228.6mm]
2.95” [75mm]
Mains connections
Operating element
connection
Motor connections
External limit switch
(optional)
Model Number
Control Unit (power cord included) = C-EL-CNTROL-120V
16 Amico Pipeline
Approval Signature Date Phone No.
Linear Drive Speci cations
The linear actuators shall be designed for seating and reclining operations. It adjusts the column to the required level by raising and lowering it.
The drive is provided by a DC motor with worm geat which acts on a trapexoidal threaded spindle drive or ball screw system.
The actuators shall be protected against overheating by the Magnetic control unit.
Amico actuators shall use a push / pull force. They shall have a self-locking function which is activated by a built-in­brake. A safety nut shall also prevent overload of the actuator.
All forces shall be transmitted through die-cast metal parts.
Amico products comply with NFPA-99 and CSA Z7396.1.
high-quality die-cast metal parts, stainless steel push tubes & ultrasonic-welded plastic parts guarantee high performance
customizable stroke & mounting lengths in increments of 1.97 - 15.75” [50 - 400mm] & increments of 3.94” [100mm] over the range of 15.75 - 27.56” [400 - 700mm]; other options are available upon request.
maintenance free with a long lifetime
push force us reduced by 50% when operated in parallel with the control unit.
quiet in running
Push / pull force 1.5N
Speed at 24 V constant 14 mm/s
Stroke 1.97 - 27.56” [50-700 mm]
Voltage 24 DC
Power Consumption 120 W
Current Consumption 24V 5 A
Duty Cycle (ON / OFF) 1 min / 9min. int.
Ambient Temperature 0 / +40 C
Degree of Protection IP66
Protective Tube 1.69” [43 mm]
Weight 3 kg
Approval Signature Date Phone No.
www.amico.com 17
Linear Drive Dimensions
4.21”
[107mm]
1.89”
[48mm]
1.57” [40mm]
6.65” [169mm]
2.6”
[66mm]
&RONTVIEW
3IDEVIEW
0.39”
0.98” [10mm]
[25mm]
1.02” [26mm]
O 0.47” +0.01” [0.15mm]
[12mm] +0.002” [0.05mm]
O 0.39” +0.01” [0.3mm]
0.34”
[8.7mm]
[10mm] +0.01” [0.2mm]
" EARING INSERT FORFORKHEAD 4OPVIEW
1.67”
1.10”
[42.5mm]
[28mm]
1.10”
0.59”
[15mm]
0.43”
[28mm]
O 0.25” [6.3mm]
[11mm]
+0.01” [ 0.3mm] O 0.47” [12mm] +0.004” [0.1mm]
RL = min. 7.68” [195mm] + stroke 0.08” [2mm]
0.47”
[12mm]
1.10” [28mm]
+0.01” [ 0.3mm] O 0.47” [12mm] +0.004” [0.1mm]
2.6”
[66mm]
+
_
4.57”
[116mm]
0.55”
[14mm]
0.59”
[15mm]
0.43” [11mm]
1.10” [28mm]
Model Number
Linear Drive = 30" [762mm] Column: C-X-ELEC-CYL-10
36" [914.4mm] Column: C-X-ELEC-CYL-14
42" [1066.8mm] Column: C-X-ELEC-CYL-18
Approval Signature Date Phone No.
18 Amico Pipeline
Servicing
WARNING
Always shut o source pressure at designated zone valves before servicing any gas outlet.
The periodic inspection of gas outlets and electrical outlet is recommended to ensure proper safety and operation.
A) Gas Outlet
* Inspect the gas outlet connection points for the following:
1. Proper and legible identi cation.
2. Worn or defective engagement mechanisms.
3. Leaks.
4. Deformed or damaged threaded connections.
5. Loose fasteners.
WARNING
Use care when wiping down electrical outlets to prevent electrical shock.
B) Electrical
* Inspect the electrical outlet connection points for the following:
1. Proper and legible identi cation.
2. Deformation or damage
3. Discoloration
Approval Signature Date Phone No.
www.amico.com 19
www.amico.com
Amico Architectural | www.amico.com