Agilent N6705A Product Overview

Agilent DC Power Analyzer
Models: N6705A, N6715A, N6731B-36B, N6741B-46B, N6751-54A, N6761A-62A, N6773A-76A
Product Overview
Ideal for R&D testing and Design Validation
Sources and measures DC voltage and current into your DUT
Combines 1 to 4 DC Power Supplies, DMM, Oscilloscope, Arb waveform
Saves time – no programming required
Eliminates need to gather and configure multiple instruments
Flexible, modular system: Can mix and match DC source
power levels and measurement performance levels to optimize investment
Connect via GPIB, USB or LAN
Fully compliant to LXI Class C specification
Agilent Technologies
For Power
Solutions
in ATE –
See back cover
Due to increasing time-to-market pressures, research and devel­opment engineers often find themselves on tight schedule to work through DUT testing. Along with being driven faster, the R&D engineers can face a high regret factor should their haste result in damaging scarce, complex or expensive DUTs during product development. This is a particular concern when tests involve applying DC power to a DUT. Furthermore, test complexity increases when testing devices that require multiple input voltages, such as printed circuit boards.
Today, when performing DC power-related tests, R&D engineers must gather and configure multiple instruments to complete DC sourcing and measurement tasks. When executing these complex tasks, which can involve simultane­ously connecting to and physi­cally interacting with multiple test instruments, the risk of error increases. In response, R&D engineers may choose to automate tests that are too complex to do manually. Unfortunately, while automating tasks reduces human error, writing and debugging programs adds more work to already overloaded R&D engineers.
The Agilent N6705A DC Power Analyzer represents an entirely new instrument category for R&D engineers. It provides unrivaled productivity gains when sourcing and measuring DC voltage and current into a DUT (Device Under Test). Using the Agilent N6705A DC Power Analyzer, R&D engineers can gain insights into the DUT’s power consumption in minutes without writing a single line of code. It provides an easy-to­use interface, with all sourcing and measuring functions available from the front panel.
The Agilent N6705A DC Power Analyzer saves time
Provides unrivaled productivity gains for sourcing and measuring DC voltage and current into your DUT by integrating up to 4 advanced power supplies with DMM, scope, arb, and datalogger features.
• Eliminates the need to gather multiple pieces of equipment, create complex test setups including transducers (such as current probes and shunts) to measure current into your DUT.
• Eliminates the need to develop and debug programs to control a collection of instruments and take useful measurements because all the functions and measure­ments are available at the front panel.
Agilent N6705A DC Power Analyzer makes these tasks easy, right from the front panel
Setup and view critical turn-on/turn-off sequences
Measure and display voltage, current versus time to visualize power into the DUT
Control DC bias supply ramp-up/down rates
Generate DC bias supply transients and disturbances
Log data for seconds, minutes, hours, or even days to see current consumption or capture anomalies
Save data and screen shots to internal storage or external USB memory devices
Save and name your setup and tests for easy re-use
Share setups with colleagues
2
New Instrument Category for R&D Engineers to Increase Productivity
R&D Engineers are Under Time Pressure
The Agilent N6705A DC Power Analyzer is a modular system that is tailorable to meet spe­cific test needs. At the heart of the DC Power Analyzer is the DC Power Module. The Agilent N6705A DC Power Analyzer is a mainframe that has four slots to accept one to four DC Power Modules. Each DC Power Module takes one slot, except for the N6754A 300 W High-Performance Autoranging DC Power Module, which occupies two slots. This modular design gives you the flexibility to mix and match from 21 different DC Power Modules to create a solution optimized to meet specific test requirements:
The N6730, N6740, and N6770 family of Basic DC Power Modules
50 W, 100 W, and 300 W; up to 100 V, up to 20 A
The N6750 family of High-Performance Autoranging DC Power Modules
50 W, 100 W and 300 W; up to 60 V, up to 20 A
The N6760 family of Precision DC Power Modules
50 W and 100 W; up to 50 V, up to 3 A
3
Modular System Based on DC Power Supply Outputs
Feature Benefit
Integrates capabilities of power supply, Saves time by eliminating the need to find and DMM, scope, arb and datalogger interconnect multiple instruments.
Provides synergistic functions not available from separately connected instruments.
Large color graphics display Fast simple quick set up and monitoring.
Ability to visualize results of multiple channels.
Connections and controls Fast set-up and control. color-coded to the display
Confidence that you are configured and testing correctly.
Intuitive, dedicated physical Fast set-up and control using a familiar controls for common functions interface, with each instrument function
behaving like its standalone counterpart.
Access all capabilities Reduce 90% of the effort associated with without programming set-up by eliminating the need for a PC,
drivers, and software.
Figure 1. The Agilent N6705A DC Power Analyzer
R&D engineers can invest in high-performance outputs where speed and accuracy are needed, or purchase basic performance outputs for sim­ple DC power requirements. In the future, as your test needs change, you can purchase dif­ferent modules and swap them into the DC Power Analyzer, thus creating a solution that protects your investment and grows with you.
Each DC Power Module output is fully isolated and floating from ground and from each other.
Each DC Power Module in the Agilent N6705A DC Power Analyzer has a fully integrated voltmeter and ammeter to measure the actual voltage and current being sourced out of the DC output into the DUT. Because this voltmeter/ammeter function is built-in, it is easy to make measurements without additional wires or the added complexity of current sense resistors or current shunts. The accuracy of the voltage and current measurements are based on the type of mod­ule that is installed (Basic, High-Performance, Precision). You can find the accuracy specification in the tables starting on page 14 under the performance parameter “Voltmeter/Ammeter Measure­ment Accuracy”.
4
Voltmeter/Ammeter
Figure 3. In Meter View, all 4 outputs can be
viewed simultaneously. The both measured values for voltage and current and setting for voltage and current are displayed for each output.
Figure 4. In Meter View, you can also view an enlarged view of one channel, displaying many settings and measured values, for that channel. A summary is shown for the other three channels.
Figure 2. DC Power Modules are easily installed into the N6705A DC Power Analyzer Mainframe.
Each DC Power Module in the Agilent N6705A DC Power Analyzer has a fully integrated digitizer to capture the actual voltage versus time and current versus time being sourced out of the DC output into the DUT. This digitized data is displayed on the large color display just like an oscilloscope. Because this oscilloscope function is built-in, it is possible to make current measurements without current sense resistors, current shunts, or current probes. This greatly reduces measurement setup complexity and provides for accurate and fully specified, calibrated measurements. The accuracy of the measurements in oscilloscope mode is based on the type of module that is in­stalled (Basic, High-Performance Autoranging, and Precision). You can find this information in the specifications tables starting on page 15 under the performance parameter “Oscilloscope Measurement Accuracy”.
The N6760 family of Precision DC Power Modules offer simul­taneous digitizing of output voltage and output current, such that you can view a voltage trace
and a current trace at the same time on the oscilloscope display. For all other module types, you can select to view either a voltage trace or a current trace on the oscilloscope display.
The digitizer in each module runs at 50 kHz and captures 4096 samples per trace. With an effective measurement bandwidth of 10 kHz, this oscilloscope function is perfectly matched to capture time varying events on the DC output, such as peak current demand, dropouts, rise times and other DC transients and disturbances.
The oscilloscope can be triggered on either voltage or current levels. Because the Agilent N6705A DC Power Analyzer is an integrated instrument, the oscilloscope can also be easily configured to trigger on the start of an arbitrary waveform or to trigger when the DC power output is enabled. For example, to make an inrush current measurement on your DUT, you can set the oscilloscope to trigger on the DC output’s on/off key, set the trigger mode to single shot, and then turn on the DC output. This will immediately capture the current flowing out of the DC module into the DUT and give a picture of the inrush current of the DUT. This integrated functionality is not available when using a collection of separate test instruments on the test bench and is an example of how the DC Power Analyzer reduces setup time and complexity.
5
Oscilloscope
Figure 5. In Scope View, voltage and current
traces are displayed. In this picture, the DC current flowing into the DUT is clearly visible as a time-varying waveform.
The Agilent N6705A DC Power Analyzer can also function as a datalogger. Using the measure­ment capability built into each DC Power Module, the N6705A can continuously log data to the large color display and to a file. Data can be logged all four DC outputs at the same time. The accuracy of the logged voltage and current measurements are based on the type of module that is installed (Basic, High­Performance, Precision).
There are two modes of operation:
• In standard mode, measure­ments are made spaced apart by the sample period, which is programmable from 75 milliseconds to 60 seconds. For each DC output, the logged measurements can be voltage measurements, current measurements, or both. Each reading is an integrated volt­age or current measurement. Standard mode datalogging is available on all DC Module types.
• In continuously sampled mode, the built-in digitizer of the DC Power Module runs continuously at 50,000 readings per second. You can specify a sample period, which is the period of time during which these continuous
readings will be accumulated. For each sample period, one average reading (and option­ally, a minimum and maximum value) will be saved. In this mode, the digitizer runs continuously as the readings are averaged and stored;
therefore, the digitizer is always making measurements and no data is missed. The sample period is programma­ble from 1 millisecond to 60 seconds. In this mode, the N6760 family of Precision DC Power Modules offer simultaneous logging of output voltage and output current. For all other module types, you can select to log either voltage or a current when in continuous sampling mode.
The maximum datalog file size is 2 gigabytes, which is approx­imately 500 million readings. The logged data file can be stored on the N6705A’s internal non-volatile RAM (limited to 64 MB) or saved externally on a USB memory device.
The datalogger display can be saved as a GIF file for use in reports. The logged data can be saved for viewing at a later time. Logged data can also be exported to a CSV file that can be read by most common data analysis software packages.
