AirSep MN110-1 User Manual

Page 1
LifeStyle™ Portable Oxygen Concentrator Service Manual
AirSep Corporation • 401 Creekside Drive • Buffalo, NY 14228-2085 USA
T elephone: 716. 691.0202 • 24-Hour Fax: 716.691.4141 • www .airsep.com
Page 2
Expedited Product Warranty Check service is always at your fingertips with AirSep:
Type www.airsep.info/warranty_check/main.asp
OR
In the US or Canada,
dial 866-873-9277
AirSep® is a registered trademark of AirSep Corporation.
LifeStyle™ is a trademark of AirSep Corporation.
MN110-1 Rev A 07/07
Page 3
T able of Contents
Section 1.0
Introduction
1.1 Equipment Provider Responsibility 1-1
1.2 Important Notice and Symbol Explanations 1-2
1.3 Definitions of Symbols 1-3
1.4 LifeStyle Product Limited W arranty 1-4
1.5 LifeStyle S pecifications 1-5
1.6 LifeStyle Classifications 1-6
Section 2.0
Operation Check and Oxygen Concentration Test
Section 3.0
Patient Instructions
2.1 Description of Operation 2-1
2.2 Operation Check 2-1
2.3 Alarm System 2-2
2.3.1 Start-up 2-2
2.3.2 Low Battery 2-2
2.3.3 Discharged Battery 2-2
2.3.4 Battery Charging 2-2
2.3.5 Battery Fully Charged 2-3
2.3.6 Cannula Disconnect 2-3
2.3.7 Performance Warning 2-3
2.3.8 LifeStyle Capacity Exceeded 2-3
2.3.9 General Malfunction 2-4
2.4 Oxygen Concentration T est and S pecification 2-4
3.1 General Instructions 3-1
3.2 Routine Maintenance by the Patient 3-1
3.2.1 Cleaning the Air Intake Gross Particle Filter 3-1
3.2.2 Checking the Hour Meter 3-2
3.3 Usage at Night or with Extended T ubing 3-2
AirSep Corporation i – table of contents
Page 4
Section 4.0
Equipment Provider Maintenance
Section 5.0
Service
4.1 Routine Maintenance 4-1
4.1.1 Air Intake / Gross Particle Filter 4-1
4.1.2 Product Filter Replacement 4-1
4.1.3 Compressor Replacement 4-2
4.1.4 Recording Maintenance 4-2
4.2 Infection Control 4-2
4.3 Cleaning and Preparing for New Patient Use 4-2
5.1 Components 5-1
5.2 Housing Removal 5- 1
5.2.1 Removing Bottom Housing 5 -1
Section 6.0
Troubleshooting
Appendix
Exploded Drawings
5.2.2 Removing T op Housing 5-2
5. 3 Compressor Replacement 5-2
5.3.1 Compressor Assembly Removal 5-2
5.3.2 Compressor Assembly Installation 5-3
6.1 General Troubleshooting 6-1
6.2 Troubleshooting Chart 6-2
Figure 1: LifeStyle — Control Panel A- 1 Figure 2: LifeStyle — Internal, T op View A- 2 Figure 3: LifeS tyle — Internal, Bottom View A- 3
ii – table of contents AirSep Corporation
Page 5
1.0 Introduction
1.1 Equipment Provider Responsibility
All Equipment Providers of the LifeStyle™ Portable Oxygen Concentrator must assume responsibilities for handling, operational check-out, patient instruction, and maintenance. These responsibilities are outlined below and throughout this manual.
As an Equipment Provider, you must do all of the following:
WARNING
This unit is not a life support device. Geriatric, pediatric, and other patients unable to communicate discomfort while using this oxygen concentrator may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with monitoring the alarms.Instruct patients to consult their physician immediately if they experience any signs of discomfort.
Inspect the condition of each LifeStyle unit immediately upon delivery to your
business location. Note any sign of damage, external or internal, on the delivery receipt, and report it directly to both the freight company and AirSep Corporation immediately .
Check the operation of each LifeStyle unit before delivery to a patient.
Always operate the unit for a reasonable length of time, and check that the oxygen concentration level is within specifications, as described in Section 2.4 of this manual.
Charge the battery initially for 12 hours. See “Battery Charging” in Section
2.3.4 of this manual.
Deliver LifeS tyle units only to patients authorized by a physician’s prescrip-
tion.
Instruct patients how to use LifeStyle in conjunction with the Patient Manual.
