Alpha Omega Instruments 9600 Operating Manual

SERIES 9600
OXYGEN AND CARBON DIOXIDE ANALYZER
USER MANUAL
CAUTION Please read this manual before starting to use the Series 9600 Oxygen and Carbon Dioxide Analyzer. Instructions within the manual are essential for the proper operation of the product.
Alpha Omega Instruments Corp. 40 Albion Road, Suite 100 Lincoln, RI 02865 Phone: 401-333-8580 Fax: 401-333-5550 Website: www.aoi-corp.com Email: salescontact@aoi-corp.com
Rev 1.4.1, October 2011
ALPHA OMEGA INSTRUMENTS CORP. ALL RIGHTS RESERVED INCLUDING THE RIGHT TO REPRODUCE THIS MANUAL, OR ANY PORTION THEREOF, IN ANY FORM.
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
© COPYRIGHT 2011. Alpha Omega Instruments Corp. All rights reserved including the right to reproduce manual or any portion thereof in any form.
WARRANTY
Alpha Omega Instruments Corp. (Seller) warrants the products delivered to be free from defects in material and workmanship at the time of delivery to the FOB. point specified in the purchase order, its liability under this warranty being limited to repairing or replacing, at Alpha Omega Instruments option, items which are returned to it prepaid within two (2) years from the date of shipment and found to Seller’s satisfaction to be defective.
Alpha Omega Instruments offers a two (2) year warranty on both the oxygen and carbon dioxide sensors. This warranty period commences on the date of shipment from Alpha Omega Instruments manufacturing facility. The oxygen sensor, should it fail under normal use, must be returned to Seller prepaid and, if such sensor is determined by Seller to be defective, Seller shall provide Buyer a replacement sensor at no charge. If the carbon dioxide sensor should fail, please contact the factory before removing the carbon dioxide sensor. Buyer must provide the serial number of the analyzer from which the sensor(s) have been removed. If a sensor is found to be defective and a new one issued, the warranty of the replacement sensor (s) is for a period of one year from the date of shipment. At times, it may be necessary to ship a replacement sensor in advance of receiving one returned for warranty claim. In such cases, if the returned sensor has been tested and not found to be defective, it will not be covered under warranty, and the user will be charged the full price of a replacement sensor. Note: replacement of the carbon dioxide sensor must be done at the factory. In no event shall Alpha Omega Instruments Corp. be liable for consequential damages.
NO PRODUCT IS WARRANTED AS BEING FIT FOR A PARTICULAR
PURPOSE AND THERE IS NO WARRANTY OF MERCHANTABILITY.
This warranty applies only if:
(i) the items are used solely under the operating conditions and manner recommended in this
manual, product specifications, or other product specific literature;
(ii) the items have not been misused or abused in any manner or unauthorized repairs were
attempted thereon;
(iii) written notice of the failure within the warranty period is forwarded to Alpha Omega
Instruments Corp. and, the directions received for properly identifying items returned under warranty are followed;
(iv) the return notice authorizes Alpha Omega Instruments Corp. to examine and disassemble
returned products to the extent the Company deems necessary to ascertain the cause of failure.
The warranties stated herein are exclusive. THERE ARE NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, BEYOND THOSE SET FORTH HEREIN, and Alpha Omega Instruments Corp. does not assume any other obligation or liability in connection with the sale or use of said products.
Disclaimer of Warranty
Alpha Omega Instruments Corp. makes no representation or warranties, either expressed or implied, by or with respect to anything in this manual, including, but not limited to, implied warranties of merchantability or fitness for a particular purpose. In no event will Alpha Omega Instruments Corp. be liable for any damages, whether direct or indirect, special, consequential, or incidental arising from the use of this manual. Some states in the USA do not allow the exclusion of incidental or consequential damages. Alpha Omega Instruments Corp. also reserves the right to make any changes to improve the performance of its products at any time and without notice.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE i
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
This page intentionally left blank
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE ii
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
Explanation of graphic symbols
The information listed below is essential to the proper operation of the analyzer. Please review the following safety precautions prior to using the analyzer. Cautionary notes are included throughout this manual.
