AirSep Centrox PSA Technical manual

AirSep® Corporation Centrox PSA Oxygen Concentrator
Instruction Manual
MN124-1 04/10 Rev. -
AirSep® Corporation x 260 Creekside Drive x Buffalo, NY 14228-2075 USA
Phone: 1-800-320-0303 x Fax: (716) 691-1255 x www.airsep.com
Ownership Data
Take a moment to note important information below about your AirSep CENTROX PSA Oxygen Concentrator. Keep this instruction manual, along with your invoice, to serve as a permanent record of your purchase.
PSA Oxygen Concentrator
Model Number:
Serial Number:
Invoice Date:
Start-Up Date:
AirSep Representative
Company:
Contact:
Address:
CENTROX
City/Town: State: Zip:
Country: Fax:
Telephone: Telex:
Before you attempt to install, operate, or repair the oxygen concentrator, read and thoroughly understand this instruction manual. Improper operation can result in severe bodily injury, damage to the system, or poor performance.
AirSep® Corporation
1.0 Introduction ..................................................................................... 1-1
1.1 General .............................................................................................1-1
1.2 Warnings, Cautions, and Notes ........................................................1-1
1.3 References to Controls and Indicators with Labels ..........................1-1
2.0 Safety................................................................................................ 2-1
2.1 General .............................................................................................2-1
2.2 Potential Hazards .............................................................................2-1
2.3 Safety Publications ........................................................................... 2-3
3.0 System Description ......................................................................... 3-1
3.1 Introduction .......................................................................................3-1
Table of Contents
4.0 Controls, Parts, and Connections.................................................. 4-1
4.1 Introduction .......................................................................................4-1
4.2 Oxygen Concentrator Controls and Indicators.................................. 4-1
4.3 Oxygen Concentrator Parts ..............................................................4-4
4.4 Auxiliary Kits Information (Optional) .................................................4-8
4.5 Connections....................................................................................4-10
5.0 Installation ....................................................................................... 5-1
5.1 Unpacking.........................................................................................5-1
5.2 Pre-Installation Guidelines................................................................5-3
5.3 Installation Instructions .....................................................................5-4
6.0 Operation ......................................................................................... 6-1
6.1 Initial Start-Up ...................................................................................6-1
6.2 Operation ..........................................................................................6-4
6.3 Shutdown..........................................................................................6-5
6.4 Normal Start-Up................................................................................6-5
6.5 Start-Up after an Extended Shutdown ..............................................6-5
CENTROX Instruction Manual i
AirSep® Corporation
7.0 Maintenance/Service .......................................................................7-1
7.1 Weekly Maintenance ........................................................................7-2
7.2 Semi-Annual Maintenance................................................................ 7-2
7.3 Annual Maintenance......................................................................... 7-4
7.4 Pressure Switch Adjustment Procedure ........................................... 7-5
8.0 Troubleshooting ..............................................................................8-1
Troubleshooting Chart.................................................................................. 8-2
A Appendix Technical Data ............................................................... A-1
B Appendix Warranty/Returns .......................................................... B-1
C Appendix Parts List ........................................................................ C-1
D Appendix Component Literature ................................................... D-1
ii CENTROX Instruction Manual
AirSep® Corporation
Figure 3.1: CENTROX Oxygen Concentrator ........................................... 3-1
Figure 4.1: Compressor Enclosure........................................................... 4-1
Figure 4.2: PSA Enclosure Controls and Indicators .............................. 4-2
Figure 4.3: Compressor Enclosure Internal Components ...................... 4-4
Figure 4.4: PSA Enclosure Internal Components.................................... 4-5
Figure 4.5: Oxygen Receiver ..................................................................... 4-9
Figure 5.1: General Arrangement of the CENTROX concentrator ......... 5-5
Figure 6.1: PSA Enclosure Supply Valve Assembly ............................... 6-4
List of Illustrations
Figure 7.1: Pressure Switch ...................................................................... 7-6
Figure A.1: Flow Schematic ..................................................................... A-5
Figure A.2: Wiring Schematic (120 V, 60 Hz) – PSA Enclosure............. A-6
Figure A.3: Wiring Schematic (220 V, 50 Hz) – PSA Enclosure............. A-7
Figure A.4: Wiring Schematic (120 V, 60 Hz) – Compressor Encl......... A-8
Figure A.5: Wiring Schematic (220 V, 50 Hz) – Compressor Encl......... A-9
CENTROX Instruction Manual iii
AirSep® Corporation
1.0 Introduction
1.1 General
This instruction manual provides a description of the AirSep CENTROX Pressure Swing Adsorption (PSA) Oxygen Concentrator as well as instructions for its installation, operation, and maintenance. Pertinent drawings and component information are also included.
