The information in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including, but not limited to, the implied warranties of merchantability and fitness for a
particular purpose. Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the furnishing, performance,
or use of this material.
This document contains or refers to proprietary information which is protected by
copyright. All rights are reserved. No part of this document may be photocopied,
reproduced, or translated to another language without the prior written consent of Agilent
Technologies Company.
Responsibility of the Manufacturer
Agilent Technologies only considers itself responsible for any effects on safety, reliability
and performance of the equipment if all the following are true:
• Assembly operations, extensions, re-adjustments, modifications or repairs are done by
persons authorized by Agilent Technologies.
• The electrical installation of the relevant room complies with the IEC or national requ irements.
• The instrument is used according to the instructions for use presented in this manual.
As with all electronic equipment, radio frequency interference between this
cardiograph and any existing RF transmitting or receiving equipment at the
installation site, including electrosurgical equipment, should be evaluated
carefully and any limitations noted before the equipment is placed in service.
Monitoring during electrosurgery should not be attempted and monitoring
electrodes should be removed from the patient to preclude the possibility of
burns. Radio frequency generation from electrosurgical equipment and close
proximity transmitters may seriously degrade cardiograph performance. Agilent
Technologies assumes no liability for failures resulting from RF interference
between Agilent medical electronics and any radio frequency generating
equipment at levels exceeding those established by applicable standards.
ii
Notice
CAUTION
CAUTION
Like all electronic devices, this cardiograph is susceptible to electrostatic discharge
(ESD). Electrostatic discharge typically occurs when electrostatic energy is transferred to
the patient, the electrodes, or the cardiograph. ESD may result in ECG artifact that may
appear as narrow spikes on the cardiograph display or on the printed report. When ESD
occurs, the cardiograph’s ECG interpretation may be inconsistent with the physician’s
interpretation.
ESD discharges to expose d m etal on th e rear of the cardiograph can o ccasion ally cau se an
error message to appear on the cardiograph display. The cardiograph returns to normal
operation after turning the power off, then on again.
The data transmission cable must have a suppression device attached to assure the
cardiograph’s compliance with the European Radiated Emissions Standard found in
CISPR 11. If your data transmission cable does not include a suppression device,
compliance can be achieved by attaching one of the following suppression devices to the
cable, near the cardiograph:
part number 0443164251Fair-Rite Products Corporation
P. O. Box J
One Commercial Row
Wallkill, New York 12589
telephone: (914) 895-2055
FAX: (914)895-2629
or
Euro-Schaffner, S. A.
1 B Avenue de Suisse - BP 16
68311 Illzach Cedex, France
telephone: 33-8-931-0400
FAX: 33-8-931-0401
part number 28B2025-0A0 Steward
East 36th Street
P. O. Box 510
Chattanooga, TN 37401
telephone: (615) 867-4100
FAX: (615) 867-4102
or
Steward EMC, S.A.
Rue Fritz-Couvoisier 40
Ch-2300 La Chaux-de-Fonds
Switzerland
telephone: 41-39-282-387
Fax: 41-39-280-277
This is to certify that this equipment is in accordance with the Radio Interference
Requirements of the EMC Directive.
iii
Notice
Medical Device Directive
The PageWriter M1770 series of cardiographs comply with the requirements of the Medical Device Directive 93/42/EEC and carries the
Agilent Technologies warrants this medical product against defects in materials and
workmanship for a period of three years in certain geographics, or one year with onsite
support.
If Agilent Technologies receives notice of such defects during the warranty period,
Agilent T echn olog ies shall, at its option, either repair or replace hardware pr odu cts which
prove to be defective.
Agilent Technologies software and firmware products that are designated by Agilent
Technologies for use with a hardware product, when properly installed on that hardware
product, are warranted not to fail to execute their programming instructions due to defects
during the warranty period. Agilent Technologies shall repair or replace software media
and firmware that do not execute their programming instructions due to such defects.
Agilent Technologies does not warrant that the operation of the software, firmware, or
hardware shall be uninterrupted or error free.
If Agilent Technologies is unable, within a reasonable time, to repair or replace any
product to a condition as warranted, Buyer shall be entitled to a refund of the purchase
upon return of the product to Agilent Technologies.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulti ng from any of the following:
1. Improper or inadequate maintenance by Buyer.
2. Buyer-supplied software or interfacing.
3. Unauthorized modification or misuse.
4. Operation outside of the environmental specifications for the product.
5. Improper site preparation and maintenance.
iv
Notice
THE WARRANTY SET FORTH ABOVE IS EXCLUSIVE AND NO OTHER
WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED.
AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
Printing History
October 1994Edition 1
May 1996Edition 2
October 1998Edition 3
December 1999Edition 4
v
Safety Summary
0Safety Summary
Safety Symbols Marked on the Cardiograph
The following symbols are used on the cardiograph or the cart:
Caution - See operating instructions
Alternating current.
Meets IEC type CF leakage current requirements and is defibrillator
protected. (Isolated ECG input.)
Equipotential (identifies independent protective earth conductor to the
cardiograph).
Fuse.
Indicates power control for cardiograph.
HzIndicates operating frequency in cycles per second.
The maximum weight that the cart can hold
Displays the configuration menu on the PageWriter
200/200i/300pi.
Please see Chapter 2, “Performance Verificati on and Maintenance,” for safety
requirements that apply to the cardiograph.
vi
Conventions Used in This Manual
WARNING
CAUTION
NOTE
0Conventions Used in This Manual
Warning statements describe conditions or actions that can result in personal
injury or loss of life.
Caution statements describe conditions or actions th at can result in damage to the
equipment or software.
Notes contain additional information on cardiog raph usage.
TEXT
.H\
Softkey
represents the labels that appear on the display.
represents keys on the key panel.
represents the temporary key labels that appear on the display.
vii
Preface
0Preface
This manual contains service information for the Agilent M1770A pageWriter 300pi,
PageWriter 200i, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs.
The information and procedures in this manual apply to all models unless otherwise
specified.
This manual is organized as follows:
• Chapter 1 — Introduction. Contains a gener al description of the cardiogr aphs, lists of
technical specifications, and lists of accessories and options.
• Chapter 2 — Performance V erification and Maintenance. Explains how to check the
cardiograph’ s perf ormance using buil t-in self-tes ts, and lists main tenance procedures and
safety requirements that apply to the cardiograph.
• Chapter 3 — Theory of Operation. Provides an overview of how the cardiograph
works and describes the operation of the m a jor subassemblies.
• Chapter 4 — Troubleshooting. Contains procedures and error codes to aid the service
person in localizing faults to a replaceable subassembly.
• Chapter 5 — Removal and Replacement. Contains procedures for removing and
replacing each of the cardiograph’s major subassemblies.
• Chapter 6 — Parts List. Lists part numbers for the cardiograph’s replaceable p arts, and
provides assembly drawings.
• Appendix A — Connector Pin Assign ment s. Identifies and defines the signals
This chapter introduces you to the M1770A PageWriter 200i/300pi, M1771A PageWriter
200, and M1772A PageWriter 100 cardiographs and lists their technical specifications.
These cardiographs are only to be serviced by Agilent qualified personnel.
Safe and effective use of medical instrumentation requires periodic inspection and
preventive maintenance. Perform the preventive maintenance procedures in Chapter 2 of
this manual at required intervals to ensure satisfactory instrument performance.
The cardiographs use a thermal printhead to record waveforms and label the ECG report.
The paper supplied with the cardiographs is a thermal paper designed to work with this
printhead and with the photo detector used to advancing the paper.
Agilent Technologies guarantees the performance of the cardiographs only when used
with Agilent Technologies supplies, accessories, and paper that meet or exceed Agilent
Technologies specifications.
0PageWriter 100, 200, 200i, and 300pi Series Cardiographs
The M1770A PageWriter 200i/300pi is Agilent Technologies’ economical, interpretive
cardiograph. The M1771A PageWriter 200 is the economical, non-interpretive
cardiograph. The M1772A PageWriter 100 is the most economical, non-interpretive
cardiograph. The cardiograph contains the user controls, the printer, and all the processing
circuitry. All models use the same enclosure. The M1705B cart is designed for these
cardiographs. Figure 1-1 shows the M1770A mounted on the cart. Option #A05 provides
transmission and storage capability for the M1770A and M1771A. The StressWriter
option provides capability for communication between a PageWriter 100/200/200i/300pi
cardiograph and a StressW riter system.
1-1
PageWriter 100, 200, 200i, and 300pi Series Cardiographs
The PageWriter 100, 200, and 300 family of cardiographs print ECGs in Auto and
Manual formats. All PageWriter cardiographs use a continuous feed, high-resolution,
thermal array printer. This produces electrocardiograms on standard- size paper that can
be included in patient records without cutting or mountin g. ECG reports clearly show the
ECG waveforms. The PageWriter 200 and 300 series cardiographs also annotate records
with patient information and ECG measurements. The PageWriter 200i and 300pi also
includes ECG interpretation for both adult and pediatric patients.
From a service standpoint, the PageWriter 100, 200, and 300 family of cardiographs are
similar except for two major hardware differences:
• The M1770A has additional firmware for ECG measurements and interpretation.
• The M1771A has additional firmware for ECG measurements.
• The M1772A keypanel do es no t hav e an L CD preview dis play and uses operation al k eys
instead of alphanumeric keys for user input.
All other model differences are found in the cardiograph’s software.
The PageWriter 100, 200, and 300 family of cardiographs is designed for low-cost, longterm reliability. The modular design makes extensive use of VLSI and gate array
technology, resulting in a minimum number of subassemblies. The modular approach
means less down-time for the user, since replacing subassemblies allows faster field
repairs. The internal self-test efficiently identifies faulty subassemblies, further speedin g
the repair process.
1-2
Inquiries
M1705B Cart
The optional M1705B Cart provides mobility for the PageWriter 100, 200, and 300 family
of cardiographs. The large wheels make the cart easy to move, yet steady. Slots in the
cardiograph’s feet and a thumbscrew secure the cardiograph to the cart rails. The tray
directly below the cardiograph provides storage for the user’ s guide. Storage for the power
cable is built in. Other built-in cable retainers hold the patient data cable ou t of the way.
Two compartmented trays provide storage for spare patient electrodes and patient cables,
additional thermal paper, and consumable supplies.
Transmission and Storage, Option #A05
Transmission and storage features are available for the M1770A and M1771A with
purchase of Option #A05. Up to th irty Auto ECGs can be stored in the car diograph’s
internal flash memory. Stored ECGs can be recalled later for editing, re-analyzing,
printing, or transmi ssi on to another PageWri te r 200 /2 00i/300pi equipped with Opt i on
#A05, a PageWriter XLi cardiograph, a TraceMaster ECG Management System, or
facsimile machine. Manual ECGs cannot be transmitted or st ored.
1Inquiries
Refer any questions or comments regarding these instruments to the nearest Agilent
Technologies Sales/Service Office or to one of Agilent Technologies’ Service Dispatch
Centers. Always identify the instrument by model number and serial number in all
correspondence. Telephone numbers for Service Dispatch Centers and Sales/Service
Offices are listed in Chapter 6, “Parts List.”
1Specification Data
The following tables list the technical specifications for the cardiographs and the mobile
cart. Specifications are the same for all models except as noted
Table 1-1Physical Specifications
Parameter Specification
Dimensions (h × w × l)
× 15 in × 17 in (11.5 cm × 39.2 cm × 43.7 cm)
cardiograph
cart
Weight
cardiograph
cart
4 in
× 17 in × 33 in (91.4 cm × 43.5 cm × 84.8 cm)
36 in
18.7 lbs. (8.5 kg) or less (includes batt ery, 200 sheets of paper, patient cable, and po w er co r d)
38 lbs. (17.24 kg)
Chemical resistance,
cleaners
Withstands the following: isopropyl alcohol (except patient cable), mild soap and water, chlorine
bleach and water (30 ml/l of water).
1-3
Specification Data
Table 1-2Electrical Specifications
Parameter Specification
Resolution
ECG (internal)
display
5 µV
128 row pixels by 24 0 column pixels
display sweep sp eed
Input impedancegreater than 2.5 MΩ@ 10 Hz, typical ly gr eat er than 100MΩ @ DC
Gain accuracy
Input biasless than 40 nA for in put leadwire, less th an 500 nA for Right Leg output lead-
Common mode rejec t io n110 dB or greater with AAMI test ci r cui t
Defibrillator recovery
Crosstalk rejection
Sample rate4.096M samples per second per input leadwire (unfiltered)
DC offset tole rance
Noise
Standardizing vo lt age
50/60 Hz notch filter50/60 Hz AC line rejection filters are always active.
