NewLife® Elite / NewLife
Oxygen Concentrator
Service Manual
®
0459
AirSep® Corporation
AirSep Corporation • 401 Creekside Drive • Buffalo, NY 14228-2085 USA
T elephone: (716) 691-0202 • 24-Hour Fax: (716) 691-4141
Page 2
NewLife
®
Elite Service Manual
Expedited Product Warranty Check service is always at your fingertips with AirSep:
Type www.airsep.info/warranty_check/main.asp
OR
In the US or Canada,
dial 866-873-9277
AirSep® is a registered trademark of AirSep Corporation.
!
NewLife® Elite and NewLife® are registered trademarks of AirSep Corporation.
MN105-1 Rev. B 03/06
AirSep® Corporation
Page 3
T able of Contents
Section 1.0
Introduction
1.1Equipment Provider Responsibility1-1
1. 2Important Notice and Symbol Explanations1-2
1.3NewLife
1.4Functional Specifications1-5
Section 2.0
Operation Check
and Oxygen
Concentration Test
2.1Description of Operation2-1
NewLife
®
Elite/NewLife® Product Warranty1-4
®
Elite Service Manual
Section 3.0
Patient Instructions
Section 4.0
Equipment Provider
Maintenance
2. 2Operation Check2-1
2.3Alarm System2-2
2.3.1Battery T est2-2
2.3.2Power Failure Alarm T est2-2
2. 4Oxygen Concentration T est and S pecification2-3
3.1General Instructions3-1
3.2Routine Maintenance by the Patient3-1
3.2.1Cleaning the Air Intake
Gross Particle Filter3- 1
3.2.2Checking the Alarm System Battery3- 2
4.1Routine Maintenance4-1
AirSep® Corporation
4.1.1Air Intake Gross Particle Filter/GPF4- 1
4.1.2Product Filter Replacement4-1
4.1.3Optional Filter Replacement4-2
4.1.4Battery Replacement4-2
4.1.5Recording Maintenance4-3
4.2Cleaning and Infection Control4-3
4.2.1Preparing for New Patient Use4-4
i – table of contents
Page 4
®
NewLife
Elite Service Manual
Section 5.0
Service
5.1Components5-1
5.2Cabinet Removal5 -1
5.2.1Removing Side Panel(s)5-1
5.2.2Removing Back Panel5-1
5.2.3Removing Lower Front Cover5-1
5.2.4Removing Control Panel5-1
5.2.5Superstructure5-1
5.2.6Caster Replacement5-2
5.3Compressor5-2
5.3.1Compressor Replacement5-3
5.3.2Capacitor Replacement5-4
5.4Solenoid Valves5-4
5.4.1Feed or Waste V alve Rebuilding5-4
5.4.2Feed or Waste V alve Reassembly5-5
5.4.3Solenoid V alve Coil Replacement5-5
5.5Sieve Bed Replacement5-6
5.5.1Sieve Bed Removal5-6
5.5.2Sieve Bed Installation5-7
5.6Cabinet Fan Replacement5-7
5.7Circuit Board Replacement5-8
5.7.1Circuit Board Removal5-8
5.7.2Circuit Board Installation5-9
5. 8Product Regulator Check and Setting5-10
5.8.1Setting Product Regulator for
Normal Operation5-10
5.8.2Product Regulator Lockout5-10
5.8.3Back Pressure Correction at 5 lpm5-11
5.8.4Product Regulator Cleaning or
Rebuilding5-12
5. 9High or Low Pressure Switch Replacement
(In Equipped Units)5-13
ii – table of contents
5.10Circuit Breaker Replacement5-14
5.10.1Circuit Breaker Removal5-14
5.10.2Circuit Breaker Installation5-14
5.11ON/OFF Power Switch Replacement5-14
5.11.1ON/OFF Power Switch Removal5-14
AirSep® Corporation
Page 5
Section 6.0
Troubleshooting
NewLife
5.11.2ON/OFF Power Switch Installation5-15
5.12Buzzer Replacement5-15
5.13Hour Meter Replacement5-15
5.14Flowmeter Replacement5-16
5.14.1Flowmeter Removal5-16
5.14.2Flowmeter Installation5-16
5.15Power Cord Replacement5-16
6.1Operating Pressure T est6-1
6.1.1High Operating Pressure6-1
6.1.2Low Operating Pressure6-2
6.2General Troubleshooting6-2
6.3Troubleshooting Chart6-4
®
Elite Service Manual
Appendix
Exploded
Drawings
6. 4T ool Kit and Pressure T est Gauge6-7
Figure 1: Control Panel AssemblyA-1
Figure 2: Main Structure AssemblyA-2
Figure 3: Base and Cabinet ComponentsA-3
Figure 4: Test Block AssemblyA-4
Figure 5: V alve Block AssemblyA-5
Figure 6: Compressor AssemblyA-6
Figure 7: Adsorption Bed AssemblyA-7
AirSep® Corporation
iii – table of contents
Page 6
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 7
1.0Introduction
1.1Equipment Provider Responsibility
All Equipment Providers of the NewLife® Elite/NewLife® Oxygen Concentrator
must assume responsibilities for handling, operational check-out, patient instruction, and maintenance. These responsibilities are outlined below and throughout this manual.
NewLife
®
Elite Service Manual
WARNING
NewLife units must not be used for or with any lifesupporting applications. Geriatric, pediatric, or any other
patients unable to communicate discomfort while using
this machine may require additional monitoring. Advise
patients to immediately notify their Equipment Providers
and/or physicians in case of an alarm or any discomfort.
As an Equipment Provider, you must do all of the following:
!Inspect the condition of each NewLife unit immediately upon delivery
to your business location. Note any sign of damage, external or internal, on the delivery receipt, and report it directly to both the freight
company and AirSep Corporation immediately.
!Check the operation of each NewLife before delivery to a patient.
Always operate the unit for a reasonable length of time and check
that the oxygen concentration level is within specifications as referred
to in Section 2.4. (Test the battery and power disconnect alarm as
described in Section 2.3 of this manual.)
!Deliver NewLife units only to patients authorized by a physician’s
prescription. The NewLife must not be used as a life-supporting
device. A backup supply of oxygen must be available.
!Instruct patients how to use the NewLife in conjunction with the
Patient Manual.
!Instruct patients to notify their physicians and/or Equipment Provid-
ers if they experience any signs of discomfort.
