NewLife® Elite / NewLife
Oxygen Concentrator
Service Manual
®
0459
AirSep® Corporation
AirSep Corporation • 401 Creekside Drive • Buffalo, NY 14228-2085 USA
T elephone: (716) 691-0202 • 24-Hour Fax: (716) 691-4141
NewLife
®
Elite Service Manual
Expedited Product Warranty Check service is always at your fingertips with AirSep:
Type www.airsep.info/warranty_check/main.asp
OR
In the US or Canada,
dial 866-873-9277
AirSep® is a registered trademark of AirSep Corporation.
!
NewLife® Elite and NewLife® are registered trademarks of AirSep Corporation.
MN105-1 Rev. B 03/06
AirSep® Corporation
T able of Contents
Section 1.0
Introduction
1.1Equipment Provider Responsibility1-1
1. 2Important Notice and Symbol Explanations1-2
1.3NewLife
1.4Functional Specifications1-5
Section 2.0
Operation Check
and Oxygen
Concentration Test
2.1Description of Operation2-1
NewLife
®
Elite/NewLife® Product Warranty1-4
®
Elite Service Manual
Section 3.0
Patient Instructions
Section 4.0
Equipment Provider
Maintenance
2. 2Operation Check2-1
2.3Alarm System2-2
2.3.1Battery T est2-2
2.3.2Power Failure Alarm T est2-2
2. 4Oxygen Concentration T est and S pecification2-3
3.1General Instructions3-1
3.2Routine Maintenance by the Patient3-1
3.2.1Cleaning the Air Intake
Gross Particle Filter3- 1
3.2.2Checking the Alarm System Battery3- 2
4.1Routine Maintenance4-1
AirSep® Corporation
4.1.1Air Intake Gross Particle Filter/GPF4- 1
4.1.2Product Filter Replacement4-1
4.1.3Optional Filter Replacement4-2
4.1.4Battery Replacement4-2
4.1.5Recording Maintenance4-3
4.2Cleaning and Infection Control4-3
4.2.1Preparing for New Patient Use4-4
i – table of contents
®
NewLife
Elite Service Manual
Section 5.0
Service
5.1Components5-1
5.2Cabinet Removal5 -1
5.2.1Removing Side Panel(s)5-1
5.2.2Removing Back Panel5-1
5.2.3Removing Lower Front Cover5-1
5.2.4Removing Control Panel5-1
5.2.5Superstructure5-1
5.2.6Caster Replacement5-2
5.3Compressor5-2
5.3.1Compressor Replacement5-3
5.3.2Capacitor Replacement5-4
5.4Solenoid Valves5-4
5.4.1Feed or Waste V alve Rebuilding5-4
5.4.2Feed or Waste V alve Reassembly5-5
5.4.3Solenoid V alve Coil Replacement5-5
5.5Sieve Bed Replacement5-6
5.5.1Sieve Bed Removal5-6
5.5.2Sieve Bed Installation5-7
5.6Cabinet Fan Replacement5-7
5.7Circuit Board Replacement5-8
5.7.1Circuit Board Removal5-8
5.7.2Circuit Board Installation5-9
5. 8Product Regulator Check and Setting5-10
5.8.1Setting Product Regulator for
Normal Operation5-10
5.8.2Product Regulator Lockout5-10
5.8.3Back Pressure Correction at 5 lpm5-11
5.8.4Product Regulator Cleaning or
Rebuilding5-12
5. 9High or Low Pressure Switch Replacement
(In Equipped Units)5-13
ii – table of contents
5.10Circuit Breaker Replacement5-14
5.10.1Circuit Breaker Removal5-14
5.10.2Circuit Breaker Installation5-14
5.11ON/OFF Power Switch Replacement5-14
5.11.1ON/OFF Power Switch Removal5-14
AirSep® Corporation
Section 6.0
Troubleshooting
NewLife
5.11.2ON/OFF Power Switch Installation5-15
5.12Buzzer Replacement5-15
5.13Hour Meter Replacement5-15
5.14Flowmeter Replacement5-16
5.14.1Flowmeter Removal5-16
5.14.2Flowmeter Installation5-16
5.15Power Cord Replacement5-16
6.1Operating Pressure T est6-1
6.1.1High Operating Pressure6-1
6.1.2Low Operating Pressure6-2
6.2General Troubleshooting6-2
6.3Troubleshooting Chart6-4
®
Elite Service Manual
Appendix
Exploded
Drawings
6. 4T ool Kit and Pressure T est Gauge6-7
Figure 1: Control Panel AssemblyA-1
Figure 2: Main Structure AssemblyA-2
Figure 3: Base and Cabinet ComponentsA-3
Figure 4: Test Block AssemblyA-4
Figure 5: V alve Block AssemblyA-5
Figure 6: Compressor AssemblyA-6
Figure 7: Adsorption Bed AssemblyA-7
AirSep® Corporation
iii – table of contents
NewLife
®
Elite Service Manual
AirSep® Corporation
1.0Introduction
1.1Equipment Provider Responsibility
All Equipment Providers of the NewLife® Elite/NewLife® Oxygen Concentrator
must assume responsibilities for handling, operational check-out, patient instruction, and maintenance. These responsibilities are outlined below and throughout this manual.
