The Ambulatory Oxygen System for Complete Patient
Mobility
Components
The complete ImPulse™ Elite OCD System includes
the following, as shown in Figure 1 below:
ImPulse Elite
Oxygen
Conserving
Device
Figure 1
Not shown:
• (1) One Alkaline “D” Energizer® Battery
• ImPulse Elite OCD System Patient Manual
Note: The carrying case is available separately.
Features
• Easy to operate with just one “D” alkaline battery
(Energizer recommended). Cannula connects in
seconds. Patient-accessible controls.
• Oxygen-conserving technology allows smaller
cylinders to last longer.
• With ImPulse Elite, the “D” battery provides the
greatest ambulation time. For most patients, 10–18
months.
• Pulse modes offer up to 6 lpm efciency.
• Fail-safe electronic circuitry and alarm system
provide added assurance of proper unit function and
oxygen delivery.
• Audible alarm when device does not detect
inhalation (inspiration), if cannula is disconnected or
tubing kinked, or if the unit is inadvertently left on
when not in use.
• Only minimal maintenance required.
• Designed for maximum patient mobility.
• Economical way for ambulatory oxygen patients to
enjoy freedom of movement.
• Developed to enhance quality of life for patients
using supplemental oxygen.
ImPulse Elite User Manual — 3
Introduction
ImPulse™ Elite is an electronic oxygen conserving
device (OCD) that provides convenience and mobility
to patients with therapeutic oxygen needs. Designed
and manufactured by AirSep Corporation, ImPulse
Elite uses portable cylinders to provide patients with
oxygen delivery at their prescribed dosages.
With this device, actual oxygen delivery to the patient
occurs at or near the beginning of the breathing cycle
when the patient starts to inhale.
Because ImPulse Elite does not deliver oxygen
during exhalation, which accounts for approximately
two-thirds of the breathing cycle, it wastes little or
no oxygen. This differs greatly from conventional
oxygen systems that deliver oxygen continuously. In
addition, many patients are more comfortable with
oxygen conserving devices than continuous ow
systems because they have fewer problems with nasal
dehydration that can result from constant ow into the
nostrils.
Maximizing Delivery with ImPulse Elite
ImPulse Elite’s advanced electronics enable it to
respond effectively to a patient’s respiratory cycle.
ImPulse Elite senses the start of inhalation. Because
only a small amount of oxygen is actually required to
oxygenate the blood effectively, ImPulse Elite delivers
the precise amount when it is most benecial — at the
beginning of inhalation.
ImPulse Elite quickly delivers a pulse of oxygen
compatible with your breathing rate and automatically
adjusts with any change in your breathing cycle —
depending on your prescribed oxygen requirement.
This method of oxygen delivery has been clinically
tested and determined to meet oxygen needs satised
by comparable ow settings on continuous ow units
up to 6 lpm.
Initial Setup
As part of your initial set up with ImPulse Elite, your
clinician may choose to assess your saturation level
through pulse oximetry. This helps to ensure proper
titration while using the oxygen conserving device.
Safety
ImPulse Elite meets applicable safety and
environmental regulations. This assurance means
you can fully enjoy all the benets of this device.
However, with all electronic devices and applications
that require oxygen, normal care and precautions must
be followed.
Before you attempt to use this oxygen conserving
device and a portable oxygen cylinder, read all
warning and caution notes in this section of the
manual. Contact your Equipment Provider should you
have further questions.
WARNING: FEDERAL (USA) LAW RESTRICTS
THIS OXYGEN CONSERVING DEVICE TO SALE
BY, OR ON ORDER OF, A LICENSED
PHYSICIAN. THIS DEVICE MUST BE USED
ONLY UNDER THE SUPERVISION OF A LICENSED
PHYSICIAN.
WARNING: HIGH PURITY OXYGEN PROMOTES RAPID
BURNING. DO NOT ALLOW SMOKING OR OPEN FLAMES
WITHIN FIVE FEET OF THIS DEVICE, THE PORTABLE
OXYGEN CYLINDER, OR ANY OXYGEN-CARRYING
ACCESSORY.
WARNING: IF YOU OBSERVE THAT THE OXYGEN
CONSERVING DEVICE IS NOT WORKING PROPERLY OR
YOU FEEL ILL, CONTACT YOUR EQUIPMENT PROVIDER
AND/OR PHYSICIAN IMMEDIATELY.
WARNING: ALTHOUGH OXYGEN IS A NON-FLAMMABLE
GAS, IT VIGOROUSLY SUPPORTS COMBUSTION. USE NO
OIL, GREASE, OR PETROLEUM-BASED PRODUCTS ON OR
NEAR THE DEVICE OR ANY OXYGEN SOURCE.
WARNING: USE THIS DEVICE AND THE OXYGEN CYLINDER
ONLY WITH CLEAN, DRY, OIL-FREE HANDS.
