Alpha Omega Instruments 1300 Operating Manual

SERIES 1300
OXYGEN DEFICIENCY MONITOR
USER MANUAL
CAUTION
Please read this manual before using the Series 1300 Oxygen Deficiency Monitor. Instructions
within the manual are essential for the proper operation of this product.
40 Albion Road, Lincoln, RI, USA 02865
Tel: (001) (401) 333-8580 Fax: (001) (401) 333-5550 Email: contact@aoi-corp.com Website: www.aoi-corp.com
Rev 1.0915, September 2017
© COPYRIGHT 2014. Alpha Omega Instruments Corp. All rights reserved including the right to reproduce this manual or any portion thereof in any form.
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
WARRANTY
Alpha Omega Instruments Corp. warrants that delivered products are free from defects in material and workmanship at the time of delivery to the FOB point specified in the purchase order. Liability under this warranty is limited to repairing or replacing, at Alpha Omega Instruments' option, items which are returned to it prepaid within four years (including sensor) from the date of shipment and found to Seller’s satisfaction to have been defective.
Alpha Omega Instrument's four (4) year sensor warranty provides protection for four years from the date of shipment of the Series 1300 Oxygen Deficiency Monitor. Any sensor from a Series 1300 Oxygen Deficiency Monitor that fails under normal use must be returned to Seller prepaid and, if such sensor is determined by Seller to be defective, Seller shall provide Buyer with a replacement sensor. Buyer must provide the serial number of the monitor from which the sensor has been removed. If a sensor is found to be defective and a new one issued, the warranty of the replacement sensor(s) shall not extend beyond the initial warranty period of the Series 1300 Oxygen Deficiency Monitor, or for a period of one year, whichever is longer. After this, all replacement sensors will be warranted for a period of one year from the date of shipment. In no event shall Alpha Omega Instruments Corp. be liable for consequential damages.
NO PRODUCT IS WARRANTED AS BEING FIT FOR A PARTICULAR
PURPOSE AND THERE IS NO WARRANTY OF MERCHANTABILITY.
This warranty applies only if:
(i) the items are used solely under the operating conditions and manner recommended in this manual, product
specifications, or other product specific literature;
(ii) the items have not been misused or abused in any manner or unauthorized repairs were attempted thereon;
(iii) written notice of the failure within the warranty period is forwarded to Alpha Omega Instruments Corp. and, the
directions received for properly identifying items returned under warranty are followed;
(iv) the return notice authorizes Alpha Omega Instruments Corp. to examine and disassemble returned products to the
extent the Company deems necessary to ascertain the cause of failure.
The warranties stated herein are exclusive. THERE ARE NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, BEYOND THOSE SET FORTH HEREIN, and Alpha Omega Instruments Corp. does not assume any other obligation or liability in connection with the sale or use of said products.
Disclaimer of Warranty
Alpha Omega Instruments Corp. makes no representation or warranties, either expressed or implied, by or with respect to anything in this manual, including, but not limited to, implied warranties of merchantability or fitness for a particular purpose. In no event will Alpha Omega Instruments Corp. be liable for any damages, whether direct or indirect, special, consequential, or incidental arising from the use of this manual. Some states in the USA do not allow the exclusion of incidental or consequential damages. Alpha Omega Instruments Corp. also reserves the right to make any changes to improve the performance of its products at any time and without notice.
Series 1300 Oxygen Deficiency Monitor PAGE i
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
Table of Contents
WARRANTY..............................................................................................................................................i
1 Overview...............................................................................................................................................1
1.1 Explanation Of Graphic Symbols...................................................................................................1
1.2 Instrument Description...................................................................................................................1
1.2.1 Features.................................................................................................................................2
1.3 Mounting........................................................................................................................................2
1.4 Oxygen Sensor..............................................................................................................................2
1.5 Description Of Optional Equipment...............................................................................................3
1.6 Standard Features.........................................................................................................................4
1.7 Optional Features..........................................................................................................................4
1.8 General Specifications...................................................................................................................5
2 Installation.............................................................................................................................................6
2.1 Unpacking The Instrument.............................................................................................................6
2.2 Electrical Installation......................................................................................................................7
2.3 Mechanical Configuration..............................................................................................................7
2.3.1 Main Enclosure.......................................................................................................................8
2.3.2 Remote Oxygen Sensor Enclosure........................................................................................9
3 Wiring..................................................................................................................................................10
3.1 Local Oxygen Sensor...................................................................................................................11
3.1.1 Local Oxygen Sensor Installation.........................................................................................12
3.2 Power (TB1).................................................................................................................................13
3.3 Remote Oxygen Sensors.............................................................................................................14
3.3.1 Remote Oxygen Sensor Installation Procedure...................................................................15
3.4 Horn And Strobe...........................................................................................................................16
3.4.1 Horn and Strobe PCB...........................................................................................................17
3.4.2 Horn and Strobe Daisy Chain Configuration........................................................................18
3.4.3 Horn and Strobe Star Configuration.....................................................................................19
3.4.4 Horn and Strobe External Power..........................................................................................19
3.4.5 Horn and Strobe Relay Wiring..............................................................................................20
3.4.6 Horn and Strobe Switch Settings..........................................................................................20
3.4.7 Horn and Strobe Maximum Distance...................................................................................21
3.5 Analog Outputs.............................................................................................................................22
3.6 Alarm Relays................................................................................................................................22
3.6.1 Fail Safe................................................................................................................................23
3.7 RS-232/485 Communications......................................................................................................23
4 Operation.............................................................................................................................................24
4.1 Power ON.....................................................................................................................................24
4.2 Panel Description.........................................................................................................................24
4.3 Navigating The Screens...............................................................................................................24
4.3.1 Home Screen........................................................................................................................24
4.3.1.1 Three Line "Home" Screen................................................................................................25
4.3.1.2 Large Font Home Screen..................................................................................................25
4.3.2 Alert Screen..........................................................................................................................25
4.3.2.1 Navigating the Alert Screens.............................................................................................26
4.3.3 Active Alarms Screen............................................................................................................28
4.3.4 Alarm Status Screen.............................................................................................................28
PAGE ii Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
4.3.5 Main Menu Screen................................................................................................................29
4.4 Alarm System...............................................................................................................................29
4.4.1 Alarm System Menu Settings...............................................................................................30
4.4.2 Default Alarm Settings..........................................................................................................31
4.4.3 Default Alarm Activation Behavior........................................................................................31
4.4.4 Low Alarms vs High Alarms..................................................................................................31
4.4.5 Alarm Latching......................................................................................................................31
4.4.6 Relay Fail-safe Operation.....................................................................................................32