THIS SYMBOL IS INTENDED TO ALERT THE USER TO THE PRESENCE OF IMPORTANT OPERATING AND MAINTENANCE INSTRUCTIONS
THIS SYMBOL IS INTENDED TO ALERT THE USER TO POTENTIALLY DANGEROUS SITUATIONS OR ITEMS THAT SHOULD BE AVOIDED
Important Safety Information
POTENTIALLY HAZARDOUS AC VOLTAGES EXIST WITHIN THE
DANGER
RISK OF
SHOCK
ANALYZER, IF NOT AVOIDED, COULD POTENTIALLY RESULT IN SERIOUS INJURY OR DEATH. DISCONNECT ALL SOURCES OF POWER AND EXTERNAL CONNECTIONS BEFORE REMOVING THE COVER TO THE ANALYZER
TO AVOID THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THE SERIES 9600 OXYGEN AND CARBON DIOXIDE ANALYZER TO RAIN, WATER SPRAY, OR ANY OTHER LIQUIDS.
RF Disclaimer
This instrument generates and uses small amounts of radio frequency energy, and there is no guarantee that interference will not occur in a particular installation. If this equipment does cause interference to radio or television reception, try to correct the interference by one or more of the following steps:
1. Reorient the receiving antenna.
2. Relocate the instrument with respect to the receiver.
3. Change the AC outlet of the instrument so the instrument and receiver are on different branch circuits.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE iii
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
1 Table of Contents
PAGE
Copyright, Warranty, and Disclaimer-------------------i
Explanation of Graphic Symbols, Important Safety Information, and RF Disclaimer-------------------------iii
Table of Contents-------------------------------------------iv
Tables and Figures-----------------------------------------vii
1 Overview--------------------------------------------------------1
1.1 Instrument Description-----------------------------------------------------------------1
1.1.1 Oxygen Sensor-----------------------------------------------------------------------------1
1.1.2 CO2 Sensor----------------------------------------------------------------------------------1
1.1.3 Water Vapor---------------------------------------------------------------------------------1
1.2 Optional Equipment Descriptions--------------------------------------------------2
1.2.1 Low Pressure Regulator-----------------------------------------------------------------2
1.2.2 High Pressure Regulator-----------------------------------------------------------------2
1.2.3 Sample Pump Option---------------------------------------------------------------------2
1.2.4 Particulate Filter----------------------------------------------------------------------------2
1.2.5 Needle Valve--------------------------------------------------------------------------------2
1.2.6 Isolated RS-232C Serial Communications------------------------------------------2
1.2.7 Isolated RS-485C Serial Communications------------------------------------------2
1.2.8 Coalescing Filter---------------------------------------------------------------------------2
1.2.9 Data Logger---------------------------------------------------------------------------------3
1.2.10 Flow Meter-----------------------------------------------------------------------------------3
1.2.11 Flow Meter with Integral Valve---------------------------------------------------------3
1.2.12 Bypass Plumbing Assembly-------------------------------------------------------------3
1.2.13 Analog Output Isolation------------------------------------------------------------------3
1.2.14 Rack / Panel Mounting Kit---------------------------------------------------------------3
1.3 Standard Features-----------------------------------------------------------------------4
1.4 General Specifications-----------------------------------------------------------------5
Oxygen Sensor-----------------------------------------------------------------------------5
Carbon Dioxide Sensor------------------------------------------------------------------6
2 Installation Procedures------------------------------------7
2.1 Unpacking the Instrument------------------------------------------------------------7
2.2 Electrical Installation-------------------------------------------------------------------7
2.3 Mechanical Configuration-------------------------------------------------------------8
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE iv
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
2.4 Wiring Alarm Relays--------------------------------------------------------------------8
2.5 Wiring the mA Outputs-----------------------------------------------------------------9
3 Operation------------------------------------------------------10
3.1 Preparation for Operation-----------------------------------------------------------10
3.1.1 Mounting Configurations---------------------------------------------------------------10
3.1.2 Initial Check--------------------------------------------------------------------------------10
3.2 Operating Procedures----------------------------------------------------------------10
