No part of this manual may be reproduced in any form or by any means
(including electronic storage and retrieval
or translation into a foreign language)
without prior agreement and written consent from Agilent Technologies, Inc. as
governed by United States and international copyright laws.
The material contained in this document
is provided “as is,” and is subject to being
changed, without notice, in future editions. Further, to the maximum extent
permitted by applicable law, Agilent disclaims all warranties, either express or
implied, with regard to this manual and
any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a
particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the
furnishing, use, or performance of this
document or of any information contained herein. Should Agilent and the user
have a separate written agreement with
warranty terms covering the material in
this document that conflict with these
terms, the warranty terms in the separate
agreement shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
Safety Notices
A
CAUTION
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in damage to the product or loss
of important data. Do not proceed
beyond a
the indicated conditions are fully
understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARNING notice until the indicated
conditions are fully understood
and met.
notice denotes a
CAUTION
notice until
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only
those rights customarily provided to end
user customers. Agilent provides this
customary commercial license in Software and technical data pursuant to FAR
12.211 (Technical Data) and 12.212
(Computer Software) and, for the Department of Defense, DFARS 252.227-7015
(Technical Data - Commercial Items) and
DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer
Software Documentation).
Agilent InfinityLab LC Series Vialsamplers User Manual
In This Book:
This manual covers the following modules:
•Agilent 1260 Infinity II Vialsampler (G7129A)
•Agilent 1290 Infinity II Vialsampler (G7129B)
•Agilent 1260 Infinity II Vialsampler (G7129C)
1Introduction
This chapter gives an introduction to the module and instrument overview.
2Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and
performance specifications.
3Using the Module
This chapter provides information on how to use the module.
4Preparing the Module
This chapter explains the operational parameters of the module.
5Optimizing Performance
This chapter provides information on how to optimize the module.
6Troubleshooting and Diagnostics
This chapter gives an overview of the maintenance, troubleshooting, and
diagnostic features available for the Agilent InfinityLab Series Vialsampler.
7Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
Agilent InfinityLab LC Series Vialsamplers User Manual3
8Maintenance
This chapter describes the maintenance of the module.
9Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.
10Identifying Cables
This chapter provides information on cables used with the Agilent 1200 Infinity
Series modules.
11Hardware Information
This chapter describes the detector in more detail on hardware and electronics.
12LAN Configuration
This chapter provides information on connecting the module to the Agilent
ChemStation PC.
13Appendix
This chapter provides additional information on safety, legal and web.
Agilent InfinityLab LC Series Vialsamplers User Manual4
Contents
1Introduction 9
Product Description (G7129A) 10
Features (G7129A) 11
Product Description (G7129B) 12
Features (G7129B) 13
Product Description (G7129C) 14
Features (G7129C) 15
Overview of the Module 16
Operating Principle 18
Leak and Waste Handling 27
2Site Requirements and Specifications 33
Site Requirements 34
Physical Specifications 38
Performance Specifications 39
Specifications of the Sample Cooler 47
Specifications of the Sample Thermostat 49
Specifications of the Integrated Column Compartment 51
3Using the Module 53
Magnets 55
Turn on/off 56
Status Indicators 58
Vial Drawers and Trays 59
Choice of Vials and Caps 70
Install the Optional Integrated Column Compartment 74
Using the Optional Integrated Column Compartment 96