4.4.7 Navigating Alarms.................................................................................................................32
4.5 Horn And Strobe...........................................................................................................................33
4.5.1 Horn and Strobe Naming......................................................................................................34
4.6 Analog Outputs.............................................................................................................................34
4.6.1 Analog Output Sensor Association.......................................................................................34
4.6.2 Analog Output Range...........................................................................................................34
4.6.3 Analog Output Offset............................................................................................................35
4.6.4 Voltage Output Mode............................................................................................................35
4.7 Navigating The Menu...................................................................................................................36
4.7.1 Timing Out.............................................................................................................................36
4.8 Main Menu...................................................................................................................................37
4.8.1 Passcode..............................................................................................................................39
4.8.2 CALIBRATE..........................................................................................................................39
4.8.3 SENSOR SETUP..................................................................................................................39
4.8.3.1 Sensor Naming..................................................................................................................39
4.8.4 ALARM SETUP.....................................................................................................................39
4.8.4.1 Setting the Alarm Sensor...................................................................................................39
4.8.4.2 Setting the Alarm Set point................................................................................................40
4.8.4.3 Setting the Alarm High or Low...........................................................................................40
4.8.4.4 Setting the Alarm Latch.....................................................................................................40
4.8.4.5 Clearing a Latched Alarm..................................................................................................40
4.8.4.6 Activating Alarm Relays and Horn and Strobes................................................................40
4.8.5 FAILSAFE SETUP................................................................................................................41
4.8.6 OUTPUTS SETUP................................................................................................................41
4.8.6.1 Sensor................................................................................................................................41
4.8.6.2 Scale Low & Scale High....................................................................................................41
4.8.6.3 Analog Output Offset.........................................................................................................42
4.8.7 HORN & STROBES..............................................................................................................42
4.8.7.1 Naming the horn and strobe..............................................................................................42
4.8.8 DATA LOGGER.....................................................................................................................43
4.8.9 MISCELLANEOUS...............................................................................................................43
4.8.10 Field Elevation Adjustment.................................................................................................44
4.8.11 System Information (“SYSTEM INFO”)...............................................................................44
5 Data Logger.........................................................................................................................................45
5.1 Logger Operation.........................................................................................................................45
5.1.1 Logger State.........................................................................................................................45
5.1.2 Logger Mode.........................................................................................................................45
5.2 Sample Interval............................................................................................................................46
5.3 Date And Time..............................................................................................................................46
5.4 Sensor Selection..........................................................................................................................46
Series 1300 Oxygen Deficiency Monitor PAGE iii
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
6 Sensor Setup.......................................................................................................................................47
6.1 Sensor Removal...........................................................................................................................47
7 Calibration Procedures........................................................................................................................48
7.1 Routine Calibration Check...........................................................................................................48
7.2 Oxygen Calibration......................................................................................................................48
7.2.1 Calibration with Ambient Air..................................................................................................48
7.2.2 Calibration With Other Gases...............................................................................................48
7.2.3 Procedure for Checking Oxygen Calibration........................................................................49
7.3 Calibration On Ambient Air...........................................................................................................49
7.4 Optional Calibration Fixture.........................................................................................................49
7.5 NO Zero Adjustment....................................................................................................................50
7.6 Change Oxygen Sensor Notice...................................................................................................50
7.6.1 Replacing a Sensor..............................................................................................................50
8 Battery Backup....................................................................................................................................52
8.1 Battery Backup HALT...................................................................................................................52
8.2 Battery Backup Time....................................................................................................................53
8.3 Battery Status Screen..................................................................................................................54
9 Warnings..............................................................................................................................................55
9.1 Alarm Warnings............................................................................................................................55
9.2 Sensor Removal Warnings..........................................................................................................56
9.2.1 Scenario 1 – Dangerous Low Oxygen Level........................................................................56
9.2.2 Scenario 2 – Disconnected Oxygen Sensor.........................................................................56
9.3 Horn And Strobe Warnings..........................................................................................................56
9.3.1 Horn and Strobe Removal....................................................................................................56
9.3.2 Horn and Strobe Address Change........................................................................................56
10 Maintenance and Troubleshooting....................................................................................................57
10.1 User Maintenance......................................................................................................................57
10.2 Serviceable Items......................................................................................................................57
10.3 System Messages.....................................................................................................................57
11 Serial Communications......................................................................................................................59
11.1 Baud Rates.................................................................................................................................59
11.1.1 Setting the Baud Rate.........................................................................................................59
11.2 Line Endings...............................................................................................................................59
11.3 RS Echo.....................................................................................................................................59
11.4 Addressing..................................................................................................................................59
11.4.1 Address...............................................................................................................................60
11.5 Standard Commands.................................................................................................................60
11.5.1 Read....................................................................................................................................60
11.5.2 Write....................................................................................................................................61
11.6 RS485.........................................................................................................................................61
11.7 RS Security................................................................................................................................61
11.8 Variable Registers Table.............................................................................................................61
