
Patient Manual
Oxygen Concentrator
Includes
Oxygen Monitor, Air Outlet,
Dual Flow, and Pediatric Flow
Options

DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF
YOU ARE UNABLE TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT YOUR EQUIPMENT
PROVIDER BEFORE ATTEMPTING TO USE THIS EQUIPMENT; OTHERWISE, INJURY OR DAMAGE MAY
RESULT.
Smoking while using oxygen is the number one cause of fire, injuries and related
deaths.
You must follow these safety warnings:
Do not allow smoking, candles, or open flames in the same room with the device or within 5 feet
(1.52 meters) of the oxygen-carrying accessories.
Smoking while wearing an oxygen cannula may cause facial burns and possibly death.
Removing the cannula and putting it on bedding, sofas, or other cushion material will cause a
flash fire when exposed to a cigarette, heat source, or flame.
If you smoke, these 3 steps may save your life: turn off the oxygen concentrator, take off the
cannula, and leave the room where this device is located.
English: A multilingual version of the manual is available through your equipment provider
.
Español
: Una versión multilingüe del manual está disponible a través de su proveedor d
e equipo.
Français:
Une version multilingue du manuel est disponible par l'intermédiaire de votre fournisseur de matériel.
Deutsche:
Eine mehrsprachige Version des Handbuchs ist in Ihrer Geräte-Anbieter.
³No Smoking
±
Oxygen in Use
´ signs must be prominently displayed in the home, or where the
oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of
smoking in the presence of, or while using, medical oxygen.

TABLE OF CONTENTS
AirSep
®
NewLife
®
Elite
Oxyge
n Concentrator
AirSep NewLife Elite
Oxygen Concentrator 1
Important Safety Rules
1
Why Your Physician Prescribed Oxygen 6
What is an Oxygen Concentrator? 7
How to Operate Your Oxygen Concentrator 8
Filters
12
Ox
ygen Without Humidifier
12
Oxygen With Humidifier
13
Nasal Cannula
14
Proper Setting of Oxygen Flowmeter
15
Cleaning, Ca
re, and Proper Maintenance
16
Cabinet
16
Filters
17
Reserve Oxygen Supply 17

TABLE OF CONTENTS
AirSep® NewLife® Elite
Oxygen Concentrator
Troubleshooting 18
Product Specifications 20
Classification 21
Oxygen Monitor Option
22
Function of the Oxygen Monitor 22
Alarm Signal 22
Air Outlet Option
23
Operating the NewLife Elite Air Outlet 23
Dual Flow and Pediatric/
Low Flow Options
30
Dual Flow Application 31
Pediatric/Low Flow Application 31
Setting the Pediat
ric Flowmeter 32
Symbols/Abbreviations
33

[Read the Important Safety Rules section before operating this equipment.]
AirSep®
NewLife® Elite Oxygen Concentrator
This Patient Manual will acquaint you with AirSep¶s NewLife
Elite Oxygen Concentrator. Make sure you
read and understand all the
information contained in this manual before you operate your unit. Should you
have any questions, your Equipment Provider will be happy
to answer them for you.
Important Safety Rules
Carefully review an
d familiarize yourself with the following important safety information about the NewLife Elite Oxygen
Concentrator.
³No Smoking
±
Oxygen in Use
´ signs must be prominently displayed in the home, or where the oxygen
concentrator is in use. Patients and
their caregivers must be informed about the dangers of smoking in the
presence of, or while using, medical oxygen.
This device supplies high
-concentration oxygen that promotes rapid burning. Do not allow smoking or open
flames within the same room
of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe
this warning can result in severe fire, property damage and / or cause physical injury or death.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate
discomfort while using this machine may require additional monitoring. Patients with hearing and/or sight
impairment(s) may need assistance with monitoring alarms.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately
Use no oil, grease, or petroleum
-based or other flammable products on or near nasal end of cannula or the
NewLife Elite unit. Oxygen accelerates the combustion of flammable substances.