Install the Optional Sample Cooler/Sample Thermostat 116
Using the Optional Sample Cooler/Sample Thermostat 123
Transporting the Sampler 132
Agilent Local Control Modules 137
Agilent InfinityLab LC Series Vialsamplers User Manual5
4Preparing the Module 139
Leak and Waste Handling 140
Preparing the Module 141
Solvent Information 142
Capillary Color Coding Guide 149
Installing Capillaries 150
Flow Connections to the Vialsampler 152
Setting up the Vialsampler 154
5Optimizing Performance 166
Optimization for Lowest Carryover 167
Fast Injection Cycle and Low Delay Volume 171
Precise Injection Volume 173
Choice of Rotor Seal 175
6Troubleshooting and Diagnostics 176
Status Indicators 177
Overview of Tests and Tools 178
Maintenance and Troubleshooting Tools 179
Diagnostic Tests 187
Agilent Lab Advisor Software 189
7Error Information 190
What are Error Messages 192
General Error Messages 193
Vialsampler Error Messages 201
Sample Cooler/Sample Thermostat Error Messages 213
Integrated Column Compartment (ICC) Heater Error Messages 219
Agilent InfinityLab LC Series Vialsamplers User Manual6
8Maintenance 222
Introduction to Maintenance 223
Warnings and Cautions 224
Overview of Maintenance 226
Cleaning the Module 227
Remove and Install Doors 228
Exchange the Needle Assembly 229
Exchange the Needle Seat Assembly 235
Exchange the Sample Loop Assembly 241
Exchange the Rotor Seal 245
Exchange the Metering Seal and Piston 250
Replace the Analytical Heads 255
Exchange the Gripper Arm 258
Replace the Peristaltic Pump Cartridge 261
Exchange the Wash Port Assembly 263
Replace the Module Firmware 269
Replace the Sample Cooler/Sample Thermostat 270
9Parts and Materials for Maintenance 275
Standard Parts For Maintenance 276
Accessory Kits 279
Vial Drawers and Trays 283
Multidraw Kits 286
Sample Thermostat Upgrade 287
Integrated Column Compartment 288
Column ID Upgrade Kit 289
Parts for 900 µL Injection Upgrade 290
Cabinet and Door Kits 291
Analytical Head Assembly (100 µL) 295
Analytical-Head Assembly (900 µL) 296
Analytical Head Assembly (40 μL) 297
2 Position/6 Port Injection Valve, 600 bar 298
2 Position/6 Port Injection Valve, 800 bar 299
2 Position/6 Port Injection Valve, 1300 bar 300
Agilent InfinityLab LC Series Vialsamplers User Manual7
10Identifying Cables 301
Cable Overview 302
Analog Cables 304
Remote Cables 306
CAN/LAN Cables 310
RS-232 Cables 311
USB 312
What You Have to Do First 333
TCP/IP parameter configuration 334
Configuration Switches 335
Initialization Mode Selection 336
Dynamic Host Configuration Protocol (DHCP) 338
Manual Configuration 341
PC and Agilent ChemStation Setup 345
13Appendix 354
General Safety Information 355
Waste Electrical and Electronic Equipment (WEEE) Directive 361
Refrigerant 362
Radio Interference 365
Sound Emission 366
Solvent Information 367
Agilent Technologies on Internet 368
Agilent InfinityLab LC Series Vialsamplers User Manual8
1Introduction
Product Description (G7129A) 10
Features (G7129A) 11
Product Description (G7129B) 12
Features (G7129B) 13
Product Description (G7129C) 14
Features (G7129C) 15
Overview of the Module 16
Operating Principle 18
This chapter gives an introduction to the module and instrument overview.
Agilent InfinityLab LC Series Vialsamplers User Manual9
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Product Description (G7129A)
Product Description (G7129A)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the
reliability and ease-of-use needed for routine pharmaceutical tasks and quality
control, as well as for environmental and food analyses. It can house optionally
the integrated column compartment for two LC columns with temperature
control up to 80
4
°C to 40 °C, all within one module.
°C as well as a sample thermostat for stable temperatures from
Agilent InfinityLab LC Series Vialsamplers User Manual10
Figure 1Overview of the Vialsampler
1Introduction
Features (G7129A)
Features (G7129A)
•High capacity – up to 132 vials (2 mL) or up to 36 vials (6 mL).
•Reliable injections – 0.1 to 100 µL injections for up to 600 bar (G7129A) and
up to 800
•Easy volume extensions – for injection volumes up to 1800 µL for applications
ranging from microbore to semipreparative chromatography.