11.9 RS Data Logger..........................................................................................................................65
12 APPENDIX A – Main Enclosure Template.........................................................................................66
13 APPENDIX B – Remote Enclosure Template...................................................................................67
14 APPENDIX C – Terminal Block Reference.......................................................................................68
14.1 Standard AC Connections..........................................................................................................68
14.2 Remote Sensor And Horn And Strobe Connections..................................................................68
PAGE iv Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
14.3 PCB Connectors........................................................................................................................69
15 APPENDIX D – Horn and Strobe(s)..................................................................................................70
15.1 Externally Powered Horn And Strobe(s)....................................................................................70
15.2 Horn And Strobe Sound Switch Settings...................................................................................70
15.3 Horn And Strobe Tone Settings..................................................................................................71
15.4 Horn And Strobe Mechanical Details.........................................................................................72
15.5 Horn And Strobe PCB Detail......................................................................................................73
16 APPENDIX E – Remote with Calibration Fixture..............................................................................73
17 APPENDIX F – Oxygen Sensor Material Safety Data Sheet............................................................74
18 APPENDIX G – Battery Backup........................................................................................................78
18.1.1 Battery Temperature...........................................................................................................78
18.1.2 Battery Testing....................................................................................................................78
18.2 Battery Installation/Replacement...............................................................................................78
Figures
Figure 1 - Oxygen Sensor........................................................................................................................2
Figure 2 - Main Enclosure........................................................................................................................8
Figure 3 - Remote Oxygen Sensor Enclosure.........................................................................................9
Figure 4 - Bottom View 1 Of Series 1300...............................................................................................10
Figure 5 - Bottom View 2 Of Series 1300...............................................................................................11
Figure 6 - Oxygen Sensor......................................................................................................................11
Figure 7 - Local Oxygen Sensor Installation..........................................................................................12
Figure 8 - Local Oxygen Sensor Installation & Wiring...........................................................................13
Figure 9 - Remote Oxygen Sensor Enclosure.......................................................................................14
Figure 10 - Remote Oxygen Sensor Enclosure & PCB.........................................................................15
Figure 11 - Remote Oxygen Sensor.......................................................................................................15
Figure 12 - Horn And Strobe With PCB..................................................................................................17
Figure 13 - Horn And Strobe PCB..........................................................................................................17
Figure 14 - Wiring Horn And Strobes In Series (Daisy Chained)...........................................................18
Figure 15 - Wiring Horn And Strobes Individually..................................................................................19
Figure 16 - Externally Powered Horn And Strobes................................................................................19
Figure 17 - Wiring Horn And Strobes Using A Relay..............................................................................20
Figure 18 - SW1 Detail On Horn And Strobe PCB.................................................................................20
Figure 19 - Three Line Home Screen.....................................................................................................25
Figure 20 - Large Font Home Screen....................................................................................................25
Figure 21 - Main Alert Screens...............................................................................................................26
Figure 22 - Active Alarms Screen...........................................................................................................28
Figure 23 - Modified Active Alarms Screen............................................................................................28
Figure 24 - Alarm Status Screens..........................................................................................................28
Figure 25 - Main Menu Screen...............................................................................................................29
Figure 26 - Setting An Alarm..................................................................................................................30
Figure 27 - Horn And Strobe Detected...................................................................................................33
Figure 28 - Horn And Strobe Removed..................................................................................................33
Figure 29 - Main Menu...........................................................................................................................36
Figure 30 - Main Enclosure Template....................................................................................................66
Figure 31 - Remote Enclosure Template................................................................................................67
Series 1300 Oxygen Deficiency Monitor PAGE v
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
Figure 32 - Horn And Strobe PCB Wiring Diagram................................................................................70
Figure 33 - Horn And Strobe Mechanical Details...................................................................................72
Figure 34 - Horn And Strobe PCB Detail...............................................................................................73
Figure 35 - Remote Oxygen Sensor Enclosure & Optional Calibration Fixture.....................................73
Tables
Table 1 - Power Connections (Terminal Block 1)...................................................................................13
Table 2 - Remote Oxygen Sensor Connections (Terminal Block 1).......................................................14
Table 3 - Horn And Strobe Connections (Terminal Block 1)...................................................................16
Table 4 - Horn And Strobe PCB Switch Settings....................................................................................21
Table 5 - Analog Output Connections.....................................................................................................22
Table 6 - Relay Connections..................................................................................................................22
Table 7 - Fail Safe Logic.........................................................................................................................23
Table 8 - RS232/485 Communications Wiring.......................................................................................23
Table 9 - Alert Screens...........................................................................................................................27
Table 10 - Overview Of Alarm Settings..................................................................................................30
Table 11 - Default Alarm Settings...........................................................................................................31
Table 12 - Alarm Status Screens............................................................................................................32
Table 13 - Analog Output Factory Default Settings................................................................................34
Table 14 - Main Menu.............................................................................................................................38
Table 15 - Data Logger Settings.............................................................................................................43
Table 16 - Miscellaneous Settings..........................................................................................................43
Table 17 - System Information...............................................................................................................44
Table 18 - Battery Backup Minimum Operating Times...........................................................................53
Table 19 - Battery Status Screens..........................................................................................................54
Table 20 - System Messages.................................................................................................................58
Table 21 - Variable Registers.................................................................................................................64
Table 22 - Logger Time / Interval Estimates...........................................................................................65
Table 23 - Logger RS Commands..........................................................................................................65
Table 24 - Horn And Strobe Sound Switch Settings...............................................................................71
PAGE vi Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
1 Overview
1.1 Explanation Of Graphic Symbols
The information listed below is essential to the proper operation of the monitor. Please review the following safety precautions prior to using the monitor. Cautionary notes are included throughout this manual.
WARNING MESSAGE
THIS SYMBOL IS INTENDED TO ALERT THE USER ABOUT GENERAL WARNINGS AND IMPORTANT OPERATING AND MAINTENANCE INSTRUCTIONS.
WARNING
HAZARDOUS VOLTAGE. CONTACT MAY CAUSE ELECTRIC SHOCK OR BURN. TURN OFF & LOCK OUT SYSTEM BEFORE SERVICING. THIS SYMBOL IS INTENDED TO ALERT THE USER TO POTENTIALLY DANGEROUS SITUATIONS.
