Alpha Omega Instruments Corp. warrants that delivered products are free from defects in material
and workmanship at the time of delivery to the FOB point specified in the purchase order. Liability
under this warranty is limited to repairing or replacing, at Alpha Omega Instruments' option, items
which are returned to it prepaid within four years (including sensor) from the date of shipment and
found to Seller’s satisfaction to have been defective.
Alpha Omega Instrument's four (4) year sensor warranty provides protection for four years from the
date of shipment of the Series 1300 Oxygen Deficiency Monitor. Any sensor from a Series 1300
Oxygen Deficiency Monitor that fails under normal use must be returned to Seller prepaid and, if such
sensor is determined by Seller to be defective, Seller shall provide Buyer with a replacement sensor.
Buyer must provide the serial number of the monitor from which the sensor has been removed. If a
sensor is found to be defective and a new one issued, the warranty of the replacement sensor(s) shall
not extend beyond the initial warranty period of the Series 1300 Oxygen Deficiency Monitor, or for a
period of one year, whichever is longer. After this, all replacement sensors will be warranted for a
period of one year from the date of shipment. In no event shall Alpha Omega Instruments Corp. be
liable for consequential damages.
NO PRODUCT IS WARRANTED AS BEING FIT FOR A PARTICULAR
PURPOSE AND THERE IS NO WARRANTY OF MERCHANTABILITY.
This warranty applies only if:
(i) the items are used solely under the operating conditions and manner recommended in this manual, product
specifications, or other product specific literature;
(ii) the items have not been misused or abused in any manner or unauthorized repairs were attempted thereon;
(iii) written notice of the failure within the warranty period is forwarded to Alpha Omega Instruments Corp. and, the
directions received for properly identifying items returned under warranty are followed;
(iv) the return notice authorizes Alpha Omega Instruments Corp. to examine and disassemble returned products to the
extent the Company deems necessary to ascertain the cause of failure.
The warranties stated herein are exclusive. THERE ARE NO OTHER WARRANTIES, EITHER
EXPRESSED OR IMPLIED, BEYOND THOSE SET FORTH HEREIN, and Alpha Omega Instruments
Corp. does not assume any other obligation or liability in connection with the sale or use of said
products.
Disclaimer of Warranty
Alpha Omega Instruments Corp. makes no representation or warranties, either expressed or implied, by or with respect to
anything in this manual, including, but not limited to, implied warranties of merchantability or fitness for a particular purpose.
In no event will Alpha Omega Instruments Corp. be liable for any damages, whether direct or indirect, special, consequential,
or incidental arising from the use of this manual. Some states in the USA do not allow the exclusion of incidental or
consequential damages. Alpha Omega Instruments Corp. also reserves the right to make any changes to improve the
performance of its products at any time and without notice.
4.8.10 Field Elevation Adjustment.................................................................................................44
4.8.11 System Information (“SYSTEM INFO”)...............................................................................44
5 Data Logger.........................................................................................................................................45
Table 24 - Horn And Strobe Sound Switch Settings...............................................................................71
PAGE viSeries 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
1 Overview
1.1 Explanation Of Graphic Symbols
The information listed below is essential to the proper operation of the monitor. Please review the
following safety precautions prior to using the monitor. Cautionary notes are included throughout this
manual.
WARNING MESSAGE
THIS SYMBOL IS INTENDED TO ALERT THE USER ABOUT GENERAL WARNINGS
AND IMPORTANT OPERATING AND MAINTENANCE INSTRUCTIONS.
WARNING
HAZARDOUS VOLTAGE. CONTACT MAY CAUSE ELECTRIC SHOCK OR BURN.
TURN OFF & LOCK OUT SYSTEM BEFORE SERVICING.
THIS SYMBOL IS INTENDED TO ALERT THE USER TO POTENTIALLY
DANGEROUS SITUATIONS.
1.2 Instrument Description
The Series 1300 Oxygen Deficiency Monitor allows for the simultaneous monitoring of breathing air
oxygen using up to three (3) separate oxygen sensors. For monitoring of a single location, the Series
1300 supports either a “Local Sensor” (located within the main electronics enclosure) or a “Remote
Sensor” (located in a separate remote electronics enclosure). For monitoring of multiple locations, up
to two (2) “Remote Sensors” can be added to the “Local Sensor” for a total of three (3) monitoring
locations. The standard input power to the Series 1300 instrument is 88-264VAC, 47-63Hz (Hertz).