1-1
Page 6
Instruct patients to notify their physicians and/or Equipment Providers if they
experience any signs of discomfort.
Instruct each patient how to perform routine maintenance of the air intake
gross particle filter and how to monitor the hour meter for the compressor. (Refer to Section 3.2.)
Be available to service each patient at any time.
Maintain LifeStyle in accordance with Section 4.0.
Repair components and replace parts only as outlined in this manual. Use
only AirSep parts for replacement in LifeS tyle Oxygen Concentrators.
Refer to the “LifeS tyle Product Limited W arranty ,” Section 1.4 of this manual,
if replacement of parts is required within the warranty period.
1.2 Import ant Notice and Symbol Explanations
As you review both the LifeStyle Patient Manual and this LifeS tyle Service Manual, pay special attention to the W ARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important information as follows:
WARNING: Describes a hazard or unsafe practice that can result in severe
bodily injury or death and/or property damage.
CAUTION: Describes a hazard or unsafe practice that can result in minor
bodily injury or property damage.
NOTE: Provides information important enough to emphasize or repeat.
1-2
Page 7
1.3 Definitions of Symbols
The definitions of symbols used with the LifeStyle Portable Oxygen Concentrator follow .
Symbol Description
Refer to Patient Manual instructions
No oil
No smoking
Type BF equipment
Class II
CE Mark
Safety agency for CAN/CSA C22.2 No. 601.1.M90 for medical electrical equipment.
FAA SF AR106
Proper disposal of waste of electrical and electronic equipment required.
1-3
Page 8
1.4 LifeStyle Product Limited W arranty
AirSep Corporation warrants the LifeStyle Oxygen Concentrator to be free from defect in parts for two years (or as specified on the original invoice provided) from the date of delivery to the original purchaser, under normal use and operation. The battery is warrantied for 90 days. AirSep Corporation’s obligations under this warranty are limited to the repair or replacement of any such item of equipment (or part thereof) shown to be defective or, at AirSep Corporation’ s option, to refund the purchase price of any such defective item of equipment.
Each item of equipment for which a warranty claim is asserted shall, at the request of AirSep Corporation, be returned on a prepaid basis with proof of purchase date to the AirSep factory at the expense of the purchaser. The purchaser will be responsible for return freight charges. Replacement parts shall be warranted as stated above for the unexpired portion of the original two-year parts warranty (as specified on the original invoice provided). This warranty does not extend to any item or part subjected to misuse, accident, improper maintenance, or application, or which has been repaired or altered outside of the AirSep Corporation factory without the express prior written authorization of AirSep Corporation.
THE FOREGOING W ARRANTY IS IN LIEU OF ANY OTHER WAR­RANTY, EXPRESSED OR IMPLIED, IN FACT OR IN LA W , INCLUDING WITHOUT LIMIT A TION THE W ARRANTY OF MERCHANTABILITY OR THE W ARRANTY OF FITNESS FOR P AR TICULAR PURPOSE. IT IS EXPRESSL Y UNDERSTOOD THAT PURCHASER’S SOLE AND EXCLUSIVE REMEDY FOR DEFECT IN P AR TS IS LIMITED T O ENFORCEMENT OF AIRSEP CORPORA TION’S OBLIGA TION AS SET FORTH ABOVE, AND AIRSEP CORPORA TION SHALL NOT BE LIABLE TO PURCHASER OR OTHERS FOR LOSS OF USE OF THE EQUIPMENT OR FOR OTHER SPECIAL, INDIRECT , INCIDENT AL, OR CONSEQUENTIAL DAMAGES.
1-4
Page 9
1.5 LifeStyle Specifications
Oxygen Concentration:* Set at 1-5 LPM: equivalent to a continuous flow of 90%
oxygen ± 3%
Dimensions: 5.5 in. high x 7.25 in. wide x 16.31 in. long
(13.9 cm high x 18.4 cm wide x 41.4 cm long) Weight: 9.75 lb (4.4 kg) Power: LifeStyle unit - 13 VDC, 42 watts
Power supply - 100 – 250 V AC, 50/60 Hz, 2.0 amp s max
Battery pack - 12 VDC, nominal Battery duration: 50 minutes nominal for fully-charged battery Battery charge time: 2 — 2½ hours Warm-up time: 3 minutes Audible alarms and Start-up – audible and visual (
visual indicators: Pulse flow – visual (GREEN pulse light)
Battery condition – audible and visual (YELLOW indicator,
warning, and alarm lights)
Cannula disconnect - audible and visual (RED alarm light)
System overdraw – audible and visual (RED warning and
alarm lights)
High and low pressure – audible and visual (RED alarm light) Sound: Less than 55 decibels T emperature range Operational temperature: 41oF to 104oF (5oC to 40oC)
intended for use: Storage temperature: -4oF to 140oF (-20oC to 60oC)
* Based on an atmospheric pressure of 14.7 psia (101 kPa) at 70oF (21oC)
GREEN indicator light)
1-5
Page 10
1.6 LifeStyle Classifications
Type of protection against electric Protection from electric shock is achieved by shock: double insulation.