1.2 Instrument Description
The Series 1300 Oxygen Deficiency Monitor allows for the simultaneous monitoring of breathing air oxygen using up to three (3) separate oxygen sensors. For monitoring of a single location, the Series 1300 supports either a “Local Sensor” (located within the main electronics enclosure) or a “Remote Sensor” (located in a separate remote electronics enclosure). For monitoring of multiple locations, up to two (2) “Remote Sensors” can be added to the “Local Sensor” for a total of three (3) monitoring locations. The standard input power to the Series 1300 instrument is 88-264VAC, 47-63Hz (Hertz). Available power options include 12VDC or 18-36VDC (24VDC) and must be specified at the time of order placement.
The Series 1300 Oxygen Deficiency Monitor is housed in a NEMA 1 equivalent enclosure rated for general purpose indoor use. Oxygen values for all sensors are displayed on a 4 line by 20 character liquid crystal display (LCD) with a total height of 0.81" (20.8 mm). Various display modes are available that allow for viewing all sensors simultaneously or having the sensor oxygen values cycle in a large font display mode. The front panel includes a membrane panel with seven buttons that provide access to the monitor's settings. The monitor is equipped with four alarm relays (see specifications) which are user configurable and can be set for fail-safe operation. Also included, is a built-in audible indication of an oxygen alarm condition.
The Series 1300 Oxygen Deficiency Monitor comes equipped with two direct current (DC) analog outputs. The analog outputs can be independently configured for 0-20mA or 4-20mA. Both are scalable over the operating range of the monitor. Using a terminating resistor on either of these outputs will provide an analog voltage up to 5 volts full scale up to a maximum of 1,000 feet.
The Series 1300 Oxygen Deficiency Monitor comes equipped standard with a data-logger and a serial interface commonly referred to as RS-232. The interface is compatible with EIA/TIA-232E. Optional addresses from 1 to 32 may be assigned to each monitor for communicating with up to 32 instruments on the RS-2321 or RS-485 bus.
1
The built in standard RS-232 by definition does not allow for more than one instrument on the same physical wires. An external RS-232 to RS-485 adapter or an “Ethernet Serial server” can be used to allow addressing multiple RS-232 units by converting the RS-232 signals to an alternate bus.
Series 1300 Oxygen Deficiency Monitor PAGE 1
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
1.2.1 Features
Standard Features:
Easy to use and configure.
Capable of monitoring up to three oxygen
sensors simultaneously.
Automatically senses and configures new
sensors.
Multiple monitoring screens: detailed or
large font versions.
Alarm status is clearly displayed on the
screen with pop-up for active alarms
Scrolling collapsible menu system is easy
to navigate.
Wide range of configurable settings.
Any sensor can be assigned to any output.
Standard Features (continued):
Any Alarm can be assigned to any sensor.
Any Alarm can activate any combination
of relays, horns, and strobes.
Isolated digital communications (RS-232
or RS485).
Built-in data-logger.
Optional Features:
Battery backup.
Isolated process outputs.
Extended memory Data-logger (capable
of almost 4x more logging capacity).
1.3 Mounting
Wall Mount (WMT) is the basic configuration for the Series 1300 Oxygen Deficiency Monitor. All electrical connections are located on the bottom of the monitor. See APPENDIX A – Main Enclosure Template on page 66 for detailed dimensional drawings.
1.4 Oxygen Sensor
The Series 1300 Oxygen Deficiency Monitor features an ambient temperature electrochemical sensor with an Enhanced Electrolyte System (EES). The EES, significantly extends the useful life of the oxygen sensor as well as providing exceptional measurement stability.
Figure 1 - Oxygen Sensor
PAGE 2 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
1.5 Description Of Optional Equipment
The Series 1300 Oxygen Deficiency Monitor incorporates a host of standard features. For certain requirements, user's may desire to augment the capabilities of the monitor by ordering the monitor with one or more available options described below.
1.5.1 Battery Backup P/N 13-BAT
The Battery Backup option provides backup power to the Series 1300 Oxygen Deficiency Monitor during a temporary loss of main power. Please see section 8 Battery Backup for more information on this option.
1.5.2 RS-485 P/N 13-485
The factory installed RS-485 is an optional upgrade to the factory default of RS-232 and offers the ability to extend the distance from 50 feet maximum up to 4,000 feet between the monitor and the host system. This option also provides serial communications (both sending and receiving ) between several monitors up to a maximum of 32 on one RS-485 three wire communications channel.
1.5.3 Replacement Oxygen Sensor P/N 13-SEN
This option consists of just the sensor (no enclosure or electronics). This option should be ordered in the event that the sensor is damaged or needs replacement due to end of life.
1.5.4 Remote Mounted Sensor Enclosure P/N 13-RSEN
This option consists of a NEMA 1 (IP 30) remote sensor enclosure containing the oxygen sensor and associated circuitry used to communicate with the read out electronics. The remote sensors may be added to the Series 1300 Oxygen Deficiency Monitor in the field by the user. See section 3.3 Remote Oxygen Sensors.
1.5.5 Analog Output Isolation P/N 13-ISO
Installed at the factory, this option provides galvanic isolation of both analog outputs. This eliminates any potential ground loops when connecting to earth grounded inputs or powering from DC where the power return could create a ground loop with the output grounds or the input and output grounds are tied to chassis ground. The isolation also allows for a higher maximum load resistance up to 1200 ohms (24V @ 20mA) as opposed to the 12V maximum of 300 ohms (6V @ 20mA).
1.5.6 Calibration Fixture P/N 13-CFN
The optional calibration fixture is recommended for checking the calibration of the Series 1300 Oxygen Deficiency Monitor. Ideally, the source of the calibration gas should be compressed air (clean, dry, and oil free). Most industrial gas suppliers can provide the compressed air in small hand-held cylinders. The calibration fixture is designed to easily install over the gas opening of the Series 1300 oxygen sensor making a tight seal. The fixture is equipped with ¼ inch tube fittings to deliver the compressed air through a 1/4" OD plastic tube (not supplied). The sample pressure to the fixture should never exceed 1.5 pounds per square inch (PSIG) at a corresponding sample flow rate of between 0.2 to 1.0 standard cubic feet per hour (SCFH) or 0.1 to 0.5 liters per minute (LPM).