Pacemaker pulse display> 0.2 mV indica tion for pace puls es be tween 0.5 an d 2 ms dur ati on at amplit ude
23.1 mm/sec
±5% of input signal or ±40 µV, whichever is greater
wire
System recovery 8 seconds after 360 j o ule discharge, no dam age
Less than 2% channel crosstalk
≥ ±300 mVdc with less than 5% gain change
≤ 30 µV peak-to-peak RTI
1 mV
2 to 250 mV
± 1%
± 5% for 100 mSec ± 5 mSec
Sampling Characteristics of Cardiograph
Physiological factors such as breathing can cause variation in amplitudes of heart beats,
independent of the heart rate. In addition, sampled systems (as op posed to continuous
systems) may show changes in the apparent height of the R-wave when sampling of the Rwave occurs slightly off-peak. Because the sampling rate and the heart rate are
asynchronous, the time between the peak and an adjacent sample can vary from one QRS
complex to the next. This results in a slight variation of displayed QRS amplitu de. This
effect is more pronounced wit h narrower signals, more commonly foun d in some pediatric
electrocardiograms and in pacemaker pulses. The M177XA family of cardiographs
minimizes this effect by:
• using an integrating-type A/D converter
• sampling all leadwires simultaneously
• oversampling the signals
• processing the oversampled signals with appropriate digital signal processing techniques
1-4
Specification Data
Table 1-3ECG to Paper Specifications
ParameterSpecification
Frequency respons e of PageWriter
200/200i/300pi
(-3dB bandwidth)
Auto ECG
Manual ECG
Frequency response of PageWriter 100
Auto ECG
Manual ECG
ECG resolution on paper
voltage axis
time axis
Recorder speed5, 10, 25, and 50 mm/sec + 1.5%
ECG visibility with pace pulseMeets AMMI EC11-1991 standard for Diagnostic Electrocardiographic
<3 dB down at (user selectable):
<3 dB down at (user selectable):
Approximately 8 dots/mm (200 dots/in)
Approximately 20 dots/mm (500 dots/in) at 25mm/sec and slower
Approximately 10 dots/mm (250 dots/in) at 50mm/sec
Devices
High PassLow Pass
0.05 Hz
0.15 Hz
0.5 Hz
0.05 Hz
0.15 Hz
0.5 Hz
0.15-150 Hz or 0.15-40 Hz
0.05-150 Hz or 0.05-40 Hz
150 Hz
100 Hz
40 Hz
150 Hz
100 Hz
40 Hz
1-5
Specification Data
Table 1-4Power Supply and Battery Specifications
Parameter Specification
AC line frequenc y
115 V setting
230 V setting
AC input power rating
Battery voltage6 V DC, 6.5 Ah
Battery recharge time with unit off
(battery fully discharged)
to 90% capacity
to full capacity
Battery capacity
Auto
continuous rhythm (Manual mode)
Battery life
AC and battery operation
battery only operation
Low battery warningAt least 2 Auto ECGs or at least 2 minutes of Man-
Table 1-5Safety Specifications
Parameter Specification
50 and 60 Hz (nomina l)
90–132 Vac
198–264 Vac
50 VA
7 hours typical
16 hours typical
40 ECGs
40 minutes
Typically 24 months
Typically 14–16 months
ual ECG is allowed after the Low Battery indication is given.
ECG leads source current to groundLess than 10 µA RMS
ECG leads sink current
120 Vac, 60 Hz
240 Vac, 50 Hz
Ground wire integrity
Shunting of defibrillator energy
Operator safety during defibrillation
Less than 20 µA RMS with patient cable
Less than 50 µA RMS with patient cable
Less than 100 mΩ @ 25 A AC
< 10%
< 100 µC
1-6
Specification Data
Table 1-6Environmental Specifications
ParameterSpecification
Temperature
operating
storage
Humidity
operating
storage
Pressure (altitude)
operating
storage
Table 1-7Miscellaneous Specifications
ParameterSpecification
Real time clock accuracy
10 to 40
° C
50
° to 104° F (10° to 40° C)
° to 122° F (0° to 50° C)
32
15 to 80% RH, non-c ondensing
15 to 90% RH, non-c ondensing
15,000 ft. (4600 m) for 2 hours
15,000 ft. (4600 m)
Less than 3 minutes deviation per month.
Printhead lifeTypically 100,000 pages
Table 1-8Storage and Transmission Specifications (Option #A05 only)
ParameterSpecification
Auto ECG storage ca pacity30 ECGs
Transmission Protocols
DT
SCP
Modem command interfaces
Data modem
FAX modem
Modem protocols
Modulation
Error corre c tion
Compression
FAX modulation
Data transmission standard use d betwe en Agile nt Pag eWriter
cardiographs and A gi le nt TraceMa st er ECG Management Systems
Standard Communications Protocol
Standard AT command set
EIA/TIA-578 Service Class 1
V.34
V.42
V.42 bis
V.17
1-7
Options and Accessories
1Options and Accessories
PageWriter 100, 200, 200i, and 300pi Cardiographs
These tables list the options and accessories available for the PageWriter 100, 200, and
300 family of cardiographs.
Country/Region Options
Each country/region option includes the appropriate power cord, voltage, printer, patient
cable, and language. See T ab le 1-9, which shows the configuration of each country/region
option.
A05Adds storage and transmission capability
M2488A-#A70, A71, A72StressWriter system includes PageWriter cardiograph
M2488A-#C70StressWriter system includes interface card for PageW riter
Standard Accessories
Accessories included are based on model number and localization.
• Using the PageWriter 200/200i Cardiograph Operator Training Video (PageWriter
200/200i, also for use with PageWriter 300pi)
• Using the PageWriter 100 Cardiograph Operator Training Video (PageWriter 100 only)
1-12
Performance Verification and Maintenance
NOTE
1Introduction
This chapter describes how to verify the cardiograph’s performance, explains preventive
maintenance, presents patient safety information, and prov ides a p erfo rmance verification
checklist.
1Performance Verification
Several procedures make up performance verification: visual inspection of the
cardiograph exterior, execution of Extended Self-test, visual examination of an ECG
recorded from an ECG patient simulator, and system safety tests using a safety analyzer.
A Performance Verification matrix which appears at the end of this section specifies the
tests and inspections which must be performed following servicing of th e PageWriter
cardiographs. The Performance Verification test results must be recorded on
Customer Service order records
Make copies of the matrix. Fill out a copy each time the cardiograph is tested. Attach the
printer test output, and simulator ECG trace to the completed matrix and file with the
cardiograph’s permanent maintenance record.
If any of the performance verification tests fail, refer to Chapter 4, Troubleshooting.
2-1
Performance Verification
Visual Inspection and Power On Self Test
Before beginning the inspection, p res s the key to put the cardiograph in
2Q6WDQGE\
Standby mode, and unplug the power cord from the wall outlet. Inspect the cardiograph
for the following:
• Worn or damaged power cord
• Loose or missing hardware
• Mechanical damage
• Evidence of liquid spill
• W orn printer drive gear
• Worn printer roller
• Corroded or damaged reusable electrodes, if present
• Damaged patient cable
• Dirt/paper residue on the thermal printhead
• Frayed or damaged wiring
Replace any damaged or missing items, and clean the printhead and patient electrodes as
necessary. Cleaning instructions are listed under “Preventive Maintenance” later in this
chapter. Connect the cardiograph to AC power. Be sure that the AC indicator comes on.
Turn the cardiograph On and observe that the display turns on (PageWriter
200/200i/300pi) or that the LED’s flash on (PageWriter 100).
Extended Self-test
The cardiograph’s extended self-test checks each major subassembly. The extended selftest consists of two sections: the testing of internal circuitry and the testing of printer,
display, keyboard, and modem capabilities.
To begin the extended self-test, press and hold both the and key s while
turning on the cardiograph with thekey.
2Q6WDQGE\
T o st op or exit extended self-test, yo u must press to p lace the cardiogr aph in
$XWR0DQXDO
2Q6WDQGE\
Standby.
Internal Circuitry Testing of PageWriter 100
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each of these tests is described in detail later in this chap ter.
When the internal circuitry test of the PageWriter 100 is performed, the LEDs show which
test is being performed. The following table shows which LED is associated with each
test.
2-2
Performance Verification
Table 2-1PageWriter 100 Test LED’s
Test Number Test Name LED(s) lit on Page-
Writer 100
1ROM-15 mm/mVOK
2ROM-210 mm/mVNormally fails for 100.
3ROM-X120 mm/mVOK for A#10.
4ROM-X2V LeadsOK for A#10.
5RAMFilterOK
6NVRAMAutoOK
7GAI II III Lead GroupWill also flash all LEDs. OK
8ECG FEaVR aVL aVF Lead GroupOK
9SIOII aVF V2 Lead GroupOK for A#10.
AREMOTEV LeadsFor future use - normally fails.
BMODEM5 mm/secOK for A#10, with modem or test tool
CFLASH110 mm/secOK for A#10.
DFLASH225 mm/secOK for A#10.
When the tests are completed, the results are also printed out in a pass/fail format
on the left-hand side of the first page of the test report.
Comments/ Ex pected Result s
attached
Table 2-2Test Results Format
Test Passes Test Fails
Printed ResultOK* *
In some cases, a test failure halts the test/test printout and an error code ind icatin g the
failure is flashed on the LED’s. See Chapter 4, “Troubleshooting,” for a list of self-test
failure codes and failure symptoms. The section “How the PageWriter 100 Communicates
Error Codes and Messages” in Chapter 4 describes how to read the LED’s.
2-3
Performance Verification
Internal Circuitry Testing of PageWriter 200/200i/300pi
These tests are performed on the internal circuitry:
•Memory
• Gate Array
• Front-End Interface
• Serial Interface
Each test is described detail later in this chapter.
When testing the internal circuitry of the PageWriter 200/200i/300pi, the numbers 1
through 9 and the characters A through D are displayed one at a time beginning with
number 1. Each number represents a particular test. If the test fails, the character X is
placed after the test number, f or example, 3X. The test descriptions below are listed in the
order the tests are performed.
Table 2-3PageWriter 200/200i/300pi Test Numbers
Test Number Test Name Character Displayed on Page-
Writer 200/200i/300pi
1ROM-11OK
2ROM-22Passes for 300p i, 200i, and 200 with
3ROM-X13Passes for Options #A05 and #A10
4ROM-X24Passes for Options #A05 and #A10
5RAM5OK
6NVRAM6OK
7GA7OK
8ECG FE8OK
9SIO9Passes for Options #A05 and #A10
AREMOTEAFor future use - normally fails
BMODEMBPasses for Options #A05 and #A10
Comments
software revision A.05.06 or later.
only.
only.
only.
only, with modem or test tool
attached
CFLASH1CPasses for Options #A05 and #A 10
only
DFLASH2DFor future use - normally fails
2-4
Performance Verification
CAUTION
If your cardiograph is equi p ped wi th Opt i on s # A05 or #A10, do not turn it off during Test
C or D. If you turn off the cardiograph while Test C or D are in process, you could lose
stored ECGs, the Log of ECGs Stored, or the Log of ECGs Taken.
When the tests are completed, the results are printed out in a pass /fai l format on the lefthand side of the first page of the test report.
Table 2-4Test Results Format
Test Passes Test Fails
Printed ResultOK* *
In some cases, a test failure halts the test/test printout and an error code ind icatin g the
failure appears on the display. See Chapter 4, “Troubleshooting,” for a list of self-test
failure codes and failure symptoms.
Memory Test
This test looks for failures in the following memory subsys tems:
• Read-Only Memory (ROM)
• Random Access Memory (RAM)
• Non-Volatile Random Access Memory (NVRAM)
Read-Only Memory (tests 1-4) . The cardiograph is designed to have up to four ROMs
installed, two on the main control board and two on an option board. The Read-Only
Memory test reads the contents and performs a 32-bit cyclic redundancy check (CRC) of
each of the four ROM address spaces. The test of a ROM address space will fail if the
ROM module is faulty or is not present. The four ROM address spaces are identified on
the report as
•ROM-1
•ROM-2
•ROM-X1
•ROM-X2
Depending on the serial number and model of the unit you are servicing, the
corresponding ROM part numbers vary. Refer to tables 2-5, 2-6, 2-7, and 2-8 for the
correct ROM replacement part numbers.
2-5
Performance Verification
Table 2-5ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number
Prefix CNA, CNB, or CNC, XXXA or , US001
Test Part NumberLanguagePart Number
Labeled on
ROM
ROM 1
ROM 2U206 - Main CPU BoardM1770-89521
ROM X1U9 - Interface BoardM2488-17901
ROM X2U10 - Interface BoardM2488-17902
Table 2-7ROM Replacement Part Numbers for PageWriter Cardiographs M1772A Serial Number
Prefix CNA, CNB, or CNC, XXXA or US001
Test ROM Part Number
ROM 1
ROM 2Not Required
ROM X1U9 - Interface Boa rdM2488-17901
ROM X2U10 - Interface BoardM2488-17902
U205 - Main CPU Boar dM 1770-89505
Table 2-8ROM Replacement Part Numbers for PageWriter Cardiographs M1772A Serial Number
Prefix CND, or US002
Test ROM Part Number
ROM 1
ROM 2
ROM X1U9 - Interface Boa rdM2488-17901
ROM X2U10 - Interface BoardM2488-17902
U205 - Main CPU Boar dM 1770-89550
If ROM-1 is defective, it is unlikely this test or any other cardiograph function will
operate.
Random-Access Memory (test 5) . The RAM test performs a read/write test of the
instrument’s memory. If this subsystem is defective, it is unlikely this test or any other
cardiograph functions will operate. The RAM is identified on the report as
•RAM
2-7
Performance Verification
Non-Volatile Random Access Memory (test 6) . Each 2-byte location of NVRAM is
written and read with a pseudo-random pattern. This is a non-destructive test. The
contents of the NVRAM are saved in RAM prior to this test, and then restored when the
test is complete. The NVRAM test result is identified on the report as
• NVRAM
System Gate Array Test (test 7) . Various registers in the System Gate Array are read
and written. The System Gate Array test result is identified on the report as
• GA
Front-End Interface Test (test 8) . The front-end interface test verifies that the ECG
front-end circuitry is responding to commands, and that the input multiplexers are
operating between ground and internal reference voltage. The front-end interface test
result is identified on the report as
•ECG FE
Manufacturing/Serial Interface Test (test 9) . The manufacturing/serial interface test
checks for proper communication with the optional expansion card UART. For this test to
pass, an extender card must be plugged into the expansion connector. The serial interface
test result is identified on the report as
•SIO
Remote Test (test A). The remote test checks the ability to transmit and receive data
through the spare remote port on the optional expansion board. The remote test is
identified on the report as
• REMOTE
Modem Test (test B) . The modem test verifies that the cardiograph is sending the
expected signals to the modem, and that the modem is receiving and responding
appropriately to signals from the cardiograph. The optional expansion board must be
installed to connect to the modem. The front end and RS-232 port test tool described in
Chapter 4 can help differentiate modem and cardiograph problems. See the section in
Chapter 4 titled “Test Tools” for instructions about using this tool. The modem test is
identified on the report as
• MODEM
2-8
Performance Verification
CAUTION
Flash Memory Tests (tests C and D) . The flash memory tests verify that the flash
memory on the optional expansion board can be erase, and that information can be stored
and retrieved. The flash memory tests are identified on the report as
•FLASH1
•FLASH2
If your cardiograph is equipped with Option #A05 or #A10, do not turn it off during Test
C or D. If you turn off the cardiograph while Test C or D are in process, you could lose
stored ECGs, the Log of ECGs Stored, or the Log of ECGs Taken.