Electrical
shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to
prevent accidental electrical shock and burn hazard. Only your Equipment Provider or a qualified service
technician should remove the covers or service the unit.
Do not leave a nasal cannula under bed coverings or chair cushions. If the unit is turned on but not in use,
the oxygen will make the material flammable. Set the
I/0
power switch to the 0 (off) position when the
NewLife Elite unit is not in use.
Do not use extension cords with this unit or connect too many plugs into the same electrical outlet. The use
of extension cords could adversely affect the performance of the device. Too many plugs in one outlet can
result in an overload to the electric
al panel causing the breaker/fuse to activate or fire if the breaker or fuse
fails to operate.
Use only voltage specified on back panel label.
Do not use your oxygen concentrator in the presence of flammable gases. This can result in rapid burning
causing property damage, bodily injuries or death.
Care should be taken to prevent NewLife Elite from getting wet or allowing water to enter the unit. This can
cause the unit to malfunction or shut down, and cause an increased risk for electrical sh
ock or burns.
Do not use liquid directly on the unit. A list of undesirable chemical agents includes but is not limited to
the following: alcohol and alcohol-based products, concentrated chlorine-based products (ethylene
chloride), and oil
-based prod
ucts (Pine
-Sol, Lestoil). These are NOT to be used to clean the plastic
housing on NewLife Elite, as they can damage the unit¶s plastic.

Clean the cabinet, control panel, and power cord only with a mild household cleaner applied with a damp
cloth or
sponge, and then wipe all surfaces dry. Do not allow any liquids to enter the concentrator.
Ensure concentrator is operated in an upright position.
AirSep recommends an alternate source of supplemental oxygen in the event of a power outage, alarm
condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve
system required.
Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care
provider.
It
is very important to follow the prescribed level of oxygen flow. Do not increase or decrease the flow until
you first consult your physician.
In the event of an alarm or you observe the NewLife Elite is not working properly; consult the
troubleshooting
section of this manual. If you cannot resolve the problem, consult your Equipment
Provider.
Do not operate this unit in a restricted or confined space where ventilation can be limited. This can cause
the device to overheat and affect performance.
Storing the device outside of its temperature specifications may affect performance (refer to the
specification section of the manual).

If the alarm is weak or does not sound at all, consult your Equipment Provider immediately.
Position the unit away from curtains or drapes, hot air registers or heaters. Be certain to place the unit on a
flat surface and make sure all sides are at least 1 foot (30 cm) away from a wall or other obstruction. Do
not place the unit in a confined area. Choose a d
ust and smoke free
-location away from direct sunlight. Do
not operate the unit outdoors unless the unit is plugged into a Ground Fault Circuit Interrupter (GFCI
)
protected outlet.
If the humidifier bottle tubing is not properly connected to the humidifier bottle fitting or the oxygen outlet,
an oxygen leak can occur.
Normally, you should not need to adjust the flowmeter on your unit. If you turn the flowmeter adjustment
knob clockwise, you will decrease and can shut off the flow of oxygen from you
r unit. For your
convenience, the flowmeter is marked in ½ lpm increments from 1 to 5 lpm. For units with the pediatric
flowmeter option, the flowmeter is marked in 1/8 lpm increments for flow settings up to 2 lpm.
To prevent a voided AirSep warranty,
follow all manufacturers¶ instructions.
Replace the disposable cannula periodically following normal usage according to your Equipment Providers
recommendations.
AirSep does not recommend the sterilization of this equipment.
Do not attempt
any maintenance other than the possible solutions listed below.

Always follow the cannula manufacturer¶s instructions for proper use. Consult your licensed health care
provider to determine how often the cannula should be replaced.
The standard NewLife Elite Oxygen Concentrator accommodates prescriptions from 1 to 5 lpm.
For prescriptions levels below 1 lpm, you must operate the NewLife Elite unit for at least 5 minutes at 2
lpm
The use of some oxygen administration accessories not
specified for use with this oxygen concentrator
may impair its performance.
The use of certain humidifiers not specified for use with this oxygen concentrator may impair its performance.
To Equipment Provider:
The following humidifier bottle is recommended for use with the NewLife Elite Oxygen
Concentrator:
AirSep Part No.
HU003-1
To Equipment Provider:
The following oxygen administration accessories are recommended for use with the
NewLife Elite Oxygen Concentrator:
•
Nasal Cannula w
ith 7 feet (2.1 m) of tubing: AirSep Part No.
CU002-1.
•
Oxygen Outlet Adapter: AirSep Part No.
F0025-1
.

