Agilent M1165/66/67/75/76/77A
Component Monitoring System
and
Agilent M1205A V24 & V26
User’s Reference Manual
Volume 2
Parameter Information
s1
Part Number M1046-900 1L
Printed 06/2000
First Edition
Notice
This document contains proprietary information which is protected by copyright. All Rights
Reserved. Reproduction, adaptation, or translation without prior w rit ten permission is
prohibited, except as a llowed under the copyright laws.
Agilent Technologies
3000 Minuteman Road
Andover, MA 01810-1085
(978) 687-1501
Publication number
M1046-9001L
Warranty
The information contained in this document is subject to change without n otice.
Agilent Technologies ma kes no warranty of any kind with regard to this material, including,
but not limited to, the implied warranties or merchantability and fitness for a particular
purpose.
Agilent Technologies shall no t be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance, or use of this
material.
New editions of this document will incorporate all material updated since the previous edition.
Update packages may be issued bet ween editions and contain replac ement and additional
pages to be merged by a re vision date at the bottom of the pa ge. Note that pages which are
rearranged due to ch anges on a previous page are not considered revised.
The documentation printing date and part number indicate its current edition. The printing
date changes when a new edition is printed. (Minor corrections and updates which are
incorporated at reprint do not cause the date to change.) The document part number changes
when extensive technical ch anges are incorporated.
United States federal law restricts these devices to sale by or on the order of a physician.
The M1165/66/75/76A Systems comply with UL544, CSA 22.2-125, IEC 601-1, EN 60601-1, and
EN 60601-1-2 and car r ies Marking to Counci l Dir ec ti ve 93/ 4 2/ EEC, E ur opean Medical
Device Directive (MDD).
The M1167/77A Systems comply with UL2601-1, CSA 22.2 No. 601.1-M90, IEC 601-1,
EN 60601-1, and EN 60601-1-2 and carries Marking to Council Directive 93/42/EEC,
0366
European Medical Device Directive (MDD).
The M1205A Systems comply with UL2601, IEC 601-1, CSA C22.2 no. 601-1, EN60601-1, and
EN60601-1-2 and carries Marking to Council Directive 93/42/EEC, European Medical
Device Directive (MDD).
iii
Electromagnetic Interference
Anomalies due to electromagnetic interference are not unique to the M1165/66/67/75/76/77A
or the M1205A but are chara cteristic of patient monitors in use today. This performance is
due to the very sensitive high gain front end amplifiers used to display the physiological
signals. Among the many similarly performing monitors already in use by customers,
interference from electromagnetic sources is rarely a problem in actual use.
Avoiding Electromagnetic Interference
When electromagnetic interference (EMI) is encountered there are a number of actions that
can be taken to mitigate the problem.
•Eliminate the source. Possible sources of EMI can be turned off or moved away to
reduce their strength.
•Attenuate the coupling. If the coupling path is through the patient leads, the
interference may be reduced by moving and/or rearranging the leads. If the coupling is
through the power cord, plugging the monitor into a dif ferent circuit may help.
•Reduce the sensitivity of th e syst em . In all o f th e EM C t esti n g the m oni to r wa s adju ste d
to maximum sensitivity. For the ECG amplifier the gain was four times what is
normally required. By reducing the gain of the system receiving the EMI, the
interference can often be eliminated.
•Add external attenuat ors. I f E MI becomes an unusually difficult problem external
devices such as an isolation transf ormer or a transient suppressor ma y be of help. An
Agilent Customer Engineer can be of help in determining the need for external devices.
iv
Intended Use
Intended Use
Description
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are ne tw ork connectable bedside patient monitoring devices.
The Agilent M1205A Models V24CT and V26 CT may powered by either AC line power or by
battery power.
Purpose
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors measure and display multiple ph ysiological parameters and w aves, and
generate alarms and recordings. They exchange information with compatible devices. The
Agilent M1165/66/67 /75/76/77A Component Monitoring System and the Agile nt M1205A V24
and V26 monitors are not therapeutic devices.
Patient Population
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors are intended to be used on adult, pediatric, an d neonatal patients.
Environment
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors are intended to be used in a clinical environment by trained healthcare
professionals. They are not intended for home use.
They communicate with de vices such as a central station through network interface ports and
a serial I/O port.
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are pres cription devices and will carry the following label, “United
States Federal law restricts this device to sale by or on the order of a physic ian.”
v
Indications for Use
Indications for Use
Condition
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are generally indicated when the clinician decides there is a need to
measure and display multiple physiological parameters and waves, to generate alarms and
recordings of adult, pediatric, or neonatal patie nts.
Part of Body or Type of Tissue with Which the Device Interacts
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors do no t c ontact the body or tissue of the patient. Signals are obtain ed
from accessory electrode, transducer, and sensor devices.
Frequency of Use
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 monitors ar e indicated for use when presc ribed by a clinician.
Physiological Purpose
The Agilent M1165/66/6 7/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are indicated when the purpose is to gain information for treatment, to
assess adequacy of treatment, or to rule out causes of symptoms. The Agilent M1165/66/67/
75/76/77A Compon e nt Mo n it or i n g Syste m an d the Agi le n t M120 5A V24 an d V26 mo n itors are
well suited for patient m onitoring.
Patient Population
Adult, pediatric, and neonatal non-ambulatory patients.
vi
Indications for Use
Prescription Versus Over-the-Counter
The Agilent M1165/66/67/75/76/77A Component Monitoring System and the Agilent M1205A
V24 and V26 Monitors are prescription devices.
vii
Indications for Use
Warnings, Cautions, and Notes
Warnings, cautions, and notes are us ed throughout this User’s Manual to give you additional
information about the Agilent M1165/66/67/75/76/77A Component Monitoring System and the
Agilent M1205A V24 and V26 monitors. The warnings and cautions included in this safety
section refer to the equipment in general.
WarningWarning
A “warning” calls attention to the user of imminent hazard to people if proper
procedures are not followed.
•For continued safe use of this equipment, it is necessary that the listed instructions are
followed. Instructions in this manual in no way supersede established medical
procedures.
•Explosion Hazard- Do n ot u se th i s equipment in the pre s en ce o f fl am ma bl e an esth etics.
•Alarms - Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monit oring may result
in patient jeopardy. Remember that the most relia ble method of patient monitoring
combines close persona l surveillance with correct operation of m onitoring equipment.
•This equipment is only intende d fo r use in healthcare facilities by trained healthcare
professionals.
•The product is not intended for outside hospital use such as a helicopt ers or
ambulances.
•This product is not intended for home use.
•To reduce the risk of elect ric al shock, do NOT remove any cover. Refer servicing to
qualified personnel.
•This equipment may interfere with ultrasound imaging equipment by causi ng
interference on the ultrasound display. Try to keep the instruments as far apart as
possible.
viii
Indications for Use
•Exposure of electrical contacts or connections to saline or other liquids and gels is
dangerous. Electrical contacts and connections such as cable connectors, power
supplies, parameter module plu g-in connections and rack connec tions must be kept
clean and dry. Thoroughly dry any electrical connections that become contaminated
with liquids. If additional decontamination is required please contact your biomedical
department or Agilent Tech nologies Response Center.
•Although this equipment is shielded against Electromagnetic Interference (EMI), it is
recommended to avoid the use of electrically radiating devices in close proximity to this
equipment.
•Connecting the Agilent monitoring network (SDN) cable when the product is powered
on is not supported. Error c odes and Agilent monitoring network (SDN) interface lockup may occur. Power cycling the product will recover the product. No permanent
damage will result. To prevent unintentional disruption in monitoring, be sure th e SDN
interface cable is properly secured at both ends when c onnecting to the Agilent
monitoring network (SDN).
•Do not connect a second rack by a cable when using a module rack docked to the back
of the Agilent V24CT or V26CT. Using a second rack connected by a cable may disrupt
module communication.
Caution
A “caution” calls attention to a c ondition or possible situation that co uld cause injury to the
user.
•Ventilation Requirements - Failure to meet ventilation requirements may cause
equipment failure an d, in turn, jeopardize the functions of automated monitoring. Do not
locate equipment in an enclosed ar ea which could restrict heat dissipation.
•Maintenance - Failure on the part of the responsible individual, hospital, or institution
employing the use of this equipment to implement a satisfactory maintenance schedule
may cause undue equipment failure and possible healt h hazards.
•Do not spray cleaning solut ions directly onto the monitor. Moisture droplets may enter
the internal components and cause equipment mal function or failure. Cleaning solutions
should be applied to a cloth and the cloth used to wipe the monitor clea n. The monitor
should be turned off during cleaning.
ix
Indications for Use
•Replacement Parts - It is highly recommended that only Agilent Technologies
recommended parts and a ccessories be used with this equipment. Failure to do so may
result in the degradation of performance. Accessorie s and parts for individual modules
and components are listed at the back of the appropriate section in th is manual.
Note—
A note gives special instructions to highlight an operating procedure or practice. Notes
may precede or follow the applicable text.
At this time, Agilent Technologies will ma ke available on request, and in E nglish only, such
circuit diagrams, component part lists, descriptions, calibration instructions, or other
information which wil l assist the user’s appropriate qualified technical pers onnel to repair
those parts of the equipment which are classified by Agilent Technologies to be repairable. A
list of Agilent Sales and Support Of fices is provided at the end of this manual.
x
Indications for Use
Using This Manual
To enable you to find information easily, there is a contents list at the front of the guide and a
comprehensive index at the back.
The User’s Reference Manual is separated into two parts; the core document and the
parameter module guides.
The Core Document
This section of the guide contains al l the general information about the sy stem. It is a good
place for new users to start because it gives an introd uction to the system and the way it
works, and shows you how to get start ed. Here is a list of the major sections:
•Introducing the Agilent M1165/66/67/75/76/77A Component Monitoring System and the
Agilent M1205A V24 and V2 6 monitors
•Getting Started
•Configuring the System
•Other Patients
•Alarm Functions
•Recording Functions
•Trends and Data Management
•Installation and Patient Safety
•Care and Cleaning
xi
Indications for Use
Monitor Setup
Parameter Module Sections
These sections each contain information for one parameter module. This covers setting-up,
monitoring, and problem solving if you encounter difficulty. Each section is separ at ed with a
white tab which has the title of the section.
Note—
The User’s Reference Manual contains information for all the parameter modules
available for the system. This means, of course, that depending on the model and number of
modules you have ordered, the screens w ill not always apply to your system. However, the
information for the param eters and functions is valid for all the system s.