!Instruct each patient how to perform routine maintenance of the air
intake gross particle filter and how to check the alarm system battery.
(Refer to Section 3.2.)
!Be available to service each patient at any time.
!Maintain the NewLife in accordance with Section 4.0.
!Repair components and replace parts only as outlined in this manual.
Use only AirSep parts for replacement in NewLife Oxygen Concentrators.
AirSep® Corporation
!Refer to the NewLife Product Warranty if parts replacement is re-
quired within the warranty period.
1-1
Page 8
NewLife
®
Elite Service Manual
1.2Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION,
and NOTE messages. They identify safety guidelines or other important information as follows:
WARNING:Describes a hazard or unsafe practice that can result in
severe bodily injury or death.
CAUTION:Describes a hazard or unsafe practice that can result in
minor bodily injury or property damage.
NOTE:Provides information important enough to emphasize or
repeat.
1-2
AirSep® Corporation
Page 9
NewLife
®
Elite Service Manual
The following harmonized symbols (pictograms), used for nonEnglish language countries, will be located on the exterior of the
NewLife unit:
Read the Patient Instruction Manual.
No smoking within five feet of this device, oxygencarrying tubing, or accessories.
No open flames within five feet of this device, oxygencarrying tubing, or accessories.
No oil, grease or petroleum-based products on or near
the unit.
Oxygen outlet
Oxygen alarm
AirSep® Corporation
Nebulizer option
1-3
Page 10
NewLife
®
Elite Service Manual
1.3NewLife Product Warranty
AirSep Corporation warrants the NewLife Elite/NewLife Oxygen Concentrator
to be free from defect in parts and workmanship for one, three, or five years (as
specified on the original invoice provided) from the date of delivery to the original purchaser, under normal use and operation. AirSep Corporation’s obligations under this warranty are limited to the repair or replacement of any such
item of equipment (or part thereof) shown to be defective or, at AirSep
Corporation’s option, to refunding the purchase price of any such defective item
of equipment.
Each item of equipment for which a warranty claim is asserted shall, at the request of AirSep Corporation, be returned on a prepaid basis with proof of purchase date to the AirSep factory at the expense of the purchaser. The purchaser
will be responsible for return freight charges. Replacement parts shall be warranted as stated above for the unexpired portion of the original (one-, three-, or
five-) year parts warranty (as specified on the original invoice provided). This
warranty does not extend to any item or part subjected to misuse, accident, improper maintenance, or application, or which has been repaired or altered outside of the AirSep Corporation Factory without the express prior written authorization of AirSep Corporation.
THE FOREGOING W ARRANTY IS IN LIEU OF ANY OTHER W ARRANTY , EXPRESSED OR
IMPLIED, IN F ACT OR IN LA W , INCLUDING WITHOUT LIMIT ATION THE WARRANTY OF
MERCHANT ABILITY OR THE WARRANTY OF FITNESS FOR P AR TICULAR PURPOSE. IT IS
EXPRESSL Y UNDERSTOOD THAT PURCHASER’S SOLE AND EXCLUSIVE REMEDY FOR
DEFECT IN P AR TS IS LIMITED TO ENFORCEMENT OF AIRSEP CORPORA TION’S OBLIGATION AS SET FORTH ABOVE, AND AIRSEP CORPORA TION SHALL NOT BE LIABLE TO
PURCHASER OR OTHERS FOR LOSS OF USE OF THE EQUIPMENT OR FOR OTHER SPECIAL,
INDIRECT , INCIDENTAL, OR CONSEQUEN-TIAL DAMAGES.
1-4
AirSep® Corporation
Page 11
1.4Functional Specifications
Dimensions:28.5 in. high, 15.7 in. wide, 14.5 in. deep
Concentration:1–3 liters per minute at 95% + 0.5/ - 3%
4 liters per minute at 92% + 3.5/ - 3%
5 liters per minute at 90% + 5.5/ - 3%
Response Time:Allow 5 minutes to attain maximum oxygen concentra-
tion.
NewLife
®
Elite Service Manual
Positioning:Operate the unit in an upright position, maintaining at
least six inches of open space on all sides for ventilation.
0459
AirSep® Corporation
1-5
Page 12
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 13
NewLife
®
Elite Service Manual
2.0Operation Check and Oxygen Concentration Test
2.1Description of Operation
Air enters the NewLife Oxygen Concentrator through an external air intake gross
particulate filter. This filtered air enters the compressor via a suction resonator,
which quiets the compressor’s suction sound. Pressurized air then exits the compressor and passes through a heat exchanger. The heat exchanger reduces the
temperature of the compressed air. Next, a two-way solenoid feed valve directs
the air into one of two sieve beds that contain molecular sieve. The unique property of molecular sieve enables it to physically attract (adsorb) nitrogen when air
passes through this material, thus producing high concentration oxygen.
There are two sieve beds: while one produces high concentration oxygen, the
other is purged of the nitrogen it adsorbed (collected) while it made oxygen. Each
adsorber produces oxygen for approximately eight seconds and delivers it to the
product tank. Oxygen exits the product tank through a pressure regulator, product filter, flow control valve, flowmeter, and finally exits the unit. The flow control valve, which is part of the flowmeter, controls the amount of oxygen delivered to the patient. The NewLife unit delivers up to 95.5% pure oxygen at flow
rates from 1–5 lpm.
NOTE: NewLife Elite OxiSilent™ option incorporates an additional felt intake
filter prior to air entering the compressor.
2.2Operation Check
AirSep tests every NewLife Oxygen Concentrator thoroughly after manufacture.
You must perform the following test to ensure that no damage occurred in shipping or handling.
1Open and inspect all cartons (that contain units) upon delivery. Un-
pack the unit and remove it from the carton. Inspect the unit itself for
damage. If the exterior of a unit’s carton is damaged, or the unit itself is
damaged, note it on the freight bill signed by the driver.
2Plug in the power cord of the unit, and set the ON/OFF switch to the
ON position. Check to see that the following occurs:
a.A continuous alarm loudly sounds for approximately five
seconds to indicate that the battery for the alarm is good.
See the troubleshooting chart in Section 6.0 of this
manual if the unit’s alarm does anything other than
sound loudly for five seconds.
AirSep® Corporation
b.The compressor runs. Listen for the sound.
c.Exhaust air flows out of the bottom of the unit.