[Read the Important Safety Rules section before operating this equipment.]
Why Your Physician Prescribed Oxygen
Many
people suffer from a variety of heart, lung, and other respiratory diseases. A significant number of these patients can benefit
from supplemental ox
ygen therapy at home, in the hospital, or at a medical facility.
Oxygen is a gas that makes up 21% of
the room air we breathe. Our bodies depend on a steady supply to function properly. Your
physician prescribed a flow or setting to address your particular respiratory condition.
Although
Oxygen i
s a non-additive drug,
unauthorized oxygen therapy can be
dangerous. You must seek medical advice before you
use this oxygen concentrator. The Equipment Provider who supplies your oxygen equipment will demonstrate how to set the
prescribed flow rate.
It is very important to select the prescribed level of oxygen flow. Do not increase or decrease the flow until you
fi
rst consult your physician.

[Read the Important Safety Rules section before operating this equipment.]
What i
s an
Oxygen Concentrator?
Oxygen concentrators were introduced in the mid-
1970¶s
and have become the most convenient, reliable source of supplemental
oxygen available today. Oxygen concentrators are the most cost-effective, efficient, and safest
alternative to using high-pressure
oxygen cylinders or l
iquid oxygen. An oxygen
concentr
ator provides all the oxygen you need with no cylinder or bottle deliveries
required.
The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gases. In the
NewLife Elite unit, room air passes
through a regenerative, adsorbent mate
rial called ³molecular sieve.´ This material separates the oxygen from the nitrogen
. The result is
a flow of high-concentration oxygen delivered to the patient.
There is never a danger of depleting the oxygen in a room when you use your NewLife
Elite
unit.

[Read the Important Safety Rules section before operating this equipment.]
How to Operate Your Oxygen Concentrator
First, become familiar with the important parts of your NewLife
Elite
Oxygen Concentrator (Figures 1a and 1
b).
A. On/Off
(I/0
)
Power Switch:
Starts and stops the operation of the unit.
B.
Circuit Break
er Reset Button:
Resets the unit after electrical overload shutdown
C. Digital Hour Meter:
Records the unit¶s total hours of operation.
D. Flowmeter/Adjustme
nt Knob:
Controls and indicates the oxygen flow rate in liters per minute
(lpm).
E. Oxygen
Outlet :
Provides connections
for a humidifier (if required) or
nasa
l cannula.
F. Top and Side Handles:
Figure 1a
Enables convenience in carrying the unit.
G. Operating Instructions:
Explains procedures to operate the unit.

[Read the Important Safety Rules section before operating this equipment.]
I. Air Intake Gross Particle Filter:
Prevents dust and other airborne particles from entering the
unit.
J
. Power Cord
:
Allows connection of unit into electrical outlet.
Do not use extension cords with this unit or connect
too many plugs into the same outlet. Too many plugs
in one outlet can result in an overload to the electrical
panel c
ausing the breaker/fuse to activate or fire if the
breaker or fuse fails to operate.
1.
Locate the unit near an electrical outlet in the room where
you spend
most of your time.
2.
Position the unit away from curtains or drapes, hot air registers, heaters,
and fireplaces.
Be certain to place the unit so all sides are at least 12 inches
(30.5 cm)
away from a wall or other obstruction. Do not place the unit in a
Figure 1b
confined area.
3.
Turn the unit so that the operating controls are within easy reach and the air intake on the back of the unit is not obstructed.
4.
Connect oxygen accessories such as a humidifier (if required),
nasal
cannula, and/or extension tubing to the oxygen outlet.
5.
Completely unwrap the
power cord (Figure 1b).

[Read the Important Safety Rules section before operating this equipment.]
6.
Insert power cord into the electrical outlet.
7.
Locate the power switch on the front of the unit, and switch it to
the
I
position (on). (Figure 2.)
A battery
-operated audible alarm must sound for a 5-
second
test to indicate a good battery and alarm.
If the alarm is weak or does not sound at all, consult
your Equ
ipment Provider immediately.
The standard New
Life Elite
Oxygen
Concentrator
accommodates
prescriptions from 1 to 5 lpm
.
8.
Dual Flow and High Flow Applications:
Set the flowmeter adjustment
knob(s) to the prescribed lpm, in any combination of flows up to a total
of 5 lpm for Dual 5 un
its (Figure
3).
or
Pediatric/Low Flow Applications:
Set the flowmeter adjustable knob(s)
to the prescribed lpm up to 2 lpm in 1/8 liter (125 ccm) increments.
Figure 3
The concentrator is now ready for use.

