Agilent E4440A User’s and Service Guide

Agilent Technologies
Z5602A Options H24/H35/H50/H51/H52
User’s and Service Guide
50 MHz Precision Power Reference
Use this manual with the following document:
E4440A Calibration and Adjustment Software Getting Started Guide
Part Number E4440-90555
Manufacturing Part Number: Z5602-90005
© Copyright Agilent Technologies, Inc. 2006
Warranty Statement
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
ii
Safety Notes
The following safety notes are used throughout this document. Familiarize yourself with each of these notes and its meaning before performing any of the procedures in this document.
WARNING
CAUTION
Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
Definitions
Specifications describe the performance of parameters covered by the product warranty (temperature –0 to 55 °C, unless otherwise noted.)
Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty.
Nominal values indicate expected performance or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
Characteristic Performance describes performance parameter that the product is expected to meet before it leaves the factory, but is not verified in the field and is not covered by the product warranty. A characteristic includes the same guard bands as a specification.
iii
iv
Contents
Z5602A Option H24/H35/H50/H51/H52
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Verifying the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Connecting the Precision Power Reference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Amplitude Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Typeface Key Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Adjustment Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety and Regulatory Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Connector Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Shipping Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Compliance with Canadian EMC Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Compliance with German Noise Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Contacting Agilent Sales and Service Offices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Contents-1
Contents
Contents-2
Z5602A Option H24/H35/H50/H51/H52
User’s and Service Guide 1
Z5602A Option H24/H35/H50/H51/H52 Description
Description
The Agilent Z5602A Precision Power Reference is a portable 50 MHz sine wave source that requires a very short warm-up time. Traditionally, the process to determine the amplitude accuracy of a source used to calibrate a device under test (DUT) involved power meter measurements and associated calculations to determine the uncertainty of the power meter, the interconnect cables and the VSWR effects of the power meter/source and source/DUT. The Agilent Z5602A Precision Power Reference can achieve the same accuracy as the calibration source built into a power meter, and mates directly with the Type-N female connectors on the spectrum analyzer inputs. The Precision Power Reference greatly reduces the measurement uncertainties mentioned above.
The Precision Power Reference is available in three different connector types: 2.5 mm,
3.5 mm and Type-N (m). Additionally, the Type-N version has 3 different power levels: 0 dBm, 25 dBm and 30 dBm. Each Precision Power Reference has a power cable that can be directly connected to the probe power outlet (12.6V, +15V) on a spectrum analyzer or other instruments with the probe power feature.
The small size and portability of the Precision Power Reference increases the potential for use in automated production test systems.
Verifying the Shipment
Verify that you have received the proper option items listed in Table 1. Inspect the shipping container. If the container or packing material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is physical damage refer to “Contacting Agilent Sales and Service Offices” on page 12. Keep the damaged shipping materials (if any) for inspection by the carrier and an Agilent Technologies representative.
Table 1 Option Numbers with Connector Types
Option Number Connector Output Power
H24 2.4 mm (f)
H35 3.5 mm (f)
H50 Type-N (m) 0 dBm
H51 Type-N (m)
H52 Type-N (m)
25 dBm
25 dBm
25 dBm
30 dBm
2 User’s and Service Guide
Z5602A Option H24/H35/H50/H51/H52 Connecting the Precision Power Reference
Connecting the Precision Power Reference
1. Turn On the spectrum analyzer.
2. Connect the Precision Power Reference to the RF Input connector.
3. Connect the probe power cable from the reference to the probe power output connector. Refer to Figure 1.
4. Perform a factory Preset on the spectrum analyzer so that it is in a known state.
Figure 1 Front Panel
User’s and Service Guide 3
Z5602A Option H24/H35/H50/H51/H52 Amplitude Accuracy
Amplitude Accuracy
The PSA Series spectrum analyzer requires the Z5602A Option H24/H35/H50/H51 or H52 when adjusting the amplitude of the internal 50 MHz reference.
The following example illustrates how the Option H51 version of the Precision Power Reference can be used to verify the amplitude accuracy of the Agilent E4440A Spectrum Analyzer at 50 MHz.
Verify that the spectrum analyzer has been turned On for a minimum of 30 minutes.
Typeface Key Conventions
The following key conventions are used throughout this document.
[HARDKEYS] are labeled front panel keys
SOFTKEYS are unlabeled key whose function is indicated on the instrument display
Adjustment Procedure
Perform the following front panel keystrokes on the spectrum analyzer.
1. Press [SYSTEM] > Alignments > Align All Now to trigger an alignment on the spectrum analyzer.
2. Connect the Precision Power Reference to the spectrum analyzer as shown in Figure 1
on page 3.
3. To view the Precision Power Reference signal, set the spectrum analyzer controls as follows:
a. Press [Frequency] > [50 MHz > [Span] > [1 MHz] > [Single] > [Peak Search]
b. Compare the signal amplitude displayed on the spectrum analyzer to the power level
shown on the calibration sticker attached to the Precision Power Reference. For example; the spectrum analyzer marker value may be 25.07 dBm and the calibration sticker on your Precision Power Reference may indicate 25.009 dBm. Therefore, the spectrum analyzer amplitude error is 0.061 dB.
4 User’s and Service Guide
Z5602A Option H24/H35/H50/H51/H52 Amplitude Accuracy
Figure 2 Agilent Z5602A Option H51 Signal
User’s and Service Guide 5
Z5602A Option H24/H35/H50/H51/H52 Specifications
Specifications
The specifications for the Agilent Z5602A Option H24/H35/H50/H51/H52 Precision Power References are listed below.
No warm up time is required for the Precision Power Reference once the power is turned on.