To ensure safe operation and proper system maintenance, AirSep Corporation recommends that you keep this instruction manual readily available for reference.
1.2 Warnings, Cautions, and Notes
As you read the instruction manual, pay special attention to the Warning, Caution, and Note messages. They identify safety guidelines or other important information as follows:
Provides information that can prevent severe bodily injury or death.
Provides information that can prevent minor bodily injury or property damage.
Provides information important enough to emphasize or repeat.
1.3 References to Controls and Indicators with Labels
This operating manual uses uppercase characters (e.g., ON/OFF switch) to refer to controls and indicators. Refer to section 4 for a description of all the controls and indicators of the oxygen concentrator.
CENTROX Instruction Manual 1-1
AirSep® Corporation
yg
2.0 Safety
2.1 General
Oxygen, the most abundant of the elements, makes up about 50 percent of the earth's crust. In its free state, it forms about one-fifth of our air by volume. Although oxygen is classified as a non-flammable gas, it supports combustion. As an active element, it combines directly or indirectly with all elements except the rare gases. It is an invisible gas that is colorless, odorless, and tasteless.
To ensure your safety, thoroughly read and familiarize yourself with this Safety section. AirSep Corporation strongly recommends that you review this section periodically.
2.2 Potential Hazards
Before you attempt to install, operate, or repair the CENTROX Oxygen Concentrator, read and thoroughly understand this instruction manual. Improper operation can result in severe bodily injury, damage to the system, or poor performance.
It is recommended to have an alternate source of oxygen supply if a power failure or equipment malfunction occurs.
Take extreme care to keep all the oxygen piping and vessel clean. To avoid a fire or an explosion, oxygen clean all surfaces that can come in contact with oxygen. Check all oxygen fittings/joints for leaks with an ox
en-compatible leak-detecting solution.
CENTROX Instruction Manual 2-1
AirSep® Corporation
Oxygen vigorously accelerates the burning of combustible materials. In an oxygen-enriched atmosphere, many materials that do not burn in normal air require only a slight spark or moderate heat to set them aflame.
To avoid a fire or an explosion, keep gasoline, kerosene, oil, grease, cotton fibers, paint, and any other combustible material away from any part of the CENTROX Oxygen Concentrator or optional auxiliary oxygen receiver.
Do not smoke or use an open flame near the oxygen concentrator or optional auxiliary oxygen receiver.
Post "NO SMOKING OR OPEN FLAMES" signs in the area where the oxygen concentrator and optional auxiliary oxygen receiver are located. AirSep STRONGLY recommends that only individuals trained and experienced in the safe handling of oxygen operate this system.
The interior of the CENTROX Oxygen Concentrator contains electrical parts that can produce an electrical hazard if not handled properly. To prevent electrical shock, read and thoroughly understand the Troubleshooting section of this instruction manual before you service the system.
Connect the oxygen concentrator power cord to a properly grounded wall outlet on a circuit that cannot be accidentally turned off. Do not use extension cords.
To prevent fire or electrical shock, locate the oxygen concentrator and the auxiliary oxygen receiver (if supplied) indoors away from rain or any other type of moisture.