[Read the Important Safety Rules section before operating this equipment.]
9.
To turn the concentrator off, press the
I/0
switch to the
0
position.
Ensure concentrator is operated in an upright position
For prescriptions levels below 1 lpm, you must operate the NewLife Elite unit for at least
5 minutes at 2 lpm
.
10.
If the NewLife Elite
unit fails to operate properly, refer to the Troubleshooting sectio
n for a list of probable causes
and
solutions.

[Read the Important Safety Rules section before operating this equipment.]
Filters
Air enters the NewLife Elite
unit through an air intake gross particle filter located on the
back off the oxygen concentrator. This filter removes dust particles and other
large particles from
the
air. Bef
ore you operate the NewLife Elite unit, make sure this filter is clean and positioned
correctly
(Figure 4).
The supplemental oxygen p
roduced by the NewLife Elite unit receives additional filtration
f
rom a product filter located within the oxygen concentrator. Your Equipment Provider performs
m
aintenance on the product filter in addition to other maintenance on the unit.
Figure 4
The use of some oxygen administration accessories not
specified for use
with this oxygen concentrator may impair its performance.
Oxygen Without Humidifier
1. I
f your physician did not prescribe a humidifier, connec
t the oxygen tubing directly to
the unit¶s
Oxygen outlet. A sep
arate outlet fitting is su
pplied
for this type of connection
(Figure 5).
Figure 5

[Read the Important Safety Rules section before operating this equipment.]
Operating With Humidifier
Following these steps if your physician prescribed an oxygen humidifier as part of your therapy:
1. Remove or unscrew the reservoir bottle from the humidifier (If you have a pre-
filled
unit, do not perform this step. Proceed directly to step 4.)
2. Fill the reservoir with cool or cold water (distilled water is p
referred) to the fill lin
e indicated
on the bottle. DO NOT OVERFILL.
3. Screw the reservoir bottle back together.
Figure 6
4. On the top of the humidifier, turn th
e thread nut counterclockwise while you connect the
humidifier to the oxygen outlet, and tighten securely (Figure 6).
5. Connect oxygen tubing from the
nasal
cannula
to the humidifier outlet fitting (Figure 7).
The use of certain humidifiers not specified for use with this oxygen concentrator may
impair its performance.
To Equipment Provider:
The following humidifier bottle is recommended for use with
the NewLife Elite Oxygen Concentrator:
AirSep Part No.
HU003-1

[Read the Important Safety Rules section before operating this equipment.]
Nasal Cannula
Your physician has prescribed either a nasal cannula
, face mask, or
other accessories
(Figure 8). In
most
cases
the manufacturer has already connected the oxygen supply tubing to the
nasal cannula,
f
ace
mask, or
other accessory
. If not, follow the manufacturer¶s instructions for proper connection.
Connect
the oxygen tubing
to the o
xygen outlet adapter or humidifier.
To Equipment Provider:
The following oxygen administration accessories ar
e
recommended fo
r use with the NewLife Elite Oxygen Concentrator:
•
Nasal Cannula with 7 feet
(2.1 m)
of tubing: AirSep Part No.
CU002-1.
•
Oxygen Outlet Adapter:
AirSep Part No.
F0025-1
.

[Read the Important Safety Rules section before operating this equipment.]
Proper Setting of Oxygen Flowmeter
To set the proper flow of supplemental oxygen, turn the flowmeter adjustment knob left or right
until the ball inside the flowmeter centers on the flow line number prescribed by your physician
(Figure 9).
To view the flowmeter at the proper angle, note that the back line and the front numbered line
m
ust give the appearance of just one line.
It is very important to
follow
the prescribed level of oxygen. Do not increase or
decrease the flow until you first consult your physician.
Normally, you should not need to ad
just the flowmeter on your unit.
If you turn the
flowmeter adjustment knob clockwise, you will decrease and can shut off the flow
of oxygen from your unit. For your convenience, the flowmeter is marked in ½ lpm
increments from 1 to 5 lpm
. For units with the pediatric flowmeter option, the
flowmeter is marked in 1/8 lpm increments for flow settings up to 2 lpm.