Available power options include 12VDC or 18-36VDC (24VDC) and must be specified at the time of
order placement.
The Series 1300 Oxygen Deficiency Monitor is housed in a NEMA 1 equivalent enclosure rated for
general purpose indoor use. Oxygen values for all sensors are displayed on a 4 line by 20 character
liquid crystal display (LCD) with a total height of 0.81" (20.8 mm). Various display modes are available
that allow for viewing all sensors simultaneously or having the sensor oxygen values cycle in a large
font display mode. The front panel includes a membrane panel with seven buttons that provide
access to the monitor's settings. The monitor is equipped with four alarm relays (see specifications)
which are user configurable and can be set for fail-safe operation. Also included, is a built-in audible
indication of an oxygen alarm condition.
The Series 1300 Oxygen Deficiency Monitor comes equipped with two direct current (DC) analog
outputs. The analog outputs can be independently configured for 0-20mA or 4-20mA. Both are
scalable over the operating range of the monitor. Using a terminating resistor on either of these
outputs will provide an analog voltage up to 5 volts full scale up to a maximum of 1,000 feet.
The Series 1300 Oxygen Deficiency Monitor comes equipped standard with a data-logger and a
serial interface commonly referred to as RS-232. The interface is compatible with EIA/TIA-232E.
Optional addresses from 1 to 32 may be assigned to each monitor for communicating with up to 32
instruments on the RS-2321 or RS-485 bus.
1
The built in standard RS-232 by definition does not allow for more than one instrument on the same physical wires. An
external RS-232 to RS-485 adapter or an “Ethernet Serial server” can be used to allow addressing multiple RS-232 units by
converting the RS-232 signals to an alternate bus.
Series 1300 Oxygen Deficiency MonitorPAGE 1
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
1.2.1 Features
Standard Features:
•Easy to use and configure.
•Capable of monitoring up to three oxygen
sensors simultaneously.
•Automatically senses and configures new
sensors.
•Multiple monitoring screens: detailed or
large font versions.
•Alarm status is clearly displayed on the
screen with pop-up for active alarms
•Scrolling collapsible menu system is easy
to navigate.
•Wide range of configurable settings.
•Any sensor can be assigned to any output.
Standard Features (continued):
•Any Alarm can be assigned to any sensor.
•Any Alarm can activate any combination
of relays, horns, and strobes.
•Isolated digital communications (RS-232
or RS485).
•Built-in data-logger.
Optional Features:
•Battery backup.
•Isolated process outputs.
•Extended memory Data-logger (capable
of almost 4x more logging capacity).
1.3 Mounting
Wall Mount (WMT) is the basic configuration for the Series 1300 Oxygen Deficiency Monitor. All
electrical connections are located on the bottom of the monitor. See APPENDIX A – Main Enclosure
Template on page 66 for detailed dimensional drawings.
1.4 Oxygen Sensor
The Series 1300 Oxygen Deficiency Monitor features an ambient temperature electrochemical sensor
with an Enhanced Electrolyte System (EES). The EES, significantly extends the useful life of the
oxygen sensor as well as providing exceptional measurement stability.
Figure 1 - Oxygen Sensor
PAGE 2 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
1.5 Description Of Optional Equipment
The Series 1300 Oxygen Deficiency Monitor incorporates a host of standard features. For certain
requirements, user's may desire to augment the capabilities of the monitor by ordering the monitor with one or
more available options described below.
1.5.1 Battery BackupP/N 13-BAT
The Battery Backup option provides backup power to the Series 1300 Oxygen Deficiency Monitor during a
temporary loss of main power. Please see section 8 Battery Backup for more information on this option.
1.5.2 RS-485P/N 13-485
The factory installed RS-485 is an optional upgrade to the factory default of RS-232 and offers the ability to
extend the distance from 50 feet maximum up to 4,000 feet between the monitor and the host system. This
option also provides serial communications (both sending and receiving ) between several monitors up to a
maximum of 32 on one RS-485 three wire communications channel.