2-2 CENTROX Instruction Manual
AirSep® Corporation
2.3 Safety Publications
Disconnect power before servicing oxygen generator.
This section is not a complete summary of required safety procedures. Review the following publications for additional information on the safe handling of oxygen:
"Installation of Bulk Oxygen Systems at Consumer Sites;" NFPA
No. 50; National Fire Protection Association; 1 Batterymarch Park; P. O. Box 9101; Quincy, Massachusetts 02269-9101 USA.
"Oxygen;" Pamphlet G-4; Compressed Gas Association; 1725
Jefferson Davis Highway; Arlington, Virginia 22202-4102 USA.
"Cleaning Equipment For Oxygen Service," Pamphlet G-4.1;
Compressed Gas Association; 1725 Jefferson Davis Highway; Arlington, Virginia 22202-4102 USA.
CENTROX Instruction Manual 2-3
AirSep® Corporation
3.0 System Description
3.1 Introduction
Air Contains 21 % oxygen, 78% nitrogen, 0.9% argon, and 0.1% other gases. AirSep Oxygen Concentrator separates this small percentage of the oxygen from the compressed air through a unique Pressure Swing Adsorption (PSA) process.
The CENTROX unit consists of two enclosures. The smaller enclosure contains two compressors that supply air for the process. Each compressor has a dedicated switch and circuit breaker. The cord from the enclosure plugs into the PSA enclosure. This cord must not be plugged into any other power supply outlet. Circuitry inside the PSA enclosure controls the power supplied to the compressors. The power cord from the PSA enclosure plugs into the wall. Figure 3.1 shows the two enclosures of the CENTROX oxygen concentrator.
PSA Enclosure
The compressor enclosure (See Figure 3.1) supplies pressurized air to the PSA enclosure (See Figure 3.1). Inside the PSA enclosure, the pressurized air enters the feed & waste manifold (See Figure 4.4). A series of valves in the feed & waste manifold controls the flow of air into each of the adsorber beds (See Figure 4.4). Valves also connect the beds to two mufflers (See Figure 4.4) that allow waste gas to be vented from the beds. The oxygen concentrator uses in its adsorber vessels an inert ceramic material called molecular sieve to separate compressed air into the oxygen and the other gases. The unique properties of molecular sieve allow it to attract, or adsorb, nitrogen physically from air under pressure.
Compressor
Figure 3.1: CENTROX Oxygen Concentrator
CENTROX Instruction Manual 3-1
AirSep® Corporation
This allows oxygen to exit the adsorbers as a product gas. Oxygen from top of the beds then flows to the check valve assembly and then to the product manifold (See Figure 4.4). The check valve assembly supplies product oxygen to the flow controller. The product manifold controls the flow of oxygen from one bed to another during various stages of the oxygen generating process. The product oxygen then flows through the product valve to the customer’s application. An oxygen analyzer continuously monitors the purity of oxygen and provides an alarm in case of low purity.
The entire oxygen generating process is completely regenerative, which makes it both reliable and virtually maintenance-free. The molecular sieve does not normally require replacement.
This instruction manual serves as the guidelines for CENTROX oxygen concentrator. Refer to the illustrations, located in the Appendix A of this instruction manual, for the detailed flow diagram and electrical schematic of the oxygen concentrator referenced in this instruction manual.
3-2 CENTROX Instruction Manual
AirSep® Corporation
4.0 Controls, Parts, and Connections
4.1 Introduction
The section describes the various parts, controls, indicators and connections required for the CENTROX oxygen concentrator.
4.2 Oxygen Concentrator Controls and Indicators
4.2.1 Compressor Enclosure
Figure 4.1 shows the compressor enclosure along with all the controls and indicators.
On/Off
Circuit Breaker
Air Intake
Hose for Compressed Air to PSA Enclosure
Figure 4.1: Compressor Enclosure
ON/OFF (Power) Switch
The individual ON/OFF switches (green color) on the compressor enclosure starts and stops the operation of the oxygen compressors. When you supply power to the system, the green indicator light of the switches turns on. It remains lit whether the switch is in the ON or OFF position (See Figure 4.1).