Note—
The screenshots displayed in this manual were generated in demo mode and may
therefore differ from what actually appears on your screen during patient monitoring.
Notice to the User
Although there may be products in your area that look similar to the Agilent M1165/66/67/75/
76/77A Component Monitoring System and the Agilent M1205A V24 and V26 monitors, their
functionality may not be the same. This User’s Reference Manual is intende d t o be used with
the Agilent M1165 / 66/67/75/76/77A Component Monitoring System, the M1026A Anest hetic
Gas Module and the Agilent M1205A V24 and V26 monitors only.
This Manual is only applicable for Release C.0 versions of the monitors listed above. A
Release C.0 monitor can be identified by:
xii
a. the Release C.0 label on the monitor, or
b. the suffix of the EPROMpack part number. To view this number, press
→ → .
Monitor RevisionShow SW Rev
The suffix of the EPROMpack part number on a Release C.0 Agilent CMS is A.
The Software R evision of a Releas e C.0 Agilent V24 or V26 monitor is L.xx.xx
Responsibility of the Manufacturer
Responsibility of the Manufacturer
Agilent Technologies only considers itself responsible for any effects on safety, reliability and
performance of the equipment if:
assembly operations, extensions, re-adjustments, modifications or repairs are carried out
by persons authorized by Agilent , and
the electrical installation of t he relevant room complies with nationa l standards, and
the instrument is used in accordance with the instructions for use.
To ensure optimum usage, we recommend that Agilent parts and accessories are used in
conjunction with the Agilent M1165/66/67/75/76/77A Co mponent Monitoring System, the
Agilent M1026A Anesthetic Gas Module and the Agilent M120 5A V24 and V26 Monitors,
wherever available. If non-Agilent parts are used, Agilent Technologies is not liable for any
damage that these parts may cause to the Agilent equipment.
Manufacturer´s Address
For South America, North America and Canada:
Agilent Technologies, Inc.
3000 Minuteman Road
Andover
MA 01810-1099
E. Supported Device Information E-1
F. Main Sales and Support Offices F-1
Contents-14
14
ECG and ECG/Respiration Module
Section
This chapter describes the ECG module and the ECG/RESP module. You
will find information on ECG and respiratory monitoring, troubleshooting,
and ST Segment monitoring. It includes the following sections:
Monitoring the ECG produc es a continuous waveform of the patient’s
cardiac electrical activity to enable an accurate assessment of his
current physical state.
•The system displays the average Heart Rate (HR), this can be
derived from the ECG or a remote arrhythmia computer.
•You can select HR or PULSE (PULSE can only be selected if you
are monitoring invasive pressure or pleth) as the alarm parameter both share the same alarm limits.
The EASI™ option all o ws yo u to de ri ve a ll 12 standard ECG leads, us in g
a 5-electrode ECG cable and a special electrode placement. This is
described on “EASI™ 12-lead ECG Display, Printing and Recording” on
page 14-33
With the ECG or the ECG/RESP modules listed below, you can use a 5or a 3-electrode ECG cable set to derive up to 8 selectable ECG leads
which can be displayed in up to 3 channels.
Note—
EASI™: With any of the ECG modules listed below and the
EASI™ option switched on, you can use a 5-electrode ECG cable set to
derive the 12 standard ECG leads of which up to 3 can be displayed on
the monitor.
•If you are monitoring with a 3-lead set you can select one lead out
of I, II or III to be displayed in 1 channel.
•If you are monitoring with a 5-lead set you can select any two
leads from I, II, III, aVR, aVL, aVF, V or MCL
14-2 ECG and ECG/Respiration Module Section
to be displayed in
1
Introduction to ECG Parameter Module
channel 1 and channel 2. Channel 3 (CH3) will only monitor the V
lead.
•EASI™: With a 5-lead set and EASI™, you can select any lead you
want to monitor in all three channels.
Automatic 3-lead versus 5-lead Detection
•If the monitor is turned on with a 3-lead cable connected to the ECG
module and your monitor is configured for 2 or 3 ECG channels, a
limb lead will automatically be monitored in channel 1, even if you
have configured a different lead (for example aVR or aVF). After 1
minute, channels 2 and 3 will be automatically turned off.
•If you unplug the 3-lea d cable and plug in the 5-lead cable, the
Monitor automatically recovers its configured state, as long as a
lead has not been changed, or a channel has not been switched ON
or OFF manually in the meantime.
a. If you are using a 5-lead set (without using EASI™ software
and placement) and the RL electrode falls off, then the Monitor
automatically switches to a limb lead and monitors a 3-lead set
in channel 1. You can recover the Monitor's original 5-lead
state as follows:
1. Reattach the RL elec trode within 1 minute.
2. unplug and plug in the cable again, or unplug and plug in
the frontend module again.
ECG and ECG/
Respiration Module
b. If you are using a 5-lead set with EASI™ and the RL (N) or A
electrode falls off, the Mon itor cannot switch to a limb lead so
an INOP will result.
The Patient Cable
The patient cable is either single-piece or consists of 2 sections:
1. The trunk cable which connects to the monitor.
2. The lead set which connects to the patient.
The Module
• The front of the module has the key(s):
– The ECG key for parameter setup, a light will appear above the
key when you are in the setup.
ECG and ECG/Respiration Module Section
14-3
Introduction to ECG Parameter Module
ECG and ECG/
Respiration Module
– (M1002A/B only) The RESP key for parameter setup, a light
will appear above the key when you are in the setup.
14-4 ECG and ECG/Respiration Module Section
Introduction to ECG Parameter Module
ECG
ECG
M1001A
12
PIN
T
80x90
ECG/RESP
M1002A
ECG
12
PIN
T
RESP
The settings are transferred with ECG and ECG/RESP modules that have
a T on the front. For more information on Parameter Settings Transfer,
refer to “Parameter Settings Transf er” on page 14-46
ECG
M1001B
Defib
T
P
!
ECG/RESP
M1002B
Defib
T
P
!
ECG and ECG/
Respiration Module
ECG
12
PIN
ECG
80x90
12
PIN
RESP
ECG and ECG/Respiration Module Section
14-5
Introduction to ECG Parameter Module
Note—
signal delay of <20ms
ECG and ECG/
Respiration Module
All ECG and ECG/RESP modules with a P on the front have a signal
delay of <30ms
waves from a high lead ECG signal. According to the
specifications defined by AAMI the peak of the synchronized
defibrillator di scharge should be deliv ered within 60msec of the
peak of the R wave.
The Agilent M1001A/B and M1002A/B ECG modules meet the
AAMI specifications, however your biomedical engineer should
verify that your ECG/Defibrillator combination does not exceed
the recommended maximum delay of 60msec.
All ECG and ECG/RESP modules without a P on the front have a
WarningWarning
!
Your ECG/Defibrillator combina tion is conf igured to detec t R
Defib
RESP
Parameter
If you want to monitor respira tion this can be done from the ECG
electrodes. See “Respiratory Monitoring Setup” on page 14-18.
14-6 ECG and ECG/Respiration Module Section
Considerations When Monitoring ECG
Considerations When Monitoring ECG
Listed below are important considerations to remember when monitoring
ECG.
•Interference from a non-grounded instrument near the patient, and
ESU interference can cause problems with the waveform. Refer to
Recommended ECG Lead Placement for Surgical Patients in this
section for more details.
•Radiated field strengths above 0.7V/m may cause noise on the ECG
and respiration waves at various frequencies, though this noise does
not influence the accuracy of the Heart Rate or Respiration Rate. If
operating under conditions according to the EMC-standard 60601-12 (Radiated Immunity 3 V/m), field strengths above 1 V/m may cause
erroneous measurements at various frequencies. Therefore it is
recommended to avoid the use of electrically radiating equipment in
close proximity to ECG/r esp i ra ti on meas ur em en ts.
•If an ECG module is inserted into the System and ECG monitoring
commenced when the patient is in as ystole, the asystole alarm will
be generated about 15 seconds later.
•If you get an INOP caused by pacemaker activity, reduce the size of
ECG-CH1 until the status message is removed, or select a lead that
produces a smaller pace pulse. (This INOP does not occur if you are
using the EASI™ option).
ECG and ECG/
Respiration Module
•After defibrillation, the screen display will recover within 10
seconds if the correct electrodes are used and applied in
accordance with the manufacturers instructions.
ECG and ECG/Respiration Module Section
14-7
Considerations When Monitoring ECG
ECG and ECG/
Respiration Module
WarningWarning
DO NOT TOUCH THE PATIENT, OR TABLE OR INSTRUMENTS
DURING DEFIBRILLATION
When connecting electrodes a nd/or patient cables, ensure that
the ECG leads and connectors do not come into contact with
other conductive parts or earth.
14-8 ECG and ECG/Respiration Module Section
ECG Monitoring Setup
ECG Monitoring Setup
1. Check that the ECG or ECG/RESP module is inserted in the rack.
2. Plug in the patient cable to the ECG module.
3. Prepare the patient’s skin prior to placing the electrodes.
The skin is a poor conductor of e lectricity, therefore preparation of
the patient’s skin is important to facilitate good electrode to skin
contact. Recommendations:
– Shave hair from sites, if necessary.
– Wash sites thoroughly with soap and water. (Never use ether
or pure alcohol, because this increases skin resistance).
– Dry briskly to increase capillary blood flow in the tissues and
remove skin cells and oil.
– Attach clip or snap to electrode prior to placement.
4. Place the electrodes on the patient. Use electrode gel prior to
placement if pre-gelle d electrodes are not used.
Remember to select a site where th e signal will not be interfered
with by either movement or bone s.
ECG and ECG/
Respiration Module
5. Attach the electrode leads to the patient cable.
6. Switch on the monitor (if it was not on before) using power ON/OFF
switch.
Note—
In both auto and manual modes, if the lead of the first chan nel is in
INOP for at least 10 seconds, the leads in channel one and two are
automatically exchanged. This means that the lead of channel two is
displayed in channel one so that the heart rate counter can continue to
count. If the situation which caused the INOP is rectified, and the user
selected lead is out of INOP for at least 10 seconds, the leads w ill be
swapped back again.
The automatic swapping of the leads can be configured on or off during
ECG and ECG/Respiration Module Section
14-9
ECG Monitoring Setup
ECG and ECG/
Respiration Module
the installation of the monitor. If automatic swapping is set to off, the
lead in channel one is only displayed when a signal is available.
Note—
EASI™: If the E, S or I electrodes fall off, the monitor shows a
non-standard lead (that is, one of the 3 EASI™ vectors) labelled "ECG"
until the electrode is reattached.