Table 2 50 MHz Precision Power Reference Specifications
Par am eter
Option H24H35H50H51H52
Connector Type 2.4 mm 3.5 mm Type-N (m) Type-N (m) Type-N (m)
Power Level 25 dbm 25 dbm 0 dbm 25 dbm 30 dbm
Frequency Drift |∆f|≤ 2.5 KHz |∆f|≤ 2.5 KHz |∆f|≤ 2.5 KHz |∆f|≤ 2.5 KHz |∆f|≤ 2.5 KHz
Typical VSWR 1.06 1.06 1.06 1.06 1.06
Output Power Variation Over 20 °C – 30 °C
Output Power Variation Over 0 °C – 55 °C
Total Harmonic Content < 45 dBc < 45 dBc < 45 dBc < 45 dBc < 45 dBc
Amplitude Error Due to Harmonic Content
± 0.004 dBm ± 0.004 dBm ± 0.004 dBm ± 0.004 dBm ± 0.004 dBm
± 0.02 dBm ± 0.02 dBm ± 0.02 dBm ± 0.02 dBm ± 0.02 dBm
< 0.0001 dB < 0.0001 dB < 0.0001 dB < 0.0001 dB < 0.0001 dB
Specifications
1
1. The recommended cal cycle on the Agilent Z5602A Precision Power Reference is 1 year.
6 User’s and Service Guide
Z5602A Option H24/H35/H50/H51/H52 Safety and Regulatory Information
Safety and Regulatory Information
Introduction
Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
Refer to the Getting Started Guide (E4440-90555) for the E4440A Performance Spectrum Analyzer.
Before Applying Power
Verify that the product is configured to match the available main power source. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
Connector Care and Cleaning
If alcohol is used to clean the connectors, the power cord to the instrument must be removed. All cleaning should take place in a well ventilated area. Allow adequate time for the fumes to disperse and moist alcohol to evaporate prior to energizing the instrument.
WARNING
To prevent electrical shock, disconnect the Agilent Technologies model product from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Declaration of Conformity
For a copy of the manufacturer’s Declaration of Conformity for this apparatus, contact your local Agilent Technologies office or sales representative. Refer to “Contacting Agilent Sales
and Service Offices” on page 12.
Statement of Compliance
This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe
condition.
User’s and Service Guide 7
Z5602A Option H24/H35/H50/H51/H52 Safety and Regulatory Information
Shipping Instructions
You must always call the Agilent Technologies Instrument Support Center to initiate service before retuning your instrument to a service office. See “Contacting Agilent Sales
and Service Offices” on page 12. Always transport or ship the instrument using the
original packaging if possible. If not, comparable packaging must be used. Attach a complete description of the failure symptoms.
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada.
Compliance with German Noise Requirements
This is to declare that this instrument is in conformance with the German Regulation on Noise Declaration for Machines (Laermangabe nach der Maschinenlaermrerordnung-3. GSGV Deutschland).
Acoustic Noise Emission/Geraeuschemission
LpA<70 dB Lpa<70 dB
Operator Position am Arbeitsplatz
Normal Operation normaler Betrieb
per ISO 7779 nach DIN 45635 t. 19
8 User’s and Service Guide
Z5602A Option H24/H35/H50/H51/H52 Safety and Regulatory Information
Warning s
WARNING
The WARNING notice denotes a hazard. It calls attention to a procedure which if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Warnings applicable to this instrument are:
WARNING
To prevent electrical shock, disconnect the Agilent Technologies
Z5602A Options H24/H35/H50/H51/H52 from the mains before
cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
WARNING
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
WARNING
For continued protection against fire hazard replace line fuse only with same type and rating:
• United States—F 0.5A/250V, Part Number 2110-0202
• Europe—F 3.15A/250V, Part Number 2110-0655
The use of other fuses or material is prohibited.
WARNING
WARNING
WARNING
WARNING
WARNING
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only into a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the product is likely to make the product dangerous. Intentional interruption is prohibited.
These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010-1: 2001.
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock do not remove covers.
User’s and Service Guide 9
Z5602A Option H24/H35/H50/H51/H52 Safety and Regulatory Information
WARNING
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
Cautions
CAUTION
Cautions applicable to this instrument are:
CAUTION
CAUTION
CAUTION
The CAUTION notice denotes a hazard. It calls attention to an procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met
Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding (by not using this cord) can cause instrument damage.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 61010 Second Edition and 664 respectively.
This instrument has autoranging line voltage input; be sure the supply voltage is within the specified range.
CAUTION
Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
10 User’s and Service Guide
Z5602A Option H24/H35/H50/H51/H52 Safety and Regulatory Information
Instrument Markings
When you see this symbol on your instrument, you should refer to
!
the instrument’s instruction manual for important information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
This symbol indicates the product meets the Australian Standards.
This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13,
2005. All electric and electronic equipment are required to be separated from normal waste for disposal (Reference WEEE Directive, 2002/96/EC).
This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is in the STANDBY position.
Safety Earth Ground. This is a Safety Class I
product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
User’s and Service Guide 11
Z5602A Option H24/H35/H50/H51/H52 Contacting Agilent Sales and Service Offices
Contacting Agilent Sales and Service Offices
Assistance with test and measurement needs, and information on finding a local Agilent office are available on the Internet at:
http://www.agilent.com/find/assist
You can also purchase accessories or documentation items on the Internet at:
http://www.agilent.com/find
If you do not have access to the Internet, contact your field engineer.
NOTE
In any correspondence or telephone conversation, refer to the product by its model number and full serial number. With this information, the Agilent representative can determine whether your unit is still within its warranty period.
12 User’s and Service Guide