Handheld Laser Particle Counter
Model: P311
Operation Manual
Ver: 1.50
AIRY TECHNOLOGY INC

Warranty
AIRY TECHNOLOGY INC warrants to the original user that this instrument shall be free from
defects in material and workm anship for one year from the dat e of shipment.
Airy's obligations under thi s warranty , and the sole reme dy for its breach, are limited t o repair or, in
Airy's sole discretion, replacement of the instrument or any of its parts. Should it become
necessary to return the instrument for repair during or beyond the warranty period, user shall
contact Airy Technology, Inc. (USA). E-mail: info@airytechnology.com
. User is responsible for
shipping charges, freight, insurance and proper pac kaging to prevent damage in tr ansit.
This warranty shall be void in the event of user actions including misuse, improper wiring,
operation outside of specification, improper maintenance or repair, unauthorized modification, or
any other defect caused b y the user' neglect or accident.
This warranty is the sole and exclusive warranty for this instrument, and no other warranty,
whether written or oral, is expressed or implied. Airy specifically disclaims any implied warranties
of merchantability or fitness for a specific purpose and will not be liable for any direct, indirect,
incidental, consequential, or punitive damages. Airy’s total liability is limited to repair or
replacement of the product.

Safety information
This section gives instruct i ons for promoting safe and proper handling of the P ar t icl e Counter .
Laser Safety
The Handheld Laser Particle Cou nt er is a Class I laser- based instrument.
During normal operation, you will not be exposed to laser radiation.
Precaution should be t aken to avoid exposure to haz ar dous radiation in the form of intens e,
focused, invisible light.
Exposure to this light may cause blindness.
Take these pr ecautions:
DO NOT remove any parts fro m the p article co unter unle ss you are specifically told to do so in
this manual.
DO NOT remove the hous ing or cover s. There are no user serviceable components insid e the
housing.
DANGER
The use of control s , a djustments, or procedures other than those
specified in this manual m ay result in exposure to hazardous optic al
radiation.
Precautions for power use
AC Adapt or
The AC adapter accommodates voltage of AC 100~240V and frequency of 50/60Hz.
Batteries
Use four AA B atteries.
DANGER
Driving voltage should be kept within specified range.
Failure to follow this instruction may cause electric
shock and instrument damage.
CAUTION
Do not start sampling when the instrument is connected to a personal
computer.
Failure to follow thi s i nstruction may cause
the instrument to work abnormally.
PROHIBITION
WARNING
WARNING

Product Overview
Carefully unpack the Handheld Laser Particle Counter from the shi pping container and v er ify that
all the items shown in the photos below and listed in the following tables are present .
Contact us immediately if any items are missin g or broken.
Handheld Laser Particle Counter parts l i st :
Qty. Item Description Reference Picture
1 Handheld Laser Particle C ounter
1 Isokinetic inlet
1 Probe for tubing
1 Cap
1 AC Power adapter
1 Zero filter
or
1 USB cable USB type A to mini USB-B
1 Battery Charger
1 Application CD
1 Quick Start Guide
1 Calibration Report
1 Carry ing Case

Getting Started
The Handheld Laser Particle Counter (particle counter) is a light w eight, handheld part icle counter
with a TFT LCD display. It operates on bat tery or AC power. This model has a 0.1 CFM (2.83 L/ min)
flow rate and counts in user-adjustable bin siz es o f 0.3/0.5, 0.5/1.0/2.0/2.5 and 5 microns (channel
1 and 2 are selectable). Up to 8000 data s ets can be stored and dow nloa ded for analy sis and
reporting using the utility inclu ded with the device.
Inlet nozzle:
User can replace the Inlet nozzle between the isokinetic inlet and the probe for tubing. The
isokinetic inlet is used for ambient air sampling. To use the isokinetic inlet, detach the red
cap from the inlet nozz le and att ach t he is o kinetic pr obe. After all the samples are completed,
please detach the isokinetic inlet and put the red cap back before placing the unit into the
carrying case. If you are planning to use a tube for samplin g, plea se cont a ct Airy Technology
or your local distributor.
Zero filter
The zero filter cleans the sensor after your sampling at contaminated places. The zero filter
also checks whether the part icl e counter is counting electrica l noises. To use the zero filter:
1. Detach the is okinetic inlet from the main unit
2. Connect the zero filter to the main unit using the tube (There is a tube in the plastic bag
containing the zero filter)
Key Pad
Inlet nozzle
LCD display
USB Port

