3.3DI POLISHER EXTERNAL FRONT VIEW ............................................................................................... 11
3.4STORAGE TANK ........................................................................................................................................ 12
4.3STORAGE TANK ........................................................................................................................................ 19
SECTION 5, SYSTEM START-UP ...................................................................................................................... 20
SECTION 14, SPARE PARTS LIST ................................................................................................................... 44
Page 4
1
SECTION 1, GENERAL INFORMATION
1.1 INTRODUCTION
This system is designed to pretreat and purify water for use in health care applications.
The system is shipped with required water treatment components. If the system will be
running without softened water, the PT401 system will need to be connected to the side of the
RO. This Operation Manual was written specifically for the RO3 & RO4 models, storage tank
and DI polisher. Your system was thoroughly tested and in excellent condition when it was
shipped to you. However, because damage during shipment is possible, please unpack and
carefully inspect the system as soon as you receive it. Please notify AmeriWater if any
problems are encountered.
WARNING: This Reverse Osmosis System (RO) contains a preservative solution to
prevent microbiological growth and freezing. Discard all product water for
at least two hours of operation before placing the RO in service.
Please read the Operations Manual before using the system. Contact AmeriWater
Customer Service with any questions at 1-800-535-5585 Monday through Friday 8:00 a.m. to
5:00 p.m. eastern standard time. For after hours emergencies call 1-800-535-5585 and follow
the instructions on the recorded message. Our on-call technician will return your call as soon
as possible. This entire Operations Manual should be read before operating or servicing the
system. This Operations Manual should then be kept near the system and used as a
reference and troubleshooting guide.
All of the above listed materials meet FDA and/or NSF standards.
Life Expectancy of Components:
The 2 Carbon Cartridge Filters will last for up to 150 hours depending upon the content of the
incoming water. These should be changed whenever chlorine breakthrough occurs.
The PT401 anti-scalant should be refilled every 150 hours as well.
DI resin packs should be changed whenever the quality indicator light changes from green to
red.
Safety Features
The RO is equipped with several safety features for the benefit the user. They consist of the
following:
Color-coded inlets and outlets are on the membrane assemblies to avoid mix-ups.
INCOMING TAP WATER, PRODUCT WATER, and REJECT WATER TO DRAIN hoses are
labeled to prevent incorrect connections.
An audible alarm sounds whenever water quality drops to an unacceptable level.
98-2005 Rev. G
Page 5
2
1.2 ELETRICAL LEAKAGE STANDARDS
The AmeriWater RO3 & RO4 water treatment systems comply with the National Electrical
Standards for Product Safety and Construction.
The cabinet of the RO is PVC for additional operator safety.
The RO3 & RO4 is compliant with ANSI/AAMI ESI-1993, Safe Current Limits for Electro
medical Apparatus. All major components of the RO (controller, pump, solenoid valve,
antiscalant pump) as well as other components are UL listed.
This product has been tested to the requirements of CAN/CSA-C22.2 No. 61010-1,
second edition, including Amendment 1, or a later version of the same standard
incorporating the same level of testing requirements.
Product marked with this symbol indicates that it is protected throughout by double
insulation or reinforced insulation.
98-2005 Rev. G
Page 6
3
1.3 CAUTIONARY SYMBOLS
Caution, risk of electrical shock!
Attention, risque de choc électrique!
Open by qualified service personnel only!
Ouverture par le personnel qualifié seulement!
Refer to this Operation and Maintenance Manual for instructions and
safety considerations. Référez-vous au manuel des Opérations et
Entretien pour instructions et mesures de sécurité.
Caution, risk of danger!
Attention, danger potentiel!
For service by qualified service personnel only!
Entretien par le personnel qualifié seulement!
120 15
Replace with 120Vac, 15amp, time-delay fuse only.
Remplacer avec 120Vac, 15 amp, fusible à retardement
seulement.
Earth Ground terminal
Borne de mise à la terre
98-2005 Rev. G
Page 7
4
SECTION 2, TECHNICAL INFORMATION
2.1 THEORY OF OPERATION
RO:
The process of osmosis can be reversed by placing pressure upon the feed water side
(concentrated solution side) of the membrane. Water will be forced through the membrane
barrier to yield water that is purer on the lower pressure side of the membrane than on the
more concentrated solution side (higher pressure side) of the membrane. The feed water will
become more “concentrated,” and will be discharged through the reject port known as “reject
water” or “concentrate”.
Hence, the liberation of purer water from its solutions is caused by the reversal of the osmotic
pressure; the operation is termed “Reverse Osmosis”. Reverse Osmosis is commonly
referred to as “RO”.
DI Polisher:
This system is designed to polish the RO product water in order to provide purified water
used in laboratory applications. The facility’s tap water first passes through the RO, then
enters the first deionizer column (worker) where the deionizer resin removes chemical
contaminants from the water through an ion exchange process. The water then passes into
the “polisher” column. As water leaves the polisher, the purified water passes through a
200,000-Ohm resistivity monitor with alarm and on to the storage tank.
Storage Tank:
The storage tank is equipped with 2 float switches, one which will turn the reverse osmosis
unit off when the water reaches the predetermined cutoff point and another to turn the RO on
when the tank is drained to the appropriate level float. The distribution pump on the storage
tank will cycle whenever there is a pressure drop detected. A flow switch has been wired into
the pump to act as a kill switch to prevent the pump from operating when there is no water
flow.
98-2005 Rev. G
Page 8
5
2.2 SPECIFICATIONS
Ideal, minimum, and maximum
incoming water temperature
Prefilter gauge pressure (when the
RO is running)
Minimum
Maximum
Pump pressure –
Minimum
Maximum
Maximum output of product water
@ 77°F (25°C), TDS<1000 ppm of
NaCl, & pump pressure of 150 psi.
Connections
Electrical Requirements
Min = 41° F (5° C) Max = 90° F (33° C)
Ideal Temperature = 77° F (25° C)
20 PSI
50 PSI (Pounds per Square Inch)
150 PSI
230 PSI
RO3 – 4,200 GPD (15,897 LPD)
RO4 – 5,600 GPD (21,196 LPD)
(Gallons Per Day / Liters Per Day)
RO Feed =3/4” FGH (Female Garden Hose)
RO Product = 3/4" MGH (Male Garden Hose)
RO Drain = 1/2" Hose Barb
Distribution Pump 1” FPT & MPT
(1) X 115V/60Hz/15A dedicated GFI (Ground Fault
Interrupter) for RO controller & DI Polisher
(2) X 220V/60Hz/15A dedicated outlets
System Dimensions 105" L x 36” D x 68" H
Shipping Weight
Operating Weight
Power Ratings
Controller 115 60 Hz Single 10
Pump 220 60 Hz Single 9.8
650 lbs.
1200 lbs.
Voltage Frequency Phase Amp Draw
98-2005 Rev. G
Page 9
6
2.3 SPECIFIC ENVIRONMENTAL/TRANSPORT CONDITIONS ANTICIPATED
ENVIRONMENTAL CONDITIONS ANTICIPATED
This device is intended to be used under the following conditions:
Indoor use;
Altitude up to 2000 m;
Temperature between 5°C and 40°C;
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%
relative humidity at 40°C;
MAINS supply voltage fluctuations up to + 10% of the nominal voltage;
Transient overvoltages present on MAINS supply = CATEGORY II;
Applicable RATED POLLUTION degree 2.