6
Datalogger
Figure 6. In Datalog View, you can log data
on multiple traces. Here, the current flowing out of output 1 and output 2 are captured over 30 seconds
Figure 7. To set up the datalogger, you use a menu screen to select the operating parameters. Menu screens like this are used throughout the DC Power Analyzer for setup
Each DC power output on the Agilent N6705A DC Power Analyzer can be modulated by the module’s built-in arbitrary waveform generator function. This permits the DC output to act as a DC bias transient generator or power arbitrary waveform generator. The max­imum bandwidth is based on the type of module that is installed (Basic, High-Performance, and Precision). See page 21 for a table that lists the bandwidth for each DC Power Module type.
The Agilent N6705A uses run length encoding, where each point in the waveform is defined by the voltage setting and the dwell time or duration to stay at that setting. Waveforms can be generated by specifying only a small number of points. For example, a pulse would only take three points to define it.
The Agilent N6705A offers the following waveform choices:
For each waveform, you can set it to repeat continuously or you can specify the number of times the waveform is repeated. For example, to generate a pulse train of 10 identical pulses, you can program the parameters for one pulse and then specify that you want it to repeat 10 times.
For the user defined voltage and current waveforms, you can download up to 512 set-points
of voltage or current. For each set-point, a dwell time is speci­fied and output will stay (i.e., dwell) at that set-point for the programmed dwell time value. For each of the 512 set-points in the user defined waveform, you can have a different dwell time from 0 to 262 seconds with 1 microsecond resolution.* The module will step thru the user defined table of values, staying at each set-point for the pro­grammed dwell time, and then it will move on to the next point. User defined waveforms can be imported from a CSV file or directly entered from the front panel of the DC Power Analyzer.
*Note that the output response time is less than 5 milliseconds per voltage change, so steps of less than 5 milliseconds will not achieve their final output voltage value before moving on to the next step. This is useful when trying to create a smooth waveform.
7
Arbitrary Waveform Generator
Waveform Number of points per waveform
Sine 100 points
Step 2 points
Ramp 100 points
Pulse 3 points
Stepped ramp (or staircase) Determined by number of steps you program
Exponential 100 points
User defined voltage waveform Up to 512 points (where the output is a voltage source)
User defined current waveform Up to 512 points (where the output is a current source)
Figure 8. The Arb Selection menu is used to select which pre-programmed waveform will be applied to the output of the DC Power Module. Each of the four outputs can have a different waveform applied.
Figure 9. Once you’ve selected a waveform, you simply fill in the blanks to describe the waveform.
Output Sequencing
Each DC power module can be individually set to turn on or to turn off with a delay. By adjusting the delay times and then commanding the Agilent N6705A to turn on, you can set the Agilent N6705A modules to sequence on in a particular order. The same sequencing capability is available to shut down the modules in a particu­lar order. Delay times can be set from no delay to one second of delay in one millisecond increments.
For applications that require more than four DC power mod­ules to be sequenced, this out­put sequencing can be extended across multiple Agilent N6705A mainframes. By wiring together the I/O ports on the rear panel of the mainframes, a pair of synchronization signals is sent between mainframes, allowing the output sequences of each mainframe to be synchronized. This capability is also supported
to link Agilent N6705A output sequences with outputs installed in N6700B, N6701A and N6702A Low-Profile Modular Power System mainframes. It is not supported on N6700A main­frames.
Programmable Voltage Slew
For some applications, like inrush limiting or powering rate-sensitive devices, it is nec­essary to slow down and con­trol the speed of the DC output to maintain a specific voltage slew rate. The Agilent N6705A provides programmable voltage slew rate, so that you can easily control the speed at which the output slews from one voltage to another. You can set the speed of a voltage change anywhere from its maximum up/down programming speed to its slowest change of up to 10 seconds.
Series and Parallel Operation
To increase available voltage and power per output, identically rated outputs can be operated directly in series. The maxi­mum series operation is 240 V. To increase the available cur­rent and power per output, identically rated outputs can be operated directly in parallel. The maximum rated parallel operation is 80 A per Agilent N6705A.
Convenient Front Panel Connections
The N6705A uses 3-way binding posts on the front panel for connection to the DUT. The
binding posts accept standard banana plugs, bare wire, and spade-lug connectors. The binding posts are rated for 20 A per connection. To avoid setup and connection errors, the binding posts are color-coded to the control keys and the display.
4-wire Sensing for Improved Measurement Accuracy
To improve the voltage measure­ment accuracy and regulation of the DC outputs, the Agilent N6705A DC Power Analyzer offers 4-wire sensing capabili­ty, also called remote sensing, on each of the four outputs of the DC Power Analyzer. 4-wire remote sensing is useful when the DUT draws high current and you want to account for voltage drop in the power leads to achieve tight regulation and high voltage measurement accuracy. To use 4-wire sens­ing, in addition to your power leads, you connect two low current sensing leads between the DUT input terminals and 4-wire sense terminal binding posts located on the front of the N6705A power supply. This permits the output mod­ule to monitor and regulate its output voltage directly at the DUT input terminals instead of the N6705A front panel output binding posts. It then automatically adjusts its output voltage to compensate for volt­age drops across the resistance in the power leads. For conve­nience, switching between 2-wire mode (local sensing)
8
Figure 10. The Output On/Off Delays screen allows you to enter the delay times for each output. A graphical representation of the set­tings is shown to visually confirm your choices.
Additional Features
and 4-wire mode (remote sensing) is done via an internal relay inside the N6705A DC Power Analyzer, eliminating the need for shorting bars or jumpers commonly found on other bench power sources.
DC Power Modules Offer Low Noise Outputs
Careful attention has been paid to this design to ensure low normal mode noise (ripple and peak-peak) as well as low common mode noise. While all DC Power Modules are switch­ing power supplies, the N6750 High-Performance Autoranging DC Power Modules and the N6760 Precision DC Power Modules are switching power supply designs that outperform most linear power supplies on the market.
DC Power Modules Provide Fast Voltage Changes
When it comes to speed, the N6750 High-Performance Auto­ranging DC Power Modules and the N6760 Precision DC Power Modules achieve performance unlike a typical DC power sup­ply. Thanks to an active down­programming circuit to rapidly pull down the output when lowering the module’s output voltage, the N6750/60 can rapidly program both up and down in voltage. Changing volt­age from 0 V to 50 V, or 50 V to 0 V, can be accomplished in less than 1.5 milliseconds. And
for smaller voltage changes, for example from 0 V to 5 V or 5 V to 0 V, the programming speed is less than 200 microseconds. These output speeds allow the N6750/60 to give maximum system throughput when your test calls for frequent changes in power supply voltage settings.
Autoranging for Flexibility
The N6750 High-Performance Autoranging DC Power Modules and the N6760 Precision DC Power Modules give you even more flexibility by providing autoranging outputs. This autoranging capability provides maximum output power at any output voltage up to 50 V. This allows one power supply to do the job of several power supplies because its operating range covers low voltage, high current as well as high voltage, low current operating points. For example, the N6751A High­Performance, Autoranging DC
Module, rated at 50 V, 5 A, and 50 W can provide full power at 10 V @ 5 A (=50 W), 20 V @ 2.5 A (= 50 W),
33.3 V @ 1.5 A (= 50 W), 50 V @ 1 A (= 50 W) or anywhere in between. There­fore, this 50 W autoranging power supply, due to its extended voltage and current range, can produce voltage and current combinations in the range of a 250 W non-autoranging power supply. See page 27 for a diagram describing the details of the autoranging output char­acteristics of the N6750 and N6760 families of DC Power Modules.
Use the N6760 DC Power Modules for Precision Low-level Performance
The N6760 Precision DC Power Modules provide dual ranges on both programming and measurement. In the low range, these power supplies provide precision in the milliampere
9
Figure 11. The Agilent N6705A DC Power Analyzer is a perfect size for the bench.
and microampere regions. They are ideally suited for semicon­ductor and passive device testing, or where a precisely controlled output and highly accurate, precise measurements are needed during test.
Use the N6730/40/70 DC Power Modules for Basic DC Bias
Not all testing require high performance power supplies. When your budget is tight and when speed and accuracy are a low consideration, the N6730 family of 50 W DC Power Modules, the N6740 family of 100 W DC Power Modules, and the N6770 family of 300 W DC Power Modules are an economical solution that give you clean, reliable DC power.
Real Time Clock
The Agilent N6705A DC Power Analyzer has a built-in battery backed real time clock. This allows for proper time-stamping of logged data. It is also used to tag files with correct creation dates.
Internal Memory
The Agilent N6705A DC Power Analyzer has 64 megabytes of non-volatile storage. This storage is shared between the four DC outputs. It can be used for saving test setups, test results, and screen images. The 64 mega­bytes of internal storage is large enough to hold up to 30 minutes of data at a reading rate of 1 millisecond per reading on all 4 channels simultaneously. External USB storage is sup­ported for increased storage capacity to log data longer (see section “Front Panel USB”).
Front Panel USB
The Agilent N6705A DC Power Analyzer provides a convenient front panel USB port designed exclusively for data storage devices, such as USB memory devices or USB hard drives. On devices connected to this USB port, you can save test setups, test results, and screen images. It is also an easy way to move test setup files between two N6705A DC Power Analyzers or test results between the DC Power Analyzer and a PC. You can also log data directly to the USB device plugged into the front panel. This extends the total storage capability of the N6705A beyond the 64 MB limit of its internal non-volatile RAM.
