The information contained in this document is subject to change
without notice.
Agilent Technologies makes no warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
The following safety symbols are used throughout this manual.
Familiarize yourself with the symbols and their meaning before
operating this equipment.
WARNINGWarn ing denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
NOTENote calls out special information for the user’s attention. It provides
operational information or additional instructions of which the user
should be aware.
Warranty
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of one year from date
of shipment. During the warranty period, Agilent Technologies will, at
its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies shall
pay shipping charges to return the product to Buyer. However, Buyer
shall pay all shipping charges, duties, and taxes for products returned
to Agilent Technologies from another country.
This user guide contains information about installation, operation,
product specifications, calibration data, maintenance and repair, and
storage conditions of the N5532A Sensor Module.
5
Page 6
General Information
Description
Description
The N5532A is intended to be used with the N5531S or N5530S
Measuring Receiver System. It is essentially a power sensor with a
power splitter at the input. See Figure 1-1 on page 7. In use, the sensor
module's two output cables are connected to a power meter (E4418/19,
E4416/17, or N1911/12 Series) and the Performance Spectrum Analyzer
(PSA) RF Input. When using an N1911 or N1912 Series power meter, it
is necessary to use an adapter (N5532A Option 019) to connect the
output cable of the sensor module to the power meter. The sensor
module is calibrated such that the power meter indicates the power
supplied by the source under test.
The N5532A Sensor Module is offered in four frequency configurations:
• Option 504 covers 100 kHz to 4.2 GHz
• Option 518 covers 10 MHz to 18 GHz
General Information
• Option 526 covers 30 MHz to 26.5 GHz
• Option 550 covers 30 MHz to 50 GHz
(with adaptor 11903D 2.4mm(f) to Type N(m))
The sensor module accurately measures power levels from -20 dBm to
+30 dBm (10 µW to 1 W). Calibration Factor (Cal Factor) information is
unique to each sensor module and is supplied on a floppy disc included
with each N5532A. Please refer to “Installing the Calibration Factors”
on page 2-12 for more information on using the Cal Factor Data.
6Chapter 1
Page 7
General Information
Description
NOTEThe N5532A Sensor Module is compatible ONLY with the newer
E4418/19, E4416/17, or N1911/12 Series power meters. They are NOT
compatible with the earlier 430 Series or 70100 power meters. If used
with an N1911/12 Series power meter, an adapter (N5532A Option 019)
is required to connect the dc cable to the power meter.
N5532A Block Diagram
Figure 1-1 N5532A Sensor Module
General Information
Chapter 17
Page 8
General Information
General Information
Instruments Covered by Manual
Instruments Covered by Manual
These instruments have a two-part serial number: the prefix (two
letters and the first four numbers), and the suffix (the last four
numbers). The two letters identify the country in which the unit was
manufactured. The four numbers of the prefix are a code identifying the
date of the last major design change incorporated in your Agilent
Technologies product. The four-digit suffix is a sequential number and
coupled with the prefix, provides a unique identification for each unit
produced. The contents of this manual apply directly to all serial
numbers unless otherwise indicated.
8Chapter 1
Page 9
2Installation
Installation
9
Page 10
Installation
Initial Inspection
Initial Inspection
Inspect the shipping container for damage. If the shipping container or
packaging material is damaged, it should be kept until the contents of
the shipment have been checked mechanically and electrically. If there
is mechanical damage notify the nearest Agilent Technologies office as
shown in Table 4-3 on page 37. Keep the damaged shipping materials (if
any) for inspection by the carrier and an Agilent Technologies
representative.
Operating Precautions
NOTEFor optimal results, the sensor module must be placed on a supportive
surface with the serial number label facing up. This orientation
provides repeatable measurements by eliminating strain on the input
connector.
Installation
The input connector should be torqued to 12 in-lb (135 Ncm) for the
type-N connector, 8 in-lb (90 Ncm) for the APC 3.5 mm connector, or 8
in-lb (90 Ncm) for the 2.4 mm connector to ensure a repeatable
connection in order to meet the published specifications.