•Lowest carryover – with an in needle flush port included, for rinsing the outside
of the needle.
•Efficient temperature control – with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30
and provides heating capacity from 5
reproducible chromatography data at optimized resolution.
•Integrated sample thermostat – available as option or upgrade, providing
cooling and heating in the range from 4
•Low internal volume – for minimum contribution to a system's total internal
volume, which can be even further reduced using "bypass" mode.
•Increased productivity – with overlapped injections.
•Customizable injection program – available for customizing advanced
injections as well as for sample preparation steps upfront injection.
bar (G7129C).
cm length,
°C above ambient up to 80 °C for
°C - 40 °C.
Agilent InfinityLab LC Series Vialsamplers User Manual11
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Product Description (G7129B)
Product Description (G7129B)
The Agilent 1290 Infinity II Vialsampler (G7129B) is an autosampler designed for
UHPLC applications up to 1300
ease-of-use needed for routine pharmaceutical tasks and quality control, as well
as for environmental and food analyses. It can house optionally the integrated
column compartment for two LC columns with temperature control up to 80
as well as a sample thermostat for stable temperatures from 4
within one module.
bar. It provides the reliability, safety, and
°C
°C to 40 °C, all
Agilent InfinityLab LC Series Vialsamplers User Manual12
Figure 2Overview of the Vialsampler
1Introduction
Features (G7129B)
Features (G7129B)
•Accurate and precise injections - within a wide and flexible range of volumes
•High capacity - up to 132 vials (2 mL) or up to 36 vials (6 mL).
•Easy volume extensions - for injection volumes up to 1500 µL for applications
ranging from microbore to semipreparative chromatography.
•Lowest carryover - with an in needle flush port included, for rinsing the outside
of the needle.
•Efficient temperature control - with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30
and provides heating capacity from 5
reproducible chromatography data at optimized resolution.
•Integrated sample thermostat - available as option or upgrade, providing
cooling and heating in the range from 4
•Low internal volume - for minimum contribution to a system's total internal
volume, which can be even further reduced using bypass mode.
•Increased productivity - with overlapped injections.
•Customizable Injection program - available for customizing advanced
injections as well as for sample preparation steps upfront injection.
°C above ambient up to 80 °C for
– 40 °C.
cm length,
Agilent InfinityLab LC Series Vialsamplers User Manual13
1Introduction
Status indicator
Needle assembly
Drawer
Column shelf
Leak drain
Power switch
Product Description (G7129C)
Product Description (G7129C)
The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the
reliability and ease-of-use needed for routine pharmaceutical tasks and quality
control, as well as for environmental and food analyses. It can house optionally
the integrated column compartment for two LC columns with temperature
control up to 80
4
°C to 40 °C, all within one module.
°C as well as a sample thermostat for stable temperatures from
Agilent InfinityLab LC Series Vialsamplers User Manual14
Figure 3Overview of the Vialsampler
1Introduction
Features (G7129C)
Features (G7129C)
•High capacity – up to 132 vials (2 mL) or up to 36 vials (6 mL).
•Reliable injections – 0.1 to 100 µL injections for up to 600 bar (G7129A) and
up to 800
•Easy volume extensions – for injection volumes up to 1800 µL for applications
ranging from microbore to semipreparative chromatography.
•Lowest carryover – with an in needle flush port included, for rinsing the outside
of the needle.
•Efficient temperature control – with an integrated column compartment as
option or upgrade available. Which holds two columns up to 30
and provides heating capacity from 5
reproducible chromatography data at optimized resolution.
•Integrated sample thermostat – available as option or upgrade, providing
cooling and heating in the range from 4
•Low internal volume – for minimum contribution to a system's total internal
volume, which can be even further reduced using "bypass" mode.
•Increased productivity – with overlapped injections.