2-9
Performance Verification
Figure 2-1An Extended Self-Test Report Example
F
G
I
D
K
L
A. Test Results
B. Character Set
C. Timing Tics
D. Diagonal Lines
E. Stepped Bars
F. Software Revision
G. Printer Speed
H. Printhead Voltage
I. Printhead Temperature
J. Battery Voltage
K. Time Stamp
L. Last Error
M. Event Log (Option #A05 only)
B
H
J
A
M
E
C
2-10
Performance Verification
How to Read the Extended Self-Test Report
Each printed page of the extended self-test contains the following information:
ATest Results - See the previous section.
BCharacter Set - Characters 0 through 255 are printed as a table 4x64 characters in
size. This checks the character tables and printhead performance.
CTiming Tics - Vertical lines are drawn from the bottom of the page. Each vertical
line is 25 mm long as measured from the horizontal line drawn ac ross the bottom o f
the page. Spacing between timing ticks is 25 mm ± 1.5%.
DDiagonal Lines - These dense diagonal lines are printed across the top 2 inches of
the report and quickly show speed variations. An 8 cm wide column of diagonal
lines is also printed down the left-hand side of the report, starting at the top-of-form
hole, to show whether the print roller is round.
EStepped Bars - The stepped bars are 1/4-inch thick and 3/4-inch tall, stepping from
the top of the page to the bottom of the page. These bars are used to show printhead
dots that are burned out.
FSoftware Revision - The software revision number for the base ROM.
GPrinter Speed - mm/sec
HPrinthead Voltage - Voltage range is 25.6 to 26.6 V
IPrinthead Temperature - degrees Celsius.
JBattery Voltage - Voltage range is 5.2 - 7.5 V.
KTime Stamp - Number of seconds that the printer test has run.
LLast Error - The last error code encountered. An error code of 65535 indicates that
no errors have occurred since the instrument left the factory. See Table 4-1 for the
table of error codes.
MEvent Log - List of the last 46 events with time and date stamp. The events are listed
with the most recent first. Event sub-codes have the same time stamp as the event
code, and appear on the line above the event code. See Table 4-2 for a table of event
codes.
2-11
Performance Verification
NOTE
Display Test (PageWriter 200/200i/300pi only)
The display test shows test patterns on the display. Each of the following patterns appear
in this order:
• all pixels lit to create a black screen
• no pixels lit to create a white screen
• a vertical bar scrolling from left to right
You must observe the display while the test pattern is running to ensure:
• no burned out pixels
• no random lines or dots in the display
• no permanent patterns visible at normal contrast. Patterns may be visible at low contrast
(darker screen); this is normal.
• no LCD flickering
It is normal for the top row of the LCD to have some pixels on during the first sweep of
the vertical bar.
This test relies on the visual inspection of the test patterns for detecting failures. There is
no failure message for the display test.
Indicator Light Test (PageWriter 100 only)
During this test, all indicator lights turn on and off at least once. At one point, the indicator
lights flash rapidly for a short time.
Printer Test
The printer test exercises the printhead and paper drive mechanism by printing test
patterns and diagnostic data on the page. The patterns consist of a rectangular area that
contains the entire character font set, timing tics, diagonal lines, and stepped bars. On the
PageWriter 200/200i/300pi cardiographs, the printer test and the display test are
conducted simultaneously. The printer test continues until the cardiograph is placed in
Standby mode.
When the printout is complete, inspect it for:
• straight diagonal lines
• even spacing between diagonal lines
• consistent print quality for all patterns
• constant width betw een timing tics (25 mm ±1%)
• consistent length of timing tics
• accurate rendition of all characters
• clean stepped bars with no dropout in black areas
• even spacing between diagonal lines at the beginning of the page
2-12
Performance Verification
NOTE
6SHHG
NOTE
&KDUW6SHHG
Keyboard Tests
The keyboard tests are not automatically performed during the extended self-test. These
must be manually performed and the results visually inspected to verify correctness.
PageWriter 200/200i/300pi Keyboard Test
The keyboard test for 200/200i/300pi models in volves pressing each key on e at a time and
observing that the key character and/or key hex number is displayed on the display. The
keyboard test can be performed anytime after the extended self-test has started.
Pressing changes the speed of the printer test pattern and restarts the extended
self-test.
PageWriter 100 Keyboard Test
To test the keyboard press each key, cycling through the available choices. Be sure that
each LED associated with a key lights when you cycle through the cho ices. This test is no t
performed as part of the extended self-test.
Pressing the key changes the speed of the printer test pattern and restarts the
extended self-test.
2-13
Performance Verification
NOTE
NOTE
ECG Simulation
T aking an ECG using a 12-lead ECG simulator allows you to verify areas of operation that
the extended self-test cannot check:
• integrity of the patient cable
• accuracy of the paper speed (not available on all simulators)
• accuracy of the gain settings (not available on all simulators)
The recorded ECG trace should look similar to the one shown in Figure 2-2. Trace
differences may result from differences in simulators, simulator settings, and from
differences in configuration and control settings on the cardiograph. To make a recording
similar to the one in Figure 2-2:
1.Connect the patient cable leadwires to the simulator.
2.On a PageWriter 200/200i/300pi, verify that all leadwires are connected by cycling
through each lead group and observing the display for flatline. Firmly pull each leadwire taut and look for excessive noise on the display.
3.Select the 12-lead Manual format on the front panel display.
If you are using a PageWriter 100, cycle through all lead group selections to verify the
performance of all leads.
4.Press twice to start the recording.
5.Print approximately 2-1/2 pages. Press .
6.Press twice to print an Auto ECG. Make sure the cardiograph advances the
0DQXDO
6WRS
$XWR
paper to the top of form.
When the recording is complete:
• Verify trace activity for all 12 leads. This assures integrity of all patient electrodes and
leadwires. Noise should measure less than 1 mm, and there should be no baseline wander.
• Verify no gross distortion of complexes or calibration pulses (no overshoot, etc.).
• Verify that calibration pulses are of pr oper duration. This assures the correct pap er speed.
With the cardiograph set to record at 25 mm/sec, the calibration pulse should measure 5
mm (calibration pulse duration is 200 ms).
• Verify that calibration pulse amplitude is correct.
An arrhythmia simulator is not an acceptable tool for verifying computerized ECG
analysis. The analysis software is biased to process human ECG data.
2-14
Performance Verification
Figure 2-212-Lead ECG from ECG Simulator (200/200i/300pi only).
2-15
Changing the Default Operating Language (200/200i/300pi Only)
$OW
$OW
$OW
$OW
$OW
$OW
2Changing the Default Operating Language (200/200i/300pi
Only)
The default operating language is selected by holding down the desired key combination
while turning on the instrument. Table 2-9 lists the key combinations.
Table 2-9Language Key Combinations
Language PageWriter
200/200i/300pi
English
French
German
Dutch
Spanish
Italian
Swedish
Finnish
Norwegian
Portuguese
Polish
$OW
$OW
$OW
$OW
$OW
$OW
–0
–1
–2
–3
–4
–5
–6*
–8*
–9*
–6*
–8*
Russian
–9*
*Only available on certain models.
2-16
Resetting the Cardiograph to the Factory Default State
2Resetting the Cardiograph to the Factory Default State
Resetting the cardiograph sets the Auto ECG report counter to 0, and resets the
PageWriter 200/200i/300pi configuration to the factory default.
Resetting the PageWriter 200/200i/300pi
To reset the PageWriter 200/200i/300pi perform the following steps:
1.Using Table 2-9, turn the cardiograph On in a language different from the current lan-
guage.
2.The cardiograph is now reset to the factory default state.
3.If it is necessary to operate the cardiograph in the original language, turn the cardio-
graph to Standby, then turn the cardiograph On using the appropriate key combination from Table 2-9.
Resetting the PageWriter 100
To reset the PageWriter 100 perform the following steps:
1.Turn the cardiograph to Standby.
2.Press and hold and while turning the cardiograph on.
&RS\
3.Turn the cardiograph to Standby.
4.Press and hold and while turning the cardiograph on.
&RS\
5.The cardiograph is now reset to the factory default state.
3DJH$GYDQFH
&KDUW6SHHG
2-17
Preventive Maintenance
CAUTION
CAUTION
2Preventive Maintenance
Routine maintenance of the cardiograph consists of cleaning and inspection. This should
be done as needed.
The cardiograph does not require lubrication. Lubricating any part of the cardiograph
could damage it or diminish its performance.
Care and Cleaning
The outside surfaces of the cardiograph and its accessories (except the patient cable) can
be cleaned by mild soap and water or isopropyl alcohol. The patient cable cannot be
cleaned with isopropyl alcohol. The patient cable can be cleaned only with mild
disinfectant or soap and water.
Cleaning the Cardiograph
1.Unplug the power cord and ensure that the cardiograph is in Standby mode (the dis-
play is off).
2.Wipe the external surfaces of the cardiograph with a soft cloth dampened with mild
soap and water or isopropyl alcohol. Avoid applying cleaning fluids to cable connectors.
Do not use any strong solvents or abrasive cleaning materials.
Do not spill any liquids on the surface of the cardiograph. Service the cardiograph
immediately if any liquids spill on the surface.
Do not use any of the following to clean the cardiograph:
• Acetone
• Chlorine bleach
• Ammonia-based cleaners
• Iodine-based cleaners
• Phenol-based cleaners
• Ethylene oxide sterilization (ETO)
• Autoclave
2-18
Preventive Maintenance
CAUTION
CAUTION
Cleaning the Keyboard Overlay
1.Carefully insert a small flat-bladed screwdriver into the notches at the front edge of
the keyboard overlay and pry the front edge of the keyboard overlay up. Refer to Figure 2-3.
2.Repeat with the rear notches of the keyboard overlay and remove the overlay.
Figure 2-3Removing the Keyboard Overlay
A. Notches
Do not wash the keyboard overlay in hot water. Do not attempt to clean the keyboard
overlay in a dishwasher.
3.Rinse off the keyboard overlay and thoroughly dry it.
The rubber keypad and/or keyboard assembly may be damaged if any key or LED
becomes trapped under the overlay while replacing the overlay. Make sure each and every
key and LED comes through its ho le in the overlay befor e snapping the over lay into place.
4.Align the overlay with each of the keyboard keys and LEDs. Position your thumbs
along the front edge of the overlay near each end. Apply pressure on the overlay
toward the rear of the unit and then press down on the overlay in one continuous
motion. The front of the overlay should now be snapped into place. Snap the rear of
the overlay into place.
2-19
Preventive Maintenance
CAUTION
Cleaning the Digital Array Printhead and Paper Sensor
If you use Agilent Technologies recording paper, you should not have to clean the
printhead for the life of the product. If you need to clean the printhead, this is the
procedure to follow. These are the materials required:
• Foam swabs — Agilent Technologies part number 9310-0468 or 9300-0767
• 90% Isopropyl Alcohol — Agilent Technologies part number 8500-0559 or equivalent
• Dry, lint-free tissue — Kimwipes® or Agilent Technologies lint-free wipes (Agilent
Technologies part number 92193W)
To clean the printhead:
Touch the equipotential connector on the back of the cardiograph to discharge any static
electricity stored on your skin before touching the printhead. The printhead can be
damaged by static electricity.
1.From the front of the cardiograph, unlatch and open the paper door. The printhead is
to the right under the paper blade, behind the ESD brush (see Figure 2-4).
Figure 2-4Cleaning the Digital Array Printhead.
A. Printhead
B. Paper Sensor
2-20
Preventive Maintenance
CAUTION
2.Wipe the printhead with a foam swab dipped in 90% isopropyl alcohol. Scrub until all
visible residue is removed.
3.Dry the printhead with a clean lint-free tissue.
To clean the paper sensor lens:
1.From the front of the cardiograph, unlatch and open the paper door. The paper sensor
lens is to the right under the printhead (see Figure 2-4).
2.Lightly wipe the paper dust off of the paper sensor lens with a dry foam swab. Do not
use alcohol.
Cleaning the Electrodes and Cables
Clean the reusable electrodes and patient cable with a soft cloth moistened with a
recommended disinfectant or cleaning agent from the following list:
• Cetylcide® (may discolor cable)
•Cidex®
• Lysol® Disinfectant
• Lysol® Deodorizing Cleaner (may discolor cable)
• Dial® Liquid Antibacterial Soap
• ammonia
• 409® (may discolor cable)
• 10% solution of Clorox® in water (may discolor cable)
• Murphy® Household Cleaner, or
• Ves-phene II®.
Wring any excess moisture from the cloth before cleaning.
• Do not clean the patient cable with alcohol. Alcoho l can cause the plastic to become brittle and may cause the cable to fail prematurely.
• Do not autoclave the cable or use ultrasonic cleaners.
• Do not immerse the patient cable.
• Do not use abrasive materials to clean metal surfaces—scratches on them can cause artifacts on the ECG report.
• Do not expose the connectors to liquids, especially the 15-pin connector.
2-21
Caring for the Battery
CAUTION
NOTE
NOTE
2Caring for the Battery
The battery must be installed for proper operation of the cardiograph. Even if the
cardiograph is plugged into AC power, it cannot print an ECG report without the battery.
For information about removing or replacing the battery, refer to Chapter 5, Removal and Replacement.