Emergency Stop
Should a hazardous situation occur during testing, you can press the large red Emergency Stop button on the front panel of the Agilent N6705A. Pressing this easy-to-find button imme­diately removes power from the DC outputs. However, any data collection (such as a scope trace or data log) that is run­ning at the time will continue to run. By doing so, the data you were collecting is saved and you will get a record of what was happening at the time of the event that caused you to press the Emergency Stop button. The measurements could aid in failure analysis, repair, or debugging of the DUT.
DUT Protection Features
Each Agilent N6705A DC Power Module is protected against over-voltage, over-current, and over-temperature. A fault con­dition in one module can be detected within 10 microseconds by other modules so that they can be quickly shut down to avoid hazardous conditions on your DUT.
10
Figure 12. The N6705A’s front panel USB port
Figure 13. The Emergency Stop button shuts
down all outputs immediately
Output Disconnect and Polarity Reversal Relays
Modules in the Agilent N6705A can be individually ordered with optional Output Disconnect Relays (option 761) or Output Disconnect/Polarity Reversal Relays (option 760). See table on page 29 for option 760 and 761 availability. All relays are built into the module, so no additional wiring is needed to get the relay function. With option 761, Output Disconnect Relays, an emergency condition or turning the DC output off causes mechanical relays disconnect both the plus and minus side of the power sup­ply, including the sense leads. With option 760, Output Disconnect/Polarity Reversal Relays, mechanical relays switch the leads on both the plus and the minus side of the power supply, including the sense leads, resulting in a voltage polarity reversal at the DUT. In addition to polarity reversal, option 760 provides the same output disconnect function as option 761.
Note: Output current is limited on some modules when option 760 Output Disconnect/Polarity Reversal Relays is installed. See the “Available options” tables at the bottom of page {29 and page 31} for more information about maximum current limitations with option 760.
Triggering
The Agilent N6705A DC Power Analyzer has hardware trigger in/trigger out signals which permit the Agilent N6705A to be synchronized with other test equipment. For example, when you turn on the outputs of the Agilent N6705A, it can generate a trigger signal to start a measurement on an RF power meter.
Connectivity
The Agilent N6705A DC Power Analyzer comes standard with GPIB, USB 2.0, and 10/100 Base-T Ethernet LAN interfaces. The Agilent N6705A is fully compliant with the LXI class C specification.
Security
All non-volatile RAM data and settings can be cleared from the front panel. For customers who have security concerns about USB access to internally stored test data and setups, the Agilent N6705A also offers option AKY, which removes the USB ports from the front and rear of the Agilent N6705A. When used in systems running GPIB, the LAN and/or USB interfaces can be disabled for extra security.
Control from any Browser
The Agilent N6705A can be controlled via a standard web browser. The Agilent N6705A contains a web server that provides a webpage containing
a graphical front panel repre­sentation of the Agilent N6705A front panel. The WebGUI oper­ation is identical to operating the real front panel on the Agilent N6705A DC Power Analyzer.
Drivers
For customers who wish to operate the DC Power Analyzer under computer control, the Agilent N6705A comes with both VXIplug&play drivers and IVI-COM drivers. LabView dri­vers are available at NI.COM.
Programming Language
The Agilent N6705A supports SCPI (Standard Commands for Programmable Instruments). Note that the Agilent N6705A’s command set is compatible with the N6700 Modular Power System for ATE, so programs written for the Agilent N6700 will work on the Agilent N6705A.
Firmware Updates
The Agilent N6705A firmware is stored in FLASH ROM and can be easily updated when new features become available. Firmware can be downloaded into the Agilent N6705A over GPIB, LAN, or USB using the supplied firmware update utili­ty program. Firmware updates can be found at
www.agilent.com/find/N6705firmware.
11
Power Management Feature Allows You to Allocate N6705A Mainframe Power
Often, a DUT requires some high power DC sources and several very low power DC sources. In this case, you may choose to configure a system where the sum of the power modules installed in the Agilent N6705A exceeds the total power avail­able from the Agilent N6705A mainframe. The power manage­ment features of the Agilent N6705A allow you to allocate mainframe power to the outputs where it’s needed, achieving maximum asset utilization and flexibility. This feature provides safety from unexpected and dangerous shutdowns that can occur with power systems with­out power management when operated in a similar way. For example, if your DUT requires 250 W on two of its inputs, but
only 10 W each two auxiliary inputs, you can configure a sys­tem consisting of two 300 W DC modules and two 50 W DC modules. Even though the sum of the module power is greater than 600 W, you can still use the Agilent N6705A. Thanks to the power management feature, you can allocate the 250 W each to the two 300 W modules while you allocate only 25 W to each of the 50 W modules.
Universal AC Input
The Agilent N6705A has a uni­versal input that operates from 100-240 VAC, 50/60/400 Hz. There are no switches to set or fuses to change when switching from one voltage standard to another. The AC input employs power factor correction.
12
Makes a great tool for ATE systems that require an advanced user interface for test and debug
While the Agilent N6705A DC Power Analyzer is designed primarily as an R&D bench tool, customers building ATE systems may find the Agilent N6705A has great utility in an ATE system. It is fully programmable, LXI class C certified, and takes the same commands as the Agilent N6700. But thanks to its large display and easy to use controls, test engineers may find the Agilent N6705A makes a great tool for visualizing test results as the tests execute, for DUT trou­bleshooting, for DUT debugging, and for ATE test development, The Agilent N6705A mounts in a stan­dard 19" rack using standard rack mount hardware for 4 U instruments.
Achieve correlation and share data between R&D and Manufacturing
The Agilent N6705A DC Power Analyzer is a modular system that uses the same DC Power Modules as the N6700 Low-Profile Modular Power System for ATE. Customers who use the N6705A in R&D and the N6700 in Manufacturing can easily achieve test correlation between R&D testing, design characterization/validation testing, and manufacturing test­ing because the DC Power Modules are common to both the bench and ATE versions of the product. Test programs can be easily shared between R&D and manufacturing since the Agilent N6705A and the Agilent N6700 share a common com­mand set.
13
N6750 Family
For applications where the power supply plays a critical role
The Agilent N6750 family of high-performance, auto ranging DC power modules provides low noise, high accuracy and output voltage changes that are up to 10 to 50 times faster than other power supplies. In addi­tion, auto ranging output capa­bilities enable one power supply to do the job of several tradi­tional power supplies. The N6750 family combines widest the arbitrary waveform genera­tor bandwidth, available power up to 300 W, and high accuracy measurements. In oscilloscope mode, they can be configured to display either a voltage or a current trace.
N6760 Family
For applications where precision is required
The Agilent N6760 family of precision DC power modules provides fast outputs changes (for wide arbitrary waveform generator bandwidth) and 16-bit voltage and current programming, and 18-bit measurements for precision in the milliampere and microam­pere region. In oscilloscope mode, the N6760 family offers the ability to view both a voltage and a current trace simultaneously.
N6730/40/70 Family
For basic DC applications
The Agilent N6730, N6740 and N6770 families of DC power modules provide programmable voltage and current, measure­ment and protection features at a very economical price. These modules offer a wide range of voltage, current, and power outputs. In oscilloscope mode, they can be configured to display either a voltage or a current trace.
Choosing the Right DC Power Modules to Meet Your Testing Needs
See detailed specifications on page 14
14
Agilent N6751A/N6752A, N6754A and N6761A/N6762A Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C after a 30-minute warm-up period, with each module’s sense terminals externally jumpered directly to their respective output terminals (local sensing).
N6751A / N6754A N6761A / N6752A N6762A
DC Output Ratings
Voltage 50 V 60 V 50 V
Current (derated 1% per °C above 40°C) 5 A / 10 A 20 A 1.5 A / 3 A
Power 50 W / 100 W 300 W 50 W / 100 W
Output Ripple and Noise (PARD)
(from 20 Hz – 20 MHz) CV peak-to-peak 4.5 mV 6 mV 4.5 mV
CV rms 350 µV 1 mV 350 µV
Load Effect (Regulation)
(for any output load change, Voltage 2 mV 2 mV 0.5 mV with a maximum load-lead Current (@ 0 - 7 V) 2 mA 5 mA 30 µA drop of 1 V per lead) (@ 0 - 50 V) 2 mA 5 mA 65 µA
Source Effect (Regulation) Voltage 1 mV 1.2 mV 0.5 mV
Current 1 mA 2 mA 30 µA
Programming Accuracy
(at 23°C ±5°C Voltage high range 0.06% + 19 mV 0.06% + 25 mV 0.016% + 6 mV after 30 minute Voltage low range (5.5 V) N/A N/A 0.016% + 1.5 mV warm-up. Applies Current high range 0.1% + 20 mA 0.10% + 8 mA 0.04% + 200 µA from min. to max. Current low range (100 mA, @ 0 - 7 V) N/A N/A 0.04% + 15 µA programming range) (100 mA, @ 0 - 50 V) N/A N/A 0.04% + 55 µA
Voltmeter/Ammeter Measurement Accuracy
(at 23°C ±5°C) Voltage high range 0.05% + 20 mV 0.05% + 25 mV 0.016% + 6 mV
Voltage low range (5.5 V) N/A N/A 0.016% + 1.5 mV
Current high range 0.1% + 4 mA 0.10% + 8 mA 0.04% + 160 µA
Current low range (100 mA, @ 0 - 7 V)
NOTE 1
N/A N/A 0.03% + 15 µA
NOTE 2
(100 mA, @ 0 - 50 V) N/A N/A 0.03% + 55 µA
Load Transient Recovery Time
(time to recover to within the settling band following a load change)
• from 60% to 100% and from 100% to 60% of full load for models N6751A & N6761A
• from 50% to 100% and from 100% to 50% of full load for models N6752A-N6754A & N6762A.
