AirSep ML-CONC0056 User Manual

eTASK User Guide
Use eTASK with SeQual® Eclipse and eQuinox™ Transportable Oxygen Concentrators
To better assist in titration knowledge, CAIRE has developed eTASK, an interactive software program and smart phone application. By targeting consistent FiO2 levels, CAIRE’s eTASK quickly illustrates the conversion between continuous flow and pulse dose therapy with autoSAT. Targeting FiO2s is critical to effective oxygenation, a challenge for ambulating patients due to their changing respiratory rates. eTASK also calculates battery durations based on the necessary flow rate and the patient’s breath rate.
Windows® Version
1
Titrate to Saturate
Variable Entry Continuous Flow Equivalent Pulse Dose
Continuous Flow (LPM): 3.0
Breaths Per Minute: 20
VT (mL): 500
Flow: 3.0 LPM
Est. Battery Life: 1.30 hrs
Pulse Dose Setting:
Pulse Setting Number: 3.5
Est. Battery Life: 3.61 hrs
3
Concentrator O2 (%): 92.0
Device Type: Eclipse
Battery Type:
For use only with the SeQual Eclipse and eQuinox. Not for clinical use with other devices. For use only under the supervision
of a medical professional. Equivalent pulse calculations are for reference only. Pulse dose should not be prescribed without
the patient being evaluated by a physician or therapist to determine its eectiveness. Battery durations are typical operating
times for a new battery. Durations may decrease with age, temperature, and breath rate.
2
1 2 3 4
Using the up/down arrows, input the individual patient scenario of continuous flow rate, BPM, tidal volume, and desired O2 concentration.
Using the up/down arrows, input the CAIRE concentrator the patient will be using (Eclipse or eQuinox). If using an eQuinox, select whether the 12-cell or 24-cell battery will be used.
The patient’s necessary continuous flow rate will be displayed here, along with the estimated battery duration. Note that if the required continuous flow rate is above
3.0 LPM, this is not available on the Eclipse or eQuinox and will be shaded in gray.
4
56 mL
Equivalent Eclipse or eQuinox pulse dose setting in mL and setting number will be displayed here along with the battery run time.
Note the pulse setting numbers are not LPM equivalents.
Battery run time on pulse settings will vary based on breath rate.
eTASK User Guide
Use eTASK with SeQual® Eclipse and eQuinox™ Transportable Oxygen Concentrators
To better assist in titration knowledge, CAIRE has developed eTASK, an interactive software program and smart phone application. By targeting consistent FiO2 levels, CAIRE’s eTASK quickly illustrates the conversion between continuous flow and pulse dose therapy with autoSAT. Targeting FiO2s is critical to effective oxygenation, a challenge for ambulating patients due to their changing respiratory rates. eTASk also calculates battery durations based on the necessary flow rate and the patient’s breath rate.
Smartphone Version
1
Select the CAIRE concentrator the patient will be using (Eclipse or eQuinox). If using an eQuinox, select whether the 12-cell or 24-cell battery will be used.
1
2
Using the +/- buttons, input the individual patient scenario of continuous flow rate, BPM, tidal volume, and desired O2 concentration.
3
The patient’s necessary continuous flow rate will be displayed here, along with the estimated battery duration. Note that if the required continuous flow rate is above 3.0 LPM, this is not available on the Eclipse or eQuinox and will be shaded in gray.
4
Equivalent Eclipse or eQuinox pulse dose setting in mL and setting number will be displayed here along with the battery run time.
Note the pulse setting numbers are not LPM equivalents.
Battery run time on pulse settings will vary based on breath rate.
2
3
4
Contact Customer Support: Contact Technical Service:
The Americas: 800-482-2473 customerservice.usa@chartindustries.com 800-242-2473 techservice.usa@chartindustries.com United Kingdom: +44 (0) 1189 367060 France: +33 (0) 561 429 411 +33 (0) 561 429 411 Germany: +49 (0) 202 739 55420 +49 (0) 202 739 55420 Italy: +39 049 879 9601 +39 049 879 9601 Australia/New Zealand: +61 2 9749 4333 customerservice.australia@chartindustries.com Asia/Pac Rim: 770.721.7759 csasia@chartindustries.com
CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence. All rights not expressly stated herein are reserved by us, as applicable.
customerservice.europe@chartindustries.com
+44 (0) 1189 367060
techservice.europe@chartindustries.com
+61 2 9749 4333 techservice.asia@chartindustries.com
Copyright © 2014 Chart Industries ML-CONC0056 B
Loading...