3.2.1 Power ON----------------------------------------------------------------------------------10
3.2.2 Front Panel Controls and Indicators------------------------------------------------10
3.3 Navigating the Front Panel and Menu System-------------------------------11
3.3.1 General Navigation----------------------------------------------------------------------11
3.3.2 Pass codes--------------------------------------------------------------------------------11
3.4 Miscellaneous Settings---------------------------------------------------------------12
3.4.1 Setting the Baud Rate------------------------------------------------------------------12
3.4.2 Enabling and Setting Instrument Addresses in Optional RS485------------12
3.4.3 Enabling and Setting the System Software Clock-------------------------------12
3.5 Analog Outputs of the Analyzer---------------------------------------------------13
3.5.1 Setting the Analog Output Sensor Association-----------------------------------13
3.5.2 Setting the Analog Output Range----------------------------------------------------13
3.5.3 Setting the Analog Output Offset----------------------------------------------------14
3.5.4 Analog Output Options Overview----------------------------------------------------14
3.6 Alarm System----------------------------------------------------------------------------14
3.6.1 Setting the Alarm Sensor Type Association--------------------------------------15
3.6.2 Setting the Alarm Set point------------------------------------------------------------15
3.6.3 Setting the Alarm Hi or Lo-------------------------------------------------------------15
3.6.4 Setting the Alarm Latch-----------------------------------------------------------------15
3.6.5 Setting the Alarm Failsafe-------------------------------------------------------------16
3.6.6 Setting the Alarm Audible--------------------------------------------------------------16
3.6.7 Alarm Options Overview---------------------------------------------------------------16
3.7 Timing Out--------------------------------------------------------------------------------17
4 Sample Gas Handling-------------------------------------18
4.1 Positive Pressure Sampling-------------------------------------------------------19
4.2 Negative Pressure Sampling------------------------------------------------------20
4.3 Gas System Pressure and Flow Limits ----------------------------------------21
4.4 Corrosive Gases------------------------------------------------------------------------22
4.5 Orientation--------------------------------------------------------------------------------22
4.6 Condensible Gas Constituents----------------------------------------------------22
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE v
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
4.7 Sample Filtration-----------------------------------------------------------------------22
5 Calibration Procedures-----------------------------------23
5.1 Routine Calibration Check----------------------------------------------------------23
5.2 Oxygen Calibration--------------------------------------------------------------------23
5.2.1 Procedure for Checking Oxygen Calibration--------------------------------------23
5.2.2 Zero Adjustment for Oxygen----------------------------------------------------------24
5.2.3 Change Oxygen Sensor Notice------------------------------------------------------24
5.3 Carbon Dioxide Calibration---------------------------------------------------------24
5.3.1 Procedure for Checking Carbon Dioxide Calibration---------------------------25
5.3.2 Zero Check of the Carbon Dioxide Channel--------------------------------------25
6 Menu System-------------------------------------------------26
6.1 Main Menu--------------------------------------------------------------------------------26
7 Maintenance and Troubleshooting-------------------28
7.1 User Maintenance----------------------------------------------------------------------28
7.2 Serviceable Items----------------------------------------------------------------------28
7.3 Troubleshooting------------------------------------------------------------------------28
8 Optional RS-232/RS-485 Serial Communications.29
8.1 Baud Rates-------------------------------------------------------------------------------29
8.1.1 To set the RS-232/485 baud rate----------------------------------------------------29
8.1.2 RS485 Commands----------------------------------------------------------------------29
8.2 Standard Commands------------------------------------------------------------------29
8.3 Detailed RS Command Descriptions--------------------------------------------31
8.4 RS232/485 Connections--------------------------------------------------------------36
Appendix A - Bench Top Configuration Drawings
Part Number BTP------------------------------------------------------------------------37
Appendix B - Oxygen Sensor Material Safety Data Sheet