•Customizable injection program – available for customizing advanced
injections as well as for sample preparation steps upfront injection.
bar (G7129C).
cm length,
°C above ambient up to 80 °C for
°C - 40 °C.
Agilent InfinityLab LC Series Vialsamplers User Manual15
1Introduction
Overview of the Module
Overview of the Module
The Agilent InfinityLab LC Series Vialsampler is designed for use with other
modules of the Agilent InfinityLab LC Series, 1200 Series, and 1100 Series LC, as
well as with other LC systems with adequate remote-control capabilities. The
Vialsampler can be controlled from any computer with a suitable
chromatography data system (for example, Agilent OpenLab CDS, MassHunter,
etc.) or via an Agilent local control module (InfinityLab Companion or 1200
Infinity Series Instant Pilot).
The 1290 Infinity II Vialsampler (G7129B) can be operated at up to 1300 bar. The
two 1260 Infinity II Vialsampler variants (G7129A and G7129C) are limited to a
maximum system pressure of 600 and 800
The Vialsampler can accommodate two vial drawers, providing space for up to
132 samples. The two cartesian vial drawers allow for the use of either 66 × 2
or 18 × 6
system is also available, each capable of hosting 50 × 2
drawers are designed to ease method transfer from older Agilent samplers. The
Vialsampler can also be equipped with an external tray, which can be beneficial
for automation. The Agilent WalkUp solution fully supports the external tray.
mL vials. A pair of classic vial drawers with a numerical vial assignment
bar, respectively.
mL vials. The classic vial
mL
The Vialsampler features a robotic gripper arm for the transport of vials, driven by
four stepper motors to ensure the highest precision and flexibility for the
transport movement.
The default configuration of the analytical head and sample loop allows an
injection range from 0.1
0.1
– 100 μL for the 1260 Infinity II Vialsamplers. An analytical head with
extended volume is also available for injection volumes from 0.1
can be operated at up to 400
volume can even reach 1800
The 2-position/6-port injection valve is driven by a high-speed, hybrid stepper
motor, ensuring smooth switching between the mainpass and bypass positions.
Thanks to the flow-through design of the Vialsampler, the eluent continuously
rinses the component parts of the sample introduction system. This, together
with the automated needle wash function for rinsing the outer surface of the
needle, ensures the lowest level of carryover for each analysis. For advanced
delay volume reduction, the injection valve can be programmed to automatically
switch back to the bypass position when the injection is completed.
Agilent InfinityLab LC Series Vialsamplers User Manual16
– 20 μL for the 1290 Infinity II Vialsampler and
– 900 μL, and
bar. With the Multidraw Kit, the maximum injection
μL.
1Introduction
Overview of the Module
The Agilent InfinityLab Integrated Column Compartment, an optional upgrade for
the Vialsampler, eliminates the need for having a standalone column thermostat
in the LC stack. It has the heating capability of reaching temperatures up to 80
and can hold a total of two columns with maximum lengths of 30
For applications requiring control over the vial temperature, the Vialsampler can
be equipped with the Agilent InfinityLab Sample Thermostat (G7167-60101). It
features a vapor-compression refrigeration system and an electric heater,
allowing the Vialsampler to reach vial temperatures down to 4
cm.
°C and up to 40 °C.
°C
Agilent InfinityLab LC Series Vialsamplers User Manual17
1Introduction
Operating Principle
Operating Principle
Sampling Sequence
The Vialsampler processor continuously monitors the movements of the
vialsampler components during the sampling sequence. The processor defines
specific time windows and mechanical ranges for each movement. If a specific
step of the sampling sequence can’t be completed successfully, an error
message is generated.
During the sampling sequence, the solvent bypasses the vialsampler via the
injection valve. The gripper arm selects the sample vial, either from a static
sample rack, or from external vial positions. The gripper arm places the sample
vial below the injection needle. The required volume of sample is drawn into the
sample loop by the metering device. Sample is applied to the column when the
injection valve returns to the mainpass (main path) position at the end of the
sampling sequence.