3. Start sampling
4. Wait until the counter doe s not det ect any particles
5. Stop sampling and detach the zero filter
If the counter keeps detecting particles after 1 minute of sampling, please contact Airy
Technology or your local distributor.
AC power, USB port, and USB cable
1. AC Power
When using AC power, must use the affiliated AC adapter, shown below.
Connect Mini USB-B plug to the instrument.
2. Data Communicat ion
When using the USB cable to transfer data records to a PC, set it up as follows:
Connect Mini USB-B plug to the instrument.
USB type A plu g to the type A receptacle of PC.
Battery
Loading the batteries:
1. The batteries have to be loaded before use.
2. Remove the battery cover on the back of the shell.
Mini USB-B
USB type A

3. Insert four AA of Ni-MH type or alkaline batteries.
4. When using Ni-MH batteries, charge the battery fully before use.
Q u i c k S t a r t
The best way to start quickly is to refer to the printed “Quick Start Guide” that comes with the
instrument. It will help you to quickly set up the instrument and start sampling. Refer to the
sections below for more detailed information on configuring and running the instrument.
Operation
Key Pad
The instrument is controlled by key pad and its functions are shown as follows:
KEY Function
ENTER Enter a menu or execute
START/STOP Start or stop sampling
PREV Return to previous screen
Or Start or stop sampling
POWER
Power On/Off
Regulate backlight
UP/DOWN/LEFT/RIGHT Move the cursor or change the values
OK Execute

Use the up and down keys to highlight a menu or a menu option. Use the left and right keys to
enter the sub item or leave the s ub item.
Use the up and down keys to perform o per at i ons such as increasing a value. Use the right key
and left keys to move right and left.
The PREV key always brings you back to the pr evious screen.
Power On/ Off
Press the POWER key to turn on the instrument.
Press the POWER key to regulate backlight of LCD.
Press the POWER key for more than one second, and the message “Power off…” will appear on
the bottom of current scre en.
Press the POWER key for more than two seconds, power will be off.
Splash Screen
To turn the instrument on, press the Power key. A splash screen will appear f or t hr ee seconds,
displaying the company logo, model number , ser ial number , and firmware version nu mb er ( see
below).
Splash Screen
The default screen will appear. The instrument is ready for operation.
Default Screen
LOC. 001 CYCLES 00/01
STIME 01:00 INT 01:00
0.3μm
1.0μm
Handheld Laser Particle Counter
Model : P311
Version : 1.00(00.00.00.00)
Serial Number : 000000

The parts of the de fau lt s creen are explained as follows:
【AUTO 】 Measure mode ( AUTO , MANUAL, ISO)
【0001/8000】 Current number of data records indicated (Max 8000)
【10:05 】 Current time indicated (hour& minute)
【▲▼ CNT 】 Change the measurem ent unit by the pressing the UP/DOWN key (CNT, m3, cf)
【 ∑ 】 Cumulative count - the number of a ll the particles equa l or larger than selected
particle size
【 △ 】 Differenti al count - the number of p arti cles betwee n the sel ected s iz e and th e next
selected size
【0.3um 】 Particle size (Channel 1 0.3μm, 0.5μm)
【0.5um 】 Particle size (Channel 2 0.5μm, 1.0μm, 2.0μm, 2.5μm)
【5.0um 】 Particle size (Channel 3 5.0μm)
【LOC. 】 Location/site number
【STIME】 Sampling time (Setting range is 1sec~99min 59sec)
【CYCLES】 Cycle co unt (1~2000)
【INT 】 Interval time (Setting range is 1sec~99min 59sec)
【Stopped】 Operating status (Stopped, Sampling, Waiting, Holding)
【START】 Sampling start or stop by using ENTER or START key
【MENU 】 Pressing PREV key to go to Main Menu
Main Menu
Main Menu Screen
1. Use the UP/DOWN key to select an item, the ENTER/OK key to select a sub item and the PREV
key to return to the default screen.
From the Main Menu you can se lect ot her me nus:
Menu Description
Measurement Setting Set measure mode, sample time, interval time, cycle count,
particle units, and channel size.
Data Process Display data records, clear data records and transmit the data.
System Setting Set time and date, screen settings and location number.
Utility Setting Set delay time, key sound and pump speed
Each of these menus is described i n the remainder of this chapter.

