AirSep NewLife Elite, NewLife User manual

NewLife
®
Elite Service Manual
NewLife® Elite / NewLife Oxygen Concentrator Service Manual
®
0459
AirSep® Corporation
AirSep Corporation • 401 Creekside Drive • Buffalo, NY 14228-2085 USA
T elephone: (716) 691-0202 • 24-Hour Fax: (716) 691-4141
NewLife
®
Elite Service Manual
Expedited Product Warranty Check service is always at your fingertips with AirSep:
Type www.airsep.info/warranty_check/main.asp
OR
In the US or Canada,
dial 866-873-9277
AirSep® is a registered trademark of AirSep Corporation.
!
NewLife® Elite and NewLife® are registered trademarks of AirSep Corporation.
MN105-1 Rev. B 03/06
AirSep® Corporation
T able of Contents
Section 1.0
Introduction
1.1 Equipment Provider Responsibility 1-1
1. 2 Important Notice and Symbol Explanations 1-2
1.3 NewLife
1.4 Functional Specifications 1-5
Section 2.0
Operation Check and Oxygen Concentration Test
2.1 Description of Operation 2-1
NewLife
®
Elite/NewLife® Product Warranty 1-4
®
Elite Service Manual
Section 3.0
Patient Instructions
Section 4.0
Equipment Provider Maintenance
2. 2 Operation Check 2-1
2.3 Alarm System 2-2
2.3.1 Battery T est 2-2
2.3.2 Power Failure Alarm T est 2-2
2. 4 Oxygen Concentration T est and S pecification 2-3
3.1 General Instructions 3-1
3.2 Routine Maintenance by the Patient 3-1
3.2.1 Cleaning the Air Intake Gross Particle Filter 3- 1
3.2.2 Checking the Alarm System Battery 3- 2
4.1 Routine Maintenance 4-1
AirSep® Corporation
4.1.1 Air Intake Gross Particle Filter/GPF 4- 1
4.1.2 Product Filter Replacement 4-1
4.1.3 Optional Filter Replacement 4-2
4.1.4 Battery Replacement 4-2
4.1.5 Recording Maintenance 4-3
4.2 Cleaning and Infection Control 4-3
4.2.1 Preparing for New Patient Use 4-4
i – table of contents
®
NewLife
Elite Service Manual
Section 5.0
Service
5.1 Components 5-1
5.2 Cabinet Removal 5 -1
5.2.1 Removing Side Panel(s) 5-1
5.2.2 Removing Back Panel 5-1
5.2.3 Removing Lower Front Cover 5-1
5.2.4 Removing Control Panel 5-1
5.2.5 Superstructure 5-1
5.2.6 Caster Replacement 5-2
5.3 Compressor 5-2
5.3.1 Compressor Replacement 5-3
5.3.2 Capacitor Replacement 5-4
5.4 Solenoid Valves 5-4
5.4.1 Feed or Waste V alve Rebuilding 5-4
5.4.2 Feed or Waste V alve Reassembly 5-5
5.4.3 Solenoid V alve Coil Replacement 5-5
5.5 Sieve Bed Replacement 5-6
5.5.1 Sieve Bed Removal 5-6
5.5.2 Sieve Bed Installation 5-7
5.6 Cabinet Fan Replacement 5-7
5.7 Circuit Board Replacement 5-8
5.7.1 Circuit Board Removal 5-8
5.7.2 Circuit Board Installation 5-9
5. 8 Product Regulator Check and Setting 5-10
5.8.1 Setting Product Regulator for Normal Operation 5-10
5.8.2 Product Regulator Lockout 5-10
5.8.3 Back Pressure Correction at 5 lpm 5-11
5.8.4 Product Regulator Cleaning or Rebuilding 5-12
5. 9 High or Low Pressure Switch Replacement (In Equipped Units) 5-13
ii – table of contents
5.10 Circuit Breaker Replacement 5-14
5.10.1 Circuit Breaker Removal 5-14
5.10.2 Circuit Breaker Installation 5-14
5.11 ON/OFF Power Switch Replacement 5-14
5.11.1 ON/OFF Power Switch Removal 5-14
AirSep® Corporation
Section 6.0
Troubleshooting
NewLife
5.11.2 ON/OFF Power Switch Installation 5-15
5.12 Buzzer Replacement 5-15
5.13 Hour Meter Replacement 5-15
5.14 Flowmeter Replacement 5-16
5.14.1 Flowmeter Removal 5-16
5.14.2 Flowmeter Installation 5-16
5.15 Power Cord Replacement 5-16
6.1 Operating Pressure T est 6-1
6.1.1 High Operating Pressure 6-1
6.1.2 Low Operating Pressure 6-2
6.2 General Troubleshooting 6-2
6.3 Troubleshooting Chart 6-4
®
Elite Service Manual
Appendix
Exploded Drawings
6. 4 T ool Kit and Pressure T est Gauge 6-7
Figure 1: Control Panel Assembly A-1 Figure 2: Main Structure Assembly A-2 Figure 3: Base and Cabinet Components A-3 Figure 4: Test Block Assembly A-4 Figure 5: V alve Block Assembly A-5 Figure 6: Compressor Assembly A-6 Figure 7: Adsorption Bed Assembly A-7
AirSep® Corporation
iii – table of contents
NewLife
®
Elite Service Manual
AirSep® Corporation
1.0 Introduction
1.1 Equipment Provider Responsibility
All Equipment Providers of the NewLife® Elite/NewLife® Oxygen Concentrator must assume responsibilities for handling, operational check-out, patient instruc­tion, and maintenance. These responsibilities are outlined below and through­out this manual.