Part Numbers: 1SEN and 2SEN ----------------------------------------------------38
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE vi
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
Tables and Figures
Tables
Table 2.1 Relay Configurations----------------------------------------8
Table 2.2 Customer Wiring---------------------------------------------9
Table 3.5.1 Analog Output Factory Default Settings--------------13
Table 3.5.2 Overview of Analog Output Settings------------------14
Table 3.6.7 Overview of Alarm Settings------------------------------17
Table 6.1 Main Menu Settings, Values, and Descriptions-- - -26
Table 8.1 Standard RS Commands--------------------------------30
Table 8.2 RS232/485 Wiring------------------------------------------36
Figures
Figure 2.1 Mechanical Dimensions------------------------------------7
Figure 2.2 Rear View------------------------------------------------------8
Figure 4.1 Simple Positive Pressure Sampling ------------------18
Figure 4.2 Simple Negative Pressure Sampling with
Internal PUMP---------------------------------------------19
Figure 4.3 Positive Pressure Sampling System------------------20
Figure 4.4 Negative Pressure Sampling System-----------------21
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE vii
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
1 Overview
1.1 Instrument Description
The Series 9600 Oxygen and Carbon Dioxide Analyzer provides an accurate and repeatable measurement of both oxygen and carbon dioxide in a variety of gas streams. The instrument is powered from a universal power supply equipped to accept 90-264 VAC, 47-63Hz. Optional equipment includes pressure regulators, particulate and coalescing filters, and a vacuum pump for conditions when the sample gas lacks sufficient pressure to move the sample through the instrument's plumbing.
The analyzer is housed in a NEMA 1 equivalent enclosure rated for general purpose indoor use. Oxygen and carbon dioxide values are displayed on a 4 line by 20 character liquid crystal display (LCD) with a total height of 0.81" (20.8 mm). The front panel also includes a membrane panel with seven buttons that provide access to the analyzer's settings. The instrument is equipped with two alarm relays (see specifications) which are user configurable and can optionally be set for fail-safe operation. In addition to the alarm relays, the Series 9600 Oxygen and Carbon Dioxide Analyzer has a built-in audible alarm and two front panel light emitting diodes (LEDs) for visual indication of an alarm condition. The audible alarm can be canceled by the user, however the alarm event will still be in effect and indications of the alarm condition will still be available through the front panel LEDs and relay contact(s). A flow meter is mounted on the front panel of the instrument and is standard equipment for the Series 9600 Oxygen and Carbon Dioxide Analyzer. A flow meter with integrated flow control needle valve is available as optional equipment. The standard Series 9600 Oxygen and Carbon Dioxide Analyzer comes equipped with two direct current (DC) analog outputs. These analog outputs can be independently configured for 0-20 mA or 4-20 mA. Both are scalable over the operating range of the instrument's sensors. Using a terminating resistor on either of these outputs can also provide any analog voltage in the range of 0-10 volts (see general specifications on page 4 for maximum resistance).
1.1.1 Oxygen Sensor
The Series 9600 Oxygen and Carbon Dioxide Analyzer features an ambient temperature electrochemical sensor with an Enhanced Electrolyte System (EES). One of the features of the EES is the use of a weak acid electrolyte in place of potassium hydroxide. This produces a sensor with a greatly extended useful life and excellent measurment stability.
1.1.2 CO2 Sensor
The Series 9600 Oxygen and Carbon Dioxide Analyzer uses a non-dispersive infrared (NDIR) sensor to determine the carbon dioxide concentrations in gases. Infrared gas sensors, which are comprised of solid state devices, do not chemically react with the gas.
1.1.3 Water Vapor
The Series 9600 Oxygen and Carbon Dioxide Analyzer incorporates a CO2 sensor that uses a high precision optical system. Therefore, in order to provide accurate and repeatable data, it is essential that the CO2 sensor not be exposed to condensed water or liquids of any type. The dew point temperature of the sample gas should always be lower than the temperature of the NDIR sensor, thus helping to prevent the condensation of water vapor on the sensor’s optics. If it is anticipated that the sample gas may be saturated with water vapor, provisions should be made to remove the water by
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 1
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
filtration/coalescing, refrigeration, vortex cooling, etc. Consult Alpha Omega Instruments Corp. for more information on this subject.