The sealed lead-acid battery used in the PageWriter 100, 200, and 300 family of
cardiographs will provide optimum life when the u nit is continuously connected to AC
power and fully charged after each use. A d epleted battery requires 16 hours of continuous
charge time to fully charge. Because it is not always possible to allow a full charge cycle
between uses, the cardiograph was designed to charge a depleted battery to 90% of its
capacity in approximately seven hours.
Repeated undercharging of the battery will damage the battery and reduce battery life.
Agilent Technologies recommends that the cardiograph be plugged into AC power
whenever possible to maximize battery life.
Battery life varies depending on frequency of use and maintenance. For improved battery
life, keep the cardiograph plugg ed in when no t in use. If the battery has been fu lly char ged
and requires recharging after a few ECGs, consider replacing it. Use only the Agilent
battery, part number M2460A.
Battery should be removed from unit and placed in storage if the cardiograph will not be
used for more than three months.
Storing the Battery
To prepare the battery for storage, charge it in the cardiograph for 16 hours. Remove the
battery and store it in a cool, dry location. Recharge a lead-acid battery in storage for at
least 16 hours every six months. This ensures that the ba ttery does not completely
discharge while in storage. The battery’s shelf life is longer with cooler temperatures, but
do not store below freezing.
2-22
Caring for the Battery
Safety Tests
Perform the following tests as par t of your prevent ive main tenance program, and aft er any
corrective maintenance.
• Chassis-to-ground resis t ance
• Ground wire leakage current
• Enclosure leakage current
• Patient lead leakage current (source leakage) to ground
• Leakage current between patient leads
• Patient lead leakage current (sink current) with line voltage applied
To perform these tests, use a safety tester or analyzer designed specifically for this
purpose. Follow the manufacturer’s operating instructions.
Record results as indicated in the Performance Verification Matrix.
2-23
PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix
2PageWriter 100, 200/200i, and 300pi Series Performance
Verification Matrix
The following Performance V erification matrix lists all the tests and inspection procedures
which must be performed on the M1770A, M1771A, or M1772A cardiographs after
servicing. Instructions for executing each test or inspection are described in this manual.
Table 2-10 summarizes performance verification for the PageWriter 300pi/200i/200/100
cardiographs; includi n g tes t name, t es t o r in sp ect io n t o perf orm , expected t es t resul t s, and
data to record. Functional tests include ROM/RAM tests, trans mi ssion tests, ECG FrontEnd tests, Flash Memory tests, Display tests, indicator LED tests, printer and keyboard
tests. Simulated ECG and Safety testing are also required.
Chassis-to-Ground Resistance
Groundwire Leakage Current
(Normal condition)
Groundwire Leakage Current
(Single Fault)
Patient Lead Leakage Current
•Source (Normal Condition)
•Source (Single Fault Conditi on)
•With Mains on applied part
(Single Fault Condition )
•Auxiliary (Normal Condition)
•Auxiliary (Single Fault Condi-
tion)
Indicate Applicable Safety
Standard as follows:
Note: All leakage current test are
Normal and Reverse Polarity
Conditions
<
200 mΩ aaa
500 uA (<300uA UL) bbb
<
1000 uA cccc
<
10 uA dd
<
50 uA ee
<
50 uA (< 20 uA UL) ff
<
10 uA gg
<
50 uA hh
<
i=I for IEC i
i=U for UL
S:aaa,bbb,ccc,dd,
ee,ff,gg,hh,i
example:
S:09,400,900,8,4
0,30,8,40,I
For the above - record the worst
case value for multiple lead
measurements.
Note: When recording test results, separate tests by a semi-colon(;). For example: V:p;PO:p;F:p;
D:p;L:p;P:p;K:p;ECG:p;S:90,400,900,8,40,30,8,40,I
2-27
PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix
2-28
Theory of Operation
This chapter contains an overvi ew of the card iograph o peration, an d circuit descriptions of
the major subassemblies. Except as noted, the information in this chapter applies to the
M1770A PageWriter 200i/300pi, the M1771A PageWriter 200, and the M1772A
PageWriter 100.
The last section in this chapter describes the Option #A05 printed circuit assembly.
2Operational Overview
This overview covers two areas of cardiograph operation: the path that ECG data follows
from collection to display, and the cardiograph’s power-on/power-off sequence.
ECG Data Path
Analog ECG data are collected through the patient cable to the ECG Front End on the
CPU Assembly Board in the cardiograph. The ECG Front End performs analog-to-digital
conversion before sending the data to the System Gate Array. The System Gate Array
receives the ECG data and processes it. The data may be interpreted (200i/30 0pi only),
displayed (200/200i/300pi only), printed, or stored in system RAM. PageWriter
200/200i/300pi cardiog raphs equi pped with Opt ion #A05 can al so store up to thirty ECGs ,
and transmit ECGs to TraceMaster ECG Management Systems, ECG Manager software,
other PageWriter cardiographs, facsimile machines, and other Agilent equipment.
Power-on and Power-off Sequences
When the cardiograph is turned on or off, it follows a sequence of events in applying
power to or removing power from its circuits. The System Gate Array controls this
sequence because the System Gate Array remains powered up even when the cardiograph
is in Standby.
Power-on
The system gate array is powered from a supply which is always turned on (Unswitched
5V). When it detects that the key has been pressed, it sends the ENBL5V
signal to the power su pp ly t o t ur n on t he Sw itch ed 5 V su ppl y and r e set t he ot her circuitry.
When the CPU is taken out of reset mode, the processor begins execution of the software
located in ROM. The software sets up the system, decides what type of keyboard is
attached, calibrates the ECG Front End, and initializes the displa y, if present.
2Q6WDQGE\
3-1
Circuit Descriptions
Power-off
The power-off sequence is software controlled. The power-off sequence is initiated when
2Q6WDQGE\
the key is pressed, battery time-out occurs, or the battery is too low to
continue. At power-off, the software begins housekeeping tasks such as turning off the
printer, disabling the keyboard, and turning off the display, if present. When the
housekeeping is complete, the software instructs the Sy stem Gate Array to turn off the
Switched 5V supply. The details of how the power supply shuts down are described later
in this chapter in the power supply desc ri ption.
2Circuit Desc ript ions
The M177XA family of cardiographs has a co mmon set of major subassemblie s: a Power
Supply board, a CPU Assembly board, and a Keyboard (including a Di splay on the
200/200i/300pi). The design also includes a thermal print mechanism. The power supply
design requires a battery for all operations. Option #A05 adds a storage and transmission
board.
Option #A10 adds an interface board for communication to the Agilent StressWriter
system.
The CPU assembly is the same for all cardiographs except that some models of the
PageWriter 200/200i CPU assembly have an extra ROM to accommodate the ECG
measurements and interpretation software. The subassemblies, including the CPU
assembly and its major circuit groups, are shown in Figure 3-1.
3-2
The Patient Cable
Figure 3-1Simplified System Block Diagram
3The Patient Cable
The patient cable used in the PageWriter 100/200/200i/300pi family is a passive
design using no active components. The leadwires each co ntain a r esistor designed to
protect the instrument from the energy used in defibrillation.
The cable simply conducts the analog ECG data from the patient to the ECG Front
End on the CPU Assembly Board.
3-3
CPU Assembly
3CPU Assembly
This discussion of the CPU assembly covers th e following circuits:
•CPU
• System Gate Array
• Serial EEROM
• System Memory (DRAM)
•ROM
• Printer Gate Array/SRAM
• Analog/Digital Converter
• System Expansion Connector
The block diagram in Figure 3-2 shows these circuits and how they interconnect.
Figure 3-2CPU Block Diagram
3-4
ROM 1,2
(2) 512K x 16
or
(1) 2MB x 16
Based on Serial #
CPU Assembly
CPU/System Oscillator
The CPU (Central Processing Unit) is the processing engine for the instrument. The
microprocessor used is the Intel 80960SA, a 32 bit microprocessor with a 32-bit
address bus and a 16-bit ext ernal data bus (3 2 bit int ernal data bus) . The lower 16 bi ts
of the address bus are multiplexed with the data bus; these lines are de-multiplexed b y
the System Gate Array. The processor operates at 16 MHz off of a 32 MHz System
Oscillator. The data and address buses are shown in the above block diagram.
Instructions for the CPU are stored in RAM and ROM external to the CPU.
System Gate Array/Real-Time Crystal
The System Gate Array provides the following functions:
• ECG Front End control
• Interrupt control
• DMA for ECG data
• DMA for LCD data
• System reset circuitry
• Watch-dog circuitry
• Real-time clock (including driving the external Real-Time crystal)
• DRAM RAS and CAS generation and refresh
• Interface to the Serial EEROM
• Address decoding for the Printer Gate Array and ROMs
• Keyboard scan and LED drive
• LCD control (for the PageWriter 200/300 series)
ECG Front End Control
The ECG Front End control interface consists of serial command data to the front
end, and serial data (ECG or status) from the front end.
The input command data consists of a single 24-bit word containing an address field
and a control field. Depending on the data in the address field, the contents of the
control field apply to either
• configuring each individual front end integrated circuit,
• configuring both front end integrated circuits together as a system, or
• setting the calibration value of the front end integrated circuit
Interrupt Control
The interrupt control interface manages two of the four processor interrupt signals.
The processor interrupt signal names are nINT0, INT1, INT2 , and nINT3. nI NT0 and
nINT3 are controlled by the System Gate Array. INT1 is for interrupts from an
expansion board (such as the Option #A05 storage and transmission board) connected
to the System Expansion Connector. INT2 is controlled by the Printer Gate Array.
3-5
CPU Assembly
DMA for ECG Data
The System Gate Array receives data from the ECG Front End in serial form. When 16
bits have been received, the System Gate Array places the CPU in a HOLD state, and
generates the necessary address and control signals needed to write the data into the
System Memory (DRAM). After the data are successfully written, the HOLD state is
discontinued and the CPU resumes its processing. The System Gate Array can also deliver
interrupts to the CPU to indicate when a frame (1 msec) and/or a buffer-full (100 msec) of
data have been written into memory.
DMA for LCD Data
In the 200/200i/300pi, the System Gate Array reads 16 b its of LCD data from System
Memory (DRAM) and delivers it to the LCD for display at an approximately 8 µsec
intervals. The process of reading the 16 bits is to first place the CPU in a HOLD state, then
to generate the necessary address and control signals to read the data, then to discontinue
the HOLD state and allow the CPU to resume processing. The data are delivered one
nibble (4 bits) at a time to the LCD, after which time the HOLD cycle is repeated.
System Reset Circuitry
At a normal power on (pres sing ), the CP U and most other circuit s are reset to
2Q6WDQGE\
a known initial state by the internally generated signal NRESET. This signal is created
within the System Gate Array, which is always active and operating.
Removing all power resets all circuits, including the System Gate Array, via the signal
NNEW5V. Removing all power also resets the real-time clock and date.
Watch-dog Circuitry
The watch-dog circuit will turn off the instrument within 7-8 seconds unless periodically
reset by the control software.
Real-Time Clock
The real-time clock maintains the current date and time. This clock operates off the
cardiograph battery and will run as long as the battery is installed and charged.
DRAM RAS and CAS Generation and Refresh
Refresh of the DRAM is accomplished using a CAS (Column Address Strobe) before
RAS (Row Address Strobe) technique. RAS and CAS are DRAM control signals used to
latch the desired memory address location. A refresh of one row in the DRAM occurs
approximately every 15 µsec.
3-6
CPU Assembly
NVRAM Interface
The NVRAM contains 1K bits of non-volatile storage for the cardiograph
configuration.
Keyboard Interface
The passive Keyboard is scanned by the System Gate Array for keypresses. The
LED(s) are also driven by the System Gate Array. The software is responsible for
determining which keyboard is loaded and configuring the instrumen t for proper
operation.
Display Control for PageWriter 200 and 300 Series
The display is a 240 x 128 dot super-twisted nematic Liquid Crystal Display module
with integrated driver circuitry. Its operation is controlled by the System Gate Array.
NVRAM
The NVRAM stores configuration information. It is written to b y the System Gate
Array under software control and requires no power to retain its data.
ECG Front End
The ECG Front end is described in “ECG Front End”.
DRAM
The DRAM (Dynamic Random Access Memory) is a 256K x 16 bit read/write
memory. The System Gate Array provides the multiplexed address information,
generates RAS and CAS, and provides the timing and control required for memory
transfers from and to the CPU. The DRAM connects directly to the Data bus.
ROM-1 and ROM-2 Address Space
Cardiographs with ser ial num ber prefix CNA, CNB, CNC, or 380 7A or earlier have a
Control Board with space for two ROMs (Read Only Memory) to provide data and
instructions for the CPU. In addition, the Control Board expansion connector and
System Gate Array provide for additional exp a ns io n. Each ROM contai ns 512 K x 16
bits of read-only memory.
Cardiographs with serial # prefix CND have one 2MB EPROM on the control board
which defines the ROM1 and ROM2 address spaces. The ROM contains 2MB x 16
bits of read-only memory.
3-7
CPU Assembly
Printer Gate Array/SRAM
The printer gate array is the principal component of the printer control circuitry. It buf fer s
print data from the system, and status data going to the system. It also controls printer
timing and the printer RAM, regulates motor speed, monitors printhead temperature,
monitors the battery condition, and provides the interface between the printer and the
system CPU.
The printer RAM consists of 32K of static RAM used to store printer data and the printer
look-up table data. The look-up table data is used to compensate the moto r driv e sig nals
and the printer enable pulses for variations in the operating environment. The look-up
table data also includes the motor settings and the motor ramp-up profile.
Motor Driver
The phase signals from the printer gate array enable the motor driver circuit, which
switches current through the stepper motor in the paper transport.
Analog/Digital Converter
This 8-bit parallel A/D converter measures printhead temperature, battery voltage, and
printer supply voltage.