1.5.3 Replacement Oxygen SensorP/N 13-SEN
This option consists of just the sensor (no enclosure or electronics). This option should be ordered in the event
that the sensor is damaged or needs replacement due to end of life.
1.5.4 Remote Mounted Sensor EnclosureP/N 13-RSEN
This option consists of a NEMA 1 (IP 30) remote sensor enclosure containing the oxygen sensor and
associated circuitry used to communicate with the read out electronics. The remote sensors may be added to
the Series 1300 Oxygen Deficiency Monitor in the field by the user. See section 3.3 Remote Oxygen Sensors.
1.5.5 Analog Output IsolationP/N 13-ISO
Installed at the factory, this option provides galvanic isolation of both analog outputs. This eliminates any
potential ground loops when connecting to earth grounded inputs or powering from DC where the power return
could create a ground loop with the output grounds or the input and output grounds are tied to chassis ground.
The isolation also allows for a higher maximum load resistance up to 1200 ohms (24V @ 20mA) as opposed to
the 12V maximum of 300 ohms (6V @ 20mA).
1.5.6 Calibration FixtureP/N 13-CFN
The optional calibration fixture is recommended for checking the calibration of the Series 1300 Oxygen
Deficiency Monitor. Ideally, the source of the calibration gas should be compressed air (clean, dry, and oil free).
Most industrial gas suppliers can provide the compressed air in small hand-held cylinders. The calibration
fixture is designed to easily install over the gas opening of the Series 1300 oxygen sensor making a tight seal.
The fixture is equipped with ¼ inch tube fittings to deliver the compressed air through a 1/4" OD plastic tube
(not supplied). The sample pressure to the fixture should never exceed 1.5 pounds per square inch (PSIG) at a
corresponding sample flow rate of between 0.2 to 1.0 standard cubic feet per hour (SCFH) or 0.1 to 0.5 liters
per minute (LPM).
1.5.7 12VDC PowerP/N 13-12V
Allows for a user supplied regulated 12V as the main source of power to the instrument. Minimum power
requirement is 25.2 Watts.
1.5.8 24VDC Power (18 To 36V)P/N 13-24V
Allows for 18 to 36VDC (nominal 24VDC) as the main source of power to the instrument. Minimum power
requirement is 35 Watts.
1.5.9 Extended LoggerP/N 13-LOG
Adds almost four times the logging capacity to the existing built in logger. This option is installed at
the factory. Consult factory.
Series 1300 Oxygen Deficiency MonitorPAGE 3
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
1.6 Standard Features
Analog Outputs:(2) DC mA (direct current milliampere) loop outputs.
User selectable for either 0 to 20mA or 4 to 20mA.
User configurable for any oxygen sensor.
Non-isolated Maximum load resistance: 300 ohms max @ 25C
Isolated Option Maximum load resistance:1200 ohms max @ 25C
Alarm Relays:(4) SPDT (single pole-double throw) Form C contacts rated at:
10 A (250 VAC) / 5A (100 VDC).
Alarms are user configurable for automatic clearing or latching
(requires manual clearing at the instruments panel).
Alarms:(9) individually configurable alarms that can be setup to activate any
relay, any Horn/Strobe combination, and assignable to any sensor.
Each alarm can be configured as either a “high alarm” or a “low alarm”.
Input Power:Universal 90-264 VAC, 47-63 Hz.
12VDC optional and must specify at time of order.
24VDC (18-36VDC) optional and must specify at time of order.
RS
Communications:
Data Logger:1 to 60 second interval time. Selectable sensors. 0-30% O2 logging.
Audible Alarms:Internal audible alarm rated at 85 decibels at 10 centimeters
Operating
Temperature:
RS-232 Communications standard. Compatible to EIA/TIA-232E. User
selectable baud from 2400 to 115.2k bps. Optional assignable address
from 1 to 32. Maximum of 50 feet from host system to monitor.
Included real time clock allows selectable time stamp.
Individually select which available sensors to log.
Output in easy to graph character separated values (CSV) for
importing into spreadsheets or data visualization programs.