Class II
Degree of protection against Equipment providing a particular degree of electric shock: protection against electric shock regarding Type BF 1) allowable leakage current
2) reliability of protective earth connection (if present). Not intended for direct cardiac application.
Independent testing for Tested by UltraTech Group of Labs to be in Medical Electrical Equipment compliance with IEC 60601-1:1988 + Standard A1:1991, A2:1995
Protection against potential Tested by UltraTech Group of Labs to be in electromagnetic or other compliance with IEC 601-1-2 (EMC) interference between the equipment and other devices:
Method of cleaning and Please refer to “Provider Maintenance,” infection control allowed: Section 4.0 of this LifeStyle Service Manual.
Degree of safety of Equipment not suitable for such application. application in the presence of flammable anesthetic gases:
Mode of operation: Continuous duty
FAA SFAR 106: RTCA DO160D, Section 21,
Category M-Compliant
1-6
Page 11
2.0 Operation Check and Oxygen Concentration Test
2.1 Description of Operation
Air enters the LifeStyle Oxygen Concentrator through an external air intake / gross particle filter . This filtered air enters the compressor via a suction resonator that quiets the compressor’s suction sound. Pressurized air then exits the compressor and passes through a heat exchanger. The heat exchanger reduces the temperature of the com­pressed air. Next, a two-way solenoid feed valve directs the air into one of two sieve beds that contain molecular sieve. The unique property of molecular sieve enables it to physically attract (adsorb) nitrogen when air passes through this material, thereby separating the nitrogen from the oxygen in ambient air.
There are two sieve beds: While one produces high-concentration oxygen, the other is purged of the nitrogen it adsorbed (collected) while it made oxygen. Each adsorber (bed) produces oxygen for approximately eight seconds and delivers it through a product tank to the regulator. From the regulator , oxygen passes through the product filter and then to the outlet fitting.
In addition to the molecular sieve beds that allow oxygen to be separated from air, LifeStyle contains an Oxygen Conserving Device (OCD) valve that controls delivery of high-concentration oxygen to the patient at the point in the breathing cycle when the patient starts to inhale.
LifeStyle delivers a gas consisting of 90% (±3%) oxygen at pulse flow rates from 1–5 lpm.
2.2 Operation Check
AirSep tests every LifeStyle Oxygen Concentrator thoroughly after manufacture. Y ou must perform the following test to ensure that no damage occurred in shipping or handling.
1. Open and inspect all cartons upon delivery . Unpack the unit and remove it
from the carton. Inspect the unit itself for damage. If the exterior of a unit’ s carton is damaged, or if the unit itself is damaged, note this on the freight bill signed by the driver .
2. Connect the power cord of the unit, and set the ON/OFF switch to the ON
position. Check to see that the following sequence occurs:
a. A brief, continuous alarm sounds to indicate that LifeStyle has
been turned on.
b. The compressor runs. c. Exhaust air flows from the unit.
3. Perform an oxygen concentration test, as described in Section 2.4.
2-1
Page 12
2.3 Alarm System
An audible alarm sounds if LifeStyle has a low battery , senses no inspiration, or if performance outside of specifications is detected. If an alarm sounds, open the cover, and observe which lights are on. The visual and audible alarm conditions are explained in detail below .
2.3.1 Start-Up
A five-second, continuous, audible alarm sounds to indicate start-up. LifeStyle begins to operate when the alarm stops.
2.3.2 Low Battery
As the battery power approaches a low level, a one-second beep is emitted every five seconds, and the yellow BA TTER Y light flashes for one second every five seconds. When this occurs, the patient is instructed to replace the battery , connect LifeStyle to a DC auto adapter power outlet or an AC power outlet, or change to another source of oxygen within five minutes. When LifeStyle is connected to a DC auto adapter outlet or to an AC outlet, power from the automobile battery or from the AC outlet operates LifeStyle and recharges LifeStyle’ s battery simultaneously .
