American Diagnostic Corporation (ADC) Diagnostix 922 User Manual

Page 1
DIAGNOSTIX
922
MERCURIAL
BLOOD
PRESSURE
INSTRUMENT
USE, CARE, &
TM
Page 2
A
Special
Thank
You...
Thank you for choosing an ADC blood pressure instrument . We’re proud of the care and quality that goes into the manufacture of each and every sphygmomanometer that bears our name.
Only the finest materials are used to assure you of a timeless instrument designed for optimum perform­ance.
You’ll quickly appreciate the results for you now own the finest sphygmomanometer that money can buy.
With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of dependable service. Please read the following instructions and general information which will prove helpful in allowing you to enjoy your ADC product.
Thank you for your patronage. It is indeed our pleasure to serve you.
Sincerely, American Diagnostic Corp.
Device Description and Intended Use
ADC®aneroid sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory blood pressure technique to determine systolic and diastolic blood pressure in humans and animals.
A warning statement in this manual identifies a condition or practice which, if not corrected or discontin­ued immediately could lead to patient injury, illness, or death.
CAUTION: Federal law restricts this device to sale by or on the order of a physician or licensed healthcare practi­tioner.
WARNING:
If luer lock connectors are used in the construction of tubing, there is a possibility that they might be inad-
vertently connected to intra-vascular fluid systems, allowing air to be pumped into a blood vessel. Immediately consult a
physician if this occurs.
WARNING:
Do not allow a blood pressure cuff to remain on patient for more than 10 minutes when inflated above 10
mmHg. This may cause patient distress, disturb blood circulation, and contribute to the injury of peripheral nerves.
WARNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is not established.
WARNING:
Do not twist connectors when joining the blood pressure cuff to tubing as this may cause the connectors to
stick and make changing the blood pressure cuff difficult.
WARNING:
When using with an infant or child cuff, extra care must be taken to prevent over-inflation. With smaller
cuffs (infant or child) the cuff can inflate to over 300mmHg with just two full compressions of the bulb. To prevent dis­comfort or injury to the patient and damage to the instrument, bulb should only be partially squeezed, so that each “stroke” inflates the cuff in 40mmHg to 60mmHg increments until inflated to the desired level.
WARNING:
This product may contain a chemical known to the state of California to cause cancer, birth defects, or other
reproductive harm.
Page 3
Parts and Assembly
This booklet contains operating and maintenance information for the DIAGNOSTIX
922 series desktop mercurial blood
pressure instrument. Please read and retain.
Your DIAGNOSTIX calibrated unbreakable plastic cartridge tube, inflation sys­tem (which includes the ADCUFF
Guide and the ADFLOW
922 consists of a main unit with 4mm
marking system, latex-free inflation bladder, bulb,
valve), luer connector, and operating in-
nylon cuff with Size
structions.
Connecting the Inflation System to Manometer
Remove the red cap which seals the reservoir containing mer­cury. Attach the free end of reservoir tube securely to air inlet. Save red safety cap for later use. When transporting mercury instrument we recommend replacing the red safety cap.
Please note: It is normal for negligible amounts of residual mercury droplets to accumulate around the air inlet beneath the red safety cap during transportation.
Warning:Before transporting this instrument, mercury must be stored in the tank and the reservoir lock MUST be switched to the off position to prevent mercury spills.
Cartridge Tube
Main Unit
Reservoir
Cuff
Bulb & Valve
Locking Mechanism
Operating Instructions
Your DIAGNOSTIX™ 922 Desktop Mercury Sphygmomanometer incorporates a special safety feature that locks the mercury within the reservoir during storage, transport, or maintenance.
To Operate the 922: To release the mercury from the reservoir, move the locking lever to the left. Mercury will flow up into the plastic cartridge tube and rest at the “0” mark. If the mercury doesn’t rise within the tube, tilt the entire unit 45° to the left. If mercury fails to rise, or doesn’t reach the “0” level, have unit serviced.
To Lock Mercury within the Reservoir: Tilt the entire unit back 45° towards the reservoir to permit mercury to flow out of cartridge tube and into reservoir. When cartridge tube is completely emptied of ALL mercury (and while it is still tilted 45°), move locking lever to the right. Mercury should be locked within the reservoir during maintenance or transport.
PLEASE NOTE: MERCURY SHOULD BE SECURED WITHIN THE RESERVOIR BEFORE CLOSING LID, DURING TRANSPORT, OR WHEN SERVICING THE CARTRIDGE TUBE, DIAPHRAGMS, OR FILTERS. NEVER DISAS­SEMBLE UNIT UNLESS MERCURY IS FIRST LOCKED WITHIN RESERVOIR.
MEASUREMENT PROCEDURE
1. Patient Position
The patient should sit or lie comfortably. The arm should be fully sup­ported on a flat surface at heart level. (If the arm’s position varies, or is not level with the heart, measurement values obtained will not be consis-
Select Smaller Cuff
tent with the patient’s true blood pressure.) Observer should view manometer in a direct line and at eye level to avoid “parallax error.”
