1. GENERAL DESCRIPTION
The 24DT is an electronic digital thermometer which displays the
current temperature and also logs the maximum and minimum
temperatures experienced. With the use of an Optional momentary
switch the min. / max. Temperatures can be displayed at the touch
of a button and reset if required. (Switch sold separately) A 9 Volt
battery can supply backup power for temperature readings when
there is a blackout. (Battery bracket and harness sold separately)
2. INSTALLATION
The instrument is equipped with fast-on type spade terminals with
probe connections separated from the power supply connections.
To install the instrument proceed as follow:
1. Attach .250 inch Fast-on connectors to the wires for the interior
light and the wires connecting the switches.
2. Wire the switch according to the drawing Fig. 1.
3. Place the probe where it has to measure the temperature. Be
sure to leave slack so as to have some strain relief. Probe bulb
should have the end up vertically or on at least a slight angle to
try to prevent water intrusion into the sensor. Were a sharp
bend is required try to loop the probe wire in a 1 ½ “circle.
4. Connect the probe terminals to the two small (.110 inch) faston connectors as indicated.
5. Connect the power supply cables to the large (.250”) fastons. (N or NEUTRAL) to the fast-on on the left and (L or
HOT) to the fast-on on the right. SEE FIG. 2
6. There should be .63mA fuse on the Line.
7. If using the optional 9V battery back up install it on the
connectors at the top left of the thermometer.
3. DRAWINGS
Fig.1
Fig. 2
N L
NOTE: POWER SUPPLY AS STATED ON UNIT
LABEL!
4. ALARM DISPLAY
MESSAGE CAUSE
PO
P1
OPEN PROBE OR SWITCH CLOSED
NO PROBE CONNECTED
SAFETY PRECAUTIONS
Always ensure that the probe is connected to the instrument
before the power supply is connected and turned ON.
Check the supply voltage is correct before connecting the
instrument. (12V, 120V, 230V) (N or NEUTRAL) to the fast-on
on the left and (L or HOT) to the fast-on on the right. SEE
FIG. 2
The probe is double insulated for safety on models with 110V
and 230V power supply.
The instrument is designed for panel mounting and electrical
connections must be positioned inside a properly protected
board/panel.
Do not expose to water or moisture: use the instrument only
within the operating limits avoiding sudden temperature changes
with high atmospheric humidity to prevent the formation of
condensation.
Warning: disconnect all electrical connections before any kind of
maintenance.
The instrument shall never be opened.
In case of failure or faulty operation send the instrument back to
the distributor or Weiss Instruments (see address) with a detailed
description of the fault.
Be sure that the wires for probes and for the power supply are
separated and far enough from each other, without crossing and
coiling.