See section Section 5 “Data Logger” for details.
50° to 104°F (10° to 40°C)
0 to 90% relative humidity (RH), non-condensing
1.7 Optional Features
The Series 1300 Oxygen Deficiency Monitor can be ordered with the following options:
Analog Outputs:Isolated, longer driving distance.
Input Power:12V@25.2W or 24V nominal @35W (18 to 36V).
RS Communications:RS-485 add on to extend the distance from host to monitor from
50 feet to 4000 feet. Also allows for up to 32 monitors on one bus.
Data Logger:Factory installed extended memory for longer logging (almost 4x).
Audible Alarms:Up to 8 horn and strobes (factory ready).
PAGE 4 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
1.8 General Specifications
Measurement Range:Percent Oxygen: 0 to 30%.
Resolution:0.1%
Sensor Type:Long-life electrochemical sensor (5 years typical).
Accuracy:±1% of full scale.
Response Time:90% of full scale in < 20 seconds.
Start Up Time:
(after initial installation)
Calibration Sample Flow Rate:0.2 to 1.0 standard cubic feet per hour (SCFH)
Calibration:Ambient air or oil-free compressed air.
2
< 1 minute.
0.1 to 0.5 liters per minute (LPM)
Optimum flow using Alpha Omega Instruments
calibration fixture.
Enclosures:
Polycarbonate, equivalent to NEMA 1 (IP 30)
(Main & Remotes)
Main Enclosure Dimensions:
Nominal
10.94 Inches (277.8 mm) – height.
6.3 Inches (160 mm) – width.
3.6 inches (91 mm) – depth.
Remote Enclosure Dimensions:
Nominal
6.2 Inches (158 mm) – height.
5.83 Inches (148 mm) – width.
3.49 inches (89 mm) – depth.
With calibration fixture:
7.3 Inches (158 mm) – height
Sample Delivery:Open diffuser-no pump required.
Display:4 Line by 20 character LCD.
Remote Sensor Max Distance:Over 1000 feet (305 meters) using 22 AWG wire.
Longer lengths available using larger diameter
wire.
2
General specifications are at standard temperature, pressure, and humidity. Unless specified otherwise.
Series 1300 Oxygen Deficiency MonitorPAGE 5
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
2 Installation
DANGER
POTENTIALLY HAZARDOUS AC VOLTAGES EXIST WITHIN THE INSTRUMENT. IF
NOT AVOIDED, THESE VOLTAGES COULD POTENTIALLY RESULT IN SERIOUS
INJURY OR DEATH. DISCONNECT ALL SOURCES OF POWER AND EXTERNAL
CONNECTIONS BEFORE REMOVING THE COVER TO THE MONITOR.
RISK OF SHOCK
TO AVOID THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THE
SERIES 1300 OXYGEN MONITOR TO RAIN, WATER SPRAY, OR ANY OTHER
LIQUIDS.
2.1 Unpacking The Instrument
Upon opening the shipping container, carefully unpack the instrument to check if the outer surfaces
have been damaged. If so, report the findings immediately to Alpha Omega Instruments who will, in
turn, provide further instructions.
If there is no apparent damage, check the contents to ensure all items were shipped. In some cases,
items may be back ordered.
NOTE: IF DAMAGE HAS BEEN FOUND, DO NOT PROCEED FURTHER, BUT INSTEAD, CONTACT THE
FACTORY.
All damage and shortage claims must be made known to Alpha Omega Instruments within 10 days
after receipt of shipment.
Carefully rotate the monitor and check to make sure no components have been loosened or
dislodged.
If there are any loose or dislodged components (rattling of any kind), contact the factory for further
instructions.
If there is no evidence of loose or dislodged components, the installation procedure can begin.
PAGE 6 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
2.2 Electrical Installation
WARNING
ELECTRICAL INSTALLATION SHOULD BE PERFORMED BY A QUALIFIED
PERSON AND SHOULD COMPLY WITH APPLICABLE FEDERAL, STATE, OR
LOCAL ELECTRICAL SAFETY CODES.