The sampling sequence occurs in the following order:
1 The injection valve switches to the bypass position.
2 The piston of the metering device moves to the initialization position.
3 The gripper arm moves from the home position, and selects the vial. At the
same time, the needle lifts out of the seat.
4 The gripper moves into the needle station and stops in the draw position.
5 The needle lowers into the vial.
6 The metering device draws the defined sample volume.
7 The needle lifts out of the vial.
1
8 The gripper arm moves out slightly and stops in the wash position1.
9 The needle moves downwards and dips into the wash well of the wash port.
Simultaneously the peristaltic pump delivers the flush solvent.
10 The needle moves back.
11 The gripper arm moves out of the needle station and the wash port snaps
back in position.
1
only if automated needle wash is selected. If this feature is disabled, the gripper arm
positions the sample vial directly below the needle (Step 4) and lowers the needle into
the vial.
Agilent InfinityLab LC Series Vialsamplers User Manual18
1
1
1Introduction
Vial
Wash
port
Needle
seat
Sampling loop
Solvent
Wash
pump
Metering
device
Operating Principle
12 The gripper arm replaces the vial, and returns to the home position.
Simultaneously, the needle lowers into the seat.
13 The injection valve switches to the mainpass (main path) position.
Injection Sequence
Before the start of the injection sequence, and during an analysis, the injection
valve is in the mainpass (main path) position (see
position, the mobile phase flows through the autosamplers metering device,
sample loop, and needle. This ensures that all parts in contact with sample are
flushed during the run, thus minimizing carryover.
Figure 4 on page 19). In this
Figure 4Mainpass (main path) Position - standard position during runs and when the sampler
idle
is
Agilent InfinityLab LC Series Vialsamplers User Manual19
1Introduction
Vial
Wash
port
Needle
seat
Sampling loop
Solvent
Wash
pump
Metering
device
Operating Principle
When the sample sequence begins, the valve unit switches to the bypass
position (see
Figure 5 on page 20). Solvent from the pump enters the valve unit at
port 1, and flows directly to the column through port 6.
Figure 5Valve in bypass - needle in vial, metering device aspirates sample volume
Agilent InfinityLab LC Series Vialsamplers User Manual20
1Introduction
Vial
Wash
port
Needle
seat
Sampling loop
Solvent
Wash
pump
Metering
device
Operating Principle
Then the vial is positioned below the needle. The needle moves down into the
vial, the metering unit draws the required sample volume into the loop, and the
needle is raised. In the next step, the needle is washed (see
Figure 6 on page 21).
Figure 6Outer face of needle getting washed in wash port
Agilent InfinityLab LC Series Vialsamplers User Manual21
1Introduction
Vial
Wash
port
Needle
seat
Sampling loop
Solvent
Wash
pump
Metering
device
Operating Principle
When the metering unit has drawn the required volume of sample into the
sample loop, the vial is replaced in the sample tray. The wash port flips into the
origin position, the needle is lowered into the needle seat, and the injection valve
switches back to mainpass (main path) position, flushing the sample onto the
column (see
Figure 7 on page 22).
Figure 7Valve switches to mainpass (main path) - sample is transferred towards the LC column
Agilent InfinityLab LC Series Vialsamplers User Manual22
1Introduction
Operating Principle
Needle Parkstation
The needle parkstation comprises two main assemblies: needle drive and wash
port.
Figure 8Needle Station
Needle-Drive
The needle movement is driven by a stepper motor connected to the spindle
assembly by a toothed belt. The circular motion of the motor is converted to
linear motion by the drive nut on the spindle assembly. The upper and lower
needle positions are detected by reflection sensors on the needle station board,
while the needle-in-vial position is determined by counting the motor steps from
the upper needle-sensor position.