WARNING: OBSERVE ALL FEDERAL AND/OR
INTERNATIONAL REGULATIONS REGARDING VEHICULAR
TRAVEL BY AIR, SEA, OR LAND WHILE USING AND/OR
TRANSPORTING OXYGEN CYLINDERS.
CAUTION: Maintain a backup supply of
oxygen, and periodically check the contents
gauge on the unit, which monitors the
consumption of oxygen in the cylinder.
CAUTION: Do not use this device with a humidifier or
expose it to water or other liquids.
CAUTION: Do not expose the device to extreme
temperatures.
CAUTION: Do not use this device
• with the batteries removed.
• if the oxygen tubing is kinked or obstructed.
• if the device was tampered with or opened by
anyone other than your Equipment Provider.
4 — MN092-1 D
Description of Parts
As you can see in Figure 2 below, all parts are within easy reach and allow you to
adjust your ImPulse Elite unit as necessary. Follow your clinician’s instructions
before making any change to the ow selector setting.
Brass CGA Sealing Washer
Securing Knob
Figure 2a
Battery Test
Button
Low Battery
(Yellow Indicator)
Pulse Flow
(Green Indicator)
Oxygen Outlet
Pulse Flow/
Continuous
Flow Switch
Cylinder
Contents Gauge
Figure 2b
Figure 2
Easily connects to any standard
post-valve cylinder.
Alarm
(Red Indicator)
Battery
Compartment
(not shown)
Flow Selector
On/Off Switch
ImPulse Elite User Manual — 5
How to Operate the Unit
To connect the ImPulse Elite to any post-valve
cylinder:
Oxygen Cylinder Connection
WARNING: HIGH PURITY OXYGEN
PROMOTES RAPID BURNING. DO NOT
ALLOW SMOKING OR OPEN FLAMES WITHIN
OXYGEN CYLINDER, OR ANY OXYGEN-CARRYING
ACCESSORY.
WARNING: USE NO OIL, GREASE, OR OTHER PETROLEUMBASED PRODUCTS ON OR NEAR THE DEVICE OR ANY
OXYGEN SOURCE.
WARNING: USE THIS DEVICE AND THE OXYGEN CYLINDER
ONLY WITH CLEAN, DRY, OIL-FREE HANDS.
WARNING: OXYGEN IN THE CYLINDER IS UNDER HIGH
PRESSURE. DO NOT OPEN THE CYLINDER VALVE WITHOUT
THE IMPULSE ELITE UNIT PROPERLY CONNECTED TO THE
CYLINDER.
Your Equipment Provider will review proper cylinder
care and connection procedures with you the rst time
you operate your unit.
Refer to the procedure on the right should you need
assistance with connecting your ImPulse Elite unit to
an oxygen cylinder.
FIVE FEET OF THIS DEVICE, THE PORTABLE
Alignment
Pins
1. Loosen the ImPulse Elite securing knob. To
ensure a leak-tight seal between the cylinder
valve and the unit, make sure the sealing
washer is present in the center of the unit before
installation (Figure 3).
Sealing Washer
Figure 3
2. Align the two parallel alignment pins
(Figure 4) on the unit with the holes on the neck
of the cylinder (Figure 5).
Cylinder
Valve Toggle
Alignment Holes
6 — MN092-1 D
Figure 4 Figure 5
3. Push the alignment pins into the holes on the
neck of the cylinder, and hold the unit in place
against the cylinder with one hand while you
turn the unit’s securing knob clockwise with
your other hand until the unit is secured rmly
on the cylinder (Figure 6).
Figure 6
Testing and Replacing Battery
ImPulse Elite requires just one 1.5 volt “D” size
alkaline battery (Energizer recommended) for
full operation. Battery life is dependent upon the
predetermined unit settings, your breathing rate, and
the length of daily ambulation time. Most patients
experience 10–18 months of battery life. AirSep
strongly recommends that you maintain a spare
battery in your carrying case for replacement in this
unit as necessary.
To install or replace the battery, follow the steps
below:
1. Make sure the ON/OFF power switch is set
to the OFF position. (It is not necessary to
disconnect the cannula or remove the ImPulse
Elite from the cylinder, although you may nd
it easier to replace the battery within the battery
compartment door if the unit is rst removed
from the cylinder.)
2. If the unit is not connected to the cylinder,
simply hold the unit with one hand while you
press in and lift up on the battery compartment
latch with your other hand. (Figure 7).
Figure 7
3. Insert a fully charged battery into the inside
door of the battery compartment. Make sure
the battery is properly positioned to match the
correct positive and negative polarity markings
(Figure 9).
