Agilent E4991A Data Sheet

Agilent E4991A RF Impedance/Material Analyzer
Data Sheet
2
Definitions
All specifications apply over a 5 °C to 40 °C range (unless otherwise stated) and 30 minutes after the instrument has been turned on.
Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Supplemental information is intended to provide information useful in applying the instrument, but that is not covered by the product warranty. The information is denoted as typical, or nominal.
Typical (typ.)
Expected performance of an average unit which does not include guardbands. It is not covered by the product warranty.
Nominal (nom.)
A general, descriptive term that does not imply a level of performance. It is not covered by the product warranty.
Measurement Parameters and Range
Measurement parameters
Impedance parameters:
|Z|, |Y|, Ls, Lp, Cs, Cp, Rs(R), Rp, X, G, B, D, Q, θz,
θy, |Γ|, Γx, Γy, θ
γ
Material parameters (option E4991A-002):
(see “Option E4991A-002 material measurement (typical)” on page 17)
Permittivity parameters: |εr|, εr', εr", tanδ Permeability parameters: |µr|, µr', µr", tanδ
Measurement range
Measurement range (|Z|):
130 mto 20 k.
(Frequency= 1 MHz, Point averaging factor ≥ 8, Oscillator level= –3 dBm; = –13 dBm; or = –23 dBm, Measurement accuracy ±10%, Calibration is performed within 23 °C ±5 °C, Measurement is performed within ±5 °C of calibration temperature)
3
1. It is possible to set more than 0 dBm (447 mV, 8.94 mA) oscillator level at frequency > 1 GHz. However, the characteristics at this setting are not guaranteed.
2. When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.
DC Bias (Option E4991A-001)
DC voltage bias
Range:
0 to ±40 V
Resolution:
1 mV
Accuracy:
±{0.1% + 6 mV + (Idc[mA] x 20 )[mV]}
(23 °C ±5 °C)
±{0.2% +12 mV + (Idc[mA] x 40 Ω)[mV]}
(5 °C to 40 °C)
DC current bias
Range:
100 µ A to 50 mA, –100 µ A to –50 mA
Resolution:
10 µ A
Accuracy:
±{0.2%+ 20 µ A+ (Vdc[V] /10 k)[mA]}
(23 °C ±5 °C)
±{0.4% + 40 µ A+ (Vdc[V] /5 k)[mA]}
(5 °C to 40 °C)
DC bias monitor
Monitor parameters:
Voltage and current
Voltage monitor accuracy:
±{0.5% + 15 mV + (Idc[mA] x 2 )[mV]}
(23 °C ±5 °C, typical)
±{1.0% + 30 mV + (Idc[mA] x 4 )[mV]}
(5 °C to 40 °C, typical)
Current monitor accuracy:
±{0.5% + 30 µ A + (Vdc[V] / 40 k Ω)[mA]}
(23 °C ±5 °C, typical)
±{1.0% + 60 µ A + (Vdc[V] / 20 k Ω)[mA]}
(5 °C to 40 °C, typical)
Source Characteristics
Frequency
Range:
1 MHz to 3 GHz
Resolution:
1 mHz
Accuracy:
without Option E4991A-1D5:
±10 ppm (23 °C ±5 °C) ±20 ppm (5 °C to 40 °C)
with Option E4991A-1D5:
±1 ppm (5 °C to 40 °C)
Stability:
with Option E4991A-1D5:
±0.5 ppm/year (5 °C to 40 °C)
Oscillator level
Range:
Power (when 50 load is connected to test port):
–40 dBm to 1 dBm (frequency 1 GHz) –40 dBm to 0 dBm (frequency > 1 GHz1)
Current (when short is connected to test port):
0.0894 mArms to 10 mArms (frequency 1 GHz)
0.0894 mArms to 8.94 mArms (frequency > 1 GHz1)
Voltage (when open is connected to test port):
4.47 mVrms to 502 mVrms (frequency 1 GHz)
4.47 mVrms to 447 mVrms (frequency > 1 GHz1)
Resolution:
0.1 dB
2
Accuracy:
(Power, when 50 load is connected to test port)
Frequency 1 GHz:
±2 dB (23 °C ±5 °C) ±4 dB (5 °C to 40 °C)
Frequency > 1 GHz:
±3 dB (23 °C ±5 °C) ±5 dB (5 °C to 40 °C)
with Option E4991A-010:
Frequency 1 GHz
±3.5 dB (23 °C ± 5 °C) ±5.5 dB (5 °C to 40 °C)
Frequency > 1 GHz
±5.6 dB (23 °C ± 5 °C) ±7.6 dB (5 °C to 40 °C)
Output impedance
Output impedance:
50 (nominal)
4
Measurement Accuracy
Conditions for defining accuracy
Temperature:
23 °C ±5 °C
Accuracy-specified plane:
7-mm connector of test head
Accuracy defined measurement points:
Same points at which the calibration is done.