NewLife
®
Elite Service Manual
WARNING
NewLife units must not be used for or with any life­supporting applications. Geriatric, pediatric, or any other patients unable to communicate discomfort while using this machine may require additional monitoring. Advise patients to immediately notify their Equipment Providers and/or physicians in case of an alarm or any discomfort.
As an Equipment Provider, you must do all of the following:
! Inspect the condition of each NewLife unit immediately upon delivery
to your business location. Note any sign of damage, external or inter­nal, on the delivery receipt, and report it directly to both the freight company and AirSep Corporation immediately.
! Check the operation of each NewLife before delivery to a patient.
Always operate the unit for a reasonable length of time and check that the oxygen concentration level is within specifications as referred to in Section 2.4. (Test the battery and power disconnect alarm as described in Section 2.3 of this manual.)
! Deliver NewLife units only to patients authorized by a physician’s
prescription. The NewLife must not be used as a life-supporting device. A backup supply of oxygen must be available.
! Instruct patients how to use the NewLife in conjunction with the
Patient Manual.
! Instruct patients to notify their physicians and/or Equipment Provid-
ers if they experience any signs of discomfort.
! Instruct each patient how to perform routine maintenance of the air
intake gross particle filter and how to check the alarm system battery. (Refer to Section 3.2.)
! Be available to service each patient at any time. ! Maintain the NewLife in accordance with Section 4.0. ! Repair components and replace parts only as outlined in this manual.
Use only AirSep parts for replacement in NewLife Oxygen Concentra­tors.
AirSep® Corporation
! Refer to the NewLife Product Warranty if parts replacement is re-
quired within the warranty period.
1-1
NewLife
®
Elite Service Manual
1.2 Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important informa­tion as follows:
WARNING: Describes a hazard or unsafe practice that can result in
severe bodily injury or death.
CAUTION: Describes a hazard or unsafe practice that can result in
minor bodily injury or property damage.
NOTE: Provides information important enough to emphasize or
repeat.
1-2
AirSep® Corporation
NewLife
®
Elite Service Manual
The following harmonized symbols (pictograms), used for non­English language countries, will be located on the exterior of the NewLife unit:
Read the Patient Instruction Manual.
No smoking within five feet of this device, oxygen­carrying tubing, or accessories.
No open flames within five feet of this device, oxygen­carrying tubing, or accessories.
No oil, grease or petroleum-based products on or near the unit.
Oxygen outlet
Oxygen alarm
AirSep® Corporation
Nebulizer option
1-3
NewLife
®
Elite Service Manual
1.3 NewLife Product Warranty
AirSep Corporation warrants the NewLife Elite/NewLife Oxygen Concentrator to be free from defect in parts and workmanship for one, three, or five years (as specified on the original invoice provided) from the date of delivery to the origi­nal purchaser, under normal use and operation. AirSep Corporation’s obliga­tions under this warranty are limited to the repair or replacement of any such item of equipment (or part thereof) shown to be defective or, at AirSep Corporation’s option, to refunding the purchase price of any such defective item of equipment.
Each item of equipment for which a warranty claim is asserted shall, at the re­quest of AirSep Corporation, be returned on a prepaid basis with proof of pur­chase date to the AirSep factory at the expense of the purchaser. The purchaser will be responsible for return freight charges. Replacement parts shall be war­ranted as stated above for the unexpired portion of the original (one-, three-, or five-) year parts warranty (as specified on the original invoice provided). This warranty does not extend to any item or part subjected to misuse, accident, im­proper maintenance, or application, or which has been repaired or altered out­side of the AirSep Corporation Factory without the express prior written autho­rization of AirSep Corporation.