TRANSPORT CONDITIONS ANTICIPATED
Altitude up to 2000 m;
Temperature between 5°C and 40°C;
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%
relative humidity at 40°C;
2.4 TEMPERATURE CORRECTED RO PRODUCTION RATES
Membrane performance is affected by water temperature. The Product Water Flow Rate and
Output decrease as the temperature of the Incoming Tap Water decreases.
PT401 Antiscalant/Scale Inhibitor
Contaminants
Chemical Feed
System
Membrane Scale Control Not to exceed 40 ppm
98-2005 Rev. G
Page 10
7
SECTION 3, COMPONENTS AND SCHEMATICS
3.1 RO EXTERNAL FRONT VIEW
98-2005 Rev. G
Page 11
8
IDENTIFICATION OF COMPONENTS (EXTERNAL VIEW)
1. SIDE ENTRY HOODS: External connection for float level switches (4 pin connection)
and PT Pump power connection (5 pin connection).
2. PREFILTER INLET GAUGE – Gauge that measures the pressure in pounds per square
inch (PSI) of the incoming tap water as it enters the carbon block prefilters.
3. CARBON CARTRIDGE FILTERS – Dual 10-micron carbon cartridges for removal of
chlorine.
4. PRODUCT WATER – Hose transmitting purified water from the RO system to the DI
Polisher or Storage Tank.
5. PT401 FEED PUMP – Injects PT401 Antiscalant / Scale inhibitor solution at a
predetermined dosage based on a water analysis. The PT401 pump runs when the
main pump runs.
6. PT401 ANTISCALANT / SCALE INHIBITOR – Plastic container filled with PT401
solution to prevent the RO membranes from scaling. The plastic container should be
refilled with PT401 solution at least monthly or when on when it reaches the half-full
level.
7. INCOMING TAP WATER – Hose feeding tap water into the RO system.
8. REJECT WATER TO DRAIN – Hose transmitting wastewater to the drain.
9. CHLORINE SAMPLE PORT – Valve with nozzle to let small amounts of water out to test
for the presence of chlorines before the RO membranes.
10. PAA QUICK CONNECT FITTING (RO) – Quick connect fitting that the PAA container is
connected to on the RO for sanitization.
11. PRODUCT WATER SAMPLE PORT – Valve with nozzle to let small amounts of water
out to test the quality of the product water.
12. PREFILTER OUTLET GAUGE – Gauge that measures the pressure in pounds per
square inch (PSI) of the water after going through the carbon cartridge filters. Change
the filters when the outlet gauge reads 15 PSI less than the inlet gauge.
13. PUMP PRESSURE – Gauge that measures the primary feed pressure in pounds per
square inch (PSI) from the pump to the RO membranes.
14. REJECT PRESSURE – Gauge that measures the pressure in pounds per square inch
(PSI) from the reject of the membranes.
15. CONTROLLER – Control mechanism for the RO.
16. RECIRCULATION GPM – Flowmeter that indicates the flow of water that is recirculated
back to the pump’s inlet. A valve is incorporated into the flowmeter to allow adjustment.
17. REJECTION GPM – Flowmeter that indicates the flow of water that is rejected to the
drain. A valve is incorporated into the flowmeter to allow adjustment.
18. PRODUCT GPM – Flowmeter that measures the flow of the product water in gallons per
minute (GPM) and liters per minute (LPM).
98-2005 Rev. G
Page 12
9
3.2 RO INTERNAL REAR VIEW
98-2005 Rev. G
Page 13
10
IDENTIFICATION OF COMPONENTS (INTERNAL VIEW)
19. PRODUCT CONDUCTIVITY CELL – Measures the conductivity of the product water.
20. CLEAN IN PLACE (CIP) SWITCH - When the CIP Switch is placed in the ON position,
all RO fail-safe modes are disabled for low pressure membrane cleaning with the
optional AmeriWater Clean In Place System (P/N 00CIP1).
21. INLET PRESSURE SWITCH – Pressure switch to protect the pump from running dry in
the event that water pressure drops too low.
22. INLET CONDUCTIVITY CELL – Measures the conductivity of the incoming water.
23. DISINFECT SOLENOID – Solenoid valve that opens during the disinfect cycle to allow
disinfectant to be drawn into the RO.
24. INLET SOLENOID – Solenoid valve that opens during normal operation to allow flow of
water into the device.
25. PUMP – Provides the pressure for the RO system. The RO ON/OFF switch controls the
pump motor.
27. MOTOR CONTACTOR / OVERLOAD RELAY - Starts and stops the pump motor and
protects the motor from overload.
98-2005 Rev. G
Page 14
11
3.3 DI POLISHER EXTERNAL FRONT VIEW
IDENTIFICATION OF DI POLISHER COMPONENTS
28. AIR VENT SCREWS – Vents provided to break the vacuum seal to facilitate resin
changes.
29. WATER INLET – Inlet for water from the RO.
30. 200,000 OHMS QUALITY LIGHT – Light with alarm that will indicate when the water is
outside of the limit.
31. DRAIN VALVES – To allow for draining each column of the DI Polisher individually.
98-2005 Rev. G
Page 15
12
3.4 STORAGE TANK
EXTERNAL FRONT VIEW
98-2005 Rev. G
Page 16
13
IDENTIFICATION OF STORAGE TANK COMPONENTS (FRONT VIEW)
32. WATER INLET – Connection for the purified water into the tank.
33. PUMP PRESSURE GAUGE – Gauge that measures the pressure in pounds per square
inch (PSI) when the distribution pump is running.
34. FLOW SWITCH - Switch that prevents the pump from running when there is no flow.
35. DISTRIBUTION PUMP – Pump that is used to distribute water to the system whenever
the internal pressure drops or there is greater than 0.3 GPM flow.
36. SYSTEM OUTLET HOSE – Hose that is provided to change the system outlet from 1”
FPT to 1” MPT.
37. DISCHARGE ISOLATION VALVE – 1” FPT connection point for the loop.
38. BYPASS QUICK CONNECT – Used to allow untreated water to completely bypass the
system in the event of downtime.
39. RETURN FROM LOOP – Connection that can be utilized to allow the loop to re-circulate
back into the storage tank to minimize opportunities for bacteria growth. The return flow
control on this is 3 GPM.
98-2005 Rev. G
Page 17
14
STORAGE TANK EXTERNAL SIDE VIEW
IDENTIFICATION OF STORAGE TANK COMPONENTS (SIDE VIEW)
40. HIGH LEVEL FLOAT – Sends a signal to the RO when the tank is full to stop water
purification.
41. LOW LEVEL FLOAT – Sends a signal to the RO when the predetermined low level has
been breached to have the RO begin water purification.