Accuracy when open/short/load calibration is performed
Probe Station Connection Kit (Option E4991A-010)
Oscillator level
Power accuracy:
Frequency 1 GHz:
±5.5 dB (5 °C to 40 °C)
Frequency > 1 GHz:
±7.6 dB (5 °C to 40 °C)
Sweep Characteristics
Sweep conditions
Sweep parameters:
Frequency, oscillator level (power, voltage,
current), DC bias voltage, DC bias current
Sweep range setup:
Start/stop or center/span
Sweep types:
Frequency sweep: linear, log, segment Other parameters sweep: linear, log
Sweep mode:
Continuous, single
Sweep directions:
Oscillator level, DC bias (voltage and current): up sweep,
down sweep
Other parameters sweep: up sweep
Number of measurement points:
2 to 801
Delay time:
Types: point delay, sweep delay, segment delay Range: 0 to 30 sec Resolution: 1 msec
Segment sweep
Available setup parameters for each segment:
Sweep frequency range, number of measurement
points, point averaging factor, oscillator level (power, voltage, or current), DC bias (voltage or current), DC bias limit (current limit for voltage bias, voltage limit for current bias)
Number of segments:
1 to 16
Sweep span types:
Frequency base or order base
|Z|, |Y|: ±(Ea+ Eb) [%]
(see Figures 1 through 4 for examples of calculated accuracy)
θ: ±
(Ea+ Eb)
[rad]
100
L, C, X, B: ±(E
a
+ Eb) x (1 + D
2 x
) [%]
R, G: ±(Ea+ Eb) x (1 + Q
2 x
) [%]
D:
at D
x
tan
E
a
+ E
b
< 1 ±
100
at D
x
0.1
±
E
a
+ E
b
100
Q:
at Qxtan
Ea+ E
b
< 1 ±
100
at
10
Q
x
10 ±Q
2
x
Ea+ E
b
Ea+ E
b
100
(1 + D
2
x
)tan
Ea+ E
b
100
1 Dxtan
E
a
+ E
b
100
(1 + Q
2
x
)tan
E
a
+ E
b
100
1 Qxtan
E
a
+ E
b
100
±
±
at Oscillator level < –33 dBm:
±1 [%] (1 MHz Frequency 100 MHz) ±1.2 [%] (100 MHz < Frequency 500 MHz) ±1.2 [%] (500 MHz < Frequency 1 GHz) ±2.5 [%] (1 GHz < Frequency 1.8 GHz) ±5 [%] (1.8 GHz < Frequency 3 GHz)
Eb =
(|Zx|: measurement value of |Z|)
Ec =
(F: frequency [MHz], typical)
Zs = (Within ±5 °C from the calibration temperature. Measurement accuracy applies when the calibration is performed at 23 °C ±5 °C. When the calibration is performed beyond 23 °C ±5 °C, the measurement accuracy decreases to half that described. F: frequency [MHz].)
at oscillator level = –3 dBm, –13 dBm, or –23 dBm:
±(13 + 0.5 × F) [m] (averaging factor 8) ±(25 + 0.5 × F) [m] (averaging factor 7)
at oscillator level –33 dBm
±(25 + 0.5 × F) [m] (averaging factor 8) ±(50 + 0.5 × F) [m] (averaging factor 7)
at oscillator level < –33 dBm
±(50 + 0.5 × F) [m] (averaging factor 8) ±(100 + 0.5 × F) [m] (averaging factor 7)
Yo = (Within ±5 °C from the calibration temperature. Measurement accuracy applies when the calibration is performed at 23 °C ±5 °C. When the calibration is performed beyond 23 °C ±5 °C, the measurement accuracy decreases to half that described. F: frequency [MHz].)
at oscillator level = –3 dBm, –13 dBm, –23 dBm:
±(5 + 0.1 × F) [µS] (averaging factor 8) ±(10 + 0.1 × F) [µS] (averaging factor 7)
at oscillator level –33 dBm:
±(10 + 0.1 × F) [µS] (averaging factor 8) ±(30 + 0.1 × F) [µS] (averaging factor 7)
at oscillator level < –33 dBm
±(20 + 0.1 × F) [µS] (averaging factor 8) ±(60 + 0.1 × F) [µS] (averaging factor 7)
± 0.06 +
0.08 × F
[%]
1000
Accuracy when open/short/load/low-loss capacitor calibration is performed
(See Figure 5)
Definition of each parameter
Dx = Measurement value of D
Qx = Measurement value of Q
Ea = (Within ±5 °C from the calibration temperature.
Measurement accuracy applies when the calibration is performed at 23 °C ±5 °C. When the calibration is performed beyond 23 °C ±5 °C, measurement error doubles.)
at oscillator level –33 dBm:
±0.65 [%] (1 MHz Frequency 100 MHz) ±0.8 [%] (100 MHz < Frequency 500 MHz) ±1.2 [%] (500 MHz < Frequency 1 GHz) ±2.5 [%] (1 GHz < Frequency 1.8 GHz) ±5 [%] (1.8 GHz < Frequency 3 GHz)
|Z|, |Y|: ±(Ea+ Eb) [%]
θ: ±
E
c
[rad]
100
L, C, X, B: ± (E
a
+ Eb)2+ (EcDx)2[%]
R, G: ± (Ea+ Eb)2+ (EcQx)2[%]
D:
at Dxtan
E
c
< 1 ±
100
at D
x
0.1 ±
E
c
100
Q:
at Qxtan
E
c
< 1 ±
100
at
10
Qx≥ 10 ±Q
2
x
E
c
E
c
100
(1 + D
2
x
)tan
E
c
100
1 Dxtan
E
c
100
(1 + Q
2
x
)tan
E
c
100
1 Qxtan
E
c
100
±
Z
s
+Yo• Z100 [%]
Z
x
5
±
±
6
Figure 2. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level –33 dBm. Point averaging factor 8 within ±5 °C from the calibration temperature.
Figure 3. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level –33 dBm. Point averaging factor 7 within ±5 °C from the calibration temperature.
Measurement Accuracy
(continued)
Examples of calculated impedance measurement accuracy
Figure 1. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level = –23 dBm,
–13 dBm, –3 dBm. Point averaging factor 8 within ±5 °C
from the calibration temperature.
7
Figure 4. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level < –33 dBm within ±5 °C from the calibration temperature.
Figure 5. Q Measurement accuracy when open/short/load/low-loss
capacitor calibration is performed (typical).
Measurement Support Functions
Error correction
Available calibration and compensation
Open/short/load calibration:
Connect open, short, and load standards to the desired reference plane and measure each kind of calibration data. The reference plane is called the calibration reference plane.
Low-loss capacitor calibration:
Connect the dedicated standard (low-loss capacitor) to the calibration reference plane and measure the calibration data.
Port extension compensation (fixture selection):
When a device is connected to a terminal that is extended from the calibration reference plane, set the electrical length between the calibration plane and the device contact. Select the model number of the registered test fixtures in the E4991A’s setup toolbar or enter the electrical length for the user’s test fixture.
