Agilent E4445A Getting Started Guide

Page 1

Getting Started Guide

PSA Series Spectrum Analyzers
This manual provides documentation for the following instruments:
E4440A (3 Hz - 26.5 GHz) E4443A (3 Hz - 6.7 GHz) E4445A (3 Hz - 13.2 GHz) E4446A (3 Hz - 44 GHz) E4447A (3 Hz - 42.98 GHz) E4448A (3 Hz - 50 GHz)
Manufacturing Part Number: E4440-90608
Supersedes: E4440-90332
Printed in USA
June 2008
© Copyright 2001-2008 Agilent Technologies, Inc.
Page 2
Notice
The information contained in this document is subject to change without notice.
Agilent T echnologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this analyzer.
WARNING Warning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of the analyzer. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
NOTE Note calls out special information for the user’s attention. It provides
operational information or additional instructions of which the user should be aware.
General Safety Information
The following general safety precautions must be observed during all phases of operation. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
WARNING This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited. (IEC 348 clauses
2
Page 3
17.3.3c & 17.3.4)
CAUTION This product is designed for use in Installation Category II and Pollution Degree 2
per IEC 61010 Second Edition and IEC 664 respectively.
WARNING If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
WARNING No operator serviceable parts inside. Refer servicing to qualified personnel.
To prevent electrical shock do not remove covers.
CAUTION The Mains wiring and connectors shall be compatible with the
connector used in the premise electrical system. Failure, to ensure adequate earth grounding by not using the correct components may cause product damage, and serious injury
Additional Information
For the latest information about this analyzer, including firmware upgrades, application information, and product information, see the following URL:
http://www.agilent.com/find/psa/
3
Page 4
4
Page 5
Contents
1. Installation and Setup
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
AC Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Instrument Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Cleaning the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Battery Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Turning on the Analyzer for the First Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Why Aren’t All the Personality Options Loaded in Memory?. . . . . . . . . . . . . . . . . . . . . . . . 15
If Option 117 is Loaded in Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Using an External Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Firmware Revision. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Running Internal Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Printer Setup and Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Protecting Against Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

T a ble of Contents

2. Front and Rear Panel Features
Front Panel Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Front-Panel Connectors and Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Display Annotations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Rear-Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Key Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Front and Rear Panel Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3. Making a Basic Measurement
Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Entering Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Using Menu Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Presetting the Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Creating a User Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Viewing a Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4. Viewing Catalogs and Saving Files
File Menu Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Locating and viewing files in the catalog. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Creating a directory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Saving a File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Step 1. Set up the analyzer trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Step 2. Save the file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Loading a file. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Renaming a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Copying a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Deleting a File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Using the Alpha Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5
Page 6
Contents
Table of Contents
5. Options and Accessories
Ordering Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
50 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
50 /75 Minimum Loss Pad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
AC Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
AC Probe (Low Frequency) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Close-Field Probes (for EMC measurements with PSA) . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
Broadband Preamplifiers and Power Amplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
GPIB Cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
USB/GPIB Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
HP/Agilent 11970 Series Harmonic Mixers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
HP/Agilent 11974 Series Preselected Millimeter Mixers. . . . . . . . . . . . . . . . . . . . . . . . . . . .71
RF and Transient Limiters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
Power Splitters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
RF Bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
50 Transmission/Reflection Test Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Static Safe Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Noise Sources (for use with PSA Option 219). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
6. In Case of Difficulty
Types of Spectrum Analyzer Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Before Calling Agilent Technologies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Check the Basics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
Read the Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Service Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Calling Agilent Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Returning an Analyzer for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
6
Page 7
Installation and Setup

1 Installation and Setup

7
Page 8
Agilent Technologies Performance
Installation and Setup
Spectrum Analyzer (PSA)
Installation and Setup
This chapter provides the following information that you may need when you first receive your spectrum analyzer:
“Initial Inspection” on page 9
“Power Requirements” on page 11
“Turning on the Analyzer for the First Time” on page 14
“Running Internal Alignments” on page 18
“Printer Setup and Operation” on page 19
“Protecting Against Electrostatic Discharge” on page 21
“Safety Information” on page 22
8 Chapter 1
Page 9
Installation and Setup

Initial Inspection

Initial Inspection
Inspect the shipping container and the cushioning material for signs of stress. Retain the shipping materials for future use, as you may wish to ship the analyzer to another location or to Agilent Technologies for service. Verify that the contents of the shipping container are complete. The following table lists the items shipped with the analyzer.
Item Description
Accessories
IntuiLink Toolbar software, CD-ROM Provides a set of connectivity tools that enable you to easily
move data from your analyzer to your PC. Power Cable Connection for power source. USB Cable (Option 111 only) Connects PSA USB to USB on a PC, part number 8121-1074. Adapter (E4446A, E4447A, and E4448A
only) Adapter (E4446A, E4447A, and E4448A
only) Agilent Automation-Ready CD Assists in quickly and properly connecting instruments to
Standard Documentation Set
Getting Started Guide Covers unpacking and setting up the analyzer, analyzer
User’s and Programmer ’s Reference Describes analyzer features in detail, including front-panel
2.4 mm(f) to K(f) front panel connector adapter, part number
1250-2187.
2.4 mm(f) to 2.4 mm(f) front panel connector adapter, part
number 1250-2188.
your PC, part number E2094-60003. Includes: Agilent IO
Libraries Suite, Agilent VEE Pro, T&M Tool Kit, and Agilent
IO Hardware information.
features, and how to make a basic measurement. Includes
information on options and accessories, and what to do if you
have a problem.
key descriptions, basic spectrum analyzer programming
information, and SCPI command descriptions.
Installation and Setup
Measurement Guide Provides details on how to measure various signals, and how
to use catalogs and files. Specifications Guide Documents specifications, safety, and regulatory information. Instrument Messages and Functional Tests Includes instrument messages (and suggestions for
troubleshooting them), and manual functional tests. Programming Conversion Guide Describes SCPI programming command compatibility for
8590, 8560, 8566, 8568, ESA/PSA series analyzers. Documentation CD-ROM Includes the documents in the standard set (listed above).
You can view and print the information as needed. See the
CD-ROM jacket for installation information.
Chapter 1 9
Page 10
NOTE If you purchased one or more optional measurement personalities, the
Installation and Setup
Installation and Setup
Initial Inspection
related guides for the options you ordered are included. Service documentation is not included in the standard documentation
set. See “Options” on page 63 for information on ordering.
If There Is a Problem
If the shipping materials are damaged or the contents of the container are incomplete:
• Contact the nearest Agilent Technologies office to arrange for repair or replacement (see “Contacting Agilent Technologies” on page 78). You will not need to wait for a claim settlement.
• Keep the shipping materials for the carrier’s inspection.
• If you must return an analyzer to Agilent Technologies, use the original (or comparable) shipping materials (see “Returning an
Analyzer for Service” on page 79).
10 Chapter 1
Page 11
Installation and Setup

Power Requirements

Power Requirements
The only physical installation of your Agilent spectrum analyzer is a connection to a power source.
Line voltage does not need to be selected. This analyzer does not contain customer serviceable fuses.
WARNING Failure to ground the analyzer properly can result in personal
injury. Before turning on the analyzer, you must connect its protective earth terminals to the protective conductor of the main power cable. Insert the main power cable plug into a socket outlet that has a protective earth contact only. DO NOT defeat the earth-grounding protection by using an extension cable, power cable, or autotransformer without a protective ground conductor.
Installation and Setup
If you are using an autotransformer, make sure its common terminal is connected to the protective earth contact of the power source outlet socket.
This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
WARNING To prevent electrical shock, disconnect the Agilent spectrum
analyzer from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
CAUTION VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
This analyzer has autoranging line voltage input. Be sure the supply voltage is within the specified range.
NOTE For detailed analyzer specifications, see the Specifications guide.
Chapter 1 11
Page 12
Installation and Setup
Power Requirements
T able 1-1. Power Requirements
Installation and Setup
NOTE The front panel switch is a standby switch only; it is not a LINE switch
Description
Voltage, Frequency 100 to 120 Vrms, 50/60/400 Hz
220 to 240 Vrms, 50/60 Hz
Power Consumption, On Base Fully Loaded
< 260 W < 450 W
Power Consumption, Standby < 20 W
Specification
a
a. For greater detail, refer to the Specifications Guide.

AC Power Cord

The analyzer is equipped with a three-wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.
(power disconnecting device).
WARNING Install the product so that the detachable power cord is readily
identifiable and easily reached by the operator . The detac hable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front panel switch is only a standby switch and is not a LINE switch. Alternatively, an externally installed switch or circuit breaker (which is readily identifiable and is easily reached by the operator) may be used as a disconnecting device.
CAUTION Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can cause product damage.
12 Chapter 1
Page 13

Instrument Maintenance

Instrument Maintenance

Cleaning the Instrument

WARNING To prevent electrical shock, disconnect the Agilent
Technologies E4440A Series from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Installation and Setup

Connector Care

CAUTION Cleaning connectors with alcohol shall only be done with the
instruments power cord removed, and in a well-ventilated area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument.

Battery Information

The analyzer uses a Lithium Polycarbon Monofloride battery to power the analyzer clock. The battery is located on the CPU board.
You can order the service documentation for Agilent spectrum analyzers through your Agilent Sales and Service office. The documentation is described under “Options” on page 63.
NOTE If the analyzer’s clock does not keep time when powered off, the
problem is the battery. See “Returning an Analyzer for Service” on
page 79.
WARNING Danger of explosion if battery is incorrectly replaced. Replace
only with the same or equivalent type recommended. Discard used batteries according to the manufacturer’s instructions.
Installation and Setup
Chapter 1 13
Page 14
WARNING If this product is to be energized through an external auto
NOTE Do not connect anything else to the analyzer yet.
NOTE The instrument requires >2 minutes to complete the start-up routine.
Installation and Setup
Information Screen
Installation and Setup

Turning on the Analyzer for the First Time

Turning on the Analyzer for the First Time
Plug in the power cord.
transformer for voltage reduction, make sure that its common terminal is connected to a neutral (earthed pole) of the power supply.
Press the power switch (located in the lower left corner of the
analyzer’s front panel) to turn the analyzer on. See “Front Panel
Overview” on page 24.
An information screen appears during the initialization process. The information screen contains the analyzer product number and a URL for accessing product support information on the Web. See “Additional
Information” on page 3.
NOTE The information screen displays for approximately 10 seconds before
the initialization process is complete. Record the firmware revision and serial number, and keep it for
reference. If you should ever need to call Agilent Technologies for service or with any questions regarding your analyzer, it will be helpful to have this information readily available. You can also obtain the firmware revision and serial number by pressing
System.
System, More, Show
If using LAN, set the IP address of the analyzer to an appropriate
number for your network; one that the network recognizes, but that is not yet in use. (Consult your local IT group):
—Press — If the current address is not appropriate, press
System, Config I/O, and note the IP address.
IPAddress and use
the keypad to change it.
— Verify the subnet mask and gateway values. If necessary, change
the values.
— Connect the LAN cable to the LAN connector located on the rear
panel of your analyzer (see “Rear-Panel Features” on page 29).
— Cycle the analyzer power.
14 Chapter 1
Page 15
Installation and Setup
Turning on the Analyzer for the First Time
NOTE It is necessary to cycle the power to the analyzer after plugging in the
LAN for the analyzer to recognize the network. Allow the spectrum analyzer to warm-up for 30 minutes before
making a calibrated measurement. To meet its specifications, the analyzer must meet operating temperature conditions.
NOTE It is normal to hear clicking when the Auto Alignment function is on.
See “Running Internal Alignments” on page 18 for more information.