The standard Series 1300 Oxygen Deficiency Monitor accepts a universal AC power input of 90-264
VAC, 47-63 Hz. In this configuration, the unit is shipped with a standard North American power cord
NEMA style 5-15P. The power cord is wired to an AC terminal block labeled “TB1” within the
enclosure. If the installation requires hard wiring, simply replace the power cord with the appropriately
rated wire. Use of #18 AWG is recommended. See section Wiring on page 10.
For DC operation the terminal block labeled “TB1” will still be used, however the labeling will indicate
the configured power.
2.3 Mechanical Configuration
The Series 1300 Oxygen Deficiency Monitor can be configured as a single stand alone enclosure with
a single (local) sensor located within the main enclosure. The sensor is mounted at the bottom of the
main enclosure and is exposed to the surrounding atmosphere the monitor has been installed in.
The Series 1300 Oxygen Deficiency Monitor can also be configured with a Remote Oxygen Sensor
that is housed in it's own separate smaller enclosure. The Series 1300 Oxygen Deficiency Monitor
can accept a maximum of two remote oxygen sensors that can be placed at different locations away
from the main electronics.
The Series 1300 Oxygen Deficiency Monitor main enclosure is fabricated from a polycarbonate
enclosure with a cover that is secured by six (6) screws. Access to the wiring and membrane switch
control panel is available by removing the cover using a standard Philips type screw driver with a #2
style point. Please exercise caution when removing the cover as it is not hinged.
Mounting holes are also accessed by removing the cover. See the following pages for more
information and also APPENDIX A – Main Enclosure Template on page 66 and APPENDIX B –
Remote Enclosure Template on page 67 for mounting templates.
Series 1300 Oxygen Deficiency MonitorPAGE 7
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
2.3.1 Main Enclosure
Figure 2 - Main Enclosure
PAGE 8 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
See section 3 “Wiring“ on page 10 for wiring details.
WARNING
THE SERIES 1300 OXYGEN DEFICIENCY MONITOR AND THE REMOTE OXYGEN
SENSOR ENCLOSURES SHOULD BE KEPT IN THEIR DESIGNED WALL
MOUNTED ORIENTATION WITH THE SENSOR FACING DOWNWARDS.
2.3.2 Remote Oxygen Sensor Enclosure
The Remote Oxygen Sensor enclosure is configured similarly in a polycarbonate enclosure with a
cover secured by four (4) screws. Remove the cover of the Remote Oxygen Sensor enclosure to gain
access to both the printed circuit board connections and the mounting holes. See APPENDIX B –
Remote Enclosure Template on page 67 for a mounting template.
Wiring will be through the strain relief located on the right side of the remote enclosure. See section
3.3 “Remote Oxygen Sensors” on page 14 for details.
Figure 3 - Remote Oxygen Sensor Enclosure
CAUTION
DO NOT OPERATE THE SERIES 1300 OXYGEN DEFICIENCY MONITOR OR THE
REMOTE OXYGEN SENSORS IN ANY POSITION OTHER THAN WITH THE
SENSOR MOUNTED DOWNWARD AS SHOWN. A DIFFERENT ORIENTATION MAY
ADVERSELY AFFECT THE OXYGEN VALUES.
Series 1300 Oxygen Deficiency MonitorPAGE 9
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
3 Wiring
The Series 1300 Oxygen Deficiency Monitor main enclosure is designed to accept a number of inputs
and outputs. Please reference Figure 4 below to wire to the following hardware. Note: Following this
page are detailed instruction for each of the following items:
Oxygen SensorA single “Local Oxygen Sensor” mounted at the bottom center of the enclosure. This is a
two (2) wire interface located on the left side, inside the main enclosure.
PowerAC or DC power via terminal block (TB1). Labeled at the factory. Figure 4 - Bottom View 1
of Series 1300 below Shows the AC version labeled “L”, “N”, and “G” for Line, Neutral, and
Ground respectively. For DC operation these will be “+”, “-”, and “G” respectively.
Remote SensorsOne (1) or Two (2) Remote Oxygen Sensors via terminal block TB1, terminals 1 thru 6.See section 3.3
Horn and StrobesUp to eight (8) horn and strobes (where four can be powered directly without external
power supplies) via terminal block TB1, terminals 7 thru 11.