2-1
Page 14
NewLife
®
Elite Service Manual
3Turn the flowmeter adjustment knob counterclockwise until it stops
(wide open).
The maximum flow should register 5.5 lpm (the 5.5 lpm line should
appear to split the middle of the flowmeter ball) with no connections at
the oxygen outlet. If not, refer to Section 5.8.1 to set the product regulator.
4 Perform an oxygen concentration test, as described in Section 2.4.
2.3Alarm System
The NewLife Oxygen Concentrator is equipped with a battery-powered alarm
system, which sounds a continuous and loud alarm when a power failure occurs. It sounds an intermittent alarm if the high or low pressure indicators are
activated or if the optional oxygen monitor detects lower than therapeutic levels
of oxygen concentration. The alarm remains on until you correct the alarm condition or you set the ON/OFF switch to the OFF position. Refer to Section 6.0 for
a list of probable alarm causes.
d.OPTIONAL: The Oxygen Monitor’s amber light remains
on until the oxygen concentration reaches 85% ± 3%
(approximately two minutes).
2.3.1Battery Test
Each time the NewLife unit is turned on, a five-second audible alarm
sounds to indicate the condition of the battery.
NOTE
The audio alarm must sound loudly for approximately five seconds each time the unit is turned to
ON to indicate the battery is in good condition.
2.3.2Power Failure Alarm Test
To test the power failure alarm, take the following step:
Unplug the unit, and set the ON/ OFF switch to the ON
position.
This should immediately activate the audio alarm. If it does not, refer to
the troubleshooting chart in Section 6.0 of this manual.
2-2
AirSep® Corporation
Page 15
NewLife
®
Elite Service Manual
2.4Oxygen Concentration Test and Specification
To ensure that the unit’s output of oxygen is within specification, you
must perform a test of oxygen concentration. Test the unit upon delivery to a patient and at periodic intervals. Equipment Providers, based
on their expertise and documentation, may establish and implement
their own protocol to check oxygen concentration. The interval established may be longer or shorter than 90 days, which is AirSep’s default
time period for providers who do not choose to establish their own protocol.
1If an oxygen humidifier bottle is used, remove it from the oxygen
outlet.
2Connect a calibrated oxygen concentration analyzer to the oxygen
outlet.
3Verify that the product flow rate delivered by the unit matches the
patient’s prescription and does not exceed the capacity of the unit.
4Set the unit’s ON/OFF power switch to the ON position. (It takes
approximately five minutes for the oxygen concentration to stabilize.) Take oxygen concentration readings every 60 seconds until
the analyzer shows that two consecutive readings are the same.
5Disconnect the oxygen analyzer, and reconnect the humidifier bottle
(if used) and any other equipment/ accessories that may be required.
6Adjust the flowmeter adjustment knob to the prescribed level.
NOTE
Do not measure oxygen concentration output
after the product stream passes through a humidifier bottle, or erroneous readings will result.
4 lpm92% + 3.5/ - 3%89% or higher
5 lpm90% + 5.5/ - 3%87% or higher
2-3
Page 16
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 17
3.0Patient Instructions
3.1General Instructions
It is important that patients thoroughly understand how to operate the AirSep
NewLife unit. This enables proper treatment as prescribed by a qualified, licensed
physician. You must explain that the purpose of this therapy is to alleviate symptoms. If patients experience any discomfort or the unit alarms, they must notify
their Equipment Provider and/or physician immediately.
You, as the Equipment Provider, are responsible to see that each patient receives
the Patient Manual. Locate the Patient Manual in the pocket on the back of each
unit. Explain each step in the operation of the unit to the patient in reference to
this manual.
3.2Routine Maintenance by the Patient
To ensure accurate output and efficient operation of the unit, the patient must
perform two simple routine maintenance tasks:
NewLife
®
Elite Service Manual
!Clean the air intake gross particle filter
!Check the alarm system battery
3.2.1Cleaning the Air Intake Gross Particle Filter
NOTE
The patient must clean this filter weekly, as
described below. The filter may require daily
cleaning if the NewLife unit operates in a harsh
environment such as a house heated by wood,
kerosene, or oil, or one with excessive cigarette
smoke.
1Remove the dirty air intake gross particle filter from the back
of the NewLife unit, and install the clean filter stored in the
pocket on the back of the unit.
AirSep® Corporation
2Wash the dirty filter in warm soapy water, and rinse.
3Use a soft absorbent towel to remove excess water.
4Place the clean air intake gross particle filter in the pocket on
the back of the unit.
3-1
Page 18
NewLife
®
Elite Service Manual
3.2.2Checking the Alarm System Battery
The alarm system battery is tested each time the ON/OFF switch is set
to the ON position. A continuous alarm sounds for approximately five
seconds to indicate a good battery. If the alarm does anything other
than sound loudly for five seconds, instruct the patient to call the Equipment Provider immediately.
3-2
AirSep® Corporation
Page 19
4.0Provider Maintenance
4.1Routine Maintenance
The NewLife Elite unit has two or three filters and a 9-volt battery that require
scheduled maintenance and replacement.
To ensure that the unit’s output of oxygen is within specification, you must perform a test of oxygen concentration. Test the unit upon delivery to a patient and
at periodic intervals. Equipment Providers, based on their expertise and documentation, may establish and implement their own protocol to check oxygen
concentration. The interval established may be longer or shorter than 90 days,
which is AirSep’s default time period for providers who do not choose to establish their own protocol.
AirSep does not require preventative maintenance on the concentrator. You do
not need to perform any maintenance as long as the NewLife unit remains within
specifications at the desired flow rate.
4.1.1Air Intake Gross Particle Filter/GPF
NewLife
®
Elite Service Manual
The external air intake gross particle filter is located on the back of the
unit. You can easily remove it by hand. Instruct the patient to clean this
filter weekly. (Refer to Section 3.2.1.)
NOTE
The filter may require more frequent cleaning if
the NewLife unit operates in a harsh environment
such as a house heated by wood, kerosene, or oil,
or one with excessive cooking or cigarette smoke.
4.1.2Product Filter Replacement
The product filter must be replaced after every 25,000 hours of use.
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the side panels to locate the product filter.
NOTE
AirSep® Corporation
Observe the position of the filter before
removal.
3Cut the tie-wraps, and separate the green tubing from both
sides of the filter.