Circuit Breaker Reset Buttons
The individual circuit breaker reset buttons on the compressor enclosure are used to reset the compressors after an electrical overload shutdown (See Figure 4.1).
CENTROX Instruction Manual 4-1
AirSep® Corporation
g
(
)
Air Intake Filters
Located on the left side panel, the air intake filter removes any foreign particles from the air that enters the compressor enclosure (See Figure 4.1).
Fuses/Circuit Breakers, if required, must be replaced with the same type and amp rating as the original.
4.2.2 PSA Enclosure
Figure 4.2 shows the controls and indicators of the PSA enclosure.
Feed Air
Pressure Gauge
Low Purity
ht
Li
Hour Meter
Product Pressure Gauge
Low Purity Alarm
On/Off
Power
Figure 4.2: PSA Enclosure Controls and Indicators
Auto/Manual Switch
ON/OFF (Power) Switch
The ON/OFF switch starts and stops the operation of the oxygen concentrator. When you supply power to the system, the switch's green indicator light turns on. It remains lit whether the switch is in the ON or OFF position (See Figure 4.2).
AUTO/MANUAL Switch
The AUTO/MANUAL switch includes an yellow indicator light. In the AUTO position, the concentrator cycles on and off to meet oxygen demand. In the MANUAL position, the concentrator cycles continuously. The concentrator produces oxygen only while the yellow light is lit (See Figure 4.2).
Hour Meter
The hour meter indicates the total number of hours the concentrator has cycled (See Figure 4.2).
4-2 CENTROX Instruction Manual
AirSep® Corporation
Low Purity Light
The unit is equipped with an oxygen sensor. If the purity supplied to the tank is below 85 % (r 3%), the red light on the PSA enclosure will be on (See Figure 4.2).
Low Purity Alarm
If the purity is below 85 % (r 3%) for more than 30 minutes, an audible alarm will sound (See Figure 4.2).
The low oxygen purity alarm will sound for 4 seconds when the unit is turned on.
FEED AIR PRESSURE Gauge
The FEED AIR PRESSURE gauge indicates the pressure of the feed air before it enters the PSA enclosure (See Figure 4.2).
PRODUCT PRESSURE Gauge
The PRODUCT PRESSURE gauge indicates the pressure of oxygen coming out of the adsorber beds (See Figure 4.2).
CENTROX Instruction Manual 4-3
4.3 Oxygen Concentrator Parts
Figure 4.3 and Figure 4.4 shows the internal components of the CENTOX oxygen concentrator
4.3.1 Compressor Enclosure
Resonators
Heat Exchanger
AirSep® Corporation
Compressors
Check Valve
Assembly
Figure 4.3: Compressor Enclosure Internal Components
Cabinet Fans
Compressors
The compressors compress the air entering the enclosure before it is supplied to the PSA enclosure (See Figure 4.3).
Cabinet Fans
The cabinet fans provide internal cooling for the air compressor. Fans also help in better circulation of the air inside the compressor enclosure (See Figure 4.3).
Compressor Intake Filter
This filter provides additional filtration for the air as it enters the air compressor. This filter is attached to the gray PVC pipe. The complete assembly of the PVC pipe and the filter is called resonator (See Figure 4.3).
Heat Exchanger
The heat exchanger coils cool the compressed air leaving the compressor enclosure (See Figure 4.3).
Check Valve Assembly
Check Valve assembly prohibits the back flow of compressed air.
4-4 CENTROX Instruction Manual
AirSep® Corporation
r
r
4.3.2 PSA Enclosure
Coalescing Filte
Product Valve
Flow Controlle
Product
Feed Air Dump Valve
Check Valve
Feed Valves
Adsorber Beds
Pressure Switch
Product Pressure Gauge
Check Valve Assembly
Oxygen Monitor Board
Equalization
Waste Valves
Main Circuit
Exhaust Mufflers
Oxygen Sample Regulator
Figure 4.4: PSA Enclosure Internal Components
Adsorber Beds
The adsorbers contain the molecular sieve that adsorbs (attracts) nitrogen from compressed air and allows oxygen to pass through as product gas (See Figure 4.4).