Digital Communications (RS-232 or RS-485) via terminal block TB1, terminals labeled “T”, “R”, and “G”.See section 3.7
See section 3.1
See section 3.2
See section 3.4
Figure 4 below shows a cut-away view of the internal terminal block TB1.
Figure 4 - Bottom View 1 Of Series 1300
PAGE 10 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
Figure 5 below shows a cut-away line drawing showing the PCB terminal blocks TB2, TB3, and TB4.
Figure 5 - Bottom View 2 Of Series 1300
3.1 Local Oxygen Sensor
If ordered with a Local Oxygen Sensor, the Series 1300 Oxygen Deficiency Monitor will have a large
gland that is centrally located on the bottom of the enclosure as shown above. This hole will be
plugged when the Series 1300 Oxygen Deficiency Monitor is ordered without a Local Oxygen Sensor.
Figure 6 - Oxygen Sensor
Series 1300 Oxygen Deficiency MonitorPAGE 11
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
3.1.1 Local Oxygen Sensor Installation
To install the replaceable local sensor, simply open the front cover and slide the sensor down into the
large gland designed to hold the sensor. Gently tighten the nut while holding the sensor in place from
the top and the bottom at the same time as shown.
Figure 7 - Local Oxygen Sensor Installation
PAGE 12 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
The sensors wired connector should then be pressed together with the mating connector already
supplied inside the main enclosure down at the bottom on the left. The mating connector is in the
form of a cable with a two pin male connection as shown below before and after connection:
BeforeAfter
Figure 8 - Local Oxygen Sensor Installation & Wiring
3.2 Power (TB1)
Power the Series 1300 Oxygen Deficiency Monitor by wiring to terminal block TB1. Please Note that
the terminal blocks are color coded. The two primary power connectors are blue and the ground
returns are a yellow/green color.
The interface to the “Remote Oxygen Sensor”s are located at terminal block TB1 (Refer to Figure 4)
inside the main enclosure. The Remote Oxygen Sensors are connected using a twisted shielded pair
of wires. The connections within the main electronics enclosure are located on a terminal block
located in the bottom of the enclosure labeled “TB1”. See Table 2 below for reference. Terminals 1-6
are used to connect to the Remote Oxygen Sensors as shown in the table below. These connections
are terminated at the Remote Oxygen Sensor enclosure by means of another terminal block (also
labeled TB1) located within the remote electronics enclosure. See Figure 10 - Remote Oxygen
Sensor Enclosure & PCB.
The sensor is connected to P1 with the red wire towards the edge of the PCB. The connector is a two
(2) pin Molex style. Note that only the “+” and “-” pins are used.
3.3.1 Remote Oxygen Sensor Installation Procedure
Follow the procedure below for connecting the first Remote Oxygen Sensor.
CAUTION
REMOVE ANY POWER FROM THE INSTRUMENT BEFORE PERFORMING ANY
WIRING ON THE MAIN INSTRUMENT OR THE REMOTE OXYGEN SENSOR.
NOTE: For best results it is recommended to use a twisted shielded pair to connect to the Remote
Oxygen Sensors. Using #22 AWG (American Wiring Gauge) will allow distances up to 1000'. Longer
lengths available with larger diameter wire (consult factory). In most cases the proper cable is
specified at time of order and will be supplied by the factory.
1. Connect the SHIELD wire of the twisted pair shielded cable to TB1, terminal 1 of the main
electronics. This wire should terminate at terminal block TB1, terminal 1 inside the Remote Oxygen
Sensor. Make sure all wires are put through the strain relief first.
2. Connect the POSITIVE wire of the twisted pair shielded cable to TB1, terminal 2 of the main
electronics. This wire should terminate at terminal block TB1, terminal 2 inside the Remote Oxygen
Sensor.
3. Connect the RETURN wire of the twisted pair shielded cable to TB1, terminal 3 of the main
electronics. This wire should terminate at terminal block TB1, terminal 3 inside the Remote Oxygen
Sensor.
Series 1300 Oxygen Deficiency MonitorPAGE 15
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
Repeat the above procedure to connect the second Remote Oxygen Sensor if applicable but
substitute the terminals as follows:
1. Connect the SHIELD (TB1, terminal 4) of the main electronics to TB1, terminal 1 inside the 2
Remote Oxygen Sensor.
