Agilent InfinityLab LC User Manual

Vialsamplers
Agilent InfinityLab LC Series
User Manual
Agilent InfinityLab LC Series Vialsamplers User Manual
Notices
CAUTION
WARNING
Document Information
Document No: SD-29000242 Rev. E Part No: G7129-90010 Rev. E EDITION 04/2020
Copyright
© Agilent Technologies, Inc. 2015-2020
No part of this manual may be repro­duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written con­sent from Agilent Technologies, Inc. as governed by United States and interna­tional copyright laws.
Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future edi­tions. Further, to the maximum extent permitted by applicable law, Agilent dis­claims all warranties, either express or implied, with regard to this manual and any information contained herein, includ­ing but not limited to the implied warran­ties of merchantability and fitness for a particular purpose. Agilent shall not be lia­ble for errors or for incidental or conse­quential damages in connection with the furnishing, use, or performance of this document or of any information con­tained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Safety Notices
A
CAUTION
hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARN­ING notice until the indicated conditions are fully understood and met.
notice denotes a
CAUTION
notice until
Restricted Rights Legend
U.S. Government Restricted Rights. Soft­ware and technical data rights granted to the federal government include only those rights customarily provided to end user customers. Agilent provides this customary commercial license in Soft­ware and technical data pursuant to FAR
12.211 (Technical Data) and 12.212 (Computer Software) and, for the Depart­ment of Defense, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commer­cial Computer Software or Computer Software Documentation).
Agilent InfinityLab LC Series Vialsamplers User Manual

In This Book:

This manual covers the following modules:
Agilent 1260 Infinity II Vialsampler (G7129A)
Agilent 1290 Infinity II Vialsampler (G7129B)
Agilent 1260 Infinity II Vialsampler (G7129C)
1Introduction
This chapter gives an introduction to the module and instrument overview.
2 Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and performance specifications.
3 Using the Module
This chapter provides information on how to use the module.
4 Preparing the Module
This chapter explains the operational parameters of the module.
5 Optimizing Performance
This chapter provides information on how to optimize the module.
6Troubleshooting and Diagnostics
This chapter gives an overview of the maintenance, troubleshooting, and diagnostic features available for the Agilent InfinityLab Series Vialsampler.
7Error Information
This chapter describes the meaning of error messages, and provides information on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series Vialsamplers User Manual 3
8 Maintenance
This chapter describes the maintenance of the module.
9 Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.
10 Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity Series modules.
11 Hardware Information
This chapter describes the detector in more detail on hardware and electronics.
12 LAN Configuration
This chapter provides information on connecting the module to the Agilent ChemStation PC.
13 Appendix
This chapter provides additional information on safety, legal and web.
Agilent InfinityLab LC Series Vialsamplers User Manual 4
Contents
1Introduction 9
Product Description (G7129A) 10 Features (G7129A) 11 Product Description (G7129B) 12 Features (G7129B) 13 Product Description (G7129C) 14 Features (G7129C) 15 Overview of the Module 16 Operating Principle 18 Leak and Waste Handling 27
2 Site Requirements and Specifications 33
Site Requirements 34 Physical Specifications 38 Performance Specifications 39 Specifications of the Sample Cooler 47 Specifications of the Sample Thermostat 49 Specifications of the Integrated Column Compartment 51
3 Using the Module 53
Magnets 55 Turn on/off 56 Status Indicators 58 Vial Drawers and Trays 59 Choice of Vials and Caps 70 Install the Optional Integrated Column Compartment 74 Using the Optional Integrated Column Compartment 96 Install the Optional Sample Cooler/Sample Thermostat 116 Using the Optional Sample Cooler/Sample Thermostat 123 Transporting the Sampler 132 Agilent Local Control Modules 137
Agilent InfinityLab LC Series Vialsamplers User Manual 5
4 Preparing the Module 139
Leak and Waste Handling 140 Preparing the Module 141 Solvent Information 142 Capillary Color Coding Guide 149 Installing Capillaries 150 Flow Connections to the Vialsampler 152 Setting up the Vialsampler 154
5 Optimizing Performance 166
Optimization for Lowest Carryover 167 Fast Injection Cycle and Low Delay Volume 171 Precise Injection Volume 173 Choice of Rotor Seal 175
6 Troubleshooting and Diagnostics 176
Status Indicators 177 Overview of Tests and Tools 178 Maintenance and Troubleshooting Tools 179 Diagnostic Tests 187 Agilent Lab Advisor Software 189
7Error Information 190
What are Error Messages 192 General Error Messages 193 Vialsampler Error Messages 201 Sample Cooler/Sample Thermostat Error Messages 213 Integrated Column Compartment (ICC) Heater Error Messages 219
Agilent InfinityLab LC Series Vialsamplers User Manual 6
8 Maintenance 222
Introduction to Maintenance 223 Warnings and Cautions 224 Overview of Maintenance 226 Cleaning the Module 227 Remove and Install Doors 228 Exchange the Needle Assembly 229 Exchange the Needle Seat Assembly 235 Exchange the Sample Loop Assembly 241 Exchange the Rotor Seal 245 Exchange the Metering Seal and Piston 250 Replace the Analytical Heads 255 Exchange the Gripper Arm 258 Replace the Peristaltic Pump Cartridge 261 Exchange the Wash Port Assembly 263 Replace the Module Firmware 269 Replace the Sample Cooler/Sample Thermostat 270
9 Parts and Materials for Maintenance 275
Standard Parts For Maintenance 276 Accessory Kits 279 Vial Drawers and Trays 283 Multidraw Kits 286 Sample Thermostat Upgrade 287 Integrated Column Compartment 288 Column ID Upgrade Kit 289 Parts for 900 µL Injection Upgrade 290 Cabinet and Door Kits 291 Analytical Head Assembly (100 µL) 295 Analytical-Head Assembly (900 µL) 296 Analytical Head Assembly (40 μL) 297 2 Position/6 Port Injection Valve, 600 bar 298 2 Position/6 Port Injection Valve, 800 bar 299 2 Position/6 Port Injection Valve, 1300 bar 300
Agilent InfinityLab LC Series Vialsamplers User Manual 7
10 Identifying Cables 301
Cable Overview 302 Analog Cables 304 Remote Cables 306 CAN/LAN Cables 310 RS-232 Cables 311 USB 312
11 Hardware Information 313
Firmware Description 314 Electrical Connections 317 Interfaces 319 Setting the 6-bit Configuration Switch 326 Instrument Layout 330 Early Maintenance Feedback (EMF) 331
12 LAN Configuration 332
What You Have to Do First 333 TCP/IP parameter configuration 334 Configuration Switches 335 Initialization Mode Selection 336 Dynamic Host Configuration Protocol (DHCP) 338 Manual Configuration 341 PC and Agilent ChemStation Setup 345
13 Appendix 354
General Safety Information 355 Waste Electrical and Electronic Equipment (WEEE) Directive 361 Refrigerant 362 Radio Interference 365 Sound Emission 366 Solvent Information 367 Agilent Technologies on Internet 368
Agilent InfinityLab LC Series Vialsamplers User Manual 8

1Introduction

Product Description (G7129A) 10 Features (G7129A) 11 Product Description (G7129B) 12 Features (G7129B) 13 Product Description (G7129C) 14 Features (G7129C) 15 Overview of the Module 16 Operating Principle 18
Sampling Sequence 18 Needle Parkstation 23 Hydraulic Box 24 Transport Assembly 26
Leak and Waste Handling 27
Leak Sensor 31 Waste Guidance 31 Waste Concept 32
This chapter gives an introduction to the module and instrument overview.
Agilent InfinityLab LC Series Vialsamplers User Manual 9
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch

Product Description (G7129A)

Product Description (G7129A)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 4
°C to 40 °C, all within one module.
°C as well as a sample thermostat for stable temperatures from
Agilent InfinityLab LC Series Vialsamplers User Manual 10
Figure 1 Overview of the Vialsampler
1Introduction

Features (G7129A)

Features (G7129A)
High capacity – up to 132 vials (2 mL) or up to 36 vials (6 mL).
Reliable injections – 0.1 to 100 µL injections for up to 600 bar (G7129A) and
up to 800
Easy volume extensions – for injection volumes up to 1800 µL for applications ranging from microbore to semipreparative chromatography.
•Lowest carryover – with an in needle flush port included, for rinsing the outside of the needle.
Efficient temperature control – with an integrated column compartment as option or upgrade available. Which holds two columns up to 30 and provides heating capacity from 5 reproducible chromatography data at optimized resolution.
Integrated sample thermostat – available as option or upgrade, providing cooling and heating in the range from 4
Low internal volume – for minimum contribution to a system's total internal volume, which can be even further reduced using "bypass" mode.
•Increased productivity – with overlapped injections.
Customizable injection program – available for customizing advanced
injections as well as for sample preparation steps upfront injection.
bar (G7129C).
cm length,
°C above ambient up to 80 °C for
°C - 40 °C.
Agilent InfinityLab LC Series Vialsamplers User Manual 11
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch

Product Description (G7129B)

Product Description (G7129B)
The Agilent 1290 Infinity II Vialsampler (G7129B) is an autosampler designed for UHPLC applications up to 1300 ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 as well as a sample thermostat for stable temperatures from 4 within one module.
bar. It provides the reliability, safety, and
°C
°C to 40 °C, all
Agilent InfinityLab LC Series Vialsamplers User Manual 12
Figure 2 Overview of the Vialsampler
1Introduction

Features (G7129B)

Features (G7129B)
Accurate and precise injections - within a wide and flexible range of volumes
High capacity - up to 132 vials (2 mL) or up to 36 vials (6 mL).
Easy volume extensions - for injection volumes up to 1500 µL for applications
ranging from microbore to semipreparative chromatography.
•Lowest carryover - with an in needle flush port included, for rinsing the outside of the needle.
Efficient temperature control - with an integrated column compartment as option or upgrade available. Which holds two columns up to 30 and provides heating capacity from 5 reproducible chromatography data at optimized resolution.
Integrated sample thermostat - available as option or upgrade, providing cooling and heating in the range from 4
Low internal volume - for minimum contribution to a system's total internal volume, which can be even further reduced using bypass mode.
•Increased productivity - with overlapped injections.
Customizable Injection program - available for customizing advanced
injections as well as for sample preparation steps upfront injection.
°C above ambient up to 80 °C for
– 40 °C.
cm length,
Agilent InfinityLab LC Series Vialsamplers User Manual 13
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch

Product Description (G7129C)

Product Description (G7129C)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 4
°C to 40 °C, all within one module.
°C as well as a sample thermostat for stable temperatures from
Agilent InfinityLab LC Series Vialsamplers User Manual 14
Figure 3 Overview of the Vialsampler
1Introduction

Features (G7129C)

Features (G7129C)
High capacity – up to 132 vials (2 mL) or up to 36 vials (6 mL).
Reliable injections – 0.1 to 100 µL injections for up to 600 bar (G7129A) and
up to 800
Easy volume extensions – for injection volumes up to 1800 µL for applications ranging from microbore to semipreparative chromatography.
•Lowest carryover – with an in needle flush port included, for rinsing the outside of the needle.
Efficient temperature control – with an integrated column compartment as option or upgrade available. Which holds two columns up to 30 and provides heating capacity from 5 reproducible chromatography data at optimized resolution.
Integrated sample thermostat – available as option or upgrade, providing cooling and heating in the range from 4
Low internal volume – for minimum contribution to a system's total internal volume, which can be even further reduced using "bypass" mode.
•Increased productivity – with overlapped injections.
Customizable injection program – available for customizing advanced
injections as well as for sample preparation steps upfront injection.
bar (G7129C).
cm length,
°C above ambient up to 80 °C for
°C - 40 °C.
Agilent InfinityLab LC Series Vialsamplers User Manual 15
1Introduction

Overview of the Module

Overview of the Module
The Agilent InfinityLab LC Series Vialsampler is designed for use with other modules of the Agilent InfinityLab LC Series, 1200 Series, and 1100 Series LC, as well as with other LC systems with adequate remote-control capabilities. The Vialsampler can be controlled from any computer with a suitable chromatography data system (for example, Agilent OpenLab CDS, MassHunter, etc.) or via an Agilent local control module (InfinityLab Companion or 1200 Infinity Series Instant Pilot).
The 1290 Infinity II Vialsampler (G7129B) can be operated at up to 1300 bar. The two 1260 Infinity II Vialsampler variants (G7129A and G7129C) are limited to a maximum system pressure of 600 and 800
The Vialsampler can accommodate two vial drawers, providing space for up to 132 samples. The two cartesian vial drawers allow for the use of either 66 × 2 or 18 × 6 system is also available, each capable of hosting 50 × 2 drawers are designed to ease method transfer from older Agilent samplers. The Vialsampler can also be equipped with an external tray, which can be beneficial for automation. The Agilent WalkUp solution fully supports the external tray.
mL vials. A pair of classic vial drawers with a numerical vial assignment
bar, respectively.
mL vials. The classic vial
mL
The Vialsampler features a robotic gripper arm for the transport of vials, driven by four stepper motors to ensure the highest precision and flexibility for the transport movement.
The default configuration of the analytical head and sample loop allows an injection range from 0.1
0.1
– 100 μL for the 1260 Infinity II Vialsamplers. An analytical head with extended volume is also available for injection volumes from 0.1 can be operated at up to 400 volume can even reach 1800
The 2-position/6-port injection valve is driven by a high-speed, hybrid stepper motor, ensuring smooth switching between the mainpass and bypass positions. Thanks to the flow-through design of the Vialsampler, the eluent continuously rinses the component parts of the sample introduction system. This, together with the automated needle wash function for rinsing the outer surface of the needle, ensures the lowest level of carryover for each analysis. For advanced delay volume reduction, the injection valve can be programmed to automatically switch back to the bypass position when the injection is completed.
Agilent InfinityLab LC Series Vialsamplers User Manual 16
– 20 μL for the 1290 Infinity II Vialsampler and
– 900 μL, and bar. With the Multidraw Kit, the maximum injection μL.
1Introduction
Overview of the Module
The Agilent InfinityLab Integrated Column Compartment, an optional upgrade for the Vialsampler, eliminates the need for having a standalone column thermostat in the LC stack. It has the heating capability of reaching temperatures up to 80 and can hold a total of two columns with maximum lengths of 30
For applications requiring control over the vial temperature, the Vialsampler can be equipped with the Agilent InfinityLab Sample Thermostat (G7167-60101). It features a vapor-compression refrigeration system and an electric heater, allowing the Vialsampler to reach vial temperatures down to 4
cm.
°C and up to 40 °C.
°C
Agilent InfinityLab LC Series Vialsamplers User Manual 17
1Introduction

Operating Principle

Operating Principle

Sampling Sequence

The Vialsampler processor continuously monitors the movements of the vialsampler components during the sampling sequence. The processor defines specific time windows and mechanical ranges for each movement. If a specific step of the sampling sequence can’t be completed successfully, an error message is generated.
During the sampling sequence, the solvent bypasses the vialsampler via the injection valve. The gripper arm selects the sample vial, either from a static sample rack, or from external vial positions. The gripper arm places the sample vial below the injection needle. The required volume of sample is drawn into the sample loop by the metering device. Sample is applied to the column when the injection valve returns to the mainpass (main path) position at the end of the sampling sequence.
The sampling sequence occurs in the following order:
1 The injection valve switches to the bypass position. 2 The piston of the metering device moves to the initialization position. 3 The gripper arm moves from the home position, and selects the vial. At the
same time, the needle lifts out of the seat.
4 The gripper moves into the needle station and stops in the draw position. 5 The needle lowers into the vial. 6 The metering device draws the defined sample volume. 7 The needle lifts out of the vial.
1
8 The gripper arm moves out slightly and stops in the wash position1. 9 The needle moves downwards and dips into the wash well of the wash port.
Simultaneously the peristaltic pump delivers the flush solvent.
10 The needle moves back. 11 The gripper arm moves out of the needle station and the wash port snaps
back in position.
1
only if automated needle wash is selected. If this feature is disabled, the gripper arm positions the sample vial directly below the needle (Step 4) and lowers the needle into the vial.
Agilent InfinityLab LC Series Vialsamplers User Manual 18
1
1
1Introduction
Vial
Wash port
Needle seat
Sampling loop
Solvent
Wash
pump
Metering device
Operating Principle
12 The gripper arm replaces the vial, and returns to the home position.
Simultaneously, the needle lowers into the seat.
13 The injection valve switches to the mainpass (main path) position.
Injection Sequence
Before the start of the injection sequence, and during an analysis, the injection valve is in the mainpass (main path) position (see position, the mobile phase flows through the autosamplers metering device, sample loop, and needle. This ensures that all parts in contact with sample are flushed during the run, thus minimizing carryover.
Figure 4 on page 19). In this
Figure 4 Mainpass (main path) Position - standard position during runs and when the sampler
idle
is
Agilent InfinityLab LC Series Vialsamplers User Manual 19
1Introduction
Vial
Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Operating Principle
When the sample sequence begins, the valve unit switches to the bypass position (see
Figure 5 on page 20). Solvent from the pump enters the valve unit at
port 1, and flows directly to the column through port 6.
Figure 5Valve in bypass - needle in vial, metering device aspirates sample volume
Agilent InfinityLab LC Series Vialsamplers User Manual 20
1Introduction
Vial
Wash
port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Operating Principle
Then the vial is positioned below the needle. The needle moves down into the vial, the metering unit draws the required sample volume into the loop, and the needle is raised. In the next step, the needle is washed (see
Figure 6 on page 21).
Figure 6 Outer face of needle getting washed in wash port
Agilent InfinityLab LC Series Vialsamplers User Manual 21
1Introduction
Vial
Wash port
Needle seat
Sampling loop
Solvent
Wash pump
Metering device
Operating Principle
When the metering unit has drawn the required volume of sample into the sample loop, the vial is replaced in the sample tray. The wash port flips into the origin position, the needle is lowered into the needle seat, and the injection valve switches back to mainpass (main path) position, flushing the sample onto the column (see
Figure 7 on page 22).
Figure 7 Valve switches to mainpass (main path) - sample is transferred towards the LC column
Agilent InfinityLab LC Series Vialsamplers User Manual 22
1Introduction
Operating Principle