There is no automatic lead swapping for EASI™.
Placing the
Electrodes
for ECG
Monitoring
Note—
When EASI™ lead placement has been configured, “EASI” is
shown beside the 1mV calib ration bar on the ECG wave in c hannel 1.
For EASI™ lead placement, see “Electrode Placement with a 5-Lead Set
for EASI™ 12-lead.” on page 14-33.
You can switch from EASI to standard lead placement in the Change
Lead Task Window of each channel. See “Switching between EASI and
Standard leads” on page 14-22 for details.
Note—
Electrode labels and colors are given for U.S. norm and in square
brackets [-] for European norm.
Standard Electrode Placement with a 5-Lead Set.
Angle of
Lewis
BLACK
[YELLOW]
WHITE
[RED]
BROWN
[WHITE]
1
2
3
4
GREEN
[BLACK]
14-10 ECG and ECG/Respiration Module Section
80x80
RED
[GREEN]
COLOR CODING
ECG Monitoring Setup
Label
AAMI
Color
Coding
IEC Color
Coding
RAWHITE(RED)place directly below the clavic le
and near the right shoulder.
LABLACK(YELLOW)place direct ly below the clavicle
and near the left shoulder.
CBROWN(WHITE)place on the chest as illustrated on
following page.
RLGREEN(BLACK)place in right lower abdomen.
LLRED(GREEN)place in left lower abdomen.
Note—
To ensure patient safety, all leads must be attached to the patient.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-11
ECG Monitoring Setup
ECG and ECG/
Respiration Module
For 5-Lead configuration. Place the V electrode at one of the locations
shown below, so as to get the best quality signal:
V1
V3R
V4R
Note—
For accurate V lead placement and monitoring it is important to
VE
V2
80x80
V7
V3
V4
V6
V5
locate the 4th intercostal space. The 4th intercostal space is determined
by first locating the first intercostal space. Because patients vary with
respect to body shape, it is difficult to palpate the 1st intercostal space
with accuracy. Thus, locate the 2nd intercostal space by first palpating
the little bony prominence, called the Angle of Lewis, where the body
of the sternum joins the manubrium. This rise in the sternum identifies
where the second rib is attach ed, and the space just below it is the
second intercostal space . Palpate and count down the chest until you
locate the 4th intercostal space.
•V1 on the 4th intercostal space at the right sternal border.
•V2 on the 4th intercostal space at the left sternal border.
• V3 midway between V2 and V4 electrodes.
14-12 ECG and ECG/Respiration Module Section
ECG Monitoring Setup
• V4 on the 5th intercostal space at the left midclavicular line.
• V5 on the left anterior axillary line, horizontal with V4 electrode.
• V6 on the left midaxillary line, horizontal with V4 electrode.
• V3R-V6R on the right side of the chest in positio ns corre spon ding to
those on the left.
• VE over the xiphoid pr ocess.
For posterior V lead placement, place the V electrode at one of the
following locations.
• V7 on posterior chest at the left posterior axillary line in the 5t h
intercostal space.
• V7R on posterior chest at the right po sterior axillary line in the 5th
intercostal space.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-13
ECG Monitoring Setup
ECG and ECG/
Respiration Module
Electrode Placement with 3-Lead Sets (Standard Configuration)
Note—
Not for simultaneous monitoring of more than one ECG lead. For
more than one simultaneou s cha nnel of ECG, use 5-lead electrode set.
WHITE
[RED]
aVL
aVF
+
+
-
III
+
I
-
-
aVR
II
80x80
BLACK
[YELLOW]
RED
[GREEN]
•White(RA) electrode - place directly below the clavicle and near
the right shoulder.
•Black(LA) electrode - place directly below the clavicle and near
the left shoulder.
•RED(LL) electrode - place in left lower abdomen.
Lead
Position
(-)(+)Reference
1 (I)RALALL
2 (II) RALLLA
3 (III)LALLRA
14-14 ECG and ECG/Respiration Module Section
Electrode Placement with 3-Lead Sets!
(MCL
and MCL6 Configurations)
1
ECG Monitoring Setup
ECG and ECG/
Respiration Module
WHITE
[RED]
1
2
3
4
BLACK
[YELLOW]
Select Lead I for mon itoring the MCL
M
80x80
1
L
C
6
L
MC
RED
[GREEN]
configuration. Electrodes need to
1
be moved or switched according to the following instructions. As you will
notice, you must attach lead wires to areas of the chest that do not
coincide with the electrode labels (that is, that do not coincide with the
RA, LA and LL labels).
•White(RA) electrode - place directly below the clavicle and near the
left shoulder.
•Black(LA) electrode - place on the 4th intercostal space at the right
sternal border.
•RED(LL) electrode - place on the left midaxillary line at the 5th
intercostal space.
Note—
This modified lead placeme nt allows monitoring of the MCL6 lead.
Simply select Lead II on the monitor for MCL
ECG and ECG/Respiration Module Section
.
6
14-15
ECG Monitoring Setup
Recommended ECG Lead Placement for Surgical Patients
ECG and ECG/
Respiration Module
The placing of the ECG leads will depend on the type of surgery that is
being performed. For example, with open heart surgery the electrodes
may be placed laterally on the chest or on the back.
In this case you will no longer be monitoring the standard leads.
In the operating room artifacts can sometimes affect the ECG waveform
due to the use of ES (electro-surgical) equipment. To help reduce this
you can place the electrodes on the right and left shoulders, the right
and left sides near the stomach, and the chest lead on the left side at
mid-chest. Avoi d p l aci n g the electrodes on th e up p er ar ms o t h er wi s e th e
ECG waveform will be too small.
WarningWarning
•To protect against the hazard of burns when using high
frequency surgical equipment in the OR, only the OR orange
lead sets should be used (part numbers can be found in the
Accessories section of this chapter). With this lead set,
respiration cannot be measured. To avoid triggering
unnecessary RESP alarms in this situation, it is possible to
configure the RESP parameter to be Off when the system is
switched on. This setting is made in the special
Configuration Mode.
•All specified patient lead cables are protected against the
effects of the discharge of a defibrillator.
14-16 ECG and ECG/Respiration Module Section
ECG Monitoring Setup
Caution
•When using ES equipment, never place an electrode near the
grounding pla te of the ES devi ce, as sign ifi cant inte rfere nce wi th the
ECG signal occurs.
The electrodes should be approximately equidistant from an axis
joining the ES knife and t he ES grounding plate.
•Do not apply leads directly to the external wall of the heart.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-17
Respiratory Monitoring Setup
Respiratory Monitoring Setup
ECG and ECG/
Respiration Module
How
Respiration
is Measured
Checklist
for
Respiratory
Monitoring
The system measures respiration from the amount of thoracic
impedance between t wo ECG electrodes. The change of impedence
between the two electrodes, (due to the thoracic mo vement), produces
a respiratory waveform on the screen.
For respiratory monitoring you do not need additional electrodes but
the placing of electrodes is important. If the patient is using the thoracic
muscles you can use the electrode placement as shown on page 14-19.
Some patients, due to their clinical condition, expand their chest
laterally causing a negative intrathoracic pressure. In these cases it is
best to place the two respiratory electrodes laterally in the right
midaxillary and left lateral chest areas at the m ax imum point of
breathing movement to optimize the respiratory waveform.
Note—
Respiratory monitoring is not recommended for patients who are
very active as this can cause false alarms to occur.
Note—
Implantable pacemakers which are minute ventilation rateadaptive can occasionally interact with the impedance measurement of
cardiac monitors causing the pacemakers to pace at their maximum
programmed rate.
1. Plug the patient cable into the ECG/RESP module.
2. Prepare the patient’s skin prior to placing the electrodes.
3. Attach snap or clip to the electrodes and attach the electrodes to
the patient as describe d above.
4. Switch on the monitor (if not already switched on).
14-18 ECG and ECG/Respiration Module Section
Respiratory Monitoring Setup
Placing the
Electrodes
for
Respiratory
Monitoring
Note—
Position the White (Red) and Red (Green) electrodes diagonally to
optimize the respiration waveform. Avoid th e liver area and the vent ricles
of the heart in the line between the respiratory electrodes so as to avoid
cardiac overlay or artifacts from pulsating blood flow.
This is particularly important for neonates.
Note—
EASI™: Respiratory monitoring is also possible with the the
EASI™ placement shown in “Electrode Placement with a 5-Lead Set for
EASI™ 12-lead.” on page 14-33.
Electrode Placement with a 5-Lead Set
WHITE
[RED]
GREEN
[BLACK]
Angle of
Lewis
R
E
S
80x80
P
BLACK
[YELLOW]
BROWN
[WHITE]
RED
[GREEN]
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-19
Respiratory Monitoring Setup
Electrode Placement with a 3-lead Set
ECG and ECG/
Respiration Module
ECG/RESP
with Analog
Respiration
Output
Angle of
Lewis
WHITE
[RED]
R
E
S
P
80x80
Note—
Respiration can only be monitored with an ICU ECG cable set, not
BLACK
[YELLOW]
RED
[GREEN]
with an OR ECG cable set. This is because of the higher internal
impedance of the OR cable set, required for use if electro-surgery is
being performed.
Option C01 of the M1002A/B ECG/RESP Module provides an analog
respiration output signal for us e with external devices. A standard BNC
cable is required to connect the device to the output socket situated at
the top left of the module.
Note—
Avoid accidental disconnection of the BNC cable by locking it
into the socket. Always turn the connector clockwi se.
Note—
The BNC Cable must not exceed 4 meters (13 feet) in length.
Note—
Always keep the cable away from working space.
Caution
The module provides an output signal only. Do not attempt to input any
other signal or data into the BNC out put socket.
14-20 ECG and ECG/Respiration Module Section
ECG Adjustments
ECG Adjustments
Information about the following ECG adjustments is provided on the
proceeding pages:
•Changing the lead
•Adjusting the waveform size
•ECG waveform settings
•Changing the filter bandwidth
•Changing the QRS detection mode
•Adjusting the QRS detection level
•Adjusting the alarm li mits
•Changing the alarm parameter
Changing
the Lead
A 5-lead set is option al for th e S yste m. Th is gives you a choice of leads for
the first 2 channels: I, II, III, aVR, aVL, aVF, V, MCL1. The third channel
displays one of the V leads, V
Note—
channels : I, II, III, aVR,aVL, aVF, V
, V2, V3, V4, V5, or V6.
1
EASI™: The EASI™ software gives you a choice of leads for all 3
, V2, V3, V4, V5, V6.