Connect the sensor module by turning only the hex nut portion of the
connector. Damage can occur if torque is applied to the sensor module
body.
The connector plastic insulator bead deteriorates when contacted by
acetone, trichloroethylene, carbon tetrachloride, benzene, etc.
WARNINGBEFORE CONNECTING THE SENSOR MODULE TO OTHER
INSTRUMENTS, ensure that all instruments are connected to
the protective (earth) ground. Any interruption of the
protective earth grounding will cause a potential shock hazard
that could result in personal injury and cause damage to the
sensor module.
10Chapter 2
Page 11
Installation
Initial Inspection
CAUTIONTo prevent physical damage to the sensor module or device-under-test
(DUT), pay careful attention to the mechanical setup. The mechanical
design of the sensor module minimizes SWR. However, the weight and
length of the module, combined with the stiffness of the interconnect
cables, enables the exertion of considerable leverage at the module’s
input connector. To prevent physical damage to the sensor module, the
DUT, and the mating connectors, and to assure best electrical
performance, observe the following precautions:
1. Do not permit the interconnect cables of the sensor module to extend
out where anyone passing by could accidentally push and exert
leverage on the cables. This consideration is especially important
when the DUT is a light-weight instrument that sits freely on a
table.
2. When possible, lay the sensor module on a supportive surface with
the serial number label facing up. This consideration is especially
important when rigid RF adapters are used to directly interconnect
the sensor module to the DUT. RF adapters lengthen the leverage
arm of the sensor module and are often fragile. When level-accuracy
requirements permit, use flexible cables to interconnect the sensor
module to the DUT.
3. Do not bend or coil the interconnect cables (more than necessary) to
a diameter circumscribing less than 150 mm (6 in.). This precaution
often applies when the instruments are rack-mounted and
interconnected to rear-panel connectors. Repeated flexing of coiled,
interconnect cables can degrade SWR and increase RFI of the RF
output cable.
4. Do not attempt to tighten the type-N connector by twisting the body
of the sensor module.
Installation
Chapter 211
Page 12
Installation
Installing the Calibration Factors
Installing the Calibration Factors
Cal factors can be entered and stored in an internal table (Automatic
Cal Factors) in the PSA’s non-volatile memory. The table is
automatically accessed during RF power measurements (after the
measuring receiver has tuned to the frequency) to correct for the
non-linear response of the power sensor. If no entry in the table directly
corresponds to the frequency being measured, the measuring receiver
derives a cal factor through linear interpolation using the nearest
calibration points in the table.
Installing the Cal Factors in the N5530S
There are two methods of installing the sensor module calibration
factors (cal factors) in the N5530S Measuring Receiver System.
The first method loads and saves the cal factors using the PC. The
second method uses the PSA to load the data.
Installation
Figure 2-1 Utilities Menu
Both methods are accessed through the Utilities drop-down menu (See
Figure 2-1, “Utilities Menu.”.)
Using the PC to Load and Save the Cal Factors
1. Insert the N5532A Sensor Module Data Disk in the PC A:\ drive.
NOTEIf your PC does not have a floppy disc drive, use the PSA to save your
cal data to your PC. Then use this procedure to load the sensor cal
factors into the system.
2. Select Calibration Factors to bring up the Calibration View Edit dialog box.
3. Press Load Cal File in the Calibration View Edit dialog box. The Open Calibration File dialog box appears.
4. Navigate to the PC A:\ drive if necessary and select the CFData.XML
file.
12Chapter 2
Page 13
Installation
Installing the Calibration Factors
NOTEIf you obtained the cal data from the PSA, go to the directory you
specified in Step 2 of the next section to load the data.
5. Press Open.
6. Press OK to apply the Cal Factors to the system.
Using the PSA to Save the Cal Factors to the PC
1. Insert the N5532A Sensor Module Data Disk into the PSA A:\ drive.
2. Select Copy Files from SA under the Utilities menu option of the
N5531S or N5530S User Interface. Browse to the location you wish
to place the SA floppy disc files.
3. Click OK to copy the calibration factor files to the specified location.
Three files appear: CALTABLE.XSL (enables viewing with web
browser), CFDATA.XML (actual Cal Factor Data), and CFCERT.DOC
(Cal Factor Certification)
1
.