Needle Parkstation

The needle parkstation comprises two main assemblies: needle drive and wash port.
Figure 8Needle Station
Needle-Drive
The needle movement is driven by a stepper motor connected to the spindle assembly by a toothed belt. The circular motion of the motor is converted to linear motion by the drive nut on the spindle assembly. The upper and lower needle positions are detected by reflection sensors on the needle station board, while the needle-in-vial position is determined by counting the motor steps from the upper needle-sensor position.
Agilent InfinityLab LC Series Vialsamplers User Manual 23
1Introduction
NOTE
Operating Principle

Hydraulic Box

The hydraulic box comprises two main assemblies: metering device, and injection valve.
Figure 9 Hydraulic Unit
The replacement hydraulic box excludes the injection valve and metering head assemblies.
Agilent InfinityLab LC Series Vialsamplers User Manual 24
1Introduction
Operating Principle
Analytical Head
The analytical head is driven by the stepper motor that is connected to the drive shaft by a toothed belt. The drive nut on the spindle converts the circular movement of the spindle to linear motion. The drive nut pushes the sapphire piston against the tension of the spring into the analytical head. The base of the piston sits on the large bearing of the drive nut, which ensures the piston is always centered. A ceramic ring guides the movement of the piston in the analytical head. The home position of the piston is sensed by an optical sensor on the hydraulic unit board while the sample volume is determined by counting the number of steps from the home position. The backward movement of the piston (driven by the spring) draws sample from the vial.
Injection Valve
The two-position 6-port injection valve is driven by a stepper motor. Only five of the six ports are used (port 3 is not used). A lever/slider mechanism transfers the movement of the stepper motor to the injection valve. Two microswitches monitor switching of the valve (bypass and mainpass (main path) end positions). No valve adjustments are required after replacing internal components.
Agilent InfinityLab LC Series Vialsamplers User Manual 25
1Introduction
X motor
Theta motor
Gripper motor
Gripper arm
Z motor
Flex board
Theta axis
X axis
Z axis
Operating Principle

Transport Assembly

The transport unit comprises an X-axis slide (left-right motion), a Z-axis arm (up-down motion), and a gripper assembly (rotation and vial-gripping).
Figure 10 Transport Assembly
The transport assembly uses four stepper motors driven in closed-loop mode for accurate positioning of the gripper assembly. The rotational movement of the motors is converted to linear motion (X- and Z-axes) by toothed belts connected to the drive spindles. The rotation (theta axes) of the gripper assembly is transferred from the motor by a toothed belt and series of gears. The opening and closing of the gripper fingers are driven by a stepper motor linked by a toothed belt to the planetary gearing inside the gripper assembly.
The stepper motor positions are determined by the optical encoders mounted onto the stepper-motor housing. The encoders monitor the position of the motors continually, and correct for position errors automatically (e.g. if the gripper is accidentally moved out of position when loading vials into the vial tray). The initialization positions of the moving components are sensed by reflection sensors mounted on the flex board. These positions are used by the processor to calculate the actual motor position. An additional six reflection sensors for tray recognition are mounted on the flex board at the front of the assembly.
Agilent InfinityLab LC Series Vialsamplers User Manual 26
1Introduction

Leak and Waste Handling

Leak and Waste Handling
The Agilent InfinityLab LC Series has been designed for safe leak and waste handling. It is important that all security concepts are understood and instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The maximum volume for an individual bottle stored in the solvent cabinet should not exceed 2 Cabinets (a printed copy of the guideline has been shipped with the solvent cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and Infinity II modules can be stacked on top of each other. Waste tubes are guided through a channel on the right hand side of the instrument, keeping the front access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks, guiding them to the leak sensor for leak detection, and passing them on to the next module below, if the leak sensor fails. The leak sensor in the leak plane stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste container:
from the detector's flow cell outlet
from the Multisampler needle wash port
from the Sample Cooler or Sample Thermostat (condensate)
from the pump's Seal Wash Sensor (if applicable)
from the pump's Purge Valve or Multipurpose Valve
L. For details, see the usage guideline for the Agilent Infinity II Solvent
Agilent InfinityLab LC Series Vialsamplers User Manual 27
1Introduction
Leak and Waste Handling
Figure 11 Infinity II Leak Waste Concept (Flex Bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual 28
1Introduction
Leak and Waste Handling
Figure 12 Infinity II Single Stack Leak Waste Concept (bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual 29
1Introduction
Leak and Waste Handling
Figure 13 Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak plane outlet on each of the bottom instruments guides the solvent to a suitable waste container.
Agilent InfinityLab LC Series Vialsamplers User Manual 30
1Introduction
CAUTION
NOTE
NOTE
Leak and Waste Handling

Leak Sensor

Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
The leak sensor in the sampler is hidden under the ICC Column Heater or Column Shelf respectively.

Waste Guidance

Agilent InfinityLab LC Series Vialsamplers User Manual 31
The waste drainage must go straight into the waste containers. The waste flow must not be restricted at bends or joints.
1Introduction
Leak and Waste Handling

Waste Concept

1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use your own waste solution, make sure that the tubes don't immerse in the liquid.
Agilent InfinityLab LC Series Vialsamplers User Manual 32

2 Site Requirements and Specifications

Site Requirements 34 Physical Specifications 38 Performance Specifications 39
Performance Specifications (G7129A) 39 Performance Specifications (G7129B) 42 Performance Specifications (G7129C) 45
Specifications of the Sample Cooler 47 Specifications of the Sample Thermostat 49 Specifications of the Integrated Column Compartment 51
This chapter provides information on environmental requirements, physical and performance specifications.
Agilent InfinityLab LC Series Vialsamplers User Manual 33
2 Site Requirements and Specifications
WARNING
WARNING
WARNING

Site Requirements

Site Requirements

Site Requirements
A suitable environment is important to ensure optimum performance of the instrument.

Power Considerations

The module power supply has wide ranging capability. It accepts any line voltage in the range described in selector in the rear of the module. There are also no externally accessible fuses, because automatic electronic fuses are implemented in the power supply.
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
Connect your instrument to the specified line voltage only.
Table 1 on page 38. Consequently there is no voltage
Agilent InfinityLab LC Series Vialsamplers User Manual 34
Electrical shock hazard
The module is partially energized when switched off, as long as the power cord is plugged in.
The cover protects users from personal injuries, for example electrical shock.
Do not open the cover.
Do not operate the instrument and disconnect the power cable in case the cover has any signs of damage.
Contact Agilent for support and request an instrument repair service.
Inaccessible power plug.
In case of emergency it must be possible to disconnect the instrument from the power line at any time.
Make sure the power connector of the instrument can be easily reached and unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the cable.
2 Site Requirements and Specifications
WARNING
WARNING
WARNING
Site Requirements

Power Cords

Country-specific power cords are available for the module. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is suitable for your particular country or region.
Unintended use of power cords
Using power cords for unintended purposes can lead to personal injury or damage of electronic equipment.
Never use a power cord other than the one that Agilent shipped with this instrument.
Never use the power cords that Agilent Technologies supplies with this instrument for any other equipment.
Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.
Agilent InfinityLab LC Series Vialsamplers User Manual 35
Absence of ground connection
The absence of ground connection can lead to electric shock or short circuit.
Never operate your instrumentation from a power outlet that has no ground connection.
Electrical shock hazard
Solvents may damage electrical cables.
Prevent electrical cables from getting in contact with solvents.
Exchange electrical cables after contact with solvents.
2 Site Requirements and Specifications
WARNING
NOTE
Site Requirements

Room Size and Ventilation

Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat and laboratory.
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m3 (1 m3 for every 8 g of R600a refrigerant inside of the Sample Thermostat).
Ensure adequate ventilation: typical air exchange of 25 m3/h per m2 of laboratory floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block the openings on the circumference of the Sample Thermostat.

Bench Space

Agilent InfinityLab LC Series Vialsamplers User Manual 36
The module dimensions and weight (see Table 1 on page 38) allow you to place the module on almost any desk or laboratory bench. It needs an additional
2.5
cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is designed to bear the weight of all modules.
The module should be operated in a horizontal position, especially if a Sample Cooler or Sample Thermostat is installed. Use a bubble level to check the leveling of the sampler.
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another laboratory.
2 Site Requirements and Specifications
WARNING
CAUTION
Site Requirements
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.