1
ECG and ECG/
Respiration Module
With a 3-lead set, (1 channel only), a ll th e lead labels are displayed on the
screen but only leads I, II, or III can be processed.
If you choose one of the other leads it will cause an INOP.
You cannot use a 3-lead set with the EASI™ option. If you try to use a
3-lead set with the EASI™ opti on, the prompt “EASI ECG requires 5
electrodes” will be issued.
If you are not receiving a good ECG waveform and the electrodes are
securely placed, you should try changing the lead in which you are
monitoring.
ECG and ECG/Respiration Module Section
14-21
ECG Adjustments
Confirm
Characteristics of a Good Signal
ECG and ECG/
Normal QRS complex
Respiration Module
• Tall and narrow with no notches (for a good HR me asurement).
• R-wave should be tall.
• Pacer pulses should be smaller than the R-wave height.
• T-wave should be less than the R-wave height (for a good HR
measurement)
• P-wave should be much smaller t han the T-wave.
R
P
T
Q
S
Switching
between EASI
In the Change Lead Task Window of each channel you can also switch
between EASI and Standard lead placement:
and Standard
leads
Procedure
1. Press to change to standard lead placement (if EASI is
active) or to EASI lead placement (if standard is active). To cancel
the procedure, press
Change Lead
.
2. You will be prompted to change the electrode placement. After
performing the change, press .
3. Monitoring will be interrupted for 15 seconds and will then resume
with the new setting.
14-22 ECG and ECG/Respiration Module Section
Confirm
ECG Adjustments
Adjusting
the
Waveform
Size
If either autoadjust or autosize ECG wave settings is selected (in
Configuration mode), a n automatic size adjustment will be performed in
the following situations:
•after power on
•after the module has been plugged in, or a lead has been off for
longer than 60 seconds
•between 7 and 15 seconds after lead change.
When the ECG wave has been ad justed, the gain i s frozen.
You can also manually change the siz e of the ECG wave form by entering
Adjust Size Task Window.
the
This Task Window also enables you to set the waveform to a specific size
in order to get a calibrated ECG wave recording of 10 mm/mV. Two wave
sizes, x1 or x2, are available. Select the wave size shown in the table for
your recorder and its sec tor size in order to get a calibrated 10 mm/mV
recording.:
RecorderSector sizeWave Size
M1116A/B
Module Recorder
M1117A Bedside
Recorder (CMS
Only)
40 mm
20 mm
50 mm
25 mm
x1
x2
x1
x2
ECG and ECG/
Respiration Module
Central Recorder40 mm and 50 mmx1
Note—
Pressing the key changes the si ze of thi s indivi dual
channel. Pressing the key changes the size of all three
channels simultaneously. For viewing and recording mult ilead ECG, the
size and filter of channel 1 is used for all leads of the multilead ECG.
Note—
EASI™: If you are using EASI™, you cannot adjust the gain at the
ECG Out using the key.
Note—
On touch-operated monitors, you can use the arrow buttons below
the size bar to adjust the size of the ECG waveform.
Adjust Size
Confirm
Adjust Size
ECG and ECG/Respiration Module Section
14-23
ECG Adjustments
ECG and ECG/
Respiration Module
Indication for Adjusting the Size
If the system is in manual mode and the pace pul se is t he same size or
larger than the R-wave, you can select a lead that produces a smaller
pace pulse, or you will need to change the size of the QRS complex so
that the R-wave is larger than the pace pulse otherwise the pace pulse
could also be counted and this will double the heart ra te count.
Notes—
•If you get an INOP caused by pacemaker activity the message
Pace pulses too large - reduce size of ECG-CH1
is
displayed in the status line. Reduce the size of ECG-CH1 slowly
until the status message is removed or select a lead that produces
a smaller pace pulse.
ECG Waveform Settings
There are four possible configuration settings which control the size and
the position of the ECG wave. The settings must be selected in a special
Configuration Mode by your biomedical engineering department, or the
Agilent Service Engineer.
Autoadjust
This is the factory default. The size o f t he wav e is au tomatically ad justed
to fit inside the channel. The position of th e wave is optimized so that
the middle of the wave occupies the middle of the channel, unless ST
monitoring is switched on. In this case the wave is not centered.
Autosize
The size of the wave is automatically adjusted to fit inside the channel,
without changing position.
Gain x 1
The size of the wa ve is not automatically adjusted. It is always displayed
with a gain of x1. (The amplitud e of the wave is magnified by 1000.)
Gain x 2
The size of the wave is always displayed with a gain of x 2. (The
amplitude of the wave is magnified by 2000.).
14-24 ECG and ECG/Respiration Module Section
ECG Adjustments
Changing
the Filter
Bandwidth
The filter bandwidth refers to a measure of electrical frequencies
contained wi thin an electrical signal. Depending on the meas urement
conditions, you may wa nt to change the ECG’s filter to remove some of
these electric al frequencies. The System offers a choice of filter
bandwidth settings:
•Filter
•Monitor
•Diag
These settings can be selected by pressing the key.
Indications for Changing the Filter Bandwidths
•Filter - use if signal is distorted. The filter bandwidth reduces
interference to the signal. In the operating room the filter reduces
the artifact and interference from electrosurgical units. In normal
monitoring situations the filter may suppress the QRS c omplexes
too much. It is important to remember that using the filter
bandwidth is not a replac ement for good skin prep.
If the monitor is configured for th e O.R. and it is in monitor or
diagnostic mode, it will automatically switch to filt er mode during
episodes of electro-surgical interference (ESU). It will return to its
original setting once the interfer ence stops.
•Monitor - use in normal monitoring situation. The monitoring
bandwidth filters out artifac ts which may cause fal se al ar ms.
Filter/Mon/Diag
ECG and ECG/
Respiration Module
•Diag - use when diagnostic quality is required. The diagnostic
bandwidth allows the monitor to display an unfiltered ECG wave.
This enables the clinician to detect changes in the ECG, for
example, R-wave notching and ST segments.
A letter indicating the filter bandwidth is shown next to the HR numeric.
is filter,
When ST monitoring is ON, the low end bandwidth for diag, monitor and
filter always defaults to a lower value of 0.05 Hz. This enables the
clinician to detect changes in the ECG, for example, discrete elevation or
depression of S-T segments. When ST monitoring is switched OFF, the
bandwidth automatically returns to the previously selected filter,
monitoring or diagnostic bandwidth.
M is monitoring, and D is diagnostic.
ECG and ECG/Respiration Module Section
14-25
F
ECG Adjustments
ECG and ECG/
Respiration Module
Note—
For viewing or recording multilead ECG, the bandwidth setting
for channel 1 is used for all leads of the multilead ECG.
Note—
EASI™: For EASI™ 12-lead ECG, the same bandwidth setting is
used for all leads of the ECG.
Changing
the QRS
Detection
Mode
You have a choice of auto or manual modes:
•In auto mode the QRS complexes are detected automatically.
•In manual mode the QRS detecti on level appears on the screen a s
a horizontal line across the ECG wave. This enables you to see
exactly what is causing the heart rate counter to count.
These two detection modes are available for adult, pediatric and
neonatal non-paced a nd paced patients.
Indications for Changing Mode
Auto Mode - use in n ormal monitoring situation.
Manual Mode - use in exceptional cases if the monit or miscounts the
heartrate. If arrhythmia monitoring is assigned, the monitor
automatically switches to Auto mode.
WarningWarning
In manual mode you may have to readjust the QRS detection
level if the waveform size or the shape of the QRS complex has
changed.
Paced Patients vs Non-Paced Patients
In the non-paced patient mode no pace pulses are expected and no
pace pulse rejection occurs.
In paced patient mode the pace pulses are annotated with a small dash
on the screen, as shown in the figure below:
ll
D HR
1mV
14-26 ECG and ECG/Respiration Module Section
ECG Adjustments
Auto/Manual pacing cannot be adjusted when arrhythmia channels are
assigned or an arrhythmia computer is being used. The arrhythmia
function cannot recognize which paced mode has been selected in the
Monitor. For this reason, whenever arrhythmia is assigned you have to
reselect paced mode. If you are using the Ag ilent Information Cent er
(M3150A/M3153A), you select paced mode in the Admit Window. If you
are using CCM (M2300), you have to select the paced mode in the arrhyth
functns Task Window at either the monitor or the Central Station. After
arrhythmia has been assigned during the use of an arrhythmia computer,
the Monitor stays in “Paced Patient” mode.
If heartrate is counte d irregularly or there are too many (unaccountable)
pacer marks on a paced pa tient monitored in “Paced Patient” mode, do
not use the instrument for further patient monitoring. Contact the
hospital biomedical engineer.
WarningWarning
•Beats with a pacer spike within the QRS complex may not be
detected by the System's QRS detector.
•The paced/non-paced selection should be based on your specific
patient situation - “paced mode” should be used for paced patients,
and “non-paced mode” used for non-paced patients. Using the non
paced mode with pace d patients may result in pace pulses being
counted as regular QRS complexes which could prevent an asystole
alarm from being activated.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-27
ECG Adjustments
Adjusting
the QRS
ECG and ECG/
Detection
Respiration Module
Level
Manually
Correct
Positioning
of the QRS
Detection
Level
The QRS detection level only needs to be adjusted if the monitor is in
manual mode - If the monitor is in a ut o mode, the QRS complexes are
detected au tomatically.
Note—
Ventricular fibrillation alarms will function in both automatic and
manual modes.
Non-Paced Patients
Set the QRS detection level at a location where it is guaranteed to cross
the QRS complex.
Paced Patients
The pace pulse in manual mode is set to a fixed size on the display. This
enables you to increase the size of the ECG wave independently and
thus set the QRS detection level above the pace pulses so that only the
QRS complex is causing the heart rate counter to count. An example of
incorrect and correct positioning of the QRS Detection Level is shown in
the diagram below.
R
QRS DETECTION LEVEL
PACING
PULSE
P
Q
S
14-28 ECG and ECG/Respiration Module Section
T
ECG Adjustments
Changing
Alarm
Parameter
Adjusting
the Alarm
Limits
In some situations it may be impossib le t o derive an accurate HR from an
ECG, for instance because of excessive ECG artifact or because of
inability to place el ectr odes on a pa tient wit h burns. If a pulsati le pressu re
is being monitored (pressure or pleth), you can choose PULSE as the
alarm parameter.
You can set the heart rate alarm limits according to the individual
patient’s clinical condition.