2.3.3 Discharged Battery
LifeStyle does not produce oxygen when the battery power falls below a certain level. When this occurs, a half-second beep sounds every half second, and the
YELLOW battery light flashes. In this case, the patient is instructed to replace the
battery , connect LifeStyle to a DC auto adapter outlet or an AC outlet, or change to another source of oxygen immediately . When LifeStyle is connected to a DC auto adapter or an AC outlet, power from the automobile battery or from the AC outlet operates LifeStyle and recharges LifeS tyle’s battery simultaneously .
2.3.4 Battery Charging
Charge the battery for the first time for a minimum of 12 hours. The BATTERY light may illuminate a steady yellow significantly before 12 hours; however, the battery should be left in LifeStyle with LifeStyle connected to the DC auto adapter or AC adapter for the full 12 hours. When LifeStyle is plugged into an automobile’ s DC outlet or an AC outlet and is turned OFF, the yellow BATTERY light flashes to indicate that the battery is being charged. When LifeStyle is turned ON and plugged into the DC auto adapter or AC adapter, the battery still charges; however , the yellow BATTERY light is not on.
2-2
Page 13
2.3.5 Battery Fully Charged
When LifeStyle is connected to an AC or DC outlet and is in the OFF setting, the yellow BATTERY light illuminates and displays continuously to indicate that the battery is fully charged. However, when char ging the battery for the first time, the battery should be left in the charger for 12 hours even if the yellow BATTERY light is on continuously .
2.3.6 Cannula Disconnect
When LifeStyle is connected to a power source and does not sense breathing within a predetermined time period, a constant alarm sounds, and the red ALARM light illuminates. If this occurs, the patient is instructed to check the connection from the cannula to the LifeStyle unit, make sure that the nasal cannula is positioned properly on the face, and ensure that breathing is through the nose. If the alarm continues to sound, the patient should change to an alternate source of oxygen and contact the Equipment Provider.
2.3.7 Performance Warning
The performance of LifeStyle can be affected by two conditions. The first is if the user’s breathing rate causes the LifeStyle unit’ s capacity to be exceeded. The LifeStyle unit supplies a fixed amount of high-concentration oxygen. The amount of oxygen that LifeStyle supplies is dependent on the flow control switch setting and the breathing rate. If the breathing rate is too fast for the flow selector setting, the capacity of LifeStyle can be exceeded.
The second condition that can result in decreased performance is a general malfunction. LifeStyle monitors its performance and alerts the patient if this occurs.
If either of these events occurs, a half-second alarm sounds every five seconds. If breathing is too fast for the flow selector setting and LifeStyle capacity is exceeded, the red ALARM light flashes. If a general malfunction occurs, a steady red
ALARM light appears. During this time, LifeStyle is still producing high-purity
oxygen and may continue to be used. The patient should reduce activity or consider locating another source of oxygen.
2.3.8 LifeStyle Capacity Exceeded
When the breathing rate has caused the capacity of LifeStyle to be exceeded, a half-second alarm sounds every half second, and the red alarm light flashes. When this occurs, oxygen concentration is outside specifications. The patient should change to another source of oxygen and contact the Equipment Provider.
2-3
Page 14
2.3.9 General Malfunction
If a general malfunction of the LifeStyle unit occurs, a half-second alarm sounds every half second, and the red ALARM light turns on. When this occurs, oxygen concentration is outside specifications. The patient should change to another source of oxygen and contact the Equipment Provider .
2.4 Oxygen Concentration T est and S pecification
T o ensure that LifeStyle’ s oxygen output is within specification, you must perform a test of oxygen concentration. T est the unit upon delivery to a patient and at periodic inter­vals. Equipment Providers, based upon their own expertise and documentation, may establish and implement their own protocol to check oxygen concentration. The interval established may be longer or shorter than 90 days, AirSep’ s default time period for providers who do not choose to establish their own protocol.
1. Connect LifeStyle to AC electrical power using only the AirSep-supplied AC adapter (AirSep Part No.PW005) to a 100-250 volt, 50/60 Hz outlet.
(Note: Do not use the battery [AirSep Part No. BT007-1] to operate the unit when measuring oxygen concentration.)