Measurement Setting
Measurement menu
1. UP/DOWN key t o select an item, and press the ENTER key to select a sub item ;
2. UP/DOWN/LEFT/RIGHT key to set and press OK to execute. PREV key to go back to Main
Menu screen.
The following table describes this menu's options and the available parameters.
Option Description
Measure Mode Auto, Manual, ISO
Sample Time Setting range from 1sec to 99 min 59 sec.
Interval T ime Setting range from 1 sec to 99 min 59 sec
Cycle Count 1~2000 times
Particle Unit CNT, /m3, /cf,
Channel 1 Size 0.3μm, 0.5μm
Channel 2 Size 0.5μm, 1.0μm, 2.0μm, 2.5μm
Channel 3 Size 5.0μm
【Data Process 】
At the data process screen you can view the number of recor ds stored in the instrument, display
data records, and clear t he data. You can also transmit the records to co m puter with a USB cable.
Data Process Menu
Data Process Menu
Sample Time 01:00
Interval Time 01:00
Cycle Count 0001
Particle Unit < CNT >
Channel 1 Size < 0.3um >
Channel 2 Size < 0.5um >
Channel 3 Size < 5.0um >

【Show Data Record】
Press the Enter key to go to secondar y screen where you can select the record (by sampling
number) to review.
When “Show Data Rec or d” is selected and the data mo de is ISO, the sampling result will be
displayed.
For example:
1. When data # 0001 is selected, the display will show as follows (In this example the total
number of samples is 4):
2. Use the UP/LEFT and DOWN/RIGHT key s t o scr oll through the records, PREV to go bac k to
the select record screen. In this example, data # 0004 is the last sample of consecutive
samplings for ISO mode.
At the bottom right of the screen, “C AL” appears. "CAL" appears only in the last sampling
result of consecutive ISO samplings.
Location 001
Count 0002/0002
Start 2009-09-15 10:26:00
Sample 00:30
Interval 00:30
0.3um 6064951 5985140
0.5um 79811 38140
5.0um 41671 41671
PREV CAL
Location 001
Count 0001/0002
Start 2009-09-15 10:25:30
Sample 00:30
Interval 00:30
0.3um 5694853 5633405
0.5um 61448 24721
5.0um 36727 36727
PREV
Mode: ISO
Start: 2009-09-15
10:25:30

3. Select “CAL” to go to the ISO calculation result sc reen.
This screen shows the average, standard deviation, and UCL. At this calculation results
display, Press “UP” or “DOWN” to select different channel size calculation results. Press
“PREV” to return to the “Show Data Record” screen.
N o t e
If the buffer is filled with more than 7900 data record sets, the instrument still continues to count and
save data but the number of data records shown on the default screen will be red and the buzzer will
beep as an alarm. When the buffer is filled with the maximum capacity of 8000 records, the instrument
will continue to count but the data will not be saved. The user must write down the data and clear the
buffer (if necessary, please download the data to PC to save before deleting data from the instrument).
【Clear Data Record】
Press the Enter key to clear all sa mpled d ata in the buffer. The screen will ask you to confirm your
request.
Measured Point s:2
AVG 100470
0.5um SD 38206
UCL 270669
PREV CAL
ISO Unit: /m3
Measured Point s:2
AVG 8985479
0.3um SD 4345502
UCL 28343704
PREV CAL
Do you want to del ete
all data records?
[ENTER] to confirm
[PREV] to cancel

【Transmit Data Recor d】
Downloading the data to PC can be performed in the screen below. Please plug the USB cable to
the instrument and your PC.
【System Setting】
System Setting Screen
1. UP/DOWN key t o select item and press the ENTER key to go to sub item ;
Waiting for the computer
to connect now!
[PREV] to cancel and return
PREV
Are you sure you want to delete
all data now?
[OK] to confirm
[PREV] to cancel
PREV
Set Time 11:28:12
Backlight < brighter >
Location 001
PREV ENTER