The high rate al ar m set ti n g i s also important in QR S detection. You sh ou l d
not set the upper limit too high otherwise it may cause the heart rate
counter to miscount. Set the limit no higher than 20 bea ts per minute
higher than the heart rate, taking into account the variability.
The alarm limits that you set are shared by the HR/PULSE parameters,
but only one of these parameters can be in active alarm function.
WarningWarning
Red alarms for Asystole, V. Fib. and Brady will not occur with
Pulse selected as the active alarm function. If Pulse is selected,
the following status message is displayed on the Agilent CMS:
“ECG Alarms Off. To switch on, select HR.”
ECG and ECG/
Respiration Module
Switching
ECG
Parameter
Alarms On/
Off
As both numerics are displayed on the screen, the parameter that has not
been chosen as the alarm parameter will have a sign next to th e
parameter label.
ECG parameter alarms can be switched on/off, like any ot her parameter.
If, however, arrhythmia is assigned to the bedside and all alarms are
switched off (the key has been pressed), then if you go into
the ECG Task Window, the softkey is hollow and there is
a sign beside the ECG alarm bar and HR reading.
Suspend
On/Off Alarms
ECG and ECG/Respiration Module Section
14-29
ECG Adjustments
Changing
Brady Limit
ECG and ECG/
Respiration Module
If the alarm source for HR is ECG, you can set a red brady alarm for
neonates and pediatrics. When the patient’s HR falls below the low
alarm limit, a yellow alarm is activated. If it then falls below the brady
limit a red alarm is activated.
When you go into the HR Alarm Adjust Task Window, in neonatal or
pediatric mode, you will see the brady alarm limit displayed above the
HR low alarm limit. F or ex am ple , the l o w al a rm l imi t m ay be 100 and the
brady limit 80. If the HR falls below 100, a yellow alarm is activated. If it
then continues to fall below 80, a red alarm is activated.
The difference between the HR lower alarm limit and the brady alarm
limit can be set in a special Configuration Mode by your biomedical
engineering de pa rtm en t . For exam p l e, yo u m ay wan t a lo we r al ar m li mit
of 80 with a brady alarm limit of 70 (Brady Limit Diff =10). You can even
set the brady alarm limit to be the same as the lower alarm limit. In this
case, you would always see a red low he art rate alarm.
A brady alarm delay time can also be set in this special configuration mode. This defines the amount of time the HR needs to be below the
brady limit, before a red alarm is activated. For example, if the delay
time is set to 3 seconds, the HR has to r em ain below the brady limit for 3
seconds before a red alarm is activated. Choosing a short brady delay
time (for example, less t han 3 seconds) may increase the probability of
issuing false brady alarms.
14-30 ECG and ECG/Respiration Module Section
ECG Adjustments
Tone
Modulation
If tone modulation is configured, the pitch of the QRS tone is related to
the SpO
level. If the SpO2 level drops, the QRS tone becomes low er.
2
The QRS tone is derived from the alarming parameter, which can be
configured as either HR or pulse in the “Adjust A larms” Task Window. If
the numerics for the alarming parameter are set to “Off” when the
monitor is switched on, the re will be no QRS tone and therefore no tone
modulation is possible.
If the alarm parameter source goes into INOP, the modulated QRS tone
does not stop. Instead, it is derived from PLETH, if the PLET H parameter
is on.
Note—
Tone modulation is selected in a spe cia l Configuration Mode, by
your biomedical engineering depart ment. You can have tone modulation
configured as either “Off”, “Standard” or “Enhanced”:
Standard:The tone decreases by a small amount for each drop in
the SpO
level. (This is equivalent to that used in
2
NELLCOR® equipment.)
Enhanced:The tone decreases more dramatically for each drop in
the SpO
level, and is therefore easier to recognize.
2
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-31
ECG Adjustments
Multilead
ECG Display
ECG and ECG/
and
Respiration Module
Recording
The Multilead ECG capability allows you to view and recor d a snapshot
of 8 ECG leads.
For multilead ECG viewing and recording, you will need to use a 5-lead
ECG cable.
The size and filter bandwidth of channel 1 (ECG-CH1) is used by the
system as a reference if the remaining channels have differing size and
filter bandwidth settings. For ex ample, if the size or filter bandwidth for
channels 2 and 3 are different to that of channel 1, the system will
change the size or filter bandwidth of channels 2 and 3 to correspond to
channel 1 for the multile ad procedure. After the mult ilead procedure,
the original settings will be returned
14-32 ECG and ECG/Respiration Module Section
ECG Adjustments
EASI™ 12lead ECG
Display,
Printing and
Recording
Note—
In the USA, EASI™ 12-lead ECG is only for use on adult and
pediatric patients.
The optional EASI™ 12-lead ECG capability allows yo u to view and
record or record a snapshot of 12 ECG leads.
For EASI™ 12-lead ECG viewing and recording, you will need to use a 5-
lead ECG cable. The electrode placement for EASI™ is given below.
Note—
You must select the EASI™ lead placement in the ECG
Configuration Task window for th is placement to work.
When EASI™ lead placement is selected, "EASI" is shown beside the 1mV
calibration bar on the ECG wave in channel 1.
Electrode Placement with a 5-Lead Set for EASI™ 12-lead.
ECG and ECG/
Respiration Module
[WHITE]
Note—
Great care must be taken to place the electrodes as accurately as
possible for the best qu ality EASI™ measurements.
•S - Black [IEC: Yellow] (LA) e lectrode - place on the upper sternum.
•E - Brown [IEC:White ] (V) electrode - place on the lower sternum at
the level of the fifth intercostal space.
ECG and ECG/Respiration Module Section
14-33
ECG Adjustments
12-Lead ECG
View ECG
ECG and ECG/
Respiration Module
•I - White [IEC: Red] (RA) electrode - place on the right midaxillary
line at the same level as the E electrode.
•A - Red [IEC: Green] (LL ) electrode - place on the lef t midaxillary
line at the same level as the E electrode.
•Green [IEC: Black] (REF) electrode - place anywhere.
If the electrode plac ement shown is not possible on your patient
(because of plasters or therapeutic equipment), the E and S electrodes
can be moved in parallel to the patient’s right, but this move should be
kept as small as possible.
If the electrode plac ement shown is not possible due to pulmonary
problems, the A and I electrodes can be moved up or down in parallel.
Moving the electrodes down produces better measurements than
moving them up.
Note—
Respiratory monitoring is also possible with the the EASI™
placement. Respiration is measured between the I and A electrodes.
T o View a 12-
1. Press in the Procedures task window,
Capture 12-Lead
Lead ECG
or
1. in the ECG Main Task window (channel 1 ).
2. Press to view/preview the 12-lead snippets.
14-34 ECG and ECG/Respiration Module Section
Example of a 12-lead E CG display
Print ECG
Record ECG
ECG Adjustments
ECG and ECG/
Respiration Module
3. You can change the size using the softkey or the up
Adjust Size
and down arrows. (Pressing t h e key changes t h e size of the 1mV
calibration bar f or the first channel . T he wa ves change size 1 sec ond
after you have finish ed adjusting the calibration bar.)
Press in the 12-lead Task window to print the 12-lead
snippets to the local or central pr inter.
Press in the 12-lead Task window to record the 12-lead
snippets on a central or local recorder.
Note—
Use diagnostic bandwidth for the opt imal signal definition.
Due to network transmission restrictions, the bandwidth printed on a
central recorder is of reduced resolution.
The same filter bandwidth is used by the system for all channels.
The 12 lead snippets are all the same size. The initial size is set in the
EASI™ ECG Configuration Task Window in Configuration Mode.
Adjust the size to "x2" to get 1cm/mV on the printout.
ECG and ECG/Respiration Module Section
14-35
ECG Adjustments
ECG and ECG/
Respiration Module
You cannot adjust the size of the waves after you have printed them. If
you need to change the size of the waves, you should capture the 12leads again.
Scheduled or
Triggered 12Lead Reports
EASI™
Configuration
Press in the 12-lead Task window.
Auto Reports
The type of alarm that trig gers a report as well as the kind of report
(printer or recorder) can be set up in the EASI™ Configuration Task
Window in Configuration Mode. You can select between
•No Reports.
•Reports On Alarm, triggered by ST alarms, or ST & HR alarms. The
type of alarm that triggers a report can be set in the EASI™ ECG
Configuration Task Window. Reports generated by an ST & HR
alarm have a symbol, reports for an ST alarm have an
symbol.
Arrhythmia alarms will also generate a 12-lead printout if you
select ST & HR alarms.
•Scheduled reports, which can be at any multiple of half an hour
between a half an hour and 48 hours (0.5, 1.0, 1.5, 2.0,...48).
Scheduled reports are ma rked with a symbol.
•All, which selects the scheduled reports you have set and any
reports triggered by an ST or S T&HR alarm.
In the ECG configuration in configuration mode, you must firs t select
EASI™ lead placement. Then, in addition to the normal parameters for
the ECG, there are a number of para meters just for the 12-lead.
•The 12-lead sequenc e lets you choose between nor mal and
Cabrera order fo r the display, printing and recordi n g o f th e 1 2- l ead
snippets.
Normal order displays the lead snippets as follows:
IV
IIV
IIIV
14-36 ECG and ECG/Respiration Module Section
1
2
3
ECG Adjustments
aVRV
aVLV
aVFV
4
5
6
Cabrera order displays the lead snippets as follows::
aVLV
IV
-aVRV
IIV
aVFV
IIIV
Note—
The 12-lead snippets on the recorder are paired so that you
1
2
3
4
5
6
can cut the strip up and stic k them on a sheet to get the Normal or
Cabrera order given above.
•You can select the size of the 12-lead window as large or normal.
The large window will cover all of the real-time traces on the
screen. The normal window leaves space for 1 full trace above the
task window.
ECG and ECG/
Respiration Module
•The next settings are for the automatic EASI™ reports, as described
in “Scheduled or Triggered 12-Lead Reports” on page 14-36. Here
you can select
– Whether th e rep o rt is gener ated for alarms, at scheduled times,
for alarms and on schedule, or no reports at all.
– Whether the alarm reports are triggered by ST or ST & HR
alarms. HR alarms are sele cted auto mati call y for mo nitor s with
no ST functionality.
– The time interval between scheduled reports.
ECG and ECG/Respiration Module Section
14-37
ECG Adjustments
– Whether alarm reports are output to the printer or recorder.
ECG and ECG/
Respiration Module
– Whether scheduled reports are out put to the printer or
recorder.
•12-lead on SDN is set to "Off". This parameter is for future use.
Note—
Only one precordial-lea d can be transmitted on SDN.
