Alpha Omega Instruments ZRO2000 Specifications

Z
R
O
2000
Fast Response Oxygen Analyzer
FEATURES BENEFITS
Advanced Solid State Sensor Provides High Precision Oxygen Measurements
Measures PPB to 100% One Instrument Does It All
Non-depleting Sensor Designed to Provide Years of Operation
User Scalable Analog Outputs Better Denes the Concentration Range of Interest
Dual User Adjustable Alarms Set Critical Process Limits
Microprocessor Controller Helps to Ensure Dependable and Accurate Reading
Built-in Sample By-Pass gas Helps to promote rapid transport of the sample gas
Product Description
The Series ZrO2000 Fast Response Oxygen Analyzer is a microproces­sor based instrument designed to measure gas phase oxygen concentra­tions over a range of 0-100%. The non-depleting sensor used in the Series ZrO2000 Fast Response Oxygen Analyzer is a zirconium type that exhibits fast response times. The analyzer is housed in a NEMA 1 enclosure rated for general purpose use. The front panel contains ve switches that pro­vide access to the analyzer’s settings. Oxygen values are displayed on a
0.7”(17.8 mm) high 6 digit liquid crystal display (LCD). The instrument is
equipped with two oxygen alarm relays. Both relays are Form C (SPDT) types rated at 10 amps at 115/230 VAC/ 30 VDC and are user congurable
for fail-safe operation. In addition to the alarm relays, the Series ZrO2000 Fast Response Oxygen Analyzer has a built-in audible alarm and red LEDs
for visual indication of an oxygen alarm condition. There are also two green LEDs to indicate whether the LCD is displaying parts per million (PPM) or
percent (%).
Specications
PERFORMANCE
Measurement Ranges: 0% to 100%
Accuracy: Percent: ±2% of reading or 0.05% absolute whichever is greater.
Trace: ±2% of reading or ±0.5 ppm absolute whichever is greater.
The audible alarm may be manually canceled at anytime. If the audible alarm is canceled and the alarm event continues, indications of this condi-
tion will still be available through the front panel LEDs and relay contact(s). Two ow meters are standard equipment for the Series ZrO2000 Fast Re­sponse Oxygen Analyzer, one for the “Sample Flow” and another for the “Bypass Flow”.
The Series ZrO2000 Fast Response Oxygen Analyzer comes equipped
with two standard analog outputs. These can be congured for 0-20 mADC or 4-20 mADC separately. Both are scalable by the user. The analyzer
accepts a universal input of 90VAC to 264VAC @ 50-60Hz. The Series ZrO2000 Fast Response Oxygen Analyzer incorporates standard features that make it suitable for many applications. However, for certain require­ments, the user may desire to augment the capabilities of the instrument by
equipping it with a sample pump or pressure regulator.
Oxygen Alarm Two (2) SPDT Form C contacts rated
Relays: 10 A @ 30 V DC /115/230 VAC.
Alarms may be cleared manually or automatically, by user selection.
Audible Alarm: Internal audible alarm
NOTE: The sample gas used with the ZRO2000 must not contain any hy­drocarbon vapors, hydrogen, carbon monoxide, or other reducing gases.
Sample tubing should be cleaned for oxygen service. Please consult the
factory for additional information.
Response Time: < 5 seconds at 0.15 liters per
minute (LPM) over 1 decade.
The response time for measurements of less than 50 PPM or less depend
to a great extent on the design of the sample delivery system including the materials used.
Sensor Type: Zirconium Oxide
Operating Temperature 50° to 104°F (10° to 40°C) of Analyzer:
Warranty: Two Years Electronics One Year Sensor
ELECTRICAL
Display: 0.7” (17.8 mm) high 6 digit liquid crystal display
Resolution: 0.01 on highest reading & 0.001
ppm on lowest reading
Input Power: Universal 90-264 VAC, 50 or 60 Hz
Standard (2) Selectable 0-20 mA DC or 4-20 Outputs: mA DC, both scalable by user
Audible Alarm: Cancel Front panel switch
SAMPLE GAS CHARACTERISTICS
Maximum Sample Temperature: 122°F (50°C)
Maximum 0.1 to 0.5 pounds per square inch gauge
Sample Pressure: (PSIG) (0.007 to 0.035 kg/cm2).
Sample Gas 85% RH
Moisture Limits:
CONSTRUCTION
Enclosure: Aluminum, rated NEMA 1.
Enclosure 13 inches (330 mm) - length
Dimensions: 10.71 inches (272 mm) - width
6.3 inches (160 mm) – height
Note: All dimensions are without optional equipment
Gas Connections:1/4” stainless steel compression ttings.
Weight: (Bench top Conguration)
9.4 pounds (4.26 kg.) / 10.4 pounds (4.72 kg.)
with optional sample pump.
Alpha Omega Instruments Corp. reserves the right to change or modify its product
specications without notice
Applications
Modied Atmoshphere Packages (MAP) Glove Boxes Biological Reactors Critical Welding Pharmaceutical Processing Heat Treating
Semiconductor Manufacturing Leak Detection
Vessel Blanketing Air Separation
Optional Equipment
Pressure Regulators Sample Pumps
40 Albion Road, Suite 100, Lincoln, RI 02865
Tel: 401.333.8580, 800.262.5977 Fax: 401.333.5550
Email: salescontact@aoi-corp.com Web: aoi-corp.com
Alpha Omega Instruments Accepts VISA, Mastercard, and AMEX