Airsep NewLife Intensity 10 Service manual

NewLife® Intensity 10 Oxygen Concentrator Service Manual
ii
Expedited Product Warranty Check service is available at 800-482-2473
MN134-1 REV D
iii
Table of Contents
Section 1.0
Introduction
1.1 Equipment Provider Responsibility 1
1.2 Important Notice and Symbol Explanations 2
1.3 Functional Specifications 4
Section 2.0
Operation Check and Oxygen Concentration Test
2.1 Description of Operation 5
2.2 Operation Check 5
2.3 Alarm System 5
2.3.1 Start Up/Battery Test 6
2.3.2 Power Failure Alarm Test 6
Section 3.0
Patient Instructions
Section 4.0
Provider Instructions
2.4 Oxygen Concentration Test and Specification 6
3.1 Instructions 7
3.2 Routine Maintenance by the Patient 7
3.2.1 Cleaning the Air Intake Gross Particle Filter 7
3.2.2 Checking the Alarm System Battery 7
4.1 Instructions 8
4.1.1 Air Intake Gross Particle Filter/GPF 8
4.1.2 Felt Filter Replacement 8
4.1.3 Product Filter 8
4.1.4 Battery Replacement 8
4.1.5 Recording Maintenance 9
4.2 Preparing for New Patient Use/Method of 9 Cleaning and Infection Control
MN134-1 REV D
iv
Section 5.0
Main Components
5.1 Components 10
5.2 Cabinet Removal 10
5.2.1 Removing Side Panel(s) 10
5.2.2 Removing Back Panel 10
5.2.3 Removing Lower Front Panel 10
5.2.4 Removing Control Panel 10
5.2.5 Superstructure 10
5.2.6 Caster Replacement 10
5.3 Compressor 11
5.3.1 Compressor Replacement 11
5.3.2 Capacitor Replacement 12
5.4 Solenoid Valves 12
5.4.1 Feed/Waste Valve Replacement 12
5.4.2 Solenoid Valve Coil Replacement 12
5.5 Sieve Beds 13
5.5.1 Sieve Bed Replacement 13
5.6 Cabinet Fan 14
5.6.1 Cabinet Fan Replacement 14
5.7 Circuit Board 14
5.7.1 Circuit Board Replacement 15
5.8 Product Tank Replacement 15
5.9 Product Regulator Check and Setting 16
5.9.1 Setting Product Regulator for Normal Operation 16
5.9.2 Product Regulator Cleaning or 17
Rebuilding
5.10 Circuit Breaker Replacement 17
5.11 I/0 Power Switch Replacement 17
5.12 Buzzer Replacement 18
5.13 Hour Meter Replacement 18
5.14 Flowmeter Replacement 18
5.15 Power Cord Replacement 18
5.16 Oxygen Monitor Circuit Board Replacement 19
MN134-1 REV D
v
Section 6.0
Troubleshooting
6.1 Operating Pressure Test 20
6.1.1 High Operating Pressure 20
6.1.2 Low Operating Pressure 20
6.2 General Troubleshooting 21
6.3 Troubleshooting Chart 22
6.4 Tool Kit and Pressure Test Gauge 25
MN134-1 REV D
1
1.0 Introduction
This unit is not a life-support device. Geriatric, pediatric, or any other patient unable to communicate discomfort while using this oxygen concentrator may require additional monitoring. Patients with hearing and/or sight impairments may need assistance with monitoring the alarms.
This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in severe fire, property damage, and/or cause physical injury or death.
Use no oil, grease, or petroleum-based or other flammable products on or near nasal end of cannula or on the Intensity 10 unit. Oxygen accelerates the combustion of flammable substances.
1.1 Equipment Provider Responsibility
All Equipment Providers of the NewLife® Intensity 10 Oxygen Concentrator must assume responsibilities for handling, operational check-out, patient instruction, and oxygen concentration checks. These responsibilities are outlined below and throughout this manual.
As an Equipment Provider, you must do all of the following:
Inspect the condition of each NewLife Intensity 10 unit immediately upon delivery
to your business location. Note any sign of damage on the delivery receipt, and report it directly to both the freight company and AirSep Corporation immediately.
Check the operation of each NewLife Intensity 10 before delivery to a patient.