•Set the initial size for the 12-lead snippets to
– AutoAdj for a best fit.
– GainCH1 to use the gain for channel 1 for all snippets.
– Gain x1 sets all snippets to 0.5cm/mV.
– Gain x2 sets all snippets to 1cm/mV.
Note—
The same gain applies to all 12 le ads.
14-38 ECG and ECG/Respiration Module Section
Respiration Adjustments
Respiration Adjustments
Changing
the
Respiration
Detection
Mode
The monitor counts the respiration using one of two modes, auto or
manual.
In auto mode the moni tor counts the respiratio n an d ad ju sts th e de tection
level automatically de pending on waveform height , presence of cardiac
artifact and absence of valid breaths. The auto mode is used in the
following situations:
•When breathing is spontaneous with or without Continuous Positive
Airway Pressure (CPAP).
•With ventilated patients, (except Intermitten t Mandatory Ven tilation
- IMV).
•When the Respiration Rate (RR) does not approach t he Heart Rate
(HR).
In manual mode you set the de tection level for counting the respiration. It
is important to remember that if the depth of breathing changes the
detection level may require adjustment. The manua l mode should be
selected in th e following situations:
•When the RR approximately equals the HR.
ECG and ECG/
Respiration Module
Adjusting
the
Respiration
Detection
Level
•When the respiration is weak, heart activity, or movements of the
chest wall caus ed by the heart, can cau se artifacts. (In this situation,
adjust electrode placement to improve the signal. Take special care
when adjusting EASI ™ pl aceme nt - see “Elec trod e Plac ement wi th a
5-Lead Set for E ASI™ 12-lead.” on p age 14-33).
In manual mode the Respiration Detection Level is shown as a horizontal
line across the respiratory waveform. Each downward stroke of the
waveform which crosses the detection line is counted as a respiration.
The manual mode is more sensitive than the auto mode to any impedance
changes, including artifact. You should check the waveform on the screen
to ensure that it represents the patient's breathing pattern.
ECG and ECG/Respiration Module Section
14-39
Respiration Adjustments
ECG and ECG/
Respiration Module
Caution
•If you move the detection level towards the waveform's baseline,
the monitor is more likely to detect impedance chan ges due to
cardiac activity. If apnea oc cu rs i n thi s sit ua ti on th e mon ito r migh t
falsely interpret cardiac activity as respiratory activity and not
detect the apnea.
The apnea alarm would not be activated.
You would remain unaware of this critical patient condition.
• Switching the RESP alarms OFF, also has the effect of switching
the apnea alarms OFF.
14-40 ECG and ECG/Respiration Module Section
ECG and RESP Alarm and INOP Messages
ECG and RESP Alarm and INOP Messages
ECG Alarm
Messages
The Heart Rate alarm messages are rated in order of severity:
*** Red
** Yellow
INOP message
Alarm tone = a single chime repeated every second.
INOP tone = a single beep repeated every 2 seconds.
Alarm limits are dependent on patient’s condition.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-41
ECG and RESP Alarm and INOP Messages
Physiological Alarms
ECG and ECG/
Alarm MessageConditionVisual Indication
Respiration Module
***ASYSTOLE
1
The interval between two
QRS complex es has
HR numeric (=0) blinks.
Red alarm lamp.
exceeded 4 seconds.
***VENT FIB
The heart is fibrillating.
Only in adult
HR numeric blinks.
Red alarm lamp.
1
configurations.
***BRADY
1, 2
60<80
HR has dropped below the
selected Brady limit.
**HR 160>150 HR has exceeded the high
alarm limit.
**HR 90<100 HR has dropped below the
low alarm limit.
1
If PULSE is the HR numeric source, there is no ASYSTOLE, VENT FIB
HR numeric blinks.
Red alarm lamp.
HR numeric blinks.
Yellow alarm lamp.
HR numeric blinks.
Yellow alarm lamp.
or BRADY red alarm.
2
Brady alarms will on ly be generated in neonatal or pediatric mode.
Technical Alarms
INOP messages appear when the monitor cannot measure or process
signals properly. This could be du e to patient-related or equipment related problems but you must always check the patient’s condition
first. The INOP message is accompanied by an audible alarm which can
be silenced with the key.
Silence/Reset
Audible
Indication
Alarm tone
Alarm tone
Alarm tone
Alarm tone
Alarm tone
14-42 ECG and ECG/Respiration Module Section
Equipment-Related INOPS
ECG and RESP Alarm and INOP Messages
INOP MessageConditionVisual Indication
LEADS OFFNot all the required leads
HR numeric displays -?-. INOP tone
are attached or electrodes
have been displaced.
LEADS OFF XXSingle electrode fro m lead
XX detached.
If INOP occurs in
channel 1, HR numeric
displayed as -?-. After 10
seconds lead in channel
2 switches to channel 1
so that heart rate
counter can operate. HR
numeric displayed as
normal.
LEADS OFF
(EL.X)
Single electrode from lead
E, S, or I is detached
(EASI™ lead set).
A non-standard ECG
signal
is shown in channel 1,
the HR numeric is
derived.
Channels 2 and 3 show
baseline.
ST is disabled.
Audible
Indication
INOP tone
INOP tone at
beginning.
ECG and ECG/
Respiration Module
ECG/RESP UN
PLUGGED
ECG EQUIP
MALF
ECG/RESP module
switched on and has been
un-plugged from the rack.
Malfunction in the ECG
hardware.
HR numeric displays -?-. INOP tone
HR numeric displays -?-. INOP tone
ECG and ECG/Respiration Module Section
14-43
ECG and RESP Alarm and INOP Messages
RESP Alarm
Messages
ECG and ECG/
Respiration Module
Alarm MessageConditionVisual Indication
***APNEARESP has exceeded the
**RESP 25>20 RESP has exceeded the
**RESP 6<8 RESP has dropped
INOP MessagesConditionVisual Indication
RESP ERRATICPoor contact between
Physiological Alarms
preset apnea alarm
limit.
high alarm limit.
below the low alarm
limit.
Patient-Related INOPS
the electrode and skin
or excessive patient
movement.
RESP numeric (=0)
blinks. Red alarm lamp.
RESP numeric blinks.
Yellow alarm lamp.
RESP numeric blinks.
Yellow alarm lamp.
? next to RESP label.None
Audible
Indication
Alarm tone
Alarm tone
Alarm tone
Audible
Indication
14-44 ECG and ECG/Respiration Module Section
Equipment-Related INOPS
ECG and RESP Alarm and INOP Messages
INOP MessageConditionVisual Indication
ECG/RESP UN
PLUGGED
ECG/RESP module
switched on and un-
RESP numeric displays -?-. INOP tone
plugged from the
rack.
RESP LEAD OFFRESP lead fallen off,
RESP numeric displays -?-. INOP tone
or electrode
detached, or patient
cable unplugged.
RESP EQUIP
MALF
Malfunction in the
RESP hardware .
RESP numeric displays -?-. INOP tone
Audible
Indication
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-45
Parameter Settings Transfer
Parameter Settings Transfer
ECG and ECG/
Respiration Module
WarningWarning
Make sure that the ECG placement (Standard or EASI™)
selected on the monitor to which you are transferring, and your
patient’s electrode pl acement are the same.
The following settings can be transferred with ECG and ECG/RESP
modules that have a T on the front. For more information on Parameter
Settings Transfer, refer to “Parameter Settings Transfer” in Chapter 3.
Setting Name Meaning
Standard Lead on CH1
/ CH2
EASI™ Lead on CH1 /
CH2 / CH3
Paced ModePatient paced / non pace d
Alarm parameterHeart rate or pulse rate alarm source
Alarm limitsHeart / Pulse and respiration rate alarm limits
Apnea time Apnea delay time
Note—
EASI™: Both the EASI™ settings and the standard settings are
transferred.
Selected ECG lead
Selected ECG lead
14-46 ECG and ECG/Respiration Module Section
Accessories and Ordering Information
Accessories and Ordering Information
Note—
All Agilent patient lead cables are protected against the effects of
the discharge of a defibrillator.
Trunk Cables
Note—
You cannot use 3-lead cab les with the EASI™ option.
3-lead
AAMI:M1500A - Trunk cable (2.7m)
M1540C - Trunk cable (0.4m)
IEC:M1510A - Trunk cable (2.7m)
M1550C - Trunk cable (0.4m)
3-lead One Piece Cables
AAMI:M1970A - OR (1.9m)
M1972A - ICU (Sna p, 1.9m)
IEC:M1980A - OR (1.9m)
M1981A - ICU (Grabber, 1.9m)
5-lead
ECG and ECG/
Respiration Module
AAMI:M1520A - Trunk cable (2.7m)
M1560A - Trunk cable (0.4m)
IEC:M1530A - Trunk cable (2.7m)
M1570A - Trunk cable (0.4m)
5-lead One Piece Cable
AAMI:M1975A - OR (2.5m)
M1977A - ICU (Sna p, 2.5m)
IEC:M1985A - OR (2.5m)
M1986A - ICU (Grabber, 2.5m)
If you are using EASI™ lead placement, replace the standard leadplacement label on the cable with the appropriate EASI™ Label.
ECG and ECG/Respiration Module Section
14-47
Accessories and Ordering Information
ECG and ECG/
Respiration Module
Since these Agilent cables were designed specifically for your Agilent
V24 and V26 Monitor, we re commend the use of only A gile nt cables and
leadsets for optimum monitor pe rformance.
Note—
The following listed accessories are not shipped with every
instrument. Contact your Agilent sales representative for details.
Accessories
3 lead 5 lead
combiner blockM1501AM1502A
cable organizer (standard and OR
For safety an d environmental specifications, please refer to the Mo nit or
Installation and Patient Safety Chapter 10.
ECGCardiotach
Adult Range15 to 300 bpm
Pedi/Neo Range15 to 350 bpm
Accuracy +
Resolution1 bpm
Sensitivity>= 200 µV peak
Pace Pulse Detection:
• 100 µs pulse width amp litude greater than 3 mV
• 50 µs pulse width amplitude greater than 5 mV
Pace Pulse Rejection:
• Meets the requirements for AAMI EC 13-1993 Standards for
cardiac monitors (automode).
1% of range
14-50 ECG and ECG/Respiration Module Section
ECG/RESP Performance Specifications
Filter Bandwidths
Mode Adult Pediatric / Neonatal
Diagnostic Mode 0.05 to 130 Hz
a
0.5 to 130 Hz
a
Monitoring Mode 0.5 to 40 Hz 0.5 to 60 Hz
Filter Mode 0.5 to 20 Hz 0.5 to 20 Hz
a. The M1001B and M1002B have a bandwidth of 0.05Hz to 150Hz.
