Ametek T-5, T-10, T-15, T-20, T-30 Specifications

...
p r e s s u r e
Specification Sheet
SS-CP-2150-US
Model Type T
Hydraulic Deadweight Tester
Pressure Range
100 to 10,000 kPa (10 to 15,000 psi)
Accuracy
to ±0.015% of Indicated reading Accuracy ±0.025 and 0.100% is also available
Repeatebility
±0.005% of Indicated reading
Available units
psi, kg/cm2, bar, kPa
Flexible configurations to meet dif­fering applications
- single column for field use
- dual column for lab use
Robust construction allows for repeated daily use
Proven in design and application to meet your needs
Complete tester with tools, cases, and fittings
PRODUCT DESCRIPTION
The M&G Type T Hydraulic Deadweight Tester offers laboratory accuracy in an instrument that is designed to withstand an industrial environment. These testers are designed to be primary pressure standards and are available in several ranges and engineering units
Laboratory accuracy
The M&G Type T Deadweight tester is available in ranges to 1,000 bar (15,000 psi). This robust tester has been engineered to withstand the stresses of daily operation in a manufacturing environment and still maintain its accuracy and repeatability. The Type T is constructed of 300 series stainless steel and Monel and is supplied with Buna N O-rings as a standard unit. Viton and EPT O-rings are optional. This deadweight can use a water/alcohol mix or hydraulic oils for the pressure medium.
Traceable certification supplied with each tester
ISO 9001 Manufacturer
Model T
®
Hydraulic Deadweight Tester
Specification Sheet
SS-2150-T
FEATURES
Three accuracies available
These instruments are available in ±0.100%, ±0.025%, and ±0.015% of reading. The accuracy stated is the over­all accuracy of the tester. The accuracy takes into account linearity, hysteresis, and repeatability. It would also account for intrinsic corrections. Site corrections are the user’s responsibility and the unit will not perform as speci­fied without them. The testers are manufactured to either International Standard Gravity of 9.80665 m/s2 or to a user specified local gravity.
Re-entrant piston / cylinders
The Type T deadweight employs a re-entrant type pis­ton/cylinder assembly. This design applies test fluid to the outside and inside of the cylinder. Unlike a simple piston/ cylinder, this configuration reduces clearance between the piston and cylinder as pressure increases. This reduces the rate of fluid leakage and increases float time, offers the technician more time for testing, and reduces the amount of pumping necessary to sustain the fluid loss.
Re-entrant system
Configurations
There are two important factors in the selection of the proper deadweight configuration for your application: ranges and location. If you have applications for a single range of tests, then a single piston/cylinder model is a good choice. However, if you have some high pressure and some low pressure testing, the Type T is available with dual piston/ cylinder assemblies that are interchangeable and operate using the same weight set (±0.100% and ±0.025% accu­racies) or utilize dedicated weight sets for each piston/cyl­inder assembly (±0.015% accuracy). If you have the need to use the tester in a location other than a lab, you may need a smaller and more mobile con­figuration. The single column configuration will be your best selection. These are available with dual piston/cylin­der assemblies. For the applications where the tester will be stationary, the dual column configuration is a good choice for those needing multiple ranges. These are supplied with two piston/cylinder assemblies and both columns are installed and ready for use. A selector valve is used to determine which column is in service at any given time.
Worldwide engineering units
The M&G Type T deadweight tester can be manufactured in four different engineering units: psi, bar, kgt/cm2, and kPa. All of the different engineering units are available in any of the configurations.
Pressure media
Because the Type T is manufactured using 300 series stainless steel and Monel, you can use either hydraulic fluids compatible with the material or a water/alcohol mixture. We also offer EPT and Viton® o-rings as options. This increases your media options. Additionally, we offer an isolation membrane to protect your piston/cylinder assemblies from abrasive media. This facilitates the use of other fluids as process media but allows for the benefits of MGAAA oil within the column, which preserves your cali­brated parts.
2 www.ametekmg.com
Model T
Weight carrier
Weights Fluid reservoirCheck valves
Deadweight piston & cylinder Instrument under test
Venier Pump
Test manifold
®
Hydraulic Deadweight Tester
Overhung weight carriers
Another feature of Type T deadweight tester is the way that the weights are positioned on the carrier. M&G utilizes an overhung weight carrier design. This design employs a tube carrier that is positioned over the column and onto the piston driver. The center of gravity for the stack of weights is lowered, reducing side thrust and friction; which lengthens the life of the piston/ cylinder assembly. This also improves measurement accu­racy.
Specification Sheet
SS-2150-T
Protected piston/cylinder
The piston and cylinder assembly are enclosed in the col­umn of the tester. The weight carrier floats on the piston driver preventing damage to the piston/cylinder assembly. Additionally, the piston/cylinder assembly is engineered with positive over-pressure protection by restricting verti­cal movement. This protects the assembly from damage in the event of weight removal under pressure.
Dual volume pump
The Type T pump offers a dual volume pumping feature that allows for large volume of fluid to be added at low pressures to fill the system and increase pressure rapidly. Once the system is filled, you can switch the output vol­ume to a smaller volume. This eases pumping and also provides for finer control in higher pressure ranges. The reservoir is kept at atmospheric pressure meaning that you can also refill the tester during a calibration run.
Easy Leveling
AMETEK floating ball testers incorporate a bulls-eye level for reference when preparing the unit for use. The tester also employs a 3-leg leveling system which is more con­venient and is superior to a 4-leg system.
Small incremental weight sets
Small incremental weight sets are avaliable to provide fractional output pressures. These are available for the Type T deadweight tester in psi, bar, kPa and kg/cm2.