4-1
Page 20
NewLife
®
Elite Service Manual
4.1.3Optional Filter Replacement
4Install the new filter with the inlet side in the same position as
before. Push the tubing together so that the tubing overlaps
the product filter connections, and secure it with the tie-wraps.
5Record information about the product filter replacement on
the History Record Label, which is discussed in Section 4.1.5.
6Reconnect the back and side panels.
The optional internal felt filter requires changing every 5,000 hours of
use. See below on instructions for changing the filter.
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove right side panel to locate the felt intake filter.
3Remove filter in the unit, and replace with a new filter.
4Remove left side panel and record information about the filter
replacement on the History Record Label.
5Reconnect both side panels.
4.1.4Battery Replacement
Each time the NewLife unit is turned on, the alarm must sound loudly
for approximately five seconds to indicate a good battery. An alarm
that does anything other than sound loudly for five seconds indicates a
weak battery and requires replacement.
To replace the battery, take the following steps:
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
4-2
2Remove the left side panel.
3Lift the battery out of the battery holder.
AirSep® Corporation
Page 21
4Install the new battery, maintaining proper polarity, and se-
cure it with the Velcro strap.
5Set the ON/OFF switch to the ON position to test the alarm.
6Record the battery replacement information on the History
Record Label.
7Reconnect the side panel.
4.1.5Recording Maintenance
As the Equipment Provider, you can record all routine maintenance
and repairs performed on the NewLife unit, including hours and dates
of service.
A History Record Label is located inside the unit. Keep this label current
to avoid unnecessary replacement of parts (i.e., product filter and battery).
NewLife
®
Elite Service Manual
4.2Cleaning and Infection Control
With the growing concern about possible cross infection from home oxygen equipment (i.e., oxygen concentrators) from one home care patient to another, a clarification on this topic is necessary.
The organisms of most concern are M. Tuberculosis, HIV, and Viral Hepatitis.
These are potentially pathogenic.
Tuberculosis can survive outside of the human body, but its mode of transmission is by droplet nuclei. When infected individuals cough, they release droplet
nuclei into the air, and these carry the Tuberculosis organism. These droplet nuclei may be breathed in by another person, but prolonged exposure to the infected person is usually necessary for infection to occur.
HIV and Viral Hepatitis are both viruses, which are not living cells themselves
but which can duplicate when in a living “host” cell. Both of these organisms are
usually passed on by person-to-person contact, and both need to be in the human body to survive. Once outside the body, viruses do not survive.
AirSep® Corporation
4-3
Page 22
NewLife
®
Elite Service Manual
4.2.1Preparing for New Patient Use
When you remove the NewLife from a patient’s home, always dispose
of the used nasal cannula and humidifier bottle. Clean the exterior of
the NewLife with a soapy water solution or commercial cleaner to remove any gross debris, organic or otherwise. Be careful not to get any
liquid into the interior of the unit.
Next, clean the exterior with either a common chemical disinfectant or
a bleach solution* before other patients use the unit. For the bleach solution, wear eye and skin protection to prevent exposure to the chlorine.
Allow the NewLife to air dry, and then retest it before you return it to
inventory.
Clean the air inlet gross particle filter with warm soapy water between
each patient’s use. Clean this filter at least once per week, depending on
the environment, during normal operation.
Change the product filter after 25,000 hours of use. It is not necessary to
change this filter between patients even if the previous patient had a
communicable disease or infection.
________________________________
*Make the bleach solution a 1:100 dilution of 5.25% sodium hypochlorite. Mix one part household bleach (e.g.,
Clorox) with 99 parts cold tap water. To measure the solution easily, take 1/4 cup of household bleach, and mix it
with a gallon of cold tap water. Allow the mixture to sit on potentially contaminated surfaces for 10 minutes.
4-4
AirSep® Corporation
Page 23
5.0Service
5.1Components
The design of the AirSep NewLife Oxygen Concentrator allows for easy access
and removal of most components. This allows you to perform scheduled maintenance, repair, and replacement of parts with minimal time and effort.
NewLife
CAUTION
For your safety, be sure to set the ON/OFF switch to the
OFF position and unplug the power cord of the unit
before you service the NewLife Oxygen Concentrator.
NOTE
Record all scheduled maintenance. (Refer to Section 4.0.)
®
Elite Service Manual
5.2Cabinet Removal
5.2.1Removing Side Panel(s)
To remove one or both side panels, unscrew the 1/4 turn fastener(s)
and remove the panel(s).
5.2.2Removing Back Panel
Remove both side panels, and lift off the back panel. Make sure the
power cord can pass freely through the power cord cutout.
5.2.3Removing Lower Front Cover
Firmly grasp panel with both hands, and slightly bow panel outward to
remove.
5.2.4Removing Control Panel
Four screws hold the control panel in place. Remove them and the control panel to re-install or replace if necessary.
5.2.5Superstructure
AirSep® Corporation
The weight and forces of the internal components rest solely on three
parts: the superstructure, compressor plate, and the base. These parts
were specially designed and formed. They should never require replacement under normal use.
5-1
Page 24
NewLife
®
Elite Service Manual
5.2.6Caster Replacement
1Remove the cabinet panels to expose the caster nut.
2Remove the caster nut with a 9/16-inch socket.
3Install the new caster and washer, and tighten the nut.
4Reconnect the cabinet panels.
5.3Compressor
The compressor is the “pump” within the oxygen concentrator that pushes the
room air into the bottom of the sieve beds. This allows oxygen to flow out of the
top.
Use an extension for the two front caster nut removals.
Two different aspects of the compressor cause concern: the output and the sound
level.
!Output
Compressor output refers to how much compressed air the compressor
can produce. This depends upon the model of the compressor, stroke
size, bore size and cup seal condition. The cup seals form the seal between the piston and the cylinder wall. As the cup seals wear, the
compressor’s output begins to gradually decrease. This reduction in compressor output results in less air for the sieve beds. Therefore, the production of oxygen decreases.
Since this drop in oxygen production occurs over a long period of
time, preventative maintenance on the compressor is not required.
You can continue a patient’s therapy on the NewLife unit as long as
that unit’s oxygen concentration level at the prescribed liter flow rate
is within AirSep’s specifications. Refer to Section 2.4.
!Sound Level
The condition of the compressor’s bearings mainly determine its sound
level.