Main Circuit Board
Main circuit board provides signal to all the valves based on the status of the oxygen generator.
Feed Air Dump Valve
The solenoid-operated valve is used to dump the compressed feed air for 3-5 sec at the initial start-up and 1 sec at the start of standby mode. This removes any pressure that builds up inside the compressor to ensure easy start-up.
Check Valve
Check Valve assembly prohibits the back flow of compressed air.
CENTROX Instruction Manual 4-5
AirSep® Corporation
Feed Valves
The automatic feed air valves control the flow of the feed air as the air enters the adsorbers (See Figure 4.4).
Waste Valves
The automatic waste valves control the flow of waste gas as it exits the adsorbers (See Figure 4.4).
Exhaust Mufflers
The mufflers muffle the noise produced by the waste gas that vents through the beds (See Figure 4.4).
Product Manifold
Product manifold facilitates the oxygen flow across adsorber beds (from top of the adsorber beds) and to the product delivery line.
Equalization Valves
The equalization valves on the product manifold controls the flow of oxygen from one bed to another during various steps of the oxygen generation process (See Figure 4.4).
Check Valves Assembly
The check valve assembly supplies the product oxygen to the flow controller. See Figure 4.4.
Product Pressure Gauge
Product pressure gauge mounted on the check valve assembly (See Figure 4.4) displays the pressure of the product oxygen.
Flow Controller
The flow controller regulates the flow of oxygen (See Figure 4.4).
Product Valve
The valve prevents the supply of oxygen when the unit is in standby mode (Refer to section 6 for the description of different types of mode). See Figure 4.4 for the location of the product valve.
Coalescing Filter
The coalescing filter serves as bacteria filter and removes any bacteria present in the oxygen supply (See Figure 4.4).
Oxygen Monitor Board
The oxygen sensor monitors the purity supplied from the PSA enclosure. If the oxygen purity decreases below the set point, low purity light on the PSA enclosure will be on. If the problem persists for more than 30 minutes, a low purity alarm will sound.
4-6 CENTROX Instruction Manual
AirSep® Corporation
Oxygen Sample Regulator
A regulator before the oxygen monitor board regulates the pressure to 4-5 psig to the oxygen monitor board. Refer to the section 7 for the regulator adjustment.
Pressure Switch
When the oxygen concentrator operates in Auto mode, the pressure switch monitors the oxygen pressure at the outlet of the oxygen concentrator. When the pressure at the oxygen concentrator outlet increases to the pressure switch upper setpoint, the pressure switch circuit closes and the oxygen concentrator starts a timed shutdown that stops the unit at the end of the shutdown sequence (After 10 additional cycles). When the pressure at the oxygen concentrator outlet decreases to the lower setpoint of the switch, the pressure switch opens to activate the oxygen concentrator and the oxygen production begins. When the oxygen concentrator operates in Manual mode, the pressure switch circuit remains open and the oxygen concentrator cycles continuously. Refer to the appendix A of the manual for lower and upper setpoint settings of the pressure switch.
CENTROX Instruction Manual 4-7
4.4 Auxiliary Kits Information (Optional)
The items discussed in this section are supplied as ordered. Listed below are the different starter kits available as per the oxygen concentrator ordered.