Thermal Printhead
The thermal printhead is a line of 1728 printing elements. Each printing element is a
resistive heater. Heat from these elements blackens the thermoreactive paper, creating an
image. A 1728-bit shift register (on the printhead) contains the d ot data for the print line.
Data is serially loaded into the shift register, then latched so that data for the next line may
be shifted into the head while the present line prints. A strobe signal enables printing. The
pulse width of the strobe is adjusted to compensate for different recording speeds and
variations in supply voltage, average resistance, and temperature. The temperature is
sensed by a thermistor on the printhead heat sink.
System Expansion Connector
The System Expansion Connector provides the instrument with future expansion
capability, such as Option #A05, #A10.
3-8
Power Supply
3Power Supply
The Power Supply connects the instrument to the AC line to charge the battery and
power the circuitry. It conditions and regulates all power supplied to the Main Board.
Figure 3-3Power Supply Block Diagram
Voltage Selector
The voltage selector connects the line input to the proper windings on the Power
Transformer to produce the correct secondary voltage at the output of the
Transformer.
Primary Line Fuses
The Primary Line Fuses protect the power supply from catastrophic failures.
Power Transformer
The power transformer reduces the input voltage to 22V AC (nominal).
3-9
Power Supply
Rectifier/Filter
This circuitry rectifies and filters the AC voltage to produce a DC voltage for the rest of
the power supply.
Switch-Mode Battery Charger
The battery charger converts the output from the Rectifier/Filter to one of two levels. The
first level is called the float voltage and equals 6.85 VDC (nominal at 25 C). The second
level, called the overcharge voltage, is 7.4 VDC (nominal at 25 C). Both of these levels
are temperature compensated. The charger is capable of delivering a maximum of 1.5A at
its output in either the float mode or overcharge mode. The charger is a pulse width
modulated switching type that has a nominal switching frequency of 100 KHz.
The charger enters the overcharge mode when any one of the following conditions are
met: VBAT drops below 6.1V (nominal at 25 C), the instrument is turned on, a print
operation starts, or the instrument is plugged into AC power. The charger changes into the
float mode when the current into the battery is less than 80 mA.
Battery
The battery is a lead-acid 6 volt unit, providing 6.5 Amp-Hours of current when fully
charged.
VPRINT Boost Regulator
The VPRINT Boost Regulator converts its input voltage to the 26 V supply required for
operation of the thermal printhead.
Switched +6V Boost Regulator
The boost regulator is activated by turning the instrument On. It eliminates the heavy
transients induced into the +5V supplies by the printing subsystem during normal printing.
The output from this switching regulator is the input to both of the +5V Linear Regulators
when the instrument is On. When the instrument is in Standby mode, the battery powers
the unswitched +5V Linear Regulator.
5V Linear Regulators
There are two regulators, both linear, described here. The first is the unswitched +5V
regulator that provides power for circuitry active in the Standby mode. The System Gate
Array circuitry active in Standby mode includes the real time clock.
The second +5V Linear Regulator provides the bulk of the operating power for the rest of
the instrument when the instrument is On. This regulator is off when the instrument is in
Standby mode.
3-10
ECG Front End
3ECG Front End
The ECG Front End provides the interface between the system and the patient. It
electrically isolates the patient to prevent excess leakage current, protects the system
from defibrillator and electrostatic discharges, and converts the analog electrical
signals present on the patient to digital samples for processing and printing.
Figure 3-4Front End Block Diagram
Isolation Power Transformer
The Isolation Power Transformer passes power to the E CG Front End and electrically
isolates the patient-connected circuitry of the Front End from the rest of the
instrument.
Front End Power Supply
The Front End Power Supply delivers separate, regulated power at 5V to th e digital
and analog circuitry.
3-11
ECG Front End
Opto-Isolators
Opto-Isolators pass the digital sample data from the patient-connected circuitry to the rest
of the instrument, providing electrical isolation.
Integrated Front End Circuits/Oscillator
The Integrated Front End Circuits sample the patient leads at 4.096M samples per second
per lead. The data are then filtered to 12 bits at 5 µV/LSB and 2 msec per sample. The
circuits include protection circuitry, gain, A/D conversion, and Oscillator control. These
circuits have a master/slave arrangement, with the master driving the Oscillator and
communicating all data to the System Gate Array through the opto-isolators.
Calibration
This circuitry is used to set ECG gain for the Integrated Front End Circuits.
Defibrillator Protection
Defibrillation Protection is provided by resist ors in the patient cable combined with
circuitry in the Integrated Front End Circuits.
3-12
Storage and Transmission (Option #A05 0nly)
3Storage and Transmission (Option #A05 0nly)
The Option #A05 PC assembly plugs into the cardiograph expansion slot, where it
connects to the main PC assembly as shown in Figure 3-1. The cardiograph main
board provides power and access to the cardiograph system electronics. The Option
#A05 PC assembly contains all the circuitry needed to store and transmit ECGs.
Figure 3-5 M177XA Cardiograph Connections to Option #A05 PC Assembly
Cardiograph
Power
Supply
Cardiograph
Main
Board
A05 Option
Board
RS-232
Connection
3-13
Storage and Transmission (Option #A05 0nly)
Figure 3-6 Option #A05 PC Assembly Block Diagram
M177X
Expansion
Connector
Gate
Array
18.432 MHz
Oscillator
Address
Dual
UART
Serial Port
Connector 1
RS-232
Drivers
Serial Port
Connector 2
VPROG
Supply
ROM-X1
256K x 16
SYS Control
GA Control
ROM-X2
256K x 16
Data
FLASH 1
1M x 16
FLASH 2
1M x 16
M177XA Expansion Connector
The expansion connector on the Option #A05 PC assembly mates with the M177XA
cardiograph expansion connecto r . This connection provides the Option #A05 PC assembly
access to the cardiograph Address Bus, Data Bus, System Control Bus, and power
supplies.
Gate Array
The gate array integrates the following functions :
• Address decoding and chip selects for the expansion ROMs (ROM-X1 and ROM-X2)
• Flash memory controller
• Interface to the Dual UART
ROM-X1, ROM-X2
ROM-X1 and ROM-X2 are ROMs provided for th e cardiograph firmware expansio n. The
cardiograph CPU controls and execute the firmware in these ROMs.
3-14
Storage and Transmission (Option #A05 0nly)
Flash1, Flash2
FLASH1 and FLASH2 are 1M x 16 bit flash ROMs. The gate array provides control
and timing functions required to interface these parts to the ad dress and data bus.
FLASH1 is used for non-volatile ECG storage. FLASH2 is pro vided for expansion
memory. It is not normally present on the Option #A05 printed circuit assembly.
VPROG Power Supply
The VPROG power supply provides a 12V DC pr ogramming volt age to FLASH1 and
FLASH2 during erase and write operations. The source for the supply is the
cardiograph printhead supp ly voltag e, which is turne d on and of f by the software. The
VPROG supply circuit contains a 12V linear regulator that drops the printhead supply
voltage to the required 12 volts for FLASH1 and FLASH2.
Dual UART
The Dual UART (Universal Asynchronous Receiver/Transmitter) contains two
independent UARTs with the following capabilities:
• modem control
• baud rate generator
• 16 byte FIFO (first-in-first-out) buffer
Each UART is used to implement an RS232 port by translating between serial and
parallel data.
18.432 MHz Oscillator
The 18.432 MHz Oscillator provides a clock for the internal functions of the Dual
UART.
RS232 Drivers
The RS232 Drivers perform the voltage level translation between TTL logic level
signals from the Dual UART and the +
10 volt RS232 signals.
Serial Port Connectors 1 and 2
These connectors are 9-pin D-subminiature connectors that are located at the back
panel of the cardiograph, and provide access to the two RS232 ports. The serial port
marked RS-232 is used for ECG transmission. The other serial port is reserved for
future use.
The #A10 interface board also, provides storage and transmission capability in
addition to the Agilent StressWriter interface.
3-15
Storage and Transmission (Option #A05 0nly)
3-16
Troubleshooting
CAUTION
3Introduction
This chapter provides information for localizing cardiograph problems to the subassembly
level. This information is designed for use with the cardiograph’s Extended Self-test to
help you efficiently repair the cardiograph with a minimum of equipment.
3Maintenance Philosophy
The maintenance philosophy for the cardiograph is subassembly replacement.
Replaceable subassemblies are identified in Chapter 6, “Parts List”.
Individual component replacement should not be attempted outside of an authorized
Agilent Technologies’ repair facility. Component level repair is extremely difficult due to
the extensive use of surface mount technology and the high parts density on the circuit
boards. Unauthorized component replacement can impair cardiograph performance,
compromise patient safety, and jeopardize credit towards a replacement assembly.
3Test Equipment
The following test equipment is required to troubleshoot the cardiograph as described in
this chapter:
• A digital voltmeter such as the HP E2373A hand-held multimeter.
• A 12-lead ECG simulator.
• A jumper wire with Pomona Micrograbber test clips or equivalent. Ordering information:
ITT Pomona 1500 East Ninth St.
P.O. Box 2767
Pomona, CA 91769
Sales: (909) 469-2900
4-1
Test Tools
3Test Tools
Two troubleshooting tools are provided with each instrument:
• patient cable test tool (M1770-87908)
• front end and transmission test tool (M1770-87909, Rev. B)
Patient Cable Test Tool (M1770-87908)
This tool is attached inside the battery compartment, beneath the battery. T o locate it, open
the battery compartment door and lift the battery.
This tool is used to short the lead wires together to test lead wire integrity and detect
potential "opens" within the cable. To test the leads, plug all ten lead wire posts into the
holes of the tool and print an ECG. If all leads print clear , solid flat lines, the lead wires are
intact.
Patient Cable Test Procedures for Internally Shorted Lead Wires
For PageWriter Models M1771A, M1770A, and M1772 (excluding software revisions
A.01.02 or A.01.08) test the patient cable for internally shorted lead wires by following
the steps in the procedure below:
1. Plug the patient cable into the cardiograph.
2. Make sure the individual lead wires do not touch each other or any conductive mate-
rial.
3. Switch the cardiograph to the ON position.
4. The top line of the display or top line of an AUTO recorded ECG [Printed Form]
should indicate RL (AHA) or N (IEC) or RL/N (combined AHA/IEC).
5. If the top line of the display or the top line of the AUTO recorded ECG [Printed
Form] shows leads other than RL, N, or RL/N, then an internally shorted lead has
been detected, and the patient cable must be replaced.
For PageWriter M1772 Models with software revisions A.01.02 or A.01.08, test the
patient cable for internally shorted lead wires by following the steps in the procedure
below:
1. Place the cardiograph in the Standby mode [if the cardiograph was ON, then you will
need to depress the button to place the cardiograph in the Standby
2Q6WDQGE\
mode.]
2. Depress and hold down keys and with fingers of one
hand, and then depress with a finger of the other hand. Thi s puts the car-
$XWR
2Q6WDQGE\
Page Advance
4-2
Test Tools
diograph in a special "QA mode" which will enable the operator to perform th is evaluation.
3. The patient cable is then securely plugged into the cardiograph’s patient cable connec-
tor and ensure the individual lead wires do not touch each other or any electrically
conductive surface. Verify that the cable is not swinging or otherwise in motion. Do
not hold or touch the trunk cable during the test.
4. Depress and then release the key. Wait approximately 15 seconds. The car-
$XWR
diograph will generate one sheet of paper with 2 or 3 rows of information on the bottom left side of the paper. [If you do not see this information printed on the paper,
then go back to step 1 and repeat the procedure to enable the QA mode]
5. Note the second line from the bottom of the information will contain 9 separate 3 digit
numbers. There may also be present a 10th group of digits which contain an @ sign.
This 10th group of digits with the @ sign and any digits in that group should be
ignored.
Example 1:
006-005-007 009-008-063-007-008-007
A.01.02 4350 25.4ºC 7.74V 26.24V Battery OK AC
Example 2:
000-000-000 000-000-000-000-000-000 000
006-005-007 009-008-063-007-008-007 1511@64
A.01.08 4350 25.4ºC 7.74V 26.24V Battery OK AC
6. Check the first nine groups of three-digit numbers in the second row from the bottom.
If ANY three-digit number is greater than 025, then an internally shorted lead has
been detected and the patient cable must be replaced. If none of the three-digit numbers exceed 025, then the cable is free of defects known as internally shorted leads
and may continue to be used. In the example above, the 6th three-digit number is
063, which exceeds 025 for an acceptable cable. Therefore, this cable is defective.
7. Place the cardiograph back into the Standby mode by depressing the
2Q6WDQGE\
button again.
8. Depress and hold down the keys and with fingers of one
hand, and then depress again with a finger of the other hand. This will
2Q6WDQGE\
$XWR
Page Advance
restore the cardiograph to its normal operating state.
Depress and then release the key. Wait approximately 15 seconds. The
$XWR
cardiograph will generate one sheet of information on recording paper. Note that on the
bottom left side of the recording there are no longer any lines of information as previously
seen in step 5. If you continue to see this information, then repeat step 7 and step 8 to
disable the "QA mode".
4-3
The Error and Event Logging
NOTE
NOTE
NOTE
Front End and RS-232 Port Test Tool (M1770-87909, Rev. B)
This tool is located on the lower side of the printer do or. To remove it, open the printer
door and reach underneath it. The tool has two test plugs, one used to test the instrument
signal path, and one used to test the c ardiograph’s RS-232 port .
Testing the Instrument Signal Path
To test the instrument signal path, plug the larger connector into the patient cable
connector on the front of the cardiograph. Print an ECG. If the front end is operating
properly, all leads will show clear, solid flat lines with little or no noise.
You can use the patient cable test tool and the front end test tool to isolate patient cable
problems and cardiograph problems.