5-2
There are four bearings located within the compressor that allow the
inner components of the compressor to rotate. If the bearings wear to
the point that they become noisy, the compressor becomes noticeably
loud and needs servicing.
AirSep® Corporation
Page 25
5.3.1Compressor Replacement
Compressor Assembly Removal
To remove the compressor assembly for exchange, follow the steps listed
below:
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the side and lower front panels.
3Disconnect the suction tube.
4Disconnect the blue compressor lead at the terminal strip and
the brown lead at the temperature switch.
5Disconnect the two leads to the capacitor and remove if neces-
sary.
NewLife
®
Elite Service Manual
6Disconnect the compression fitting for the heat exchanger lo-
cated at the bottom center of the compressor.
7Remove the two screws that connect the compressor plate to
the base of the unit, and lift up and slide out the compressor
assembly.
8If needed, remove compressor from the plate by removing the
four compressor bolts.
Compressor Assembly Installation
To install the rebuilt compressor assembly, follow the steps listed below:
1Perform the compressor removal procedure in reverse order.
2Make sure to position the compressor’s lead wires behind the
braided suction tube.
3Leak test all connections.
AirSep® Corporation
5-3
Page 26
NewLife
®
Elite Service Manual
5.3.2Capacitor Replacement
The capacitor starts the compressor. If the compressor cannot start, the
capacitor may be defective and require replacement. To replace the
capacitor, take the following steps:
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the side and lower front panels.
3Disconnect the two leads to the capacitor and slide capacitor
out of the tie wrap holding it in place.
4To install the new capacitor, connect the leads and slide the
capacitor into the tie wrap holding it in place.
5.4Solenoid V alves
The NewLife uses 5 two-way solenoid valves: two feed, two waste, and one equalization. Each valve has an open (energized) and closed (de-energized) position.
As the NewLife operates, two valves are always energized.
The solenoid valves of the NewLife unit require no scheduled maintenance. If a
valve becomes noisy, you can easily replace the internal valve parts. To identify
the noisy valve, observe which green circuit board light illuminates at the time
of the noise. The lighting matrix on the circuit board corresponds to the valve
location.
5.4.1Feed or Waste Valve Rebuilding
The two feed valves and two waste valves are located on the unified
valve block. This valve block does not require removal to rebuild a feed
or waste valve.
1Set the ON/OFF switch to the OFF position, and unplug the
power cord.
2Remove the side and back panels.
5-4
3Remove the retaining cap from the appropriate valve with a
slotted-head screwdriver.
4Lift off the solenoid coil.
AirSep® Corporation
Page 27
5Loosen and remove the solenoid base with a one-inch deep
well socket.
6Install the rebuild kit, which contains all parts of the solenoid
assembly except the solenoid coil.
5.4.2Feed or Waste Valve Reassembly
To reassemble the feed or waste valves, follow the feed or waste valve
rebuilding procedure in reverse order, and test for leaks.
5.4.3Solenoid Valve Coil Replacement
If a solenoid valve coil does not operate, its corresponding green light on
the circuit board does not illuminate.
1Set the ON/OFF switch to the OFF position, and unplug the
power cord.
NewLife
®
Elite Service Manual
2Remove the side and back panels.
3Remove the retaining cap with a slotted-head screwdriver.
4Disconnect the solenoid leads, and lift off the solenoid coil.
5Replace with the new coil.
6Reinstall the retaining cap back on top of the coil, and recon-
nect the solenoid leads.
7Reconnect the back and side panels.
AirSep® Corporation
5-5
Page 28
NewLife
®
Elite Service Manual
5.5Sieve Bed Replacement
WARNING
Do not refill the snap-ring-style sieve beds. Special
tooling is required to remove and re-install the springloaded top cap and snap-ring.
CAUTION
Do not expose molecular sieve (contents of bed) to air for
an extended period of time. Prolonged exposure of
molecular sieve to the moisture in room air results in
contamination and permanent damage to the sieve
material. Keep all openings to the sieve beds sealed.
NOTE
If replacement is necessary, you must replace both sieve
beds at the same time.
5.5.1 Sieve Bed Removal
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the side and back panels.
3Cut the tie-wraps at the brass “T” fitting, and disconnect the
green 1/4-inch product tubes.
4Cut the tie-wraps at the top of the sieve beds, and disconnect
the green 1/4-inch equalization tubes.
5-6
5Remove the 9/16-inch compression fitting on the bottom of
each sieve bed.
6Cut the tie-wraps, and remove the sieve beds. Proper removal
of the sieve beds includes the product tubes.
7Plug the opening on the top and bottom of the sieve beds.
AirSep® Corporation
Page 29
5.5.2Sieve Bed Installation
To install the sieve beds, follow the sieve bed removal procedure in reverse order. It is very important to tighten all tubes to eliminate leaks.
However, do not over tighten.
To check for leaks, take the following steps:
1Plug in the unit.
2Set the unit’s ON/OFF switch to ON for three minutes with
the flowmeter closed to pressurize the system.
3Set the unit’s ON/OFF switch to OFF, and unplug the power
cord.
4Apply soapy water around the sieve bed tube connections, and
check for leaks.
NewLife
®
Elite Service Manual
NOTE
Leaks can be so small in air loss that
oxygen concentration is not affected
immediately. The sieve material can
become contaminated gradually. Careful
leak testing is important.
5.6Cabinet Fan Replacement
The cabinet fan for the NewLife is located in the back of the unit. Refer to the
troubleshooting chart in Section 6.0 of this manual for instances where replacement of the fan may be required.
To replace the cabinet fan in the NewLife unit, take the following steps:
1Set the unit’s ON/OFF switch to the OFF position, and unplug the power
cord.
2Remove the side and back panels.
AirSep® Corporation
3Remove the two screws that hold the fan to the superstructure, and
remove the fan.
4Disconnect the fan leads.
5Position the new cabinet fan so that the air flow arrow points toward
the compressor and the electrical connections are in the bottom right
corner.
5-7
Page 30
NewLife
®
Elite Service Manual
6Connect the fan leads, and install the cabinet fan screws.
7Reconnect the back and side panels.
5.7Circuit Board Replacement
The solid-state printed circuit board controls the sequential timing operation of
the five solenoid valves and the alarm system functions. The five-green-light
matrix corresponds to the valve configuration of the NewLife unit. Two green
lights should always illuminate and rotate (cycle) during normal operation.
