Agilent Technologies
PSA Series Spectrum Analyzers
Option 015
6 GHz Return Loss Measurement Accessory Kit
Agilent Part Number E4440-90615 Supercedes E4440-90295
Printed in USANovember 2007
Page 2
Notice.
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited
to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies
shall not be liable for errors contained herein or for incidental or consequential damages in connection with
the furnishing, performance, or use of this material.
Option 015 Accessory Kit has been assembled to ensure that you have the necessary
accessories to set up return loss and loss/gain measurements. These accessories can be used to
make measurements with the PSA using option 215. The option 215 software allows you to
synchronize the tracking between an external source and the spectrum analyzer.
Because your device under test and system configuration are unique, you must provide any
additional connectors, adaptors, and cables to make the final connections to your equipment.
This document describes common measurement and normalization configurations for these
measurements. For more information about measurements and normalization, refer to the
External Source Control Personality Guide PSA Series Spectrum Analyzers Option 215.
Option 015 Accessory Kit Contents
The Option 015 Accessory Kit contains the devices listed below.
DescriptionSpecificationsQuantityPart Number
RF Bridge 300 kHz to 6 GHz
Type N
50 ohm
Power DividerDC to 18 GHz
Type N
50 ohm
Coaxial TerminationDC to 18 GHz
Type N(m)
50 ohm
Coaxial ShortDC to 18 GHz
Type N(m)
Coaxial Cable16 inches
Type N-N
50 ohm
6 dB Coaxial AttenuatorDC to 18 GHz
Type N(m)
50 ohm
186205A
111636A
1909A Option 012
111512A
28121-0101
20955-2031
User’s Guide E4440-90615 3
Page 4
Return Loss Measurement Setup
Purpose
To make return loss measurements using the Option 015 Accessory Kit, set up the equipment
as shown below.
Measurement Normalization Setup
Connect the signal generator output, RF bridge, coaxial short (or open), 6 dB attenuator
(optional), and the spectrum analyzer RF input as shown in Figure 1. Leave port 1 of the RF
bridge open for the initial setup.
For more information on measurement normalization, refer to the External Source Control Personality Guide, PSA Series Spectrum Analyzers Option 215.
Figure 1 Normalizing the Measurement
4 User’s Guide E4440-90615
Page 5
Measurement Setup
Connect the signal generator output, RF bridge, device under test (DUT), 50 ohm load (if the
device is a two-port device), 6 dB attenuator (optional), and the spectrum analyzer RF input
as shown in Figure 2. For more measurement information, refer to the External Source Control Personality Guide, PSA Series Spectrum Analyzers Option 215.
Figure 2 Measurement Setup
User’s Guide E4440-90615 5
Page 6
Loss/Gain Measurement Setup
Purpose
To make loss/gain measurements using the Option 015 Accessory Kit, set up the equipment as
shown below.
Measurement Normalization Setup
Connect the signal generator output, 6 dB attenuator (optional), and the spectrum analyzer
RF input as shown in Figure 3. For more information on measurement normalization, refer to
the External Source Control Personality Guide, PSA Series Spectrum Analyzers Option 215.
Figure 3 Normalizing the Measurement
6 User’s Guide E4440-90615
Page 7
Measurement Setup
Connect the signal generator output, device under test, 6 dB attenuator (optional), and the
spectrum analyzer RF input as shown in Figure 4. For more measurement information, refer
to the External Source Control Personality Guide, PSA Series Spectrum Analyzers Option 215.
Figure 4 Measurement Setup
User’s Guide E4440-90615 7
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.