THE FOREGOING W ARRANTY IS IN LIEU OF ANY OTHER W ARRANTY , EXPRESSED OR IMPLIED, IN F ACT OR IN LA W , INCLUDING WITHOUT LIMIT ATION THE WARRANTY OF MERCHANT ABILITY OR THE WARRANTY OF FITNESS FOR P AR TICULAR PURPOSE. IT IS EXPRESSL Y UNDERSTOOD THAT PURCHASER’S SOLE AND EXCLUSIVE REMEDY FOR DEFECT IN P AR TS IS LIMITED TO ENFORCEMENT OF AIRSEP CORPORA TION’S OBLIGA­TION AS SET FORTH ABOVE, AND AIRSEP CORPORA TION SHALL NOT BE LIABLE TO PURCHASER OR OTHERS FOR LOSS OF USE OF THE EQUIPMENT OR FOR OTHER SPECIAL, INDIRECT , INCIDENTAL, OR CONSEQUEN-TIAL DAMAGES.
1-4
AirSep® Corporation
1.4 Functional Specifications
Dimensions: 28.5 in. high, 15.7 in. wide, 14.5 in. deep
(72.4 cm high, 40.0 cm wide, 36.8 cm deep)
Weight: 54.0 lb; shipping weight - 59.0 lb
(24.5 kg; shipping weight - 29.0 kg)
Electrical 120 VAC, 60 Hz, 4 amps, 350 watts Requirements: 230 VAC, 50 Hz, 4 amps, 350 watts
Capacity: 5 liters per minute at 90% oxygen
(Based on 70°F [21°C] at sea level)
Accuracy: Flowmeter ±5% full scale. (Based on
precalibrated analysis.)
Concentration: 1–3 liters per minute at 95% + 0.5/ - 3%
4 liters per minute at 92% + 3.5/ - 3% 5 liters per minute at 90% + 5.5/ - 3%
Response Time: Allow 5 minutes to attain maximum oxygen concentra-
tion.
NewLife
®
Elite Service Manual
Positioning: Operate the unit in an upright position, maintaining at
least six inches of open space on all sides for ventilation.
0459
AirSep® Corporation
1-5
NewLife
®
Elite Service Manual
AirSep® Corporation
NewLife
®
Elite Service Manual
2.0 Operation Check and Oxygen Concentration Test
2.1 Description of Operation
Air enters the NewLife Oxygen Concentrator through an external air intake gross particulate filter. This filtered air enters the compressor via a suction resonator, which quiets the compressor’s suction sound. Pressurized air then exits the com­pressor and passes through a heat exchanger. The heat exchanger reduces the temperature of the compressed air. Next, a two-way solenoid feed valve directs the air into one of two sieve beds that contain molecular sieve. The unique prop­erty of molecular sieve enables it to physically attract (adsorb) nitrogen when air passes through this material, thus producing high concentration oxygen.
There are two sieve beds: while one produces high concentration oxygen, the other is purged of the nitrogen it adsorbed (collected) while it made oxygen. Each adsorber produces oxygen for approximately eight seconds and delivers it to the product tank. Oxygen exits the product tank through a pressure regulator, prod­uct filter, flow control valve, flowmeter, and finally exits the unit. The flow con­trol valve, which is part of the flowmeter, controls the amount of oxygen deliv­ered to the patient. The NewLife unit delivers up to 95.5% pure oxygen at flow rates from 1–5 lpm.
NOTE: NewLife Elite OxiSilent™ option incorporates an additional felt intake filter prior to air entering the compressor.
2.2 Operation Check
AirSep tests every NewLife Oxygen Concentrator thoroughly after manufacture. You must perform the following test to ensure that no damage occurred in ship­ping or handling.
1 Open and inspect all cartons (that contain units) upon delivery. Un-
pack the unit and remove it from the carton. Inspect the unit itself for damage. If the exterior of a unit’s carton is damaged, or the unit itself is damaged, note it on the freight bill signed by the driver.
2 Plug in the power cord of the unit, and set the ON/OFF switch to the
ON position. Check to see that the following occurs:
a. A continuous alarm loudly sounds for approximately five
seconds to indicate that the battery for the alarm is good. See the troubleshooting chart in Section 6.0 of this manual if the unit’s alarm does anything other than sound loudly for five seconds.
AirSep® Corporation
b. The compressor runs. Listen for the sound. c. Exhaust air flows out of the bottom of the unit.
2-1
NewLife
®
Elite Service Manual
3 Turn the flowmeter adjustment knob counterclockwise until it stops
(wide open).