42. SIDE ENTRY HOOD – Provides connection between storage tank floats and the RO.
98-2005 Rev. G
Page 18
15
3.5 ELECTRICAL DIAGRAM
3.6 SIDE ENTRY HOOD
The external wire connection for High/Low Tank signals has a maximum voltage of 24 V and a
maximum current of 500 mA. The connection allows signals from float switches in the storage
tank to be sent to dry contacts within the RO controller. The high tank float signals the RO to
shut off in a tank-full condition, and the tank low float signals the RO to turn on when the water
level in the storage tank is low.
The second external wire connection is a quick connection to provide power to the PT 401
pump that is supplied with this unit. This connection is 115VAC that can carry up to 10 amps.
These side entry connections cannot be mixed as they are each “keyed” for the appropriate
type of connection.
98-2005 Rev. G
Page 19
16
3.7 FLUID DIAGRAM
98-2005 Rev. G
Page 20
17
SECTION 4, INSTALLATION
CAUTION:Local plumbing and electrical codes must be observed.
4.1 RO INSTALLATION
1. The electrical source must be single phase, 3-conductor type provided with a hospital
grade receptacle and a ground fault interrupter (GFI) at 115V, 15amp, and 60Hz. The
proper polarity and ground integrity must be initially checked and thereafter maintained.
Failure to do so may result in electrical shock to the operator. It is suggested that the
RO be placed on an electrical supply with emergency backup. The pump for the R0
must be plugged into a dedicated 220V/60Hz/15A outlet.
2. The RO controller must only be plugged directly into a GFI receptacle. It must not be
plugged into an extension cord or power strip that could cause low amperage.
To avoid electrical shock, always unplug the RO system before opening the face
of the electrical controller.
3. Incoming water should be between 41° F and 90° F (5° C and 33° C). It is not
recommended to use water at temperatures below 41° F (5° C) as it will reduce
membrane performance significantly. Use only the cold water supply unless using an
automatic blending valve to get 77° F (25° C) water. Never use water warmer than 90°
F (33° C).
98-2005 Rev. G
Page 21
18
4. Water with silt density index (SDI) above 5 SDI will foul the membrane.
5. The RO system is equipped with a pretreatment system to remove chlorine. It is
important to test for chlorine at the CHLORINE SAMPLE PORT periodically. Chlorine
will deteriorate the membrane and cause system failure. It is recommended to use a
Chlorine test strip.
6. Incoming tap water pH should be within EPA National Secondary Drinking Water
Regulations of 6.5 - 8.5. Incoming tap water with pH higher or lower than the
regulation will cause higher conductivity in the product water. If the water changes
drastically, the membrane will be harder to clean. Periodically check the pH of the
incoming tap water to verify that it is within the specified range (IBT pH Water Test
Strips P/N 97PH20901).
7. Always maintain water flow and pressure to avoid damage to the pump.
8. Minimum feed pressure is 20 PSI @ 6 GPM (while the RO is in operation, with flow).
Maximum feed pressure is 50 PSI.
9. Lock the two front casters so that the device will remain stationary during startup.
10. Remove the caps from the PRODUCT WATER, REJECT WATER TO DRAIN, and
INCOMING TAP WATER fittings on the side of the RO.
11. Connect the INCOMING TAP WATER, PRODUCT WATER and REJECT WATER TO
DRAIN hoses to the appropriate fittings on the RO. All hoses should run to the rear of
the device. This will allow you to move the RO forward to access the rear panel.
12. If the supply water is not softened, you will need to install the PT 401 pump by
engaging fitting on the PT 401 pump with the 5 terminal side entry hood on the side of
the RO.
13. Plug the power cord into a dedicated 115-volt, 15-amp GFI receptacle.
14. Plug the pump power cord into a 220-volt, 15-amp receptacle.
4.2 DI POLISHER
1. Locate the DI Polisher to the right of the RO, just before the storage tank.
2. Route the product water hose from the RO behind the RO to connect to the DI Polisher
hose inlet after disinfection.
3. Plug the power cord into a 115-volt outlet.
98-2005 Rev. G
Page 22
19
4.3 STORAGE TANK
1. Locate the storage tank on a firm, level foundation. For seismic requirements, drill (4)
5/8” diameter holes into the concrete through the mounting holes in the storage tank
feet a minimum of 4” deep. Install (4) 5/8” diameter, HILTI KB-TZ Expansion anchors
through the 4 holes on the pads to anchor into the ground. A minimum of 4 threads for
each anchor must be below the concrete prior to application of 60 ft-lbs of torque.
2. Connect the DI Polisher outlet hose to the storage tank’s inlet.
3. Connect the discharge from the pump to the distribution piping.
4. Run the float – switch quick disconnect to the RO and lock in place.
5. Prime the pump by filling the tank with water and opening the valve into to the pump’s
inlet. Loosen the priming plug on the pump to allow the air to escape. Once water
starts to escape, re-tighten the priming plug.
6. Plug the pump directly into a dedicated 220-volt, 15-amp receptacle.
98-2005 Rev. G
Page 23
20
SECTION 5, SYSTEM START-UP
WARNING: This Reverse Osmosis System (RO) contains a preservative solution to
prevent microbiological growth and freezing. Discard all product water for
at least two hours of operation before placing the RO in service.
1. Connect the RO INCOMING TAP WATER hose to the potable cold water supply using
the Incoming Tap Water hose and fittings supplied. If blending both warm and cold
water to improve product flow rate, do not exceed 90° F (33° C).
2. The REJECT WATER TO DRAIN hose coming out of the RO is for reject water. The
water from this hose will go down the sink or drain. Leave at least a 2" air gap between
the hose and the drain to prevent contamination or siphoning.
3. The PRODUCT WATER hose should also be secured to the sink or drain.
4. Relieve the pressure from the system by opening the CHLORINE sample port.
5. Open the rear access cover and turn the CIP SWITCH to off.
6. Verify that the carbon cartridges are installed in the filter bowls.
7. Close the CHLORINE sample port and turn on the potable cold water supply to the RO.
Allow the membranes to completely fill with water.
8. Turn on the RO and turn the reject flow control knob counter clockwise to allow the
device to run in full flush for 15 minutes.
NOTE: The RO conductivity alarm may sound, which is normal when it is in FLUSH.
Press the ALARM SILENCE key on the RO controller to silence the alarm. The
alarm will restart after a 3 minute delay.
9. After flushing for 15 minutes, adjust the reject flow control knob so that there is an
equal amount of product and reject water passing through their respective flow meters.
10. Run the product water to drain for at least 2 HOURS to remove the preservative from
the system.
11. Verify that the PRODUCT GPM FLOWMETER indicates at least 2.00 GPM for a RO3
and 2.50 GPM for a RO4.
12. Refer to the Operation Log in Section 7.0 MONITORING. Complete the log; making
sure that the system is operating within all the required ranges.
WARNING: Do not use the RO until all specifications are met.
13. When the log shows that all start-up conditions are met, the system is now ready for
use. Press the POWER key (the display will show STANDBY). Connect the
PRODUCT WATER hose to the DI Polisher inlet.
98-2005 Rev. G
Page 24
21
14. Install the resin packs in the DI Polisher:
a) Lock the front wheels. Remove the cover
plates by unscrewing the four black knobs on
each plate. To break the seal between the
cover plates and the upper plates, use a
slotted screw driver in the pry notch located in
the top of each upper plate. Open the drain
valve at the bottom of each column.
b) Remove the resin pack from the shipping box
and open the tied end of the plastic sleeve.