2. Turn on LifeStyle, and adjust the flow selector to any of 1 through 5 lpm settings.
3. Remove the gross particle filter.
4. Activate the test mode by depressing the switch located in the hole labeled “T ,” next to the hour meter. The unit should produce a pulse dose of oxygen equivalent to a continuous flow of 4 lpm at a rate of 18 breaths-per-minute while in the test mode.
5. When the unit supplies a pulse dose of oxygen, the green PULSE light is on, and you will be able to hear the pulse of oxygen. Verify that the pulse light is turning on and that you can hear the pulse dose.
6. Allow LifeStyle to run for five minutes before recording the oxygen concen­tration. Connect a calibrated T eledyne TED or equivalent to the oxygen
outlet on the unit. Turn on the analyzer , and record the concentration. The concentration specification in test mode is 90% ±3%.
2-4
Page 15
3.0 Patient Instructions
3.1 General Instructions
It is important that patients thoroughly understand how to operate AirSep’ s LifeStyle unit. This enables proper treatment as prescribed by a qualified, licensed physician. You must explain that the purpose of this therapy is to alleviate symptoms. If patients experience any discomfort or if the unit alarms, they must notify their Equipment Pro­vider and/or physician immediately .
Y ou, as the Equipment Provider , are responsible to ensure that each patient receives the LifeStyle Patient Manual. Explain each step in the operation of the unit to the patient with reference to the patient manual.
3.2 Routine Maintenance by the Patient
T o ensure accurate output and efficient operation of the unit, the patient must perform two simple, routine maintenance tasks:
Clean the air intake / gross particle filter
Check the number of hours displayed on the hour meter .
3.2.1 Cleaning the Air Intake / Gross Particle Filter
NOTE
The patient must clean this filter weekly , as described below . The filter may require daily cleaning if the LifeStyle unit operates in a harsh environment; for example, a house heated by wood, kerosene, or oil, or one with excessive cigarette smoke.
1. Lift the corner of the LifeStyle unit to expose the air intake / gross
particle filter .
3-1
Page 16
2. Remove the filter from the unit and wash it in a solution of warm water
and soap.
3. Rinse the filter thoroughly , and remove excess water with a soft,
absorbent towel. Ensure that the filter is dry before replacing it.
4. Replace the dry filter.
3.2.2 Checking the Hour Meter
The compressor, product filter , and felt filter must be replaced after 3,000 hours of unit use. Instruct the patient to take note of the reading on the hour meter, located beneath the air intake gross particle filter , whenever cleaning the air intake / gross particle filter . When the reading approaches 3,000 hours, instruct the patient that it is important to notify the Equipment Provider.
3.3 Usage at Night or with Extended T ubing
Instruct patients that AirSep recommends that LifeStyle should be used while sleeping or with extended tubing (up to 25 ft, 7.5 m) only under the direction of their clinician.
3-2
Page 17
4.0 Equipment Provider Maintenance
4.1 Routine Maintenance
The LifeStyle unit includes two filters and a compressor that require scheduled mainte­nance or replacement.
T o ensure that the output of oxygen is within specification, you must perform a test of oxygen concentration. (Refer to Section 2.4, “Oxygen Concentration T est and Specifi­cation,” of this manual.) T est the unit upon delivery to a patient and at periodic inter­vals. Equipment Providers, based on their expertise and documentation, may establish and implement their own protocol to check oxygen concentration The interval estab­lished may be longer or shorter than 90 days, AirSep’ s default time period for providers who do not choose to establish their own protocol.
4.1.1 Air Intake / Gross Particle Filter
The external air intake gross particle filter is located under the panel covering the LifeStyle controls. You can easily remove it by hand. Instruct the patient to clean this filter weekly . (Refer to Section 3.2.1, “Cleaning the Air Intake / Gross Particle Filter,” of this manual.)
NOTE
The filter may require more frequent cleaning if the LifeStyle unit operates in a harsh environment — a house heated by wood, kerosene, or oil, or one with excessive cooking or cigarette smoke.
4.1.2 Product Filter Replacement
The product filter must be replaced after every 3,000 hours of use.
1. Set the unit’ s ON/OFF switch to the OFF position, and disconnect
the power cord.
2. Remove the bottom housing, and locate the product filter . (Refer to
Section 5.2.1 of this manual.)