Why Aren’t All the Personality Options Loaded in Memory?

Many measurement personality options are available for use with this instrument. If the option is loaded in the instrument, you must also have a license key entered, to use it.
Some versions of instrument hardware may not have enough memory to accommodate all the options that you have ordered. If this is the case you will need to swap the applications in and out of memory, as needed. It may also be possible to upgrade your hardware to have more memory. Contact your local sales or service office.

If Option 117 is Loaded in Memory

If you ordered Option 117, the Secure Memory Erase option was loaded into the memory of your instrument and licensed at the factory but not enabled. DO NOT enable this option until you have fully configured your instrument (set the removed options and licenses, and configured the power-up state you desire). After the security mode is enabled, the main Flash memory becomes read-only, rather than read/write memory. This makes it impossible to add or remove personalities or their license keys, or otherwise change the instrument’s configuration, unless you do a
System, Security, Secure Erase All procedure. This procedure will render
the instrument inoperable. You will then need to run the upgrade procedure (on the Agilent Web site) to reinstall the firmware and completely reconfigure your instrument. Refer to “Managing Security” in Chapter 4 of the “User’s and Programmer’s Reference, Volume 1.”
System, Config I/O settings as needed, added or
Installation and Setup
When you are satisfied with the configuration of your PSA analyzer, enable the Secure Memory Erase option by pressing
Security, Enabled
. The security mode will then be enabled the next time
System, Security,
the analyzer’s power is cycled.
NOTE When Option 117 is present, the number of personalities that can be
loaded into memory is limited to 3 or 4.
Chapter 1 15
Page 16
Installation and Setup
Turning on the Analyzer for the First Time

Using an External Reference

1. To use an external frequency reference, connect it to the EXT REF IN connector on the rear panel (see “Rear-Panel
Features” on page 29).
2. Enter the frequency of the external reference into the analyz er:
Installation and Setup
a. Press b. Select the
System, Reference, Freq Ref
Ext function.
c. Use the keypad to enter the frequency of the external frequency
reference.
16 Chapter 1
Page 17
Installation and Setup

Firmware Revision

Firmware Revision
To view the firmware revision of your analyzer, press System, More,
Show System. If you call Agilent Technologies regarding your analyzer,
it is helpful to have this revision and the analyzer serial number available.
TIP You can get automatic electronic notification of new firmware releases
and other product updates/information by subscribing to the Agilent Technologies Test & Measurement E-Mail Notification Service for the
PSA Series at http://www.agilent.com/find/notifyme
Installation and Setup
Chapter 1 17
Page 18
Installation and Setup

Running Internal Alignments

Running Internal Alignments
Each time the analyzer is powered on, the internal alignment routine runs automatically.
The analyzer was shipped from the factory with the Alignments mode set to Auto. This setting enables the alignment routine to run automatically either every 24 hours, or when the internal analyzer temperature changes by ±3°C.
NOTE When the Alignment routine runs, you will hear the attenuator settings
changing, which generates noise. This is not an indication of trouble.
Manually Performing an Alignment
If
Auto Align, Off is selected, refer to the Specifications guide for the
conditions required to maintain calibration.
Installation and Setup
NOTE For detailed information regarding the alignment routine (including
how to run alignments, and how the analyzer can alert you when alignments are required) refer to the User’s guide for your analyzer.
18 Chapter 1
Page 19
Installation and Setup

Printer Setup and Operation

Printer Setup and Operation
A printer can be connected to your analyzer if it is equipped with an external I/O interface. Supported printers accept Hewlett-Packard Printer Control Language Level 3 (PCL3) or 5 (PCL5). Refer to the documentation or specifications supplied with your printer, or contact the manufacturer to identify your printer’s language.
Equipment
• IEEE 1284 compliant printer cable.
• Supported and tested printers are listed below. There are many PCL3/5 printers that may work with your analyzer, however, they have not been tested.
— PCL3 printers include most HP DeskJet printers. — PCL5 printers include most HP LaserJet printers.
NOTE The following printers are not compatible with your analyzer.
• HP Deskjet 720C, 722C, 820C and 1600C
• Epson MX-80, FX-85, Stylus, and LQ-570
Printer Models Language
Type
HP DeskJet 310, 320, 350C, 400L PCL3 yes HP DeskJet 500C, 550C, 600, 660C, 672C,
680C, 682C, 690C, 693C HP DeskJet 840C, 850C, 870C, 890C, 895C PCL3 yes HP DeskJet 935C, 970C, 990C PCL3 yes
PCL3 yes
Installation and Setup
Color
Capable
HP DeskJet 1120C, 1150C PCL3 yes HP Inkjet 2000C PCL 3 yes HP LaserJet 4L, 4P, 4 Plus PCL3/5 no HP LaserJet 5L, 5M, 5N, 5P, 5SI PCL3/5 no HP LaserJet 6L, 6MP, 6P PCL5 no HP LaserJet 2100 Series, PCL3/5 no HP LaserJet 4050N PCL3/5 yes HP LaserJet 5000GN PCL3/5 yes HP Professional Series 2500CM PCL3 yes HP Professional Series 2500CM PCL3 yes
Chapter 1 19
Page 20
Installation and Setup
Printer Setup and Operation
Interconnection and Setup
1. Turn off the printer and the analyzer.
2. Using an IEEE 1284 compliant parallel printer cable, connect the printer to the analyzer parallel I/O interface connector on the rear panel (see “Rear-Panel Features” on page 29).
3. Turn on the analyzer and printer.
Installation and Setup
NOTE There may be some small discrepancies in the color mapping of the
4. On the front panel, press
Print Setup, then press the Printer Setup
menu key.
5. Select the printer language and color capability appropriate for your printer. See the table above this section for a list of some compatible printers.
Testing Printer Operation
When you have completed the printer setup for the analyzer, press the front panel
Print key. If the printer is ready and the printer setup was
successful, a printout of the analyzer display will be printed. If the printer is not ready, the message “Printer not responding” appears at the bottom of the screen.
analyzer display to your color printer. Due to differences in display and printer technologies, the default display colors do not map exactly to the printer colors. For example trace 1 is yellow on your analyzer display while it maps to green on your printer.
20 Chapter 1
Page 21
Installation and Setup

Protecting Against Electrostatic Discharge

Protecting Against Electrostatic Discharge
Electrostatic discharge (ESD) can damage or destroy electronic components (the possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used).
Test Equipment and ESD
To help reduce ESD damage that can occur while using test equipment:
• Before connecting any coaxial cable to an analyzer connector for the first time each day, momentarily short the center and outer conductors of the cable together.
• Personnel should be grounded with a 1 M resistor-isolated wrist-strap before touching the center pin of any connector and before removing any assembly from the analyzer.
• Be sure that all instruments are properly earth-grounded to prevent build-up of static charge.
Installation and Setup
WARNING Do not use these first three techniques when working on
circuitry with a voltage potential greater than 500 volts.
• Perform work on all components or assemblies at a static-safe workstation.
• Keep static-generating materials at least one meter away from all components.
• Store or transport components in static-shielding containers.
• Always handle printed circuit board assemblies by the edges. This reduces the possibility of ESD damage to components and prevent contamination of exposed plating.
For information on ordering static-safe accessories, see “Accessories” on
page 69.
Additional Information about ESD
For more information about ESD and how to prevent ESD damage, contact the Electrostatic Discharge Association (http://www.esda.org). The ESD standards developed by this agency are sanctioned by the American National Standards Institute (ANSI).
Chapter 1 21
Page 22
Installation and Setup

Safety Information

Safety Information
WARNING This is a Safety Class 1 Product (provided with a protective
earthing ground incorporated in the power cord). The mains plug shall be inserted only in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
Installation and Setup
22 Chapter 1
Page 23

2 Front and Rear Panel Features

This chapter gives you an overview of the front and rear panels of your analyzer. For details on analyzer keys and remote programming, refer to the User’s and Programmer’s Reference. For connector specifications (including input and output levels), see the Specifications guide.
Front and Rear Panel Features
23
Page 24
Front and Rear Panel Features

Front Panel Overview

Front Panel Overview
This section provides information on the analyzer’s front panel, including:
• Front Panel Connectors and Keys, see below
“Display Annotations” on page 26

Front-Panel Connectors and Keys

Front and Rear Panel Features
Item
Description
# Name
1
AUDIO INPUT High-impedance input for audio signal path (Option 107). Maximum input: 7 Vrms,
20 Vdc.
Softkeys Menu labels identifying the current function of each softkey appear to the left of each
2
key. Softkey menus depend on the active menu. Also see “Using Menu Keys” on page
37.
ESC Key Exits any function without modifying the current parameters. Also blanks the active
3
function annotation and aborts the power-on alignments.
4
Control Keys Set parameters used for making measurements in the current measurement mode.
5
Measure Keys Select and set up specific measurements within the current application. Mode Keys Select measurement mode and mode parameters.
6
24 Chapter 2
Page 25
Front and Rear Panel Features
Front Panel Overview
Item
Description
# Name
7 System Keys Access features used with all analyzer modes and affect the state of the entire spectrum
analyzer.
8 Vol. Control/
Earphones
9
PROBE PWR Supplies power for external high frequency probes and accessories (see “Accessories” on
Not currently implemented.
page 69).
10
Marker Keys Enable markers to obtain specific information about the displayed measurement.
EXT TRIGGER
11
INPUT
12
POWER ON/OFF
Enables you to externally trigger measurements.
Turns the analyzer on. A green light indicates power on. A yellow light indicates standby mode.
NOTE The front-panel switch is a standby switch, not a LINE switch
(disconnecting device); the analyzer continues to draw power even when the line switch is in standby. Use the detachable power cord to disconnect the analyzer from the mains supply.
Help Key Press the Help key, then any other key to get a short description of that key and the
13
associated SCPI command. The next key you press will remove the help window from the display.
Window Keys Next Window: On displays with multiple windows, changes the highlighted window that
14
is currently active.
Zoom: Zooms in on the highlighted window.
Front and Rear Panel Features
15
Navigation Keys
16
Return Key Exits the current menu and returns to the previous menu. Data Controls Change the numeric value of an active function. Entries appear in the active function
17
Move cursor between fiel ds on the display.
area of the display. Also see “Entering Data” on page 37.
Floppy
18
Disk Drive
19
RF Input Input for an external signal. Make sure that the total power of all signals at the
Accepts a 3.5 inch 1.44 MB floppy disk.
analyzer input does not exceed +30 dBm (1 watt). The E4446A, E4447A, and E4448A input connector is 2.4 mm.
20 IF Input, 1st LO
Output
Allows connections for external mixing (Option AYZ)
Chapter 2 25
Page 26
Front and Rear Panel Features
Front Panel Overview