Open/short compensation:
When a device is connected to a terminal that is extended from the calibration reference plane, make open and/or short states at the device contact and measure each kind of
compensation data.
Calibration/compensation data measurement point
User-defined point mode:
Obtain calibration/compensation data at the same frequency and power points as used in actual device measurement, which are determined by the sweep setups. Each set of calibration/compensation data is applied to each measurement at the same point. If measurement points (frequency and/or power) are changed
by altering the sweep setups, calibration/ compensation data become invalid and calibration or compensation data acquisition is again required.
8
Measurement Support Functions
(continued)
Fixed frequency and fixed power point mode:
Obtain calibration/compensation data at fixed frequency and power points covering the entire frequency and power range of the E4991A. In device measurement, calibration or compensation is applied to each measurement point by using interpolation. Even if the measurement points (frequency and/or power) are changed by altering the sweep setups, you don’t need to retake the calibration or compensation data.
Fixed frequency and user-defined power point mode:
Obtain calibration/compensation data at fixed frequency points covering the entire frequency range of the E4991A and at the same power points as used in actual device measurement which are determined by the sweep setups. Only if the power points are changed, calibration/ compensation data become invalid and calibration or compensation data acquisition is again required.
Trigger
Trigger mode:
Internal, external (external trigger input
connector), bus (GPIB), manual (front key)
Averaging
Types:
Sweep-to-sweep averaging, point averaging
Setting range:
Sweep-to-sweep averaging: 1 to 999 (integer) Point averaging: 1 to 100 (integer)
Display
LCD display :
Type/size: color LCD, 8.4 inch (21.3 cm) Resolution: 640 (horizontal) × 480 (vertical)
Number of traces:
Data trace: 3 scalar traces + 2 complex traces
(maximum)
Memory trace: 3 scalar traces + 2 complex traces
(maximum)
Trace data math:
Data – memory, data/memory (for complex
parameters), delta% (for scalar parameters), offset
Format:
For scalar parameters: linear Y-axis, log Y-axis For complex parameters: Z, Y: polar, complex; Γ: polar,
complex, Smith, admittance
Other display functions:
Split/overlay display (for scalar parameters), phase expansion
Marker
9
3. Refer to the standard for the meaning of each function code.
Number of markers:
Main marker: one for each trace (marker 1) Sub marker: seven for each trace (marker 2 to
marker 8)
Reference marker: one for each trace (marker R)
Marker search:
Search type: maximum, minimum, target, peak Search track: performs search with each sweep
Other functions:
Marker continuous mode, marker coupled mode, marker list, marker statistics
Equivalent circuit analysis
Circuit models:
3-component model (4 models), 4-component model (1 model)
Analysis types:
Equivalent circuit parameters calculation,
frequency characteristics simulation
Limit marker test
Number of markers for limit test:
9 (marker R, marker 1 to 8)
Setup parameters for each marker:
Stimulus value, upper limit, and lower limit
Mass storage
Built-in flexible disk drive:
3.5 inch, 720 KByte or 1.44 MByte, DOS format
Hard disk drive:
2 GByte (minimum)
Stored data:
State (binary), measurement data (binary, ASCII or CITI file), display graphics (bmp, jpg), VBA program (binary)
Interface
GPIB
Standard conformity:
IEEE 488.1-1987, IEEE 488.2-1987
Available functions (function code)3:
SH1, AH1, T6, TE0, L4, LE0, SR1, RL0, PP0, DT1, DC1, C0, E2
Numerical data transfer format:
ASCII
Protocol:
IEEE 488.2-1987
Printer parallel port
Interface standard:
IEEE 1284 Centronics
Connector type:
25-pin D-sub connector, female
LAN interface
Standard conformity:
10 Base-T or 100 Base-TX (automatically switched), Ethertwist, RJ45 connector
Protocol:
TCP/IP
Functions:
FTP
10
Rear panel connectors
External reference signal input connector
Frequency:
10 MHz ±10 ppm (typical)
Level:
0 dBm to +6 dBm (typical)
Input impedance:
50 (nominal)
Connector type:
BNC, female
Internal reference signal output connector
Frequency:
10 MHz (nominal)
Accuracy of frequency:
Same as frequency accuracy described in “Frequency” on page 3
Level:
+2 dBm (nominal)
Output impedance:
50 (nominal)
Connector type:
BNC, female
High stability frequency reference output connector (Option E4991A-1D5)
Frequency:
10 MHz (nominal)
Accuracy of frequency:
Same as frequency accuracy described in “Frequency” on page 3
Level:
+2 dBm (nominal)
Output impedance:
50 (nominal)
Connector type:
BNC, female
Measurement Terminal (At Test Head)
Connector type:
7-mm connector
11
General Characteristics
Environment conditions
Operating condition
Temperature:
5 °C to 40 °C
Humidity:
(at wet bulb temperature ≤ 29 °C, without condensation)
Flexible disk drive non-operating condition:
15% to 90% RH
Flexible disk drive operating condition:
20% to 80% RH
Altitude:
0 m to 2,000 m (0 feet to 6,561 feet)
Vibration:
0.5 G maximum, 5 Hz to 500 Hz
Warm-up time:
30 minutes
Non-operating storage condition
Temperature:
–20 °C to +60 °C
Humidity:
(at wet bulb temperature ≤ 45 °C, without condensation)
15% to 90% RH
Altitude:
0 m to 4,572 m (0 feet to 15,000 feet)
Vibration:
1 G maximum, 5 Hz to 500 Hz
External trigger input connector
Level:
LOW threshold voltage: 0.5 V HIGH threshold voltage: 2.1 V Input level range: 0 V to +5 V
Pulse width (Tp):
2 µ sec (typical). See Figure 6 for definition of Tp.