Voltage settling band ± 75 mV
NOTE 2
± 90 mV ± 75 mV
Time < 100 µs < 100 µs < 100 µs
1
Applies when measuring 4096 data points (SENSe:SWEep:POINts = 4096).
2
Settling band is ±125 mV for Model N6752A when relay option 761 is installed.
15
Agilent N6751A/N6752A, N6754A and N6761A/N6762A Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or type testing. All supplemental characteristics are typical unless otherwise noted.
N6751A / N6754A N6761A / N6752A N6762A
Programming Ranges
Voltage high range 20 mV – 51 V 25 mV – 61.2 V 15 mV – 51 V
Voltage low range (5.5 V) N/A N/A 12 mV – 5.5 V
Current high range
10 mA – 5.1 A/ 20 mA – 20.4 A 1 mA – 1.53 A/ 10 mA – 10.2 A 1 mA – 3.06 A
Current low range (0.1 A) N/A
N/A
0.1 mA – 0.1 A
NOTE 1
Programming Resolution
Voltage high range 3.5 mV
NOTE 2
4.2 mV
NOTE 2
880 µV
NOTE 3
Voltage low range (5.5 V) N/A N/A 90 µV
Current high range 3.25 mA
NOTE 4
6.5 mA
NOTE 4
60 µA
Current low range (0.1 A) N/A N/A 2 µA
Measurement Resolution
Voltage high range 1.8 mV
NOTE 5
2.2 mV
NOTE 5
440 µV
NOTE 6
Voltage low range (5.5 V) N/A N/A 44 µV
Current high range 410 µA 820 µA 30 µA
Current low range (0.1 A) N/A N/A 1 µA
Programming Temperature Coefficient per °C
Voltage high range 18 ppm + 160 µV 20 ppm + 50 µV 18 ppm + 140 µV
Voltage low range (5.5 V) N/A N/A 40 ppm + 70 µV
Current high range 100 ppm + 45 µA 60 ppm + 200 µA 33 ppm + 10 µA
Current low range (0.1 A) N/A N/A 60 ppm + 1.5 µA
Measurement Temperature Coefficient per °C
Voltage high range 25 ppm + 35 µV 20 ppm + 50 µV 23 ppm + 40 µV
Voltage low range (5.5 V) N/A N/A 30 ppm + 40 µV
Current high range 60 ppm + 3 µA 60 ppm + 12 µA 40 ppm + 0.3 µA
Current low range (0.1 A) N/A N/A 50 ppm + 0.3 µA
Output Ripple and Noise (PARD)
CC rms 2 mA 4 mA 2 mA
Common Mode Noise
(from 20 Hz – 20 MHz; rms 500 µA 750 µA 500 µA from either output to chassis) peak-to-peak < 2 mA 3 mA < 2 mA
Over-voltage Protection
Accuracy 0.25% + 250 mV 0.25 % +300 mV
NOTE 9
0.25% + 250 mV
Maximum setting 55 V 66 V 55 V
Response time
50 µs from occurrence of over-voltage condition to start of output shutdown
Oscilloscope Measurement Accuracy
(at 23°C ±5°C, Voltage 0.05% + 32 mV 0.05% + 34 mV 0.016% + 16 mV Accuracy of any Current high range with current compensation on
NOTE 7
0.1% + 14 mA 0.1% + 16 mA 0.04% + 10 mA
individual point Current high range with current compensation off
NOTE 8
0.1% + 8 mA 0.1% + 16 mA 0.04% + 1 mA
in the trace) Current low range with current compensation on
NOTE 7
N/A N/A 0.03% + 10 mA
Current low range with current compensation off
NOTE 8
N/A N/A 0.03% + 175 µA
1
If you are operating the unit below 255 µA in constant current mode, the output may become unregulated with the following load conditions: The load resistance is <175 mand the load inductance is >20 µH. If this occurs, an UNRegulated flag will be generated and the output current may rise above the programmed value but will remain less than 255 µA.
2
Based on 14-bit DAC, with DAC range adjusted by software calibration
3
Based on 16-bit DAC, with DAC range adjusted by software calibration
4
Based on 12-bit DAC, with DAC range adjusted by software calibration
5
Based on 16-bit ADC (15 bits plus sign), with ADC range adjusted by software calibration
6
Based on 18-bit ADC (17 bits plus sign), with ADC range adjusted by software calibration
7
Current compensation on means the selection labeled “Compensate current measurements during voltage transients” is checked on the Meter Properties screen
8
Current compensation off means the selection labeled “Compensate current measurements during voltage transients” is not checked on the Meter Properties screen
9
Over-voltage Protection Accuracy is 0.25% + 600 mV with relay option 760 or 761 installed
Agilent N6751A/N6752A, N6754A and N6761A/N6762A Supplemental Characteristics
(Continued)
N6751A / N6754A N6761A / N6752A N6762A
Maximum Up-programming Time with full resistive load
(time from 10% to 90% For voltage change of 0 to 10 V 0 to 15 V 0 to 10 V of total voltage excursion) Up-programming time 0.2 ms 0.35 ms 0.6 ms
For voltage change of 0 to 50 V 0 to 60 V 0 to 50 V Up-programming time 1.5 ms 2.0 ms 2.2 ms
Maximum Up-programming Settling Time with full resistive load
(time from start of voltage change For voltage change of 0 to 10 V 0 to 15 V 0 to 10 V to within 50 mV of final value) Up-programming settling time 0.5 ms 0.8 ms 0.9 ms
For voltage change of 0 to 50 V 0 to 60 V 0 to 50 V
Up-programming settling time 4.0 ms 4.2 ms 4.0 ms
Maximum Down-programming Time with no load
(time from start of voltage change For voltage change of 10 to 0 V 15 to 0 V 10 to 0 V to output voltage < 0.5 V) Down-programming time 0.3 ms 0.6 ms 0.3 ms
For voltage change of 50 to 0 V 60 to 0 V 50 to 0 V
Down-programming time 1.3 ms 2.2 ms 1.3 ms
Maximum Down-programming Settling Time with no load
(time from start of voltage For voltage change of 10 to 0 V 15 to 0.8 V 10 to 0 V change to output voltage Down-programming settling time 0.45 ms 0.8 ms 0.45 ms within 50 mV of final value)
For voltage change of 50 to 0 V 60 to 0 V 50 to 0 V
Down-programming settling time 1.4 ms 2.3 ms 1.4 ms
Down-programming Time with capacitive load
NOTE 1
(time from start of voltage change Capacitive load 1000 µF
NOTE 1
680 µF
NOTE 3
1000 µF
NOTE 1
to output voltage < 0.5V)
For voltage change of 10 to 0 V 15 to 0 V 10 to 0 V
Down-programming time 0.3 ms 2.3 ms 0.3 ms
For voltage change of 50 to 0 V 60 to 0 V 50 to 0 V
Down-programming time 1.3 ms 10.0 ms 1.3 ms
Down-programming Capability
Continuous power 7 W 12.5 W 7 W
Peak current 7 A 6 A 3.8 A
Remote Sense Capability
Outputs can maintain specifications with up to a 1-volt drop per load lead.
Series and Parallel Operation
Identically rated outputs can be operated directly in parallel or can be connected for straight series operation. Auto-series and auto-parallel operation is not available.
1
Modules can discharge a 1000 µF capacitor from 50 V to 0 V at a rate of 4 times/second.
2
Modules can discharge a 4700 µF capacitor from 20 V to 0 V at a rate of 4 times/second.
3
Modules can discharge a 680 µF capacitor from 60 V to 0 V at a rate of 4 times/second.
16
17
Agilent N6731B - N6736B and N6741B - N6746B Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C after a 30-minute warm-up period, with each module’s sense terminals externally jumpered directly to their respective output terminals (local sensing)
N6731B/ N6732B/ N6733B/ N6734B/ N6735B/ N6736B/ N6741B N6742B N6743B N6744B N6745B N6746B
DC Output Ratings:
Voltage 5 V 8 V 20 V 35 V 60 V 100 V
Current
NOTE 1
10 A / 20 A 6.25 A / 12.5 A
NOTE 3
2.5 A / 5 A 1.5 A / 3 A 0.8 A / 1.6 A 0.5 A / 1 A
Power 50 W / 100 W 50 W / 100 W 50 W / 100 W 52.5 W / 105 W 50 W / 100 W 50 W / 100 W
Output Ripple and Noise (PARD)
(from 20 Hz – 20 MHz) CV peak-to- peak 10 mV / 11 mV 12 mV 14 mV 15 mV 25 mV 30 mV
CV rms 2 mV 2 mV 3 mV 5 mV 9 mV 18 mV
Load Effect (Regulation)
(with output change Voltage 5 mV 6 mV 9 mV 11 mV 13 mV / 16 mV 20 mV / 30 mV from no load to Current 2 mA 2 mA 2 mA 2 mA 2 mA 2 mA full load, up to a maximum load-lead drop of 1 V/lead)
Source Effect (Regulation) Voltage 1 mV 2 mV 2 mV 4 mV 6 mV 10 mV
Current 1 mA 1 mA 1 mA 1 mA 1 mA 1 mA
Programming Accuracy
(@ 23 °C ±5°C after
Voltage 0.1% + 19 mV 0.1% + 19 mV 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% + 100 mV
30 minute warm-up.
Current 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 20 mA 0.15% + 10 mA
Applies from minimum to maximum program­ming range)
Voltmeter/Ammeter Measurement Accuracy
(at 23°C ±5°C)
Voltage 0.1% + 20 mV 0.1% + 20 mV 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% + 100 mV Current 0.15% + 20 mA 0.15% + 10 mA 0.15% + 5 mA 0.15% + 4 mA 0.15% + 4 mA 0.15% + 2 mA
Load Transient Recovery Time
(time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load.)