Display Annotations

Front and Rear Panel Features
Ite
Description Associated Function Keys
m
1
Detector type
a
Det/Demod, Detector
2 Reference level Amplitude, Ref Level 3 Active function block Refer to the description of the activated function. 4 Screen title
Display, Title
5 Time and date display System, Time/Date 6
RF attenuation
a
Amplitude, Attenuation
7 Display line Display, Display Line 8Marker Marker 9 Active marker with frequency and
Marker
amplitude displayed
26 Chapter 2
Page 27
Front and Rear Panel Features
Front Panel Overview
Ite
Description Associated Function Keys
m
10 Remote operation annunciators
System [LOCAL] Also see programming documentation
O - indicates USB connected; flashing green center indicates bus activity
R - remote operation L - GPIB listen T - GPIB talk S - GPIB SRQ
11 Data invalid indicator Asterisk (*) means some or all trace data may not match
annotation due to possible analyzer setting changes.
12 Status message area Displays status messages (see “Types of Spectrum
Analyzer Messages” on page 75).
13 Key menu title Dependent on menu selection. 14 Softkey menu Softkey labels 15 Frequency span or stop frequency 16
Sweep time
17
Video bandwidth
a
a
Span or Frequency, Stop Freq
Sweep, Sweep Time
BW, Video BW
18 Frequency offset Frequency, Freq Offset 19 Display status line Displays informational and error messages (see “Types of
Spectrum Analyzer Messages” on page 75).
20
Resolution bandwidth
a
BW, Res BW
21 Center frequency or start frequency Frequency, Center Freq or Start Freq 22 Internal preamp Amplitude, Int Preamp 23 Auto alignment routine status System, Alignments, Auto align
(If auto alignment is off, no indicator appears)
AA: auto alignment of all
analyzer parameters.
AL: analyzer alerts you when alignment is needed.
24 Trigger or Trace
Trig or Trace/Avg
Trigger & sweep modes:
F: free-run trigger
X: external (front) trigger
R: external (rear) trigger
L: line trigger
V: video trigger
C: continuous sweep
S: single sweep
Trace modes:
W: clear-write S: store blank m: minimum hold M: maximum hold V: view 1: trace 1 2: trace 2 3: trace 3
Front and Rear Panel Features
Chapter 2 27
Page 28
Front and Rear Panel Features
Front Panel Overview
Ite
Description Associated Function Keys
m
25
A verage
a
Trace/Avg, Average: The
number shown indicates current average number.
VAvg: voltage average LgAv: log power average PAvg power average
26 Display line annotation Display, Display Line 27 Amplitude offset Amplitude, Ref Lvl Offst 28 Amplitude scale Amplitude, Scale Type or Amplitude, Scale/Div 29
Phase noise optimization
a
Auto Couple, PhNoise Opt
Phase noise can be optimized for:
f<50k: frequencies less than 50 kHz from carrier. f>50k: frequencies greater than 50 kHz from carrier. FTun: LO for fast tuning.
30
Sweep type
31 Signal track
Could also be CP (Continuous Peak)
a
Sweep, Sweep Type Frequency, Signal Track
Peak Search
, Continuous Pk
a. This function can be auto-coupled. A pound ( #) symbol in front of this annotation
indicates an uncoupled function.
Front and Rear Panel Features
28 Chapter 2
Page 29

Rear-Panel Features

Front and Rear Panel Features
Rear-Panel Features
Item Description
# Name
1 NOISE SOURCE
DRIVE OUT +28 V (PULSED)
2 PRESEL TUNE
OUT
3
10 MHz OUT (SWITCHED)
EXT REF IN Input for a 1 to 30 MHz external frequency reference signal.
4
5
321.4 MHz IF Out
6
TRIGGER 1 OUT A trigger output used to synchronize other test equipment with the analyzer. Trigger
Provides 28 V to drive a noise source. Used on Option 219 Noise Figure personality.
Outputs tuning voltage for preselected harmonic mixer.
A switched output of the analyzer’s internal 10 MHz frequency reference signal used to lock the frequency reference of other test equipment to the analyzer.
Provides 321.4 MHz IF output. Not available on E4447A.
1 is the High=Sweeping (HSWP) signal.
7
TRIGGER 2 IN Allows external triggering of measurements.
8
VIDEO OUT Provides a y-axis video out signal. Note that this signal is not available when gating
is being used (Option 124).
9
21.4 MHz IF OUT Provides 21.4 MHz IF output (Option HB2). Not available on E4447A.
Or, provides 21.4 MHz IF out Option HYX, which is available for all PSA models including E4447A.
Front and Rear Panel Features
USB Port Provides USB interface (Option 111). USB device side I/O only.
10
Chapter 2 29
Page 30
Front and Rear Panel Features
Rear-Panel Features
Item Description
# Name
11 KYBD Enables connection of an external PS-2 keyboard using a 6-pin mini-DIN connector.
Always turn off power before plugging in keyboard.
This feature not fully implemented. Currently used only for firmware upgrades.
12 Line power
input
13
GPIB Allows the connection of a General Purpose Interface Bus (GPIB, IEEE 488.1) cable,
The AC power connection. Also see “Power Requirements” on page 5.
which enables remote analyzer operation.
14
LAN A TCP/IP Interface.
Also see “Turning on the Analyzer for the First Time” on page11 for information on setting the IP address.
For information on using the analyzer remotely, refer to the User’s and Programmer’s Reference.
TRIGGER 2 OUT A trigger output used to synchronize other test equipment with the analyzer.
15
Reserved for future applications.
16
PARALLEL Supports remote printing (Printing is supported only from the parallel port).
17
70 MHZ IF OUT Provides 70 MHz IF output (Option H70). Not available on E4447A.
Other IF output options may occupy this rear-panel location.
18
MONITOR Allows connection of an external VGA monitor. RS-232 Allows connection of a PC, printer, or modem.
19
This feature is not fully implemented.
20
HNQ IN H7L IN H70 IN
These features are only available when special options are installed.
Requires a BNC jumper cable from the rear-panel “321.4 MHz IF Out” to one of these inputs, depending on the option. Not available on E4447A.
Front and Rear Panel Features
30 Chapter 2
Page 31
Front and Rear Panel Features

Key Overview

Key Overview
The keys labeled FREQUENCY Channel, System, and Marker are all examples of front-panel keys. The front-panel keys are dark gray, light gray, green, or white in color. Front-panel keys that are white perform an immediate action rather than bringing up a menu. The only green key is the all front panel keys and their related menu keys can be found in user’s guide for your analyzer). Pressing most of the dark or light gray front-panel keys accesses menus of functions that are displayed along the right side of the display. These are called menu keys.
Menu keys list functions other than those accessed directly by the front panel keys. To activate a menu key function, press the key immediately to the right of the annotation on the screen. The menu keys that are displayed depend on which front-panel key is pressed and which menu level is enabled.
If a menu key function’s value can be changed, it is called an active function. The function label of the active function is highlighted after that key has been selected. For example, press calls up the menu of related amplitude functions. Note the function labeled highlighted. indicating that it is the active amplitude function and can now be changed using any of the data entry controls.
Preset key, which performs an analyzer reset (A summary of
AMPLITUDE Y Scale. This
Ref Level (the default selected key in the Amplitude menu) is
Ref Level also appears in the active function block,
Front and Rear Panel Features
A menu key with On and Off in its label can be used to turn the menu key’s function on or off. To turn the function on, press the menu key so that On is underlined. To turn the function off, press the menu key so that Off is underlined. In the manual, when On should be underlined, it will be indicated as
Function (On).
A function with Auto and Man in the label can either be auto-coupled or have its value manually changed. The value of the function can be changed manually using the numeric keypad, knob, or step keys. To auto-couple a function, press the menu key so that Auto is underlined.
In some key menus, one key label will always be highlighted to show which key has been selected. For example, when you press
Marker, you
will access a menu of keys in which some of the keys are grouped together by a blue bar on the left side of the menu. The which is the
Marker menu default key, will be highlighted. When you
press another key within the blue bar region, such as
Normal key,
Delta, the
highlight will move to that key to show it has been selected.
Chapter 2 31
Page 32
Front and Rear Panel Features
Key Overview
In other key menus, one key label will always be highlighted to show which key has been selected but the menu is immediately exited when a selection is made. For example, when you press the the
Print Setup menu), it will bring up its own menu of keys. The Portrait
Orientation key (on
key, which is the Orientation menu default key, will be highlighted. When you press the to show it has been selected and the screen will return to the
Landscape key, the highlight will move to that key
Print Setup
menu. The arrow keys located below the analyzer display (sometimes referred
to as Tab keys) can be used to navigate within tables, for example the Limit-Line table. These keys are used to move between rows. The Left-arrow key moves up, while the right-arrow key moves down. While navigating through the table, the cursor (inverse video highlight) stays in the same column. Navigating left or right in the table is accomplished by choosing the desired field using the front-panel keys.
Front and Rear Panel Features
32 Chapter 2
Page 33

Front and Rear Panel Symbols

This symbol is used to indicate power ON (green LED).
This symbol is used to indicate power STANDBY mode
(yellow LED).
This symbol is used to indicate the ON position of the
power button.
This symbol is used to indicate the OFF position of the
power button.
This symbol indicates the input power required is AC.
The instruction documentation symbol. The product is
marked with this symbol when it is necessary for the user to refer to instructions in the documentation.
Front and Rear Panel Features
Front and Rear Panel Symbols
The CE mark is a registered trademark of the
European Community. This product complies with all relevant directives.
The C-Tick mark is a registered trademark of the
Australian/New Zealand Spectrum Management Agency. This product complies with all relevant EMC regulations.
This is a marking of an Industrial Scientific and
Medical Group 1 Class A product, and to indicate product compliance with the Canadian Interference-Causing Equipment Standard (ICES-001).
The CSA mark is a registered trademark of CSA
International. This product complies with all relevant safety requirements.
Front and Rear Panel Features
Chapter 2 33
Page 34
Front and Rear Panel Features
Front and Rear Panel Symbols
This product complies with the WEEE Directive
(2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste.
Product Category: W ith reference to the equipment types in the WEEE Directive Annex I, this product is classed as a "Monitoring and Control instrumentation" product.
Do not dispose in domestic household waste. To return unwanted products, contact your local
Agilent office, or see
http://www.agilent.com/environment/product/
for more information.
Indicates the time period during which no hazardous or
toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
Hipot Indicates that the product passes dielectric strength
hipot test.
Front and Rear Panel Features

Packaging

This symbol on all primary and secondary packaging indicates compliance to China standard GB 18455-2001.
34 Chapter 2
Page 35

3 Making a Basic Measurement

This chapter provides information on basic analyzer operation. For more information on making measurements, see the measurement guide for your analyzer.
Making a Basic Measurement
35
Page 36
Making a Basic Measurement
This chapter is divided into the following sections:
“Using the Front Panel” on page 37
“Presetting the Spectrum Analyzer” on page 38
“Viewing a Signal” on page 39
CAUTION Ensure that the total power of all signals at the analyzer RF input does
not exceed +30 dBm (1 watt).
Basic Assumption
The material in this chapter is presented with the assumption that you understand the front and rear panel layout, and display annotations of your analyzer. If you do not, refer to “Front and Rear Panel Features”
on page 23.
NOTE The display examples in this book are made using various analyzer
models, you may see some variations depending upon your analyzer.
Making a Basic Measurement
36 Chapter 3
Page 37
Making a Basic Measurement
ff
k
z
p
B n
n

Using the Front Panel

Using the Front Panel

Entering Data

When setting measurement parameters, there are several ways to enter or modify the value of the active function:
Knob Arrow Keys Numeric Keypad
Unit Softkeys Enter Key
Increments or decrements the current value. Increments or decrements the current value. Enters a specific value. Then press the desired terminator (either
a unit softkey, or the Terminate a value that requires a unit-of-measurement.
Terminates an entry when either no unit of measure is needed, or you want to use the default unit.
Enter key).