2. Apply the cuff ADCUFF™ nylon cuffs, with proprietary Size Guide™marking system,
are specially designed to promote the precise, accurate determination of blood pressure. Index and range markings ensure use of the correct cuff size. The artery mark indicates proper cuff positioning.Place the cuff over the bare upper arm with “artery” mark positioned directly over the brachial artery. The bottom edge of the cuff should be positioned approximately one inch above the antecubital fold. Wrap the end of the cuff, not containing the bladder, around arm snugly and smoothly and engage adhesive strips. To verify a correct fit, check that the ADCUFF the precise, accurate determination of blood pressure. Index and range markings ensure use of the correct cuff size.
nylon cuffs, with proprietary Size Guide™marking system, are specially designed to promote
The artery mark indicates proper cuff positioning.Place the cuff over the bare upper arm with “artery” mark positioned directly over the brachial artery. The bottom edge of the cuff should be positioned approximately one inch above the
CORRECT
Select Larger Cuff
Page 4
antecubital fold. Wrap the end of the cuff, not containing the bladder, around arm snugly and smoothly and engage ad-
°
F
hesive strips. To verify a correct fit, check that the INDEX line falls between the two RANGElines.
3. Inflate the cuff Close the ADFLOW™valve by turning thumbscrew clockwise.
Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing bulb rapidly. Inflate cuff 20­30mmHg above the point at which the radial pulse disappears.
4.Position the Stethoscope Position the chestpiece in the antecubital space below the cuff, distal to the brachium. Do not place chestpiece underneath the cuff, as this impedes accurate measurement. Use the bell side of a combination stethoscope for clearest detection of the low pitched Korotkoff (pulse) sounds.
5. Deflate the cuff Open the valve to deflate the cuff gradually at the AHA recommended rate of 2-3mmHg per second.
6. Measurement Record the onset of Korotkoff sounds (Phase 1) as the systolic pressure, and the disappearance of these sounds (Phase
5) as the diastolic pressure. (Some healthcare professionals prefer recording diastolic 1 and diastolic 2. Diastolic one oc­curs at phase 4).
After measurement is completed, open valve fully to release any remaining air in the cuff. Remove cuff.
Care and Maintenance
Warning:Before transporting this instrument, mercury must be stored in the tank and the reservoir lock MUSTbe switched
to the off position to prevent mercury spills.
Storage After measurement, wrap cuff around bulb and store in base of main unit. IT IS NOT NECESSARY TO CLOSE UNIT AFTER EACH USE - HOWEVER, ALWAYS REMEMBER TO LOCK MERCURY WITHIN THE RESERVOIR BE­FORE CLOSING LID.
Manometer Periodically replace the filter plug and diaphragm, and clean the inside of the plastic cartridge tube using alcohol
and a lint free pipe brush. Units should be cleaned and serviced periodically depending on frequency of use, or whenever in­side of cartridge tube appears dirty. Proper servicing of your mercury unit according to a strict maintenance schedule will as­sure maximum measurement accuracy and extend the life of your instrument.
For directions on disassembling the unit refer to the enclosed schematic diagram and accompanying instruc­tions.
PLEASE NOTE: MERCURY IS TOXIC. SECURE MERCURY WITHIN THE RESERVOIR BEFORE SERVICING UNIT. ALWAYS WEAR PROTECTIVE CLOTHING (GLOVES AND GOGGLES) WHEN CLEANING OR SERVICING THE UNIT. CLEAN ANY SPILLED MERCURY PROMPTLY USING APPROVED MERCURY CONTAINMENT KITS. ONLY PROPERLY TRAINED SERVICE TECHNICIANS SHOULD ATTEMPT TO SERVICE THE UNIT. CONSULT THE EN­CLOSED MANUAL ENTITLED “JUST IN CASE” FOR PROCEDURES ON DEALING WITH A MERCURY SPILL.
Symbol Definition
Important Warning/Caution
Latex-Free
Circumference Size
Conforms to EU Standards
Symbol Definition
Authorized European
Represenative’s Information
Manufacturer’s Information
Temperature Limit
Humidity Limitation
Page 5
Limited Warranty
American Diagnostic Corporation (ADC) warrants its products against defects in materials and workmanship under normal use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.
2. The entire sphygmomanometer is warranted for one year.
3. The inflation system is warranted for three years.
4. The calibrated plastic cartridge tube is warranted for life and will be replaced FREE OF CHARGE when re­turned freight prepaid with the complete unit.
What Is Covered: Calibration, repair or replacement of parts, and labor. What is Not Covered: Transportation charges to and from ADC. Damages caused by abuse, misuse, accident,
or negligence. Incidental, special, or consequential damages. Some states do not allow the exclusion or limita­tion of incidental, special, or consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service: Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Drive, Haup­pauge, NY 11788. Please include your name and address, phone no., proof of purchase, with a brief note explaining the problem, and $2.00 to cover the cost of return shipping and handling.
Implied Warranty: Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond the original selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other rights which vary from state to state.
To register your product visit us at
www.adctoday.com/register
FOR QUESTIONS, COMMENTS, OR SUGGESTIONS CALL TOLL FREE:
This manual is available online in a variety of languages, follow the links
1-800-ADC-2670
www.adctoday.com/feedback
OR VISIT
for language options.
www.adctoday.com/care