A green light that fails to illuminate indicates a disconnected or faulty solenoid
coil or an electrical malfunction in that valve circuit. The two red lights indicate
high and low pressure.
Consult the troubleshooting chart in Section 6.0 to determine when to replace
the printed circuit board.
CAUTION
The Printed Circuit Boards (PCBs) contain components
that are sensitive to electrostatic discharge (ESD) and can
damage the board if not handled properly. As when
handling any ESD-sensitive PCB, observe standard ESD
safety procedures. These procedures include the following:
!Handle the PCB by the edges only.
!Work on a grounded ESD mat.
!Wear a grounded wrist strap.
!Store PCBs only in anti-static bags.
5.7.1Circuit Board Removal
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
5-8
2Remove the side and back panels.
3Disconnect the main power 10-pin connector from the circuit
board.
AirSep® Corporation
Page 31
NewLife
®
Elite Service Manual
NOTE
If a pressure transducer circuit board, cut
tie-wrap at circuit board and disconnect
green 1/4-inch tube. Plug/cap open end
of tube until new circuit board is replaced.
4Push in on the locking tab to the alarm 6-pin connector to dis-
connect it.
5With a slotted screwdriver, push in on the board support tabs,
while you lift each corner of the circuit board.
6Remove the circuit board.
NOTE
Handle the new circuit board only by the
edges to prevent electrostatic damage to
the unit.
5.7.2Circuit Board Installation
NOTE
If a pressure transducer circuit board,
first reinstall 1/4-inch green tube with
new tie-wrap.
1Push the circuit board on to the support tabs.
AirSep® Corporation
2Firmly plug in the 6-pin connector to the circuit board so that
the connector’s locking tab locks against the circuit board.
3Plug in the 10-pin connector.
4Reconnect the back and side panels.
5-9
Page 32
NewLife
®
Elite Service Manual
5.8Product Regulator Check and Setting
The product regulator enables you to set the maximum flow of oxygen output by
the NewLife unit. To check for proper adjustment of the product regulator, take
the following steps:
1Set the ON/OFF switch to the ON position.
2Allow the unit to run for three minutes.
3Turn the flowmeter adjustment knob counterclockwise until it stops
(wide open).
4The flowmeter ball centers itself on the 5.5 lpm line. If not, the product
regulator needs to be reset.
5.8.1Setting Product Regulator for Normal Operation
Use the following procedure to reset the product regulator:
1Disconnect the humidifier bottle, if used, and the tubing from
the oxygen outlet.
2Plug in the unit.
3Set the unit’s ON/OFF switch to the ON position, and allow
the unit to run at least three minutes to build up pressure.
4Remove the right side panel.
5Turn the flowmeter adjustment knob counterclockwise until it
stops (wide open).
6Pull outward on the regulator knob to unlock it.
7Turn the regulator knob or set screws until the flowmeter ball
centers on the 5.5 lpm line (clockwise to increase).
8Push in the regulator knob to lock it.
5-10
9Reconnect the side panel.
5.8.2Product Regulator Lockout
Take the following steps to limit the maximum flow of oxygen concentrator:
AirSep® Corporation
Page 33
NewLife
®
Elite Service Manual
NOTE
It may be desirable to provide one additional liter
of flow if you use the lockout feature.
1Plug in the unit.
2Set the unit’s ON/OFF switch to the ON position, and allow
the unit to run at least three minutes to build up pressure.
3If used, connect the humidifier bottle, oxygen tubing, and can-
nula to the oxygen outlet.
4Remove the right side panel.
5Turn the flowmeter adjustment knob counterclockwise until it
stops (wide open).
6Pull outward on the regulator knob to unlock it.
NOTE
When this lockout feature is no longer
required, follow the normal regulator
setting procedure, as described in Section
5.8.1.
7Turn the regulator knob counterclockwise to decrease the set-
ting until the flowmeter registers the maximum flow desired.
8Push in the regulator knob to lock the setting.
9Reconnect the side panel.
5.8.3Back Pressure Correction at 5 LPM
AirSep® Corporation
In some cases, additional lengths of oxygen tubing with a humidifier
bottle can increase back pressure and limit oxygen flow below 5 lpm.
To achieve 5 lpm, use the following procedure:
1Plug in the unit.
5-11
Page 34
NewLife
®
Elite Service Manual
2Set the ON/OFF switch to the ON position, and allow the unit
to run at least three minutes to build up pressure.
3Connect the humidifier bottle, oxygen tubing, and cannula to
the oxygen outlet.
4Remove the right side panel.
5Turn the flowmeter adjustment knob counterclockwise until it
stops (wide open).
6Pull outward on the regulator knob to unlock it.
7Turn the regulator knob clockwise until the flowmeter ball cen-
ters on the 5.5 lpm line.
8Push in the regulator knob to lock the setting.
9Turn the flowmeter adjustment knob clockwise until the flow-
meter indicates 5 lpm.
10Reconnect the side panel.
NOTE
When this back pressure correction at 5
lpm is no longer required, follow the
normal regulator setting procedure as
described in Section 5.8.1.
5.8.4Product Regulator Cleaning or Rebuilding
Clean or rebuild the product regulator if the flowmeter ball fluctuates
more than 1/4 of a liter or if the regulator cannot be adjusted for lockout.
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
5-12
2Remove the right side panel.
3Use large pliers to unscrew the bonnet of the product regula-
tor, which contains a large spring.
AirSep® Corporation
Page 35
NewLife
NOTE
Adjust the product regulator fully counterclockwise to unload the spring. This
makes disassembly and reassembly
easier.
4Remove the diaphragm. (Clean or replace it.)
5Use a Phillips-head screwdriver to unscrew the diaphragm stem
guide located in the center of the regulator body to gain access
to the seat.
6Remove the seat. Be careful not to lose the spring located
behind the seat.
®
Elite Service Manual
7Clean or replace the seat.
8With the spring behind the seat, screw the diaphragm stem
guide back into the body of the regulator. (Do not over tighten.)
9Install a clean or replacement diaphragm.
10Put the large spring and slip ring into the bonnet, and screw
the bonnet onto the regulator body.
11Reset the product regulator as described in Section 5.8.1.