Using Menu Keys

Menu Keys (which appear along the right side of the display) provide access to many analyzer functions. Here are examples of menu key types:
Toggle Allows you to activate or deactivate states.
Example: Toggles the selection (underlined choice)
Submenu Displays a new menu of softkeys.
Signal Trac
On O
each time you press the key.
Example: A submenu key allows you to view a new
Catalog
Choice Allows you to make a selection from a list of values.
Example: A choice key displays the currently selected
Adjust Highlights the softkey and sets the active function.
Examples: Press this type of key and enter a value.
Y Axis Units
CF Ste
13.2550000 GH
Attenuatio
10.00 d
Auto Ma
menu of softkeys related to the submenu key category.
submenu choice, in this example, dBm.
dBm
When the choice is made, the submenu automatically returns.
The default for softkeys with an automatic
Auto) or manual (Man) choice is automatic.
( After you enter a value, the selection changes to manual. You can also press the softkey twice to change to manual.
Making a Basic Measurement
Chapter 3 37
Page 38
Making a Basic Measurement

Presetting the Spectrum Analyzer

Presetting the Spectrum Analyzer
Preset provides a known starting point for making measurements. The analyzer has three types of preset:
Factory Preset Restores the analyzer to its factory-defined state. User Preset Restores the analyzer to a user-defined state. Mode Preset This type of preset restores the currently selected mode
to a known state. For details, see the User’s and Programmer’s Reference. When
triggers a factory preset. When
Preset displays the softkeys Factory Pr eset, User Preset, and Mode Preset
(if applicable); you then select the preset you want. If to
Preset Type is set to Factory, pressing the green Preset key
Preset Type is set to User, pressing
Preset Type is set
Mode, but a personality is not installed, pressing Preset triggers a
factory preset.

Creating a User Preset

If you constantly use settings which are not the factory defaults, use the following steps to create a user-defined preset:
1. Set analyzer parameters as desired.
2. Press
parameters as the user preset state.
3. Press
preset as an option.
Disabling User Preset
Go to the Power On/Preset menu (press select a
System, Power On/Preset, Save User Preset to set the current
Preset to select User in the same softkey menu to enable user
System, Power On/Preset) and
Preset Type of Factory or Mode.
Making a Basic Measurement
38 Chapter 3
Page 39
Making a Basic Measurement

Viewing a Signal

Viewing a Signal
1. Press Preset. If th e softkeys Factory Preset and User Preset appear, select
Factory Preset.
2. Press
System, Reference, then note the 10 MHz Out setting. If Off is
active (underlined), press the softkey to toggle the reference on.
3. Connect the analyzer’s rear panel
10 MHz OUT (SWITCHED) to the
front-panel input.
Setting Reference Level and Center Frequency
NOTE The display examples in this section are made using various analyzer
models, you may see some variations depending upon your analyzer.
4. Set the reference level to 10 dBm: Press
5. Set the center frequency to 30 MHz: Press
Freq
, 30, MHz.
AMPLITUDE, 10, dBm.
FREQUENCY, Center
The 10 MHz reference signal spectrum appears on the display, as shown in Figure 3-1.
Figure 3-1 10 MHz Internal Reference Signal and Associated Spectrum
10 MHz Peak
Chapter 3 39
Making a Basic Measurement
Page 40
Making a Basic Measurement
Viewing a Signal
Setting Frequency Span
6. Set the frequency span to 50 MHz: Press This displays the signal as shown in Figure 3-2.
Figure 3-2 Span Changed to 50 MHz
10 MHz Peak
SPAN, 5, 0, MHz.
Making a Basic Measurement
Reading Frequency & Amplitude
7. Place a marker (labeled 1) on the 10 MHz peak, as shown in Figure
3-3.
Press
Peak Search.
The frequency and amplitude of the marker appears both in the active function block, and in the upper-right corner of the screen. You can use the knob, the arrow keys, or the softkeys in the Peak Search menu to move the marker. Pressing
Esc removes the value
from the display.
8. If you have moved the marker, return it to the peak of the 10 MHz signal.
Changing Reference Level
9. Press function. Press
AMPLITUDE. The reference level (Ref Level) is now the active
Marker , Mkr Ref Lvl.
Changing the reference level changes the amplitude value of the top graticule line.
Figure 3-4 shows the relationship between the center frequency and
40 Chapter 3
Page 41
the reference level. The box represents the analyzer display. Changing the center frequency changes the horizontal placement of the signal on the display. Changing the reference level changes the vertical placement of the signal on the display. Increasing the span increases the frequency range that appears horizontally across the display.
Figure 3-3 A Marker on the 10 MHz Peak
Making a Basic Measurement
Viewing a Signal
Marker AnnotationActive function block
Figure 3-4 Relationship Between Frequency and Amplitude
Chapter 3 41
Making a Basic Measurement
Page 42
Making a Basic Measurement
Viewing a Signal
Improving Frequency Accuracy
10.While not all of the zeros following the decimal in the active function block are significant, the numbers after the decimal in the marker annotation (upper-right corner of screen) are significant. To increase the accuracy of the frequency reading in the marker annotation, turn on the frequency count function.
a. Press
•The
b. Press
•The
The condition of the softkey
Mkr Fctn.
Marker Fctn softkeys appear.
Marker Count.
Marker Count softkeys appear.
Marker Count On Off. If Off is underlined,
press the softkey to toggle marker count on. As shown in Figure 3-5 and Figure 3-6:
• The marker annotation changes from Mkr1 to Cntr1.
• The displayed resolution in the marker annotation improves.
NOTE When you use the frequency count function, if the ratio of the resolution
bandwidth to the span is less than 0.002, the following message appears on the display: Freq Count: Reduce Span/RBW ratio
This indicates that the resolution bandwidth is too narrow.
11.Press
Marker ➞, Mkr CF to move the 10 MHz peak to the center of
the display.
V alid Marker Count Range
Making a Basic Measurement
12.Move the marker down the skirt of the 10 MHz peak. Although the readout in the active function changes, as long as the marker is at least 26 dB above the noise, the counted value (upper-right corner of display) does not change (see Figure 3-6). For an accurate count, the marker does not have to be at the exact peak.
NOTE Marker count functions properly only on CW signals or discrete peaks.
For a valid reading, the marker must be 26 dB above the noise.
13.Press
BW/Avg, Res BW, then enter a new value. This action makes
the resolution bandwidth the active function and allows you to experiment with different resolution bandwidth (RBW) values.
14.Press
NOTE After properly setting the analyzer to display your signal, you can save
the settings as either a user preset (press
User Preset), or to a file (see “Saving a File” on page 51).
42 Chapter 3
Marker, Off to turn the marker off.
System, Power On/Preset, Save
Page 43
Figure 3-5 Increasing Marker Frequency Accuracy
Frequency Count increases accuracy
Making a Basic Measurement
Viewing a Signal
Figure 3-6 Using Marker Counter
Making a Basic Measurement
Chapter 3 43
Page 44
Making a Basic Measurement
Viewing a Signal
Making a Basic Measurement
44 Chapter 3
Page 45

Viewing Catalogs and Saving Files

4 Viewing Catalogs and Saving
Files
The analyzer stores and retrieves data in a way that is similar to the way that a personal computer (PC) does. They both have internal storage and a diskette drive. While a PC has an internal drive, the
45
Page 46
Viewing Catalogs and Saving Files
analyzer’s internal storage is nonvolatile (flash) memory, which acts as an internal drive. As with a PC, both the internal storage and the diskette drive have directory and sub-directory capability.
This chapter tells you how to:
• locate catalogs and view files, on page 47.
Viewing Catalogs and Saving Files
• save a file, on page 51.
For more information on catalogs and files, see the Measurement guide.
46 Chapter 4
Page 47
Viewing Catalogs and Saving Files

File Menu Functions

File Menu Functions
This chapter describes how to use the functions located under the front-panel that of personal computers (PCs). Like PCs, these analyzers include an internal storage drive and a diskette drive, both of which have directory and sub-directory capability.
You will learn how to do all of the following:
• locate and view files in the catalog
•save a file
•load a file
• rename a file
•copy a file
• delete a file
File key. Data storage and retrieval are handled similarly to
Viewing Catalogs and Saving Files
NOTE The display examples in this book are made using various analyzer
models, you may see some variations depending upon your analyzer.

Locating and viewing files in the catalog

Techniques for locating files and directories are consistent throughout the various file menu functions. Although this section provides specific information about navigation in the catalog, you may wish to refer back to this section when performing other operations of the file menu.
NOTE T race files are n ot bitmap files. Files saved as Trace cannot be imported
into other applications or converted. If you require the use of other applications, for example Microsoft® Word or Microsoft® Excel, b e sure to save your files as .gif or .wmf formats using Comma separated value (csv) files can also be imported into Microsoft® Excel.
Screen for your file Type.
Chapter 4 47
Page 48
Viewing Catalogs and Saving Files
File Menu Functions
Press File, Catalog to bring up a screen display as shown in Figure 4-1.
Figure 4-1 Catalog Menu
Viewing Catalogs and Saving Files
NOTE Use the front-panel Step Keys or Knob and the Dir Up or Dir Select keys
to move up and down within the directory information box.
1. The entry fields show the parameters for the files viewed. Entry fields that are being modified or are modifiable are indicated by highlighting (inverse video). Each file menu function has its own unique set of entry fields pertaining to that function. Below are the fields listed under the catalog key:
a. Name: field states the filename. b. Show Type: field displays the type of files viewed. c. Path: field displays the drive and directory location of the files.
2. The directory information box displays the drives, directories, and/or files currently described in the entry fields. Highlighting a row of information selects that location and updates the entry fields. There are four column headings in this area of the display. They are described as follows:
a. Name: column states the drive ([-A-] or [-C-]) if at the top level
48 Chapter 4
Page 49
Viewing Catalogs and Saving Files
File Menu Functions
on the drive, directory name, or filename.
b. Type: column displays the extension of the filename, for example,
.STA or .TRC. Directories are displayed as DIR and drives have nothing listed under Type:. They are only designated by the Name: column ([-A-] or [-C-]).
c. Size: column displays the size of the file in bytes. d. Modified: column displays the date and time the last change
occurred.
3. The softkey menu has four options:
Type: allows you to choose the type of file you wish to view. (You
a.
may view all types by using the
Sort: you can sort by any of the four columns and you can choose
b.
All key under Type.
up (ascending) or down (descending).
Dir Up: moves you up one directory level. If you are already at the
c.
top level, this key moves you up to the drive level, displaying the available diskette drives.
Viewing Catalogs and Saving Files
Dir Select: moves you down into the highlighted directory or up
d.
into the next level (directory or drive) if the “. .” under the Name: column is highlighted.
NOTE Navigation keys: Use the front-panel Step Keys or Knob and the Dir Up
or Dir Select keys to move around in the directory information box.
4. The drive space line shows the number of bytes used on the drive and the number of bytes still free on the drive.