Condensation

Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 37
If your module was shipped in cold weather, leave it in its box and allow it to
warm slowly to room temperature to avoid condensation.
2 Site Requirements and Specifications

Physical Specifications

Physical Specifications
Tab le 1 Physical Specifications G7129A, G7129B, G7129C
Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample thermostat
Dimensions (height × width × depth)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA, 350 W, 1195 BTU/h
Ambient operating tempera­ture
320 x 396 x 468 mm (12.8 x 15.6 x 18.4 inches)
4 - 40 °C (39 - 104 °F); without sample ther­mostat up to 55 °C (131 °F)
Ambient non-operating tem­perature
Humidity
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point ≥56 °C (133 °F).
1
If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab conditions to stay beyond dew point values for non–condensing operation.
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Installation category II, Pollution degree 2 For indoor use only.
Auto-ignition temperature ≥200 °C (394
°F).
1
Non-condensing
Agilent InfinityLab LC Series Vialsamplers User Manual 38
2 Site Requirements and Specifications

Performance Specifications

Performance Specifications

Performance Specifications (G7129A)

Tab le 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Injection range 0.1 – 100 µL in 0.1 µL incre-
ments with 100 µL up to 60
MPa
0.1 – 900 µL in 0.1 µL incre­ments with 900 µL up to 40
MPa
Injection precision <0.25 % RSD of peak areas from
5
µL to 100 µL
Pressure range 0 – 60 MPa (0 – 600 bar,
– 8702 psi)
0 0 – 40 MPa (0 – 400 bar, 0
– 5801 psi)
Sample viscosity range
Sample capacity 132 x 2 mL vial (two trays
0.2 – 5.0 cP
default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
Up to 1800 µL with multiple draw (hardware modification required) Requires 900 µL analytical head
Measured caffeine
for 900 µL Analytical Head
Agilent InfinityLab LC Series Vialsamplers User Manual 39
2 Site Requirements and Specifications
Performance Specifications
Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Carryover <0.004 % (40 ppm) with needle
wash
Injection cycle time 18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min Injection volume: 1 µL
Minimum sample vol­ume
1 µL from 5 µL sample in 100 µL microvial, or 1 sample in 300
µL from 10 µL µL microvial.
Needle height offset has to be adapted to ensure that needle doesn’t touch vial bottom. Default needle height = 0 equates to 2 bottom.
mm above the vial
Using the following conditions:
Column: Agilent Pursuit XRs 3 C18, 2.0 x 50
Mobile Phase:
A: 0.1 % TFA in water
B: 0.1 % TFA in acetoni-
trile
Isocratic : %B = 40 %
Flow rate: 0.5 mL/min
Temperature: Ambient
Wavelength: 257/4 nm
Injection volume: 1 µL
Sample: 1200 ng/µL Chlor-
hexidine for UV (dissolved with mobile phase A), 1 injected and measured with Agilent UV detector
Wash solution: H2O with
0.1
% TFA (3 s)
mm
µL
Instrument Control LC & CE Drivers A.02.12 or
above Instrument Control Framework (ICF) A.02.03 or above Instant Pilot (G4208A) with firm­ware B.02.19 or above Lab Advisor B.02.07 or above
Communication Controller Area Network
(CAN),Local Area Network (LAN) ERI: ready, start, stop and shut-down signals
For details about supported software versions refer to the compatibility matrix of your ver sion of the LC & CE Drivers
-
Agilent InfinityLab LC Series Vialsamplers User Manual 40
2 Site Requirements and Specifications
Performance Specifications
Table 2 Performance Specifications G7129A
Type Specification Comments Method/Conditions
Maintenance and safety-related features
GLP features Early maintenance feedback
Housing All materials recyclable.
Extensive diagnostics, error detection and display with Agi lent Lab Advisor software Leak detection, safe leak han­dling, leak output signal for shutdown of pumping system, and low voltages in major main tenance areas
(EMF) for continuous tracking of instrument usage with user-settable limits and feed back messages. Electronic records of maintenance and errors.
-
-
-
Agilent InfinityLab LC Series Vialsamplers User Manual 41
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications (G7129B)

Tab le 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Injection range 0.1 – 20 µL in 0.1 µL increments
(default)
0.1 – 40 µL in 0.1 µL increments µL loop is installed)
(if 40
0.1 – 120 µL in 0.1 µL incre-
ments with 1290 Infinity large volume injection kit (hardware modification required)
0.1 – 100 µL in 0.1 µL (if 100 µL
loop and 100 installed)
Injection precision <0.25 % RSD of peak areas from
µL to 100 µL
5
Pressure range Up to 130 MPa (1300 bar,
18854
Sample viscosity range
Sample capacity 132 x 2 mL vial (two trays
0.2 – 5.0 cP
default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
µL head is
psi)
Up to 1500 µL (with 1400 µL multi-draw kit and 100 lytical head)
up to 130 MPa (1300 bar, 18854
psi)
up to 60 MPa (600 bar,
psi)
8702
Measured caffeine
µL ana-
Agilent InfinityLab LC Series Vialsamplers User Manual 42
2 Site Requirements and Specifications
Performance Specifications
Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Carryover <0.004 % (40 ppm) with needle
wash
Injection cycle time 18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min Injection volume: 1 µL
Minimum sample vol­ume
1 µL from 5 µL sample in 100 µL microvial, or 1 sample in 300
µL from 10 µL µL microvial.
Needle height offset has to be adapted to ensure that needle doesn’t touch vial bottom. Default needle height = 0 equates to 2 bottom.
mm above the vial
Using the following conditions:
Column: Agilent Pursuit XRs 3 C18, 2.0 x 50
Mobile Phase:
A: 0.1 % TFA in water
B: 0.1 % TFA in acetoni-
trile
Isocratic : %B = 40 %
Flow rate: 0.5 mL/min
Temperature: Ambient
Wavelength: 257/4 nm
Injection volume: 1 µL
Sample: 1200 ng/µL Chlor-
hexidine for UV (dissolved with mobile phase A), 1 injected and measured with Agilent UV detector
Wash solution: H2O with
0.1
% TFA (3 s)
mm
µL
Instrument control LC & CE Drivers A.02.12 or
above Instrument Control Framework (ICF) A.02.03 or above Instant Pilot (G4208A) with firm­ware B.02.19 or above Lab Advisor B.02.07 or above
Communication Controller Area Network (CAN),
Local Area Network (LAN), ERI: ready, start, stop and shut-down signals
For details about supported software versions refer to the compatibility matrix of your ver sion of the LC & CE Drivers
-
Agilent InfinityLab LC Series Vialsamplers User Manual 43
2 Site Requirements and Specifications
Performance Specifications
Table 3 Performance Specifications G7129B
Type Specification Comments Method/Conditions
Maintenance and safety-related features
GLP features Early maintenance feedback
Housing All materials recyclable.
Extensive diagnostics, error detection and display with Agi lent Lab Advisor software Leak detection, safe leak han­dling, leak output signal for shutdown of pumping system, and low voltages in major main tenance areas
(EMF) for continuous tracking of instrument usage with user-settable limits and feed back messages. Electronic records of maintenance and errors.
-
-
-
Agilent InfinityLab LC Series Vialsamplers User Manual 44
2 Site Requirements and Specifications
Performance Specifications

Performance Specifications (G7129C)

Tab le 4 Performance Specifications G7129C
Type Specification Comments Method/Conditions
Injection range 0.1 – 100 µL in 0.1 µL incre-
ments with 100 µL up to 80
MPa
Injection precision <0.25 % RSD of peak areas from
5
µL to 100 µL
Pressure range 0 – 80 MPa (0 – 800 bar,
– 11603 psi)
0
Sample viscosity range
Sample capacity 132 x 2 mL vial (two trays
Carryover <0.004 % (40 ppm) with needle
0.2 – 5.0 cP
default) 100 x 2 mL vial (two classic trays optional) 36 x 6 mL vials (two trays optional)
wash
Up to 1800 µL with multiple draw (hardware modification required)
Measured caffeine
Using the following conditions:
Column: Agilent Pursuit XRs 3 C18, 2.0 x 50
Mobile Phase:
A: 0.1 % TFA in water
B: 0.1 % TFA in acetoni-
trile
Isocratic : %B = 40 %
Flow rate: 0.5 mL/min
Temperature: Ambient
Wavelength: 257/4 nm
Injection volume: 1 µL
Sample: 1200 ng/µL Chlor-
hexidine for UV (dissolved with mobile phase A), 1 injected and measured with Agilent UV detector
Wash solution: H2O with
0.1
% TFA (3 s)
mm
µL
Injection cycle time 18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min Injection volume: 1 µL
Agilent InfinityLab LC Series Vialsamplers User Manual 45
2 Site Requirements and Specifications
Performance Specifications
Table 4 Performance Specifications G7129C
Type Specification Comments Method/Conditions
Minimum sample vol­ume
Instrument Control LC & CE Drivers A.02.17 or
Communication Controller Area Network (CAN),
Maintenance and safety-related features
1 µL from 5 µL sample in 100 µL microvial, or 1 sample in 300
above Instrument Control Framework (ICF) A.02.05 or above Instant Pilot (G4208A) with firm­ware B.02.22 or above InfinityLab LC Companion (G7108A) Lab Advisor B.02.10 or above
Local Area Network (LAN), ERI: ready, start, stop and shut-down signals
Extensive diagnostics, error detection and display with Agi lent Lab Advisor software Leak detection, safe leak han­dling, leak output signal for shutdown of pumping system, and low voltages in major main tenance areas
µL from 10 µL µL microvial.
Needle height offset has to be adapted to ensure that needle doesn’t touch vial bottom. Default needle height = 0 equates to 2 bottom.
For details about supported software versions refer to the compatibility matrix of your ver sion of the LC & CE Drivers
-
-
mm above the vial
-
GLP features Early maintenance feedback
(EMF) for continuous tracking of instrument usage with user-settable limits and feed back messages. Electronic records of maintenance and errors.
Housing All materials recyclable.
-
Agilent InfinityLab LC Series Vialsamplers User Manual 46
2 Site Requirements and Specifications