4. To load this data into the system, follow steps 2 through 6 in the
section entitled, “Using the PC to Load and Save the Cal Factors” on
page 2-12.
Installing the Cal Factors in the N5531S, N5530SE26,
and N5530SE50
1. Insert the floppy disk in to PSA floppy disk drive. Make sure the disk
is for the sensor module you are using.
2. Press
Calibration Factor as the file type. Press Dir Up, Up Arrow, Dir Select to
ensure the “A” drive is selected. Then press
factor file (CFDATA.XML) to the PSA spectrum analyzer.
MODE, Measuring Receiver, File, Load, Type, and select
Load Now to load the cal
Installation
1. . Provides traceability to National Institute of Standards and Technology or
NIST
Chapter 213
Page 14
Installation
Installing the Calibration Factors
Installation
14Chapter 2
Page 15
3Specifications
The specifications listed in Table 3-1, “Specifications,” are the
performance standards or limits against which the sensor module may
be tested. These specifications are valid ONLY after proper calibration
of the power meter.
Specifications
15
Page 16
Specifications
N5532A Sensor Module Specifications
N5532A Sensor Module Specifications
Table 3-1 Specifications
DescriptionSpecificationComments
Frequency Range
Option 504
Option 518
Option 526
Option 550
Power Range+30 dBm (1 watt) to
100kHzto4.2GHz
10 MHz to 18 GHz
30 MHz to 26.5 GHz
30 MHz to 50 GHz
−20 dBm (10 µW)
Maximum Safe Power
Sensor Module
Linearity
Opt. 504
+30 to +20 dBm
< +20 dBm
Sensor Module
Linearity
Opt. 518
+30 to +20 dBm
< +20 dBm
Sensor Module
Linearity
Opt. 526
+30 to +20 dBm
< +20 dBm
Sensor Module
Linearity
Opt. 550
+30 to +20 dBm
< +20 dBm
+30 dBm
±3%
negligible
±3%
negligible
±3%
negligible
±3%
negligible
Averag e Total
Power
Input SWR
Opt. 504
100kHz to 2GHz
2GHz to 4.2GHz
Specifications
16Chapter 3
a
< 1.10:1
< 1.28:1
Page 17
Table 3-1 Specifications
DescriptionSpecificationComments
Input SWRa
Opt. 518
Specifications
N5532A Sensor Module Specifications
10 MHz to 2 GHz
2 GHz to 18 GHz
Input SWR
a
< 1.10:1
< 1.28:1
Opt. 526
30 MHz to 2 GHz
2 GHz to 18 GHz
18 GHz to 26.5 GHz
Input SWR
a
< 1.10:1
< 1.28:1
< 1.40:1
Opt. 550
30 MHz to 2 GHz
2 GHz to 18 GHz
18 GHz to 26.5 GHz
26.5 GHz to 33 GHz
33 GHz to 40 GHz
40 GHz to 50 GHz
a. From input connector to end of plastic casing.
b. From input connector to end of cable.
Table 3-3 Maximum Cal Factor Uncertainty
Maximum Cal Factor Uncertainty
Frequency Range
100 kHz to ≤ 30 MHz± 1%
30 MHz to ≤ 2 GHz± 1%± 1.5± 1.5± 1.5
Opt 504Opt 518Opt 526Opt 550
Comments
2 GHz to ≤ 4.2 GHz± 1%± 2.0%± 2.0%± 2.0%
2 GHz to ≤ 10 GHz± 2.0%± 2.0%± 2.0%
10 GHz to ≤ 18 GHz± 2.5%± 2.0%± 2.0%
18 GHz to ≤ 26.5 GHz± 2.5%± 2.5%
26.5 GHz to ≤ 50 GHz± 4.2%
Specifications
Chapter 319
Page 20
Specifications
N5532A Sensor Module Supplemental Information
Specifications
20Chapter 3
Page 21
Maintenance and Repair
4Maintenance and Repair
21
Page 22
Maintenance and Repair
Maintenance
Maintenance
Cleaning Solutions
Maintenance and Repair
A solution of pure isopropyl or ethyl alcohol can be used to clean the
connector (keeping in mind its flammable nature).