1.5.7 12VDC Power P/N 13-12V
Allows for a user supplied regulated 12V as the main source of power to the instrument. Minimum power requirement is 25.2 Watts.
1.5.8 24VDC Power (18 To 36V) P/N 13-24V
Allows for 18 to 36VDC (nominal 24VDC) as the main source of power to the instrument. Minimum power requirement is 35 Watts.
1.5.9 Extended Logger P/N 13-LOG
Adds almost four times the logging capacity to the existing built in logger. This option is installed at the factory. Consult factory.
Series 1300 Oxygen Deficiency Monitor PAGE 3
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
1.6 Standard Features
Analog Outputs: (2) DC mA (direct current milliampere) loop outputs.
User selectable for either 0 to 20mA or 4 to 20mA. User configurable for any oxygen sensor. Non-isolated Maximum load resistance: 300 ohms max @ 25C Isolated Option Maximum load resistance: 1200 ohms max @ 25C
Alarm Relays: (4) SPDT (single pole-double throw) Form C contacts rated at:
10 A (250 VAC) / 5A (100 VDC). Alarms are user configurable for automatic clearing or latching (requires manual clearing at the instruments panel).
Alarms: (9) individually configurable alarms that can be setup to activate any
relay, any Horn/Strobe combination, and assignable to any sensor. Each alarm can be configured as either a “high alarm” or a “low alarm”.
Input Power: Universal 90-264 VAC, 47-63 Hz.
12VDC optional and must specify at time of order. 24VDC (18-36VDC) optional and must specify at time of order.
RS Communications:
Data Logger: 1 to 60 second interval time. Selectable sensors. 0-30% O2 logging.
Audible Alarms: Internal audible alarm rated at 85 decibels at 10 centimeters
Operating Temperature:
RS-232 Communications standard. Compatible to EIA/TIA-232E. User selectable baud from 2400 to 115.2k bps. Optional assignable address from 1 to 32. Maximum of 50 feet from host system to monitor.
Included real time clock allows selectable time stamp. Individually select which available sensors to log. Output in easy to graph character separated values (CSV) for importing into spreadsheets or data visualization programs. See section Section 5 “Data Logger” for details.
50° to 104°F (10° to 40°C) 0 to 90% relative humidity (RH), non-condensing
1.7 Optional Features
The Series 1300 Oxygen Deficiency Monitor can be ordered with the following options:
Analog Outputs: Isolated, longer driving distance.
Input Power: 12V@25.2W or 24V nominal @35W (18 to 36V).
RS Communications: RS-485 add on to extend the distance from host to monitor from
50 feet to 4000 feet. Also allows for up to 32 monitors on one bus.
Data Logger: Factory installed extended memory for longer logging (almost 4x).
Audible Alarms: Up to 8 horn and strobes (factory ready).
PAGE 4 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
1.8 General Specifications
Measurement Range: Percent Oxygen: 0 to 30%.
Resolution: 0.1%
Sensor Type: Long-life electrochemical sensor (5 years typical).
Accuracy: ±1% of full scale.
Response Time: 90% of full scale in < 20 seconds.
Start Up Time: (after initial installation)
Calibration Sample Flow Rate: 0.2 to 1.0 standard cubic feet per hour (SCFH)
Calibration: Ambient air or oil-free compressed air.
2
< 1 minute.
0.1 to 0.5 liters per minute (LPM) Optimum flow using Alpha Omega Instruments calibration fixture.
Enclosures:
Polycarbonate, equivalent to NEMA 1 (IP 30)
(Main & Remotes)
Main Enclosure Dimensions: Nominal
10.94 Inches (277.8 mm) – height.
6.3 Inches (160 mm) – width.
3.6 inches (91 mm) – depth.
Remote Enclosure Dimensions: Nominal
6.2 Inches (158 mm) – height.
5.83 Inches (148 mm) – width.
3.49 inches (89 mm) – depth.
With calibration fixture:
7.3 Inches (158 mm) – height
Sample Delivery: Open diffuser-no pump required.
Display: 4 Line by 20 character LCD.
Remote Sensor Max Distance: Over 1000 feet (305 meters) using 22 AWG wire.
Longer lengths available using larger diameter wire.
2
General specifications are at standard temperature, pressure, and humidity. Unless specified otherwise.
Series 1300 Oxygen Deficiency Monitor PAGE 5
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
2 Installation
DANGER
POTENTIALLY HAZARDOUS AC VOLTAGES EXIST WITHIN THE INSTRUMENT. IF NOT AVOIDED, THESE VOLTAGES COULD POTENTIALLY RESULT IN SERIOUS INJURY OR DEATH. DISCONNECT ALL SOURCES OF POWER AND EXTERNAL CONNECTIONS BEFORE REMOVING THE COVER TO THE MONITOR.
RISK OF SHOCK
TO AVOID THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THE SERIES 1300 OXYGEN MONITOR TO RAIN, WATER SPRAY, OR ANY OTHER LIQUIDS.
2.1 Unpacking The Instrument
Upon opening the shipping container, carefully unpack the instrument to check if the outer surfaces have been damaged. If so, report the findings immediately to Alpha Omega Instruments who will, in turn, provide further instructions.
If there is no apparent damage, check the contents to ensure all items were shipped. In some cases, items may be back ordered.
NOTE: IF DAMAGE HAS BEEN FOUND, DO NOT PROCEED FURTHER, BUT INSTEAD, CONTACT THE FACTORY.
All damage and shortage claims must be made known to Alpha Omega Instruments within 10 days after receipt of shipment.
Carefully rotate the monitor and check to make sure no components have been loosened or dislodged.
If there are any loose or dislodged components (rattling of any kind), contact the factory for further instructions.
If there is no evidence of loose or dislodged components, the installation procedure can begin.
PAGE 6 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
2.2 Electrical Installation
WARNING
ELECTRICAL INSTALLATION SHOULD BE PERFORMED BY A QUALIFIED PERSON AND SHOULD COMPLY WITH APPLICABLE FEDERAL, STATE, OR LOCAL ELECTRICAL SAFETY CODES.