Confirm the oxygen concentration level is within specifications, as referred to in Section 2.4. (Test the battery and power disconnect alarm, as described in Section
2.3 of this manual.)
Deliver NewLife Intensity 10 units only to patients authorized by a licensed health
care provider or physician’s prescription. The NewLife Intensity 10 must not be used as a life-supporting device. A backup supply of oxygen must be available.
Instruct patients how to use the NewLife Intensity 10 in conjunction with the Patient
Manual.
Instruct patients to notify their licensed health care provider/physician if they
experience any signs of discomfort.
Instruct each patient how to perform routine maintenance of the air intake gross
particle filter and how to check the alarm system battery. (Refer to Section 3.2.)
Be available to provide service to each patient at any time.Maintain the NewLife Intensity 10 in accordance with Section 4.0.Establish and implement a protocol to check oxygen concentration.Repair components and replace parts only as outlined in this manual. Use only
AirSep parts for replacement in NewLife Intensity 10 Oxygen Concentrators.
Refer to the NewLife Intensity 10 Product Warranty if parts replacement is required
within the warranty period.
1/
MN134-1 REV D
2
1.2 Important Notice and Symbol Explanations
Describes a hazard or unsafe practice that can result in severe bodily injury or death.
Describes a hazard or unsafe practice that can result in minor bodily injury or property damage.
Provides information important enough to emphasize or repeat.
Symbol
Description
Symbol
Description
On (power switched on)
Do not disassemble
Off (power switched off)
Consult instructions for use
Type BF device
Keep in vertical position Class II device
Fragile – handle with care
Do not smoke
Oxygen concentration warning LED
Do not expose to open flames
Caution: US federal law restricts this device for sale or rental by or on order of a licensed healthcare provider
Do not grease
Gas outlet, connection to the circuit
Consult the accompanying documents
WEEE Directive
Keep unit and accessories dry
As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important information as follows:
Symbols are frequently used on equipment in preference to words with the intention of lessening any possibility of misunderstanding caused by language differences. Symbols can also permit easier comprehension of a concept within a restricted space.
The following table is a list of symbols and definitions that may be used with the NewLife Intensity 10 Oxygen Concentrator. These symbols are referenced from the appropriate International Electrotechnical Commission (IEC) standards:
MN134-1 REV D
3
Method for disposing of waste: All waste from the NewLife Intensity 10 (patient circuit, etc.) must be disposed of using appropriate methods.
Method for disposing of the device: In order to preserve the environment, the concentrator must only be disposed of using the appropriate methods.
Conformity with EN 60-601 (§ 6.8.2 b):
“The manufacturer, assembler, installer or importer are not considered to be responsible
themselves for the consequences on the safety, reliability and characteristics of a device unless:
- The assembly, fitting, extensions, adjustments, modifications, or repairs have been performed by persons authorized by the party in question.
- The electrical installation of the corresponding premises complies with IEC regulations.
- The device is used in accordance with the instructions for use.” If the replacement parts used for the periodic servicing by an approved technician do not comply with the manufacturer’s specifications, the latter is absolved from all responsibility in the event of an accident. Do not open the device while in operation: risk of electrical shock. This device complies with the requirements of the 93/42/EEC European directive but its operation may be affected by other devices being used close by, such as diathermy and high frequency electro-surgical equipment, defibrillators, short wave therapy equipment, mobile telephones, CB and other portable devices, microwave ovens, induction plates or even remote control toys, and more generally electromagnetic interferences, which exceed the levels specified by the EN 60601-1-2 standard.
Classification
Type of protection against electric shock: Class II Protection from electric shock is achieved by DOUBLE INSULATION.
Protective earthing or reliance upon installation conditions are not required.
Degree of protection against electric shock:
Type BF Equipment providing a particular degree of protection against electric
shock, particularly regarding:
1) allowable leakage current;
2) reliability of protective earth connection (if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of water:
Drip-proof equipment – IPX1. Equipment provided with an enclosure preventing entry of such an amount of falling liquid as might interfere with the satisfactory and safe operation of the equipment.
Method of cleaning and infection control allowed:
Please refer to the Maintenance section in the NewLife Intensity 10 Service Manual.