Connector Cleaning
Clean the connector face using a cotton swab dipped in isopropyl
alcohol. If the swab is too big, use a round wooden toothpick wrapped in
a lint free cotton cloth dipped in isopropyl alcohol.
22Chapter 4
Page 23
Repair
Power sensor replacement
If the power sensor requires replacement, 2 screws must be removed
from the RF input connector flange of the replacement assembly.
Prepare the replacement power sensor by removing the 2 screws using
a 1.5-mm hex driver. Refer to Figure 4-1 for location of screws. Note
that the screws you need to remove are on the opposite end and
opposite side of the notch in the power sensor cable connector.
Figure 4-1 Location of Screws to be Removed
Maintenance and Repair
Repair
Maintenance and Repair
Chapter 423
Page 24
Maintenance and Repair
Repair
Disassembly Hints for Options 504, 518, and 526
Figure 4-2 Options 504, 518, and 526; Top View
Maintenance and Repair
Figure 4-3 Options 504, 518, and 526; Bottom View
Hints for disassembling Option 504, 518, or 526. Refer to Figure
4-2 and Figure 4-3:
NOTEOption 526 will appear slightly different than shown in Figure 4-2 and
Figure 4-3, due to the different connectors (APC 3.5 mm instead of
Type N ).
• The bulkhead connector (A) must be removed before the power
sensor to 3 dB attenuator cable (B, H) can be removed.
• To remove the power sensor to 3 dB attenuator cable (B, H), hold
24Chapter 4
Page 25
Maintenance and Repair
Repair
connector (C) with 3/4-inch open-end wrench while loosening nut (D)
and power sensor connector (E).
• To remove power splitter (J), loosen connectors but do not remove
the 2 cables or the 3 dB attenuator until the hardware that fastens
the power splitter to the main deck (F) has been removed. This will
keep from damaging the center conductors.
Re-assembly Hints for Options 504, 518, and 526
Hints for assembling Option 504, 518, or 526 after repair. Refer to
Figure 4-2 and Figure 4-3.
• Place the power sensor to 3 dB attenuator cable (B, H) through the
rectangular hole in the main deck (F) before installing the bulkhead
connector (A). Place lock washer and nut on bulkhead connector and
torque to 25 inch-pounds.
• When fastening the power sensor (G) to the main deck (F), torque
the 2 screws holding it to 3 inch-pounds.
Maintenance and Repair
• Connect the power sensor connector (E) to the cable (B) and torque to
8 inch-pounds while holding connector (C) with a 3/4-inch open-end
wrench. Then, while still holding connector (C), tighten the nut (D)
against the power sensor connector (E).
• When installing the power splitter (J), connect the 3 dB attenuator
(I) and the 2 cables before fastening the power splitter to the main
deck (F).
• All coax connectors should be torqued to 8 inch-pounds.
Disassembly Hints for Option 550
Figure 4-4 Option 550; Top View
Chapter 425
Page 26
Maintenance and Repair
Repair
Figure 4-5 Option 550; Bottom View
Maintenance and Repair
Hints for disassembling Option 550. Refer to Figure 4-4 and Figure
4-5.
• When removing the power sensor to 3 dB attenuator cable (B), hold
connector (C) with a 1/4-inch open-end wrench while loosening the
power sensor connector (D) using a 5/16-inch open-end wrench.
• When removing the power splitter (J), loosen the 3 connectors but do
not disconnect the cables until the hardware that fastens the power
splitter to the main deck (G) has been removed. This will keep from
damaging the center conductors.
• When removing the 3 dB attenuator (E), hold connector (F) with a
1/4-inch open-end wrench while loosening the attenuator’s connector
using a 5/16-inch open-end wrench.
Re-assembly Hints for Option 550
Hints for assembling Option 550 after repair. Refer to Figure 4-4
and Figure 4-5.
• To install the bulkhead connector (A), place lock washer and nut (H)
on bulkhead connector and torque to 25 inch-pounds.