2. UP/DOWN/LEFT/RIGHT key to set and press OK key to execute; Press the PREV key to
go back to Main Menu.
Option Description
Set Date System Date( year, month , day )
Set Time System Time (24 hours, hour, minute, second)
Backlight Brighter, Normal, Darker
Location 1~199
【Utility Setting】
You can set delay time, key sound, pump speed and battery mode in this screen.
Utility Menu
Option Description
Delay Time Period from starting pump to beginning sample ( 5~99s).
Key Sound Sound of key. (On/Off)
Pump Control
Press the Enter key to increase or decrease the pump speed. (The
pump may slow down with age, or it may be necessary to increase
speed if there are flow restrictions – such as long tubing). Use a flow
meter to check the flow.
When taking critical measurements, adjust the pump speed as
necessary. (10~83)
Battery Mode Select the battery mode. (Ni-MH or Alkaline)
Sample Procedure
Note: when sampling
* press UP/DOWN key to convert the particle unit.
Utility Menu
Pump Control < ** >
Battery Mode < Ni-MH >

Manual mode
Press START/ENTER to start sampling.
STOP Waiting MENU
LOC. 001
DELAY 00:05
0
0
0
CNT ∑ △
0.3μm
1.0μm
5.0μm
0
0
0
MANUAL 0004/8000 09:57
STOP Sampling MENU
LOC. 001
0.3μm
1.0μm
5.0μm
1640
234
6
1400
228
6
CNT ∑ △
START Stopped MENU
MANUAL
0005/8000 10:01
CNT ∑ △
0.3μm
1.0μm
5.0μm
23452
1058
24
22394
1034
24
Stop and data are saved automatically)

Auto mode
AUTO 0005/8000 10:05
CNT ∑ △
0.3μm
1.0μm
5.0μm
0
0
0
0
0
0
LOC. 001 CYCLES 00/01
STIME 10:00 INT 15:00
START Stopped MENU
STOP Waiting MENU
LOC. 001 CYCLES 00/02
DELAY 00:05
0.3μm
1.0μm
5.0μm
0
0
0
0
0
0
AUTO 0005/8000 10:05
CNT ∑ △
STOP Sampling MENU
LOC. 001 CYCLES 00/02
STIME 10:00
0.3μm
1.0μm
5.0μm
1427
53
3
1374
50
3
CNT ∑ △
AUTO 0005/8000 10:05
START
Caution: When the
difference between i nt er val
time and sampling time is
longer than 15 sec, the
pump will stop after each
sampling and restart before
the next sampling.

ISO mode
Caution: There are on ly m3/cf units in this mode.
STOP Holding MENU
LOC. 001 CYCLES 01/02
INT 15:00
0.3μm
1.0μm
5.0μm
27543
1436
67
CNT ∑ △
AUTO 0006/8000 10:15
START Stopped MENU
LOC. 001 CYCLES 00/02
STIME 10:00 INT 15:00
26107
1369
67
0.3μm
1.0μm
5.0μm
27543
1436
67
26107
1369
67
CNT ∑ △
AUTO 0008/8000 10:25
START Stopped MENU
LOC. 001 CYCLES 00/01
STIME 10:00 INT 15:00
ISO 001 0008/8000 10:25
0.3μm
1.0μm
5.0μm
/ m3 ∑ △
0
0
0
0
0
0
The data is saved
automatically after ev er y
sampling. Pressing
“STOP” will finish sampling
at anytime except during
“Waiting” period.
The sampling stops
automatically when
all cycles have
finished.

NEXT Stopped FINISH
LOC. 001 CYCLES 00/01
STIME 10:00 INT 15:00
0.3μm
1.0μm
5.0μm
234645
1453
43
220144
1411
43
/ m3 ∑ △
ISO 002 0012/8000 10:25
ISO Unit:/m3
Measured Points: 2
Sample Time: 10:00
AVG
234645
0.3μm SD 12345
UCL 2134
After the current
sampling is
finished, press
“ENTER” to start
the next sampling.
Press the “PREV”
key to return to
default screen.
Press UP/DOWN to
select the channel
size shown.
FINISH
The measurement process of ISO
executes as AUTO mode. Please
refer to the above instruction.
Press the “PREV”
key to finish this
measurement. It
will calculate and
run into the ISO
result interface.
ISO 001 0008/8000 10:25
0
0