Testing the Cardiograph’s RS-232 Port
T o t es t the cardiograph’s RS -23 2 po rt, pl ug the s mall er conn ect or int o the R S-232 port, in
place of the modem. Run the extended self-test described in Chapter 2. The modem test
will pass if the cardiograph RS-232 port is operating properly. This can help differentiate
between cardiograph problems, and cable or modem problems.
You can use the front end and RS-232 port test tool to isolate modem problems and
cardiograph problems.
The RS-232 port test connector is available only on Revision B of the M1770-87909.
3The Error and Event Logging
The last error that was displayed can be printed by starting the Extended Self-test. The
error code is useful in diagno si n g what was las t wron g wi th t he car diograph. See “How to
Read the Extended Self-Test Report” in Chapter 2 for the location of the last error on the
Extended Self-test report.
The last 46 events recorded in the Events Log are also printed on the Extended Self-Test
Report for cardiographs equipped wit h Opt ion #A 05.
4-4
Using Extended Self-test in Troubleshooting
$XWR
3Using Extended Self-test in Troubleshooting
The cardiograph’s Extended Self-test is a simple-to-use, looping test that gives you
pass/fail status for each of these major subassemblies:
• CPU assembly
•Printer
• Keyboard display (M1772A excluded)
• Modem (Option #A05 only)
Instructions for entering and using Extended Self-t est are found in Chapter 2.
3How the PageWriter 100 Communicates Error Codes and
Messages
The PageWriter 100 uses its LEDs to communicate error codes and messages.
PageWriter 100 Error Code Communication
When an error is detected, all LED’s light for 1/2 second, then the LED’s above the error
code digits light in turn for 1/2 seconds each, beginning with the leftmost digit. For
example, error code 3527 will cause the following LED’s to blink in the order given: 3
)LOWHU
, 5 , 2 , 7 (all Manual lead group LED’s).
9/HDGV
4-5
Troubleshooting Tables
PageWriter 100 Error Message Communication
The PageWriter 100 communicates the following erro r messages with these corresponding
LED’s:
Leads off All Manual lead group LED’s blink alternately with
$XWR
LED.
Check Paper SupplyThe current LED blinks.
Printer Door OpenThe current LED blinks.
3Troubleshooting Tables
The troubleshooting tables in this section help you to localize a fault and correct it. The
troubleshooting fl owchart in Figure 4-1 gu ides you to the tr oubleshooti ng table that cove rs
a particular functional area of the cardiograph.
Table 4-1 lists errors that can occur during operation and will appear only as numbers on
the display . T he table defines and explains each er ror, and suggests one or more cor rective
actions for each.
&KDUW6SHHG
&KDUW6SHHG
Figure 4-1Troubleshooting Flowchart
Start
Powers up
Yes
Error displayed?
No
Patient Cable &
ECG Front End
ECG problems
See Table 4-3
CPU
System
inoperative
See T able 4-4
Printer
No paper
movement.
Poor print
quality.
See Table 4-5
No
Yes
Keyboard &
Display
Keys don’t
function or
garbled display
See Table 4-6
See Table 4-2
See Table 4-1
Storage
Cannot store
or retrieve
ECGs.
See Table 4-7
Transmission
Cannot send
or receive ECGs.
See Table 4-8
4-6
Troubleshooting Tables
CAUTION
Before removing or inserting any board or connector, make sure AC power is off and the
battery is removed.
The error codes listed below appear on the extended self-test report in the last error
column or in the event log (Option #A05 only) described in Chapter 2.
1101ECG front end could not calibrate due to exces-
2000-2003, 5221Power supply fault.1. Test the power supply (refer to “Testing the Power
3517-3527Advisory codes appear during powe r-on and
3510, 7000, 7001Defective keyboard or keyboard cableReplace keyboard if the error persists.
3531Defective FLASH memory.1. Perform th e extended self-test (refer to Chapter
5033-5077Defective CPU ass em bly.1. Turn off, wait 30 seconds, turn on.
5155Defectiv e UART.1. Perform the exte nded self-test (refe r t o Chapter
Defective ROMs or o th er rel a te d hardware problem on CPU assembly.
sive ambient nois e.
indicate an abnor m al cardiograph sh ut down.
1. Turn off, wait 30 seconds, tur n on.
2. Replace CPU assembly, if error code persists.
If the cardiograph uses softwa re release A.01.02,
modify the card i o graph using upgr ade kit M1770-
89520.
Supply). Replace power supply if defective.
2. Inspect cabling between power supply and CPU.
3. Replace CPU assembly, if error code persists.
Turn off, wait 30 seconds, turn on.
2).
2. If the extended sel f -test fails, replac e th e O ption
#A05 PCA.
2. Replace CPU assembly, if error code persists.
2).
2. If the extended sel f -test fails, replac e th e O ption
#A05 PCA.
All other error codesBad con figuration.1. Reset configurati on. Refer to”Resetting th e Car-
Defective CPU assembly
diograph to the Factory Default Stat e. ”
2. Replace CPU assembly, if error code persists.
4-7
Troubleshooting Tables
2
Table 4-2Power Supply and Battery
Symptom Possible Cause Corrective Action
AC indicator not lit.Power cord unplugged.
AC line voltage switch in wrong position.
One or both AC fuses are blown.
Power supply to CPU interface cables or key-
board cable not fully seated.
Bad power supply to CPU interface cable(s).
Bad keyboard circuit board.
Bad power supply.
Bad CPU assembly.
Cardiograph doesn’t run on
Battery connecto r i s l oose or defective.
battery, but runs on AC.
Low battery or defective battery.
Software has stoppe d.
Battery fuse is blown.
Battery charger is defective.
Cardiograph won’t powe r up
on AC or batter y.
Power supply to CPU interface cables or keyboard cable not fully seated.
Power supply to CP U int erface cable(s) is defective.
Software has stoppe d.
Keyboard circuit board is defective.
Power supply is defective.
CPU assembly is defective.
Plug in power cord.
Move to correct position.
Replace AC fuse(s).
Reseat cables.
Replace cable(s).
Replace keyboard as s em bly.
Replace power suppl y.
Replace CPU assembly.
Make sure battery conn ector is properly seated
and not defective.
Charge or replace battery.
Unplug from AC power. Remove battery. Wait
2 minutes. Reinstall battery and turn instrument on.
Disconnect the battery and visually inspect the
battery fuse on the battery. Replace the battery,
if necessary.
Replace power suppl y assembly.
Reseat cables.
Replace the power su ppl y to CPU interface
cable(s).
Unplug from AC power. Remove battery. Wait
2 minutes. Reinstall battery and turn instrument on.
Replace the keyboard assembly.
Replace power suppl y assembly.
Replace CPU assembly.
Battery does not charge.AC line voltage switch in wrong position.
Defective power supply.
Defective batte ry.
Battery capacity too low.Battery is not fully charged.
AC line voltage switch in wrong position.
Defective or worn-out battery.
Cardiograph turns itself off
while plugged into AC
power.
AC line voltage switch in wrong position.
Power supply to CPU interface cables or key-
board cable not fully seated.
Defective power supply.
4-8
Move to correct position
Remove battery. Using a voltmeter, measure
for approxima t ely 6.8 Vdc across the battery
connector in the unit. Replace power supply, if
absent.
Replace battery.
Fully charge battery.
Move to correct position.
Replace battery.
Move to correct position.
Reseat cables.
Replace power suppl y.
Troubleshooting Tables
Table 4-2Power Supply and Battery
Symptom Possible Cause Corrective Action
Low Battery ligh t or Low
Battery message remains on
during extended self-test.
Low Battery ligh t or Low
Battery message blinks.
One or more power supply
voltages missing (see “Testing the Power Supply” at the
end of this chapter).
Low battery or defective battery.
Battery fuse is blown.
Battery charger is defective.
Low battery.Fully charge battery.
Defective regul ato r circuit(s).Replace power supply assembly.
Charge or replace battery.
Disconnect the battery and visually inspect the
battery fuse on the battery. Replace the battery,
if necessary.
Replace power suppl y assembly
Table 4-3Patient Cable and ECG Front End
Symptom Possible Cause Corrective Action
”Leads off” not indicated
when wire is off.
Dirty contacts on p at i ent cable connector.
Defective or di rt y leadwire.
Defective ECG front end.
Ensure the contacts on the patient ca bl e connector are clean and dry. Reseat the patient cable
connector.
Clean leadwire. Replace pati ent cable if leadwire
defective.
Replace CPU assembly.
”Leads off” indica te d w hen
wire is not off.
Poor electrode contact.
Defective lead wire.
Improve patien t preparation.
Clean leadwire. Replace pati ent cable if leadwire
defective.
4-9
Troubleshooting Tables
NOTE
Table 4-3Patient Cable and ECG Front End
Symptom Possible Cause Corrective Action
Bad ECG:
Good calibration pulse in
channels where traces are not
good.
A lead is missing — dotted
line on trace.
Noisy lead(s): “AC” (regular
pattern) — for 60 Hz, 1 2
peaks/5 mm at 25 mm/sec;
for 50 Hz, 10 peaks/5 mm.
Defective lead wires.
Defective ECG front end.
Faulty operator tec hnique; poor electrode
contact.
Defective lead wires.
Defective ECG front end.
Faulty operator tec hnique; poor electrode
contact.
Patient or patient cabl e nea r AC power.
Lead wires may be picki ng up interference
from poorly grou nded equipment near the
patient.
Refer to Table 6-2 in user’s manual.
Replace patient cable.
Replace CPU assembly.
Refer to Table 6-2 in user’s manual.
Replace patient cable.
Replace CPU assembly.
Refer to Table 6-2 in user guide.
Reposition patient ca bl e. Refer to Table 6-2 in
user guide.Try unplugging cardiograph from the
AC outlet.
Route lead wires alon g lim bs and away from
other electrical equipment. Fix or move poorly
grounded equipm ent .
Patient cable is too close to the cardiograph
or other power cords.
Noisy leads: Muscle artifact.Patient is not relaxed or skin has been irri-
tated.
Low quality disposable electrodes, leadw i r e
adapters.
Noisy lead(s): I, II I, aVLLA leadwire defective.Replace patient cab le .
Noisy lead(s): II , III, aVFLL leadwire defective.Replace patient cable.
Noisy lead(s): I, II , aVRRA lead w i re defective.Replace patient cable.
Noise in only one V l ead.Faulty ope rator technique; po or electrode
contact.
Defective lead wires
Noise in general.Faulty operator tec hnique; poor electrode
contact.
Defective lead wires.
Move the cardiogra ph away from the patient .
Unplug the cardiograph and operate on battery.
Unplug the electric bed.
Refer to Table 6-2 in User Guide.
Replace disposable electrodes or leadwire adapters.
Test with ECG simulator. Replace patient cable.
Refer to Table 6-2 in user guide.
Replace patient cable.
Refer to Table 6-2 in user guide.
Replace patient cable.
Remember that most noise results from poor patient-electrode connections or poor quality
electrodes. If noise persists after checking electrode placement and lead-electrode
connections, the noise may be due to poor quality electrodes. Some electrodes have a shelf
life of 48 hours or less once the foil package is opened. Eliminate electrode placement,
connections, and freshness as the causes of the noise before replacing the patient cable.
4-10
Troubleshooting Tables
NOTE
Chart Speed
Chart Speed
Page Advance
You can use the patient cable test tool and the ECG front end test tool to determine
whether a problem is caused by a poor patient to electrode connection, a defective cable or
a defective cardiograph. See the section titled “Test Tools” in this chapter for more
information about using the patient cable test tool.
Table 4-4CPU Assembly
Symptom Possible Cause Corrective Action
System turns on but won’t
run. Power supply OK.
System won’t turn off.Software has stopp ed.
A cable is not fully seated.
Defective CPU assembly.
Defective CPU assembly
Reseat all cables.
Replace CPU assembly.
Unplug from AC power. Remove battery. W ait 2
minutes. Replace battery and turn unit on.
Replace CPU assembly.
Table 4-5Printer
Symptom Possible Cause Corrective Action
Paper doesn’t move; printer erro r message
displayed
the current LED is flashing.
Paper moves then st ops and displays error
message
the current LED is flashing.
Check paper supply or
Check paper supply or
A printer cable is no t fully seated.
Defective sensor assembly.
Defective motor assembly.
Defective motor drive assembly.
Defective roller.
Defective roller.
Paper loaded incorrec tly.
Dirty sensor lens .
A printer cable is no t fully seated.
Defective sensor assembly.
Defective sensor circuit.
Defective roller.
Replace paper or clear paper jam.
Reseat connectors J303 through J306.
Replace sensor assembly.
Replace motor assembly.
Replace CPU assembly.
Replace printer door assembly.
Make sure cardiograph is loaded properly
with approved thermal paper.
Clean sensor lens.
Reseat connectors J303 through J306.
Replace sensor assembly.
Replace CPU assembly.
Replace printer door assembly.
Paper stops in the wrong place after pressing
Auto
or . No error
message.
Wrong type of paper.
Sensor hole not cut out properly on
paper.
A printer cable is no t fully seated.
Dirty sensor lens .
Defective sensor.
Defective sensor circuit
Make sure cardiograph is loaded properly
with approved thermal paper.
Remove defective s heet of paper.
Reseat connectors J303 through J306.
Clean sensor lens.
Replace sensor assembly.
Replace CPU assembly.
4-11
Troubleshooting Tables
Chart Speed
Table 4-5Printer
Symptom Possible Cause Corrective Action
Message: Printer door open or
the current LED is flash-
ing.
Wavy diagonal lines on printer self- t est or
distortion of printouts in the time axis.
Printing is dark on one side of page but fai nt
on the other side.
Paper moves but nothing prints.Paper loaded incorrectly, or non-
Door is ajar.
A printer cable is no t fully seated.
Printhead contact switch clip or leaf
spring touching contact post.
Defective sensor assembly.
Defective door-detect circuit.