Specifications of the Sample Cooler

Specifications of the Sample Cooler
The Agilent Infinity II Sample Cooler is a vapor-compression refrigeration system that uses a fluorinated greenhouse gas (HCF-134a) as the refrigerant. For information on carbon dioxide equivalency (CDE) and global warming potential (GWP), see the instrument label.
Tab le 5 Physical Specification of the Sample Cooler
Type Specification Comment
Weight < 6 kg (< 13.2 lbs)
Dimensions (height × width × depth) 205 x 340 x 370 mm
(8.1 x 13.4 x 14.6 inches)
Refrigerant gas HFC-134a (0.042 kg) Ozone depletion poten-
tial (ODP) = 0
Supply voltage 24 VDC
Current 10 A max.
Ambient operating temperature 4 – 40 ° C (39 – 104 ° F)
Ambient non-operating temperature -40 – 70 ° C (-40 – 158 ° F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
ISM Classification ISM Group 1 Class B According to CISPR 11
Installation category II, Pollution degree 2
For indoor use only
Agilent InfinityLab LC Series Vialsamplers User Manual 47
2 Site Requirements and Specifications
CAUTION
NOTE
Specifications of the Sample Cooler
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat
must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local
regulations.
Please contact your local Agilent Service Center in regard to safe
environmental disposal of the appliance or check www.agilent.com for more info.
Tab le 6 Performance Specifications of the Sample Cooler
Type Specifications
Operating principle High performance, low-energy consumption
micro-compressor based cooler with ozone-friendly HFC-134a coolant (42 g), user-upgradable.
Agilent InfinityLab LC Series Vialsamplers User Manual 48
Temperature range from 4 °C to 5 °C below ambient
Temperature settable from 4 – 40 °C in 1 ° increments
Temperature accuracy (<25 °C, <50 % r.H.) 2 °C to 6 °C at a setpoint of 4 °C
The Agilent Infinity II Sample Cooler is not available for trade sales anymore and has been replaced by the Agilent InfinityLab Sample Thermostat.
2 Site Requirements and Specifications

Specifications of the Sample Thermostat

Specifications of the Sample Thermostat
The Agilent InfinityLab Sample Thermostat is the combination of an electric heater and a vapor-compression refrigeration system. It uses isobutane as a non-Freon refrigerant, which is harmless to the environment and does not affect the ozone layer and global warming but is combustible. Please adhere to the warnings listed in the manual.
Tab le 7 Physical Specifications of the Sample Thermostat
Type Specification Comments
Weight <6 kg (< 13.2 lbs)
Dimensions (height x width x depth)
Refrigerant gas R600a (0.030 kg) Ozone depletion potential (ODP) =0
Supply voltage 24VDC
205 x 340 x 370 mm (8.1 x 13.4 x 14.6 inches)
Global warming potential (GWP) =3
Current 10 A max.
Ambient operating temperature
Ambient non-operat­ing temperature
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN, CSA, UL
ISM Classification ISM Group 1 Class B According to CISPR 11
4 – 40 °C (39 – 104 °F)
-40 – 70 °C (-40 – 158 °F)
Installation category II, Pollution degree 2
For indoor use only
Agilent InfinityLab LC Series Vialsamplers User Manual 49
2 Site Requirements and Specifications
CAUTION
NOTE
Specifications of the Sample Thermostat
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat
must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local
regulations.
Please contact your local Agilent Service Center in regard to safe
environmental disposal of the appliance or check www.agilent.com for more info.
Tab le 8 Performance Specifications for the Sample Thermostat
Type Specifications
Operating principle High performance, low-energy consumption micro-compressor based
cooler with natural R600a coolant (Butane 30
Temperature range from 4 – 40 °C
Temperature settable from 4 – 40 °C in 1 ° increments
g), user-upgradable
Agilent InfinityLab LC Series Vialsamplers User Manual 50
Temperature accuracy (<25 °C, <50 % r.H.)
2 °C to 6 °C at a setpoint of 4 °C
Minimum firmware revision for the Sample Thermostat is D.07.22.
Minimum LC driver revision for the Sample Thermostat is A.02.14.
2 Site Requirements and Specifications

Specifications of the Integrated Column Compartment

Specifications of the Integrated Column Compartment

Physical Specifications Agilent InfinityLab LC Series Integrated Column Compartment (G7130A)

Tab le 9 Physical Specifications (G7130A)
Type Specification Comment
Weight 1.8 kg (4.0 lbs)
Dimensions (height x width x depth)
Supply Voltage 24 V DC
Power consumption 110 W
Ambient operating temperature 4 – 55 °C (39 – 131 °F)
Ambient non-operating tem­perature
Humidity
Operating altitude Up to 3000 m (9842 ft)
Safety standards IEC, EN, CSA, UL
ISM Classification ISM Group 1 Class B According to CISPR 11
1
If a sample cooler is included the upper value for humidity can be reduced. Please check your lab condi­tions to stay beyond dew point values for non–condensing operation.
86.5 x 396.0 x 106.5 mm (3.4 x 15.6 x 4.2 inches)
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Installation category II, Pollution degree 2
1
Maximum outside
Non-condensing
For indoor use only
Agilent InfinityLab LC Series Vialsamplers User Manual 51
2 Site Requirements and Specifications
Specifications of the Integrated Column Compartment

Performance Specifications Agilent InfinityLab LC Series Integrated Column Compartment (G7130A)

Tab le 10 Performance Specifications G7130A
Type Specification Comments
Temperature range 5 °C above ambient to 80 °C
Column capacity 2 columns up to 30 cm and
mm ID
4.6
Temperature stability ±0.10 °C at sensor
Temperature accuracy ±0.8 K (±0.5 K with calibration)
Warm up time 20 – 40 °C in 5 min at sensor
Agilent InfinityLab LC Series Vialsamplers User Manual 52

3Using the Module

Magnets 55 Turn on/off 56 Status Indicators 58 Vial Drawers and Trays 59
List of Drawers and Trays 59 Exchange Drawers 62 Insert a Vial Into the Sampler 65 Install the External Tray 67
Choice of Vials and Caps 70
Compatible Vials and Caps for the 2 mL Vial Drawers 70 Compatible Vials and Caps for the 6 mL Vial Drawer 73
Install the Optional Integrated Column Compartment 74
Unpacking the Unit 74 Install the Integrated Column Compartment 76 Install a Column in the ICC 82 Remove a Column from the ICC 86 Install the Column Shelf 88 Install the Column ID Upgrade Kit 92 Connect a Column Identification Tag to the Tag Reader 94
Using the Optional Integrated Column Compartment 96
Dashboard 96 Control Interface 97 Control 98 Method Parameters 99 Online Signals 101 Column Assignment 102 Column Tag Information Table 103 Using Column Identification Tags 104
Install the Optional Sample Cooler/Sample Thermostat 116
Unpacking the Unit 116 Install the Sample Cooler/Sample Thermostat 117
Using the Optional Sample Cooler/Sample Thermostat 123
Dashboard 123 Control Interface 124 Control 125 Temperature Mode 127 Online Signals 128
Agilent InfinityLab LC Series Vialsamplers User Manual 53
3Using the Module
Specifications of the Integrated Column Compartment
Reporting Sample Temperature 129 Operation Information 130
Transporting the Sampler 132
Prepare a Sampler Without Cooler/Thermostat for Transportation 132 Prepare a Sampler with Cooler/Thermostat for Transportation 133 Install the Transport Protection Foam 135
Agilent Local Control Modules 137
This chapter provides information on how to use the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 54
3Using the Module

Magnets

Magnets
1 Magnets in doors of pumps, autosamplers, detectors, and fraction collectors.
Agilent InfinityLab LC Series Vialsamplers User Manual 55
3Using the Module

Turn on/off

Turn on/off
This procedure exemplarily shows an arbitrary LC stack configuration.
1 2
Power switch: On
Agilent InfinityLab LC Series Vialsamplers User Manual 56
3Using the Module
Turn on/off
3 Turn instrument On/Off with the control software. 4
Power switch: Off
5
Agilent InfinityLab LC Series Vialsamplers User Manual 57
3Using the Module