• When fastening the power sensor to the main deck (G), torque the 2
screws holding it to 3 inch-pounds.
• Connect the power sensor connector (D) to the cable (B) and torque
to 8 inch-pounds while holding connector (C) with a 1/4-inch
open-end wrench.
• When connecting the 3 dB attenuator (E) to the 3 dB attenuator to
power splitter cable, torque to 8 inch-pounds while holding connector
(F) with a 1/4-inch open-end wrench.
26Chapter 4
Page 27
Maintenance and Repair
Repair
• When installing the power splitter (J), connect the 3 cables before
fastening the power splitter to the main deck (G). Hold the nut with
a 3/16-inch open-end wrench while tightening the corresponding
screw (2 each) to secure the power splitter to the main deck.
• All coax connectors should be torqued to 8 inch-pounds.
Maintenance and Repair
Chapter 427
Page 28
Maintenance and Repair
Repair
Replaceable Parts
Table 4-1 N5532A Series Sensor Module Replaceable Parts
Maintenance and Repair
DescriptionPart
Number
Power sensor
adapter (same as
N5532A Opt 019)
Screw, metric,
3.0 x 0.50
Front screws for
cover attachment
Screw, socket head
cap, metric 1.6 x
0.35, 10 mm long
For sensor mounting
Washer, lock, helical,
#0, 0.062-in-ID,
0.1-in-OD
For sensor mounting
Screw, tapping, 4-20,
0.835 inch long, pan
head, TORX
For support block
mounting and rear
screws for cover
attachment
N5532-60013XXXX
0515-0667XXXX
0515-1185XXXX
2190-0572XXXX
0624-0824XXXX
Opt 504
(4 GHz)
Opt 518
(18 GHz)
Opt 526
(26 GHz)
Opt 550
(50 GHz)
Screw, 4-40, 0.562
inch long, pan head,
Pozidrive
For power splitter
mounting
Washer, lock, #4,
helical, 0.115-in-ID
For power splitter
mounting
Washer, lock, 3/8
inch, 0.384-in-ID
For mounting
2.4 mm input
connector
28Chapter 4
2200-0091XXX
2190-0003XXX
2190-0022X
Page 29
Maintenance and Repair
Repair
Table 4-1 N5532A Series Sensor Module Replaceable Parts
Maintenance and Repair
DescriptionPart
Number
Hex nut, dbl
chamfer, M10 x 1.0,
6 mm thick
For mounting
2.4 mm input
connector
Nut, 4-40,
0.063-in-thick
For power splitter
mounting
Washer, lock, 7/16
inch, 0.439-in-ID
For Type-N
connector
Hex nut,
UNEF-2B-THD,
0.094-in-thick
For Type N
connector
0535-0992X
2260-0002XXX
2190-0104XX
2950-0132XX
Opt 504
(4 GHz)
Opt 518
(18 GHz)
Opt 526
(26 GHz)
Opt 550
(50 GHz)
Washer, lock, helical,
#2, 0.088-in-ID
For cable bracket
Screw, socket head
cap, 2-56,
0.25-in-long
For cable bracket
Setscrew, 4-40,
0.25-in-long
For sensor cable
bracket (see Figure
4-5)
Setscrew, 6-32,
0.125-in-long
For input connector
Hex nut, 2-56,
0.062-in-thick
For power splitter
mounting
Deck, mainN5532-00001XXX
Deck, main (526) N5532-20001X
2190-0112XXXX
3030-0053XXXX
3030-0824XXXX
3030-0022X
0610-0002X
Chapter 429
Page 30
Maintenance and Repair
Repair
Table 4-1 N5532A Series Sensor Module Replaceable Parts
DescriptionPart
Number
Maintenance and Repair
Deck support block
(right)
Deck support block
(left)
Spacer
For splitter
Bracket, cable (top)
(see Figure 4-4)
Bracket, cable
(bottom) (see Figure
4-5)
Cover, plastic (top)11722-40001XXXX
Cover, plastic
(bottom)
Cable clip
For sensor and RF
output cables
N5532-20021XXXX