Data Handling
USB Computer Communication
Mini USB Port
The Model P311 is equipped with a USB-compatible cable for uploading and downloading information
to a PC. Plug the cable into the right side of the instrument as shown below.
Software Installation
The Airy P311 Software (Data Transfer Utility) comes as a CD including software and USB driv ers
for the particle counter.
Note: this version of the softw ar e only guarantee use in the Windows XP (SP2) or Windows Vista
or Windows7 (32bits) OS.
Installation consists of two parts:
Installation of Airy P311 Software.
Installation of Custom USB Device.
Downloading Data
Connect a USB cable to the instrument and PC. Go to [Data Process] and then to [Transmit Data
Record]
. Double click [Airy P311 Softw ar e] icon, and the main appl ication screen will show up.
Waiting for the computer
to connect now!
[PREV] to cancel and return

【Read Data】
【Read records】
Click Read Records. It takes a few moments to dow nload all the dat a dep endin g on the number o f
samples stored in instrument. W hen it is finished, the main application screen will display as
follows:
【Save Data】
【Save all records】
After downloading data to PC, the other functions will be enabled. If you want to save all records,
click Save All. You can select the file location and decide the file name.

To save the file, click Save, and then the data will be saved in the selected lo cat ion.
To cancel the transfer, select Cancel.
Data is stored in a .CSV file format that can be opened by most spreadsheet programs
such as Microsoft® Excel®.
【Save selected records 】
When you need to save part of the record s, you can select the data to save.
Click Save Selected
You can specify the record number by t yping in the window.
Note: the start number cannot exceed the finish number.
Click OK to continue and the spe ci fi ed data range will be sav ed.
【Delete Data】
【Delete Records】
This operation will delete all the records saved in device. Please double-check before
you delete data. When you click Delete Records, the below dialog show s up.

Click OK to continue, Click Cancel to stop.
Click OK to continue, Click Cancel to stop.
Click Cancel to stop. Click OK to delete data.
WARNING!
Deleting data is an irreversible operation. Download and save data before deleting in
order to have a copy for future use.

Calibration
Please return the Airy Technology P311 to the manufacturer or the authorized service center for
annual calibration.

A P P E N D I X A
Specifications
Specifications
Size Range 0.3 -5.0 μm
Channel Sizes
channel 1: 0.3/0.5μm
channel 2:
0.5/1.0/2.0/2.5μm
channel 3: 5.0
μm
Counting Efficiency 50% at 0.3 μm; 100% for particles > 0.45 μm (per JIS)
Concentration Limits 4,000,000 particles / ft3 at 10% coincidence loss
Light Source Laser diode
Zero Count Level <1 count / 5 minutes Meets JIS B9921
Flow Rate 0.1 CFM (2.83 LPM)
Calibration NIST traceable
Sample Probe/Tubing Isokinetic sampling probe, probe for tubing
Sampling Modes Manual, Automatic and ISO
ISO Certification
ISO 5-9
@ 0.3-5.0μm*
Sampling Time 1 second to 99 minutes 59 seconds(Configurable)
Sampling Frequency 1 to 2000 cycles or continuous(Configurable)
Sample Output Internal HEPA filter
Vacuum Source Internal pump
Communication Mode USB
Data Storage 8000 sample records
Status Indicators battery used, over range alarm
Display 3.5-inch 320 x 240 Color LCD
Power DC 5V 0.7A (Mini USB TYPE-B)
Battery 4 x AA
Battery Life
Up to 4.5 hours of continuous use (LCD Backlight low, Included Ni-
Battery)
Dimensions (L x W x H) 178x90x47mm (without isokinetic inlet)
Weight 480g (without battery)
Standards CE, JISB9921, ISO 21501-4
Warranty 1 year limited warranty
Operating Conditions 5° to 35°C 20% to 95%RH non-condensing
Storage Conditions -20° to 50°C Up to 98%RH non-condensing
Included Accessories
AC adapter, Isokinetic inlet, USB cable, Zero filter, Software,
4 x AA batteries with charger, Carrying case
* Exclude ISO 5 @ 5.0 μm