NewLife
®
Elite Service Manual
WARNING
NewLife units must not be used for or with any lifesupporting applications. Geriatric, pediatric, or any other
patients unable to communicate discomfort while using
this machine may require additional monitoring. Advise
patients to immediately notify their Equipment Providers
and/or physicians in case of an alarm or any discomfort.
As an Equipment Provider, you must do all of the following:
!Inspect the condition of each NewLife unit immediately upon delivery
to your business location. Note any sign of damage, external or internal, on the delivery receipt, and report it directly to both the freight
company and AirSep Corporation immediately.
!Check the operation of each NewLife before delivery to a patient.
Always operate the unit for a reasonable length of time and check
that the oxygen concentration level is within specifications as referred
to in Section 2.4. (Test the battery and power disconnect alarm as
described in Section 2.3 of this manual.)
!Deliver NewLife units only to patients authorized by a physician’s
prescription. The NewLife must not be used as a life-supporting
device. A backup supply of oxygen must be available.
!Instruct patients how to use the NewLife in conjunction with the
Patient Manual.
!Instruct patients to notify their physicians and/or Equipment Provid-
ers if they experience any signs of discomfort.
!Instruct each patient how to perform routine maintenance of the air
intake gross particle filter and how to check the alarm system battery.
(Refer to Section 3.2.)
!Be available to service each patient at any time.
!Maintain the NewLife in accordance with Section 4.0.
!Repair components and replace parts only as outlined in this manual.
Use only AirSep parts for replacement in NewLife Oxygen Concentrators.
AirSep® Corporation
!Refer to the NewLife Product Warranty if parts replacement is re-
quired within the warranty period.
1-1
NewLife
®
Elite Service Manual
1.2Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION,
and NOTE messages. They identify safety guidelines or other important information as follows:
WARNING:Describes a hazard or unsafe practice that can result in
severe bodily injury or death.
CAUTION:Describes a hazard or unsafe practice that can result in
minor bodily injury or property damage.
NOTE:Provides information important enough to emphasize or
repeat.
1-2
AirSep® Corporation
NewLife
®
Elite Service Manual
The following harmonized symbols (pictograms), used for nonEnglish language countries, will be located on the exterior of the
NewLife unit:
Read the Patient Instruction Manual.
No smoking within five feet of this device, oxygencarrying tubing, or accessories.
No open flames within five feet of this device, oxygencarrying tubing, or accessories.
No oil, grease or petroleum-based products on or near
the unit.
Oxygen outlet
Oxygen alarm
AirSep® Corporation
Nebulizer option
1-3
NewLife
®
Elite Service Manual
1.3NewLife Product Warranty
AirSep Corporation warrants the NewLife Elite/NewLife Oxygen Concentrator
to be free from defect in parts and workmanship for one, three, or five years (as
specified on the original invoice provided) from the date of delivery to the original purchaser, under normal use and operation. AirSep Corporation’s obligations under this warranty are limited to the repair or replacement of any such
item of equipment (or part thereof) shown to be defective or, at AirSep
Corporation’s option, to refunding the purchase price of any such defective item
of equipment.
Each item of equipment for which a warranty claim is asserted shall, at the request of AirSep Corporation, be returned on a prepaid basis with proof of purchase date to the AirSep factory at the expense of the purchaser. The purchaser
will be responsible for return freight charges. Replacement parts shall be warranted as stated above for the unexpired portion of the original (one-, three-, or
five-) year parts warranty (as specified on the original invoice provided). This
warranty does not extend to any item or part subjected to misuse, accident, improper maintenance, or application, or which has been repaired or altered outside of the AirSep Corporation Factory without the express prior written authorization of AirSep Corporation.