ECG Output
Signal Gain:640 to 6400 depending upon display gain
ECG and ECG/
Respiration Module
Full Scale Display:6.4V
pp
Gain ErrorConstant deviation <20%
Baseline Offset<180mV + ECG offset
Signal Delay:<30ms
a
a. ECG Output Signal Delay range for ECG and ECG/RESP modules:
All ECG and ECG/RESP modules without a P on the front have a
signal delay of <20ms
All ECG and ECG/RESP modules with a P on the fron t have a sign al
delay of <30ms
WarningWarning
EASI™ 12-Lead: If you are using EASI™, the signal gain is fixed at
3400. You cannot adjust the gain at the ECG Out using the
Adjust Size
.
Certain defibrillators may double trigger on the R-wave and Twave, depending on the size and shape of the signal.
Use an alternative lead which does not double trigger, or change
to non-EASI mode for cardioversion procedures.
ECG and ECG/Respiration Module Section
14-51
ECG/RESP Performance Specifications
ECG and ECG/
Respiration Module
Defib
waves from a high lead ECG signal. According to the
specifications defined by AAMI the peak of the synchronized
defibrillator di scharge should be deliv ered within 60msec of the
peak of the R wave.
The Agilent M1001A/B and M1002A/B ECG modules meet the
AAMI specifications, however your biomedical engineer should
verify that your ECG/Defibrillator combination does not exceed
the recommended maximum delay of 60msec.
WarningWarning
!
Your ECG/Defibrillator combina tion is conf igured to detec t R
14-52 ECG and ECG/Respiration Module Section
ECG Output Impedance
ECG/RESP Performance Specifications
Tip (ECG & Marker)2.6kΩ (typically)
Ring (ECG only)2.2kΩ (typically)
Marker Input Requirements
Signal Type0 to -12V (source < 7kΩ)
Current1mA
Falltime<100µs
Duration>4ms
Alarms
Adult Pediatric Neonatal
HR Alarm 15 to 250 bpm 15 to 300 bpm 15 to 300 bpm
Limit Range
Bradycardia
not applicable 20 to 295 bpm 40 to 295 bpm
Alarm
Limit Range
ECG and ECG/
Respiration Module
Adjustment:Steps of 5bpm
High/Low delay:10 seconds according to AAMI EC 13-1993
standards.
Asystole:•no detected beat within 4 seconds
•alarm within 3 seconds after this
criterion is met according to AAMI
EC 13-1993 standards.
INOP Alarms:Triggered if the offset potential rises above
500mV or leads off.
ECG and ECG/Respiration Module Section
14-53
ECG/RESP Performance Specifications
ECG and ECG/
Miscellaneous
Respiration Module
Baseline RecoveryLess than 1 second after defibrillation
Display Update 2 seconds nominal.
14-54 ECG and ECG/Respiration Module Section
RESPRespiration Rate
ECG/RESP Performance Specifications
Adult/Pedi Range:0 to 120 rpm
Neonatal Range0 to 180 rpm
Accuracy+
Resolution1 rpm
Alarm
Adult/Pedi Neonatal
High Alarm Limit 10 to 100 bpm 30 to 150 bpm
Low Alarm Limit 0 to 95 bpm 0 to 145 bpm
AdjustmentSteps of 1 under 20 rpm
RESP Alarm DelayIf limit is between 0 and 20 rpm:
1 rpm at 60 rpm
Steps of 5 above 20 rpm.
As soon as value goes below limit setting.
If limit is above 20 rpm:
10 seconds after limit exceeds limit setting.
ECG and ECG/
Respiration Module
Apnea Alarm Delay• Range of 10 to 40 seconds .
• Adjustment in Steps of 5 seconds.
• No detected breath within the adjusted
apnea delay time.
Alarm2 seconds after this criterio n is met.
INOP Alarms:Triggered if the impedance rises above 2 kΩ
or leads off.
Miscellaneous
Bandwidth0.3 to 2.5 Hz (- 6 dB)
ECG and ECG/Respiration Module Section
14-55
ECG/RESP Performance Specifications
NoiseLess than 25 mΩ (input reference).
ECG and ECG/
Respiration Module
ECG/RESP
with Analog
Respiration
Output Performance
Specifications
(CMS Only)
Output Range-3.5V to +3.5V
Amplitude1.0V/Ω change of transthoracic impedance
Offset0.0V ±1V
Delay<90ms
PolarityRising (positive) edge at inspiration
Load Capacitance<100nF
Load Resistance>1k
GroundFloating, isolated
ConnectorBNC
Output Delay
Actual delay of the respiration output signal depends upon the
respiration rate. When used with a critical timing device, careful
evaluation of the device’s compatibility is essential.
Cable Specification
Proper operation of the analog output is only guaranteed if the interface
cable meets the following specifications:
1. Remove any adhesive used to secure the electrodes to the patient
2. Sponge cables with warm water and soap, or another suitable
3. Check each cable for corrosion, cracks and deterioration.
Do not autoclave cables or electrodes or heat them above 75°C (167°F). If
the metal surfaces become tarnished, they may be cleaned with any
cleaner which does not leave a residue. Do not use any metallic abrasive
such as steel wool. The cables should be stored in an environmental
temperature between -2 0°C to 75°C (-68°F to 167°F). They should be hung
up or laid flat to prevent damage to the cable.
ECG and ECG/
Respiration Module
and wipe off any remaining electrolyte fr om the electrodes. (If
adhesive tape residue has t o be removed, use a plaster remover
solution or the Scholl Mfg. Co. Double seal tape remover. Acetone,
alcohol, ammonia, chl oroform or other strong solvent s are not
recommended, becau se th ey will even tu all y da mage the vinyl
cabling.)
cleaning solution, and dry. Do not immerse them in water.
ECG and ECG/Respiration Module Section
14-57
ST Segment Monitoring
ST Segment Monitoring
ECG and ECG/
Respiration Module
In this section you will find information on setting up ST segment
monitoring, adjusting alarm limits, and troubleshooting.
If you have problems with the monitoring (for example, an INOP
message appears on the screen), you will find information on alarms and
INOPs on page 12-53.
For a list of the parameter settings related to ST segment monitoring
that are stored inside the ECG and ECG/RESP modules with a T on
front, turn to page 12-55.
To find specific information check either the contents list at the front of
the user guide or the alphabetical index at the back.
This application provides ST level change information; the
clinical significance of this ST level change information should be
determined by a physician.
WarningWarning
ST Segment
Using EASI™, ST monitoring can be done on multiple V-leads.
Monitoring
Note—
with EASI™
14-58 ECG and ECG/Respiration Module Section
If an electrode falls off, the monitor shows a non-standard lead
(that is, one of the 3 EASI™ vectors) labelled "ECG" until the electrode
is reattached.
No ST analysis is possible in this time, and ST analysis is switched off
automatically.
1. Assessment of EASI derived 12-Lead ST measurements is recommended for patients that meet the
following requirements:
Ages: 33 to 82
Heights: 147 to 185 cm (58 to 73 in)
Weights: 53 to 118 kg (117 to 261 lb)
Height-to-Weight Ratios: 1.41 to 2.99 cm/kg (0.25 to 0.54 in/lb)
1
Introduction to ST Segment Monitoring
Procedures
Introduction to ST Segment Monitoring
The ST segment monitoring for System measures the elevation or
depression of the ST segment on up to three leads of a patient’s ECG.
The corresponding ST measurements:
•are presented as numeri cs o n t he Standard Display. These ST v al ues
are as follows:
ST 1 is the ST measurement from the channel 1 ECG lead.
ST 2 is the ST measurement from the channel 2 ECG lead.
ST 3 is the ST measurement from the channel 3 ECG lead.
•are shown graphically in the ST Analysis Task Win dow, an
integrated display of current ST values and waveforms, HR a nd ST
trends, reference ST values and waveforms, and net change
•are available in Vital Signs and Graph Trends displays and reports in
Patient Data
• can be recorded with reference beats at the bedside or at a central
recorder. Recordings made at the Patient Information Center do not
include referenc e beats.
ST Analysis is accessed through the ECG Channel 1 Task Window, or
through . The ST Analysis Task Window provides quick
access to information needed to make an initial assessment of the
significance of ST changes.
ECG and ECG/
Respiration Module
In the left side of the window, up to three ST trends are shown, with the
HR trend graphed in the ST1 plot area. Trends can be scaled at +
5.0, or +10.00 mm; the current scale is displayed in the box above the
+
Change Scale
softkey label. The trend time span choices are 1, 3, 6 or
1.0, +2.0,
12 hours for the standard database, 1 or 3 hours with high resolution
storage; the span for the c urrent trends is noted in the box above the
Change Time
label. The bar cursor provides a mechanism for
retrieving stor ed values for disp lay in the cursor spot on the right side of
the window.
ECG and ECG/Respiration Module Section
14-59
Introduc tion to ST Segment Monitoring
ECG and ECG/
Respiration Module
In the right side of the window, the current or recalled ST complexes are
displayed in the cursor spot at the right. To their left in the reference spot the reference set of complexes for use in analysis is displayed. Up
to 242 sets of realtime algorithm outputs (ST values, beats and
measurement points) can be trended and recalled. The default reference
beat can be updated to any of the stored values.
The net change between the current and reference values is listed at the
extreme right of the Task Window.
The arrow keys can be used to move the cursor to different time
columns within the trend area. On touch-operated systems, the arrow
buttons below the trends can be used to perform the task of the arrow
keys. Alternatively, the trend area can be directly selected by touch or
mouseclick. This will bring u p th e ST st rored for the selected time.
Operation
Notes
14-60 ECG and ECG/Respiration Module Section
•ST measurements a re not performed on ventricular or A-V
sequential paced beats. ST measurements are performed on
atrially paced beats.
•Arrhythmia and pulse monitoring are fully compatible with ST
segment monitoring.
Introduction to ST Segment Monitoring
•Usually VPBs do not affect the accuracy of the ST segment
measurement. However, rhythms that are primarily ventricular can
cause incorrect ST measurements.
•If you are using a touchscreen, selecting the trend area will bring up
the ST segment stored for that particular time.
Note—
The System automatically marks the following ST-related
adjustments as events in Patient Data:
•Whenever an ECG lead is changed.
•Whenever ST parameter measurement points are changed.