DO NOT remove the resin pack from the plastic
sleeve at this time.
c) Stretch the resin pack over your arm
to elongate the pack, and feed it into
the column allowing the pack to slide
out of the plastic sleeve. Save the
plastic bag for future exchange.
d) Verify that the O-ring is clean and in
place and replace the cover plates.
Tighten the black knobs evenly by
turning the knobs at opposite
corners at the same time. Repeat
for the other two corners.
CAUTION:Failure to tighten the knobs completely may cause the cover plate to
separate from the system when pressurized.
e) Close the drain valve at the bottom of the column.
AIR VENT
f) Repeat steps a. through e. for each of the columns.
15. Open the air vents located on the top of the cover plates and
turn on the water supply and the RO. Tighten the air vents
closed when water begins to escape through the vents.
16. Ensure that the DI Polisher quality light is plugged into a GFI
receptacle.
17. When the resin is changed, the light may indicate red
initially, but should switch to green, indicating good quality.
18. Lock the casters on the DI Polisher.
19. Verify that the distribution pump is plugged into a 220V/60Hz/15A dedicated outlet.
98-2005 Rev. G
Page 25
22
20. To adjust the pressure that the pump will deliver, you can
depress the up or down arrows on the distribution pumps
control panel. This is located on top of the pump, under the
storage tank.
5.1 PT401 PRIMING PROCEDURE
AmeriWater RO products featuring the PT401 antiscalant system are equipped with a priming
valve. The following procedure should be followed during the initial start up of the RO system,
and whenever the PT401 pump loses its prime:
1. Verify that the RO is on and running.
2. Be sure the injection pump’s suction line is immersed in solution in the PT401
container.
3. Press the ON key to turn on the PT401 injection pump.
4. Press the “DOWN” key to cycle the STROKE RATE to approximately 360.
5. Remove the lid form the container and verify that there is discharge of PT401.
6. Close the adjustment knob completely by turning clockwise until the knob stops,
making sure there is no liquid flow out of the air bleed tubing.
If the pump does not prime, repeat the foregoing
steps.
Once the pump has been primed, and is pumping the
chemical through the head into the water stream,
adjust the stroke rate to “10”.
At the stroke rate of 10, the PT401 injection pump will
put about 300 milliliters (approximately 10 liquid
ounces) per hour into the incoming water stream.
Repeat these steps as necessary when the system is
started after sitting for extended periods of time or the
PT401 Bottle is empty and air has drawn into the
pump. The priming may not take as long as the initial
time.
WARNING: Overdosing PT401 will adversely affect (blind) the membrane causing a
decrease and/or loss of permeate flow.
WARNING: If softened water is supplied, turn off the PT401 system by depressing the
on/off button on the chemical feed pump and emptying the PT401 jug.
98-2005 Rev. G
Page 26
23
5.2 REFILLING PT401
HARDNESS (Grains Per Gallon)
STROKE SETTINGS
The amount of PT401 in the container should be checked weekly. When it is less than half
full, you should add more PT401.
Remove the cap from the PT401 Container. It is not necessary to remove the rubber stopper
with the hose; this will break the prime of the PT401 pump. Refill the container with new
PT401 purchased from AmeriWater (P/N 95810125, sold in (4) 1 gallon containers). Close
the container with the cap. If the PT401 Container is run dry, follow the priming procedures
above.
5.3 PT401 ANTISCALANT SETTINGS
The digital readout on the PT401 injection pump can be adjusted from 0 to 360 strokes per
minute. Based on the hardness of the feed water, the following stroke settings are
recommended:
1-6 GPG 9
7-20 GPG 11
21-40 GPG 13
For water with hardness greater than 40 GPG, contact AmeriWater for consultation
1-800-535-5585
98-2005 Rev. G
Page 27
24
SECTION 6, OPERATION
6.1 OPERATION
Once the system has been started, the RO will continue to make water until the high-level
float switch is tripped on the storage tank. At this point, the RO will be placed into standby
(tank full) until the low-level float switch is tripped.
Once the low level float is tripped, the RO will re-initialize and begin to produce water again.
In the event that the conductivity is above the pre-determined set-point, an alarm will sound.
The storage tank will still operate as normal.
6.2 BYPASS PROCEDURE
1. Turn off the POWER to the RO and Distribution Pump.
2. Shut off the water supply to the system.
3. Remove the RO water inlet hose from the source.
4. Retrieve the bypass hose and install to the incoming water supply.
5. Connect the CPC connector into the fitting just after the pump on the storage tank.
6. Turn on the water to begin bypass operation.
98-2005 Rev. G
Page 28
25
SECTION 7, MONITORING
Fill out the monitoring log prior to each use. Having this information available will help to
quickly diagnose issues related to performance. Failure to carry out the daily monitoring and
maintenance at the indicated intervals will result in reduced performance of the RO system
and may void the warranty.
1) DATE
RO PERFORMANCE (DURING WATER FLOW)
2) Temperature (41ºF - 90ºF)
3) Product flow (RO3 = 2.00 – 4.00 GPM)
(RO4 = 2.50 – 4.50 GPM)
4) Pump PSI (100 - 230 PSI)
5) Filter Pressure Drop (maximum of 15 PSI)
6) Hour Meter Reading
7) Chlorine Test (maximum 0.1 PPM)
8) Product Water Quality Reading
9) Verify light on DI Polisher is Green
10) Verify that the distribution pump is not in alarm
EXCHANGE
11) Carbon Block Filter Cartridges
12) Clean or Replace the Membrane
13) Add PT401 Antiscalant
14) DI Polisher Resin Change
INITIALS
98-2005 Rev. G
Page 29
26
SECTION 8, MAINTENANCE
WARNING: If any component of the water treatment system is changed or replaced,
the user should conduct appropriate tests to ensure that the revised
system meets all standards to which it was initially tested.
WARNING: Maintenance shall be performed by qualified personnel only.
1. AmeriWater has provided an Operation Log for the RO. The recorded information may
be useful in troubleshooting problems encountered with the system.
2. The DI resin, membrane and carbon cartridges are non-durable components and will
need to be exchanged periodically.
8.1 RO NORMAL MAINTENANCE ITEMS
Preparation
1. Turn off the RO and remove from the power source.
2. Turn off the incoming tap water supply.
3. Place a container under the CHLORINE TEST PORT. Slowly open the sample port to
relieve the pressure and let the water drain.
4. Remove the rear cover.
To Exchange the Carbon Block Filters:
1. Use the filter wrench to unscrew the filter housing and discard the used filter(s).
2. Partially unwrap the plastic from a brand new filter. Holding the end covered with
plastic, place the new filter in the housing. Discard the plastic after installation.
3. Screw the filter housing back on making sure the o-ring is in the groove and not
pinched. Hand-tighten, only.
To Exchange the Membrane:
1. Mark the location on each compression fitting prior to loosening the nut.
2. Disconnect the fittings that secure the tubing to the sides of the membrane.
3. Remove the Unistrut clamp holding the membrane assembly.
4. Lift out the membrane assembly.
5. Loosen the nuts on the retaining clamp on the feed side to remove the cap and slide
old membrane out of the housing and discard.