Creating a directory

Directories and sub-directories can be created on both the A: diskette and the internal C: drive. This allows maximum flexibility in organizing files. For this example, we will create a directory on the C: drive.
1. Press
Figure 4-1except the entry fields will be Dir Name: and Path: and
the Heading will read: Directories.
2. Navigate through the file system until the Path: field displays C:\. (Use the Step Keys, Knob, Dir Up, and/or Dir Select keys.)
File, More 1 of 2, Create Dir. Your screen should look similar to
3. Press (The numeric keypad is also available for the filename.) Press
Name and enter the name, “START” using the Alpha Editor.
Enter
(hardkey) when the Dir Name: field contains this directory title. (Use the
Chapter 4 49
Bk Sp hardkey on the front panel to make corrections.)
Page 50
Viewing Catalogs and Saving Files
File Menu Functions
4. Press Create Dir Now to execute the operation. The message: Creating Directory appears on the screen. Then the status bar displays: Directory C:\START created.
5. To make a sub-directory, scroll down to the START directory you just created and press
Viewing Catalogs and Saving Files
C:\START\. Repeat Steps 3 and 4 above, using a new name. The status bar now should read: Directory C:\START\new name
Dir Select. The Path: field should now read:
created.
50 Chapter 4
Page 51
Viewing Catalogs and Saving Files

Saving a File

Saving a File
You can save files (setups, states, traces, limits, corrections, measurement results, or screens) to a diskette (A:\), or the internal drive (C:\). In this example you will save a trace to the internal C: drive.
NOTE T race files are n ot bitmap files. Files saved as Trace cannot be imported
into other applications or converted. If you require the use of other applications, for example Microsoft® Word or Microsoft® Excel, b e sure to save your files as .gif or .wmf formats using Screen for your file Type.

Step 1. Set up the analyzer trace

1. Perform a factory preset by pressing Preset, Factory Preset (if present).
Viewing Catalogs and Saving Files
2. Turn on the internal 50 MHz alignment signal of the analyzer as follows:
•Press
3. Set the center frequency to 50 MHz by pressing
Freq
4. Set the span to 75 MHz by pressing
Input/Output, Input Port, Amptd Ref.
FREQUENCY, Center
, 50, MHz.
SPAN, Span, 75, MHz. The
reference signal will appear on the display.
5. View trace 1 and put it into maximum hold by pressing
Max Hold. (Trace 1 should already by underlined, but if not, press Trace 1 2 3 until 1 is underlined and then press Max Hold.) Your
View/Trace,
analyzer display should look similar to Figure 4-2.
Chapter 4 51
Page 52
Viewing Catalogs and Saving Files
Saving a File
Figure 4-2 Viewing Trace 1
Viewing Catalogs and Saving Files
6. Activate trace 2 and put it into minimum hold by pressing
Min Hold. Your analyzer display should look similar to Figure 4-3.
Figure 4-3 Viewing Traces 1 and 2
Trace (2),
7. Activate trace 3 by pressing
Trace (3), Clear Write. Your analyzer
display should look similar to Figure 4-4.
52 Chapter 4
Page 53
Figure 4-4 Viewing Traces 1, 2, and 3
Viewing Catalogs and Saving Files
Viewing Catalogs and Saving Files
Saving a File

Step 2. Save the file

1. To access the Save menu, press, File, Save.
2. Select the type of file you want to save. In this example you are saving a trace; press
NOTE T race files are n ot bitmap files. Files saved as Trace cannot be imported
into other applications or converted. If you require the use of other applications, for example Microsoft® Word or Microsoft® Excel, b e sure to save your files as .gif or .wmf formats using Comma separated value (csv) files can also be imported into Microsoft® Excel.
3. Select the trace you wish to save (1, 2, 3, or all traces). In this example you will save trace 3; press menu),
Trace 3.
4. Select the format you wish to use for saving the trace, either trace plus state (
Trace + State) or comma separated value (CSV). In this
example you will save as trace plus state; press
5. Enter a filename by pressing this example you will name the file, “TEST1”. (The numeric keypad is also available for the filename.) The file extension is always set by the system and for this file type the extension is TRC. You must now press the
Enter (hardkey) to get back to the Save menu.
Type, Trace.
Screen for your file Type.
Source (softkey in the Save
Format, T rac e + State.
Name. The Alpha Editor appears. For
Chapter 4 53
Page 54
Viewing Catalogs and Saving Files
Saving a File
NOTE You need to use a filename that does not already exist in the current
directory. The filename is limited to eight characters, alpha (A-Z) or numeric (0-9) in any combination. The analyzer will not allow you to overwrite an existing file. If you select a filename that already exists, the status bar will display the message: File already exists. If you do not choose a filename, the analyzer will automatically generate a
Viewing Catalogs and Saving Files
name based on the type of file you are saving (Setup: SETUP, State: STATE, Trace: TRACE, Limits: LIMIT, Corrections: COREC, Screen: SCREN).
It also generates a three digit integer (starting at 000 and extending through 999, remembering the previously saved value through a power cycle) which it adds to the name, for example: TRACE056.TRC. This three digit integer increments upon each attempted save until a unique filename is created, without regard to the success of the save.
6. The destination for the saved file is shown in the Path: field. In this example, you will select the path as C:\START\. If the correct location is not listed in the Path: field, change directories as follows:
a. Press
Dir Up or Dir Select and use the step keys or knob, to
highlight the desired destination directory.
b. Press
Dir Select and confirm your choice displayed in the Path:
field.
7. Press
Save Now or Save (hardkey) to save the file to the C:\Start\
location. The message Saving Now is displayed during this operation. For this example, the status line displays: C:\START\TEST1.TRC file saved.
NOTE When saving to drive (A:), never remove the diskette during the save
operation. To do so could corrupt all data on the diskette.
54 Chapter 4
Page 55
Viewing Catalogs and Saving Files
Saving a File

Loading a file

1. Reset the analyzer by pressing Preset, Factory Preset (if present).
Viewing Catalogs and Saving Files
2. To access the Load menu, press
File, Load.
3. Select the type of file you want to load (setup, state, trace, limits, or corrections). In this example you are loading a trace file; press
Trace
.
NOTE Not all file types can be loaded back into the analyzer. For example,
Type,
Screen files and CSV (comma separated value) files cannot be loaded. CSV and Screen files are designed for use with a PC.
4. Select the directory where your file is located. In this example, select the C:\START\ directory . (Use the front-panel step keys, knob, and/or
Dir Select keys to locate the directory.)
Dir Up,
5. Select the file you want to load into the analyzer by moving the cursor with the front-panel knob to highlight the file name. In this example, the file is TEST1.TRC. Select the trace into which you wish to load the file. In this example, load it into Trace 2. (
Trace 2).
6. Press
Load Now to load the specified file. Your display should look
Destination,
similar to Figure 4-5. The status bar reads: C:\START\TEST1.TRC file loaded.
Figure 4-5 File Loaded
Chapter 4 55
Page 56
Viewing Catalogs and Saving Files
Saving a File
NOTE You should notice that the trace you saved, in this example trace 3, is
loaded to trace 2. If you wish to verify this condition, remove the signal input. Press
Input/Output, Input Port, Amptd Ref (Off).
When a trace is loaded, it is placed in view mode. To see that trace 2 is
Viewing Catalogs and Saving Files
Blank, Trace 3, Blank. Then press Trace 2. Notice that trace 2 is the only
in view mode, blank traces 1 and 3 by pressing
View/Trace, Trace 1,
trace displayed in view mode (the view key is selected, and the trace is not sweeping). Your display should look similar to Figure 4-6.
Figure 4-6 Verifying Trace 1
CAUTION When operating in DC coupled mode, ensure protection of the input
mixer by limiting the input level to 0 Vdc and +30 dBm.
NOTE When loading Limits files, be sure you have selected the appropriate
X-Axis Units: frequency or time (
Display, Limits, Properties, X Axis Units).
If you are in time X-Axis Units, and you load frequency limits, all current limit line data will be erased and the analyzer will switch to the frequency domain. The reverse of the this situation also holds true.
56 Chapter 4
Page 57
Viewing Catalogs and Saving Files
Saving a File

Renaming a File

1. To access the Rename menu, press File, Rename.
2. Select the type of file you want to rename (setup, state, trace, limits, screens, or corrections). In this example you are renaming a trace file; press
3. Select the drive and directory where your file is located. In this example, choose drive C:\START.
4. Select the file you want to rename. In this example, choose the file TEST1.TRC.
Type, Trace.
Viewing Catalogs and Saving Files
5. Press
Name to open the Alpha Editor menu. For this example,
rename the file to RENAME using the Alpha Editor softkeys. (The Name: field is limited to eight characters.) Your display should look
similar to Figure 4-7.
6. Press
Enter, Rename Now: your file is now renamed and visible within
the directory displayed on your analyzer . The status line displ ays the message: C:\START\TEST1.TRC file renamed to C:\START\RENAME.TRC.
Figure 4-7 Renaming a File
Chapter 4 57
Page 58
Viewing Catalogs and Saving Files
Saving a File

Copying a File

1. T o access the Copy menu, press File, Copy. This menu function brings up 2 directory boxes as described by the From: and To: fields located directly above the boxes. Refer to Figure 4-8.
Viewing Catalogs and Saving Files
2. Put a formatted diskette in the A: drive.
Figure 4-8 Copy Menu
3. Select the type of file you want to copy (setup, state, trace, limits, screens, or corrections). In this example you are copying a trace file from the C: drive to the A: drive; press,
4. Press
Dir (in the key menu) to underline From:. Notice the From:
Type, Trace.
field (just above the top directory box) highlights the location from which you will be copying. Using the navigation keys (
select
, step keys, or front panel knob), locate the C:\START\
Dir Up, Direct
directory.
5. Scroll to highlight the file displayed in the directory box from which you wish to copy. In this example, you will need to highlight the file RENAME.
6. Press
Dir (in the key menu) to underline To:. The To: field (just
above the bottom directory box) is highlighting the location to which you will be copying. Using the navigation keys as described above, highlight the A:\ drive.
NOTE When pressing Dir (To), the file chosen in the From: location is no longer
highlighted, however the Name: field above the From directory box
58 Chapter 4
Page 59
Viewing Catalogs and Saving Files
displays the chosen file; in this example: RENAME.TRC.
Viewing Catalogs and Saving Files
Saving a File
Chapter 4 59
Page 60
Viewing Catalogs and Saving Files
Saving a File
NOTE Highlighting a filename in the To: directory box has no significance.
The file cannot be placed inside another file.
7. Press
Copy Now to execute the operation. The message: “Copying
file” is displayed during the copying process. Upon completion, the
Viewing Catalogs and Saving Files
status bar reads: “C:\START\RENAME.TRC file copied”. You have now copied the file RENAME from the C: drive to the A: drive.