2. Connect the POSITIVE (TB1, terminal 5) of the main electronics to TB1, terminal 2 inside the 2
Remote Oxygen Sensor
3. Connect THE RETURN (TB1, terminal 6) of the main electronics to TB1, terminal 3 inside the 2
Remote Oxygen Sensor
nd
nd
nd
After connection to the Remote Oxygen Sensor(s) has been completed, power the instrument and
wait for it to initialize. The display should now show each sensor being initialized. Please reference
section 6 “Sensor Setup” on page 47 for more information on setting up the sensor(s).
NOTE: Remote Oxygen Sensors do not need to be connected to the unit sequentially. They will be
detected based on the terminals they are connected to.
3.4 Horn And Strobe
The interface to the Horn and Strobes is located at terminal block TB1 (Refer to Figure 4) inside the
main enclosure. The Series 1300 Oxygen Deficiency Monitor is capable of supporting up to 8 horn and
strobes. The horn and strobes use a 3 wire connection3 consisting of power, ground return, and data.
The connections within the main electronics enclosure are located on a terminal block located in the
bottom of the enclosure labeled “TB1”. See Table 3 below for reference.
Main EnclosureHorn and Strobe PCB
Terminal
Block
TB 1SignalName on PCBDetail
7Shield
8Shield“SH”Shield
9Positive“V+”Positive Power (typically 12V)
10Data“DAT”Data communications line
11Return“GND”Power Return
Table 3 - Horn And Strobe Connections (Terminal Block 1)
Follow the procedure below for connecting a horn and strobe that will be powered from the Series
1300 Oxygen Deficiency Monitor.
1. Connect the 1st wire of the shielded cable to TB1 terminal 9 (POSITIVE), inside the main enclosure.
Terminate this wire to one of the terminals on the horn and strobe PCB labeled “V+”.
2. Likewise connect the 2nd wire of the shielded cable to TB1 terminal 10 (DATA). Terminate this wire
to one of the terminals on the horn and strobe PCB labeled “DAT”.
3. Connect the 3rd wire of the shielded cable to TB1 terminal 11 (RETURN). Terminate this wire to one
of the terminals on the horn and strobe PCB labeled “GND“.
4. Connect the shield of the shielded cable to TB1 terminal 8 (SHIELD). Terminate this shield to one of
the terminals on the horn and strobe PCB labeled “SH”.
Note: it is only necessary to connect the shield at the horn and strobe PCB when daisy-chaining4. In a
star-point configuration, simply leave it disconnected.
3
Horn and Strobes can be wired directly to the relay contacts using only two (2) wires. The Series 1300 Oxygen Deficiency
Monitor will not be able to warn the user of any disconnect or wiring problem without the use of the “DAT” line.
PAGE 16 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
Follow the same procedure to connect multiple horn and strobe units to the Series 1300. The main
enclosure and the horn and strobe unit both have multiple terminals that can accept more than a
single wire twisted together if necessary. If daisy-chaining, then simply connect to the second set of
terminals supplied on the horn and strobe unit.
When powered from 90-264 VAC, 47-63 Hz., the instruments default DC output power is capable of
driving up to 4 horn and strobes simultaneously. If more than 4 are required, the additional horn and
strobes need to be powered from either an independent DC power source or an upgraded power
supply that can handle the extra horn and strobes. The maximum length of cable for a single horn
and strobe wired with the recommended #18 AWG wire can be up to 1,000 feet away from the main
electronics. However, each additional horn and strobe wired in SERIES (daisy-chained) on the SAME
wire will reduce the maximum length by approximately half. The maximum voltage drop in any single
chain of wire must be less than or equal to 3V. There must be a minimum of 9V at the end of the wire.
CAUTION
THE MAX CABLE CAPACITANCE OF 0.15UF MUST BE ADHERED TO OR THE
HORN AND STROBES MAY NOT WORK CORRECTLY AND/OR CAUSE
UNWANTED BEHAVIOR. PLEASE CONTACT THE FACTORY FOR ANY
APPLICATIONS REQUIRING LONGER LENGTHS.