Polarity:
Positive or negative (selective)
Connector type:
BNC, female
Figure 6. Definition of pulse width (Tp)
12
General Characteristics
(continued)
Other specifications
EMC
European Council Directive 89/336/EEC
IEC 61326-1:1997+A1
CISPR 11:1990 / EN 55011:1991 Group 1, Class A
IEC 61000-4-2:1995 / EN 61000-4-2:1995
4 kV CD / 4 kV AD
IEC 61000-4-3:1995 / EN 61000-4-3:1996
3 V/m, 80-1000 MHz, 80% AM
IEC 61000-4-4:1995 / EN 61000-4-4:1995
1 kV power / 0.5 kV Signal
IEC 61000-4-5:1995 / EN 61000-4-5:1995
0.5 kV Normal / 1 kV Common
IEC 61000-4-6:1996 / EN 61000-4-6:1996
3 V, 0.15-80 MHz, 80% AM
IEC 61000-4-11:1994 / EN 61000-4-11:1994
100% 1cycle Note: When tested at 3 V/m according to EN 61000-4-3:1996, the measurement accuracy will be within specifications over the full immunity test frequency range of 80 MHz to 1000 MHz except when the analyzer frequency is identical to the transmitted interference signal test frequency.
AS/NZS 2064.1/2 Group 1, Class A
Safety
European Council Directive 73/23/EEC IEC 61010-1:1990+A1+A2 / EN 61010-1:1993+A2
INSTALLATION CATEGORY II, POLLUTION DEGREE 2 INDOOR USE
IEC60825-1:1994 CLASS 1 LED PRODUCT
CAN/CSA C22.2 No. 1010.1-92
Power requirements
90 V to 132 V, or 198 V to 264 V (automatically switched), 47 Hz to 63 Hz, 350 VA maximum
Weight
Main unit: 17 kg (nominal) Test head: 1 kg (nominal)
Dimensions
Main unit: See Figure 7 through Figure 9 Test head: See Figure 10 Option E4991A-007 test head dimensions: See Figure 11 Option E4991A-010 test head dimensions: See Figure 12
13
Figure 8. Main unit dimensions
(rear view, in millimeters, nominal)
Figure 9. Main unit dimensions
(side view, in millimeters, nominal)
Figure 7. Main unit dimensions
(front view, in millimeters, nominal)
14
Figure 10. Test head dimensions
(in millimeters, nominal)
General Characteristics (continued)
Figure 11. Option E4991A-007 test head
dimensions (in millimeters,nominal)
15
General Characteristics (continued)
Figure 12. Option E4991A-010 test head
dimensions (in millimeters,nominal)
16
Furnished accessories
Model/option number Agilent part number Description Qty
Agilent E4991A - Agilent E4991A impedance/material analyzer (main unit) 1
- Test head 1
- Agilent 16195B 7-mm calibration kit 1 8710-1766 Torque wrench 1
- Mouse
4
1
- Keyboard
5
1
- Power cable 1 E4991-1610x CD-ROM (firmware and VBA software)
6
1
E4991-905xx CD-ROM (for manual)
6
1
Option 007 E4991-60005 Fixture station 1
E4991-60006 Test head stand 1 E4991-60032 7-mm OPEN standard 1 E4991-60031 7-mm SHORT standard 1 04287-61651 RF extenstion cable assembly 1 1250-2879 N (m) to SMA (f) adapter 3 1250-3157 N (f) to SMA (f) adapter 3 E4991-18001 Floppy disk
(Contains temperature characterisitc 1
measurement software) 04291-09001 Pad 1 1400-0584 Mount cable tie 1
Option 010 - Test head 1
(Including 1 m extension cable) 1250-2879 N (m) to SMA (f) adapter 3 1250-1747 3.5-mm to 7-mm adapter 1 0515-1551 Screw 4 3050-0891 Washer 4
Option ABA E4991-900x0 Operation manual
6
1
E4991-900x1 Installation and quick start guide
6
1
E4991-900x2 Programming manual
6
1
E4991-180x0 Sample program disk (3.5 inch floppy disk)
6
1
Option 1D5 8120-1838 BNC(m)-BNC(m) cable
7
1 Option 1CM 5063-9216 Rackmount kit 1 Option 1CN 5063-9229 Handle kit 1 Option 1CP 5063-9223 Rackmount & handle kit 1
4. Not furnished if Option 1CS (without mouse) is designated.
5. Not furnished if Option 1A2 (without keyboard) is designated.
6. The number indicated by “x” in the part number of each manual, sample
program disk, or CD-ROM, is incremented by 1 each time a revision is made. The latest edition comes with the product (0 for the first edition).
7. This cable is furnished if Option 1D5 high stability frequency reference is designated.
Typical accuracy of permittivity parameters:
εr' accuracy
(at tanδ < 0.1)
Loss tangent accuracy of εr(= tanδ):
±(Ea+ Eb) (at tanδ <0.1)
where,
E
a
=
at Frequency ≤ 1 GHz:
E
b
=
f = Measurement frequency [GHz]
t = Thickness of MUT (material under test) [mm] ε'
rm
= Measured value of ε'
r
tanδ = Measured value of dielectric loss tangent
17
=
∆ε'
rm
:
ε'
rm
Option E4491A-002 Material Measurement (Typical)
Measurement parameter
Permittivity parameters:
|εr|, εr', εr", tanδ
Permeability parameters:
|µr|, µr', µr", tanδ
Frequency range
Using with Agilent 16453A:
1 MHz to 1 GHz (typical)
Using with Agilent 16454A:
1 MHz to 1 GHz (typical)
Measurement accuracy
Conditions for defining accuracy:
Calibration:
Open, short, and load calibration at the test
port (7-mm connector)
Calibration temperature:
Calibration is performed at an environmental
temperature within the range of 23 °C ± 5 °C. Measurement error doubles when calibration temperature is below 18 °C or above 28 °C.
Temperature:
Temperature deviation: within ±5 ˚C from the
calibration temperature
Environment temperature: Measurement accuracy
applies when the calibration is performed at 23 °C ±5 °C. When the calibration is below
18 °C or above 23 °C, measurement error doubles.