The maximum flow should register 5.5 lpm (the 5.5 lpm line should appear to split the middle of the flowmeter ball) with no connections at the oxygen outlet. If not, refer to Section 5.8.1 to set the product regula­tor.
4 Perform an oxygen concentration test, as described in Section 2.4.
2.3 Alarm System
The NewLife Oxygen Concentrator is equipped with a battery-powered alarm system, which sounds a continuous and loud alarm when a power failure oc­curs. It sounds an intermittent alarm if the high or low pressure indicators are activated or if the optional oxygen monitor detects lower than therapeutic levels of oxygen concentration. The alarm remains on until you correct the alarm con­dition or you set the ON/OFF switch to the OFF position. Refer to Section 6.0 for a list of probable alarm causes.
d. OPTIONAL: The Oxygen Monitor’s amber light remains
on until the oxygen concentration reaches 85% ± 3% (approximately two minutes).
2.3.1 Battery Test
Each time the NewLife unit is turned on, a five-second audible alarm sounds to indicate the condition of the battery.
NOTE
The audio alarm must sound loudly for approxi­mately five seconds each time the unit is turned to ON to indicate the battery is in good condition.
2.3.2 Power Failure Alarm Test
To test the power failure alarm, take the following step:
Unplug the unit, and set the ON/ OFF switch to the ON position.
This should immediately activate the audio alarm. If it does not, refer to the troubleshooting chart in Section 6.0 of this manual.
2-2
AirSep® Corporation
NewLife
®
Elite Service Manual
2.4 Oxygen Concentration Test and Specification
To ensure that the unit’s output of oxygen is within specification, you must perform a test of oxygen concentration. Test the unit upon deliv­ery to a patient and at periodic intervals. Equipment Providers, based on their expertise and documentation, may establish and implement their own protocol to check oxygen concentration. The interval estab­lished may be longer or shorter than 90 days, which is AirSep’s default time period for providers who do not choose to establish their own pro­tocol.
1 If an oxygen humidifier bottle is used, remove it from the oxygen
outlet.
2 Connect a calibrated oxygen concentration analyzer to the oxygen
outlet.
3 Verify that the product flow rate delivered by the unit matches the
patient’s prescription and does not exceed the capacity of the unit.
4 Set the unit’s ON/OFF power switch to the ON position. (It takes
approximately five minutes for the oxygen concentration to stabi­lize.) Take oxygen concentration readings every 60 seconds until the analyzer shows that two consecutive readings are the same.
5 Disconnect the oxygen analyzer, and reconnect the humidifier bottle
(if used) and any other equipment/ accessories that may be re­quired.
6 Adjust the flowmeter adjustment knob to the prescribed level.
NOTE
Do not measure oxygen concentration output after the product stream passes through a humidi­fier bottle, or erroneous readings will result.
AirSep NewLife Oxygen Concentration Specifications
AirSep® Corporation
Liter Flow Specification In Spec
1–3 lpm 95% + 0.5/ - 3% 92% or higher
4 lpm 92% + 3.5/ - 3% 89% or higher 5 lpm 90% + 5.5/ - 3% 87% or higher
2-3
NewLife
®
Elite Service Manual
AirSep® Corporation
3.0 Patient Instructions
3.1 General Instructions
It is important that patients thoroughly understand how to operate the AirSep NewLife unit. This enables proper treatment as prescribed by a qualified, licensed physician. You must explain that the purpose of this therapy is to alleviate symp­toms. If patients experience any discomfort or the unit alarms, they must notify their Equipment Provider and/or physician immediately.
You, as the Equipment Provider, are responsible to see that each patient receives the Patient Manual. Locate the Patient Manual in the pocket on the back of each unit. Explain each step in the operation of the unit to the patient in reference to this manual.
3.2 Routine Maintenance by the Patient
To ensure accurate output and efficient operation of the unit, the patient must perform two simple routine maintenance tasks:
NewLife
®
Elite Service Manual
! Clean the air intake gross particle filter ! Check the alarm system battery
3.2.1 Cleaning the Air Intake Gross Particle Filter
NOTE
The patient must clean this filter weekly, as described below. The filter may require daily cleaning if the NewLife unit operates in a harsh environment such as a house heated by wood, kerosene, or oil, or one with excessive cigarette smoke.
1 Remove the dirty air intake gross particle filter from the back
of the NewLife unit, and install the clean filter stored in the pocket on the back of the unit.
AirSep® Corporation
2 Wash the dirty filter in warm soapy water, and rinse. 3 Use a soft absorbent towel to remove excess water. 4 Place the clean air intake gross particle filter in the pocket on
the back of the unit.
3-1
Loading...
+ 39 hidden pages