[Read the Important Safety Rules section before operating this equipment.]
Cleaning, Care, and Proper Maintenance
Cabinet
Electrical shock hazard. Disconnect the power cord from the electric outlet before you clean the unit to
prevent accidental electrical shock and burn hazard. Only your Equipment Provider or a qualified service
technician should remove the covers or service the unit.
Do not use liquid direc
tly on the NewLife
Elite
unit to clean it. A list of
undesirable
chemical agents
includes but is not limited to, the following, according to the plastics
manufacturer: alcohol and
alcohol-
based products, concentrated chlorine-based
products (ethylene chlor
ide), and oil-based products (Pine
-
Sol
, Lestoil). These are NOT to be used to clean the plastic housing
on NewLife Elite, as
they can
damage the unit¶s plastic.
Clean the cabinet, control panel, and power cord only with a mild household cleaner applied with a damp
cloth or sponge, and then wipe all surfaces dry. Do not allow any liquids to enter the concentrator.
To prevent a voided AirSep warranty, follow all
manufacturers¶ instructions
.
AirSep does not recommend the sterilization of t
his equipment.

[Read the Important Safety Rules section before operating this equipment.]
Filters
Do not operate the unit without the air intake gross particle filter in place.
At least one time each week, wash the air intake gross particle filter, which is located in the back of the unit. Your
Equipment Provider may advise you to clean it more often, depending upon your operating conditions. Follow these
steps to properly clean the air intake filter:
1. Remove the filter
and wash it in a warm
solution of soap and water.
2. Rinse the filter thoroughly, and remove excess water with a soft, adsorbent towel. Ensure that the filter is dry
before replacing it.
3. Replace the dry filter.
Reserve Oxygen Supply
Your Equipment Provider may recommend ano
ther
source of supplemental oxygen therapy in case there is a mechanical
f
ailure or a power outage.

Troubleshooting
If your NewLife
Elite
Oxygen Concentrator fails to operate properly, refer to the chart on the following pages for possible
causes and solutions and, if needed, consult your Equipment Provider.
If you cannot get the unit to operate, connect your nasal
cannula
, face mask, or other accessories to the reserve supplemental
oxygen supply (if provided).
Do not attempt any maintenance other than the possible solutions listed below.
Unit does not operate
. Power failure condition
causes a continuous alarm to sound.
Power cord not connected into electrical
outlet.
No power at electrical outlet.
O
xygen concentrator circuit breaker is
activated.
Check power cord plug at the electrical outlet for a
proper connection.
Check power source, wall switch, fuse, or circuit
breaker in-house.
Contact your Equipme
nt Provider for service.
Dirty or obstructed humidifier bottle.
Defective nasal cannula,
face mask,
catheter, and/or oxygen delivery tube
,
or other accessory
.
Other leak or restriction.
Remove the humidifier bottle (if used) from the
oxygen outlet
. If
flow is
rest
ored, clean or replace
with a new humidifier bottle.
Remove nasal cannula
, face mask, or
other
accessories from oxygen tubing. If proper flow is
restored, replace with new nasal
cannula
, face
mask,
or
other accessories
.
Disconnect delivery tubing at oxyg
en outlet (front
of unit). If proper flow is restored, check oxygen
tubing for kinks or obstructions. Replace if needed.
Contact your Equipment Provider.

Condensation collects in the oxygen tubing when
you use the humidifier bottle.
Unit not properly ventilated. Elevated
operating temperature.
Make sure unit is positioned away from curtains or
drapes, hot air registers, heaters, and fireplaces. Be
certain to place the unit so all sides are at least 12
inches
(30
.5 cm)
away from a wall or other
obstruction. Do not place the unit in a confined
area.
Allow oxygen tubing to dry out, or replace with
new tubing.
Refill humidifier bottle with COLD
water. DO NOT OVERFILL.
Intermittent alarm sounds at one second
inte
rvals.
Set
I/0
power switch to 0
position, use your reserve
oxygen supply (if provided), and consult your
Equipment Provider immediately.
Unit does not alarm, or weak alarm sounds for 5
seconds during start-up.
Call your Equipment Provider to replace 9-volt
battery.
Set
I/0
power switch to the 0
position, use your
reserve oxygen supply (if provided), and consult
your Equipment Provider immediately
.