N5532-20022XXXX
N5532-20003X
11722-20006XXXX
11722-20007XXXX
11722-40002XXXX
11722-40003XXXX
Opt 504
(4 GHz)
Opt 518
(18 GHz)
Opt 526
(26 GHz)
Opt 550
(50 GHz)
Label, option ID 5185-1867XXXX
Label, product IDN5532-80004XXXX
3 dB attenuator,
26 GHz
3 dB attenuator,
50 GHz
Input connector,
Type-N
Input connector,
3.5 mm
Input connector,
2.4 mm
Power splitter,
26.5 GHz
Power splitter,
50 GHz
Power sensor,
4.2 GHz
08493-60013XXX
08490-60010X
5064-3806XX
08673-60402X
N5532-20015X
5086-7408XXX
N5532-60002X
N5532-60011X
30Chapter 4
Page 31
Maintenance and Repair
Repair
Table 4-1 N5532A Series Sensor Module Replaceable Parts
Maintenance and Repair
DescriptionPart
Number
Power sensor,
18 GHz
Power sensor,
26.5 GHz
Power sensor,
50 GHz
Cable, input
connector to splitter
Cable, input
connector to splitter
Cable, RF input,
2.4 mm
Cable, splitter to
output
Cable, splitter to
output
Cable, splitter to
attenuator
N5532-60012X
N5532-60004X
N5532-60005X
N5532-20004XX
N5532-20007X
N5532-20009X
N5532-20018XXX
N5532-20010X
N5532-20011X
Opt 504
(4 GHz)
Opt 518
(18 GHz)
Opt 526
(26 GHz)
Opt 550
(50 GHz)
Cable, attenuator to
sensor
Cable, attenuator to
sensor
Cable, attenuator to
sensor
Cable, sensor
output, shielded
Cable, RF output,
3.5 mm, 66 inches
long
Cable, RF output,
2.4 mm, 66 inches
long
N5532-20012X
N5532-20006X
N5532-20019XX
N5532-60010XXXX
N5532-20013XXX
N5532-20017X
Chapter 431
Page 32
Maintenance and Repair
Calibration
Calibration
Agilent Technologies recommends that repair and calibration of the
Maintenance and Repair
N5532A Sensor modules be performed at an Agilent Technologies
service center. The service center has the required calibration
standards and the ability to produce a calibration data disk in the
format that the N5531S or N5530S Measuring Receiver System
requires.
However, if your cal lab currently calibrates Agilent thermocouple
power sensors such as the 8481A, 8482A, or 8487A, and the calibration
standards and test metrology used meet the N5532A specifications
listed in Table 3-1 on page 16, you may wish to develop your own
calibration process. Refer to Agilent Technologies Application Notes
1449-1 through 1449-4 for information regarding the fundamentals of
RF and microwave power measurements. Application Note 1449-1
includes an overview of the theory and practice of sensor calibration.
Figure 1-1 on page 7 shows that the N5532A Sensor Modules consist
basically of a power splitter and an Agilent 848xA Power Sensor with a
padded input. The module is calibrated as a standard power sensor
with the RF OUTPUT TO PSA line terminated with a matched 50 Ω
load.
The calibration power at the input should be 0 dBm ±2 dB. This is to
ensure enough power is provided so the sensing element in the N5532A
Sensor Module stays above its noise floor. The maximum calibration
must not exceed +10 dBm to avoid non-linearities in the calibration
standards (reference sensor or standard sensor).
The presence of the splitter and cable create a ripple in the sensor’s
input efficiency, so many frequency calibration points are required to
accurately characterize the sensor module.
