AirSep MN083-1 User Manual

AirSep FreeStyle5 POC
Reading and Recording Hours
1.) Insert the USB to TTL Serial Converter Cable (MI347) into an open USB port on your PC.
2.) Start the program by clicking on the Focus­FreeStyle 5 Data Interface icon on your
3.) When the program starts, you will see the Focus/FreeStyle 5 Data Interface box appear.
desktop.
4.) Unscrew the three screws on the front of the FreeStyle 5 and remove the front cover.
Screw
5.) Lay the FreeStyle 5 unit on its back, and then connect the
Black wire
on left
other end of the data cable to J11 on the circuit board. The
black wire goes to the left as shown.
Screws
6.) The Data Cable is now connecting the PC and the FreeStyle 5. Next plug the FreeStyle 5 unit into AC power, and then put the unit into test mode.
Note: Communication is only possible when the FreeStyle 5 unit is in test mode.
To activate the unit’s test mode:
Turn the unit on in flow setting 4.
After the 10 second start up time, press flow settings 1 and 5 simultaneously for 3 seconds.
All 5 flow setting LED’s will light after the 3 seconds. The unit will then start to auto pulse when in test mode.
If unit is not pulsing, turn unit off and repeat all steps.
7.) When the FreeStyle 5 unit enters Test Mode, the Focus/FreeStyle 5 Data Interface Program on your PC will display hours in the Unit Run Time box, and the Compressor Run Time box within 2 minutes.
8.) You may now record the hours from the Unit Run Time box. When recorded, turn the FreeStyle 5 off, disconnect AC power, disconnect data cable, and re-assemble.
MN083-1 Rev-
Loading...