Voltage settling band
±0.08 V / 0.1 V
NOTE 2
±0.08 V / 0.1 V
NOTE 2
± 0.2 V / 0.3 V ± 0.2 V / 0.3 V ± 0.4 V / 0.5 V ± 0.5 V / 1.0 V
Time < 200 µs < 200 µs < 200 µs < 200 µs < 200 µs < 200 µs
1
Output current is derated 1% per °C above 40°C.
2
Settling band is ±0.10 V/0.125 V for 5 V and 8 V Models when relay options 760 and 761 are installed.
3
For N6742B, output current is limited to 10 A when option 760 Output Disconnect/Polarity Reversal Relays is installed.
N6731B/ N6732B/ N6733B/ N6734B/ N6735B/ N6736B/ N6741B N6742B N6743B N6744B N6745B N6746B
Programming Ranges
Voltage 15 mV – 5 .1 V 15 mV – 8 .16 V 30 mV – 20.4 V 40 mV – 35.7 V 70 mV – 61.2 V 100 mV – 102 V Current 60 mA – 10.2 A/ 40 mA –6.375 A/ 10 mA – 2.55 A/ 5 mA – 1.53 A/ 2.5 mA – 0.85 A/ 1.5 mA – 0.51 A/
60 mA – 20.4 A 40 mA – 12.75 A 10 mA – 5.1 A 5 mA – 3.06 A 2.5 mA – 1.7 A 1.5 mA – 1.02 A
Programming Resolution
NOTE 1
Voltage 3.5 mV 4 mV 7 mV 10 mV 18 mV 28 mV Current 7 mA 4 mA 3 mA 2 mA 1 mA 0.5 mA
Measurement Resolution
NOTE 2
Voltage 3 mV 4 mV 10 mV 18 mV 30 mV 50 mV Current 10 mA 7 mA 3 mA 2 mA 1 mA 0.5 mA
Programming Temperature Coefficient per °C Voltage 0.005% + 0.1 mV 0.005% + 0.1 mV 0.005% + 0.2 mV 0.005% + 0.5 mV 0.005% + 0.5 mV 0.005% + 1 mV
Current 0.005% + 1 mA 0.005% + 0.5 mA 0.005% + 0.1 mA 0.005% + 0.05 mA 0.005% + 0.02 mA 0.005% + 0.02 mA
Measurement Temperature Coefficient per °C Voltage 0.01% + 0.1 mV 0.01% + 0.1 mV 0.01% + 0.2 mV 0.01% + 0.2 mV 0.01% + 0.5 mV 0.01% + 0.5 mV
Current 0.01% + 1 mA 0.01% + 0.5 mA 0.01% + 0.1 mA 0.01% + 0.05 mA 0.01% + 0.02 mA 0.01% + 0.02 mA
Output Ripple and Noise (PARD) CC rms 8 mA 4 mA 2 mA 2 mA 2 mA 2 mA
Common Mode Noise
(from 20 Hz – 20 MHz; from either output to chassis) rms 1 mA 1 mA 1 mA 1 mA 1 mA 1 mA
peak-to-peak < 15 mA < 10 mA < 10 mA < 10 mA < 10 mA < 10 mA
Over-voltage Protection
Accuracy 0.25% + 50 mV 0.25% + 50 mV 0.25% + 75 mV 0.25% + 100 mV 0.25% + 200 mV 0.25% + 250 mV Accuracy w/opt 760 0.25% + 600 mV 0.25% + 600 mV 0.25% + 350 mV 0.25% + 250 mV 0.25% + 300 mV 0.25% + 300 mV Accuracy w/opt 761 0.25% + 600 mV 0.25% + 600 mV 0.25% + 350 mV 0.25% + 250 mV 0.25% + 300 mV 0.25% + 300 mV
Maximum setting 7.5 V 10 V 22 V 38.5 V 66 V 110 V
Response time 50 µs from occurrence of over-voltage condition to start of output shutdown
Maximum Up-programming and Down-programming Time with full resistive load
(time from 10% to 90% of total voltage excursion)
Voltage setting from 0 V to 20 ms 20 ms 20 ms 20 ms 20 ms 20 ms full scale and full scale to 0 V
Maximum Up-programming and Down-programming Settling Time with full resistive load
(time from start of voltage change until voltage settles within 0.1% of the full-scale voltage of its final value)
Voltage setting from 0 V to 100 ms 100 ms 100 ms 100 ms 100 ms 100 ms full scale and full scale to 0 V
Remote Sense Capability Outputs can maintain specifications with up to a 1-volt drop per load lead.
Series and Parallel Operation Identically rated outputs can be operated directly in parallel or can be connected for straight series operation.
Auto-series and auto-parallel operation is not available.
Oscilloscope Measurement Accuracy
(at 23°C ±5°C, Accuracy of any individual point in the trace)
Voltage 0.1% + 25 mV 0.1% + 30 mV 0.1% + 45 mV 0.1% + 75 mV 0.1% + 130 mV 0.1% + 190 mV
Current with current compensation on
NOTE 3
0.15% + 70 mA 0.15% + 40 mA 0.15% + 20 mA 0.15% + 14 mA 0.15% + 12 mA 0.15% + 7 mA
Current with current compensation off
NOTE 4
0.15% + 50 mA 0.15% + 30 mA 0.15% + 15 mA 0.15% + 10 mA 0.15% + 9 mA 0.15% + 5 mA
1
Based on 12-bit DAC, with DAC range adjusted by software calibration
2
Based on 12-bit ADC (11 bits plus sign), with ADC range adjusted by software calibration
3
Current compensation on means the selection labeled “Compensate current measurements during voltage transients” is checked on the Meter Properties screen
4
Current compensation off means the selection labeled “Compensate current measurements during voltage transients” is not checked on the Meter Properties screen
Agilent N6731B - N6736B and N6741B - N6746B Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or type testing. All supplemental characteristics are typical unless otherwise noted.
18
19
Agilent N6773A - N6776A Performance Specifications
Unless otherwise noted, specifications are warranted over the ambient temperature range of 0 to 55°C after a 30-minute warm-up period, with each module's sense terminals externally jumpered directly to their respective output terminals (local sensing).
N6773A N6774A N6775A N6776A
DC Output Ratings
Voltage 20 V 35 V 60 V 100 V
Current
NOTE 1
15 A
NOTE 3
8.5 A 5 A 3 A
Power 300 W 300 W 300 W 300 W
Output Ripple and Noise (PARD)
(from 20 Hz – 20 MHz) CV peak-to- peak 20 mV 22 mV 35 mV 45 mV
CV rms 3 mV 5 mV 9 mV 18 mV
Load Effect (Regulation)
(with output change from no load Voltage 13 mV 16 mV 24 mV 45 mV to full load, up to a maximum Current 6 mA 6 mA 6 mA 6 mA load-lead drop of 1 V/lead)
Source Effect (Regulation) Voltage 2 mV 4 mV 6 mV 10 mV
Current 1 mA 1 mA 1 mA 1 mA
Programming Accuracy:
(@ 23°C ±5°C after 30 minute Voltage 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% +100 mV warm-up. Applies from minimum Current 0.15% + 60 mA 0.15% + 60 mA 0.15% + 60 mA 0.15% + 30 mA to maximum programming range)
Voltmeter/Ammeter Measurement Accuracy
(at 23°C ±5°C) Voltage 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% +100 mV
Current 0.15% + 15 mA 0.15% + 12 mA 0.15% + 12 mA 0.15% + 6 mA
Load Transient Recovery Time
(time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load.)
Voltage settling band ± 0.3 V
NOTE 2
± 0.3 V
NOTE 2
± 0.5 V ± 1.0 V
Time < 250 µs < 250 µs < 250 µs < 250 µs
1
Output current is derated 1% per °C above 40°C.
2
Settling band is ±0.35 V for 20 V and 35 V Models when relay options 760 and 761 are installed.
3
For N6773A, output current is limited to 10 A when option 760 Output Disconnect/Polarity Reversal Relays is installed.
Agilent N6773A - N6776A Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or type testing. All supplemental characteristics are typical unless otherwise noted
N6773A N6774A N6775A N6776A
Programming Ranges
Voltage 30 mV – 20.4 V 40 mV – 35.7 V 70 mV – 61.2 V 100 mV – 102 V Current 30 mA – 15.3 A 15 mA – 8.67 A 7.5 mA – 5.1 A 4.5 mA – 3.06 A
Programming Resolution
NOTE 1
Voltage 7 mV 10 mV 18 mV 28 mV Current 9 mA 6 mA 3 mA 1.5 mA
Measurement Resolution
NOTE 2
Voltage 10 mV 18 mV 30 mV 50 mV Current 9 mA 6 mA 3 mA 1.5 mA
Programming Temperature Coefficient per °C Voltage 0.01% + 0.2 mV 0.01% + 0.5 mV 0.01% + 0.5 mV 0.01% + 1 mV
Current 0.01% + 0.5 mA 0.01% + 0.5 mA 0.01% + 0.1 mA 0.01% + 0.1 mA
Measurement Temperature Coefficient per °C Voltage 0.01% + 0.2 mV 0.01% + 0.2 mV 0.01% + 0.5 mV 0.01% + 0.5 mV
Current 0.01% + 0.5 mA 0.01% + 0.5 mA 0.01% + 0.05 mA 0.01% + 0.05 mA
Output Ripple and Noise (PARD)
CC rms 6 mA 6 mA 6 mA 6 mA
Common Mode Noise
(from 20 Hz – 20 MHz; from Rms 2 mA 2 mA 2 mA 2 mA either output to chassis) Peak-to- peak < 20 mA < 20 mA < 20 mA < 20 mA
Over-voltage Protection
Accuracy 0.25% +100 mV 0.25% + 130 mV 0.25% + 260 mV 0.25% + 650 mV Accuracy w/opt 760 0.25% + 700 mV 0.25% + 700 mV 0.25% + 400 mV 0.25% + 650 mV Accuracy w/opt 761 0.25% + 500 mV 0.25% + 350 mV 0.25% + 350 mV 0.25% + 650 mV
Maximum setting 22 V 38.5 V 66 V 110 V
Response time 50 µs from occurrence of over-voltage condition to start of output shutdown
Maximum Up-programming and Down-programming Time with full resistive load
(time from 10% to 90% of Voltage setting from 20 ms 20 ms 20 ms 20 ms total voltage excursion) 0 V to full scale
and full scale to 0 V
Maximum Up-programming and Down-programming Settling Time with full resistive load
(time from start of voltage Voltage setting from 100 ms 100 ms 100 ms 100 ms change until voltage settles 0 V to full scale within 0.1% of the full-scale and full scale to 0 V voltage of its final value)
Remote Sense Capability
Outputs can maintain specifications with up to a 1-volt drop per load lead.
