Installing and Using the 
Focus Data Interface Software on Your PC 
Focus Data Interface Software Installation 
1. If you are installing the software from the included CD insert the CD into the CD 
or DVD drive of your computer. 
2. If the installation does not automatically begin, open My Computer. Within the My 
Computer window, click on the CD or DVD drive that contains the installation 
files.  
3. Click “setup.exe”. 
4. Click “Run”. (Fig. 1) 
Fig. 1 
MN187-1                                                                                                 Rev - 
Page 1 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 2

5. If you do not have Microsoft .NET Framework 4 (x86 and x 64) installed on your 
computer it will be installed at this time. This may take a few minutes. Click 
“Accept”. (Fig. 2) 
Fig. 2 
The Focus / FreeStyle 5 Data Interface Installation Program will now begin. 
6. Click “Next”. (Fig. 3) 
Fig. 3 
MN187-1                                                                                                 Rev - 
Page 2 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 3

7. Click “Next” if this is where you would like the software installed. If you would 
like it installed in a different location click “Browse” and then select the location. 
(Fig. 4) 
Fig. 4 
8. Click “Next” to confirm installation. (Fig. 5) 
Fig. 5 
MN187-1                                                                                                 Rev - 
Page 3 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 4

9. After the Focus Data Interface program has installed successfully, the drivers for 
the Focus interface cable (MI347) will be installed. To do this the cable must be 
plugged into an open USB port. (Fig. 6) 
Fig. 6 
10. When you plug the Focus interface cable (MI347) into your computer a message 
will pop-up on your taskbar. (Fig. 7) 
Fig. 7 
11. When the Focus interface cable (MI347) is finished installing you will get another 
pop-up on your taskbar. (Fig. 8) 
Fig. 8 
MN187-1                                                                                                 Rev - 
Page 4 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 5

12. Now click “Close” on the Installation Complete screen and you are ready to run the 
Focus Data Interface software. (Fig. 9) 
Fig. 9 
MN187-1                                                                                                 Rev - 
Page 5 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 6

Reading Focus hours using the Focus Data Interface Software 
1. Insert the Focus interface cable (MI347) into an open USB port on your PC. 
2. Click the Focus Data Interface icon on your desktop to start the program. (Fig. 10). 
The Focus Data Interface program box will appear (see figure 11). 
Fig. 10 
Fig. 11 
3. Unscrew the two screws located on the bottom of the Focus unit (Fig.12). 
Carefully remove covers and place Focus on its back (Fig. 13). 
                    Fig. 12 
2 screws 
     Fig. 13 
MN187-1                                                                                                 Rev - 
Page 6 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 7

4. Connect the other end of the MI347 cable to J9 on the Focus circuit board. The 
black wire goes to the top pin. (Fig. 14) 
Fig. 14 
5. Put the unit in test mode as follows: 
Note: Communication is only possible when the Focus unit is in test mode. 
 5.1. Connect the Focus Oxygen Concentrator to an AirSep-supplied AC power 
supply (AirSep Part No. PW019-1). 
Focus units 
power 
connector 
  Fig. 15 
Focus units 
Power Inlet 
   Fig. 16 
5.2.  To activate the unit’s test mode:  
5.2.1 Turn the unit on. 
            5.2.2 After the 10 second start up time, turn the unit off & on consecutively 
3 times (within 4 seconds). 
5.2.3 The unit will then start to auto pulse and is now in test mode. 
  5.3.  Verify Focus is pulsing by observing the intermittent green LED. 
When the unit pulses, you should be able to hear the pulse of oxygen and   
see the LED turn off momentarily. 
    If unit is not pulsing, turn unit off and repeat steps 5.2.1 and 5.2.2. 
6. The Focus Data Interface program screen will display hours in the Unit Run Time 
box and the Compressor Run Time box within two minutes. (Fig. 17) 
MN187-1                                                                                                 Rev - 
Page 7 of 8  
PDF created with pdfFactory Pro trial version www.pdffactory.com
 
Page 8

Fig. 17 
7. Pressing the “Clear “ button will clear the compressor run time. A warning box 
will appear asking if you are sure you want to clear the compressor time. If this is 
what you want to do click “Yes”, if not click “No”. 
8. Figure 18 shows the Focus Data Interface program screen after pressing “Yes” to 
clear compressor time. 
Fig. 18 
   9. To exit program, click “Exit” button 
MN187-1                                                                                                 Rev - 
Page 8 of 8  