Status Indicators

Status Indicators
This procedure exemplarily shows an arbitrary LC stack configuration. 1 The module status indicator indicates one of six possible module conditions:
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or completed.
4. Error mode - interrupts the analysis and requires attention (for example a leak or defective internal components).
5. Resident mode (blinking) - for example during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a cold-start. Then try a firmware update.
Agilent InfinityLab LC Series Vialsamplers User Manual 58
3Using the Module
NOTE

Vial Drawers and Trays

Vial Drawers and Trays

List of Drawers and Trays

Supported drawers for the module:
p/n Description
G7129-60010 Drawer for 66 x 2 mL Vials
G7129-60110 Drawer for 18 x 6 mL Vials
G7129-68210 Classic Vial Drawer Kit
G7129-60210 Classic Drawer for 50 x 2 mL Vials, Left
G7129-60220 Classic Drawer for 50 x 2 mL Vials, Right
G7129-60000 External Tray for 5 x 2 mL Vials
Install all drawers for best cooling performance.
Drawer Combinations
Drawers can be installed in any combination enabling both 2 mL- and 6 mL-vials to be used simultaneously. The only exception is the usage of the classical drawer option (100 x 2
mL). This option can't combine with the other drawers.
Numbering of Vial Positions
The standard 2*66 vial drawers have 132 vial positions from P1-A1-P2-F11. However, when using two drawers, the numbering convention is slightly different. The vial positions of the right-hand drawer begin at position P2-A1 as follows:
Left-hand Drawer for 66 x 2 mL Vials: P1-A1 to P1-F11
Left-hand Drawer for 18 x 6 mL Vials: P1-A1 to P1-C6
Right-hand Drawer for 66 x 2 mL Vials: P2-A1 to P2-F11
Right-hand Drawer for 18 x 6 mL Vials: P2-A1 to P2-C6
Drawer for 50 x 2 mL Vials Classic Left: Vial 1-50
Agilent InfinityLab LC Series Vialsamplers User Manual 59
3Using the Module
P1-A1
P1-F11
P2-A1
P2-F11
P1-A1
P1-C6
P2-A1
P2-C6
Vial Drawers and Trays
Drawer for 50 x 2 mL Vials Classic Right: Vial 51-100
External Tray 5-position: 201 – 205 Position
(The disposal tube is installed into the external tray by turning and pushing it into the backside of the hole position, No. 206)
Figure 14 Numbering of drawer position (left-hand Drawer for 66 x 2 mL Vials)
Figure 15 Numbering of drawer position (right-hand Drawer for 66 x 2 mL Vials)
Figure 16 Numbering of drawer position (left-hand Drawer for 18 x 6 mL Vials)
Figure 17 Numbering of drawer position (right-hand Drawer for 18 x 6 mL Vials)
Agilent InfinityLab LC Series Vialsamplers User Manual 60
3Using the Module
1
50
51
100
201
202
203
204
205
206 (Waste Disposal)
Vial Drawers and Trays
Figure 18 Numbering of drawer position (Drawer for 50 x 2 mL Vials Classic)
Agilent InfinityLab LC Series Vialsamplers User Manual 61
Figure 19 Numbering of tray position (External tray)
3Using the Module
NOTE
NOTE
NOTE
Vial Drawers and Trays

Exchange Drawers

Do not operate the sampler without drawers installed.
Install all drawers for best cooling performance.
Do not mix standard and classic drawers.
Install classic drawer 1-50 to the left, classic drawer 51-100 to the right side.
Agilent InfinityLab LC Series Vialsamplers User Manual 62
3Using the Module
Vial Drawers and Trays
1 Open the doors of the module. 2 Remove the drawer.
a Pull the drawer out.
b Lift the front of the drawer.
c Lift the drawer out.
d Remove the drawer.
Agilent InfinityLab LC Series Vialsamplers User Manual 63
3Using the Module
Vial Drawers and Trays
3 Install the drawer.
a Insert the back of the drawer.
b Align the drawer.
c Push in the drawer.
4 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 64
3Using the Module
NOTE
NOTE
Vial Drawers and Trays

Insert a Vial Into the Sampler

Attempting to insert a new vial into the sampler while the gripper arm is moving might lead to aborting the ongoing analysis.
1 Open the doors of the module.
For accessing the left drawer, it is sufficient to open the left door only.
3 Insert the vial into an appropriate position of the drawer. 4 Push the vial drawer back into place.
2 Pull out the vial drawer into which you want to insert the
vial.
Agilent InfinityLab LC Series Vialsamplers User Manual 65
3Using the Module
Vial Drawers and Trays
5 Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual 66
3Using the Module
NOTE
Vial Drawers and Trays

Install the External Tray

Tools required Description
Flathead screwdriver
Parts required p/n Description
G7129-60000 External Tray for 5 x 2 mL Vials G1313-27302 Disposal tube
Preparations
Turn off the sampler
Open the doors
Remove the samples
Keep foam and plastic cover in a safe place.
For best temperature performance, and if the external tray is not in use, it is best to cover the opening for the external tray with the original parts.
Agilent InfinityLab LC Series Vialsamplers User Manual 67
3Using the Module
Vial Drawers and Trays
1 Remove the foam and the plastic cover.
a Lift up the front part of the foam with a flathead
screw driver.
b Remove the foam.
2 Install the external tray and the disposal tube.
a Mount the external tray in the mounting holes on
the left side of the sampler and ensure that it is pushed in all the way.
b Install the disposal tube.
c Push out the plastic cover.
Agilent InfinityLab LC Series Vialsamplers User Manual 68
3Using the Module
Vial Drawers and Trays
3 Close the doors. 4 Configure the External Tray in the online view of the
Chromatographic Data System: right-click on the sam pler dashboard and select Modify> External Tray. In the dialog, select the External Tray installed check box.
-
Agilent InfinityLab LC Series Vialsamplers User Manual 69
3Using the Module

Choice of Vials and Caps

Choice of Vials and Caps

Compatible Vials and Caps for the 2 mL Vial Drawers

List of Compatible Vials and Caps
For reliable operation vials used with the autosampler must not have tapered shoulders or caps that are wider than the body of the vial. The vials in
Vials” on page 70, “Snap Top Vials” on page 71 and “Screw Top Vials” on page 71
and caps in “Crimp Caps” on page 72, “Snap Caps” on page 72 and “Screw
Caps” on page 72 (shown with their Part numbers) have been successfully tested
using a minimum of 15,000 injections with the autosampler.
Crimp Top Vials
p/n Description
“Crimp Top
5181-3375 Crimp Top Vial, 2 mL, clear glass, 100/Pack
5183-4491 Crimp Top Vial, 2 mL, clear glass, 1000/Pack
5182-0543 Crimp Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-4492 Crimp Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-4494 Crimp Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5181-3376 Crimp Top Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-4493 Crimp Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-4495 Crimp Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
5182-0567 Crimp Top Vial, 1 mL, polypropylene, wide opening, 100/Pack
5183-4496 Crimp Top Vial, 1 mL, polypropylene, wide opening, 100/Pack
(silanized)
9301-0978 Crimp top vial, 250 µL, polypropylene, wide opening, 1000/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 70
3Using the Module
Choice of Vials and Caps
Snap Top Vials
p/n Description
5182-0544 Snap Top Vial, 2 mL, clear glass, 100/Pack
5183-4504 Snap Top Vial, 2 mL, clear glass, 1000/Pack
5183-4507 Snap Top Vial, 2 mL, clear glass, 100/Pack (silanized)
5182-0546 Snap Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-4505 Snap Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-4508 Snap Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5182-0545 Snap Top Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-4506 Snap Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-4509 Snap Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
Screw Top Vials
p/n Description
5182-0714 Screw Cap Vials, 2 mL, clear glass, 100/Pack
5183-2067 Screw Top Vial, 2 mL, clear glass, 1000/Pack
5183-2070 Screw Top Vial, 2 mL, clear glass, 100/Pack (silanized)
5182-0715 Screw Top Vial, 2 mL, clear glass, write-on spot, 100/Pack
5183-2068 Screw Top Vial, 2 mL, clear glass, write-on spot, 1000/Pack
5183-2071 Screw Top Vial, 2 mL, clear glass, write-on spot, 100/Pack (silanized)
5182-0716 Screw Cap Vial, 2 mL, amber glass, write-on spot, 100/Pack
5183-2069 Screw Top Vial, 2 mL, amber glass, write-on spot, 1000/Pack
5183-2072 Screw Top Vial, 2 mL, amber glass, write-on spot, 100/Pack (silanized)
Agilent InfinityLab LC Series Vialsamplers User Manual 71
3Using the Module
Choice of Vials and Caps
Crimp Caps
p/n Description
5181-1210 Crimp Cap, silver aluminum, septum (clear PTFE/red rubber), 100/Pack
5183-4498 Crimp Cap, silver aluminum, septum (clear PTFE/red rubber), 1000/Pack
5181-1215 Crimp Cap, blue aluminum, septum (clear PTFE/red rubber), 100/Pack
5181-1216 Crimp Cap, green aluminum, septum (clear PTFE/red rubber), 100/Pack
5181-1217 Crimp Cap, red aluminum, septum (clear PTFE/red rubber), 100/Pack
Snap Caps
p/n Description
5182-0550 Snap Cap, clear polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3458 Snap Cap, blue polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3457 Snap Cap, green polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-3459 Snap Cap, red polypropylene, septum (clear PTFE/red rubber), 100/Pack
Screw Caps
p/n Description
5182-0717 Screw Cap, blue polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0718 Screw Cap, green polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0719 Screw Cap, red polypropylene, septum (clear PTFE/red rubber), 100/Pack
5182-0720 Screw Cap, blue polypropylene, septum (clear PTFE/silicone), 100/Pack
5182-0721 Screw Cap, green polypropylene, septum (clear PTFE/silicone), 100/Pack
5182-0722 Screw Cap, red polypropylene, septum (clear PTFE/silicone), 100/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 72
3Using the Module
NOTE
Choice of Vials and Caps