1-3 lpm: 95.5% -
92.0%
4 lpm
: 92
% ± 3%
5 lpm: 90% ± 3%
Dimensions:
28.5 in. high x 15.7 in. wide x 14.5 in. deep
(72.4 cm high x 40 cm wide x 36.8 cm deep)
120 VAC, 60 Hz, 4.0 amps
350 watts ±
5 lpm model
Two-prong polarized plug
Double insulated cabinet
CSA-Approved to IEC 601-1
Export Models:
220-240 VAC, 50 Hz, 2.0 amps
220 VAC, 60 Hz, 2.0 amps
Double-insulated cabinet
Battery test
Power failure
High and low pressure
Low oxygen concentration (
with Oxygen M
onitor option
)
4F to 140F (-20C to 60C)
* Based on an atmospheric pressure of 14.7 psi (101 kPa) at 70F (21C)
* Operating unit outside of operational temperature range can affect performance.

Classification
Type of protection against electric shock:
Class II
Protection from electric shock is achieved by double insulation.
Protective earthing or reliance upon install
ation conditions are not required.
Degree of protection against electric shock:
Type B
Equipment providing a particular degree of protection against electric shock
particularly
regarding:
1) allowable leakage current;
2) reliability of protective earth co
nnection (if present).
Not intended for direct cardiac application.
Degree of protection against harmful ingress of water:
Drip-
proof equipment ±
IPX1.
Equipment provided with an enclosure preventing of such an amount of falling liquid as might interfer
e with the
satisfactory and safe operation of the equipment.
Method of cleaning and infection control allowed:
Please refer to Maintenance section in the NewLife
Elite Service Manual.
Degree of safety of application in the presence of flammable anesthe
tic gases:
Equipment not suited for such application.
Mode of operation:
Continuous duty.

[Read the Important Safety Rules section before operating this equipment.]
Oxygen Monitor Option
The following information will acquaint you with Oxy
gen Monitor option of the
NewLife
Elite Oxygen Concentrator. Make sure you
read and understand all the information contained in this manual before you operate your
unit. Should you
have any questions, your
Equipment Provider will be happy to answer them for
you.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate
discomfort while using this
machine may require additional monitoring. Patients with hearing and/or sight
impairment(s) may need assistance with monitoring alarms.
Contact your Equipment Provider immediately if the amber OXYGEN MONITOR light remains on for more
than 15 minutes.
Function of the Oxygen Monitor
The oxygen monitor is a small electronic device within the NewLife
Elite
Oxygen
Concentrat
or that monitors the
concentration
of oxygen produced by the unit.
Alarm Signal
If oxygen
concentration
falls below the acceptable therapeutic level,
an amber
OXYGEN MONITOR light on the Oxygen Concentrator turns on (Fi
gure 10). If the
light
remains on fo
r more than 15 minutes, an intermittent alarm sounds.
Figure 10
When you turn the unit on, it¶s normal for the amber OXYGEN MONITOR light to
turn on and remain on for up to five minutes.

[Read the Important Safety Rules section before operating this equipment]
Air Outlet Option
The following information will acquaint you with the Air Outlet option for the NewLife Elite Oxygen Concentrator.
Make sure you read and understand all the information in this NewLife Elite Patient Manual before you operate your unit.
Should you have any questions, your Equipment Provider will be happy to answer them for you.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate
discomfort while using this m
achine may require additional monitoring. Patients with hearing and/or sight
impairment(s) may need assistance with monitoring alarms.
Operating the NewLife Elite Air Outlet
1. Read and understand all information contained in the NewLife
Elite Patient Manual¶s How to Operate Your Oxygen
Concentrator section before you operate your unit.

[Read the Important Safety Rules section before operating this equipment.]
Air Valve
Part No.
VA007-1
Air Outlet
Barb Fitt
ing
Part No.
F0032-1
Figure 1
Figure 2
Air Outlet option
* Shown with hand held nebulizer
The NewLife Elite Air Outlet option allows you to connect a hand held nebulizer
(Figure 1)
.
You may continue to receive oxygen from the unit while you use the Air Outlet
option
.