Deleting a File

1. To access the Delete menu, press File, Delete.
2. Select the type of file you want to delete (setup, state, trace, limits, screens, or corrections). In this example, you are deleting a trace file; press
Type, Trace.
3. Select the drive and directory where the file you wish to delete is located. In this example, choose C:\START\. Use the navigati on keys
Dir Up, Dir Select, step keys, front panel knob) to highlight this
( location in the Path: field. (Refer to “Locating and viewing files in
the catalog” on page 47 for more instructions.)
4. Select the file you want to delete. In this example, choose the file RENAME.TRC. Move the cursor with the navigation keys to highlight the file name.
5. Press display during the operation. Upon completion, the status bar reads: C:\START\RENAME.TRC file deleted. Your file is now deleted and is no longer visible in the directory displayed on your analyzer.
Delete Now. The message: Deleting file pops up on the
60 Chapter 4
Page 61
Viewing Catalogs and Saving Files

Using the Alpha Editor

Using the Alpha Editor
The Alpha Editor enables you to select letters to enter a name for a file or directory (
File, Rename, for example).
Viewing Catalogs and Saving Files
1. Use the
More keys to display the softkey that lists the character you
want. Each softkey has several characters on it.
2. When you press a softkey, its characters are displayed on individual softkeys.
3. Select the one you want.
4. Digits are entered using the front panel numeric key pad.
Chapter 4 61
Page 62
Viewing Catalogs and Saving Files
Using the Alpha Editor
Viewing Catalogs and Saving Files
62 Chapter 4
Page 63
Options and Accessories

5 Options and Accessories

This chapter lists options and accessories available for your analyzer.
61
Page 64
Options and Accessories
Options and Accessories

Ordering Options and Accessories

Ordering Options and Accessories
Options and accessories help you configure the analyzer for your specific applications, and enable you to use the analyzer.
Options (see page 63)
Unless specified otherwise, all options are available when you order a spectrum analyzer; some options are also available as kits that you can order and install after you receive the analyzer . Order kits through your local Agilent Sales and Service Office.
For the latest information on Agilent Spectrum Analyzer options and upgrade kits, go to:
http://www.agilent.com/find/sa_upgrades
Accessories (see page 70)
Order accessories through your local Agilent Sales and Service Office.
62 Chapter 5
Page 65

Options

Options and Accessories
Options
Option
Number
0B0
0B1
0BW
1CM
1CN Front Handle Kit Available as P/N 5063-9228.
1CP
1CR
Manual Set on CD-ROM
Only
Additional Manual Set An additional set of the standard documentation shipped with
Service Documentation The Service guide describes assembly-level troubleshooting
Rack Mount Kit Includes rack mount flange s and hardware. Used to rack mount
Rack Mount Kit with Handles
Rack Slide Kit Available as P/N E4406-60115. Rack slides allow the analyzer to
Name Description
The documentation CD-ROM contains the standard documentation set as well as Adobe Acrobat Reader with Search.
the analyzer. See “Initial Inspection” on page 9 for information about the standard documentation set.
procedures, provides a parts list, and documents post-repair procedures.
analyzers without front handles (available as P/N 5063-9215).
Includes the parts necessary to rack mount an analyzer with front handles attached (available as P/N 5063-9222). (Includes handles.)
be pulled out while the analyzer is secured to the rack.
Options and Accessories
1DS
015 6 GHz Measurement
107 Audio Input, 100 k Audio signal analysis capability that works with the Option 233
110 RF/µW Internal
111 USB Device Side I/O
Preamplifier An internal preamplifier assembly. Not required when Option
110 is installed. Option 110 cannot be installed if Option 1DS is installed.
Frequency Range: 100 kHz to 3 GHz
Includes accessories necessary to set up return loss and loss/gain
Accessory Kit
Preamplifier
Interface
measurements. Used with Option 215.
Measuring Receiver System. Available in all models. Option 107 cannot be installed if Option 122 or 140 is installed. Requires Option 233.
An internal preamplifier assembly that covers all frequency bands. Available in all models. Option 110 cannot be installed if Option 1DS is installed.
Frequency Range: 10 MHz to 50 GHz
Provides a rear-panel USB interface connector (device side I/O only).
Chapter 5 63
Page 66
Options and Accessories
Options
Options and Accessories
Option
Number
115 512 Mb User Memory Provides additional memory to accommodate a greater number
117 Secure Memory Erase Provides the ability to quickly and easily perform a security
122 Not
available for
E4447A
123 Switchable Preselector
80 MHz Bandwidth Digitizer
Bypass
Name Description
of applications. Available in all models. Option 115 cannot be installed if Option 117 is installed.
erase, deleting all user-stored data, without affecting the usability of the analyzer. Option 117 cannot be installed if Option 115 is in stalled.
Adds an alternative 80 MHz wide IF path for making measurements. Cannot be installed if Option 107 is installed. Additional measurement accuracy can be achieved using the External Calibration Wizard Option 235, or with the 89600 Vector Signal Analyzer Application calibration. The 80 MHz functionality is available as part of the added Basic Mode, WLAN Personality (Option 217), and Flexible Digital Modulation (Option 241). Basic Mode includes the measurements:
spectrum (frequency domain) waveform (time domain)
Requires Option 123.
Allows you to bypass the preselector filter at the input of the spectrum analyzer when making measurements in the 3 to 50 GHz range. This improves the sensitivity and increases the available measurement dynamic range. However it allows the analyzers internal mixing products and image response signals to be visible on the display, in addition to the actual desired signal response. This option is available in all models. Option 123 cannot be installed if Option AYZ is installed.
124
140
Not
available for
E4447A
Y-axis Video Output Provides a 0 to 1 V signal that corresponds to the bottom and top
of the display graticule, respectively. The output voltage tracks the Y-axis position of the trace data while the analyzer is sweeping. (This output is similar to the 8566B Video Out.)
40 MHz Bandwidth Digitizer
Adds an alternative 40 MHz wide IF path for making measurements. Cannot be installed if Option 107 is installed. The 40 MHz functionality is available as part of the added Basic Mode, WLAN Personality (Option 217), and Flexible Digital Modulation (Option 241). Basic Mode includes the measurements:
spectrum (frequency domain) waveform (time domain)
Requires Option 123.
64 Chapter 5
Page 67
Options and Accessories
Options
Option
Number
Name Description
202 GSM with EDGE
Measurement Personality
204 1xEV-DO Measurement
Personality
General transmitter measurement:
transmit power
Provides GSM (GMSK modulation) transmitter measurements:
power vs. time phase and frequency error output RF spectrum transmit band spurs
Provides EDGE (8PSK modulation) transmitter measurements:
EVM power vs. time output RF spectrum transmit band spurs
Requires Option B7J hardware.
Provides 1xEV transmitter measurements:
channel power ACPR intermodulation spectrum emission mask occupied BW code domain power (pilot) EVM (pilot) QPSK EVM power statistics CCDF
Requires Option B7J hardware.
Options and Accessories
210 HSDPA/HSUPA
Measurement Personality
Provides High-Speed Downlink/Uplink Packet Access to support 16QAM and modulation accuracy. Requires Option BAF.
Requires Option B7J hardware.
211 TD-SCDMA Power
Measures the power profile of TD-SCDMA BTS and MS.
Measurement Personality
212 TD-SCDMA Modulation
Analysis Measurement Personality
213 HSDPA/8PSK for
TD-SCDMA Modulation
Provides modulation analysis for TD-SCDMA BTS and MS, such as EVM, code domain analysis, and time domain or frequency domain analysis. Requires Option B7J.
Provides measurements for high-speed downlink packet access
over TD-SCDMA to support HSDPA/8PSK. Requires Option 212. Analysis Measurement Personality
214 1xEV-DV Supports 8PSK and 16QAM modulation analysis. Requires
Option B78 and B7J. Option 1DS is recommended. Includes
measurements:
Composite Rho Code Domain tests
215 External Source Control
Measurement Personality
Adds external source control to the PSA analyzer for
stimulus-response measurements.
Chapter 5 65
Page 68
Options and Accessories
Options
Options and Accessories
Option
Number
217 Not
WLAN Measurement Personality
Name Description
available for
E4447A
219 Noise Figure Measurement
Personality
226 Phase Noise Measurement
Personality
233 Measuring Receiver
Personality and PC software
Provides WLAN transmitter measurements for 802.11 a/b/g. Requires Option 122 or 140. Also requires Option 123 for center frequencies above 3 GHz.
Provides noise figure measurements:
Limit lines Loss compensation Extended frequency (measures up/down converters)
You need the preamp option: Option 1DS to have specified performance below 3 GHz, or Option 110 to have nominal performance over the entire frequency range.
Provides phase noise measurements:
Monitor spectrum Spot frequency Log plot
This option allows you to use your PSA, an Agilent power meter and N5532A Sensor module as a measuring receiver system through the PSA user interface (UI). It also provides PC software, as an extension of the built-in measuring receiver personality, offering graphic and remote UI on an external PC. A sensor module to power meter adapter, N5532A Option 019, is required for use with N1911A/12A power meters. Option 233 is available in all models.
Requires Option 107 for Audio measurements.
23A AM/FM/PM Triggering for
use with the Measuring Receiver Personality
23B CCITT Audio Filter for use
with the Measuring Receiver Personality
235 Not
available for
80 MHz Digitizer External Calibration Wizard and License
E4447A
Option 233 must be installed before Option 23A Provides the trigger support in the AM Depth, FM Deviation,
and PM Deviation measurements of the Measuring Receiver personality to detect the settling time for carrier frequency change. The triggering types supported are Free Run, Video (IF Envelope), External Front, External Rear, and RF Burst.
Option 233 must be installed before Option 23B Adds the CCITT weighting filter, a 30 kHz and an 80 kHz low
pass filter, and a 400 Hz high pass filter to the audio measurements of the Measuring Receiver personality. (These include Audio Frequency, Audio AC Level, Audio Distortion, and Audio SINAD.)
This calibration software runs on an external PC to generate a file of correction factors for PSAs with Option 122. You get improved measurement accuracy using this correction file.
66 Chapter 5
Page 69
Options and Accessories
Options
Option
Number
Name Description
239 EMC Analyzer Personality Adds EMC measurement capability. With the addition of the
N9039A RF Preselector, allows compliant EMC measurements.
Selected signal sources are required for system alignment.
241 Flexible Digital Modulation
Analysis
Provides flexible demodulating analysis for measuring many
different formats of digital modulation, including:
MSK type 1 and type 2 EDGE QPSK and offset QPSK 8PSK BPSK DQPSK and Pi/4 DQPSK D8PSK QAM 16, 32, 64, 128, 256 FSK 2, 4, 8 states
Requires Option 122 or 140 for symbol rates > 10 MHz
266 Programming Code
Compatibility Suite
Lets you select one of three different modes that let you use a
limited set of 8560 Series, 8566B/8568B, or 8590 Series
programming commands. These modes let you control the PSA
Series analyzers with code from other spectrum analyzer
families. When these modes are enabled, the instruments native
SCPI commands are disabled if the remote language selected is
not SCPI.
Options and Accessories
ANE
Enhanced Memory Upgrade For early instruments, replaces existing 48 Mb with 64 Mb
memory on CPU, providing additional memory for storing
optional measurement personalities. The 64 Mb memory was
shipped standard in all PSA instruments with serial numbers
US/MY 42220000.
It is recommended that Option 115 (512 Mb user memory) be
purchased to accommodate a greater number of applications.
A YZ External Mixing Provides the ability to use the HP/Agilent 1 1970 Series harmonic
mixers, and HP/Agilent 11974 preselected harmonic mixers with
the Agilent E4440A, E4446A, E4447A, or E4448A analyzer to
extend the frequency range to 11 0 GH z. Oper ation to 325 GHz i s
also possible with non-HP/Agilent mixers. Option AYZ cannot be
installed if Option 123 is installed.
B78 cdma2000 Measurement
Personality
Provides cdma2000 transmitter measurements:
channel power ACPR intermodulation spectrum emission mask occupied BW code domain power modulation accuracy (composite rho) QPSK EVM power statistics CCDF
Requires Option B7J hardware.
Requires Option 214 for 1xEV-DV functionality.
Chapter 5 67
Page 70
Options and Accessories
Options
Options and Accessories
Option
Number
Name Description
B7J Digital Demod Hardware Adds 1 dB step, electronic attenuator in addition to the existing
manual attenuator. This is required for the personality options: BAC, BAE, BAF, B78, 202, 204, 210, 212, 213, and 214. This option also adds the functionality of Basic Mode, including the measurements:
spectrum (frequency domain) waveform (time domain)
BAB
(E4440A
APC 3.5 Input Connector Replaces the standard type-N female connector with an APC
3.5-mm male connector.
only)
BAC cdmaOne Measurement
Personality
Provides IS-95 and J-STD-008 transmitter measurements:
channel power modulation accuracy (rho) code domain power ACPR spur close
Requires Option B7J hardware.
BAE NADC/PDC Measurement
Personality
Provides NADC transmitter measurements:
ACP EVM (error vector magnitude)
Provides PDC transmitter measurements:
ACP EVM (error vector magnitude) occupied BW
Requires Option B7J hardware.
BAF W-CDMA Measurement
Personality
UK6
Commercial Calibration with Test Data
a
A6J ANSI Z540 Compliant
Calibration with Test Data
68 Chapter 5
Provides 3GPP W-CDMA transmitter measurements:
channel power ACPR (ACLR) intermodulation multi carrier power spectrum emission mask occupied BW code domain power modulation accuracy (composite EVM) QPSK EVM power statistics CCDF
Requires Option B7J hardware. Requires Option 210 for addition of HSDPA/HSUPA
functionality.
Provides the factory calibration test data and the standard commercial calibration certificate at initial shipment.
Provides compliant calibration test data and certificate at initial
a
shipment
Page 71
Options and Accessories
Options
Option
Number
1A7 ISO 17025 Compliant
Calibration with Test Data
R-51-001-3C 3-Year Warranty Service
Support
Name Description
a
a. Available only at time of purchase
Provides compliant calibration test data and certificate at initial
a
shipment
1-year return-to-Agilent warranty extended to 3 years of
warranty service support. This adds a 2-year service contract to
the base analyzer 1-year warranty.
Options and Accessories
Chapter 5 69
Page 72
Options and Accessories
Options and Accessories