32Chapter 4
Page 33
Maintenance and Repair
Calibration
Figure 4-6 Cal Factor (Ripple Due to Splitter and PSA Line Loading)
Maintenance and Repair
Chapter 433
Page 34
Maintenance and Repair
Calibration
Table 4-2 Recommended Calibration Points
N5532A Power SensorFrequency Points
Option 504100 kHz
Maintenance and Repair
Option 51810 MHz
300 kHz
1 MHz
3 MHz
5 MHz
10 MHz
30 MHz
50 MHz
100 MHz to 4.2 GHz
in 100 MHz steps
30 MHz
50 MHz
100 MHz to 18 GHz
in 100 MHz steps
Option 52630 MHz
50 MHz
100 MHz to 26.5 GHz
in 100 MHz steps
Option 55030 MHz
50 MHz
100 MHz to 50 GHz
in 100 MHz steps
Typical Cal Factor ranges for the 848xA Power Sensors are between
90% and 100%, depending on the efficiency of the sensor’s thermocouple
over the specified frequency range. With the added loss from the other
passive components in front of the power sensor, there is approximately
another 9 to 11 dB loss (or another 25% loss) from 50 MHz to 26.5 GHz.
This, coupled with the fact that typical power sensor Cal Factor is
defined with a constant power level at the input, and this application
has a sloped power input, an additional 25% efficiency component is
added. As a result, the N5532A Sensor Module will have about a 40%
Cal Factor at 26.5 GHz. This is acceptable because the sensor is
characterized with this behavior and thus will be corrected in the
N5531S/30S Measuring Receiver System.
34Chapter 4
Page 35
Maintenance and Repair
Calibration
Upon calibrating the sensor module, the cal factor data is stored on a
floppy disk. This disk allows you to transfer the cal data to the N5531S
or N5530S Measuring Receiver System. Since there are hundreds of
data points, manual entry is highly impractical.
Cal Factor Data File Format
The N5532A data, “CFData.xml”, is stored in an xml format. On each data
disk, the data file is accompanied with an xsl file, “CalTable.xsl”.
This file contains the xml formatting instructions which are used by an
xml viewer, such as Microsoft Internet Explorer. However, the xml file
may be viewed by any text editor and will appear as shown below.
The system software does not need the xsl file because it’s not trying
to display the data. The system, however, does look for the specific
“bookmarks” or “tags” used by the xml format. So, for the system to
properly read the data from the file, the calibration data needs to be
stored as shown in the following sample file.
Figure 4-7 File Format for the N5532A Sensor Module Data Disk (Two Data
Points Shown)
Maintenance and Repair
<?xml version="1.0" encoding="utf-8"?>
<!--This file is used with the Agilent N5532A power sensor module.-->
<?xml-stylesheet type="text/xsl" href="CalTable.xsl"?>
<Calibration>
• The fourth line marks the start of the calibration data, while the last
line marks the end of the calibration data.
• The next 5 lines contain sensor-specific information: data format
version, model number, option, serial number, and date calibrated.
Chapter 435
Page 36
Maintenance and Repair
Calibration
• The next line is the reference calibration factor.
• The rest of the data file is filled with frequency-dependent data.
There are 6 lines for every data point and each data point is marked
by the beginning <CalFactor> and the ending </CalFactor>. The
Maintenance and Repair
data set includes frequency, cal factor, cal factor uncertainty,
reflection magnitude, reflection phase, and reflection magnitude
uncertainty.
36Chapter 4
Page 37
Contacting Agilent Technologies
Agilent Technologies has offices around the world to provide you with
complete support for your accessories. To obtain servicing information
or to order replacements, contact the nearest Agilent Technologies office
listed in Table 4-3 on page 37. In any correspondence or telephone
conversations, refer to your accessory by its product number and full
serial number.
Table 4-3 Contacting Agilent Technologies
Online assistance: www.agilent.com/find/assist
Maintenance and Repair
Contacting Agilent Technologies
Maintenance and Repair
United States
(tel) 1 800 452 4844
Canada
(tel) 1 877 894 4414
(fax) (905) 282-6495
Europe
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
New Zealand
(tel) 0 800 738 378
(fax) 64 4 495 8950
Latin America
(tel) (305) 269 7500
(fax) (305) 269 7599
Japan
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Australia
(tel) 1 800 629 485
(fax) (61 3) 9210 5947
Asia Pacific
(tel) (852) 3197 7777
(fax) (852) 2506 9284
Chapter 437
Page 38
Maintenance and Repair
Contacting Agilent Technologies
Maintenance and Repair
38Chapter 4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.