The standard Series 1300 Oxygen Deficiency Monitor accepts a universal AC power input of 90-264 VAC, 47-63 Hz. In this configuration, the unit is shipped with a standard North American power cord NEMA style 5-15P. The power cord is wired to an AC terminal block labeled “TB1” within the enclosure. If the installation requires hard wiring, simply replace the power cord with the appropriately rated wire. Use of #18 AWG is recommended. See section Wiring on page 10.
For DC operation the terminal block labeled “TB1” will still be used, however the labeling will indicate the configured power.
2.3 Mechanical Configuration
The Series 1300 Oxygen Deficiency Monitor can be configured as a single stand alone enclosure with a single (local) sensor located within the main enclosure. The sensor is mounted at the bottom of the main enclosure and is exposed to the surrounding atmosphere the monitor has been installed in.
The Series 1300 Oxygen Deficiency Monitor can also be configured with a Remote Oxygen Sensor that is housed in it's own separate smaller enclosure. The Series 1300 Oxygen Deficiency Monitor can accept a maximum of two remote oxygen sensors that can be placed at different locations away from the main electronics.
The Series 1300 Oxygen Deficiency Monitor main enclosure is fabricated from a polycarbonate enclosure with a cover that is secured by six (6) screws. Access to the wiring and membrane switch control panel is available by removing the cover using a standard Philips type screw driver with a #2 style point. Please exercise caution when removing the cover as it is not hinged.
Mounting holes are also accessed by removing the cover. See the following pages for more information and also APPENDIX A – Main Enclosure Template on page 66 and APPENDIX B – Remote Enclosure Template on page 67 for mounting templates.
Series 1300 Oxygen Deficiency Monitor PAGE 7
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
2.3.1 Main Enclosure
Figure 2 - Main Enclosure
PAGE 8 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
See section 3 “Wiring“ on page 10 for wiring details.
WARNING
THE SERIES 1300 OXYGEN DEFICIENCY MONITOR AND THE REMOTE OXYGEN SENSOR ENCLOSURES SHOULD BE KEPT IN THEIR DESIGNED WALL MOUNTED ORIENTATION WITH THE SENSOR FACING DOWNWARDS.
2.3.2 Remote Oxygen Sensor Enclosure
The Remote Oxygen Sensor enclosure is configured similarly in a polycarbonate enclosure with a cover secured by four (4) screws. Remove the cover of the Remote Oxygen Sensor enclosure to gain access to both the printed circuit board connections and the mounting holes. See APPENDIX B – Remote Enclosure Template on page 67 for a mounting template.
Wiring will be through the strain relief located on the right side of the remote enclosure. See section
3.3 “Remote Oxygen Sensors” on page 14 for details.
Figure 3 - Remote Oxygen Sensor Enclosure
CAUTION
DO NOT OPERATE THE SERIES 1300 OXYGEN DEFICIENCY MONITOR OR THE REMOTE OXYGEN SENSORS IN ANY POSITION OTHER THAN WITH THE SENSOR MOUNTED DOWNWARD AS SHOWN. A DIFFERENT ORIENTATION MAY ADVERSELY AFFECT THE OXYGEN VALUES.
Series 1300 Oxygen Deficiency Monitor PAGE 9
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
3 Wiring
The Series 1300 Oxygen Deficiency Monitor main enclosure is designed to accept a number of inputs and outputs. Please reference Figure 4 below to wire to the following hardware. Note: Following this page are detailed instruction for each of the following items:
Oxygen Sensor A single “Local Oxygen Sensor” mounted at the bottom center of the enclosure. This is a
two (2) wire interface located on the left side, inside the main enclosure.
Power AC or DC power via terminal block (TB1). Labeled at the factory. Figure 4 - Bottom View 1
of Series 1300 below Shows the AC version labeled “L”, “N”, and “G” for Line, Neutral, and Ground respectively. For DC operation these will be “+”, “-”, and “G” respectively.
Remote Sensors One (1) or Two (2) Remote Oxygen Sensors via terminal block TB1, terminals 1 thru 6. See section 3.3
Horn and Strobes Up to eight (8) horn and strobes (where four can be powered directly without external
power supplies) via terminal block TB1, terminals 7 thru 11.
Analog Outputs Removable printed circuit board (PCB) terminal block TB2 See section 3.5
Alarm Relays Removable printed circuit board (PCB) terminal blocks TB3 & TB4 See section 3.6
Digital Communications (RS-232 or RS-485) via terminal block TB1, terminals labeled “T”, “R”, and “G”. See section 3.7
See section 3.1
See section 3.2
See section 3.4
Figure 4 below shows a cut-away view of the internal terminal block TB1.
Figure 4 - Bottom View 1 Of Series 1300
PAGE 10 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
Figure 5 below shows a cut-away line drawing showing the PCB terminal blocks TB2, TB3, and TB4.
Figure 5 - Bottom View 2 Of Series 1300
3.1 Local Oxygen Sensor
If ordered with a Local Oxygen Sensor, the Series 1300 Oxygen Deficiency Monitor will have a large gland that is centrally located on the bottom of the enclosure as shown above. This hole will be plugged when the Series 1300 Oxygen Deficiency Monitor is ordered without a Local Oxygen Sensor.
Figure 6 - Oxygen Sensor
Series 1300 Oxygen Deficiency Monitor PAGE 11
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
3.1.1 Local Oxygen Sensor Installation
To install the replaceable local sensor, simply open the front cover and slide the sensor down into the large gland designed to hold the sensor. Gently tighten the nut while holding the sensor in place from the top and the bottom at the same time as shown.
Figure 7 - Local Oxygen Sensor Installation
PAGE 12 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
The sensors wired connector should then be pressed together with the mating connector already supplied inside the main enclosure down at the bottom on the left. The mating connector is in the form of a cable with a two pin male connection as shown below before and after connection:
Before After
Figure 8 - Local Oxygen Sensor Installation & Wiring
3.2 Power (TB1)
Power the Series 1300 Oxygen Deficiency Monitor by wiring to terminal block TB1. Please Note that the terminal blocks are color coded. The two primary power connectors are blue and the ground returns are a yellow/green color.