“+”Positive
Polarity
“-” Negative
Polarity
Figure 9
To test the battery to ensure that there is proper
power available to operate the unit:
1. Make sure there is a battery in the unit’s battery
compartment and it is properly positioned to
match the correct positive and negative polarity
markings.
2. Make sure the ON/OFF power switch is set to
the ON position.
3. Press and hold down rmly the center of the
BAT. TEST button with your nger. (Figure
10.) (You can test the battery regardless of
whether or not a cannula is connected to the
oxygen outlet.)
If the unit is connected to the cylinder, steady
the unit by grasping the neck of the cylinder
just below the unit with one hand while you
press in and lift up on the battery compartment
latch with your other hand (Figure 8).
Figure 8
Figure 10
The green pulse ow indicator illuminates to
indicate a good battery. If the battery requires
replacement, the green PULSE FLOW and
yellow LOW BATTERY indicators illuminate.
ImPulse Elite User Manual — 7
Turning On the Unit for Proper Oxygen
Usage
Note: Do not use a mask with the ImPulse Elite unit. A mask
may not conform to the face properly enough for the unit to
fully sense inhalation.
Note: Do not use pediatric or low flow cannulas with the
ImPulse Elite unit as their smaller tubing diameters negatively
affect delivery volumes.
6. Turn the cylinder valve slowly one full turn
counter-clockwise to begin oxygen ow.
7. Make sure the cylinder contents gauge on the
ImPulse Elite unit registers within the green
shaded/green zoned areas.
This indicates an adequate oxygen supply in
the cylinder. (If the cylinder contents gauge
indicates the red REFILL zone, the cylinder
needs to be replaced.)
Note: Use ImPulse Elite during sleep only under the
recommendation of a qualified clinician.
1. Check the energy level of the battery each time
you use the unit.
Refer to “Testing and Replacing Battery” for
step-by-step instructions.
2. Check the oxygen cylinder setup and the
cylinder contents gauge.
If you need to change cylinders, which includes
removal and reconnection of the ImPulse Elite
unit, refer to Oxygen Cylinder Connection for
step-by-step instructions.
3. Make sure the PULSE FLOW / CONTINUOUS
FLOW switch is moved completely right (for
pulse ow setting).
4. Set the ON/OFF switch to the ON position.
All three lights illuminate for less than 1
second. The ImPulse Elite unit automatically
calibrates and reaches sensitivity within
approximately 15 seconds. Then the green light
shuts off, and the unit emits a short burst of
oxygen. This indicates ImPulse Elite is ready
for use.
5. Connect your cannula to the oxygen outlet.
(You may use up to a 25 ft [7.5 m] cannula, if
desired.) (Figure 11.)
Listen for gas leaks coming from the ImPulse
Elite and oxygen cylinder. If you hear a leak,
turn off the cylinder valve and refer to the
Troubleshooting Guide on page 11.
8. Position the nasal cannula with the prongs
inserted into your nostrils.
9. Set the ow selector between 1 and 6 depending
on your prescribed dosage (Figure 12).
Figure 12
Turning Off the Unit
1. Turn the oxygen cylinder valve fully clockwise
to stop oxygen ow.
2. Set the ON/OFF power switch on the ImPulse
Elite unit to the OFF position.
Note: If you do not set the unit’s ON/OFF power switch to OFF
when the unit is not in use, the alarm sounds until the battery
fully discharges.
8 — MN092-1 D
Figure 11
Alarm Light Indicators
The following chart provides an explanation of what
each alarm light indicates:
Alarm Light Indicators
SettingLightAlarmIndicates
OFFNoNoUnit is not
ON +
(Flow
Selection)
BAT.
TEST
GreenNoPulse ow.
Yellow
&
Green
RedYes
GreenNoSufcient
Yellow
&
Green
NoNoNo battery or
Yes
(Intermittent)
(Continuous)
NoBattery
on, battery
life is
preserved.
Battery
almost
discharged.
Prepare
to replace
battery
within 20-30
minutes.
No inspiration
detected for
approximately
30
seconds
(cannula
may be
disconnected,
tubing
kinked, or
unit remains
on when not
in use).
battery life.
almost
discharged.
Prepare
to replace
battery
within 20–30
minutes.
discharged
battery in
unit.
Switching to Continuous Flow
If your unit is equipped with a continuous ow switch:
For very short time periods, this PULSE FLOW/
CONTINUOUS FLOW switch at the base of the unit
can be moved completely left for CONTINUOUS FLOW to force the oxygen conserving unit to perform
as a continuous ow unit (Figure 13).
Figure 13
A typical use for the CONTINUOUS FLOW setting
occurs when the battery can no longer provide power.
This setting provides 2 lpm of continuous ow
oxygen.
Before setting ImPulse Elite to CONTINUOUS FLOW, make sure you set the ON/OFF power switch
to the OFF position to preserve battery life as this
is a manual override and no audible or visual alarm
indicators are available in this “temporary” setting.