THE FOREGOING W ARRANTY IS IN LIEU OF ANY OTHER W ARRANTY , EXPRESSED OR
IMPLIED, IN F ACT OR IN LA W , INCLUDING WITHOUT LIMIT ATION THE WARRANTY OF
MERCHANT ABILITY OR THE WARRANTY OF FITNESS FOR P AR TICULAR PURPOSE. IT IS
EXPRESSL Y UNDERSTOOD THAT PURCHASER’S SOLE AND EXCLUSIVE REMEDY FOR
DEFECT IN P AR TS IS LIMITED TO ENFORCEMENT OF AIRSEP CORPORA TION’S OBLIGATION AS SET FORTH ABOVE, AND AIRSEP CORPORA TION SHALL NOT BE LIABLE TO
PURCHASER OR OTHERS FOR LOSS OF USE OF THE EQUIPMENT OR FOR OTHER SPECIAL,
INDIRECT , INCIDENTAL, OR CONSEQUEN-TIAL DAMAGES.
1-4
AirSep® Corporation
1.4Functional Specifications
Dimensions:28.5 in. high, 15.7 in. wide, 14.5 in. deep
Concentration:1–3 liters per minute at 95% + 0.5/ - 3%
4 liters per minute at 92% + 3.5/ - 3%
5 liters per minute at 90% + 5.5/ - 3%
Response Time:Allow 5 minutes to attain maximum oxygen concentra-
tion.
NewLife
®
Elite Service Manual
Positioning:Operate the unit in an upright position, maintaining at
least six inches of open space on all sides for ventilation.
0459
AirSep® Corporation
1-5
NewLife
®
Elite Service Manual
AirSep® Corporation
NewLife
®
Elite Service Manual
2.0Operation Check and Oxygen Concentration Test
2.1Description of Operation
Air enters the NewLife Oxygen Concentrator through an external air intake gross
particulate filter. This filtered air enters the compressor via a suction resonator,
which quiets the compressor’s suction sound. Pressurized air then exits the compressor and passes through a heat exchanger. The heat exchanger reduces the
temperature of the compressed air. Next, a two-way solenoid feed valve directs
the air into one of two sieve beds that contain molecular sieve. The unique property of molecular sieve enables it to physically attract (adsorb) nitrogen when air
passes through this material, thus producing high concentration oxygen.
There are two sieve beds: while one produces high concentration oxygen, the
other is purged of the nitrogen it adsorbed (collected) while it made oxygen. Each
adsorber produces oxygen for approximately eight seconds and delivers it to the
product tank. Oxygen exits the product tank through a pressure regulator, product filter, flow control valve, flowmeter, and finally exits the unit. The flow control valve, which is part of the flowmeter, controls the amount of oxygen delivered to the patient. The NewLife unit delivers up to 95.5% pure oxygen at flow
rates from 1–5 lpm.
NOTE: NewLife Elite OxiSilent™ option incorporates an additional felt intake
filter prior to air entering the compressor.
2.2Operation Check
AirSep tests every NewLife Oxygen Concentrator thoroughly after manufacture.
You must perform the following test to ensure that no damage occurred in shipping or handling.
1Open and inspect all cartons (that contain units) upon delivery. Un-
pack the unit and remove it from the carton. Inspect the unit itself for
damage. If the exterior of a unit’s carton is damaged, or the unit itself is
damaged, note it on the freight bill signed by the driver.
2Plug in the power cord of the unit, and set the ON/OFF switch to the
ON position. Check to see that the following occurs:
a.A continuous alarm loudly sounds for approximately five
seconds to indicate that the battery for the alarm is good.
See the troubleshooting chart in Section 6.0 of this
manual if the unit’s alarm does anything other than
sound loudly for five seconds.
AirSep® Corporation
b.The compressor runs. Listen for the sound.
c.Exhaust air flows out of the bottom of the unit.
2-1
NewLife
®
Elite Service Manual
3Turn the flowmeter adjustment knob counterclockwise until it stops
(wide open).