Universal AC
90-264 VAC, 47-63 Hz.
TB1
LabelACSignal
L Line + Positive (12V) Positive (24V)
N Neutral - Negative (12V Return) Negative (24V Return)
G Ground G Chassis Ground Chassis Ground
TB1
Label
12V Signal
Table 1 - Power Connections (Terminal Block 1)
DC
24V (18 to 36V) Signal
Series 1300 Oxygen Deficiency Monitor PAGE 13
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
3.3 Remote Oxygen Sensors
The interface to the “Remote Oxygen Sensor”s are located at terminal block TB1 (Refer to Figure 4) inside the main enclosure. The Remote Oxygen Sensors are connected using a twisted shielded pair of wires. The connections within the main electronics enclosure are located on a terminal block located in the bottom of the enclosure labeled “TB1”. See Table 2 below for reference. Terminals 1-6 are used to connect to the Remote Oxygen Sensors as shown in the table below. These connections are terminated at the Remote Oxygen Sensor enclosure by means of another terminal block (also labeled TB1) located within the remote electronics enclosure. See Figure 10 - Remote Oxygen Sensor Enclosure & PCB.
Connects
from:
Main
Electronics
TB1
Label Signal
1 Shield
2 Positive 2 Positive, source of current loop
3 Return 3 Return, Negative return of current loop
4 Shield
5 Positive 2 Positive, source of current loop
6 Return 3 Return, Negative return of current loop
Connects to:
Remote Oxygen Sensor 1
Remote Oxygen Sensor 2
TB1
Label Detail
1 Shield
1 Shield
Table 2 - Remote Oxygen Sensor Connections (Terminal Block 1)
Figure 9 - Remote Oxygen Sensor Enclosure
Inside the Remote Oxygen Sensor enclosure is a PCB as shown in Figure 10 below.
PAGE 14 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
3
TB1
2 1
Figure 10 - Remote Oxygen Sensor Enclosure & PCB Figure 11 - Remote Oxygen Sensor
The sensor is connected to P1 with the red wire towards the edge of the PCB. The connector is a two (2) pin Molex style. Note that only the “+” and “-” pins are used.
3.3.1 Remote Oxygen Sensor Installation Procedure
Follow the procedure below for connecting the first Remote Oxygen Sensor.
CAUTION
REMOVE ANY POWER FROM THE INSTRUMENT BEFORE PERFORMING ANY WIRING ON THE MAIN INSTRUMENT OR THE REMOTE OXYGEN SENSOR.
NOTE: For best results it is recommended to use a twisted shielded pair to connect to the Remote Oxygen Sensors. Using #22 AWG (American Wiring Gauge) will allow distances up to 1000'. Longer lengths available with larger diameter wire (consult factory). In most cases the proper cable is specified at time of order and will be supplied by the factory.
Connecting Remote Oxygen Sensor #1 (refer to Table 2 - Remote Oxygen Sensor Connections (Terminal Block 1) on page 14 above):
1. Connect the SHIELD wire of the twisted pair shielded cable to TB1, terminal 1 of the main electronics. This wire should terminate at terminal block TB1, terminal 1 inside the Remote Oxygen Sensor. Make sure all wires are put through the strain relief first.
2. Connect the POSITIVE wire of the twisted pair shielded cable to TB1, terminal 2 of the main electronics. This wire should terminate at terminal block TB1, terminal 2 inside the Remote Oxygen Sensor.
3. Connect the RETURN wire of the twisted pair shielded cable to TB1, terminal 3 of the main electronics. This wire should terminate at terminal block TB1, terminal 3 inside the Remote Oxygen Sensor.
Series 1300 Oxygen Deficiency Monitor PAGE 15
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
Repeat the above procedure to connect the second Remote Oxygen Sensor if applicable but substitute the terminals as follows:
1. Connect the SHIELD (TB1, terminal 4) of the main electronics to TB1, terminal 1 inside the 2 Remote Oxygen Sensor.
2. Connect the POSITIVE (TB1, terminal 5) of the main electronics to TB1, terminal 2 inside the 2 Remote Oxygen Sensor
3. Connect THE RETURN (TB1, terminal 6) of the main electronics to TB1, terminal 3 inside the 2 Remote Oxygen Sensor
nd
nd
nd
After connection to the Remote Oxygen Sensor(s) has been completed, power the instrument and wait for it to initialize. The display should now show each sensor being initialized. Please reference section 6 “Sensor Setup” on page 47 for more information on setting up the sensor(s).
NOTE: Remote Oxygen Sensors do not need to be connected to the unit sequentially. They will be detected based on the terminals they are connected to.
3.4 Horn And Strobe
The interface to the Horn and Strobes is located at terminal block TB1 (Refer to Figure 4) inside the main enclosure. The Series 1300 Oxygen Deficiency Monitor is capable of supporting up to 8 horn and strobes. The horn and strobes use a 3 wire connection3 consisting of power, ground return, and data. The connections within the main electronics enclosure are located on a terminal block located in the bottom of the enclosure labeled “TB1”. See Table 3 below for reference.
Main Enclosure Horn and Strobe PCB
Terminal
Block
TB 1 Signal Name on PCB Detail
7 Shield
8 Shield “SH” Shield
9 Positive “V+” Positive Power (typically 12V)
10 Data “DAT” Data communications line
11 Return “GND” Power Return
Table 3 - Horn And Strobe Connections (Terminal Block 1)
Follow the procedure below for connecting a horn and strobe that will be powered from the Series 1300 Oxygen Deficiency Monitor.
1. Connect the 1st wire of the shielded cable to TB1 terminal 9 (POSITIVE), inside the main enclosure. Terminate this wire to one of the terminals on the horn and strobe PCB labeled “V+”.
2. Likewise connect the 2nd wire of the shielded cable to TB1 terminal 10 (DATA). Terminate this wire to one of the terminals on the horn and strobe PCB labeled “DAT”.