98-2005 Rev. G
Page 30
27
6. Install the new membrane in the housing with the brine seal at the Inlet end of the
housing, and replace the cap and retaining clamp.
7. Connect the feed (brown) fittings at the bottom of the new membrane housing in the
same way the system was initially connected. Secure the new membrane assembly
into place, and connect the remaining fittings at top of membrane housing (Reject =
red, Product = blue, Inlet = Brown).
8. Reconnect compression fittings and tighten them to the initial locations. You may need
to go slightly beyond this mark to get a good seal. Please use caution when doing this
as excessive tightening will thin the wall section of the tubing, which may lead to leaks
or a break.
9. Re-install the membrane using the Unistrut clamps.
10. Repeat this for each membrane to be changed.
Rinse Out Cycle
WARNING: This Reverse Osmosis (RO) membrane contains a preservative solution to
prevent microbiological growth and freezing. Discard all product water for
at least two hours of operation before placing the RO back into service.
After the exchange is complete, it is important to put the RO through a rinsing out cycle to
flush the preservative out of the new membrane.
1. Put the PRODUCT WATER hose at a sink or drain.
2. Turn on the RO and turn the reject flow control knob counter clockwise to allow the
device to run in full flush for 2 hours. This will rinse the preservative out of the new
membrane.
3. Adjust the reject flow control knob so that there is an equal amount of product and
reject water passing through their respective flow meters and allow the RO to run until
the conductivity is below the set-point and no longer alarming. Turn off the RO.
98-2005 Rev. G
Page 31
28
4. Reconnect the PRODUCT WATER hose to the DI Polisher.
5. Turn on the RO. The rinse out cycle is now complete, and the RO is ready for use.
8.2 DI POLISHER RESIN PACK EXCHANGE
1. Turn off the water supply to the RO system.
2. Open the air vents located in the center of the cover plates.
3. Place a container under the Drain Valves located on the bottom plates of the system.
4. Open the valves and allow the water to drain into the container. The container may be
emptied at a drain.
5. Remove the cover plates by unscrewing the black knobs located at each corner of the
cover plates.
6. Pull the exhausted resin pack part of the way out of the column and lay it over the top
of the system.
NOTE: If the Drain Valve located on the bottom plate is not open, a vacuum will be
created making it difficult to remove the resin pack.
7. Slowly pull the exhausted pack over the side of the column while sliding it out of the
column and into the plastic sleeve that it was shipped in.
8. Remove the new resin pack from the shipping box and open the tied end of the plastic
sleeve.
9. Stretch the resin pack over your arm to elongate the pack.
10. Feed the pack into the column allowing it to slide out of the plastic sleeve. Save the
plastic sleeve for the return shipment of the resin pack at the next exchange.
11. Carefully wipe the O-rings and the grooves on the top plate and cover plate to remove
any debris from the surfaces that must seal together.
12. Replace the cover plates and tighten the knobs. To ensure proper tightening and a
good seal, hand-tighten the knobs in opposite corners simultaneously. Repeat for the
other two corners.
13. Turn on the water supply to the RO and remove trapped air by loosening the air vents
located in the center of the cover plates.
14. Tighten the air vents when water begins to escape through the vent. This ensures that
all of the air has been removed from the system and maximizes the life and efficiency
of the resin packs.
98-2005 Rev. G
Page 32
29
8.3 DI POLISHER RESIN PACK REGENERATION
Regeneration certificates are used to simplify the exchange of your exhausted DI Polisher
system packs. The certificate has been purchased and will be used as payment for the resin
pack regeneration service. Simply follow these steps:
Fill out a regeneration certificate by printing your name, phone number, company and
address.
When you send each pack to AmeriWater, enclose the regeneration certificate in the box.
AmeriWater will ship you a freshly regenerated resin pack.
To order more certificates, contact the company that originally sold you the deionizer. Be sure
to keep regeneration certificates on hand to ensure uninterrupted service.
It is recommended to send the exhausted resin to AmeriWater immediately after exchange.
This will allow time for shipment and will maintain your uninterrupted supply of deionized
water.
Certificates Make Exchanging Packs a Breeze!
Fill out and enclose in box for return shipment
• NO Paperwork
• NO Phone Calls
•NO Purchase Orders
98-2005 Rev. G
Page 33
30
SECTION 9, ALARMS
9.1 RO OPERATING STATUS MESSAGES & ALARMS
The operating status of the unit is shown on the top line of the display. The following list
describes the items shown for the operating status and alarms.
STANDBY - The unit is in the STANDBY mode.
DELAY 99 - The unit is in the RO start delay. The number is the seconds remaining before the
pump starts.
OPERATING - The RO unit is operating.
TANK FULL - The unit is shut down due to a tank full condition.
PERCENT REJECTION – Shows the % of TDS rejected by the RO.
TEMPERATURE – Indicated the temperature of the incoming water.
CONDUCTIVITY – Measurement of the quality of the water.
PRESS FAULT - If the pressure fault input becomes active (closes) and stays active for the
delay programmed in the PF Delay Setpoint, the unit will shut down for a pressure fault. The
display will show PRESS FAULT, the alarm lamp will flash and the audible alarm will sound.
The pressure fault can be cleared by pressing the Alarm Silence/Reset key twice.
MEMB FLUSH 99 – Membrane Flush is active. The number is the minutes remaining in the
flush cycle. This setting is not normally used.
HI COND - If the Conductivity reading exceeds the limit programmed the Cond Limit Setpoint
for the delay programmed in the Cond Delay Setpoint, the alarm lamp will light and the HI
COND warning message will show on the display. This warning will clear when the
Conductivity drops below the Setpoint.
9.2 DI POLISHER ALARMS
WATER QUALITY INDICATOR LIGHT – Will illuminate green when the water is within the
required parameters. When the light changes to red and there is an audible alarm during
normal product flow, this indicates that the resin packs will need to be exchanged. The
audible alarm can be silenced by depressing the detent on the face of the alarm.
98-2005 Rev. G
Page 34
31
SECTION 10, STANDARD SETPOINT TABLE
VIEWING THE SETPOINTS;
Press the LEFT ARROW key to view the set points.
Press the ALARM SILENCE/RESET key to cancel the operation.
SET POINT FACTORY SETTING
TDS / CONDUCTIVITY LIMIT 50
TDS / CONDUCTIVITY DELAY 10
RO START DELAY 10
PRESS FAULT DELAY 10
AUTO RESET (PSI) 0
ALARM SILENCE 3
TF RESTART DELAY 5
TF RESTART 0
TFO TIME 3
TANK LOW RESTART 0
FLUSH TYPE 0
FLUSH TIME 0
FLUSH INTERVAL 0
FLUSH MODE 0
MAXIMUM HOURS 0
CURRENT HOURS 0
EXPANDER MODE 0
TEMPERATURE OFFSET 0
TEMPERATURE UOM 0
SWITCH SELECT 0
TDS / CONDUCTIVITY UOM 0
TDS / CONDUCTIVITY RANGE 1
C2 RANGE 4
C2 LIMIT 0
% REJECTION 1
98-2005 Rev. G
Page 35
32
SECTION 11, OPTIONAL DISINFECTION
AmeriWater recommends that all HCRO systems should be disinfected at least annually. The
following sections will outline the disinfection steps for the RO, optional Silex deionizer and
Storage Tank.