5.9High or Low Pressure Switch Replacement
(In Equipped Units)
The high pressure switch is located closer than the low pressure switch to the
pressure test port. Remove it first to gain access to the low pressure switch.
1Set the unit’s ON/OFF switch to the OFF position, and unplug the
power cord.
AirSep® Corporation
2Remove the right side panel.
3Disconnect the leads to the pressure switch.
4Unscrew the pressure switch from the pressure test block.
5-13
Page 36
NewLife
®
Elite Service Manual
5Install the new pressure switch, connect the leads, and leak test.
6Reconnect the side panel.
5.10Circuit Breaker Replacement
5.10.1 Circuit Breaker Removal
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the left side panel.
3Disconnect the circuit breaker leads.
4Unscrew the circuit breaker while you apply pressure to the
circuit breaker retaining ring.
5.10.2 Circuit Breaker Installation
Follow the removal procedure for the circuit breaker in reverse order to
install the new circuit breaker.
5.11ON/OFF Power Switch Replacement
5.11.1 ON/OFF Power Switch Removal
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove both side panels, the lower front panel, and the back
cover.
3Unscrew the four Phillips-head screws that hold the control
panel to the superstructure.
4Cut off the two white tie-wraps from the terminal strip leads.
5-14
5Disconnect the ON/OFF switch leads from the back of the
switch.
6Push on the back of the power switch, while holding in its
retaining tabs, and remove the switch through the front panel.
AirSep® Corporation
Page 37
5.11.2 ON/OFF Power Switch Installation
Follow the removal procedure for the ON/OFF power switch in reverse
order to install a new power switch. Make sure to reinstall the switch
properly by having the “O” on the switch located on the bottom when
finished.
5.12Buzzer Replacement
1Set the unit’s ON/OFF switch to the OFF position, and unplug the power
cord.
2Remove the left side panel.
3Unscrew the two Phillips-head screws that hold the buzzer to the su-
perstructure, and solder to disconnect the buzzer leads.
4Solder the leads to the new buzzer, and screw the buzzer to the super-
structure.
NewLife
®
Elite Service Manual
5Reconnect the side panel.
5.13Hour Meter Replacement
1Set the unit’s ON/OFF switch to the OFF position, and unplug the power
cord.
2Remove the left side panel.
3Disconnect the hour meter leads.
4Push hour meter mounting tabs to remove the hour meter through the
control panel.
5Install the new hour meter into the control panel, making sure that the
“hourglass” icon displays on the bottom.
6Reconnect the hour meter leads.
AirSep® Corporation
7Reconnect the side panel.
8Plug unit in and set the unit’s ON/OFF switch to ON. Hourglass icon
should flash showing it is functioning correctly.
5-15
Page 38
NewLife
®
Elite Service Manual
5.14Flowmeter Replacement
5.14.1 Flowmeter Removal
1Set the unit’s ON/OFF switch to the OFF position, and unplug
the power cord.
2Remove the left side panel.
3Cut the tie-wraps, and remove the 3/8-inch green oxygen tub-
ing from the flowmeter fittings.
4Unscrew the flowmeter nuts with a wrench.
5Remove the flowmeter through the control panel.
5.14.2 Flowmeter Installation
To install a new flowmeter, follow the flowmeter removal procedure in
reverse order. Then perform a leak test on the connections.
5.15Power Cord Replacement
1Set the unit’s ON/OFF switch to the OFF position, and unplug the power
cord.
2Remove the side and back panels.
3Open the plastic twist clamps.
4Cut the tie-wraps (located near the power cord push-on connectors).
5Disconnect the white and black power cord leads.
6Lay the unit face down, and unscrew the two Phillips-head screws that
secure the power cord receptacle to the base of the unit.
5-16
7Use a slotted-head screwdriver to break the receptacle seal.
8Pull the receptacle forward until the back of the strain relief is visible.
9Press both ends of the strain relief together with needle-nose pliers to
remove and save it for reuse.
AirSep® Corporation
Page 39
NewLife
®
Elite Service Manual
10Pull the power cord through the opening on the power cord receptacle.
11Using same distance located on old power cord, slide the strain relief
onto the new power cord with the lip of the strain relief facing the ends
of the power cord leads.
12Insert the new power cord through the opening on the receptacle by
squeezing the strain relief together with pliers.
13Position the strain relief upward, and push it into the rear of the power
cord opening on the receptacle.
14Reconnect the power cord leads, close the plastic twist clamps, and re-
attach the tie-wraps.
15Reconnect the power cord receptacle to the base of the unit with the
two Phillips-head screws.
16Reconnect the side and back panels.
AirSep® Corporation
5-17
Page 40
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 41
6.0Troubleshooting
6.1Operating Pressure Test
Testing the operating pressure is a useful diagnostic tool when a concentrator
has low oxygen concentration and requires servicing. Units functioning normally do not require operating tests.
Use the following procedure to test the operating pressure of the unit.
1Set the unit’s ON/OFF switch to the OFF position, and unplug the power
cord.
2Remove the right side panel.
3Connect the pressure test gauge to appropriate test port location. (Some
models may require an additional test adapter (KI257-1) which allows
connecting directly to the 1/4-inch black pressure tubing from the valve
block.)
NewLife
®
Elite Service Manual
4Plug in the power cord, and set the unit’s ON/OFF power switch to the
ON position. Set the flowmeter to 5 lpm, and allow the unit to run at
least five minutes.
5Observe the maximum and minimum readings on the pressure test gauge.
The maximum reading should not exceed 32 psig (220.6 kPa). The minimum reading should not be less than 10 psig (69 kPa).
NOTE
When you turn the unit on, the system pressure
always registers higher than normal for the first
few minutes of operation. Make sure the unit is
not in economy mode of operation if equipped.
6.1.1High Operating Pressure
Higher than normal operating pressure may indicate any of the following:
AirSep® Corporation
! A restrictive blowdown muffler, which does not allow the waste
(purge) gas to exit the system freely.
Operate the unit with the blowdown muffler disconnected to see
if the operating pressure returns to normal.
6-1
Page 42
NewLife
®
Elite Service Manual
6.1.2Low Operating Pressure
! An improperly operating circuit board or solenoid valve.
Confirm that the circuit board and solenoid valves function
properly.
! Contaminated sieve beds.
Change the sieve beds.
Lower than normal operating pressure may indicate any of the following:
! A restriction in the suction resonator or optional air intake filter,
which limits the amount of room air available to the compressor.
