Air Cadet 07530-60, 07530-65, 400-2901 Instruction Manual

OPERATING MANUAL
DUAL HEAD AIR CADET®PUMP
MODEL NO. 07530-60 MODEL NO. 07530-65 MODEL NO. 400-2901
Cole-Parmer Instrument Company
1-800-323-4340 (U.S. and Canada only)
11 (847) 549-7600 (outside U.S.)
www.coleparmer.com
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (outside U.S.)
www.barnant.com
A-1299-0328
Edition 06
12
WARRANTY
The Manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the Manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.
The warranty period for this product is noted on the Warranty Card.
PRODUCT RETURN
To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any dam­ages resulting from improper packaging are your responsibility.
TECHNICAL ASSISTANCE
If you have any questions about the use of this product, contact the Manufacturer or authorized seller.
Printed in U.S.A.
Cole-Parmer Instrument Co.
625 East Bunker Court
Vernon Hills, Illinois U.S.A. 60061-1844
1-800-323-4340 (U.S. and Canada only)
11 (847) 549-7600 (outside U.S.)
Fax: (847) 247-2929 (U.S. and Canada only)
11 (847) 549-1700 (Fax outside U.S.)
www.coleparmer.com
28W092 Commercial Ave.
Barrington, Illinois U.S.A. 60010-2392
1-800-637-3739 (U.S. and Canada only)
11 (847) 381-7050 (outside U.S.)
11 (847) 381-7053 (Fax outside U.S.)
(847) 381-7053 (Local Fax)
www.barnant.com
MAINTENANCE
DANGER: Unplug power cord before any cleaning operation is started.
For diaphragm replacement, disassemble top part of the Pump Head. Then unscrew the diaphragm and install the new diaphragm. Reassemble the Pump Head. It is good practice to replace the Pump Head when the diaphragm is replaced. This ensures new, clean valves and maximum per­formance. Dirty valves will reduce performance.
Do not disassemble the pump housing from the motor. Proper assembly (factory-adjusted) is critical for proper pump performance.
Keep the pump enclosure clean by using a mild detergent solution. Never immerse nor use excessive fluid when cleaning the pump.
PUMP HEAD REPLACEMENT PARTS LIST
Item Part No.
Service Kit (Nitrile diaphragm, two retainers, 07530-06 two valves and plastic head)
Service Kit (FLUOREL®diaphragm, two retainers, 07530-04 two valves and plastic head)
Eccentric Assembly (eccentric and bearing) 07530-75 Set Screw for Eccentric Assembly B-1079-0121 Diaphragm Clamp B-1352-0012
NOTE:
When rebuilding both heads order two each of the required part.
SPECIAL APPLICATIONS
Technical information and advice concerning the use of this product in spe­cific applications may be obtained from our Engineering Department. If vol­ume justifies, modifications can be made to adapt the unit to special cus­tomer applications. OEM inquiries are welcome and encouraged.
112
EU Declaration of Conformity
Name of Apparatus: AIR CADET®Dual Head
Vacuum/Pressure Pump
Model Number: 07530-65 Description of Apparatus: Dual Head Vacuum/Pressure Pump
Barnant Company declares that the above model is in conformity to the following harmonized standards and directives:
Applicable Applicable Manufacturer’s Directives Specifications Report Number
73/23/EEC EN61010-1/A2:1995 TR9528 93/68/EEC
89/336/EEC EN61326-1/A1:1998 TR9529 92/31/EEC 93/68/EEC
The last two digits of the year in which the current configuration of the above model was assessed per the Low Voltage Directive is: 00.
Manufacturer: Barnant Company Division
Cole-Parmer Instrument Company 28W092 Commercial Avenue Barrington, IL 60010-2392 USA Tel.: 847-381-7050
Manufacturer’s Signature: 27 October, 2000
Larry Jones Date Vice President, Engineering