11.1 RO DISINFECTION
1. Switch off the RO by pressing the POWER key (the display will show STANDBY).
2. Disconnect the RO PRODUCT WATER hose from the next point of use.
3. Place the PRODUCT WATER hose with the REJECT WATER TO DRAIN hose at a
sink. Secure PRODUCT WATER hose in place so that the entire length of the hose
will “flood” during disinfection.
4. Put on rubber gloves, apron, and goggles.
5. Unscrew the cap assembly of the PAA container.
6. Add 400 ml of 100% hydrogen peroxide/peroxyacetic acid disinfecting solution to the
PAA Container and fill with water to the red line. (You may use tap water or treated
water). Screw the cap assembly back on the PAA container.
CAUTION: Exposure to hydrogen peroxide/peroxyacetic acid (PAA) concentrate or
solution may cause severe chemical burns to skin or eyes. Additional
information regarding PAA is found in this section, on your PAA bottle, and
the Material Safety Data Sheet. Please read carefully before using.
98-2005 Rev. G
Page 36
33
7. Lightly shake or tilt the container to mix the solution for 10 seconds.
8. Connect the PAA tubing to the PAA connection on the front of the RO.
9. Turn the reject flow control knob counter clockwise to allow the RO to run in full flush.
10. Turn on the feed water supply to the RO.
11. Press and hold the LEFT ARROW key, and then, press the RIGHT ARROW key. This
will access the DISINFECT MODE. The controller display will show DISINFECT
ENABLED, and then the keys can be released.
12. Press and hold the ENTER key. Pressing and holding the ENTER key will turn on the
Disinfect Draw function and THE DISPLAY WILL READ DISINFECT ENABLED DRAW. Continue to hold the ENTER key, until all of the PAA is drawn in.
13. If the PAA tubing begins to suck air, immediately release the ENTER key. Adjust the
reject flow control knob so that there is an equal amount of product and reject water
passing through their respective flow meters. Avoid extra time running because PAA
will be diluted.
CAUTION: Avoid sucking air when the container is near empty.
NOTE: The ENTER key must be held until the PAA is drawn container is
nearly drawn empty into the RO.
A. AVOID STOPPING AND STARTING THE DISINFECT FUNCTION WHICH
MAY CAUSE A THERMAL OVERLOAD OF THE RO PUMP. IF A THERMAL
OVERLOAD OCCURS, IT WILL AUTOMATICALLY RESET FROM 1 TO 5
MINUTES – WAIT UNTIL THE PUMP OVERLOAD RESETS TO RESUME
THE PAA DRAW.
B. RELEASING THE ENTER KEY MAY CAUSE THE SYSTEM TO LOSE ITS
PRIME PREVENTING ALL OF THE DISINFECTANT FROM BEING DRAWN
INTO THE MRO.
C. APPROXIMATELY ¼” OF THE SOLUTION WILL REMAIN IN THE BOTTOM
OF THE PLASTIC CONTAINER. THIS IS NORMAL AND MAY BE EMPTIED
DOWN THE DRAIN AFTER THE ENTIRE PROCESS IS COMPLETE.
14. At this point, the RO is filled with the PAA disinfecting solution. To ensure that this has
passed through the system, use a PAA test strip (Peracid Test Strips, P/N 97HP20401)
at the REJECT WATER TO DRAIN hose; the results should be at least 1% (500 ppm).
Use another PAA test strip (Peracid Test Strips, P/N 97HP20401) at the PRODUCT
WATER hose; the result here should be at least 0.5% (250 ppm).
NOTE: You may need to press the ENTER key to sample the flow being discharged
from the REJECT WATER TO DRAIN and PRODUCT WATER hoses. If the test
strip does not show PAA disinfecting solution at the specified levels in the
Reject Water and Product Water, press the ENTER key again to continue flow;
repeat the PAA test until both test positive (at the specified levels). If the test
strip still does not show the disinfecting solution at the required levels, repeat
disinfect draw procedure until the tests show positive at the required levels.
98-2005 Rev. G
Page 37
34
15. Allow the PAA mixture to soak in the RO for at least 60 minutes.
WARNING: Soaking longer than twelve hours may cause damage to the membranes.
16. After the soaking time has elapsed, turn on the RO by pressing the ALARM/SILENCE
RESET to restart the RO in the operating mode. Allow the machine to run at its current
setting for at least 15 minutes.
17. After rinsing the RO in the flush mode for 15 minutes (minimum), turn the reject flow
control knob clockwise to allow the RO to run in operation for another 15 minutes.
18. Adjust the reject flow control knob to full flush for 15 additional minutes, followed by 15
additional minutes in normal operation mode.
19. Continue rinsing for until the water from the PRODUCT WATER hose, REJECT
WATER TO DRAIN hose and the product sample port test negative for PAA.
WARNING: Continue rinsing and testing until both test strips show a negative result
(no color change) to ensure that there is less than 3 PPM of PAA in the
system. AmeriWater recommends using Renal Check PX Test Strips (P/N
97PX20501).
20. Disinfection is now complete. Press the POWER key to turn off the RO. The display
will show STANDBY.
11.2 A WORD ABOUT HYDROGEN PEROXIDE/PEROXYACETIC ACID
Do not use hydrogen peroxide/peroxyacetic acid concentrate (PAA) after the expiration date.
Using outdated PAA concentrate may cause incomplete disinfection.
PAA loses effectiveness if not kept out of direct sunlight and/or the cap is not tightly sealed.
Using ineffective disinfecting solution will cause incomplete disinfection.
Using less than the required volume of PAA will result with incomplete disinfection.
Disposal of Outdated Hydrogen Peroxide/Peroxyacetic Acid:
Supplies Needed - a sink with a supply of tap water
- rubber gloves, lab apron, and goggles
- a supply of paper towels
1. Put on rubber gloves, apron and goggles.
CAUTION: Exposure to PAA concentrate or solution can cause severe chemical burns
to skin or eyes.
2. Start a flow of cold tap water to dilute the PAA as it flows down the sink drain.
3. Slowly and carefully pour the disinfecting solution down the drain, taking care to avoid
spills, splashes, or breathing the vapors.
98-2005 Rev. G
Page 38
35
4. Rinse the emptied PAA container with tap water to remove all traces of the chemical.
Rinsing emptied containers is needed to protect waste handlers from accidental
exposure to the chemical.
5. Rinse the sink surfaces with tap water to remove residual disinfecting solution and to
flush the chemical from the sink traps.
6. Discard the emptied and rinsed container in a waste receptacle or set aside for
recycling.
7. Inspect the area for spilled or dripped disinfecting solution. Wipe up small spills with a
damp paper towel. Larger spills should be either flushed to drain with water or
removed with a water bucket and floor mop.
WARNING: Verify that there is no chlorine (bleach) in the water bucket or floor mop.
Chlorine (bleach) will cause a severe chemical reaction when it comes in
contact with hydrogen peroxide/peroxyacetic acid concentrate!