Series and Parallel Operation
Identically rated outputs can be operated directly in parallel or can be connected for straight series operation. Auto-series and auto-parallel operation is not available.
Oscilloscope Measurement Accuracy
(@ 23°C ±5°C, Accuracy of any individual point in the trace)
Voltage 0.1% + 45 mV 0.1% + 75 mV 0.1% + 120 mV 0.1% + 160 mV
Current with current compensation on
NOTE 3
0.15% + 45 mA 0.15% + 27 mA 0.15% + 22 mA 0.15% + 12 mA
Current with current compensation off
NOTE 4
0.15% + 35 mA 0.15% + 22 mA 0.15% + 19 mA 0.15% + 9 mA
1
Based on 12-bit DAC, with DAC range adjusted by software calibration
2
Based on 12-bit ADC (11 bits plus sign), with ADC range adjusted by software calibration
3
Current compensation on means the selection labeled “Compensate current measurements during voltage transients” is checked on the Meter Properties screen
4
Current compensation off means the selection labeled “Compensate current measurements during voltage transients” is not checked on the Meter Properties screen
20
21
Agilent N6705A DC Power Analyzer Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or by type testing. All supplemental chraracteristics are typical unless otherwise noted.
Arbitrary Waveform Generator Maximum Bandwidth
Maximum bandwidth is based on a sine wave into a resistive load and applies to any output current level. In the tables below, THD means Total Harmonic Distortion, Vp-p means Volts peak-to-peak, 3 dB Max Frequency is the frequency where the output voltage drops to 3 dB below the programmed value, and 6 dB Max Frequency is the frequency where the output voltage drops to 6 dB below the programmed value.
N6751A
Frequency below
N6752A 3 dB Max THD at 3 dB which THD is
Frequency Max Frequency less than 1.5%
0.5 Vp-p 4000 Hz 12.0% 440 Hz
1 Vp-p 2200 Hz 21.0% 440 Hz
2.5 Vp-p 900 Hz 25.0% 265 Hz
5 Vp-p 500 Hz 27.0% 160 Hz
50 Vp-p 340 Hz 22.0% 25 Hz
N6750 High-Performance Autoranging DC Power Modules and N6760 Precision DC Power Modules
N6754A
Frequency below 3 dB Max THD at 3 dB which THD is Frequency Max Frequency less than 1.5%
0.6 Vp-p 3600 Hz 6.0% 2100 Hz
1.2 Vp-p 2600 Hz 10.0% 1280 Hz
3 Vp-p 1700 Hz 17.0% 800 Hz
6 Vp-p 1000 Hz 17.0% 480 Hz
60 Vp-p 340 Hz 22.0% 30 Hz
N6761A
Frequency below
N6762A 3 dB Max THD at 3 dB which THD is
Frequency Max Frequency less than 1.5%
0.5 Vp-p 4500 Hz 14.0% 450 Hz
1 Vp-p 3600 Hz 14.0% 450 Hz
2.5 Vp-p 1300 Hz 25.0% 340 Hz
5 Vp-p 600 Hz 25.0% 250 Hz
50 Vp-p 350 Hz 22.0% 30 Hz
22
Agilent N6705A DC Power Analyzer Supplemental Characteristics (Continued)
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or by type testing. All supplemental chraracteristics are typical unless otherwise noted.
Arbitrary Waveform Generator Maximum Bandwidth
Maximum bandwidth is based on a sine wave into a resistive load and applies to any output current level. In the tables below, THD means Total Harmonic Distortion, Vp-p means Volts peak-to-peak, 3 dB Max Frequency is the frequency where the output voltage drops to 3 dB below the programmed value, and 6 dB Max Frequency is the frequency where the output voltage drops to 6 dB below the programmed value.
N6732B N6742B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
0.08 Vp-p 125 Hz 1.0% 200 Hz 3.0%
0.16 Vp-p 125 Hz 1.0% 180 Hz 3.0%
0.4 Vp-p 75 Hz 6.0% 100 Hz 6.0%
0.8 Vp-p 40 Hz 8.5% 60 Hz 8.5%
8 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6731B N6741B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
0.05 Vp-p 175 Hz 1.0% 260 Hz 3.0%
0.1 Vp-p 125 Hz 1.0% 175 Hz 3.0%
0.25 Vp-p 75 Hz 6.0% 100 Hz 6.0%
0.5 Vp-p 40 Hz 9.0% 55 Hz 9.0%
5 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6733B N6743B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
0.2 Vp-p 110 Hz 1.0% 190 Hz 3.0%
0.4 Vp-p 110 Hz 1.0% 160 Hz 3.0%
1 Vp-p 72 Hz 6.0% 95 Hz 6.0%
2 Vp-p 40 Hz 8.0% 55 Hz 8.5%
20 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6730 50 W and N6740 100 W Basic DC Power Modules
23
N6735B N6745B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
0.6 Vp-p 100 Hz 1.0% 180 Hz 1.0%
1.2 Vp-p 100 Hz 1.0% 160 Hz 3.0%
3 Vp-p 70 Hz 5.5% 92 Hz 5.5%
6 Vp-p 40 Hz 8.0% 55 Hz 8.0%
60 Vp-p 20 Hz 8.0% 37 Hz 8.0%
N6734B N6744B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
0.35 Vp-p 125 Hz 1.0% 200 Hz 1.0%
0.7 Vp-p 125 Hz 1.0% 175 Hz 3.5%
1.75 Vp-p 72 Hz 6.0% 100 Hz 6.0%
3.5 Vp-p 40 Hz 8.0% 55 Hz 8.5%
35 Vp-p 20 Hz 8.0% 37 Hz 8.5%
N6736B N6746B
3 dB Max THD at 3 dB 6 dB Max THD at 6 dB Frequency Max Frequency Frequency Max Frequency
1 Vp-p 90 Hz 1.0% 160 Hz 1.5%
2 Vp-p 90 Hz 1.0% 150 Hz 3.0%
5 Vp-p 62 Hz 4.5% 85 Hz 6.0%
10 Vp-p 37 Hz 8.0% 50 Hz 8.0%
100 Vp-p 20 Hz 8.0% 35 Hz 8.0%
N6730 50 W and N6740 100 W Basic DC Power Modules (Continued)
Agilent N6705A DC Power Analyzer Supplemental Characteristics (Continued)
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or by type testing. All supplemental chraracteristics are typical unless otherwise noted.
Arbitrary Waveform Generator Maximum Bandwidth
Maximum bandwidth is based on a sine wave into a resistive load and applies to any output current level. In the tables below, THD means Total Harmonic Distortion, Vp-p means Volts peak-to-peak, 3 dB Max Frequency is the frequency where the output voltage drops to 3 dB below the programmed value, and 6 dB Max Frequency is the frequency where the output voltage drops to 6 dB below the programmed value.
24
N6773A 3 dB Max THD at 3 dB 6 dB Max THD at 6 dB
Frequency Max Frequency Frequency Max Frequency
0.2 Vp-p 125 Hz 1.5% 210 Hz 4.0%
0.4 Vp-p 125 Hz 1.5% 180 Hz 4.0%
1 Vp-p 75 Hz 6.0% 95 Hz 6.0%
2 Vp-p 42 Hz 9.0% 60 Hz 9.0%
20 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6774A 3 dB Max THD at 3 dB 6 dB Max THD at 6 dB
Frequency Max Frequency Frequency Max Frequency
0.35 Vp-p 125 Hz 1.0% 200 Hz 1.0%
0.7 Vp-p 125 Hz 1.0% 160 Hz 3.0%
1.75 Vp-p 75 Hz 6.0% 95 Hz 6.0%
3.5 Vp-p 40 Hz 8.5% 55 Hz 8.5%
35 Vp-p 20 Hz 10.0% 37 Hz 10.0%
N6775A 3 dB Max THD at 3 dB 6 dB Max THD at 6 dB
Frequency Max Frequency Frequency Max Frequency
0.6 Vp-p 120 Hz 1.0% 200 Hz 1.0%
1.2 Vp-p 120 Hz 1.0% 160 Hz 3.0%
3 Vp-p 70 Hz 5.0% 95 Hz 6.0%
6 Vp-p 40 Hz 8.5% 55 Hz 8.5%
60 Vp-p 20 Hz 10.0% 35 Hz 10.0%
N6776A 3 dB Max THD at 3 dB 6 dB Max THD at 6 dB
Frequency Max Frequency Frequency Max Frequency
1 Vp-p 75 Hz 1.0% 160 Hz 1.0%
2 Vp-p 75 Hz 1.0% 150 Hz 3.0%
5 Vp-p 55 Hz 4.0% 75 Hz 6.0%
10 Vp-p 35 Hz 8.0% 45 Hz 8.0%
100 Vp-p N/A N/A 35 Hz 8.0%
N6770 300 W Basic DC Power Modules
Agilent N6705A DC Power Analyzer Supplemental Characteristics (Continued)
Supplemental characteristics are not warranted but are descriptions of performance determined either by design or by type testing. All supplemental chraracteristics are typical unless otherwise noted.