AirSep® Corporation
Unit # Kit #
(Customer
Supplied Tank)
AS074-1 120 VAC/60 Hz
AS074-2 220 VAC/50 Hz
AS074-3 220 VAC/60 Hz
KI046-1 KI046-1 TA150-1
Table 4.1: Optional Starter Kits for the Concentrator Purchased
Kit # Description
KI046-1
Primary/Secondary Ball Valve Assembly 1 Hose Assembly, Oxygen-clean, (Secondary Hose) 1
KI470-1
Regulator-Flowmeter Assembly 1 Hose Assembly, Oxygen-clean (Main Oxygen Outlet Hose) 1
TA150-1
60 Gallon Tank Assembly, O
Table 4.2: Description of the Parts included in the Starter Kits
KI470-1 KI470-1 KI046-1 KI046-1
KI470-1 KI470-1 KI046-1 KI046-1
KI470-1 KI470-1
Kit #
(AirSep
Supplied Tank)
TA150-1
TA150-1
Qty.
Cleaned 1
2
Please contact AirSep Corporation Sales representative for ordering the starter kits. Below is a brief description of item included in the starter kits.
Primary Oxygen Ball Valve
The primary oxygen ball valve controls the flow of oxygen from the oxygen concentrator to your oxygen distribution system. See Figure 5.1.
Secondary Oxygen Ball Valve
The secondary oxygen ball valve controls the flow of a backup source of oxygen to your oxygen distribution system. This ball valve should always be closed unless a backup source is connected to it. See Figure 5.1.
Regulator-Flowmeter Assembly
This assembly consists of a pressure regulator and a flowmeter. The pressure regulator attaches to the outlet port of the oxygen receiver to regulate the pressure of the oxygen. The flowmeter is used to regulate
4-8 CENTROX Instruction Manual
AirSep® Corporation
the flow of the oxygen at the outlet of the oxygen receiver. Refer to figure
5.1 for the installation location of this assembly.
Oxygen Isolation Ball Valve Assembly
The oxygen isolation ball valve stops the flow of oxygen to the oxygen receiver during troubleshooting. The oxygen relief valve in the assembly prevents excess pressure from building in the oxygen receiver if a system malfunction occurs. This assembly is shipped mounted on the oxygen receiver. See Figure 4.5.
Main Oxygen Hose
An oxygen-clean hose is provided that connects the outlet of the supply valve assembly (Refer to section 4.5) on the oxygen concentrator to the oxygen isolation ball valve assembly on the oxygen receiver.
Secondary Oxygen Hose
Secondary hose is provided to connect the outlet of the regulator­flowmeter assembly to the inlet of the primary oxygen ball valve. Refer to Figure 5.1.
Oxygen Receiver
The oxygen receiver stores oxygen produced by the oxygen concentrator. It also provides stable flow and purity for short-term surges of oxygen that exceed the rated capacity of the oxygen concentrator.
Oxygen Outlet
Port
Oxygen
Receiver
Main Oxygen
Oxygen Isolation Ball
Valve Assembly
Hose
Figure 4.5: Oxygen Receiver
CENTROX Instruction Manual 4-9
4.5 Connections
Supply Valve Assembly
This assembly is shipped loose and must be connected to the PSA enclosure oxygen outlet. The supply valve assembly limits the amount of flow to the oxygen receiver during purge. Refer to Figure 5.1 and 6.1.
Power Cord
This power cord and its grounded electrical plug supply power to the CENTROX when connected to a grounded electrical outlet.
AirSep® Corporation
4-10 CENTROX Instruction Manual
AirSep® Corporation
5.0 Installation
5.1 Unpacking
AirSep Corporation ships CENTROX Oxygen Concentrators on a wooden skid covered with corrugated cardboard. This container includes an accessory kit with an instruction manual and all the items necessary to properly install the oxygen concentrator. The oxygen receiver (if supplied) is shipped separately. Contact your AirSep Corporation’s Sales representative for ordering the auxiliary kits supplied with the CENTROX oxygen concentrator.
AirSep recommends that you follow these unpacking guidelines carefully to protect yourself against loss from any damage caused during shipment.
1. Inspect the exterior for damage. If you observe any damage, note it
on the freight bill or the express receipt before you sign it.
Failure to note exterior damage on the freight bill or the express receipt at the time of delivery can result in the carrier's refusal of a damage claim.