[Read the Important Safety Rules section before operating this equipment.]
2. Locate the air outlet barb fitting on the fro
nt of the unit
(Figure 2)
. If a hand held nebulizer will be used, connect one end of
the air supply tubing to the air outlet barb fitting and the other end to the bottom of the nebulizer.
3. Fill the nebulizer cup with medication as prescribed by your physician
(Refer to the Nebulizer With Medi
cation section for filling instructions).
4. Operate the NewLife Elite unit for at least five minutes, and then
open the air valve completely
(Figure 3)
.
5. Begin your treatment (Refer to the Inhaling Medication/Treatment Instruction sections).
Nebulizer medication will now be visible as a fine mist.*
Figure 3
Oxygen
-enriched air is not delivered at the air outlet.
In high humidity environments or during extended period of non
-use, open the air
valve completely
(Figure 3)
to purge/flush the system.

[Read the Important Safety Rules section before operating this equipment.]
6. When treatment is complete, turn the air valve to the
OFF
position.
(Figure
4)
Figure 4
7. Disconnect the nebulizer and the air supply tubing from the air outlet barb fitting.
8. Clean the nebulizer. (Refer to the Cleanin
g the Nebulizer section).
* If you think that your nebulizer is not operating properly, contact your Equipment Provider.
Filling the Nebulizer with Medication
1. Wash your hands thoroughly.
2. Use an eyedropper, syringe, or other measuring device to measure out the proper amount of medication, as prescribed by
your physician.
The New
Life Elite Air Outlet regulator is preset to deliver 6
liters per minute ( lpm) at 12 psig (85 kPa).
Use only the amount of medication and frequency of treatment that your physician
prescribed

[Read the Important Safety Rules section before operating this equipment.]
3. Remove or unscrew the medication cup on the nebulizer, and place your prescribed measured dosage into the medication
cup
(Figure 5).
4. Connect the medication cup to the nebulizer, and then connect the ³T´ piece or mouthpiece to the nebulizer
(Figure 6).
Figure 5
Figure 6
5. Connect one end of the air supply tubing to the air outlet barb fitting and the other end to the bottom of the nebulizer, and
open the air valve completely as shown in
Figure 3.
6.
Begin your treatmen
t. (Refer to the Inhaling Medication/Treatment Instruction section
)

[Read the Important Safety Rules section before operating this equipment.]
Inhaling Medication/Treatment Instructions
1. Close your mouth around the mouthpiece, but do not hold it with
your teeth
(Figure 7)
.
2.
Take a slow, deep breath, and pause at the end of the inhalation for 1 -
2 seconds,
then exhale slowly and completely.
3.
Repeat this procedure until the prescribed amount of medication nebulizes or the
Prescribed amount of treatment time elapses (whichev
er occurs first).
4. If your physician or health care professional instructed you to take short rest periods
Figure 7
During your treatment, make sure you turn the air valve to the
OFF
position, as shown
in
Figure 4.
This will conserve your medication.
The following instructions for inhaling medication are o
ften recommended. If your physician or health
care professional has given you special instructions, make sure you follow them instead, as prescribed.
Prolonged treatment time can indicate a defective nebulizer. Contact your Equipment Provider if this
condition exists.

[Read the Important Safety Rules section before operating this equipment.]
Cleaning the Nebulizer
1. After each treatment, separate the nebulizer and the ³T´ piece or mouthpiece assembly.
2. Remove or unscrew the nebulizer cu
p, and rinse each component thoroughly in warm
water.
3. Once a day, clean all nebulizer parts (excluding air supply tubing) with a mild detergent
or soap solution in warm water. Rinse thoroughly, and soak all parts in a solution of
Figure 8
one (1) par
t white vinegar and three (3) parts water for 30 minutes to disinfect.
4. Rinse thoroughly in warm water to remove the disinfectant solution.
5. Place all nebulizer parts on a paper towel or soft absorbent material to air dry.
DO NOT
WIPE DRY.
6. When dry, store the nebulizer parts in a clean container or plastic bag.
7. Repeat the above procedure after each treatment/patient use.
Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care
provider.
Perfo
rm steps 1 and 2 below after each treatment to prevent medication from collecting and hardening
inside the nebulizer parts.