Accessories

Accessories
A number of accessories are available from Agilent Technologies to help you configure your analyzer for your specific applications. They can be ordered through your local Agilent Sales and Service Office and are listed below.
50 Load
The Agilent 909 series of loads come in several models and options providing a variety of frequency ranges and VSWRs. Also, they are available in either 50 or 75 Ω. Some examples include the:
909A: DC to 18 GHz 909C: DC to 2 GHz 909D: DC to 26.5 GHz
50 Ω/75 Ω Minimum Loss Pad
The HP/Agilent 11852B is a low VSWR minimum loss pad that allows you to make measurements on 75 devices using an analyzer with a 50 input. It is effective over a frequency range of DC to 2 GHz.

AC Probe

The Agilent 85024A high frequency probe performs in-circuit measurements without adversely loading the circuit under test. The probe has an input capacitance of 0.7 pF shunted by 1 M of resistance and operates over a frequency range of 300 kHz to 3 GHz. High probe sensitivity and low distortion levels allow measurements to be made while taking advantage of the full dynamic range of the spectrum analyzer.

AC Probe (Low Frequency)

The Agilent 41800A low frequency probe has a low input capacitance and a frequency range of 5 Hz to 500 MHz.

Close-Field Probes (for EMC measurements with PSA)

These handheld probes are designed to measure magnetic-field radiation from surface currents, slots, cables, and ICs for EMC diagnostics and troubleshooting. Five antenna factors appear on each probe for calculating absolute magnetic-field strength from the analyzer reading.
11940A - 30 MHz to 1 GHz range
70 Chapter 5
Page 73
Options and Accessories
Accessories
11941A - 9 kHz to 30 MHz range 11945A - Close-field probe set; includes both 11940A and 11941A
probes for full coverage from 9 kHz to 1 GHz. Option E51 adds the 11909A preamplifier.

Broadband Preamplifiers and Power Amplifiers

Preamplifiers and power amplifiers can be used with your spectrum analyzer to enhance measurements of very low-level signals.
• The Agilent 8447D preamplifier provides a minimum of 25 dB gain from 100 kHz to 1.3 GHz.
• The Agilent 83006A preamplifier provides a minimum of 26 dB gain from 10 MHz to 26.5 GHz.
• The 11909A low noise preamplifier provides a minimum of 32 dB gain from 9 kHz to 1 GHz and a typical noise figure of 1.8 dB.

GPIB Cable

Options and Accessories
The Agilent 10833 Series GPIB cables interconnect GPIB devices and are available in four different lengths (0.5 to 4 meters). GPIB cables are used to connect controllers to a spectrum analyzer.

USB/GPIB Interface

The Agilent 82357B USB/GPIB interface provides a direct connection from the USB port on your laptop or desktop PC to GPIB instruments. It comes with the SICL and VISA software for Windows® 2000 SP4 /XP SP2 (or later)/Vista (Business, Ultimate, Enterprise). Using VISA software, your existing GPIB programs work immediately, without modification. The 82357B is a standard Plug and Play device and you can interface with up to 14 GPIB instruments.

HP/Agilent 11970 Series Harmonic Mixers

The 11970 Series harmonic mixers are available to extend the frequency range of analyzers with Option AYZ up to 110 GHz. The following six models are available:
Table 5-1
HP/Agilent Model Number
Frequency Range
11970K 18.0 to 26.5 GHz 11970A 26.5 to 40.0 GHz 11970Q 33.0 to 50.0 GHz
Chapter 5 71
Page 74
Options and Accessories
Accessories
Table 5-1
Options and Accessories
HP/Agilent Model Number
11970U 40.0 to 60.0 GHz 11970V 50.0 to 75.0 GHz 11970W 75.0 to 110 GHz
Frequency Range

HP/Agilent 11974 Series Preselected Millimeter Mixers

The 11974 Series preselected millimeter mixers are available to extend the frequency range of analyzers with Option AYZ (external mixing) up to 75 GHz. Preselection reduces mixer overload from broadband signals, reduces radiation of local oscillator harmonics back to the device under test, and reduces the level of image and multiple responses displayed. The following four models are available:
Table 5-2
HP/Agilent Model Number
11974A 26.5 to 40.0 GHz
Frequency Range
11974Q 33.0 to 50.0 GHz 11974U 40.0 to 60.0 GHz 11974V 50.0 to 75.0 GHz

RF and Transient Limiters

The Agilent 11867A and N9355B RF Limiters protect the analyzer input circuits from damage due to high power levels. The 11867A operates over a frequency range of dc to 1800 MHz and begins reflecting signal levels over 1 milliwatt up to 10 watts average power and 100 watts peak power. The N9355B microwave limiter (0.1 to 18 GHz) guards against input signals over 1 milliwatt up to 1 watt average power and 10 watts peak power.
The Agilent 11947A Transient Limiter protects the analyzer input circuits from damage due to signal transients. It specifically is needed for use with a line impedance stabilization network (LISN). It operates over a frequency range of 9 kHz to 200 MHz, with 10 dB of insertion loss.

Power Splitters

The Agilent 11667A/B/C power splitters are two-resister type splitters
72 Chapter 5
Page 75
Options and Accessories
Accessories
that provide excellent output SWR, at 50 impedance. The tracking between the two output arms, over a broad frequency range, allows wideband measurements to be made with a minimum of uncertainty.
11667A: DC to 18 GHz 11667B: DC to 26.5 GHz 11667C: DC to 50 GHz
Options and Accessories
Chapter 5 73
Page 76
Options and Accessories
Options and Accessories
Accessories

RF Bridges

The Agilent 86205A 50 RF bridge can be used with the PSA series to make reflection measurements using an analyzer with the external source control, Option 215.
These external directional bridges offer high directivity and excellent port match. The 86205A operates over a frequency range of 300 kHz to 6 GHz.
50 Transmission/Reflection Test Kit
The Agilent 87512A provides the accessories to measure the reflection and transmission characteristics of 50 devices up to 2 GHz. These accessories use type N connectors. The kit contains a directional bridge, 2 cables and precision short/load terminations for calibration.

Static Safe Accessories

9300-0797 3M static control mat 0.6 m × 1.2 m (2 ft. × 4 ft.) with a
4.6 cm (15 ft.) ground wire. Does not include wrist strap or cord.
9300-1367 Wrist-strap, color black, stainless steel. Four adjustable
links and a 7 mm post-type connection. 9300-0980 Wrist-strap cord 1.5 m (5 ft.). 9300-1308 ESD heel-strap (reusable 6 to 12 months).

Noise Sources (for use with PSA Option 219)

Agilent 346A 10 MHz to 18 GHz noise source with 5 dB ENR
(nominal) Agilent 346B 10 MHz to 18 GHz noise source with 15 dB ENR
(nominal) Agilent 346C 10 MHz to 26.5 GHz noise source with 15 dB ENR
(nominal)
Agilent 346CK01 1 GHz to 50 GHz noise source with 21 dB ENR
(nominal)
74 Chapter 5
Page 77

6 In Case of Difficulty

In Case of Difficulty
73
Page 78
In Case of Difficulty
This chapter includes information on how to check for a problem with your Agilent Technologies spectrum analyzer, and how to return it for service.
If you experience a problem or would like additional information about your analyzer, Agilent T ech nologies’ worldwide organi zation is ready to provide the support you need. Before calling Agilent Technologies, however (or returning an analyzer for service), perform the quick checks listed in “Check the Basics” on page 76. This check may eliminate the problem.
If a problem persists, you may choose to:
• Repair the analyzer yourself. See “Service Options” on page 78.
• Return the analyzer to Agilent Technologies for repair. See
“Returning an Analyzer for Service” on page 79, for more
information.
WARNING No operator serviceable parts inside. Refer servicing to
qualified personnel. To prevent electrical shock, do not remove covers.
NOTE If the analyzer is still under warranty or is covered by a maintenance
contract, it will be repaired under the terms of the warranty or plan. If the analyzer is no longer under warranty or is not covered by an
Agilent T echnologies maintenance plan, Agilent Technologies will notify you of the cost of the repair after examining the analyzer.
In Case of Difficulty
74 Chapter 6
Page 79

Types of Spectrum Analyzer Messages

The analyzer can generate various messages that appear on the display during operation.
For a complete list of spectrum analyzer messages, see the Instrument Messages and Functional Tests manual. The following table describes the four types of spectrum analyzer messages.
T able 6-1 Types of Messages
Type of Message Location Notes
In Case of Difficulty
Types of Spectrum Analyzer Messages
Informational messages typically provide verification that an action has occurred. In general, no user intervention is required.
Pop-up messages prompt you to verify that an action will be performed or tell you that an action is currently being performed.
Status messages indicate a condition that may result in erroneous data being displayed. Multiple status messages may be displayed at the same time.
User Error messages appear when an attempt has been made to set a parameter incorrectly or an operation has failed (such as saving a file).
Bottom of the display in the status line (in green on analyzers with a color display).
Middle of the display in a framed box.
Right side of the display and/or in the SCPI Status Register system.
Bottom of the display in the status line (in yellow on analyzers with a color display) and in the SCPI Error Queue.
Messages will remain until the message is cleared by pressing
Esc or Preset, or is overwritten
by another message.
Message remains until you acknowledge the prompt, or the action is completed.
Most messages remain only until the erroneous condition is corrected.
Messages remain until you clear the error (press
Show Errors, Clear Error Queue
or another message is displayed in the status line.
Pressing the will clear error messages from the display, but the messages will remain in the error queue.
Esc or Preset keys
System,
),
In Case of Difficulty
Chapter 6 75
Page 80
In Case of Difficulty