Disconnect the braided suction tube at the compressor, and
allow the unit to operate without the suction resonator to see if
normal operating pressure returns.
! An improperly operating circuit board or solenoid valve.
Confirm that the circuit board and solenoid valves function
properly.
! A leak in the unit, which allows system pressure to escape.
Leak test the unit.
! A compressor with reduced output.
Ensure that the oxygen concentration level at the desired liter
flow is within AirSep’s specifications. If it is below specifications,
replace or repair the compressor.
! An obstructed or vertically positioned (ECOnomy mode)
EcoCheck valve.
For the EcoCheck valve (available only as an option for units
equipped with an oxygen monitor): confirm that the EcoCheck
valve is not obstructed or in the ECOnomy mode.
6.2General T roubleshooting
6-2
Before reviewing the troubleshooting chart, the following steps may be useful
to isolate any malfunctions:
1Turn the concentrator on. If unit does not turn on, refer to
troubleshooting chart.
AirSep® Corporation
Page 43
NewLife
®
Elite Service Manual
2Make sure all filters are clean.
3Turn flowmeter completely open, up to the highest setting. The
ball should rest at 5.5 lpm. If setting is not 5.5 lpm, adjust
regulator so that the ball is set at 5.5 lpm with the flowmeter
completely open.
4Make sure the unit is cycling properly by following the green
valve lights on the circuit board. The lights should follow the
label below the circuit board and two lights should be activated at all times. If the unit is not cycling properly, refer to
troubleshooting chart.
5If concentrator is not meeting specifications, make sure that
the unit is leak free by testing all tubing connections and fittings with leak testing solution. Protect circuit board from solution and start leak test at the heat exchanger, following air
flow of unit to oxygen outlet. Repair all leaks by tightening
connections and fittings.
6Set the concentrator at 5 lpm and connect pressure test gauge
to unit. Determine pressure parameters by observing high and
low pressure points on the gauge. If pressures are high or low,
refer to Section 6.1.
7Review troubleshooting chart to isolate and repair any other
malfunctions.
AirSep® Corporation
6-3
Page 44
®
NewLife
Elite Service Manual
6.3T roubleshooting Chart
Problem Probable Cause Solution
Compressor does not run. ConstantNo power to unit.Check wall outlet for power.
audio alarm with ON/OFF power
switch in ON position.Unit circuit breaker tripped orReset or replace circuit breaker .
Defective circuit board.Check circuit board for two-light
rotation/replace circuit board.
Defective ON/OFF powerReplace ON/OFF power switch.
switch.
Compressor runs with intermittentDefective high or lowReplace and retest pressure
alarm. Oxygen concentration atpressure switch.switch.
5 LPM is within specifications.
Compressor shuts downRestricted air flow throughClean inlet filter, or remove
intermittently.unit.obstruction.
Unit overheating due toLocate unit away from heating
improper location.source, providing adequate
ventilation on all sides.
Defective cabinet fan.Replace cabinet fan.
Defective high temperatureReplace high temperature switch.
switch.
Compressor does not start. ON/OFF Extreme cold start.Allow unit to reach room
power switch in ON position,temperature.
intermittent alarm, and cabinet fan
turns.Compressor thermally cut offBlocked air intake or defective
due to excessive heat. NOTE:cabinet fan/clear obstruction, or
It will not restart unit until itreplace cabinet fan.
cools down.
Defective capacitor.Replace capacitor.
6-4
Defective circuit board.Check circuit board for two-light
rotation/replace circuit board.
Faulty electrical connectionCheck electrical connections for
for compressor.compressor.
Defective high temperatureReplace high temperature switch.
switch.
AirSep® Corporation
Page 45
NewLife
®
Elite Service Manual
Problem Probable Cause Solution
Compressor runs with intermittentLeak.Leak test and repair leak.
low pressure alarm and low oxygen
concentration.
Defective circuit board.Check circuit board for two-light
rotation/replace circuit board.
Compressor runs with intermittentDefective sieve beds.Replace sieve beds.
high pressure alarm and low
oxygen concentration.
Faulty solenoid valve.Repair or replace solenoid valve.
Restriction in exhaust muffler .Replace or clean muffler.
Weak compressor.Check system pressure, and rebuild
or replace compressor.
Unit’s temperature is too high.Blocked air intake or defective
cabinet fan.
Defective circuit board.Check circuit board for two-light
rotation/replace circuit board.
6-6
Contaminated sieve beds.Replace sieve beds.
Defective solenoid valve.Repair or replace solenoid valve.
Restriction in exhaust muffler.Replace or clean exhaust muffler.
Restriction in suction resonator.Check suction resonator and suction
tube for obstruction and remove.
Restriction in EcoCheck valve.Clean EcoCheck valve.
AirSep® Corporation
Page 47
6.4Tool Kit and Pressure Test Gauge
The tools needed for you to properly service the NewLife unit are listed below:
Large pliers, wire cutters, needle-nose pliers, slotted-head screwdriver, Phillips-
head screwdriver, ratchet, 8-inch adjustable wrench, 1-inch deep well socket, 9/
16-inch socket, 2-inch extension, 7/16-inch combination wrench, 9/16-inch combination wrench, and 5/8-inch combination wrench.
In addition, channel locks and a small slotted-head screwdriver may be recommended for some service procedures.
A pressure test gauge to read operating system pressures on the NewLife unit
should be kept available at all times.
NewLife
®
Elite Service Manual
AirSep® Corporation
6-7
Page 48
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 49
Appendix
!!
!Exploded Drawings
!!
Control Panel AssemblyA-1
Main Structure AssemblyA-2
Base and Cabinet ComponentsA-3
Test Block AssemblyA-4
Valve Block AssemblyA-5
Compressor AssemblyA-6
Adsorption Bed AssemblyA-7
NewLife
®
Elite Service Manual
AirSep® Corporation
Page 50
NewLife
®
Elite Service Manual
Ref. No.
Ref. No.
Ref. No.
Part No.
Part No.
Part No.
Description
Description
Description
Ref. No.
Ref. No.
Ref. No.
Part No.
Part No.
Part No.
Description
Description
Description
A-1
Figure A.1: Control Panel Assembly
AirSep® Corporation
Page 51
Part No.Ref. No.DescriptionDescriptionPart No.Ref. No.