[Read the Important Safety Rules section before operating this equipment.]
Dual Flow and Pediatric/Low Flow Options
The following information will acquaint you with the
5 liter
dual flow and pediatric/low flow options of the
NewLife Elite Oxygen
Conce
ntrator
(see Figure 9)
.
Make sure you
read and understand all the information contained in this manual before you operate
your unit. Should you
have any questions, your Equipment Provider will be happy to answer them for you.
³No Smoking
±
Oxygen in
Use
´ signs must be prominently displayed in the home, or where the
oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of
smoking in the presence of, or while using, medical oxygen.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to
communicate discomfort while using this machine may require additional m
onitoring. Patients with
hearing and/or sight impairment(s) may need assistance with monitoring alarms.
This device supplies high
-concentration oxygen that promotes rapid burning. Do not allow smoking or
open flames within the same room of (1) this device, or (2) any oxygen
-carrying accessory. Failure to
observe thi
s warning can result in severe fire, property damage and / or cause physical injury or death.
Use no oil, grease, or petroleum-based or other flammable products on or near nasal end of cannula or
the NewLife Elite unit. Oxygen accelerates the combustion of flammable substances.

[Read the Important Safety Rules section before operating this equipment.]
Dual Flow Application
The NewLife Elite unit¶s dual flow option allows a single concentrator to meet
the high flow requirements of one patient
(Figure 9)
or the needs of two patients,
in any combination of flows up to the maximum capacity of the concentrator.
Excellent for use in the home, extended care facility, hospital, or physician¶s
waiting room.
Figure 9
Pediatric/Low Flow Application
The pediatric flowmeter (available for use with the dual flow NewLife Elite unit)
meets low flow requir
ements up to 2 lpm in 1/8 liter (125 ccm) increments
Figure 10).
Figure 10
Two Patients/
Dual Flow
Applications

[Read the Important Safety Rules section before operating this equipment.]
Setting the Pediatric Flowmeter
When using the pediatric flowmeter, the unit will not reach proper concentration at the pediatric setting (less than 2 lpm) until
you bleed off a portion of the oxygen by opening the primary flowmeter (on the left side of the unit). Follow the procedure
below when using the pediatric flowmeter.
1. Follow the start-
up instru
ctions 1
-7 as outlined on page 9 & 10.
2. Set the pediatric flowmeter to the prescribed flow.
3. Set the primary flowmeter to 2 lpm to bleed off excess product, and allow the unit to achieve maximum concentration.
The NewLife Elite is appropriate for usage by two patients,
provided the combined flow is a minimum of 2 lpm and does
not exceed the maximum capacity of the concentrator.
The
NewLife
Elite
Oxygen Concentrator must be operated
for at least five minutes at 2 lpm before using the unit.

Symbols/Abbreviations
Symbols are frequently used on equipment and/or the manual in preference to words with the intention of decreasing the
possibility of misunderstanding caused by language differences. Symbols can also permit easier comprehension of a concept
within a restricted space.
The following table is a list of symbols and definitions that may be used with the NewLife Elite Oxygen Concentrator.
Consult instructions for use
Warning
±
Describes a hazard or unsafe
practice that if not avoided can result in
sev
ere bodily injury, death or property
damage
Caution ±
Describes a hazard or unsafe
practice that if not avoided can result in minor
bodily injury or property damage
Complies with the 93/42/EEC directive drawn up by
the
approved organization No. 0459
Note ± Provides information important
enough to emphasize or repeat
Safety agency for CAN/CSA C22.2 No. 601.1 M90
for medical electrical equipment

Safety agency for CAN/CSA C22.2 No.
601.1 M90 for medical electrical
equipment.
Oxygen concentration warning LED
Keep unit and accessories dry
Consult the accompanying documents
Proper disposal of waste of electrical
and electronic
equipment required
Oxygen outlet connection to the cannula
Do not expose to open flames
Fragile ±
handle with care
Keep in the vertical position
Caution: Federal law (USA) restricts this
for sale or rental by or o
n the order of a
physician or licensed health care
provider.

MN078-1 Rev F
8/12
For European representative:
Gavin Ayling
9 Bungham Lane
Penkridge Stafford
Staffordshire ST19 5NH England
E-
mail:
eurorepcontact@
airsep.com

MN078-1 Rev F
8/12
For se
rvice on your NewLife Elite
Oxygen Concentrator, please contact your local Equipment Provider at:
Manufactured by:
AirSep Corporation
Buffalo, NY 14228-
2085 USA
www.airsep.com
|MN078-
1_FÍ~