Agilent xCELLigence RTCA CardioECR User Manual

Application Note
Cardiovascular Research
Assessment of Cardiomyocyte Disease Models Using the Agilent xCELLigence CardioECR System
Authors
Xiaoyu Zhang, Jeff Li, and Yama Abassi Agilent Technologies, Inc.
Abstract
The Agilent xCELLigence RTCA CardioECR system combines simultaneous measurement of field potential signal using extracellular recording (ECR) electrodes as well as contractility and viability using impedance electrodes. The real-time multiplexed evaluation of the functional activity of beating cardiomyocytes using the CardioECR system allows for depicting and recapitulating disease phenotype of cardiomyocytes differentiated from patient-specific induced pluripotent stem cell (PS-iPSC) and investigating their pharmacological responses, which could potentially provide information about underlying disease mechanisms.
Introduction
B
ECR/FP
In 2006, Yamanaka and his colleagues established the first induced pluripotent stem cells (iPSC) from murine fibroblasts via transduction of four specific transcriptional factors. This novel discovery was also translated to human somatic cells and has created a new frontier in the medical field in many respects. The characteristics of iPSCs, including infinite self-renewal and multipotency, allow them to be potentially used in a wide variety of applications, including disease modeling, drug screening, and regenerative therapy. iPSCs obtained from patients maintain patient-specific genetic lesions, providing a rationale for disease modeling using patient-specific iPSCs (PS-iPSCs). Even though primary patient cells have previously been used for studying disease mechanisms, certain cell types such as neurons and cardiomyocytes, are rather difficult to obtain. The scalability of primary diseased cells and the duration of maintaining these cells in culture has also been a hurdle for investigating disease mechanisms. PS-iPSC differentiated target cells provide researchers with a stable, scalable, and physiologically relevant cell source for disease modeling.5 To date, numerous studies have reported that these target cells can recapitulate disease phenotypes similar to those observed from patients. The pathophysiological cellular phenotypes of genetically heritable heart diseases, such as cardiomyopathies and channelopathies, have been successfully modeled invitro using patient-specific iPSC-derived cardiomyocytes (PS-iPSC-CMs).6 These model systems are promising tools for understanding disease mechanisms and can potentially be used in drug discovery, personalized
1,2
It is well established that
3,4
medicine, and cardiac liability assessment in specific populations that
may carry geneticanomalies.
This application note used the
AgilentxCELLigence RTCA CardioECR
system (CardioECR system) to compare functional profiles of contractility and electrophysiology between a PS-iPSC­CM disease model and its isogenic control. After identifying baseline phenotypes at the cellular level, this study investigated pharmacological responses of diseased and isogenic cardiomyocytes to compounds with established mechanisms. As a further approach to assessing the responses of the wild type and diseased CM, both cell lines were subjected to electrical stimulation followed by treatment with isoproterenol.
components, CardioECR control unit (laptop), CardioECR analyzer, CardioECR station, and E-plate CardioECR 48 (CardioECR plate) (Figure 1A). Two sets of electrodes, interdigitated impedance (IMP) micro-electrode arrays as well as two individual field potential electrodes are integrated into the bottom of each well of the CardioECR plate (Figure 1B). Similar to other Agilent xCELLigence platforms, the CardioECR System uses IMP electrodes to measure cellular impedance, which is affected by the number of cells covering the electrode, the morphology of the cells, and the extent of cell attachment. The fast sampling rate of
IMP measurement (2ms) allows for
capturing of the temporal rhythmic changes in cell morphology and degree of cell attachment to the plate,
Assay principle
The xCELLigence RTCA CardioECR system is a dual-mode instrument that includes both simultaneous monitoring of hiPSC-CM viability, contraction, and field potential (FP) in real time as well as directed electrical pacing of hiPSC-CMs. It consists of four
A
Figure 1. (A) The Agilent xCELLigence RTCA CardioECR system consists of four components: Control unit/laptop, CardioECR analyzer (left), CardioECR station (right), and E-PlateCardioECR 48. (B) A close-up image of E-Plate CardioECR48. Inset: a close-up of the wells reveals the layout of the electrodes, impedance (IMP) electrode assays, and two field potential (FP) electrodes.
which is a hallmark of contraction of cardiomyocytes. Therefore, the physical contraction of cardiomyocytes is monitored and recorded in real time with high temporal resolution. Additionally, the millisecond time resolution can be performed at regular intervals over a prolonged duration of
IMP
2
time providing beating and viability
Da
information of cardiomyocytes in real time. Furthermore, two field potential electrodes are used to measure integrated ion channel activities at a data acquisition rate of 10 kHz simultaneously while IMP recording is taking place via IMP electrodes. A typical workflow of cardiomyocyte assay using CardioECR system is shown in Figure 2.
One of the other critical features of the CardioECR system is its ability to electrically pace cardiomyocytes. During electrical pacing, electrical pulses are directly applied to the cells through the IMP electrodes. For most cardiomyocytes, the length of each electrical pulse employed by the IMP electrode is less than 2 ms, which allows the contractile activities of cells to be immediately captured and recorded
while the cells are being paced by IMP electrodes. The optimal conditions for electrical pacing are dependent on the cell type, the inherent beating frequency, and the experimental context. The pacing functionality of the CardioECR system is suitable for both acute pacing regimens to evaluate compound effects on contractility under controlled beating rate and long-term stimulation used for the improvement of functional maturation of hiPSC-CMs.