Arbitrary Waveform Generator Maximum Bandwidth
Maximum bandwidth is based on a sine wave into a resistive load and applies to any output current level. In the tables below, THD means Total Harmonic Distortion, Vp-p means Volts peak-to-peak, 3 dB Max Frequency is the frequency where the output voltage drops to 3 dB below the programmed value, and 6 dB Max Frequency is the frequency where the output voltage drops to 6 dB below the programmed value.
25
N6705A
Maximum Total Output Power N6705A 600 W when operating from 100 – 240 VAC input
(= Sum of Total Module Output Power)
Command Processing Time From receipt of command to start 1 ms
of the output change
Protection Response Characteristics INH input 5 µs from receipt of inhibit to start of shutdown
Fault on coupled outputs < 10 µs (from receipt of fault to start of shutdown)
Digital Control Characteristics Maximum voltage ratings 16.5 VDC/- 5 VDC between pins (pin 8 is internally connected to chassis ground).
Pins 1 and 2 as FLT output Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA Typical high-level leakage current = 0.14 mA @ 16.5 VDC
Pins 1 - 7 as digital/trigger outputs Maximum low-level output voltage = 0.5 V @ 4 mA; 1 V @ 50 mA; 1.75 V @ 100 mA (pin 8 = common) Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Pins 1 - 7 as digital/trigger inputs and Maximum low-level input voltage = 0.8 V pin 3 as INH input Minimum high-level input voltage = 2 V (pin 8 = common) Typical low-level current = 2 mA @ 0 V (internal 2.2 k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Trig out BNC Maximum low-level output voltage = 0.8 V @ 1.25 mA
Minimum high-level output voltage = 4.0 V @ 1.25 mA Typical output pulse width = 8 µs
Trig in BNC Maximum low-level input voltage = 1.3 V (with internal 100 Kohm pull-up)
Minimum high-level input voltage = 2.7 V Minimum input pulse width = 2 µs
Interface Capabilities
GPIB: SCPI - 1993, IEEE 488.2 compliant interface
LXI Compliance Class C
USB 2.0 Requires Agilent IO Library version M.01.01 and up, or 14.0 and up
10/100 LAN Requires Agilent IO Library version L.01.01 and up, or 14.0 and up
Built-in Web server Requires Internet Explorer 5+ or Netscape 6.2+
Environmental Conditions
Operating environment Indoor use, installation category II (for AC input), pollution degree 2
Temperature range 0°C to 55°C (current is derated 1% per °C above 40°C ambient temperature)
Relative humidity Up to 95%
Altitude Up to 2000 meters
Storage temperature -30°C to 70°C
LED statement Any LEDs used in this product are Class 1 LEDs as per IEC 825-1
Agilent N6705A DC Power Analyzer Mainframe
N6705A
Regulatory Compliance EMC Complies with the European EMC directive for Class A test and
measurement products.
Complies with the Australian standard and carries the C-Tick mark.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Electrostatic discharges greater than 1 kV near the I/O connectors may cause the unit to reset and require operator intervention.
Safety Complies with the European Low Voltage Directive and carries the CE-marking.
This product also complies with the US and Canadian safety standards for test and measurement products.
Acoustic Noise Declaration This statement is provided to Sound Pressure Lp < 70 dB(A),
comply with the requirements At Operator position and normal operation, according to EN 27779 (Type Test). of the German Sound Emission Schalldruckpegel Lp <70 dB(A) Directive, from 18 January 1991. Am arbeitsplatz und normaler betrieb, nach EN 27779 (Typprüfung).
Output Terminal
Isolation
Maximum Rating No output terminal may be more than 240 VDC from any other terminal or chassis ground.
AC Input
Nominal Input Ratings 100 VAC – 240 VAC; 50/60 Hz/400 Hz
Input Range 86 VAC – 264 VAC
Power Consumption 1500 VA typical
Fuse Internal fuse (not customer accessible)
Weight
N6705A with 4 installed modules Net: 17.3 kg; 38 lbs.
Single-wide power module Net: 1.23 kg; 2.71 lbs
Double-wide power module Net: 2.18 kg; 4.8 lbs
Agilent N6705A DC Power Analyzer Mainframe (Continued)
26
27
Outline Diagram
16.9 mm
0.663 in
425.6 mm
16.756 in
23.5 mm
0.923 in
17.7 mm
0.697 in
177.0 mm
6.968 in
272.6 mm
10.733 in
Autoranging Characteristic
Voltage
Autoranging 50 W Output
Current
50 V
10 V
0 1 A 5 A
50 W Curve
Voltage
Precision Outputs
Current
50 V
33 V
0 1 A 1.5 A – 50 W output
2 A 3 A – 100 W output
Voltage
N6754A Autoranging 300 W Output
Current
60 V
15 V
0 5 A 20 A
300 W Curve
Voltage
Autoranging 100 W Output
Current
50 V
12 V
8.5 V
0 2 A 8.33 A 10 A
100 W Curve
28
The DC Power Analyzer System is available 2 ways:
(1) You can order an N6705A
mainframe and various mod­ules as separate products. (See steps below.) Each item will arrive in a separate box such that you can assemble the analyzer yourself when you need to.
(2) You can order an N6715A sys-
tem, which is a build-to-order DC Power Analyzer system that is shipped as a fully assembled multiple-output power supply. (See pages 30 and 31 for N6715A ordering information.)
When ordering the DC Power Analyzer as a mainframe and modules, follow these steps:
Step 1:
Select the appropriate documenta­tion and line cord options.
Step 2:
Order 1 to 4 modules (see next page). If the sum of the power of the mod­ules exceeds the available output power rating on the mainframe, see
page 12 for an explanation of
Agilent’s
power management capa­bility. Note that each mainframe has 4 module slots to hold mod­ules. Each module occupies one module slot, except for the N6754A, which occupies 2 module slots per module.
Ordering Information
Mainframe
N6705A DC Power Analyzer Mainframe
Holds 1 to 4 modules. Total available output power = 600 W.
Available options to the N6705A Mainframe
908 Rack Mount Kit
AKY Delete Front/Rear USB
This option removes all USB capability from the DC Power Analyzer. Both the front panel USB port and the rear panel USB port are removed.
055 Delete Datalogger
This option disables the Datalogger functionality in the DC Power Analyzer firmware. The DC Power Analyzer hardware is unchanged. To enable the Datalogger functionality at a later time, order the N6705U upgrade kit.
ABA Full documentation on CD-ROM and printed Users Guide
0B0 Full documentation on CD-ROM only; no printed documentation package
900 Power Cord, United Kingdom, P/N 8120-1351
901 Power Cord, Australia, New Zealand, P/N 8120-1369
902 Power Cord, Europe, Korea, P/N 8120-1689
903 Power Cord, USA, Canada, 120 V, P/N 8120-4383
904 Power Cord, USA, Canada, 240 V, P/N 8120-0698
906 Power Cord, Switzerland, P/N 8120-2104
912 Power Cord, Denmark, P/N 8120-2956
917 Power Cord, South Africa, India, P/N 8120-4211
918 Power Cord, Japan, 100 V, P/N 8120-5342
919 Power Cord, Israel, P/N 8120-6800
920 Power Cord, Argentina, P/N 8120-6869
921 Power Cord, Chile, P/N 8120-6980
922 Power Cord, China, P/N 8120-8376
927 Power Cord, Thailand, Brazil, P/N 8120-8871
29
Modules
Order 1 to 4 modules to be installed in each N6705A DC Power Analyzer mainframe. (To order modules as part of the N6715A Build-to-order Systems, see page 31). If the sum of the power of the modules exceeds the available output power rating on the mainframe, see page 12 for an explanation of Agilent’s power management capability.
You can individually specify each option for each module. For exam­ple, you can order one module with Option 761 Output Disconnect Relays while the remaining modules have no relay option.
As your needs change and you want to change configuration or add more
modules to the existing N6705A mainframe, use this ordering information to order the required modules.