Measurement frequency points:
Same as calibration points
Oscillator level:
Same as the level set at calibraiton
Point averaging factor: 8 Electrode pressure setting of 16453A: maximum
± 5 + 10 +
0.1 t + 0.25
ε'
rm
+
100
[%]
f
ε'
rm
t
1–
13
2
f √ε'
rm
0.002 +
0.001•t + 0.004f +
0.1
f ε'
rm
1–
13
2
f √ε'
rm
∆ε'
rm
• 1+ ε'
rm
0.002
tanδ
ε
'
rm
100 t
Typical accuracy of permeability parameters:
µ
r
' accuracy
(at tanδ <0.1)
Loss tangent accuracy of µr(= tanδ):
±(Ea+ Eb) (at tanδ <0.1)
where,
Ea=
E
b
=
f = Measurement frequency [GHz]
F =
h = Height of MUT (material under test) [mm]
b = Inner diameter of MUT (material under
test) [mm]
c = Outer diameter of MUT (material under
test) [mm]
µ'
rm
= Measured value of µ'
r
tanδ = Measured value of loss tangent
18
=
µ'
rm
:
µ'
rm
µrm' • tanδ
µ'
rm
100
4 +
0.02 25 + Fµ'
rm
1 +
15
2
f
2
[%]
fFµ'
rm
Fµ'
rm
0.002 +
0.001 + 0.004 f
Fµ'rmf
hln c[mm]
b
×
19
Option E4491A-002 Material Measurement (typical)
(continued)
Examples of calculated permittivity measurement accuracy
Figure 13. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 0.3 mm, typical)
Figure 14. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 1 mm, typical)
Figure 15. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 3 mm, typical)
20
Figure 16. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 0.3 mm, typical)
8
Figure 17. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 1 mm, typical)
8
Figure 18. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 3 mm, typical)
8
8. This graph shows only frequency dependence of Eato simplify it.
The typical accuracy of tanδ is defined as E
a
+ Eb; refer to “Typical
accuracy of permittivity parameters” on page 17.
Figure 19. Permittivity (ε'r) vs.
frequency (at t = 0.3 mm, typical)
Figure 20. Permittivity (ε'r) vs.
frequency (at t = 1 mm, typical)
Figure 21. Permittivity (ε'r) vs.
frequency (at t = 3 mm, typical)
Option E4991A-002 Material Measurement (typical)
(continued)
Examples of calculated permeability measurement accuracy
Figure 22. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 0.5, typical)
Figure 23. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 3, typical)
Figure 24. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 10, typical)
21
22
Figure 25. Permeability loss tangent (tanδ)
accuracy vs. Frequency (at F = 0.5, typical)
9
Figure 26. Permeability loss tangent (tanδ)
accuracy vs. frequency (at F = 3, typical)
9
Figure 27. Permeability loss tangent (tanδ)
accuracy vs. frequency (at F = 10, typical)
9
Figure 28. Permeability (µ'r) vs.
frequency (at F = 0.5, typical)
Figure 29. Permeability (µ'r) vs.
frequency (at F = 3, typical)
Figure 30. Permeability (µ'r) vs.
frequency (at F = 10, typical)
9. This graph shows only frequency dependence of Eato simplify it.
The typical accuracy of tanδ is defined as E
a
+ Eb; refer to “Typical
accuracy of permeability parameters” on page 18.
23
Option E4991A-007 Temperature Characteristic Test Kit
This section contains specifications and supplemental information for the E4991A Option E4991A-007. Except for the contents in this section, the E4991A standard specifications and supplemental information are applied.
Operation temperature
Range:
–55 °C to +150 °C (at the test port of the high
temperature cable)
Source characteristics
Frequency
Range:
1 MHz to 3 GHz
Oscillator level
Source power accuracy at the test port of the high temperature cable:
Frequency 1 GHz:
+2 dB/–4 dB (23 °C ±5 °C) +4 dB/–6 dB (5 °C to 40 °C)
Frequency > 1 GHz:
+3 dB/–6 dB (23 °C ±5 °C) +5 dB/–8 dB (5 °C to 40 °C)
Measurement accuracy (at 23 °C ±5 °C)
Conditions
10
The measurement accuracy is specified when the following conditions are met:
Calibration: open, short and load calibration is
completed at the test port (7-mm connector) of the high temperature cable
Calibration temperature: calibration is performed at
an environmental temperature within the range of 23 °C ±5 °C. Measurement error doubles when calibration temperature is below 18 °C or above 28 °C.
Measurement temperature range: within ±5 °C of
calibration temperature
Measurement plane: same as calibration plane Oscillator level: same as the level set at calibration
Impedance, admittance and phase angle accuracy:
10. The high temperature cable must be kept at the same position throughout calibration and measurement.
|Z|, |Y| ± (Ea+ Eb) [%]
(see Figure 31 through Figure 34 for calculated accuracy)
θ ±
(E
a
+ Eb)
[rad]
100
where, Ea= at oscillator level ≥ –33 dBm:
±0.8 [%] (1 MHz ƒ 100 MHz) ±1 [%] (100 MHz < ƒ 500 MHz) ±1.2 [%] (500 MHz < ƒ 1 GHz) ±2.5 [%] (1 GHz < ƒ 1.8 GHz) ±5 [%] (1.8 GHz < ƒ 3 GHz)
at oscillator level ≥ –33 dBm:
±1.2 [%] (1 MHz ƒ 100 MHz) ±1.5 [%] (100 MHz < ƒ 500 MHz) ±1.5 [%] (500 MHz < ƒ 1 GHz) ±2.5 [%] (1 GHz < ƒ 1.8 GHz) ±5 [%] (1.8 GHz < ƒ 3 GHz) (Where, ƒ is frequency)
E
b
= ±
Z
s
+ Yo× |Zx| × 100
[%]
|Z
x
|
Where,
|Z
x
|= Absolute value of impedance
Zs= At oscillator level = –3 dBm, –13 dBm, or –23 dBm:
± (30 + 0.5 × F) [m] (point averaging factor 8) ± (40 + 0.5 × F) [m] (point averaging factor 7)
At oscillator level ≥ –33 dBm:
± (35 + 0.5 × F) [m] (point averaging factor 8) ± (70 + 0.5 × F) [m] (point averaging factor 7)
At oscillator level < –33 dBm:
± (50 + 0.5 × F) [m] (point averaging factor 8) ± (150 + 0.5 × F) [m] (point averaging factor 7)
(Where, F is frequency in MHz)
Yo= At oscillator level = –3 dBm, –13 dBm, –23 dBm:
± (12 + 0.1 × F) [µS] (point averaging factor 8) ± (20 + 0.1 × F) [µS] (point averaging factor 7)
At oscillator level ≥ –33 dBm:
± (15 + 0.1 × F) [µS] (point averaging factor 8) ± (40 + 0.1 × F) [µS] (point averaging factor 7)
At oscillator level < –33 dBm:
± (35 + 0.1 × F) [µS] (point averaging factor 8) ± (80 + 0.1 × F) [µS] (point averaging factor 7)
(Where, F is frequency in MHz)
24
Calculated Impedance/ Admittance Measurement Accuracy
Figure 31. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level = –23 dBm, –13 dBm, –3 dBm. Point averaging factor 8 within ±5 °C of calibration temperature.
Figure 32. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level –33 dBm. Point averaging factor 8 within ±5 °C of calibration temperature.
Figure 33. |Z|, |Y| measurement accuracy when open/short/load
calibration is perfomed. Oscillator level –33 dBm. Point averaging factor 7 within ±5 °C of calibration temperature.
Figure 34. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level < –33 dBm. Point averaging factor 8 within ±5 °C of calibration temperature.
25
Typical Effects of Temperature Change on Measuement Accuracy
When the temperature at the test port (7-mm connector) of the high temperature cable changes from the calibration temperature, typical measurement accuracy involving temperature dependence effects (errors) is applied. The typical measurement accuracy is represented by the sum of error due to temperature coefficients (Ea´, Yo´and Zs´), hysteresis error (E
ah
, Y
oh
and Zsh) and the specified accuracy.
Conditions
The typical measurement accuracy is applied when the following conditions are met:
Conditions of Ea’, Zs’ and Yo’:
Measurement temperature: –55 °C to 5 °C or
40 °C to 150 °C at test port. For 5 °C to 40 °C, Ea´, Yo´ and Zs´ are 0 (neglected).
Temperature change: ≥ 5 °C from calibration
temperature when the temperature compensation is off. 20 °C from calibration temperature when the temperature compensation is set to on.
Calibration temperature: 23 °C ±5 °C Calibration mode: user calibration Temperature compensation: temperature
compensation data is acquired at the same temperature points as measurement temperatures.
Conditions of Eah, Zshand Yoh:
Measurement temperature: –55 °C to 150 °C at
the test port
Calibration temperature: 23 °C ±5 °C Calibration mode: user calibration
Figure 35. Typical frequency characteristics of temperature
coefficient, (Ec+Ed)/T, when |Zx|= 10 and 250 Ω, Eah= Zsh= Yoh= 0 are assumed12.
26
Typical measurement accuracy (involving temperature dependence effects)
11
:
11. See graphs in Figure 35 for the calculated values of (Ec+Ed) exclusive of the hysteresis errors E
ah
, Zshand
Yoh
, when
measured impedance is 10 and 250 Ω.
12. Read the value of |Z|%/°C at the material measurement frequency and multiply it by T to derive the value of (Ec+Ed) when E
ah
= Yoh= Zsh= 0.
|Z|, |Y|: ± (Ea+ Eb+ Ec+ Ed) [%]
θ : ±
(E
a
+ Eb+ Ec+ Ed)
[rad]
100
Where,
Ec= Ea´ ×∆T + E
ah
Ed=±
Zs´× ∆T + Z
sh
+ (Y× ∆T + Yoh) × |Zx|× 100 [%]
|Zx|
Where,
|Z
x
| = Absolute value of measured impedance
Here, Ea´, Zs´ and Yo´ are given by the following equations:
Without temperature With temperature compensation compensation
1 MHz ƒ < 500 MHz 500 MHz ƒ 3 GHz
Ea´ 0.006 + 0.015 × ƒ [%/°C] 0.006 + 0.015 × ƒ [%/°C] 0.006 + 0.015 × ƒ [%/°C]
Zs´ 1 + 10 × ƒ [m/°C] 1 + 10 × ƒ [m/°C] 5 + 2 × ƒ [m/°C]
Yo´ 0.3 + 3 × ƒ [µS/°C] 0.3 + 3 × ƒ [µS/°C] 1.5 + 0.6 × ƒ [µS/°C]
ƒ = Measurement frequency in GHz
Eah, Zshand Yohare given by following equations:
E
ah
= E×∆T
max
× 0.3 [%]
Z
sh
= Z×∆T
max
× 0.3 [m]
Y
oh
= Y×∆T
max
× 0.3 [µS]
T = Difference of measurement temperature from calibration temperature
T
max
= Maximum temperature change (°C) at the test port from calibration
temperature after the calibration is performed.
27
Typical Material Measurement Accuracy When Using Options E4991A-002 and E4991A-007
Material measurement accuracy contains the permittivity and permeability measurement accuracy when the E4991A with Option E4991A-002 and E4991A-007 is used with the 16453A or 16454A test fixture.
Measurement parameter
Permittivity parameters:
|εr|, ε'r, ε", tanδ
Permeability parameters:
|µr|, µ'r, µ", tanδ
Frequency
Use with Agilent 16453A:
1 MHz to 1 GHz (typical)
Use with Agilent 16454A:
1 MHz to 1 GHz (typical)
Opertation temperature
Range:
–55 °C to +150 °C (at the test port of the high
temperature cable)
Typical material measurement accuracy (at 23 °C ±150 °C)
Conditions The measurement accuracy is specified when the following conditions are met:
Calibration: Open, short and load calibration is
completed at the test port (7-mm connector) of the high temperature cable
Calibration temperature: Calibration is performed at
an environmental temperature within the range of 23 °C ±5 °C. Measurement error doubles when calibration temperature is below 18 °C or above 28 °C.