CAUTION: The continuous flow setting uses
substantially more oxygen than the
normally used pulse flow setting. Make sure
the cylinder contents gauge is observed
often when operating the unit as a continuous flow
device.
Removing/Replacing the Cylinder
To easily remove the ImPulse Elite unit from a
cylinder:
1. Close the oxygen cylinder valve by rotating the
valve completely clockwise.
2. Press the BAT. TEST button OR temporarily
set the PULSE FLOW/CONTINUOUS FLOW
switch to the CONTINUOUS FLOW setting
until no oxygen can be heard venting from the
unit.
3. Turn the unit’s securing knob counterclockwise
to lift the unit off the cylinder.
ImPulse Elite User Manual — 9
ImPulse Elite as a Portable Unit
The ImPulse Elite unit and its accessories are light
weight and portable. This means that ImPulse Elite
can go virtually anywhere you go.
CAUTION: Do not expose the device to, or
submerge it in, water.
CAUTION: Do not expose the device to extreme
temperatures.
The universal-style carrying case, as shown in
Figure 14, can be used with the unit. During use,
ImPulse Elite should remain in your carrying case.
This ensures added protection for the unit and allows
discreet use. Simply check periodically to ensure
that your cannula connection is secure and that the
cylinder shutoff valve remains in the open position
(one full turn counterclockwise) to enable oxygen
ow.
Figure 14
Proper Care & Maintenance
Each time you use the ImPulse Elite, check the
following as described on page 8; 1.) battery level,
2.) cylinder contents, 3.) pulse ow / continuous
ow setting, 4.) the green pulse light shuts off in
approximately 15 seconds after turning on switch, 5.)
listen for gas leaks coming from the ImPulse Elite
and oxygen cylinder. For any failure, refer to the
Troubleshooting Guide on page 11.
Keep the ImPulse Elite clean and free from moisture,
dust, and extreme temperatures. To help prevent
possible damage to the unit, keep ImPulse Elite in its
carrying case.
Clean the unit periodically by wiping with a lint-free
cloth. Pay special attention to the oxygen outlet for
the cannula connection to make sure it remains free
of dust, water, etc. Do not expose or submerge the
ImPulse Elite device in water.
We recommend using a one-part bleach to 99 parts
water solution or using Fantastik® spray cleaner as a
safe cleaner for the ImPulse Elite’s outer housing.
A list of undesirable chemical cleaning agents
includes, but is not limited to the following, according
to the plastic manufacturer: alcohol and alcohol-based
products (isopropyl alcohol), concentrated chlorinebased products (Ethylene Chloride), and oil-based
products (Pine-Sol, Lestoil). These are NOT to be
used to clean the plastic housing on the ImPulse Elite
OCD. These chemicals and the cleaners that contain
them will crystallize the plastic and break down the
plastic’s molecular properties.
Replace the disposable cannula periodically following
normal usage. Additional supplies are available from
your Equipment Provider or AirSep.
10 — MN092-1 D
Return empty oxygen cylinders to your supplier for
replacement. Remove the unit from the empty cylinder
before returning the cylinder to your supplier.
Note: Do not open the electronic compartment to service the
device. Open only the battery compartment when necessary.
To prevent a voided AirSep warranty, follow all manufacturer’s
instructions.
Troubleshooting Guide
If your unit fails to operate as described in this manual, rst review the following chart for possible causes and
solutions. If necessary, contact your Equipment Provider.
Pulse Flow — Visual (green); indicates good battery
Continuous Flow2 lpm oxygen; manual switch
Contents GaugeRed/Refill Zone: 0–500 psi
Green Shaded Zone: 500–2,200 psi
Green Outlined Zone: 2,200–3,000 psi
Operating
Temperature
Maximum
Cylinder
Pressure
32°F to 104°F (0°C to 40°C)
3,000 psi
12 — MN092-1 D
CAIRE Inc.
®
2200 Airport Industrial Drive, Ste. 500
Ball Ground, GA 30107 USA
1-800-482-2473
Worldwide: www.CAIREmedical.com
Medical Product Services GmbH
Borngasse 20
35619 Braunfels, Germany
Worldwide: www.CAIREmedical.com
CAIRE Inc. reserves the right to discontinue its products,
or change the prices, materials, equipment, quality,
descriptions, specifications and/or processes to its products
at any time without prior notice and with no further
obligation or consequence. All rights not expressly stated
herein are reserved by us, as applicable.
As of the date of this publication, ImPulse Elite, AirSep and their related logos
(as applicable) are trademarks registered by CAIRE Inc.
Energizer® is a registered trademark of Eveready Battery Company, Inc.
Fantastik® is a registered trademark of S.C. Johnson & Son, Inc.
MN092-1 Rev D
|MN092-1_D^~
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.