Before Calling Agilent Technologies

Before Calling Agilent Technologies

Check the Basics

Is there power at the receptacle? Make sure there is no disk in the front-panel disk drive. Is the analyzer turned on? Check to see if the green LED above the
power switch is on. Also, listen for internal fan noise to determine if the analyzer cooling fans are running.
If other equipment, cables, and connectors are being used with your
spectrum analyzer, make sure they are connected properly and operating correctly.
Review the measurement procedures being performed when the
problem first appeared. Are all of the settings correct?
If the analyzer is not functioning as expected, return the analyzer to
a known state by pressing
NOTE Some analyzer settings are not affected by a Preset. If you wish to reset
Preset, and Factory Preset (if displayed).
the analyzer configuration to the state it was in when it was originally sent from the factory, press
Factory. Then press the green Preset key on the front panel.
System, Power On/Preset, Preset Type,
Is the measurement being performed, and the results that are
expected, within the specifications and capabilities of the analyzer? Refer to the Specifications guide for your analyzer.
If the analyzer is not communicating through the LAN connection,
check for the presence of blinking yellow LEDs on the rear panel
In Case of Difficulty
NOTE The analyzer must be powered on with the LAN already connected in
LAN connector. If the Rx (receive) LED is not blinking, check the LAN cable and LAN integrity.
order to recognize the LAN port. T o meet specifications, the analyzer must be aligned. Either the Auto
Align (On) feature must be selected (press
Align
, On), or the analyzer must be manually aligned at least once
System, Alignments, Auto
every 24 hours, or whenever the temperature changes more than 3°C. When
Auto Align, On is selected, the AA screen annotation
appears on the left edge of the display.
Perform an Alignment. Press
System, Alignments, Align All Now.
If the previously performed alignments did not resolve the problem,
System, Alignments, Restore Align Defaults. Then press System,
press
76 Chapter 6
Page 81
In Case of Difficulty
Before Calling Agilent Technologies
Alignments, Align All Now.
Is the analyzer displaying an error message? If so, refer to the
Instrument Messages and Functional Tests manual.
If the necessary equipment is available, perform the functional tests
in the Instrument Messages and Functional Tests document for your analyzer.
If the 10 MHz output is not functioning, verify that it is turned on by
pressing
TIP You can get automatic electronic notification of new firmware releases
System, Reference, 10 MHz Out to select On.
and other product updates/information by subscribing to the Agilent Technologies Test & Measurement E-Mail Notification Service for the
PSA Series at:
http://www.agilent.com/find/notifyme‘
In Case of Difficulty
Chapter 6 77
Page 82
In Case of Difficulty
Before Calling Agilent Technologies

Read the Warranty

The warranty for your analyzer is in the front of your Specifications Guide. Please read it and become familiar with its terms.
If your analyzer is covered by a separate maintenance agreement, please be familiar with its terms.

Service Options

Agilent Technologies offers several optional maintenance plans to service your analyzer after the warranty has expired. Call your Agilent Technologies office for full details.
If you want to service the analyzer yourself after the warranty expires, you can purchase the service documentation that provides all necessary test and maintenance information.
You can order the service documentation, Option 0BW (assembly level troubleshooting information) through your Agilent Technologies office.

Calling Agilent Technologies

Agilent Technologies has offices around the world to provide you with complete support for your analyzer. To obtain servicing information or to order replacement parts, contact the nearest Agilent Technologies office listed below. In any correspondence or telephone conversations, refer to your analyzer by its product number, full serial number, and firmware revision.
Press
System, More (1 of 3), Show System, and the product number , serial
number, and firmware revision information will be displayed on your analyzer screen. A serial number label is also attached to the rear panel of the analyzer.
In Case of Difficulty
Contacting Agilent Technologies
Online assistance: http://www.agilent.com/find/assist United States
(tel) 1 800 829 4444
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
New Zealand
(tel) 0 800 738 378 (fax) 64 4 495 8950
Asia Pacific
(tel) (852) 3197 7777 (fax) (852) 2506 9284
Europe
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
78 Chapter 6
Australia
(tel) 1 800 629 485 (fax) (61 3) 9210 5947
Page 83
In Case of Difficulty

Returning an Analyzer for Service

Returning an Analyzer for Service
Service Tag
When you are returning an analyzer to Agilent Technologies for service, fill out and attach one of the blue service tags provided at the end of this chapter. Please be as specific as possible about the problem. If you have recorded any error messages that appeared on the display, have completed a functional test, or have any other specific data on the performance of your analyzer, please include a copy of this information.
Original Packaging
Before shipping, pack the unit in the original factory packaging materials if they are available. If the original materials were not retained, see "Other Packaging" (below).
NOTE Install the transportation disk into the diskette drive to reduce the
possibility of damage during transportation. If the original transportation diskette is not available, a blank diskette may be substituted.
Other Packaging
CAUTION Analyzer damage can result from using packaging materials other than
those specified. Never use styrene pellets in any shape as packaging materials. They do not adequately cushion the equipment or prevent it from shifting in the carton. They cause equipment damage by generating static electricity and by lodging in the analyzer louvers, blocking airflow.
You can repackage the analyzer with commercially available materials, as follows:
1. Attach a completed service tag to the analyzer.
2. Install the transportation diskette or a blank diskette into the disk drive.
3. Wrap the analyzer in antistatic plastic to reduce the possibility of damage caused by electrostatic discharge.
4. Use a strong shipping container. The carton must be both large enough and strong enough to accommodate the analyzer. A double-walled, corrugated cardboard carton with 159 kg (350 lb) bursting strength is adequate. Allow at least 3 to 4 inches on all sides of the analyzer for packing material.
In Case of Difficulty
Chapter 6 79
Page 84
In Case of Difficulty
Returning an Analyzer for Service
5. Surround the analyzer with three to four inches of packing material and prevent the analyzer from moving in the carton. If packing foam is not available, the best alternative is S.D.-240 Air Cap™ from Sealed Air Corporation (Hayward, California, 94545). Air Cap looks like a plastic sheet filled with 1-1/4 inch air bubbles. Use the pink-colored Air Cap to reduce static electricity. Wrapping the analyzer several times in this material should both protect the analyzer and prevent it from moving in the carton.
6. Seal the shipping container securely with strong nylon adhesive tape.
7. Mark the shipping container “FRAGILE, HANDLE WITH CARE” to assure careful handling.
8. Retain copies of all shipping papers.
In Case of Difficulty
80 Chapter 6
Page 85

Index

Numerics
10 MHz reference, turning on 50 ohm load 50 ohm RF bridge 50 ohm/75 ohm minimum loss pad
69
75 ohm RF bridge
A
AC probe 69 Accessories accessories
50 ohm directional bridge 50 ohm load 50 ohm short/load 50 ohm/75 ohm minimum loss
AC probe broadband preamplifiers close-field probes GPIB cable harmonic mixers power splitters preselected millimeter mixers
RF bridge RF limiters transient limiters
ACPR
card slot location 29
active function address, IP Agilent Technologies, calling alignments, internal alpha editor, using amplifiers annotations, display application information arrow keys, using
B
battery information Bluetooth
card slot location bridge broadband preamplifiers
C
cable
GPIB Catalog Menu catalogs
definition
naming clock, battery close-field probes
69, 72
72
72
69
69
72
pad 69
69
69
70
70
71
71
72
71
71
31
14
18
60
70
26
37
13
29
72
70
48
45
60
13
69
39
72
70
78
3
70
connectivity
tools
9
connectors, front panel copy files
D
data, entering from front panel 37 DC probes
use of 69
delete files Digital Signal Processing and
card slot location Dir Select key Dir Up key directional bridge directories, definition directory
creating display annotations display, information screen documentation
additional set 63
CD-ROM
CD-ROM only
service
standard set
E
electrostatic discharge (ESD)
protecting against EMC probes Enter key, using ESD safety accessories examples
frequency accuracy
signals, viewing external reference (10 MHz),
F
factory preset, description Fast ADC
card slot location file copying file deleting file formats file menu functions
copy
delete
rename
save file renaming file saving files
naming
58
59
Fast ADC
49
49
9
63
69
turning on
58
59
47, 51, 53
58
59
57
51, 53, 55
51, 53, 55
60
29
49
72
45
26
63
9
21
37
41
39
39
29
47
57
24
72
14
38
firmware upgrade information FM Demodulation
card slot location frequency accuracy, increasing frequency count, using Frequency Extension
card slot location 29 front panel
connectors and keys 24
display annotations
entering data
symbols
G
GPIB
card slot location GPIB cable
I
information screen 14 information, additional internal alignments IntuiLink software IP address
K
key overview keypad, using keys knob, using
L
LAN, setting IP address 14 limiters
RF and transient load, 50 ohm
M
manuals, ordering 63 manuals, standard set marker
frequency and amplitude,
moving
turning off marker annotation
change with frequency count
location menu keys missing options mixers
33
70
14
24
37
reading
to peak to reference level with knob or arrow key
40
40
31
29
42
26
37
29
3
18
9
31
37
71
69, 72
9
40
40
40
42
15
3
41
42
Index
81
Page 86
Index
External Mixing (Option AYZ)
67
harmonic preselected millimeter
moving
data from analyzer 9 data to PC
N
navigating
catalogs files tables
Navigation keys noise figure noise figure measurement
noise source numeric keypad, using
O
options not in instrument memory overview, keys and key menus
P
packaging packaging symbol parallel interface connector PC connectivity personality options not in
power amplifiers power splitters preamplifiers preset
factory types user, creating
printer setup probes
AC and DC
product information
R
rear panel
symbols
rear panel features reference level, setting reference, turning on 10 MHz rename files
Index
RF bridge RF limiters RPG, using RS-232
card slot location
70
9
48
48
32
48, 49
72
accessories 72
72
15
79
9
instrument
15
71
70
38
38
38
19
69
33
57
72
71 37
71
37
31
34
30
70
3
29
40
39
29
RS-232 interface connector
S
save files saving
bitmaps 47, 51, 53
trace files screen annotation screen, information service options short, 50 ohm signals, increasing accuracy signals, viewing softkeys, auto and man mode softkeys, basic types software
IntuiLink 9 splitters static safety accessories symbols in titles symbols, on front and rear panels
T
tab key table
navigation text, entering titles, editing transient limiter turning on the analyzer for the
U
unit softkeys, using 37 URL, PSA analyzers URL, sales and service user preset
creating 38
description
disabling
W
warm-up time warranty
51, 53, 55
47, 51, 53
26
14
78
72
39
71
60
33
32
32
60
60
71
first time 14
38
38
14
78
30
41
37
37
72
3
78
82
Loading...