Improper mesh ing of gears in printer
drive assembly or door assembly.
Printhead is not free to float and provide even (unifor m ) pressure distribution across platen roller.
approved or non-thermal paper
installed.
A printer cable is no t fully seated.
VPRINT supply defective.
Insufficient leaf spring tension or
printhead out of position.
Defective printhe ad or printhead
cables.
Defective component in printer circuitry.
Close door.
Reseat connectors J303 through J306.
Check leaf spring behind printhead and
contact post.
Replace sensor assembly.
Replace CPU assembly.
Replace printer dri ve assembly or door
assembly.
Make sure the printer drive assembly is
properly mounted o n th e t op cover.
Make sure there is slack in the printhead
cables.
Replace printer door.
Replace printhead as sembly.
Make sure cardiograph is properly loaded
with approved thermal paper.
Reseat connectors J303 through J306.
T est VPRINT. If de fect ive, re place po wer
supply.
Check leaf spring for printhead platen
pressure, door latch.
Replace printhead assembly or printh ead
cables.
Replace CPU assembly.
Paper moves but printing is faint.Door improperly latched.
Paper loaded incorrectly, or nonapproved or non-thermal paper
installed.
A printer cable is no t fully seated.
VPRINT supply defective.
Insufficient leaf spring tension or
printhead out of position.
Defective printhe ad or printhead
cables.
Defective component in printer circuitry.
Capacitor assembly defective.
Close door properly.
Make sure cardiograph is properly loaded
with approved thermal paper.
Reseat connectors J302 through J306, and
connectors on the capacitor board
T est VPRINT. If de fect ive, re place po wer
supply.
Check leaf spring for printhead platen
pressure, door latch.
Replace printhead as sembly.
Replace CPU assembly.
Replace capacito r assembly.
4-12
Troubleshooting Tables
Table 4-5Printer
Symptom Possible Cause Corrective Action
Printed data are garbled.A printer cable is not fully seated.
Defective printhead.
Defective component in data path.
Poor print quality or some dots not printing.Dirty printhead.
Loose ESD brush fibers.
Incorrect printhead or sensor cable
routing.
Defective printhe ad or printhead
cables.
Defective component in printer circuitry.
Some dots always on.Defective printhead.
Defective printhe ad or printhead
cables.
Reseat connector J304.
Replace printhead as sembly.
1.Replace printhe ad c ont rol cable.
2. Replace CPU assembly.
Clean printhead.
Remove loose brush f ib er s and clean
printhead.
Make sure there is slack in the printhead
cables.
Replace printhead assembly or printh ead
cables.
Replace CPU assembly.
Replace printhead as sembly.
Replace printhead assembly or printh ead
cables.
Table 4-6Keyboard and Display (LCD)
Symptom Possible Cause Corrective Action
Display sta ys blank or all
black. (M1770A / M1771A
only)
Contrast misadjusted.
Keyboard cable i s not fully seated.
No power being su pplied to keyboard assem-
bly.
Defective keyboar d a ssembly.
Defective LCD assembly
Defective contrast control in CPU ass embly.
Adjust contrast using the and . or
keys.
Reseat keyboa rd cable.
Test power supply. Replace if defective.
Replace keyboard as s em bly.
Replace LCD assembly.
Replace CPU assembly.
Shift
Keys won’t work. Unable to
enter data or operate controls.
Garbled data on display.Keyboard cable not seated correctly.
Defective LCD.
Defective keyboar d a ssembly.
Defective CPU assembly.
Reconnect keyboard cable.
Replace keyboard as s em bly.
Replace CPU assembly.
Make sure cable from keyboard to CPU assembly is properly seat ed.
Make sure flex cable is fully inserted and locked
in each connector.
Replace LCD assembly.
Replace keyboard as s em bly.
Replace CPU assembly.
4-13
Troubleshooting Tables
Table 4-7Storage (Option #A05 0nly)
MessagePossible CausePossible Solutions
ECG too noisy to sto r e
(This message appears briefly on the
screen)
"Storage system full" message appears
when fewer than 30 ECGs are stored.
Unable to store ECG
(This message appears briefly on the
screen)
Unable to retrieve ECG
(This message appears briefly on the
screen)
Poor electrode cont act . D ry or dirty electrodes.
Patient moving or not relaxed.
Lead wires may be picki ng up interference from poorl y grounded equipm ent
near the patient.
Patient cable is too close to the cardiograph or other power cor ds.
30 ECGs stored in memory.
Storage memory gradually wears out after
many thousand st ore/erase cycles. Consequently, ECG storage capacity decreases
gradually over the lif e of th e pr oduct.
A fault exists in the storage hardware.Call Agilent service
A fault exists in the storage hardware.Call Agilent service
Use new electrodes. Abrade skin. Reapply electrodes. Check expiration date on
disposable electr odes.
Reassure and relax the patien t. Press the
Filter
fact.
Route lead wires alon g lim bs and away
from other electrical equipment. Fix or
move poorly grounded equipment.
Move the cardiograph away from the
patient. Unplug the cardiograph and operate on battery power. Move other electrical equipment away from patient. Unplug
electric bed.
Delete some ECGs.
If under warranty, call Agilent service.
Generally, stored ECGs are retrievable. If
remaining ECG storage capacity is unacceptable, call Agilent serv ice .
key if it is configured for Arti-
4-14
Troubleshooting Tables
Table 4-8Transmission (Option #A05 only)
MessagePossible CausePossible Solutions
Telephone busy, waiting to redialBusy telephon e ay remo te site.Normal event. The cardiograph will
redial, waiting 30 seconds between
attempts, until the line is o pen or un til th e
operator stops the transmission.
No answer, re-dialingRemote mo dem not connected, or the car-
diograph modem is set to wait for an
answer for too few rings
Check telephone cab l eNo dial tone.Check the telephon e cable connection to
Check modem and ca bl eNo power to modem, or poor modem
cable connection .
Defective cable.
Defective modem.
Check cablePoor cable conn ection between cardio-
graph and TraceMaster system.
Defective cable.
Check modem configurationIncompatible or imp roperly initialized
modem
Be sure the remote si te is r eady t o r ecei ve
the transmission. If necessary, extend the
time the modem will wait for a connec-
tion by adding the string “S7=90” to the
modem initialization string.
the modem. Attach the telephone cable to
a telephone to veri fy a dial tone on the
line.
Check that the data cable is attached to
the modem and to the cardiograph’s RS232 port. Check op eration of the RS-232
port.
Replace the data cabl e.
Run the modem self-test. Repla ce modem
if defective.
Check that the data cable is attached to
the cardiograph’s RS-232 port. Ch ec k
operation of the RS-232 port.
Replace the data cabl e.
Check the modem initialization string for
legal modem commands. See you modem
documentation for legal commands for
your modem.
4-15
Troubleshooting Tables
Table 4-8Transmission (Option #A05 only)
MessagePossible CausePossible Solutions
No modem at remote siteThe loc al modem could not complete the
connection. The mos t li ke ly causes are:
1. Telephone cable disconnected from
local modem.
2. Remote modem turned off or disconnected from telephone line.
3. Telephone prematurely disconnected at
remote site.
4. Excessive noise on the telephone line.
5. Incompati ble transm ission pr otocols o n
remote and local modems.
6. Incompatible modem setup strings on
the remote and local modems.
7. The remote modem did not answer the
phone cal within the time specified in
the modem’s S7 register, and the @
symbol is not present in the tel ep ho ne
number string in the cardiograph telephone directory.
1. Check all cable connections.
2. Be sure the remote mo dem is tu rned on
and connected to the telephone cable.
3. Retry the transmission.
4. Retry the transmission.
5. Check the cardiograph telephone directory and verify te le phone number,
device type, and transmission speed.
6. Check the modem initialization string
commands that allow compatibility
with the remote modem.
7. Be sure the remote site is rea d y to
receive the transmission. Consider
extending the time the modem waits
for a connection by adding the string
“S7=90” to the modem initialization
string.
Transmission stopped unexpectedly. X of
N ECGs received.
Cable/ modem problem, Press any key to
continue.
Transmission stopped unexpectedly. X of
N ECGs received.
Modem was disc onnected. Press any k ey
to continue.
Transmission stopped unexpectedly. X of
N ECGs sent. Remote site stopped communication (nnnn). Press any key to continue.
or
Transmission stopped unexpectedly. X of
N ECGs received. Remote site stopped
communication (nnnn). Press any key to
continue.
No power to modem, or poor modem
cable connection .
Problem with telephone line or remote
modem.
Communication speed of the remote
device does not match that of the cardiograph, or the remote site modem malfu nctions. The (nnnn) is the event code
associated with the message, and is
recorded in the event log when this message appears. These events relate to the
transmissi on prot oco l. See Table 4 -9 for a
list of event codes.
Check that at least on of the modem’s
LEDs is on. Check that th e data cable is
attached to the modem and to the cardiograph’s RS-232 port. Check operation of
the RS-232 port.
Check that at least on of the modem’s
LEDs is on. Check that th e data cable is
attached to the modem and to the cardiograph’s RS-232 port. Check operation of
the RS-232 port.
Be sure the remote si te is read y to r ecei ve
the transmission.
Retry the transmission at 2400 bps.
Be sure the remote si te is read y to r ecei ve
the transmission.
Be sure the remote site is ready to send
the transmission.
4-16
Troubleshooting Tables
In addition to the error codes listed above, the following event codes are displayed on the
extended se lf-test report for cardiographs equipped with Option #A05.
Table 4-9Event Codes (Option #A05 only)
Event #DefinitionCorrective Action
1223Record corru pt ed during transmissio n. This code is
followed by an sub-error code.
35731. The battery w as removed when the cardiograph
was not plugged in to A C pow er.
2. The battery fuse is defective.
3575The cardiograph turned off due to inactivity.Normal event. No a ct ion required.
3577The cardiograph turned off due to low battery.Plug cardiograph into AC power.
3700The power w as cycled quickly.Normal event. No actio n required.
3701Errors were detected in the front end communication
link.
5160, sub code 3No modem at remote site.1. Check all cable connections.
Retry the transmission .
Replace the battery.
Run the extended sel f-test to check oper at ion of the
front-end.
2. Be sur e the remo te modem is turned o n and connected to the telephone cable.
3. Check the cardiograph telephone directory and
verify telephon e number, device type, and tran smission speed.
4. Check the modem initialization string commands
that allow compatibility with the remote modem.
5. Be sure the remote site is ready to receive the
transmission. Consider extending the time the
modem waits for a connection by adding the
string “S7=90” to the modem initia lization string.
6. Retry the transmission
5160, sub code 4Check modem co nfiguration.Check the modem initia lization string for l egal
modem commands. See your modem documentati on
for legal commands for your modem.
5160, sub code 5Check telephone cable.Check the telephone cable connection to the
modem. Attach the telephone cab le to a telepho ne to
verify a dial tone on t he line.
5160, sub code 6Telephone busy, waiting to redial...
(after 30 seconds, this message appears:)
Telephone busy, redialing...
5160, sub code 7No answer, waiting to red ia l .. .
(after 30 seconds, this message appears:)
No answer, redialing...
Normal event. The cardiograph will redial, waiting
30 seconds between attempts , un til the line is open
or until the operator stops the transmission.
Be sure the remote site is ready to receive a transmission. If necessary, extend the time the modem
will wait for a conn ection by adding the st ring
“S7=90” to the modem initialization string.
4-17
Troubleshooting Tables
Event #DefinitionCorrective Action
5160, sub codes 10
and 11
5161Transmission stopped unexpectedl y.
5162Transmission stopped unexpectedl y.
5163Transmission stopped unexpectedl y.
Check cable. (This message appear s when the transmission type is Direct or Direct SCP)
Check modem and cable. (This message appears
when the transmission type is Modem or Modem
SCP.)
x of n ECGs sent
Cable/modem problem
Press any key to cont inue
or
Transmission stopped unexpectedly.
x of n ECGs received
Cable/modem problem
Press any key to cont inue
x of n ECGs received
Modem was disc onnected
Press any key to cont inue
x of n ECGs sent
Remote site stopped communication
Press any key to cont inue
or
Transmission stopped unexpectedly.
x of n ECGs sent
Remote site stopped communication
Press any key to cont inue
Check that at least one of the modem’s LEDs is on.
Check that the data cable is attached to the modem
and to the cardiogr aph’s RS-232 port. Check operation of the RS-232 po rt .
Check that at least one of the modem’s LEDs is on.
Check that the data cable is attached to the modem
and to the cardiogr aph’s RS-232 port. when transmitting or receiving ECGs. Check operation of the
RS-232 port.
1. Check that at least one of the modem’s LEDs is
on. Check that the data cable is attached to the
modem and to the car di ograph’s RS-232 port
when transmitting or receiving ECGs. Check
operation of the RS-2 32 port.
2. Retry transmissi on at 2400 bps.
Be sure the remote site is read y to send or re ceive
the transmission.
7127-7141Fax transmission pr obl ems.1. Verify that the remote fax is a group III fa x
machine.
2. Check the fax machine initialization string.
9001-9007Sender time-out while waiting for response from
receiver during transmission using Direct transmis-
sion protocol.
9010-9015Receiver time-out waiting for response from sender.Retry transmission at 2400 bps.
9021Normal Direct transmission link termination.No action. Normal event. Transmission completed
9022Operator stopped tran smis sion at remote site.Contact the remote site.
9023Cardiograph stopped transmission due to a system
error at the local or remote site. The most common
remote site error is missing paper on the receiving
cardiograph.
Retry transmission at 2400 bps.
successfully
Check the receiving device to make sure it is ready
to receive the transmission.
4-18
Troubleshooting Tables
Event #DefinitionCorrective Action
9030-9035, 90429045, 9050, 9051,
9069-9085
9038A non-speci fic error occurred during transmission.Retry tra n s mission at 2400 bps.