Agilent InfinityLab LC Series Vialsamplers User Manual23
1Introduction
NOTE
Operating Principle
Hydraulic Box
The hydraulic box comprises two main assemblies: metering device, and
injection valve.
Figure 9Hydraulic Unit
The replacement hydraulic box excludes the injection valve and metering head
assemblies.
Agilent InfinityLab LC Series Vialsamplers User Manual24
1Introduction
Operating Principle
Analytical Head
The analytical head is driven by the stepper motor that is connected to the drive
shaft by a toothed belt. The drive nut on the spindle converts the circular
movement of the spindle to linear motion. The drive nut pushes the sapphire
piston against the tension of the spring into the analytical head. The base of the
piston sits on the large bearing of the drive nut, which ensures the piston is
always centered. A ceramic ring guides the movement of the piston in the
analytical head. The home position of the piston is sensed by an optical sensor
on the hydraulic unit board while the sample volume is determined by counting
the number of steps from the home position. The backward movement of the
piston (driven by the spring) draws sample from the vial.
Injection Valve
The two-position 6-port injection valve is driven by a stepper motor. Only five of
the six ports are used (port 3 is not used). A lever/slider mechanism transfers the
movement of the stepper motor to the injection valve. Two microswitches
monitor switching of the valve (bypass and mainpass (main path) end positions).
No valve adjustments are required after replacing internal components.
Agilent InfinityLab LC Series Vialsamplers User Manual25
1Introduction
X motor
Theta motor
Gripper motor
Gripper arm
Z motor
Flex board
Theta axis
X axis
Z axis
Operating Principle
Transport Assembly
The transport unit comprises an X-axis slide (left-right motion), a Z-axis arm
(up-down motion), and a gripper assembly (rotation and vial-gripping).
Figure 10 Transport Assembly
The transport assembly uses four stepper motors driven in closed-loop mode for
accurate positioning of the gripper assembly. The rotational movement of the
motors is converted to linear motion (X- and Z-axes) by toothed belts connected
to the drive spindles. The rotation (theta axes) of the gripper assembly is
transferred from the motor by a toothed belt and series of gears. The opening
and closing of the gripper fingers are driven by a stepper motor linked by a
toothed belt to the planetary gearing inside the gripper assembly.
The stepper motor positions are determined by the optical encoders mounted
onto the stepper-motor housing. The encoders monitor the position of the
motors continually, and correct for position errors automatically (e.g. if the
gripper is accidentally moved out of position when loading vials into the vial tray).
The initialization positions of the moving components are sensed by reflection
sensors mounted on the flex board. These positions are used by the processor to
calculate the actual motor position. An additional six reflection sensors for tray
recognition are mounted on the flex board at the front of the assembly.
Agilent InfinityLab LC Series Vialsamplers User Manual26
1Introduction
Leak and Waste Handling
Leak and Waste Handling
The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should not
exceed 2
Cabinets (a printed copy of the guideline has been shipped with the solvent
cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and
Infinity II modules can be stacked on top of each other. Waste tubes are guided
through a channel on the right hand side of the instrument, keeping the front
access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
•from the detector's flow cell outlet
•from the Multisampler needle wash port
•from the Sample Cooler or Sample Thermostat (condensate)
•from the pump's Seal Wash Sensor (if applicable)
•from the pump's Purge Valve or Multipurpose Valve
L. For details, see the usage guideline for the Agilent Infinity II Solvent
Agilent InfinityLab LC Series Vialsamplers User Manual27
1Introduction
Leak and Waste Handling
Figure 11 Infinity II Leak Waste Concept (Flex Bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual28
1Introduction
Leak and Waste Handling
Figure 12 Infinity II Single Stack Leak Waste Concept (bench installation)
Agilent InfinityLab LC Series Vialsamplers User Manual29
1Introduction
Leak and Waste Handling
Figure 13 Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak plane outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Agilent InfinityLab LC Series Vialsamplers User Manual30
Loading...
+ 339 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.