Aiphone NH-2SA-2L Instruction Manual

NHX
Nurse Center System
Dual Patient Station with separate LED annunciators for each Call Cord
- INSTRUCTIONS -
The NH-2SA-2L is modified from the standard NH-2SA/A, and provides individual LED indication of which call cord placed a call. The green LED on the left will light when a call is placed from the NHR-8A plugged into the left jack. The red LED on the right will light when a call is placed from the NHR-8A plugged into the right jack. There will be no LED indication when the master station answers the call. (Green TALK LED on a standard station.)
NOTE:
WIRING DIAGRAM:
Trunkage Line
Connection
Terminals
An NHX system installed with an NH-2SA-2L included does not carry the UL 1069 listing.
NH-2SA-2L
CN1:
Brn
Red
Org Yel Grn Blu
CN2:
Brn
Red
Org Yel Grn Blu Vio Gry
NHR-4A
Brn
Red
Org
NHR-7A
Yel
Grn
Blu
NHX-80X
11 12 13 14 15 16
21 22 23 24 25 26
~ 51 52 53 54 55 56
A+
A-
PS-2420UL
+
-
NH-2SA-2L
CN1:
Brn
Red
Org Yel Grn Blu
CN2:
Brn
Red
Org Yel Grn Blu Vio Gry
NHR-4A
Brn
Red
Org
NHR-7A
Yel
Grn
Blu
NH-2SA-2L
Corridor Lamp
(Shared by both patient stations)
Bathroom Pullcord
(Wired to one station only)
B+
B-
Aiphone Communication Systems 1700 130th Ave. N.E. Bellevue, WA 98005 (425) 455-0510 FAX (425) 455-0071
+
-
PS-2420UL
Green LED for Left call cord
(no markings above)
Red LED for Right call cord
(marked "CALL")
TOLL FREE TECHNICAL SUPPORT:
(800) 692-0200
TOLL FREE FAX LINE:
(800) 832-3765
E-MAIL: tech-serv@aiphone.com
NOTE:
Only wiring concerning the NH-2SA-2L is shown here. For complete installation and wiring information, please see the NHX Installation instructions.
NH-2SA-2L Instr.
0103JS
WIRING:
1. Follow the wiring diagram on front page of these instructions.
2. Wiring is the same as a standard patient station.
3. Plug in two NHR-8A bedside call cords.
4. Initialize the system.
OPERATION:
1. When pressing the call button plugged into the left side jack, the green LED will light, indicating a call has been placed from that call cord.
2. When the master answers the call, the LED will turn off.
3. When pressing the call button plugged into the right side jack, the red LED will light, indicating a call has been placed from that call cord.
4. The red LED will turn off when that call is answered.
5. If both call buttons are pressed, both LED's will be lit.
6. There is no indication at the patient station when the master has the patient station selected and is in communication mode. (TALK LED function is disabled.)
Aiphone Communication Systems 1700 130th Ave. N.E. Bellevue, WA 98005 (425) 455-0510 FAX: (425) 455-0071
TOLL FREE TECHNICAL SUPPORT:
(800) 692-0200
Extensions 8107, 8118, 8111, 8125
TOLL FREE FAX LINE:
(800) 832-3765
NH-2SA-2L Instr.vsd
0103JS