Reading and Recording Hours
1.) Insert the Focus data cable (MI347) into an open USB port on your PC.
2.) Start the program by clicking on the Focus Data Interface icon on your desktop.
3.) When the program starts, you will see the Focus Data Interface box appear.
4.) Unscrew the two screws on the bottom of the Focus and remove the covers.
5.) Lay the Focus unit on its back, and then
black wire
on top
connect the other end of the data cable to J9 on the circuit board.
Theblack wire goes to the top pin as
2 screws
shown.
6.) The Data Cable is now connecting the PC and the Focus. Next plug the Focus unit into AC power, and then put the unit into test mode.
Note: Communication is only possible when the Focus unit is in test mode.
To activate the unit’s test mode:
Turn the unit on.
After the 10 second start up time, turn the unit off & on consecutively 3 times (within 4 seconds).
The unit will then start to auto pulse and is now in test mode.
If unit is not pulsing, turn unit off and repeat all steps.
7.) When the Focus unit enters Test Mode, the Focus Data Interface Program on your PC will display hours in the Unit Run Time box, and the Compressor Run Time box within 2 minutes.
8.) You may now record the hours from the Unit Run Time box. When recorded, turn the Focus off, disconnect AC power, disconnect data cable, and re-assemble.
MN028-1 Rev-