The maximum flow should register 5.5 lpm (the 5.5 lpm line should
appear to split the middle of the flowmeter ball) with no connections at
the oxygen outlet. If not, refer to Section 5.8.1 to set the product regulator.
4 Perform an oxygen concentration test, as described in Section 2.4.
2.3Alarm System
The NewLife Oxygen Concentrator is equipped with a battery-powered alarm
system, which sounds a continuous and loud alarm when a power failure occurs. It sounds an intermittent alarm if the high or low pressure indicators are
activated or if the optional oxygen monitor detects lower than therapeutic levels
of oxygen concentration. The alarm remains on until you correct the alarm condition or you set the ON/OFF switch to the OFF position. Refer to Section 6.0 for
a list of probable alarm causes.
d.OPTIONAL: The Oxygen Monitor’s amber light remains
on until the oxygen concentration reaches 85% ± 3%
(approximately two minutes).
2.3.1Battery Test
Each time the NewLife unit is turned on, a five-second audible alarm
sounds to indicate the condition of the battery.
NOTE
The audio alarm must sound loudly for approximately five seconds each time the unit is turned to
ON to indicate the battery is in good condition.
2.3.2Power Failure Alarm Test
To test the power failure alarm, take the following step:
Unplug the unit, and set the ON/ OFF switch to the ON
position.
This should immediately activate the audio alarm. If it does not, refer to
the troubleshooting chart in Section 6.0 of this manual.
2-2
AirSep® Corporation
NewLife
®
Elite Service Manual
2.4Oxygen Concentration Test and Specification
To ensure that the unit’s output of oxygen is within specification, you
must perform a test of oxygen concentration. Test the unit upon delivery to a patient and at periodic intervals. Equipment Providers, based
on their expertise and documentation, may establish and implement
their own protocol to check oxygen concentration. The interval established may be longer or shorter than 90 days, which is AirSep’s default
time period for providers who do not choose to establish their own protocol.
1If an oxygen humidifier bottle is used, remove it from the oxygen
outlet.
2Connect a calibrated oxygen concentration analyzer to the oxygen
outlet.
3Verify that the product flow rate delivered by the unit matches the
patient’s prescription and does not exceed the capacity of the unit.
4Set the unit’s ON/OFF power switch to the ON position. (It takes
approximately five minutes for the oxygen concentration to stabilize.) Take oxygen concentration readings every 60 seconds until
the analyzer shows that two consecutive readings are the same.
5Disconnect the oxygen analyzer, and reconnect the humidifier bottle
(if used) and any other equipment/ accessories that may be required.
6Adjust the flowmeter adjustment knob to the prescribed level.
NOTE
Do not measure oxygen concentration output
after the product stream passes through a humidifier bottle, or erroneous readings will result.
4 lpm92% + 3.5/ - 3%89% or higher
5 lpm90% + 5.5/ - 3%87% or higher
2-3
NewLife
®
Elite Service Manual
AirSep® Corporation
3.0Patient Instructions
3.1General Instructions
It is important that patients thoroughly understand how to operate the AirSep
NewLife unit. This enables proper treatment as prescribed by a qualified, licensed
physician. You must explain that the purpose of this therapy is to alleviate symptoms. If patients experience any discomfort or the unit alarms, they must notify
their Equipment Provider and/or physician immediately.
You, as the Equipment Provider, are responsible to see that each patient receives
the Patient Manual. Locate the Patient Manual in the pocket on the back of each
unit. Explain each step in the operation of the unit to the patient in reference to
this manual.
3.2Routine Maintenance by the Patient
To ensure accurate output and efficient operation of the unit, the patient must
perform two simple routine maintenance tasks:
NewLife
®
Elite Service Manual
!Clean the air intake gross particle filter
!Check the alarm system battery
3.2.1Cleaning the Air Intake Gross Particle Filter
NOTE
The patient must clean this filter weekly, as
described below. The filter may require daily
cleaning if the NewLife unit operates in a harsh
environment such as a house heated by wood,
kerosene, or oil, or one with excessive cigarette
smoke.
1Remove the dirty air intake gross particle filter from the back
of the NewLife unit, and install the clean filter stored in the
pocket on the back of the unit.
AirSep® Corporation
2Wash the dirty filter in warm soapy water, and rinse.
3Use a soft absorbent towel to remove excess water.
4Place the clean air intake gross particle filter in the pocket on
the back of the unit.
3-1
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.