8. Rinse rubber gloves with tap water to remove any residues from handling.
9. Remove and return rubber gloves, apron, and goggles to their storage area.
11.3 SILEX DISINFECTION
1. Turn off the water supply to the Silex system and remove the discharge from the
storage tank.
2. Remove the resin packs from the system.
3. Pour 5 oz (150 mL) of PAA into each Silex column.
4. Close all drain valves and sample ports and open the air vents located in the center of
the top plates.
5. Turn on the water supply to the Silex and allow it to fill with water until the presence of
PAA can be detected at the outlet of the system. Verify the concentration of PAA at the
outlet using a high level concentration PAA test strip (P/N 9778339-000).
6. Turn off the water supply and allow the system to soak for 1 hour.
7. After the soak time, drain the water out of the system by placing a container under the
drain valves and opening the valves.
8. Close all valves and turn on the water supply to the system.
9. Direct the product water to a drain and allow system to flush until PAA is no longer
detected. (Use low-level PAA test strips, P/N 9778338-000, to test for residual PAA).
10. Drain the water out of the system by placing a container under the drain valves and
opening the valves.
98-2005 Rev. G
Page 39
36
11. Close all drain valves and sample ports and open the air vents located in the center of
the top plates.
12. Replace the resin packs and return system to service.
11.4 TANK DISINFECTION
Storage Tank Disinfection Procedure:
1. The storage tank can be disinfected using household bleach or a peracetic acid
disinfectant (PAA).
WARNING: Never allow Chlorine (bleach) to come into contact with PAA. Doing so
will cause a severe chemical reaction!
a. For a 55 gallon tank, fill to the RO shut-off level mark and add 1 quart of
household bleach or PAA.
b. For a 100 gallon tank, fill to the RO shut-off level mark and add 2 quarts of
household bleach or PAA.
c. For a 185 gallon tank, fill to the RO shut-off level mark and add 4 quarts of
household bleach or PAA.
NOTE: The values above will provide a 250 PPM concentration of disinfectant. If
desired, the disinfectant can be doubled to provide a stronger kill (500 PPM).
However, when this is done, the rinse-up time will be significantly longer.
2a. If your system is installed with a recirculation loop back to the storage tank, recirculate
the system for 30 minutes.
2b. If your system is a demand feed (where the pump only operates when your equipment
calls for water), connect the bypass hose to the quick disconnect and place the other
end of the bypass hose into the storage tank, close the distribution valve and allow the
pump to recirculate for 30 minutes.
3. After recirculation, drain the tank, fill with RO water, and recirculate for 5 minutes.
4. Drain the tank, refill and recirculate for 5 an additional minutes. Use a chlorine test
strip to verify the residual chlorine level or a PAA test strip to verify the residual PAA
level.
5. Repeat step 3 and 4 until the test strip indicates a residual level at or below 1.0 ppm.
98-2005 Rev. G
Page 40
37
SECTION 12, CLEAN IN PLACE (CIP)
1. Move the CLEAN IN PLACE SWITCH (inside the RO
on the back of the controller) to the ON position.
To avoid electrical shock, always unplug the RO
system before opening the access panel or the
face of the electrical controller.
2. Place the FLUSH VALVE in the FLUSH position.
3. Connect the RO PRODUCT WATER hose to the
PRODUCT RETURN connection (female garden
hose thread) on the CIP Cleaning Drum.
4. Connect the RO REJECT TO DRAIN HOSE to the
REJECT RETURN connection (hose barb) on the
CIP Cleaning Drum.
5. Connect the RO INCOMING TAP
WATER HOSE to the CIP PUMP
OUTLET (male garden hose thread).
6. Leave the micron filter in the filter
housing, but remove the carbon block.
7. Add 1 cup of AmeriClean B (base
cleaning) to the CIP Cleaning Drum
and fill with RO water (if available).
Stir until dissolved.
⌧
8. Plug in the CIP Pump and turn on the
RO. Verify recirculation flow inside the
cleaning drum.
9. Allow the system to recirculate the cleaning solution for 30 minutes.
10. Turn off the RO and the CIP Pump. Allow the system to soak for 30 minutes.
11. Turn on the RO and CIP Pump and allow them to run an additional 5 minutes.
ALARM
ALARM
POWER
SILENCE/
ENTER
RESET
S150
12. Turn off the RO and CIP Pump.
13. Disconnect the RO REJECT TO DRAIN HOSE and direct it to a drain.
14. Turn on the RO and CIP Pump and run to drain until the CIP Cleaning Drum is
emptied.
15. Turn off the RO and CIP Pump and fill the CIP Cleaning Drum with approximately 2
98-2005 Rev. G
Page 41
38
gallons of clean water.
16. Connect the RO REJECT TO DRAIN HOSE to the REJECT RETURN connection on
the CIP Cleaning Drum.
17. Turn on the RO and CIP Pump and allow the water to recirculate for at least 1 minute.
18. Turn off the RO and CIP Pump and disconnect the REJECT TO DRAIN HOSE from the
CIP Cleaning Drum. Direct the REJECT TO DRAIN HOSE to a drain.
19. Turn on the RO and CIP Pump and continue running to drain until the CIP Cleaning
Drum is emptied.
CAUTION: Verify that the AmeriClean B has been thoroughly rinsed from the system
prior to proceeding to the next step. Mixing the chemicals may result in an
exothermic reaction that may damage the RO+ system.
20. Repeat steps 7 to 19 using AmeriClean A (acid cleaning) and then proceed to step 21.
21. Disconnect the RO hoses from the CIP system and run the PRODUCT WATER hose
and REJECT TO DRAIN hose to a drain. Connect the INCOMING TAP WATER hose
to a water supply.
22. Replace carbon block filter and replace micron filter, if necessary.
23. Turn on the feed water supply and the RO, and allow running for 10 minutes.
24. Turn the TEMPORARY FLUSH VALVE to the OPERATION position, and continue
running to drain until the conductivity drops below the setpoint (not in alarm).
WARNING: DO NOT place the RO in service until all operational parameters are within
specification. Consult the STARTUP LOG to verify that all specifications
are met.
98-2005 Rev. G
Page 42
39
SECTION 13, TROUBLESHOOTING GUIDE
WARNING: Only those persons who have read the complete operations manual or who
have received authorization from the medical facility director should
attempt to troubleshoot and/or repair the RO+ system.
NOTE: At times it may become necessary to replace parts on your RO. AmeriWater
stocks all parts and components for the system, and can ship the necessary
replacement parts to you if you call Customer Service at 1-800-535-5585
between 8:00 a.m. and 5:00 p.m. EST.
To assist you in quickly restoring your system into service, AmeriWater will
send your replacement part out immediately and check your bad part when it
comes in to verify if it is covered under your equipment warranty.
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
RO will not start RO not plugged in
Circuit breaker blown
RO in a FAULT condition
System has power but no
water flow
Feed source not open
Feed pressure < 20 PSI
Incoming hose kinked
Prefilter clogged
Plug into electrical outlet.
Reset the breaker.
Check RO controller display
for FAULT condition and
correct the FAULT.
Open Incoming Tap Water
valve.
Increase pressure to > 20
PSI.