Raw data collected from IMP electrodes and ECR electrodes were analyzed offline using Agilent xCELLigence RTCA CardioECR data analysis software. The software provides users with more than 25 analysis parameters based on IMP and FP signals to assess cardiac cell beating and electric activities. The fundamental parameter for contractility
using cellular impedance measurement are beating amplitude (BAmp) and beating rate (BR). The beating rate is defined as the number of beats per unit of time and is expressed as beats/minute. BAmp is defined as the absolute (delta) Cell Index (CI) value between the lowest and highest points within a beating waveform (Figure 3A). For analysis of FP signal, the FP spike amplitude (FP-Amp) is derived, which is the absolute (delta) value in mV from the lowest point of the initial spike to the highest point of the spike. The FP Duration (FPD) is defined as the period between the negative peak of the FP spike to the maximum or minimum point of the reference wave. The reference wave can be negative or positive depending on how the cells are situated to the FP electrodes (Figure 3B).
Cell viability
Thaw cellsand plateinto
E-plateCardioECR 48
ys in culture
Figure 2. (A) The workflow of hiPSC cardiomyocyte assay using the Agilent xCELLigence RTCA CardioECR system: cell seeding on day 0, start to measure cell performance on day 3, start cell status QC before compound addition, add compounds to the cells if they pass QC. Alternatively, after cells start to generate stable and robust functional activity, approximately 5 days after seeding, progressive electrical pacing is applied to the cells for consecutive 15 days to achieve functional maturation before compound addition.15 After treatment, cell viability, contraction, and electrophysiology are evaluated via IMP and FP readouts measured by CardioECR system.
0
Signal acquisition
3
Medium change
Cell status QC
5 to 14
Progressive
electrical pacing
Compound
addition
Contractility/IMP
Cell Index
(impedance)
ECR(mV)
Fieldpotential
3
Materials and methods
Beating amplitude
A
Cell culture
Patient-specific iPSC-derived cardiomyocytes (PS-iPSC CMs) were purchased from FUJIFILM Cellular Dynamics International (FCDI, Madison, WI, USA). The cells were stored in liquid nitrogen until thawed and cultured according to manufacturer instructions. Briefly, each well of the E-Plate CardioECR 48 (Agilent Technologies, Santa Clara, CA, USA) was coated with 50 µL of a 1:100 diluted fibronectin (FN) solution at 10 µg/mL (F1114, Sigma-Aldrich, St. Louis, MO, USA) and
incubated at 37 °C for at least 1hour.
Next, the fibronectin solution was replaced with 50 µL of prewarmed iCell cardiomyocyte plating medium. Cells were thawed and diluted in prewarmed plating medium at the manufacture’s recommended concentration. 50 µL of the cell suspension was transferred using a multichannel pipette and seeded directly onto a precoated E-Plate CardioECR 48 (20,000 viable and plateable cells/well) in a laminar hood. The plates containing PS-iPSC CMs were kept in the hood at room temperature for 30 minutes, then placed and cultivated in a humidified incubator with 5% CO2 at 37 °C. The plating medium was replaced with iCell cardiomyocyte maintenance medium 48 hours postseeding. A medium change was performed every other day afterward.
Chemical reagents
All the chemical reagents were purchased from Tocris (Minneapolis, MN, USA), Sigma-Aldrich (St. Louis, MO, USA), or provided by the Chemotherapeutic Agents Repository of the National Cancer Institute. The 1,000-fold chemical stock solutions were prepared in DMSO and stored at –20 °C. The serially diluted chemicals (1,000-fold) were further prepared in DMSO immediately before compound addition. The 10-fold final dilution of the chemicals was prepared with the culture medium for a single-time use only. The final concentration of DMSO in the treated well was 0.1%.
Results and discussion
Evaluation Brugada channelopathy model using CardioECR system
Brugada syndrome (BrS) is one of the more common forms of familial arrhythmic syndromes characterized by a dynamic or persistent ST segment elevation, an enhanced risk of syncope, and sudden cardiac death in young adults without structural heart disease. genes identified so far, L-type Ca2+ channelopathy of loss-of-function mutations in the CACNA1C (Cav1.2a1) has been reported to give rise toBrS Type 3(BrS3).
7,8
Among several responsible
8,9
This study used used BrS3 cardiomyocytes (BrS3 CMs) purchased from Fujifilm CDI (Fujifilm CDI, part number R1136). This BrS3 CM line possesses a change of amino acid 490 from glycine-to-arginine (G490R), which was genetically engineered into the genome of iCell cardiomyocytes (iCell CMs) derived from a healthy donor with no known disease-related genotypes. iCell CMs (Fujifilm CDI, part number R11320) were used as the isogenic control for BrS3 CMs. To investigate the functional consequences of the mutation, the BrS3 CMs and WT iCell CMs were seeded first on the same E-Plate CardioECR 48 at the same seeding density (20,000 cells/well). The cell performance, including attachment, growth, and functional activities were monitored and recorded immediately after cell seeding. Figure 4A shows that BrS3 had slower kinetics of cell attachment and spreading than the control line reflected by the lower slope at the exponential phase of the Cell Index curve. Overall, Cell Index values of BrS3 were generally smaller than the control line. In addition to the attachment and growth profiles, the excitation-contraction profiles of the BrS3 CM and WT iCell CM displayed significant differences. As shown in Figures 4B and 4D, the BAmp of BrS3 CMs was smaller than control cells by 27 ±13% while the BR of BrS3 was faster
Figure 3. Definition of main parameters used to evaluate cell contractile and field potential activities. (A) The typical IMP waveform/contraction pattern and (B) the typical field potential (FP) waveforms were obtained from diseased and WT CM. The reference point can be negative or positive.
+
–––
4
+
B
Loading...
+ 8 hidden pages