Compatible Vials and Caps for the 6 mL Vial Drawer

For reliable operation of the sampler, the combined height of the vial and cap should never exceed 40 mm.
Screw Top Vials and Caps
p/n Description
9301-1377 Screw Top Vial, 6 mL, clear glass, flat bottom, 100/Pack
5188-5369 Screw Top Vial, 5 mL, clear glass, high recovery, 100/Pack
9301-1379 Screw Cap, 16 mm, w/o septum, for 6 mL vials, 100/Pack
9301-1378 Septum, PTFE/silicone, for 16 mm caps, 100/Pack
5188-2758 Septum, preslit, PTFE/silicone, for 16 mm caps, 100/Pack
Crimp Top Vials and Caps
p/n Description
9301-1419 Crimp Top Vial, 6 mL, clear glass, flat bottom, 100/Pack
9301-1425 Crimp Cap, silver aluminum, septum (PTFE/silicone), 100/Pack
Agilent InfinityLab LC Series Vialsamplers User Manual 73
3Using the Module
CAUTION
CAUTION

Install the Optional Integrated Column Compartment

Install the Optional Integrated Column Compartment

Unpacking the Unit

Damaged Packaging
Damaged Packaging
If the delivery packaging shows signs of external damage, please call your Agilent Technologies sales and service office immediately. Inform your service representative that the instrument may have been damaged during shipment.
"Defective on arrival" problems
If there are signs of damage, please do not attempt to install the module. Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged.
Agilent InfinityLab LC Series Vialsamplers User Manual 74
Notify your Agilent sales and service office about the damage.
An Agilent service representative will inspect the instrument at your site and
initiate appropriate actions.
Condensation
Condensation within the module
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to
warm slowly to room temperature to avoid condensation.
3Using the Module
Install the Optional Integrated Column Compartment
Delivery Checklist
Ensure all parts and materials have been delivered with your module. The delivery checklist is shown below. For parts identification please check the illustrated parts breakdown in report any missing or damaged parts to your local Agilent Technologies sales and service office.
Tab le 11 Delivery checklist for the Integrated Column Compartment (ICC)
Description Quantity
Integrated Column Compartment (G7130-60030 or G7130-60060) 1
Accessory Kit for ICC 1
“Integrated Column Compartment” on page 288. Please
Agilent InfinityLab LC Series Vialsamplers User Manual 75
3Using the Module
WARNING
Install the Optional Integrated Column Compartment

Install the Integrated Column Compartment

Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required # p/n Description
1 G7130-60030 Integrated Column Compartment, 3 µL Heater
OR 1 G7130-60060 Integrated Column Compartment, 6 µL Heater
1 Accessory Kit for ICC
(includes capillaries and column holder clips)
For more information on recommended capillary types refer to “Capillary
Connections” on page 278.
Preparations
Switch off the pump.
Switch off the sampler.
Remove the column.
Open the doors of the sampler.
Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column Compartment are not installed.
Install either column shelf or Integrated Column Compartment before starting the module.
Agilent InfinityLab LC Series Vialsamplers User Manual 76
3Using the Module
Click
Install the Optional Integrated Column Compartment
1 Remove the column shelf.
a Press in the release buttons on the left and right
side.
b Pull out the column shelf.
2 Install the Integrated Column Compartment.
a Push in the Integrated Column Compartment.
b Ensure that the release buttons are clicked back in
their original position.
The column shelf is removed.
The Integrated Column Compartment is installed.
Agilent InfinityLab LC Series Vialsamplers User Manual 77
3Using the Module
Install the Optional Integrated Column Compartment
3 Push the latch of the Integrated Column Compartment
to the left.
5 Connect the injector valve (port 6) of the Vialsampler to
the inlet port of the ICC heat exchanger with a suitable capillary (see recommendations).
“Capillary Connections” on page 278 for
4 Open the flap door.
6 Connect a suitable capillary to the respective outlet port
of the ICC heat exchanger (see
Connections” on page 278 for recommendations).
“Capillary
Agilent InfinityLab LC Series Vialsamplers User Manual 78
3Using the Module
NOTE
NOTE
max.
2.5 Nm
max.
2.5 Nm
Install the Optional Integrated Column Compartment
7 Tighten the capillary fitting to make the connection
pressure tight.
Do not overtighten the fitting (maximum torque 2.5 Nm).
The use of capillaries with SL or SX fittings are recom­mended for a better handling.
8 Mount the column holders on the heat exchanger.
Agilent InfinityLab LC Series Vialsamplers User Manual 79
3Using the Module
Install the Optional Integrated Column Compartment
9 Connect the column to the free end of the capillary
coming from the ICC heat exchanger. For best perfor mance, use capillaries with the InfinityLab Quick Con­nect Fitting.
a Turn the column onto the Quick Connect Fitting
until the first resistance is observed.
b Close the lever to establish a pressure tight connec-
tion.
10 Connect a suitable capillary to the outlet port of the col-
-
umn. For best performance, use capillaries with the InfinityLab Quick Turn Fitting.
a Connect the capillary with the Quick Connect Fitting
and tighten it until finger-tightness.
The fitting should be pressure tight up to 600 bar.
b If needed, tighten the fitting further with the help of
two suitable wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 80
3Using the Module
NOTE
Click
Install the Optional Integrated Column Compartment
11 Mount the column in the column holder clip(s). 12 Route the capillary coming from the column through the
opening between the flap door and the ICC body towards the detector module.
Alternatively, the capillary can also be routed through the incisions at the right side of the ICC dedicated for this pur pose.
13 Close the flap door until the latch is locked. 14 Close the doors.
-
15 Turn on the sampler and the pump.
Agilent InfinityLab LC Series Vialsamplers User Manual 81
3Using the Module
NOTE
Install the Optional Integrated Column Compartment

Install a Column in the ICC

Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required p/n Description
Column (up to 30 cm length)
G7116-68003 Column Holder Lamella, 2/pk
For more information on recommended capillary types, refer to “Capillary
Connections” on page 278.
Preparations
Switch off the pump
Switch off the sampler
Open the doors of the sampler
A maximum of two columns can be installed.
1 Push the latch of the Integrated Column Compartment
to the left.
2 Open the flap door.
Agilent InfinityLab LC Series Vialsamplers User Manual 82
3Using the Module
NOTE
NOTE
max.
2.5 Nm
max.
2.5 Nm
Install the Optional Integrated Column Compartment
3 Connect the injector valve (port 6) of the Vialsampler to
the inlet port of the ICC heat exchanger with a suitable capillary (see recommendations).
5 Tighten the capillary fitting to make the connection
pressure tight.
Do not overtighten the fitting (maximum torque 2.5 Nm).
The use of capillaries with SL or SX fittings are recom­mended for a better handling.
“Capillary Connections” on page 278 for
4 Connect a suitable capillary to the respective outlet port
of the ICC heat exchanger (see
Connections” on page 278 for recommendations).
6 Mount the column holders on the heat exchanger.
“Capillary
Agilent InfinityLab LC Series Vialsamplers User Manual 83
3Using the Module
Install the Optional Integrated Column Compartment
7 Connect the column to the free end of the capillary
coming from the ICC heat exchanger. For best perfor mance, use capillaries with the InfinityLab Quick Con­nect Fitting.
a Turn the column onto the Quick Connect Fitting
until the first resistance is observed.
b Close the lever to establish a pressure tight connec-
tion.
8 Connect a suitable capillary to the outlet port of the col-
-
umn. For best performance, use capillaries with the InfinityLab Quick Turn Fitting.
a Connect the capillary with the Quick Connect Fitting
and tighten it until finger-tightness.
The fitting should be pressure tight up to 600 bar.
b If needed, tighten the fitting further with the help of
two suitable wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 84
3Using the Module
NOTE
Click
Install the Optional Integrated Column Compartment
9 Mount the column in the column holder clip(s). 10 Route the capillary coming from the column through the
opening between the flap door and the ICC body towards the detector module.
Alternatively, the capillary can also be routed through the incisions at the right side of the ICC dedicated for this pur pose.
11 Close the flap door until the latch is locked. 12 Close the doors.
-
Next Steps:
13 If applicable, connect the column identification tag to the tag reader (see “Connect a Column Identification Tag to the
Tag Reader” on page 94).
14 Configure the connected column in the CDS (see “Column Assignment” on page 102).
Agilent InfinityLab LC Series Vialsamplers User Manual 85
3Using the Module
Install the Optional Integrated Column Compartment

