Altek 2500T User Manual

The 2500T offers simplistic set-up to make your job easier.
Features:
Two Built-In Pumps –
Vacuum and Pressure
High Accuracy – 0.035% FS
1-51 mA Inputs
Switch Test Inputs
100 PSID Internal Sensor
Read up to 10,000 PSI With
Optional External Sensor
Built-In 24 VDC Loop Power
Temperature Compensated
From -10˚ to 50˚C (13˚ to 122˚F)
Selectable Engineering Units
Just 18 Pounds
Dual Digital Display –
PSI and mA
Gauge, Differential, or
Vacuum Modes
PneuCal III
The Altek PneuCal III is a highly accurate, digital calibrator for pneumatic field instrumentation, including valve actuators, P/I transmitters, controllers, gauges, switches and recorders. It is especially suitable for checking 3 to 15 PSI systems. Its dual precision regulators enable output of set and variable pressures to control devices, while the switching manifold allows fast selection among the pressure ports. The unit simultaneously displays pressure and mA, and has a built-in loop power supply. Accuracy is ± 0.035% of full scale for all ranges. Optional external sensors measure other pressure ranges.
Each Unit Includes: Built-In Switch Testing, AC Adapter/Charger, Test Leads, Hard Shell Carrying Case, Instructions, Built-in Pressure and Vacuum Pumps.
663 PNEUMATIC CALIBRATOR SPECIFICATIONS
Pressure Ranges/Resolution: -5.00 to 100.00 PSI, -34.47 to 680.00 kPa, VAC to
10,000 PSI with optional modules
Pressure Accuracy: ±0.035% FS PSI (over -10 to +50˚C range)
2
Pressure Engineering Units: PSI; Pa; kPa; MPa; BAR; mBAR; atm; kgf/cm
in Hg @ 0˚C and 15˚C, in H
H20 @ 4˚C and 23˚C; cm H2O @ 4˚C and 23˚C Electrical Engineering Units: mA; % of 4-20 mA; % of 10-50 mA Electrical Input Connections: Double insulated, full-size banana jacks mA Input Range: 1-51 mA mA Resolution: 0.001mA mA Accuracy: ±0.01% FS Temperature Effect: 0.005%/˚C (mA Input only) Loop Power: 24 VDC, ±20%, 25 mA max. (When loop power is enabled,
the discharge rate on the batteries exceeds the charging
capacity of the 16 VDC adapter.) Switch Test: Dry contacts only. (Display will freeze when switch test inputs
change state. Press RESET key to resume normal operation.) Operating Temperature: -10˚C to 50˚C (13˚F to 122˚F) Storage Temperature: -30˚C to 85˚C (-22˚F to 185˚F) Zero Adjustment: Via front panel push-button (pressure only) Display: 2-line,16 character alpha-numeric display Power: Internal Ni-Cd batteries or 16 VDC adapter. (To charge battery
the unit has to be powered off.) Battery Life (typical): 23 hrs. (1 hr. with 24 VDC transmitter power at 20 mA) Dimensions (h x w x d): 7"x18.5"x15.5"
O: 4˚C,15˚C, 20˚C, 23˚C; mm
2
; mm Hg;
Unless otherwise indicated, all specifications are referred to an ambient temperature of 23˚C ±1˚C (72˚F ±2˚F).
PHONE: 1-800-322-5835 • FAX: 716-349-3510
E-MAIL: alteksales@altekcalibrators.com • WEB: www.altekcalibrators.com
MODULES
PRESSURE MODULE SPECIFICATIONS
Gauge Pressure Modules
Model Range Accuracy Media
90-5G 5 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-10G 10 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-30G 30 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-100G 100 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-300G 300 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-500G 500 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-1,000G 1,000 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-2,500G 2,500 PSIG ±0.025% Rdg +0.01% FS 316SS Compatible 90-5KG 5,000 PSIG ±0.1% FS 316SS Compatible 90-10KG 10,000 PSIG ±0.1% FS 316SS Compatible
“Dual Sensor” Pressure Module
Model Range Accuracy Media
90-10G/2500G 10/2,500 PSIG ±0.025% Rdg +0.01%FS 316SS Compatible
Differential Pressure Modules
Model Range Accuracy Media
90-10WD ±10 "H 90-5D 5 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low 90-15D 15 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low 90-30D 30 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low 90-100D 100 PSID ±0.025% Rdg +0.01% FS 316SS High/Dry Gas Low
O Diff. ±0.025 in. H2O Clean Dry Gasses
2
Absolute Pressure Modules
Model Range Accuracy Media
90-15A 15 PSIA ± 0.05% FS 316SS Compatible 90-30A 30 PSIA ± 0.05% FS 316SS Compatible
Vacuum Modules
Model Range Accuracy Media
90-15V VAC to 0 PSIG ± 0.05% FS 316SS Compatible 90-15C VAC to 15 PSIG ± 0.05% FS 316SS Compatible
MODPAK 91 MODULE HOLDER
PHONE: 1-800-322-5835 • FAX: 716-349-3510
E-MAIL: alteksales@altekcalibrators.com • WEB: www.altekcalibrators.com