9046Remote device not supported.Verify that the remote site is compatible with the
9047-9049Requested an ECG from an unrecognized device.Verify that the remote site is a TraceMaster ECG
9052-9065Premature quit command received from remote site.Contact remote site.
9066Defective FLASH memory1. Perform the extended s el f- t est (refer to Chapter
9067Memory allocation fault duri ng transmission .1. Turn off, wait 30 seconds, turn on.
9086Unrecognized device for t ra ns missionVerify that the remote site is compati ble with th e
9087-9089Printer problem during transmission.1. Check the paper supply.
Direct trans m i ssion protocol error.Retry transmission .
PageWriter cardiograph.
Management System.
2).
2. If the extended sel f -test fails, replac e th e O ption
#A05 PCA.
2. Retry transmissi on.
PageWriter cardiograph.
2. Be sure the printer door is closed.
3. Retry the transmission.
9163Transmission stoppe d unexpectedly.
x of n ECGs sent
Remote site stopped communication
Press any key to cont inue
or
Transmission stopped unexpectedly.
x of n ECGs received
Remote site stopped communication
Press any key to cont inue
9200An ECG was requested from a FAX machine.Select a connection type of Modem, Direct, Modem-
9500-9674Unable to build an SCP record for transmission.Turn off, wait 3 0 seconds, turn on. Ret ry the trans-
9575-9895SCP transmission errors.Retry the transmission. Call Agilent if the error per-
9585, 9586The ECG transmission stalled before completion,
either because the file transfer proto col is absent or
is in an unexpected state, or another communication
error occurred.
1. Verify the trans mission speed matches that of the
remote site.
2. Verify the remote system communication is operating correctly
SCP, or Direct SCP to re quest a n ECG from a remo te
site.
mission. If the error persists, call Agilent.
sists.
W ait one minute, then retry th e transmission. Check
or restart the communication software at the remote
site. If the error persists, call Agilent.
4-19
Troubleshooting Tables
Event #DefinitionCorrective Action
9640, 9769The receiving device rejected a transmitted ECG
because of data error s.
Wait one minute, then retry the transmission. If the
error persists, call Agilent.
Table 4-10Recorder Problems
ProblemCorrective Action
Is the cardiograp h turned on?The LCD screen should be on. Use the shift and
up/down key to incr ease/decrease intensity.
Is the AC power ligh t on?If the cardiograph is plugged in and the AC light is
not on, check fu ses.
Is the battery adequately cha rged?The low battery message (in t he upper- left corner of
the screen) will appear if the bat t ery is not charged.
Is the cardiograp h out of paper or is paper
jammed?
Is the paper sensor lens dirty or obstruct ed?Clean sensor le ns.
Is the paper door com pletely closed?Open paper door slightly and close it. Li sten for the
Is there an erro r m essage?The error messages that display on the screen will
The cardiograph will not record an ECG unless
paper is loaded and no jam exists.
door safety latch to lock.
instruct y ou as to wha t action to take. If it i s something that you can correct, the message will instruct
you as to what to do. If an error number disp la ys,
perform the follo w ing steps:
1. Turn the cardiograph to STANDBY from the
front panel.
4-20
2. Wait 20 seconds or more and turn the unit on
again.
3. Press AUTO or MANUAL. If the cardiograph
turns itself to STANDBY, the battery is not
operating properly.
4. If error persists, call Agilent Medical Call Cen-
ter
1 800 548-8833
Testing the Power Supply
WARNING
CAUTION
NOTE
4Testing the Power Supply
Use this procedure to check the voltages produced by the power supply. The cardiograph
must be connected to AC power to test all the power supp ly voltages. All the voltage
measurements are taken with respect to ground, unless indicated. Refer to Figure 4-2 for
voltage test point locations.
When the cardiograph is connected to AC power, there are dangerous voltages
present in the areas indicated in Figure 4-2 and on the Capacitor Assembly. Do
not touch any exposed metal while the cardiograph is open and AC power is
connected.
1. Open the top cover assembly as described in Chapter 5.
2. Without replacing the top cover, plug the keyboard cable into J201.
3. With the battery removed, plug the unit into AC power.
Be very careful to not short any signal to ground or to other pins. This will result in
cardiograph failure.
4. Verify AC LED is lit. If it is not lit, measure ACON between ground and the back-
plane pin 5 of J3. It should read approximately 5 Vdc. The presence of this voltage
confirms the proper operation of the primary transformer, fuses and ACON regulator
circuit.
5. Measure VSEC between the anode of CR1 (-VSEC) and the cathode of CR8
(+VSEC). It should read approximately 25 Vdc. The presence of this voltage confirms the proper operation of the mains transformer/primary power circuit and the
secondary rectifier filter.
6. Measure VBAT between ground and the point marked VBAT. It should read approxi-
mately 6.7 (minimum) to 7.03 (maximum) Vdc @ 25° C. This vo ltage confirms the
proper operation of the battery charger.
The voltage range for environments with temperatures other than 25° C can be calculated
from the following equations:
Vmin = [2.220 - (T - 25)3.9×10
-3
]3.016, where T = temperature in °C
Vmax = [2.315 - (T - 25)3.9×10-3]3.024
7. Measure UNSW+5V between ground and the point marked UNSW+5V. It should
read 5 Vdc. Its presence confirms the proper operation of the unswitched 5 volt regulator.
4-21
Testing the Power Supply
8. Turn the cardiograph On by pressing .
2Q6WDQGE\
9. Measure SW+5V between ground and the point marked SW+5V. It should read +5
Vdc. Its presence confirms the proper operation of the switched 5 volt regulator and
boost regulator.
10. Turn the cardiograph to Standby.
11. Unplug the ca rdiograph from AC power.
12. Look on the first page of the Extended Self-test printout and verify that the printhead
voltage is approxim ately 26.1 Vdc. If th e volt age is co rrect th en the power su pply test
is complete and the cardiograph may be reassembled. If the voltage is incorrect or the
printer is inoperative then proceed with the following steps.
13. Disconnect the keyboard connector from J201. Place the upper case on top of the bot-
tom case assembly. Turn the unit bottom-side up.
14. Install the battery and turn the unit top-side up.
15. Remove the upper case.
16. Connect the keyboard cable to J201.
17. Turn the cardiograph On.
18. Connect one end of a jumper wire with grabber clips to pin 1 of VR5. The jumper
wire is now connected to +5V. Connect the other end to the lead of R37. The other
end is now connected to PPWRON.
19. Measure VPRINT between ground and the point marked VPRINT. It should read
approximately 26.1 V DC. Its presence confirms the proper operation of the printing
supply circuit.
20. Remove the jumper wire.
21. Turn the cardiograph to Standby.
22. Close the top cover assembly.
4-22
Testing the Power Supply
Figure 4-2 Voltage Test Locations
4-23
Testing the Power Supply
4-24
Removal and Replacement
NOTE
4Introduction
This chapter contains procedures for removing and replacing these cardiograph
subassemblies:
• battery
• keyboard and the keyboard display assembly
• top cover assembly
• printer drive assembly
• printhead assembly
• CPU assembly
• power supply assembly
• internal cabling
• #A05 storage and transmission assembly
The way in which wires and cables are routed and dressed inside the main chassis plays an
important part in reducing electromagnetic and radio freq uency interference emitted by
the cardiograph. When you disassemble any part of the cardiograph, pay special attention
to the way cables and wires are routed and dressed. When you reassemble the cardiograph ,
be sure to route and dress all cables and wires as they were originally.
Tool Requirements
These are the tools you need to remove and replace the cardiograph’s subassemblies:
• 7-mm or 9/ 32-inch wrench
• 7/32-inch wrench
• 2 flat-bladed screwdrivers
• #0 Phillips screwdriver
• T-10 Torx drivers (or Torx driver kit, Agilent part number 5181-1933)
• Long-nose pliers
• IC Extraction Tool (AMP 821591-1 size 44)
5-1
The Battery
NOTE
WARNING
NOTE
The directions in this chapter generally assume you are facing the front of the cardiogr aph
as you work. Thus, for example, the “left side o f the cardiog raph” means “y our left side as
you face the front of the unit.”
4The Battery
This section explains how to remove and replace the battery . These pro cedures apply to all
units.
Removing the Battery
To remove the battery:
1. Turn the cardiograph to Standby.
2. Unplug the ca rdiograph from AC power.
3. Turn the cardiograph bottom-side up.
4. Slide the battery door in the direction of the arrow until it unlatches (ap proxim a tely
1/2 inch), as shown in Figure 5-1. Lift off the door.
5. Unplug the battery connector from the cardiograph by squeezing the locking tabs of
the connector and pulling it straight out from the cardiograph.
6. Remove the battery.
To replace the battery, perform the procedure Removing the Battery in reverse.
Dispose of or recycle depleted batteries according to local regulations. Do not
disassemble, puncture, or incinerate the battery assembly.
To insure full battery capacity, connect the cardiograph to AC power and charge the
battery for 16 hours with the cardiograph in Standby mode.
5-2
The Battery
Figure 5-1 Removing the Battery Door.
A Battery door
5-3
The Keyboard Assembly
CAUTION
5The Keyboard Assembly
This section explains how to remove and replace the keyboard assembly. The keyboard
procedures apply to all models. The keyboard display procedures only apply to
PageWriter 200/200i/300pi.
Note that replacement part numbers for keyboard assembly compo nents vary with the unit
serial number. Refer to the Part List section of this manual to verify replacement part
numbers.
Removing the Keyboard
The keyboard assembly consists of the keyboard overlay, the keyboard assembly, the
flexible circuit, and the keyboard display.
Removing the Keyboard Assembly
Use an electrostatic wrist band or other approved method for protection against
electrostatic discharge when handling the keyboard assembly.
To remove the keyboard assembly:
1.Perform the procedure Removing the Battery.
2. Turn the cardiograph top-side up.
3. Carefully insert a flat-bladed screwdriver into the notches at the front edge of the key-
board overlay and pry the front edge of t he key boar d over l ay up. Refer to F igure 5 -2.
4. Repeat with the rear notches of the keyboard overlay and remove the overlay.
5-4
The Keyboard Assembly
Figure 5-2 Removing the Keyboard Overlay
A. Overlay notches
5. Using two screwdrivers, place the blade of one screwdriver on the leftmost latch of
the keyboard assembly and push toward the rear of the cardiograph. See Figure 5-3.
Carefully insert the blade of the second screwdriver in a notch to the right or left of
the latch and lift up on the keyboard assembly. The leftmost latch should unlatch.
Repeat with the center and rightmost latches.
Figure 5-3 Removing the Keyboard.
A. Notches
B. Latches
5-5
The Keyboard Assembly
CAUTION
6. Carefully lift the assembly and tilt it until the keyboard and control board are accessi-
ble.
7. Disconnect the keyboard ribbon cable from J201 on the control board.
8. Disconnect the green wire from the keyboard.
9. Remove the keyboard.
Do not remove the keyboard circuit board from the keyboard bezel. Removing the board
can impair the long-term reliability of the cardiograph.
Removing the Keyboard Flexible Circuit
To remove the keyboard flexible circuit:
1.Perform the procedures Removing the Battery and Removing the Keyboard Assembly.
2. Place the keyboard assembly with the keys down on a flat surface.
3. Grasp the tabs at the ends of one of the con nectors and g ently pull to ward the flexible
circuit. The locking sleeve should now be extended.
4. Repeat with the other connector.
5. Remove the keyboard flexible circuit.
To replace the keyboard flexible circuit, perform the above procedure in reverse.
Removing the Keyboard Display
The keyboard display consists of the LCD, display shield, and mounting hardware.
1.Perform the procedure Removing the Keyboard Assembly.
2. Using a #0 Phillips screwdriver, remove the 4 screws that secure the LCD display.
3. Remove the display shield and 4 rubber o-rings.
4. To remove the flexible circuit, grasp the tabs at the ends of one of the connectors and
gently pull toward the flexible circuit. The locking sleeve should now be extended.
Repeat with the other connector. Lift the flexible circuit out.
5. Remove the LCD.
To replace the keyboard display, perform the above procedure in reverse.
5-6
The Keyboard Assembly
CAUTION
Replacing the Keyboard
To replace the keyboard, perform the following steps:
1.Connect the green wire to the spade connector on the keyboard.
2. Connect the keyboard ribbon cable to J201 on the control board.
3. Position the top of the keyboard so the four tabs on the top of the keyboard frame are
placed under the lip on keyboard well.
4. Press down on the base of the keyboard until each latch is latched.
The rubber keypad and/or keyboard assembly may be damaged if any key or LED
becomes trapped under the overlay while replacing the overlay. Make sure each and every
key and LED comes through its hole in the overlay.
5. Align the overlay with each of the keyboard keys and LEDs. Position your thumbs
along the front edge of the overlay near each end. Apply pressure on the overlay
toward the rear of the unit and then press down on the overlay . This is one continuous
motion. The front of the overlay should now be snapped into place. Snap the rear of
the overlay into place.
5-7
The Top Cover Assembly
5The Top Cover Assembly
This section describes how to remove and replace the top cover assembly. The top cover
assembly includes the upper chassis, the printer door, and the printer assembly.
Figure 5-4Interior Details of Top Cover Removal
A. Chassis screw
B. Control board screw
C. Printer drive assembly
Opening and Removing the Top Cover Assembly
To open the top cover assembly:
1.Perform the procedures Removing the Battery and Removing the Keyboard Assembly.
2. Turn the cardiograph bottom-side up.
3. Using a Torx T-10 screwdriver, remove the five screws that secure the upper and
lower parts of the chassis.
4. Turn the chassis upright. This will cause the five screws to fall out of the chassis. Per-
form this step slowly to avoid losing the screws.
5. Using a Torx T-10 screwdriver, remove the chassis screw seen through the keyboard
opening. Refer to Figure 5-4.
5-8
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.