Ordering Information
Modules
N6730 50 W N6731B 5 V, 10 A, 50 W DC Power Module DC Power Modules
N6732B 8 V, 6.25 A, 50 W DC Power Module
N6733B 20 V, 2.5 A, 50 W DC Power Module
N6734B 35 V, 1.5 A, 50 W DC Power Module
N6735B 60 V, 0.8 A, 50 W DC Power Module
N6736B 100 V, 0.5 A, 50 W DC Power Module
N6740 100 W DC Power Modules N6741B 5 V, 20 A, 100 W DC Power Module
N6742B 8 V, 12.5 A, 100 W DC Power Module
N6743B 20 V, 5 A, 100 W DC Power Module
N6744B 35 V, 3 A, 100 W DC Power Module
N6745B 60 V, 1.6 A, 100 W DC Power Module
N6746B 100 V, 1 A, 100 W DC Power Module
N6750 High-Performance, Autoranging DC Power N6751A 50 V, 5 A, 50 W High-Performance Modules
Autoranging DC Power Module
N6752A 50 V, 10 A, 100 W High-Performance
Autoranging DC Power Module
N6754A 60 V, 20 A, 300 W High-Performance
Autoranging DC Power Module (occupies 2 module slots in the mainframe)
N6760 Precision DC Power Modules N6761A 50 V, 1.5 A, 50 W Precision DC Power Module
N6762A 50 V, 3 A, 100 W Precision DC Power Module
N6770 300 W DC Power Modules N6773A 20 V, 15 A, 300 W DC Power Module
N6774A 35 V, 8.5 A, 300 W DC Power Module
N6775A 60 V, 5 A, 300 W DC Power Module
N6776A 100 V, 3 A, 300 W DC Power Module
Available options to DC Power Modules
N6731B-N6736B N6741B-N6746B N6751A-N6752A N6754A N6761A-N6762A N6773A-N6776A 50 W DC 100 W DC High-Performance 300 W High- Precision DC 300 W DC Power Power Modules Power Modules Autoranging DC Performance Power Modules Modules
Power Modules Autoranging DC
Power Module
Output Disconnect Relays 761 761 761 761 761 761
Output Disconnect and Polarity Reversal Relays 760 760 (see note 1, 2) Not available 760 Not available 760 (see note 2)
Large Size FPGA Installed Not available, Not available, LGA (Required, Not available, Not available, Not available,
not required not required see note 3) not required not required not required
Commercial calibration with test results data UK6 UK6 UK6 UK6 UK6 UK6
ISO 17025 Cal certificate 1A7 1A7 1A7 1A7 1A7 1A7
1
Option 760 is not available on N6741B.
2
When Option 760 is installed in N6742B or N6773A, the maximum output current is limited to 10 A.
3
You must order Option LGA on N6751A and N6752A modules.
4
High Speed Test Extensions (HSTE) is not needed on modules for use in the DC Power Analyzer.
30
N6715A Build-to-order Systems
To purchase a DC Power Analyzer system, order a N6715A. This model number is a build-to-order system that is shipped as a fully tested and assembled instrument. Each system consists of one mainframe plus optionally 1 to 4 modules. Each mainframe has 4 module slots to hold modules. Each module occupies one module slot, except for the N6754A, which occupies 2 module slots per module. To specify which modules you want installed in the system, the mod­ules are ordered as options to the system model number. You must order at least 1 module.
If the sum of the power of the modules exceeds the available output power rating on the mainframe, see page 12 for an explanation of Agilent’s power management capability.
If you prefer to purchase a DC Power Analyer mainframe and modules as separate pieces, see page 28.
Ordering Information
Build-to-order System
N6715A System Build-to-order DC Power Analyzer System
Consists of 1 N6705A mainframe with total available power of 600 W.
Available options to the N6715A System
908 Rack Mount Kit
AKY Delete Front/Rear USB
This option removes all USB capability from the DC Power Analyzer. Both the front panel USB port and the rear panel USB port are removed.
055 Delete Datalogger
This option disables the Datalogger functionality in the DC Power Analyzer firmware. The DC Power Analyzer hardware is unchanged. To enable the Datalogger functionality at a later time, order the N6705U upgrade kit.
ABA Full documentation on CD-ROM and printed Users Guide
0B0 Full documentation on CD-ROM only; no printed documentation package
900 Power Cord, United Kingdom, P/N 8120-1351
901 Power Cord, Australia, New Zealand, P/N 8120-1369
902 Power Cord, Europe, Korea, P/N 8120-1689
903 Power Cord, USA, Canada, 120 V, P/N 8120-4383
904 Power Cord, USA, Canada, 240 V, P/N 8120-0698
906 Power Cord, Switzerland, P/N 8120-2104
912 Power Cord, Denmark, P/N 8120-2956
917 Power Cord, South Africa, India, P/N 8120-4211
918 Power Cord, Japan, 100 V, P/N 8120-5342
919 Power Cord, Israel, P/N 8120-6800
920 Power Cord, Argentina, P/N 8120-6869
921 Power Cord, Chile, P/N 8120-6980
922 Power Cord, China, P/N 8120-8376
927 Power Cord, Thailand, Brazil, P/N 8120-8871
31
Modules as options to N6715A
To order 1 to 4 module as options to an N6715A, specify its model number, followed by “–ATO”. For example, to order an N6731B as an option to the N6715A, you would specify “N6731B–ATO” as the option. (To order modules as separate products, see page 29.) If the sum of the power of the modules exceeds the available output power rating on the mainframe, see page 12 for an explanation of Agilent’s power management capability.
You can individually specify each option for each module. For exam­ple, you can order one module with Option 761 Output Disconnect Relays while the remaining modules have no relay option.
Ordering Information
Module options for N6715A System
N6730 50 W DC Power Modules
N6731B–ATO 5 V, 10 A, 50 W DC Power Module
N6732B–ATO 8 V, 6.25 A, 50 W DC Power Module
N6733B–ATO 20 V, 2.5 A, 50 W DC Power Module
N6734B–ATO 35 V, 1.5 A, 50 W DC Power Module
N6735B–ATO 60 V, 0.8 A, 50 W DC Power Module
N6736B–ATO 100 V, 0.5 A, 50 W DC Power Module
N6740 100 W DC Power Modules
N6741B–ATO 5 V, 20 A, 100 W DC Power Module
N6742B–ATO 8 V, 12.5 A, 100 W DC Power Module
N6743B–ATO 20 V, 5 A, 100 W DC Power Module
N6744B–ATO 35 V, 3 A, 100 W DC Power Module
N6745B–ATO 60 V, 1.6 A, 100 W DC Power Module
N6746B–ATO 100 V, 1 A, 100 W DC Power Module
N6750 High-Performance Autoranging DC
Power Modules
N6751A–ATO 50 V, 5 A, 50 W High-Performance
Autoranging DC Power Module
N6752A–ATO 50 V, 10 A, 100 W High-Performance
Autoranging DC Power Module
N6754A-ATO 60 V, 20 A, 300 W High-Performance
Autoranging DC Power Module (occupies 2 module slots in the mainframe)
N6760 Precision DC Power Modules
N6761A–ATO 50 V, 1.5 A, 50 W Precision DC Power Module
N6762A–ATO 50 V, 3 A, 100 W Precision DC Power Module
N6770 300 W DC Power Modules
N6773A–ATO 20 V, 15 A, 300 W DC Power Module
N6774A–ATO 35 V, 8.5 A, 300 W DC Power Module
N6775A–ATO 60 V, 5 A, 300 W DC Power Module
N6776A–ATO 100 V, 3 A, 300 W DC Power Module
Available options to N6700 Modules
N6731B-ATO N6741B-ATO N6751A-ATO N6754A-ATO N6761A-ATO N6773A-ATO to N6736B-ATO to N6746B-ATO to N6752A-ATO 300 W High- to N6762A-ATO to N6776A-ATO 50 W DC Power 100 W DC Power High-Performance Performance Precision DC 300 W DC Power Modules Modules Autoranging DC Autoranging DC Power Modules Modules
Power Modules Power Module
Output Disconnect Relays 761 761 761 761 761 761
Output Disconnect and Polarity Reversal Relays 760 760 (see note 1, 2) Not available 760 Not available 760 (see note 2)
Large Size FPGA Installed Not available, Not available, LGA (Required, Not available, Not available, Not available,
not required not required see note 3) not required not required not required
Commercial calibration with test results data UK6 UK6 UK6 UK6 UK6 UK6
ISO 17025 Cal certificate 1A7 1A7 1A7 1A7 1A7 1A7
1
Option 760 is not available on N6741B-ATO.
2
When Option 760 is installed in N6742B-ATO or N6773A-ATO, the maximum output current is limited to 10 A.
3
You must order Option LGA on N6751A and N6752A modules.
4
High Speed Test Extensions (HSTE) is not needed on modules for use in the DC Power Analyzer.
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
www.agilent.com/find/agilentdirect
Quickly choose and use your test equipment solutions with confidence.
Agilent Email Updates
www.agilent.com/find/open
Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development.
www.lxistandard.org
LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium.
Remove all doubt
Our repair and calibration services will get your equipment back to you, performing like new, when prom­ised. You will get full value out of your Agilent equipment through­out its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements.
Agilent offers a wide range of ad­ditional expert test and measure­ment services for your equipment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management.
For more information on repair and calibration services, go to:
www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894-4414 Latin America 305 269 7500 United States (800) 829-4444
Asia Pacific
Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 81 426 56 7832 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Thailand 1 800 226 008
Europe
Austria 0820 87 44 11 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700 Germany 01805 24 6333*
*0.14 /minute Ireland 1890 924 204 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland (French) 41 (21) 8113811(Opt 2) Switzerland (German) 0800 80 53 53 (Opt 1) United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus
Revised: May 7, 2007
Product specifications and descriptions in this document subject to change without notice. For the latest and complete product information on the Agilent N6705, go to
http://www.agilent.com/find/N6705.
© Agilent Technologies, Inc. 2007 Printed in USA, June 15, 2007 5989-6319EN
Agilent Technologies
Need a power solution for ATE?
The Agilent N6700 Modular Power System is small, flexible, and fast
• Ideal for ATE systems in R&D, Design Validation, and Manufacturing
• Small size: up to 4 outputs in 1U of rack space
• Flexible, modular system: Can mix and match power levels and performance levels to optimize investment
• Uses same modules as N6705A DC Power Analyzer
• Fast command processing times to improve throughput
• Connect via GPIB, LAN, or USB
• Fully compliant to LXI Class C specification
Complete specifications can be found in the N6700 Modular Power System Data Sheet, publication 5958-1411EN. For more information, go to
www.agilent.com/find/N6700