3. Disconnect the tubing from the inlet and outlet of the product filter .
4. Connect the tubing to the new filter , and secure with tie wraps.
5. Replace the bottom housing.
4-1
Page 18
4.1.3 Compressor Replacement
As the Equipment Provider, you are responsible for monitoring the life of the compressor, which must be replaced after 3,000 hours of use.
Refer to Section 5.3 of this manual for compressor replacement procedures.
4.1.4 Recording Maintenance
As the Equipment Provider, you should record all routine maintenance and repairs performed on the LifeStyle unit, including hours and dates of service.
4.2 Infection Control
With growing concern about possible cross infection from home oxygen equipment from one home care patient to another, a clarification on this topic is necessary . The or gan­isms of most concern are M. Tuberculosis, HIV, and V iral Hepatitis. These are poten­tially pathogenic.
Tuberculosis can survive outside the human body , but its mode of transmission is by droplet nuclei. When infected individuals cough, they release droplet nuclei into the air, and these carry the Tuberculosis or ganism. These droplet nuclei may be breathed in by another person, but prolonged exposure to the infected person is usually necessary for infection to occur .
HIV and V iral Hepatitis are both viruses that are not living cells themselves but can duplicate when in a living “host” cell. Both of these organisms are usually passed on by person-to-person contact, and both need to be in the human body to survive. Once outside the body , viruses do not survive.
4.3 Cleaning and Preparing for New Patient Use
When you remove LifeStyle from a patient’ s home, always detach and dispose of the used nasal cannula. Clean the exterior of LifeStyle with a mild household cleaner applied with a damp cloth or sponge. Pay special attention to the oxygen outlet for the cannula connection to make sure it remains free of dust, water, and particles. Be careful not to get any liquid into the interior of the unit.
Next, clean the exterior with a common chemical disinfectant before any other patient
uses the unit. Do not use liquid directly on the LifeStyle unit to clean it.
4-2
Page 19
A list of undesirable chemical agents includes, but is not limited to, the following, according to the plastic manufacturer: alcohol and alcohol-based products (isopropyl alcohol), concentrated chlorine-based products (ethylene chloride), and oil-based products (Pine-Sol, Lestoil). These are not to be used to clean the plastic housing of LifeStyle. These chemicals and the cleaners that contain them will crystallize the plastic and break down the plastic’s molecular properties.
Clean the air intake / gross particle filter with warm soapy water between each patient’s use. The patients should be instructed to clean this filter at least once per week, de­pending on the environment, during normal operation.
Change the felt filter after 3,000 hours of use.
Change the product filter after 3,000 hours of use. It is not necessary to change this filter between patients even if the previous patient had a communicable disease or infection.
Allow LifeStyle to air dry , and then retest oxygen concentration before you return the unit to your inventory .
4-3
Page 20
Page 21
5.0 Service
5.1 Components
The design of AirSep’ s Lifestyle Portable Oxygen Concentrator allows for easy access and removal of most components. This allows you to perform scheduled maintenance and to repair and replace parts with minimal time and effort.
CAUTION
For your safety , be sure to disconnect the LifeStyle AC or DC adapter from the power outlet and remove the battery before you service the concentrator.
NOTE
5.2 Housing Removal
5.2.1 Removing Bottom Housing
1. Place the unit on its top, upside-down, and locate the four rubber
feet on the bottom housing.
2. Insert a Phillips-head screwdriver through each hole in the feet, and
loosen the screw at the bottom of each hole. Repeat this for all four feet.
3. Unscrew the two screws in the center of the bottom housing, and
remove the screws.
4. Lift off the bottom housing.
Record all scheduled maintenance. (Refer to Section 4.0 of this manual.)
5.2.2 Removing Top Housing
1. Using a pair of pliers, gently pull the selector knob off.
2. T urn the unit over so that it rests on the top housing.
5-1
Page 22
3. Use a Phillips-head screwdriver to remove the two screws on the
center housing near the handle and the screw in the hole in the center housing near the muffler .
4. T urn the unit over, and gently remove the top housing.
5.3 Compressor Replacement
The compressor is the “pump” within the oxygen concentrator that pushes room air into the bottom of the sieve beds. This allows oxygen to flow out the top.
The LifeStyle compressor uses a brush-type DC motor . Over time, the brushes wear, and the compressor does not operate properly . The expected life of the compressor is 3,000 hours.
5.3.1 Compressor Assembly Removal
T o remove the compressor assembly for exchange, follow the steps listed below:
1. Disconnect the LifeStyle unit from the AC or DC outlet, and remove the
battery .