2. Carefully cut and remove any banding straps from the container. Then
remove the corrugated cardboard.
Remove the corrugated cardboard very carefully. You may need to return the oxygen concentrator if it was damaged during shipment
3. Remove the oxygen concentrator from the wooden skid.
4. Thoroughly inspect the oxygen concentrator interior and exterior for
damage caused during shipment. Pay special attention to the cabinet switches, gauges, brackets, etc.
5. Remove the accessory kit and inspect the contents for damage.
6. Although the CENTROX is carefully inspected, tested, and packed, it
can be damaged during shipment due to improper handling. If you find any concealed damage (loss or damage not found until the
CENTROX Instruction Manual 5-1
AirSep® Corporation
concentrator is unpacked), immediately call the delivery carrier and file a concealed-damage claim. Keep ALL container material and interior packing for the carrier's inspection.
YOU MUST MAKE A CONCEALED-DAMAGE CLAIM WITHIN 24 HOURS OF DELIVERY. Only the consignee can file this claim.
Follow these unpacking guidelines carefully to protect yourself against loss from any damage caused during shipment.
5-2 CENTROX Instruction Manual
AirSep® Corporation
5.2 Pre-Installation Guidelines
Before you install the CENTROX oxygen concentrator, and the oxygen receiver, if supplied, refer to the Specifications section in the Appendix of this instruction manual to determine the applicable space, and the power requirements for your particular model.
A backup source of oxygen must be available if a power failure or system malfunction occurs.
Make sure the area that surrounds the oxygen concentrator is well ventilated, and provide sufficient space around the unit [at least three feet] to allow for cool air flow as well as to allow safe operation and maintenance.
Locate the oxygen concentrator in an area where the ambient air temperature remains between 4 and 44
under the AirSep Corporation Product Warranty.
Connect the oxygen concentrator power cord only to a properly grounded electrical outlet on a circuit that cannot be accidentally turned off. Do not use extension cords.
Provide proper voltage to the oxygen concentrator to prevent damage not covered under the AirSep Product Warranty.
Do not plug in the power cord until you complete the installation of the oxygen concentrator.
qC (112qF) to prevent damage not covered
qC (40qF)
CENTROX Instruction Manual 5-3
p
5.3 Installation Instructions
To assure proper installation and safe operation of your CENTROX PSA Oxygen Concentrator, AirSep Corporation recommends that you review this entire section before you attempt to install the unit.
Do not turn off power to any component unless you are sure the medical facility does not require any oxygen, or there is a sufficient alternative/backup source of oxygen.
Before you attempt to install, operate, or repair the oxygen concentrator, read and thoroughly understand this instruction manual. Improper operation can result in severe bodily injury, damage to the system, or poor
erformance.
AirSep® Corporation
Refer to the Figure 5.1 as you follow the installation instructions:
1. Place the oxygen concentrator near the inlet of your oxygen
distribution system.
2. Mount both the enclosures securely to a wall, ensuring that there is a
minimum of two feet between the two enclosures and minimum of three feet between the compressor enclosure and a wall. The compressor enclosure can be on either side of the PSA enclosure.
3. Open the compressor enclosure and remove the tie wrap from each
compressor. The tie wrap holds the compressor to a bracket support for shipping purpose.
4. Attach the red air hose from the compressor enclosure to the fitting on
the left hand side of the PSA enclosure (See Figure 4.5 and 5.1).
The supply valve assembly at the outlet of the oxygen concentrator has a small hole drilled in it and allows a controlled amount of flow during purging process (See Section 6) of the 60 gallon tank (If supplied).
5. Mount the supply valve assembly (shipped loose) at the outlet of the
PSA enclosure. Refer to Figure 5.1 and 6.1.
6. Attach the main oxygen hose from the supply valve to the inlet of the
oxygen isolation ball valve assembly at the oxygen tank (See Figure
4.5 and 5.1).
5-4 CENTROX Instruction Manual
Loading...
+ 76 hidden pages