Measurement temperature range: Within ±5 °C of
calibration temperature
Measurement frequency points: Same as calibration
points (User Cal)
Oscillator level: Same as the level set at calibration Point averaging factor: 8
Typical permittivity measurement accuracy
13
:
13. The accuracy applies when the electrode pressure of the 16453A is set to maximum.
εr´ accuracy Eε=
∆ε´
rm
:
ε
´
rm
± 5 + 10 +
0.5
× t+ 0.25 ×
ε´
rm
+
100
f
ε
´
rm
t
|
1–
13
2
|
f √ε´
rm
[%] (at tanδ < 0.1)
Loss tangent accuracy of εr(= tanδ) :
±(Ea+ Eb) (at tanδ < 0.1)
where,
Ea=
at Frequency ≤ 1 GHz
0.002 +
0.0025
× t+ (0.008 × f ) +
0.1
f ε´
rm
|
1–
13
2
|
f √ε´
rm
Eb=
∆ε´
rm
×
1
+ ε´
rm
0.002
× tanδ
ε´
rm
100 t
f = Measurement frequency [GHz]
t = Thickness of MUT (material under test) [mm]
ε´rm= Measured value of ε´
r
tanδ = Measured value of dielectric loss tangent
.
28
µr´ accuracy Eµ=
µ´
rm
:
µ´
rm
4 +
0.02
×
25
+ F × µ´
rm
× 1 +
15
2
× f
2
f F × µ´
rm
F × µ´
rm
[%] (at tanδ < 0.1)
Loss tangent accuracy of µr(= tanδ) :
±(Ea+ Eb) (at tanδ < 0.1)
where,
E
a
= 0.002 +
0.005
+ 0.004 × f
F × µ´
rm
× f
E
b
=
µ´
rm
×
tanδ
µ´
rm
100
f = Measurement frequency [GHz]
F = hln c[mm]
b
h = Height of MUT (material under test) [mm]
b = Inner diameter of MUT [mm]
c = Outer diameter of MUT [mm]
µ´
rm
= Measured value of µ´
r
tanδ = Measured value of loss tangent
Typical permeability measurement accuracy :
.
29
Examples of Calculated Permittivity Measurement Accuracy
14. This graph shows only frequency dependence of Eafor simplification. The typical accuracy of tanδ is defined as E
a
+ Eb;
refer to “Typical permittivity measurement accuracy” on page 27.
Figure 39. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 0.3 mm, typical)
14
Figure 38. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 3 mm, typical)
Figure 36. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 0.3 mm, typical)
Figure 40. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 1 mm, typical)
14
Figure 37. Permittivity accuracy
(∆ε'
r
)
vs.
ε'
r
frequency (at t = 1 mm, typical)
30
Examples of Calculated Permittivity Measurement Accuracy
(continued)
Figure 41. Dielectric loss tangent (tanδ)
accuracy vs. frequency (at t = 3 mm, typical)
14
Figure 42. Permittivity (ε'r) vs.
frequency (at t = 0.3 mm, typical)
14. This graph shows only frequency dependence of Eafor
simplification. The typical accuracy of tanδ is defined as E
a
+ Eb;
refer to “Typical permittivity measurement accuracy” on page 27.
Figure 43. Permittivity (ε'r) vs.
frequency (at t = 1 mm, typical)
Figure 44. Permittivity (ε'r) vs.
frequency (at t = 3 mm, typical)
31
Examples of Calculated Permeability Measurement Accuracy
15. This graph shows only frequency dependence of Eafor simplification. The typical accuracy of tanδ is defined as E
a
+ Eb;
refer to “Typical permeability measurement accuracy” on page 28.
Figure 49. Permeability loss tangent (tanδ)
accuracy vs. Frequency (at F = 3, typical)
15
Figure 48. Permeability loss tangent (tanδ)
accuracy vs. Frequency (at F = 0.5, typical)
15
Figure 46. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 3, typical)
Figure 50. Permeability loss tangent (tanδ)
accuracy vs. Frequency (at F = 10, typical)
15
Figure 47. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 10, typical)
Figure 45. Permeability accuracy
(µ'
r
)
vs.
µ'
r
frequency (at F = 0.5, typical)
32
Examples of Calculated Permeability Measurement Accuracy
(continued)
Figure 52. Permeability (µ'r) vs.
frequency (at F = 3, typical)
Figure 51. Permeability (µ'r) vs.
frequency (at F = 0.5, typical)
Figure 53. Permeability (µ'r) vs.
frequency (at F = 10, typical)
33
Typical Effects of Temperature Change on Permittivity Measurement Accuracy
When the temperature at the test port (7-mm connector) of the high temperature cable changes more than 5 °C from the calibration temperature, the typical permittivity measurement accuracy involving temperature dependence effects (errors) is applied. The typical permittivity accuracy is represented by the sum of error due to temperature coefficient (Tc), hysteresis error (Tc×∆T
max
) and the accuracy at 23 °C ± 5 °C.
ε
r
´ accuracy =
∆ε´
rm
:
ε´
rm
±(Eε+ Ef+ Eg) [%]
Loss tangent accuracy of ε (= tanδ) :
±
(Eε+ Ef+ Eg)
100
where,
Eε= Permittivity measurement accuracy at 23 °C ± 5 °C
Ef= Tc×∆T
Eg= Tc×∆T
max
× 0.3
Tc = K1+ K2+ K
3
See Figure 54 through Figure 56 for the calculated value of T
c
without temperature compensation
K1= 1 × 10-6× (60 + 150 × ƒ)
K2=
× ƒ
3 × 10-6× (1 + 10 × ƒ) ×
ε´
rm
×
1
+10
t
f
2
1–
f
o
K3=
1
5 × 10-3× (0.3 + 3 × ƒ) ×
ε´
rm
×
1
+10 × ƒ
t
f
2
1–
f
o
Typical accuracy of permittivity parameters:
.