2. Remove the top and bottom housing. (Refer to Section 5.2.)
3. Disconnect the suction tube.
4. Disconnect the orange 2-pin connector with the red and black wires
from the circuit board.
5. Remove the heat exchanger and clamp.
6. Remove the 3 hex nuts and split washers that hold the compressor to its
mounts.
7. Lift the compressor assembly , and stand up in the well of chassis.
8. Carefully disconnect the red compressor wires from the temperature
switch.
9. Remove compressor assembly . Inspect the vibration mounts for damage.
Replace if necessary .
10. Transfer the brackets to the new compressor , leaving the screws loose.
5-2
Page 23
5.3.2 Compressor Assembly Installation
T o install the new compressor assembly, perform the compressor removal procedure in reverse order .
NOTE
All washers (6) must be in their proper place before tightening the mounting nuts. Once the compressor is in place, adjust it left and right to ensure that it is in a level position before tightening the clamps.
5-3
Page 24
Page 25
6.0 Troubleshooting
6.1 General Troubleshooting
Before reviewing the Troubleshooting Chart, the following steps may be useful to isolate
any malfunctions:
1. Make sure all filters are clean.
2. Set the unit’s ON/OFF switch to the ON position. If the unit does not turn on,
refer to the Troubleshooting Chart that follows.
3. Activate the test mode by inserting a paper clip through the hole marked “T” to the
right of the hour meter.
4. Make sure that the unit is leak-free by testing all tubing connections and fittings with
leak-testing solution. Protect the circuit board from solution, and start the leak test at the heat exchanger, following the air flow of the unit to the oxygen outlet. Repair all leaks by tightening connections and fittings.
5. After three minutes, measure the oxygen concentration.
6. Review the Troubleshooting Chart that follows to isolate and repair any other
malfunctions.
6-1
Page 26
6.2 T roubleshooting Chart
Problem Possible Cause Solution
Unit does not turn on, Wall switch power is off. Turn on wall switch power . and AC adapter is properly connected to Defective AC adapter. Replace AC adapter. a wall outlet.
AC adapter is not plugged into Check connection from the LifeStyle. AC adapter to LifeStyle.
General malfunction Compressor failure. Replace compressor. alarm is activated.
Battery lasts less Battery not fully charged. Ensure that battery than 50 minutes. charges for at least 2 to
2 ½ hours.
Battery failure. Replace battery.
Oxygen concentration is Compressor has reached its Replace compressor. below specifications in expected life (~3,000 hrs).
test mode.
Unit does not deliver Cannula kinked or not Check cannula and the oxygen. connected. cannula connection.
LifeStyle is not turned on. Turn LifeStyle on to the
prescribed setting.
Battery is dead. Replace battery .
6-2
Page 27
Appendix
Exploded Drawings
Figure 1: LifeStyle — Control Panel A-1 Figure 2: LifeStyle — Internal, T op V iew A-2 Figure 3: LifeStyle — Internal, Bottom V iew A-3
Page 28
Ref No. Part No. Description
1 BT007-1 Battery, Assy, 3,000 mAh 2 CA186-1 Cabinet, Housing, Lower, Light Gray 3 CA188-1 Cabinet, Handle Cover, Dark Gray 4 CA199-1 Cabinet, Housing, Cover, Assy 5 CA211-1 Cabinet, Top Housing, Assy (logo, serial #,
label, etc.)
Ref No. Part No. Description
6 FI109-1 Filter, Foam, 15PPI, 2.35 x 2.35 x 1/2” 7 FI110-1 Filter, Felt, #10, Intake 8 MI187-1 Misc Knob, Selector Switch 9 MI195-1 Bumper, Rubber
Figure A.1: LifeStyle — Control Panel
A-1
Page 29
Ref No. Part No. Description
1 CB064-1 Circuit Board, Assy, Tubing 2 CO207-1 Compressor, Assy 3 CL028-1 Clamp, AL, 3/16” OD Tube 4 FN016-1 Fan, Assy (12VDC), Brushless
Ref No. Part No. Description
5 HX011-2 Heat Exchanger 6 MO016-2 Mount, Sandwich, Male-Female 7 SW002-1 Switch, Temperature, 145F NC 8 VA319-1 Valve, Assy, 12VDC
Figure A.2: LifeStyle — Internal, Top View
A-2
Page 30
Figure A.3: LifeStyle — Internal, Bottom View
A-3