Figure 12 below shows a horn and strobe and the associated printed circuit board (PCB).
Figure 12 - Horn And Strobe With PCB
3.4.1 Horn And Strobe PCB
The printed circuit board has 8 terminals located on the front edge as shown in Figure 13 - Horn and
Strobe PCB on page 17. The signals are labeled on the board as “DAT” for data, “V+” for positive
power, “GND” for power return, and “SH” for shield. Each signal has two connections to allow daisychaining. If daisy-chaining multiple horn and strobes, the connectors are sized to allow for twisting
two wires together before inserting into the screw terminal.
Figure 13 - Horn And Strobe PCB
4
Daisy-chaining is when another Horn and Strobe is connected to the first in series. As opposed to being connected to the
main enclosure in parallel (or star-point).
Series 1300 Oxygen Deficiency MonitorPAGE 17
USER MANUALALPHA OMEGA INSTRUMENTS CORP.
Series 1300
Optionally wire
up to (4) more.
This requires
external Power
or a higher
rated internal
power supply.
CAUTION
J1 TO J4 SUPPORT A MAXIMUM OF FOUR (4) HORN AND STROBES UNLESS
USING A SPECIAL HIGH POWERED MAIN POWER SUPPLY. NEVER CONNECT
AN EXTERNAL POWER SUPPLY TO THE “V+” TERMINALS.
Note, that if powering the instrument from an external power supply, it is imperative that the power
supplied be sufficient for the number of horn and strobes connected at any given time. (5) or more
horn and strobes will require remote power or a higher wattage upgrade to the internal power supply.
The additional external power needed to handle up to four (4) standard horn and strobes is
approximately 20W. Therefore, using a 45W, 12V power supply would be able to power the Series
1300 Oxygen Deficiency Monitor as well as all eight (8) horn and strobes.
When externally powered, connect only the “DAT” and “GND” terminals back to the main electronics
enclosure. The “V+” terminal can be omitted in this case as it is not powered from the main
electronics. The external power supply connects to “EXT PWR” terminal located on the right hand
side labeled “J5”.
CAUTION
REMOVE POWER BEFORE PERFORMING ANY WIRING ON THE INSTRUMENT
OR HORN AND STROBE(S). IF EQUIPPED WITH BATTERY BACKUP, INSURE THE
BATTERY TERMINAL INSIDE THE MAIN ENCLOSURE IS DISCONNECTED.
3.4.2 Horn And Strobe Daisy Chain Configuration
This configuration shows how the wiring can be configured in a serial fashion or “daisy chain”. The
source wiring comes in on one set of connectors and the next horn and strobe is wired to the other
set of connectors as shown below.
Figure 14 - Wiring Horn And Strobes In Series (Daisy Chained)
As shown above, using four (4) horn and strobes and #18 AWG cable wire, the maximum distance to the last
horn and strobe in the chain is 250 feet.
PAGE 18 Series 1300 Oxygen Deficiency Monitor
ALPHA OMEGA INSTRUMENTS CORP.USER MANUAL
Series 1300
Series 1300
12V
3.4.3 Horn And Strobe Star Configuration
Shown below is a Series 1300 Oxygen Deficiency Monitor wired to four (4) horn and strobes.
Figure 15 - Wiring Horn And Strobes Individually
As shown above, using four (4) horn and strobes and #18 AWG cable wire, the maximum distance fore each
single horn and strobe is 1,000 feet.
3.4.4 Horn And Strobe External Power
Adding more than four (4) horn and strobes with the standard power supply requires external power.
Shown below is the wiring for externally powered horn and strobes. Note the added 12V power
supply that is connected to a separate connector.
CAUTION
DO NOT CONNECT EXTERNAL POWER TO THE V+ TERMINAL. EXTERNAL
POWER SHOULD ONLY BE APPLIED TO THE “EXT PWR” CONNECTOR
LABELED “J5”
Figure 16 - Externally Powered Horn And Strobes
Wiring of the V+ is optional and shown as a dashed line in the picture.
Series 1300 Oxygen Deficiency MonitorPAGE 19
Loading...
+ 60 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.