You can review the ST adjustment markers on the Graphs Display. The
label is “ST” or, if part of a graphs group, “ST Adjust”. See “Marking and
Reviewing Events” in Chapter 8 for more information.
ECG and ECG/
Respiration Module
Data
Transfer
ST Analysis functions are available during transfers to the M1235A Data
Transfer Module, but they are deactivated during a transfer to the
monitor.
ST trends of transferred data can be viewed in the ST Analysis Task
Window and in the Vital Signs and Graphs Trends screens accessible
under the key. Transferred values are indicated by a “T”
on the trend time line. However, ST waveforms are not transferred; thus
transferred measurements cannot be used as the reference beat.
Trends/Calcs
ECG and ECG/Respiration Module Section
14-61
ST Segment Monitoring Setup
ST Segment Monitoring Setup
ECG and ECG/
Respiration Module
•Standard Lead placement:
– ECG channel 1 should display a good cardiotach lead. This is
– ECG channel 2 should display a good lead f or monitoring the
typically lead II.
ST level. This is typically aVF.
– ECG channel 3 is always lead V, which is typically V
for
5
monitoring the ST level.
•EASI™ Lead placement:
– ECG channel 1 should display a good cardiotach lead. This is
typically lead II.
– ECG channel 2 should display a good lead f or monitoring the
ST level. This is typically V
– ECG channel 3 is always lead V, which is typically V
.
2
for
5
monitoring the ST level.
Notes—
•If monitoring with only one lead, choose the lead wh ere both beat
detection and ST segment evaluation are possible. If monitoring
with two leads, leads II and V
are typically chosen. Reme mb er ,
5
the number of ECG channels that you monitor is independent of
the number that you display. For example, you can tur n on ECGCH1, CH2, and CH3 (and thus monitor three ST parameters) but
display only ECG-CH1.
•When you turn an ST parameter on, the monitor automatically
adjusts the low end of the bandwidth (filter, monitor, or
diagnostic) to 0.05 Hz, the diagnos tic quality need ed for ST
monitoring. There is no change to the high end of the bandwidth.
When you turn ST monitoring off, the monitor automatically
returns to the original low end of the selected bandwidth.
14-62 ECG and ECG/Respiration Module Section
ST Segment Monitoring Setup
•If the monitor is configured to switch to the filter bandwidth during
ESU interference, you will get an ST INOP (Cannot A nalyze STx)
during ESU interference. When the ESU interference stops, the low
end of the bandwidth range automatically changes back to.05 and
ST monitoring resumes.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-63
ST Segment Monitoring Setup
ST Segment
Measure-
ECG and ECG/
ment Points
Respiration Module
The ST measurement for each QRS complex is the vertical difference
between where two measurement points on the wave:
The isoelectric point provides the baseline for the measurement.
The ST point provides the othe r measurement point.
R-WAVE PEAK
AT 0 MSEC
J POINT
P
ISOELECTRIC
POINT
DEFAULT =
- 80 MSEC
Q
S
ST MEASUREMENT
POINT
DEFAULT =
+ 108 MSEC
T
DIFFERENCE =
}
ST VALUE
You can adjust both measurement points. Your monitor is configured so
that you set the ST Point in reference to:
the R-wave peak.
or
the J Point, which is the transition between the QRS and the ST
segment.
Note—
The ST measurement point may need to be adjusted if the
patient’s heart rate or ECG morphology changes significantly. As you
move the cursor, the ST values at the right of the ST Measurement
Points Task Window are updated to show the effects of moving the
cursors. These values may differ from those in the Standard Display,
since the Standard Display S T va l ues are onl y u pdated every 15 seconds.
14-64 ECG and ECG/Respiration Module Section
ST Segment Monitoring Setup
Waveform
Storage/
Recall
Adjusting
the
Reference
Up to 242 sets of ST beats can be stored, each with up to three channels
of data. After 242 sets of data are stored, the database rep l aces the oldest
entry with new data. The dura tion and frequency of stor age is dependent
on the frequency of parameter storage in Patient Data. If patient data is
being collected every 12 seconds, ST wave storage occurs at 1 minute
intervals. If p a tien t data is collected on ce p er mi n ut e, S T st or ag e occurs at
5 minute intervals. During an ST alarm, wave storage is once per mi nut e.
All stored waves are cleared after a discharge procedure, when the
monitor is turned off for more tha n 3 hours, or when monitoring mode is
reset.
Setting the reference values enables you to select new or temporary
reference beats. When ST analysis is begun, the second valid beat is
selected as the default reference beat and is displayed in the reference
spot. Any trended beat can then be compared with the reference. You can
change the reference in the Adjust Reference Task Window.
In addition, any trended beat can be used as a temporary reference,
without affecting the saved reference. A temporary reference remains in
effect until you change it or you leave the Task Window. A temporary
reference can be retained as the saved reference.
If the monitor is turned off for more than 3 hours, the saved reference will
be reset to the second valid ST beat after restart. The reference is also
reset whenever the ST or ISO point is changed.
ECG and ECG/
Respiration Module
Saved Referencethe stored reference beat (This may be the
default beat, or a trended beat selected by
the user.)
Temp Reference a trended beat selected by the user and
displayed in the reference spot on a
temporary basis for analysi s
Change Referencea way for the user to replace the defa ul t
reference wit h a temporary reference
complex
ECG and ECG/Respiration Module Section
14-65
ST Segment Monitoring Setup
ECG and ECG/
Respiration Module
Adjusting
the Alarm
Limits
Each ST parameter has its own alarm limit. The alarm is triggered when
the value of the ST parameter exceeds its alarm limit for more than 1
minute. The alarm will be a yellow alarm.
Turning ST alarms on or off affects all of the ST parameters. (If you need
to disable alarms for an indiv idual parameter, adjust the alar m limits so
that reaching an alarm condition is virtually impossible.)
Caution
When more than one ST parameter is in alarm, only one alarm message
is displayed.
Note—
The displayed scales for ST alarm limits (in the Adjust ST Alarms
and Alarm Limits Task Windows, and in Vital Signs and Graph Trends)
are pre-configured to +
you can set ST alarms within is -9.8 to +9.8 mm.
1, +2, +5, or +10 mm. However, the actual range
14-66 ECG and ECG/Respiration Module Section
ST Alarm and INOP Messages
ST Alarm and INOP Messages
The ST alarm messages are rated in order of severity:
**Yellow
INOP
Alarm tone = a single chime repeated every second.
INOP tone = a single beep repeated every 2 seconds. Alarm limits are
dependent on patient’s condit ion.
The alarm and INOP messages are listed below, with the “x” representing
ST1, ST2, or ST3.
ECG and ECG/
Respiration Module
ECG and ECG/Respiration Module Section
14-67
ST Alarm and INOP Messages
Physiological Alarms
ECG and ECG/
Alarm Message Condition
Respiration Module
Visual
Indication
**STx -1.2<-1.0ST below the low alarm limit ST numeric
Audible
Indication
Alarm tone
blinks. Yellow
alarm lamp.
**STx 1.3>1.0ST above the high alarm limit ST numeric
Alarm tone
blinks. Yellow
alarm lamp.
ST Alarm Limit Ranges
-9.8mm to +9.8mm
Technical Alarms
INOP messages appear when the monitor cannot measure or process
signals properly. This could be du e to patient- related or equipment related problems but you must always check the patient’s condition
first. Some INOP messages are accompanied by an audible alarm which
can be silenced with the
Silence/Reset
key.
14-68 ECG and ECG/Respiration Module Section
Patient-Related INOPS
ST Alarm and INOP Messages
INOP Message Condition Visual Indication
ERRATIC STxThe variation between
STx numeric displays -? - None
measured ST values
exceeded the limits for
valid data.
ST PACED BEATS The algorithm recognizes
STx numeric displays -? - None
an unacceptable number
of paced beats. This could
be caused by an
exceptionally noisy ECG
signal.
CANNOT ANALYZE
STx
The algorithm cannot
generate a valid value
STx numeric displays -? - None
(e.g., noisy ECG) and
none of the above
conditions apply.
Equipment-Related INOPS
INOP Message Condition Visual Indication
Audible
Indication
Audible
Indication
ECG and ECG/
Respiration Module
NO STx SOURCEThe channel of ECG
corresponding to this ST
parameter is off.
STx OVERRANGEThe alg orithm calculated
a value >
25 mm or < -25
mm
*If the Unit Type for the Configuration Set is OR, there is no I NOP to ne.
ECG and ECG/Respiration Module Section
STx numeric displays -? - INOP tone *
STx numeric displays -? -INOP tone *
14-69
Parameter Settings Transfer
Parameter Settings Transfer
ECG and ECG/
Respiration Module
The following settings related to ST Segment Monitoring can be
transferred with ECG and ECG/ RESP modules that have a T on the
front. For more inform ation on Parameter Settings Tr ansfer, refer to
“Parameter Setting Transfer,” Chapter 3.
Setting Name Meaning
Alarm limits Heart/Pulse and respiration rate alarm limits
ISO points Isoel ect ri c measurement poi n t
ST pointsST measurement point
14-70 ECG and ECG/Respiration Module Section
Performance Specifications
Performance Specifications
ST SegmentMeasurement Range
Measurement Range-25.0 mm to 25 mm
Resolution Range+
Intermediate
Resolutions
LeadsFirst and second channels select able from I,
Measurement Points
Measurement PointsReferenced to peak of QRS complex or set
Measurement
Resolution
Isoelectric Point Rang e-460 ms to 460 ms from QRS peak
0.1 mm at highest ECG gain (50 mm/mV)
0.1 mm at 40 mm/mV
+
0.1 mm at 20 mm/mV
+
0.3 mm at 10 mm/mV
+
II, III, aVR, aVF, aVL, MCL1, V. Third cha nnel
is V.
60/80 ms from the user-selected J-point
4 ms
ECG and ECG/
Respiration Module
ST Point Range-380 ms to 460 ms from QRS peak
J-Point Range-460 ms to 460 ms from QRS peak.
ECG and ECG/Respiration Module Section
14-71
Performance Specifications
Alarms
ECG and ECG/
Range-9.8 mm to 9.8 mm
Respiration Module
AdjustmentSteps of 0.2 mm
Alarm delay1 minute af ter th e val ue exce eds the se t li mit
value
INOP AlarmsRefer to Patient and Equipment related
INOP tables.
Miscellaneous
Median ValueUpdated every 15 seconds
Automatic AnnotationMeasurement point changes and lead
changes
Manual Annotation Available to user using Events A to D.
14-72 ECG and ECG/Respiration Module Section
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.