1.2 Optional Equipment Descriptions
The Series 9600 Oxygen and Carbon Dioxide Analyzer incorporates standard features that make it immediately suitable for many applications. However, for certain requirements, the user may desire to augment the capabilities of the instrument by equipping it with one or more of the available options, as described below.
1.2.1 Low Pressure Regulator P/N 96-LPR
Aluminum body with a maximum input pressure of 145 PSIG.
1.2.2 High Pressure Regulator P/N 96-PRR
High capacity stainless steel pressure regulator with 3000 PSIG inlet capacity and an adjustable output pressure range of 0-5 PSIG. Note: Typically used with the optional needle valve (P/N 9NV).
1.2.3 Sample Pump Option P/N 96-PMP
Sample pump designed for applications where the sample pressure is insufficient to transport the sample gas through the instrument's plumbing.
1.2.4 Particulate Filter P/N 96-95S
Miniature T-type 316 stainless steel filter with 1/4” compression fittings. Recommended when particle sizes are greater than 5 microns.
1.2.5 Needle Valve P/N 96-NV
Stainless steel needle valve for high pressure sample source (greater than 1 PSIG). Also used in bypass style plumbing arrangements for applications where the instrument will be used to measure both positive and negative pressure sources using a sample pump.
1.2.6 Isolated RS-232C Serial Communications P/N 96-RS2
The RS-232C Serial Communications option is installed at the factory and is designed for applications where enhanced serial communications is required between the Series 9600 Oxygen and Carbon Dioxide Analyzer and a host system. The maximum distance from the host system is 50 feet.
1.2.7 Isolated RS-485C Serial Communications P/N 96-RS4
The RS-485 Serial Communications option is installed at the factory and is designed for applications where enhanced serial communications is required between the Series 9600 Oxygen and Carbon Dioxide Analyzer and a host system over a distance greater than 50 feet. The maximum distance from the host system is 4000 feet.
1.2.8 Coalescing Filter P/N 96-CF
Filter with aluminum housing. Rated for nominal 0.3 microns (95%).
Spare filter elements for P/N 96-CF. P/N 96-CFE
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 2
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
1.2 Optional Equipment Descriptions (continued)
1.2.9 Data Logger P/N 9DL
Internally mounted four channel data logger (no display). The data logger can be used to record the O and CO2 concentrations and operates in a wrap-around mode so that when the memory is full, the oldest data will be replaced with the newest. Recording frequency is adjustable from once every 0.5 seconds to once every 9 hours (user selectable). Memory is non-volatile EEPROM with 32,000 readings kept in storage. Software will be provided that allows downloading of the data to a PC via serial port.
1.2.10 Flow Meter P/N 9FLMA
Durable one-piece acrylic flow meter without flow control adjustment. Standard equipment.
1.2.11 Flow Meter with Integral Valve P/N 9FLMC
Durable one-piece acrylic flow meter with flow control adjustment. Optional equipment.
1.2.12 Bypass Plumbing Assembly P/N 9BYPASS
Plumbing arrangement used when the instrument will be used for both positive and negative pressure source applications. Includes a needle valve and a bypass “tee” on the inlet of the instrument. This is typically used in conjunction with 9PMP Sample Pump Option. Some high pressure applications will also make use of such a plumbing arrangement. Details are discussed later in the manual.
2
1.2.13 Analog Output Isolation. P/N 9AOISO
Installed at the factory, galvanic isolation of both analog outputs.
1.2.14 Rack / Panel Mounting Kit P/N 96-RMK
Used for mounting the instrument in either a Rack configuration or inside a panel cutout.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 3
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
1.3 Standard Features
1
Analog Outputs: Two 0/4-20 mA, configurable for voltage or direct current
Individually configurable for either O2 or CO2 monitoring
Maximum load: less than 950 ohms @ 25C
Alarm Relays: Two (2) SPDT Form C contacts rated 10 A (250 VAC) / 5A (100 VDC).
Alarms may be cleared manually or automatically, by user selection.
Individually configurable to be set as high or low alarms for either O2 or CO2 levels.