3. Connect the 3rd wire of the shielded cable to TB1 terminal 11 (RETURN). Terminate this wire to one of the terminals on the horn and strobe PCB labeled “GND“.
4. Connect the shield of the shielded cable to TB1 terminal 8 (SHIELD). Terminate this shield to one of the terminals on the horn and strobe PCB labeled “SH”.
Note: it is only necessary to connect the shield at the horn and strobe PCB when daisy-chaining4. In a star-point configuration, simply leave it disconnected.
3
Horn and Strobes can be wired directly to the relay contacts using only two (2) wires. The Series 1300 Oxygen Deficiency
Monitor will not be able to warn the user of any disconnect or wiring problem without the use of the “DAT” line.
PAGE 16 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
Follow the same procedure to connect multiple horn and strobe units to the Series 1300. The main enclosure and the horn and strobe unit both have multiple terminals that can accept more than a single wire twisted together if necessary. If daisy-chaining, then simply connect to the second set of terminals supplied on the horn and strobe unit.
When powered from 90-264 VAC, 47-63 Hz., the instruments default DC output power is capable of driving up to 4 horn and strobes simultaneously. If more than 4 are required, the additional horn and strobes need to be powered from either an independent DC power source or an upgraded power supply that can handle the extra horn and strobes. The maximum length of cable for a single horn and strobe wired with the recommended #18 AWG wire can be up to 1,000 feet away from the main electronics. However, each additional horn and strobe wired in SERIES (daisy-chained) on the SAME wire will reduce the maximum length by approximately half. The maximum voltage drop in any single chain of wire must be less than or equal to 3V. There must be a minimum of 9V at the end of the wire.
CAUTION
THE MAX CABLE CAPACITANCE OF 0.15UF MUST BE ADHERED TO OR THE HORN AND STROBES MAY NOT WORK CORRECTLY AND/OR CAUSE UNWANTED BEHAVIOR. PLEASE CONTACT THE FACTORY FOR ANY APPLICATIONS REQUIRING LONGER LENGTHS.
Figure 12 below shows a horn and strobe and the associated printed circuit board (PCB).
Figure 12 - Horn And Strobe With PCB
3.4.1 Horn And Strobe PCB
The printed circuit board has 8 terminals located on the front edge as shown in Figure 13 - Horn and Strobe PCB on page 17. The signals are labeled on the board as “DAT” for data, “V+” for positive power, “GND” for power return, and “SH” for shield. Each signal has two connections to allow daisy­chaining. If daisy-chaining multiple horn and strobes, the connectors are sized to allow for twisting two wires together before inserting into the screw terminal.
Figure 13 - Horn And Strobe PCB
4
Daisy-chaining is when another Horn and Strobe is connected to the first in series. As opposed to being connected to the
main enclosure in parallel (or star-point).
Series 1300 Oxygen Deficiency Monitor PAGE 17
USER MANUAL ALPHA OMEGA INSTRUMENTS CORP.
Series 1300
Optionally wire up to (4) more.
This requires
external Power
or a higher
rated internal
power supply.
CAUTION
J1 TO J4 SUPPORT A MAXIMUM OF FOUR (4) HORN AND STROBES UNLESS USING A SPECIAL HIGH POWERED MAIN POWER SUPPLY. NEVER CONNECT AN EXTERNAL POWER SUPPLY TO THE “V+” TERMINALS.
Note, that if powering the instrument from an external power supply, it is imperative that the power supplied be sufficient for the number of horn and strobes connected at any given time. (5) or more horn and strobes will require remote power or a higher wattage upgrade to the internal power supply.
The additional external power needed to handle up to four (4) standard horn and strobes is approximately 20W. Therefore, using a 45W, 12V power supply would be able to power the Series 1300 Oxygen Deficiency Monitor as well as all eight (8) horn and strobes.
When externally powered, connect only the “DAT” and “GND” terminals back to the main electronics enclosure. The “V+” terminal can be omitted in this case as it is not powered from the main electronics. The external power supply connects to “EXT PWR” terminal located on the right hand side labeled “J5”.
CAUTION
REMOVE POWER BEFORE PERFORMING ANY WIRING ON THE INSTRUMENT OR HORN AND STROBE(S). IF EQUIPPED WITH BATTERY BACKUP, INSURE THE BATTERY TERMINAL INSIDE THE MAIN ENCLOSURE IS DISCONNECTED.
3.4.2 Horn And Strobe Daisy Chain Configuration
This configuration shows how the wiring can be configured in a serial fashion or “daisy chain”. The source wiring comes in on one set of connectors and the next horn and strobe is wired to the other set of connectors as shown below.
Figure 14 - Wiring Horn And Strobes In Series (Daisy Chained)
As shown above, using four (4) horn and strobes and #18 AWG cable wire, the maximum distance to the last horn and strobe in the chain is 250 feet.
PAGE 18 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP. USER MANUAL
Series 1300
Series 1300
12V
3.4.3 Horn And Strobe Star Configuration
Shown below is a Series 1300 Oxygen Deficiency Monitor wired to four (4) horn and strobes.
Figure 15 - Wiring Horn And Strobes Individually
As shown above, using four (4) horn and strobes and #18 AWG cable wire, the maximum distance fore each single horn and strobe is 1,000 feet.
3.4.4 Horn And Strobe External Power
Adding more than four (4) horn and strobes with the standard power supply requires external power. Shown below is the wiring for externally powered horn and strobes. Note the added 12V power supply that is connected to a separate connector.
CAUTION
DO NOT CONNECT EXTERNAL POWER TO THE V+ TERMINAL. EXTERNAL POWER SHOULD ONLY BE APPLIED TO THE “EXT PWR” CONNECTOR LABELED “J5”
Figure 16 - Externally Powered Horn And Strobes
Wiring of the V+ is optional and shown as a dashed line in the picture.
Series 1300 Oxygen Deficiency Monitor PAGE 19
Loading...
+ 60 hidden pages