Remove a Column from the ICC

Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Preparations
1 Push the latch of the Integrated Column Compartment
to the left.
3 Remove the column from the column holder clips. 4 Loosen the Quick Turn Fitting either with your fingers or
Switch off the pump
Switch off the sampler
Open the doors of the sampler
2 Open the flap door.
with the help of two wrenches.
Agilent InfinityLab LC Series Vialsamplers User Manual 86
3Using the Module
Click
Install the Optional Integrated Column Compartment
5 Remove the capillary from the outlet port of the column. 6 Lift up the lever to release the tension on the Quick Con-
nect Fitting.
7 Unscrew and remove the Quick Connect Fitting. 8 Remove the column from the ICC.
9 Close the flap door until the latch is locked.
Agilent InfinityLab LC Series Vialsamplers User Manual 87
3Using the Module
WARNING
Install the Optional Integrated Column Compartment

Install the Column Shelf

Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required p/n Description
G7129-60023 Column Shelf
Preparations
Switch off the pump.
Switch off the sampler.
Open the doors of the sampler.
Open the ICC flap door.
Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column Compartment are not installed.
Install either column shelf or Integrated Column Compartment before starting the module.
1 Remove the capillary connecting the injection valve and
the ICC heat exchanger.
2 Remove the column from the column holder clips.
Agilent InfinityLab LC Series Vialsamplers User Manual 88
3Using the Module
Click
Install the Optional Integrated Column Compartment
3 Loosen the Quick Turn Fitting either with your fingers or
with the help of two wrenches.
5 Lift up the lever to release the tension on the Quick Con-
nect Fitting.
4 Remove the capillary from the outlet port of the column.
6 Unscrew and remove the Quick Connect Fitting.
7 Remove the column from the ICC. 8 Close the flap door until the latch is locked.
Agilent InfinityLab LC Series Vialsamplers User Manual 89
3Using the Module
Click
Install the Optional Integrated Column Compartment
9 Remove the Integrated Column Compartment.
a Press the release buttons on the left and right side.
b Pull out the Integrated Column Compartment.
10 Install the column shelf
a Push in the column shelf.
b Ensure that the release buttons are clicked back in
their original position.
The Integrated Column Comparment is removed.
The column shelf is installed.
Agilent InfinityLab LC Series Vialsamplers User Manual 90
3Using the Module
Install the Optional Integrated Column Compartment
11 Close the doors. 12 Turn on the sampler and the pump.
Agilent InfinityLab LC Series Vialsamplers User Manual 91
3Using the Module
NOTE
Install the Optional Integrated Column Compartment

Install the Column ID Upgrade Kit

Parts required # p/n Description
1 G4751A Column ID Upgrade Kit for the Integrated Column Compartment
(includes the tag reader, 2 column identification tags and a tech­nical note)
OR 1 5067-6153 Column ID Tag Reader
1 5067-5917 InfinityLab Column Identification Tag
Preparations
Power off the instrument.
Install the Integrated Column Compartment (see “Install the Integrated Column
Compartment” on page 76).
To use the G4751A Column ID Upgrade Kit, the Agilent Integrated Column Compartment must be installed in the Vialsampler.
1 Place the two mounting clips onto the left side of the
Vialsampler. Make sure of the correct orientation.
3 Connect the cable of the tag reader to the respective
socket on the back of the Vialsampler.
2 Mount the tag reader to the mounting clips.
4 Fix the cable to the back of the sampler with the cable
clips provided with the tag reader.
Agilent InfinityLab LC Series Vialsamplers User Manual 92
3Using the Module
Install the Optional Integrated Column Compartment
Next Steps:
5 Power on the instrument.
6 Place a column with an identification tag in the ICC and plug the tag into the tag reader (see “Connect a Column Identi-
fication Tag to the Tag Reader” on page 94).
Agilent InfinityLab LC Series Vialsamplers User Manual 93
3Using the Module
NOTE
Install the Optional Integrated Column Compartment

Connect a Column Identification Tag to the Tag Reader

Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
Parts required p/n Description
5067-5917 InfinityLab Column Identification Tag
Column (up to 30 cm length)
Preparations
1 If the column is without a preinstalled identification tag,
fit a new tag onto it by slipping the lanyard loop over one end of the column and pulling the cord tight through the plastic holder.
Once the lanyard is fixed, the tag can no longer be removed from the column.
Install the Column ID Tag Reader (see “Install the Column ID Upgrade Kit” on page 92).
Remove the column if installed (see “Remove a Column from the ICC” on page 86).
2 Install the column in the ICC (see “Install a Column in
the ICC” on page 82).
Agilent InfinityLab LC Series Vialsamplers User Manual 94
3Using the Module
Click
Install the Optional Integrated Column Compartment
3 Route the cord to the left and plug the column identifi-
cation tag into the corresponding socket of the tag reader.
5 Close the doors. 6 Configure the connected column with identification tag
4 Close the flap door until the latch is locked.
in the CDS (see
“Using Column Identification Tags” on page 104).
“Column Assignment” on page 102 and
Agilent InfinityLab LC Series Vialsamplers User Manual 95
3Using the Module
1
2
3
4
5
6
7
8

Using the Optional Integrated Column Compartment

Using the Optional Integrated Column Compartment
The following section describes how to operate the Agilent Integrated Column Compartment.

Dashboard

When the Integrated Column Compartment (ICC) is installed in a Vialsampler and configured in the chromatography data system (CDS), a new tile appears on the dashboard, comprising a standalone graphical user interface (GUI) for the ICC.
1 ICC: Turn On/Off
2 ICC: Heater Status (On/Off)
3 ICC: Actual Temperature
4 ICC: Set Temperature
5 ICC: Column Information
6 ICC: Status Indicator
7 ICC: Early Maintenance Feedback (EMF) Status
8 ICC: Column Identification Tag Recognised
Agilent InfinityLab LC Series Vialsamplers User Manual 96
3Using the Module
Using the Optional Integrated Column Compartment

Control Interface

Right-clicking on the sampler GUI opens the control interface, where control and method parameters can be edited, column assignment modified, and special commands executed.
Agilent InfinityLab LC Series Vialsamplers User Manual 97
3Using the Module
Using the Optional Integrated Column Compartment

Control

The Control dialog box enables the user to change some control-specific settings for the ICC:
•Thermostat:
On: The ICC turns on and the system starts to regulate the temperature of
the column oven towards the setpoint.
Off: The ICC turns off.
•At Power On:
Turn on thermostat: The ICC turns on automatically upon powering on the
sampler.
•Automatic Turn On:
Turn on at: The ICC turns on automatically at the specified time and date.
Agilent InfinityLab LC Series Vialsamplers User Manual 98
3Using the Module
NOTE
NOTE
Using the Optional Integrated Column Compartment

Method Parameters

The Method dialog box enables the user to change the method parameters relevant for the ICC:
Temperature:
Not Controlled: The temperature of the column oven stays uncontrolled.
At specified temperature (°C): The temperature of the column oven is
regulated to stay at the setpoint temperature specified.
Valid inputs are 10 to 80 °C in increments of 0.1 °C.
Unchanged: The temperature setting stays unchanged, meaning that the
setting currently stored in the system is used by the method.
•Valve Position/Column: This section provides information about the location of the column that is being used for the given method.
Enforce column for run: Marking this checkbox restricts the execution of
the method to a specific column type, identified by its product number.
•Stoptime:
As Pump/Injector: The ICC uses the same stoptime setting as specified for
the pump or the sampler.
After specified time (min): The stop time defines a time limit for the
analysis, after which all method parameters are reset to initial values.
•Posttime:
Off: No post-run period is applied.
For specified time (min): After the analysis, the system stays in the
post-run state for the specified time. Using a post-time period can be beneficial when the column needs to be re-equilibrated before starting the next analysis (for example, gradient elutions).
For Stoptime and Posttime, valid inputs are 0.01 to 99999 minutes in increments of 0.1 min.
Agilent InfinityLab LC Series Vialsamplers User Manual 99
3Using the Module
Using the Optional Integrated Column Compartment
Enable Analysis:
With any temperature: The analysis starts regardless of the actual column
oven temperature.
When temperature is within: The analysis starts only when the actual
column oven temperature is within the specified range of the setpoint temperature.
•Timetable: This feature enables the user to use programmed changes to control the oven temperature during the curse of the analysis.
Agilent InfinityLab LC Series Vialsamplers User Manual 100
Loading...