Straighten kinks from the
INCOMING TAP WATER
hose.
Check the prefilter gauges for
pressure drop; replace the
carbon blocks if the pressure
drop is 15 PSI or greater.
Circuit board relay is not
operating
Feed solenoid is not
operating
Replace the controller circuit
board
Test the solenoid. Replace
the valve if it is defective.
98-2005 Rev. G
Page 43
40
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Low product flow rate Low pressure feeding
membrane
Low pump PSI
Reject GPM flow rate too high
Excessive PRODUCT line
backpressure
Low temperature incoming
tap water
Prefilter clogged
Verify that the incoming tap
water supply is fully open.
The pressure on the prefilter
gauges should be > 20 PSI
when the RO is operating.
Pump should be operating at
160 – 230 PSI.
Turn throttle valve to balance
with production flow.
Check for restrictions in the
PRODUCT WATER hose.
Consult the Temperature
Correction to determine if the
flow rate is normal in relation
to the feed water
temperature.
Check the prefilter gauges for
pressure drop. Replace the
prefilter and carbon block
cartridges if the pressure drop
is > 15 PSI.
Membrane needs replaced
Disinfect cycle will not
operate when holding the
DISINFECT MODE has not
been accessed correctly.
ENTER key
Circuit board relay not
operating in DISINFECT
MODE.
Disinfect Solenoid Valve not
operating
Pressure fault Low incoming pressure to the
system
Replace the membrane.
Access DISINFECT MODE
Replace the controller circuit
board.
Test solenoid valve. Replace
the valve if it fails.
Verify that the incoming
pressure to the system is
above the minimum
requirements.
98-2005 Rev. G
Page 44
41
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
a. Exchange all water
RO Pump making excessive
noise
Low pressure or flow rate
feeding the RO
Feed solenoid is not
operating
Pump motor or impeller failing
Poor quality product water High Chlorine levels
RO not rinsed thoroughly
Reject valve adjusted
incorrectly
Check the prefilter outlet
gauge PSI (must be > 20
PSI), and verify that the
product flow (flowmeter) > 0.4
GPM.
Test the solenoid. Replace
the valve if it is defective.
Check PUMP PSI GAUGE to
verify that it is within
operating parameters.
Replace the pump assembly
if necessary.
Replace the Carbon Cartridge
filter.
Rinse membrane.
Turn throttle valve to balance
with production flow.
Fouled membrane
Anti-scalant not working
treatment components.
b. Check and correct the
antiscalant system.
Verify the conductivity cell
accuracy with a known good
meter. Follow the calibration
procedures or replace cell if
necessary.
Verify anti-scalant system is
in operation.
98-2005 Rev. G
Page 45
42
RO CONTROLLER TROUBLESHOOTING
CAUTION: Hazardous voltages are present when power is applied to the unit. Care
should be taken when troubleshooting any of the input power or output
circuits. When disconnecting or connecting any board or accessory, be
sure power is unplugged.
Before contacting AmeriWater for technical help, verify the programming of all Setpoints,
check the display and check the status of all lights and indicators. The more information
available when you contact us, the easier it will be to determine the source of the problem.
Standard setpoints, and drawings of the controller and pc boards can be found in Section
10.0, RO Controller.
PROBLEM INVESTIGATION CORRECTIVE ACTION
System Inoperative Is the yellow CPU active LED
blinking?
If no, is the green power
LED, DS1 Lit?
If no, is the fuse OK?
If no, replace the fuse.
If yes, with a voltmeter, verify
power is applied to the power
terminals L1 and L2.
Display Blank Is the green power LED, DS1
lit?
If yes, is the CPU active LED,
DS9 blinking?
If yes, adjust the display
contrast adjustment, R3. Is
the display still blank?
If power is applied to the
power terminals and the
other checks are OK, the pc
board is defective and should
be replaced.
If no power is applied to the
board, check the power
wiring to the system.
If no, refer to the system
inoperative section.
If no, replace the board.
If yes, replace the board.
Inlet Valve Will Not Operate Is the system in standby?
If no, are any shut down
conditions active?
If no, is the inlet LED, DS8
lit?
If yes, with a voltmeter, verify
if there is power on the inlet
terminals. Is there power?
If no, replace the board.
If no, replace the board.
If yes, check the valve and
wiring.
98-2005 Rev. G
Page 46
43
PROBLEM INVESTIGATION CORRECTIVE ACTION
RO Pump Will Not Operate
(Cont.)
No or incorrect conductivity
reading
Is the system in standby?
If no, are any shut down
conditions active?
If no, is the RO LED, DS6 lit?
If yes, with a voltmeter, verify
if there is power on the RO
pump terminals. Is there
power?
Is sensor wired correctly?
If yes, is sensor installed
inline?
If yes, verify correct
Conductivity range. Range
correct?
Does unit calibrate OK?
If no, replace the board.
If no, replace the board.
If yes, check the pump and
wiring.
If no, correct wiring.
If no, install correctly.
If no, correct range.
If yes, calibrate unit.
If no, disconnect green and
white wires of sensor. Does
reading show 0?
If yes, reconnect wires and
remove sensor from piping
and dry. Does reading show
0?
If yes, short terminals of cell
together. Does reading show
‘^^^’ ?
If no, replace board.
If no, replace cell.
If no, replace board.
98-2005 Rev. G
Page 47
44
SECTION 14, SPARE PARTS LIST
Consumables
Part # Name
002101C Regeneration Certificate, Standard DI
20-5105 Carbon Block Filter (10 micron block, 4.5” X 20”)
95-0007 Peracidin Disinfectant, 4 Quarts
95810125 PT401 Antiscalant, 4 Gallons
97HP20401 High Range Peracetic Acid Test Strips
97PX20501 Low Range Peracetic Acid Test Strips
R22-4026 Membrane
Replacement Parts
Part # Name
0119-0007 Drain Port Assembly
0119-0008 Drain Tee Assembly
0119-0010 Column Assembly
13-1075 Distribution Tank O-Ring
19300001-00 Column O-Ring
19530901 Air Vent
19560310 Knob Assembly
21675175 Filter Housing(20”)
21530238 Filter Housing O-Ring
24-0026 Membrane Housing O-Ring Set
43-0019 0 – 100 PSI Pre-Filter Inlet Pressure Gauge
43821445 0 – 160 PSI Pressure Pre-Filter Outlet Gauge
43530714 0 – 300 PSI Pump / Reject Pressure Gauge
67511004 Floatswitch
66932109 Wire Harness for Solenoid Valves
75679118 Resistivity High Trans with Patch
75679122 Hydro-Check, 200K Light with Alarm
92-0042 Hilti Anchor
94-0013 Caster, Twin Wheel with Brake
94-0014 Caster, Twin Wheel no Brake
R59-0002 Normally Closed Solenoid Valve - Inlet
R59-0015 Normally Open Solenoid Valve – Disinfect Draw
R65-0014 Incoming Pressure Switch
R67-0014 Flow Switch, CME Distribution Pump
R69446010 Controller
R80-0148 CME Distribution Pump
R080-0003 RO Pump Assembly
R84-0002 PT401 Pump Assembly
98-2005 Rev. G
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.