34
with temperature compensation
K
1
= 1 × 10-6× (60 + 150 × ƒ)
K
2
= 1 MHz f < 500 MHz
3 × 10-6× (1 + 10 × ƒ) ×
ε´
rm
×
1
+10
t
f
2
× ƒ
1–
f
o
500 MHz ƒ 1 GHz
3 × 10
-6
× (5 + 2 × ƒ) ×
ε´
rm
×
1
+10
t
f
2
× ƒ
1–
f
o
K
3
= 1 MHz ≤ ƒ < 500 MHz
1
5 × 10-3× (0.3 + 3 × ƒ) ×
ε´
rm
×
1
+10× ƒ
t
f
2
1–
f
o
500 MHz ƒ ≤ 1 GHz
1
5 × 10-3× (1.5 + 0.6 × ƒ) ×
ε´
rm
×
1
+10× ƒ
t
f
2
1–
f
o
ƒ = Measurement frequency [GHz]
ƒ
o
=
13
[GHz]
ε´
r
t = Thickness of MUT (material under test) [mm]
ε´rm= Measured value of ε´
r
T = Difference of measurement temperature from calibration temperatureT
max
= Maximum temperature change (°C) at test port from calibration
temperature after the calibration is performed.
Typical accuracy of permittivity parameters (continued):
35
Figure 55. Typical frequency characteristics of temperature coefficient
of ε'
r
(Thickness = 1 mm)
Figure 54. Typical frequency characteristics of temperature coefficient
of ε'
r
(Thickness = 0.3 mm)
Figure 56. Typical frequency characteristics of temperature coefficient
of
ε'
r
(Thickness = 3 mm)
36
Typical Effects of Temperature Change on Permeability Measurement Accuracy
When the temperature at the test port (7-mm connector) of the high temperature cable changes more than 5 °C from the calibration temperature, the typical permeability measurement accuracy involving temperature dependence effects (errors) is applied. The typical permeability accuracy is represented by the sum of error due to temperature coefficient (Tc), hysteresis error (Tc ×∆T
max
) and the accuracy at 23 °C ±5 °C.
µ
r
´ accuracy =
µ´
rm
:
µ´
rm
±(Eµ+ Eh+ Ei) [%]
Loss tangent accuracy of µ
r
(= tanδ) :
±
(Eµ+ Eh+ Ei)
100
where,
E
µ
= Permeability measurement accuracy
at 23 °C ± 5 °C
E
h
= Tc×∆T
E
i
= Tc×∆T
max
× 0.3
T
c
= K4+ K5+ K
6
See Figure 57 through Figure 59 for the calculated value of T
c
without temperature compensation
K
4
=1 × 10-6× (60 + 150 × ƒ)
K
5
=
1 × 10-2× (1 + 10 × ƒ) ×
|1–0.01 × {F × (µ´rm–1) +10} × ƒ2|
{F × (µ´rm–1) + 20} × ƒ
K
6
=
2 × 10-6× (0.3 + 3 × ƒ) ×
{F × (µ´
rm
–1) + 20} × ƒ
|1–0.01 × {F × (µ´
rm
–1) +10} × ƒ
2
|
with temperature compensation
K
4
=1 × 10-6× (60 + 150 × ƒ)
K
5
=
1 MHz ≤ ƒ < 500 MHz
1 × 10-2× (1 + 10 × ƒ) ×
|1–0.01 × {F × (µ´rm–1) +10} × ƒ2|
{F × (µ´rm–1) + 20} × ƒ
500 MHz ƒ ≤ 1 GHz
1 × 10-2× (5 + 2 × ƒ) ×
|1–0.01 × {F × (µ´rm–1) +10} × ƒ2|
{F × (µ´rm–1) + 20} × ƒ
Typical accuracy of permeability parameters:
.
37
K
6
= 1 MHz ≤ ƒ < 500 MHz
2 × 10-6× (0.3 + 3 × ƒ) ×
{F × (µ´rm–1) + 20} × ƒ
|1–0.01 × {F × (µ´
rm
–1) +10} × ƒ
2
|
500 MHz ƒ ≤ 1 GHz
2 × 10-6× (1.5 + 0.6 × ƒ) ×
{F × (µ´rm–1) + 20} × ƒ
|1–0.01 × {F × (µ´rm–1) +10} × ƒ
2
|
ƒ = Measurement frequency [GHz]
F = hln c[mm]
b
h = Height of MUT (material under test) [mm]
b = Inner diameter of MUT [mm]
c = Outer diameter of MUT [mm]
µ´ = Measured value of µ´
r
T = Difference of measurement temperature from
calibration temperature
T
max
= Maximum temperature change (°C) at test
port from calibration temperature after the calibration is performed.
Typical accuracy of permeability parameters (continued):
38
Figure 58. Typical frequency characteristics of temperature coefficient
of µ'r(at F = 3)
Figure 57. Typical frequency characteristics of temperature coefficient
of µ'r(at F = 0.5)
Figure 59. Typical frequency characteristics of temperature coefficient
of µ'r(at F = 10)
39
By internet, phone, or fax, get assistance with all your test & measurement needs
Phone or Fax United States:
(tel) 800 452 4844
Canada:
(tel) 877 894 4414 (fax) 905 282 6495
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115
Latin America:
(tel) (305) 269 7500 (fax) (305) 269 7599
Taiwan: (tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6836 0252 Email: tm_asia@agilent.com
Online Assistance:
www.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
© 2003 Agilent Technologies, Inc. Printed in USA August 13, 2003 5980-1233E
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Agilent T&M Software and Connectivity
Agilent’s Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit
www.agilent.com/find/connectivity
for more information.
Agilent Email Updates
Loading...