Degree of safety of application in the presence of flammable anesthetic gases:
Equipment not suitable for such application.
Mode of operation: Continuous duty.
MN134-1 REV D
4
1.3 Functional Specifications
Oxygen Concentration: 2-9 LPM: 92% +3.5/-3%
10 LPM: 90% +5.5/-3% (Based on 70°F [21°C] at sea level)
Accuracy: Flowmeter ± 10% or ± 200ml of indicated flow, whichever is
greater.
Response Time: Allow 5 minutes to attain maximum oxygen concentration. Positioning: Operate the unit in an upright position, maintaining at least
12 inches (30.5cm) of open space on all sides for ventilation.
Dimensions: 27.5 in. high x 16.5 in. wide x 14.5 in. deep
(69.9 cm high x 41.9 cm wide x 36.8 cm deep)
Weight: 58 lb; shipping weight 65 lb
(26.3 kg; shipping weight – 29.5 kg)
Electrical: 120 VAC, 60 Hz, 6.0 amps, 590 watts
220-240 VAC, 50 Hz, 3.0 amps, 590 watts
Two-prong polarized plug Double-insulated cabinet
Alarms: Power Failure
Low Oxygen Concentration (with optional Oxygen Monitor) Low Pressure High Pressure High Temperature Battery Test
Operating Temperature 41°F to 104°F (5°C to 40°C) for 120 VAC 60 Hz models Range: 41°F to 95°F (5°C to 35°C) for 220-240 VAC 50 Hz models
Storage Temperature -4°F to 140°F (-20°C to 60°C) Range:
Humidity: Up to 95% (non-condensing)
MN134-1 REV D
5
2.0 Operational Check and Oxygen Concentration Test
Do not use extension cords with this unit or connect too many plugs into the same electrical outlet. The use of extension cords could adversely affect the performance of the device. Too many plugs into one outlet can result in an overload to the electrical panel, causing the breaker/fuse to activate or fire if the breaker or fuse fails to operate.
2.1 Description of Operation
Air is drawn into the NewLife Intensity 10 Oxygen Concentrator through an external air intake gross particulate filter. Before this air enters the compressor, it passes through the unit’s suction resonator which, quiets the compressor’s suction sound. Pressurized air then exits the compressor and passes through a heat exchanger. The heat exchanger reduces the temperature of the compressed air. Next, a two-way solenoid feed valve directs the air into one of two sieve beds that contain molecular sieve. The unique property of molecular sieve enables it to physically attract (adsorb) nitrogen when air passes through this material, thus producing high concentrated oxygen.
There are two sieve beds: while one produces high concentration oxygen, the other is purged of the nitrogen it adsorbed (collected) during this pressure swing adsorption (PSA) cycle. Each adsorber produces oxygen and delivers it to the product tank. Oxygen exits the product tank through a pressure regulator, flow control valve, and flowmeter. The NewLife Intensity 10 unit delivers up to 95.5% oxygen at flow rates from 2-10 lpm.
2.2 Operation Check
AirSep tests every NewLife Intensity 10 Oxygen Concentrator thoroughly after manufacture. You must perform the following test to ensure that no damage occurred in shipping or handling.
1. Open and inspect all cartons (that contain units) upon delivery. Unpack the unit and remove it from the carton. Inspect the unit itself for damage. If the exterior of
a unit’s carton is damaged, or the unit itself is damaged, note it on the freight bill
signed by the driver.
2. Connect unit to power, and set the I/0 power switch to the “I” position. Check to see that the following occurs:
a. A continuous alarm sounds for approximately five seconds. See the
troubleshooting chart in Section 6.3 of this manual if the unit’s
alarm does anything other than sound for five seconds. b. The compressor runs and flow is indicated in flowmeter. c. OPTIONAL: The Oxygen Monitor’s amber light remains on until
the oxygen concentration reaches 85% ±3% (approximately two
minutes). Perform an oxygen concentration test, as described in
Section 2.4.
2.3 Alarm System
The NewLife Intensity 10 Oxygen Concentrator is equipped with a battery-powered alarm system, which sounds a continuous and loud alarm when a power failure occurs. It sounds an intermittent alarm if the high or low pressure indicators are activated or if the optional
MN134-1 REV D
Loading...
+ 21 hidden pages