Input Power: Universal 90-264 VAC, 47-63 Hz
Audible Alarms: Internal audible alarm (user configurable per alarm)
Audible Alarm Canceling: Individual front panel buttons
Operating Temperature: 50° to 104°F (10° to 40°C) / 99% max humidity (non-condensing)
Enclosure: Aluminum, equivalent to NEMA 1
Enclosure Dimensions: 13 inches (330 mm) – length
11 Inches (280 mm) – width
7 Inches (178 mm ) - height
Sample Delivery: Pressure or optional pump
Display: 4 Line by 20 character LCD
1
All specifications are based on a temperature of 77°F (25°C) at standard pressure of 29.9 inches of mercury
(1 atmosphere).
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 4
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
1.4 General Specifications

Oxygen Sensor

Measurement Range: Percent: 0-100
Resolution: Greater than 10% = 0.1%
Less than 10% = 0.01%
Accuracy: ±1% for full scale ranges >10% .
±5% for full scale ranges <10%
Sensor Type: Long-life electrochemical sensor
Recommended Sample
<1.0 PSIG
1
Pressure:
Response Time: 90% of full scale response
in < 20 seconds.
Warm Up Time: Instantaneous
Sample Flow Rate: 0.1-0.5 liters per minute (LPM) optimum
Operating Humidity Range: 0-99% RH (non-condensing)
Calibration: Ambient air or calibration gas
1
The actual pressure of the sample can be much higher, however higher pressure sampling requires different
plumbing arrangements that are described later in the manual.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 5
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP
General Specifications (continued)

Carbon Dioxide Sensor

Measurement Range1: Percent: 0-20
Trace: 0-5,000 PPM
(determined at order placement)
Resolution: Greater than 10% = 0.1%
Less than 10% = 0.01% Less than 10,000ppm = 1 ppm
Accuracy: Percent: ±0.5% full scale or ±5% of reading,
whichever is greater Trace: ±30ppm or ±2% of reading, whichever is greater.
Sensor Type: Non Dispersive Infrared (NDIR)
Recommended Sample
<1.0 PSIG (see footnote 1 opposite page)
Pressure :
Response Time: < 35 seconds to 63% step change at a sample
flow of 0.5 liters per minute
Warm Up Time: < 90 seconds
Sample Flow Rate: 0.1-0.5 liters per minute (SLPM) optimum
Operating Humidity Range: 5-95% RH (non-condensing)
Calibration: Calibration gas
1
Other ranges for CO2 covering ppm and percent to 100% are available. Please contact the factory for details.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 6
ALPHA OMEGA INSTRUMENTS CORP USER MANUAL
2 Installation Procedures
Figure 2.1 Mechanical Dimensions
2.1 Unpacking the Instrument
Upon opening the shipping container, carefully unpack the instrument to check if the outer surfaces have been damaged. If so, report the findings immediately to Alpha Omega Instruments who will, in turn, provide further instructions.
NOTE: IF DAMAGE HAS BEEN FOUND, DO NOT PROCEED FURTHER, BUT INSTEAD, CONTACT THE FACTORY.
If there is no apparent damage, check the contents to ensure all items were shipped. In some cases, items may be back ordered. All damage and shortage claims must be made known to Alpha Omega Instruments within 10 days after receipt of shipment. Carefully rotate the analyzer and check to make sure no components have been loosened or dislodged. If there are loose or dislodged components (rattling of any kind), contact the factory for further instructions. If there is no evidence of loose or dislodged components, the installation procedure can begin.
2.2 Electrical Installation
ELECTRICAL INSTALLATION SHOULD BE PERFORMED BY A
WARNING
The Series 9600 Oxygen and Carbon Dioxide Analyzer is shipped with a standard North American power cord NEMA style 5-15P. The mating receptacle on the rear panel of the analyzer is an IEC style 60320 C13/C14. The analyzer accepts a power input of 90 VAC to 264 VAC @ 50/60 Hz. There is an On/Off switch located on the rear of the instrument.
QUALIFIED PERSON AND SHOULD COMPLY WITH APPLICABLE FEDERAL, STATE, OR LOCAL ELECTRICAL SAFETY CODES.
SERIES 9600 Oxygen and Carbon Dioxide Analyzer PAGE 7
Loading...
+ 34 hidden pages